MAPPING CURRENT DENSITY DUE TO ELECTRICAL STIMULATION USING MAGNETIC RESONANCE ELECTRICAL IMPEDANCE TOMOGRAPHY By ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY UNIVERSITY OF FLORIDA 2016

Upload: others

Post on 24-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

MAPPING CURRENT DENSITY DUE TO ELECTRICAL STIMULATION USING

MAGNETIC RESONANCE ELECTRICAL IMPEDANCE TOMOGRAPHY

By

ADITYA KUMAR KASINADHUNI

A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL

OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

OF THE REQUIREMENTS FOR THE DEGREE OF

DOCTOR OF PHILOSOPHY

UNIVERSITY OF FLORIDA

2016

Page 2: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

© 2016 Aditya Kumar Kasinadhuni

Page 3: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

To my parents, K A Shiva Prasad, K Mangala Gowri, and brother, Girija Kumar Raghava

Page 4: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

4

ACKNOWLEDGMENTS

I would like to extend my heartfelt thanks to my advisor Dr. Thomas Mareci for his

guidance and mentorship throughout my journey in graduate school. Dr. Mareci has been

instrumental in molding my thought process to pay attention to detail when planning

experiments. His meticulous nature taught me to analyze every problem closely and cultivated in

me the attitude to “question everything, assume nothing” which is critical to the development of

science. His astounding depth and breadth on all things MRI has and continues to inspire me to

constantly push the boundaries of my own knowledge in the field of magnetic resonance. I would

like to thank Dr. Rosalind Jane Sadleir for her constant support and for providing me a platform

to demonstrate my abilities in MRI. Dr. Sadleir has been a wonderful mentor who constantly

kept me on my toes and helped me develop an understanding of magnetic resonance electrical

impedance tomography (MREIT). I would like to thank Dr. Paul R. Carney who was highly

motivating and always willing to discuss current trends and literature pertaining to the field of

neuromodulation and neuroplasticity. His constant zeal and enthusiasm for the field kept me

inquisitive about the mysteries of the human brain and pushed me to understand the importance

of the work I embarked upon in my doctoral study. I would like to express my gratitude to Dr.

Stephen J Blackband, who was welcoming and allowed me unrestricted access to his lab so that I

could use his weighing balance, distilled water and hot plate when I needed them the most. Very

early in my Ph.D., Dr. Malisa Sarntinoranont instilled in me the importance of selecting a

practical and feasible project. Her advice to continue working towards my goals and make

whatever little progress in the wake of obstacles helped me stay ahead in my doctoral pursuits.

Dr. Aysegul Gunduz was highly accommodating and ever ready to listen to my efforts at

mapping electrical current density and was willing to help me out with any administration

problems.

Page 5: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

5

I would also like to thank Dr. Michael Schär for his help with the Philips pulse sequence

and for his prompt replies to my queries requesting clarity on the subtleties of the 3 T Philips

Achieva system at the McKnight Brain Institute. William Triplett was extremely supportive and

helpful with his ideas on image reconstructions, and provided the skeleton code to reconstruct

images from raw Philips data, which I built upon, for the MREIT data processing. Dr. Munish

Chauhan provided excellent guidance towards developing the imaging protocol for the human

imaging work and also helped with the reconstruction of current density from the measured

magnetic field maps.

I would like to extend my heartfelt thanks to the members of the Radio Frequency Lab at

the AMRIS facility in the McKnight Brain Institute who helped me with all problems

electronics. Malathy Elumalai was highly forthcoming and always ready to help me with

soldering small circuits and stimulated my thought process on numerous occasions when I

needed to troubleshoot my experiments. Kelly Jenkins was extremely encouraging and provided

valuable guidance during my efforts at repairing RF coils. Joshua Slade provided unquestioned

mechanical engineering support and guidance to help me construct numerous contraptions and

phantoms for research. Dr. Huadong Zeng and Tammy Nicholson were extremely

accommodating to work with my imaging schedules and always ready to help troubleshoot

problems pertaining to my experiments.

I would also like to thank my former and current labmates who played a vital role in

shaping my career in graduate school. Dr. Garrett Astary (General Electric, South Carolina) was

a great mentor and taught me how to use the Agilent system for imaging experiments. Dr. Luis

Colon Perez (Febo Lab, University of Florida) provided help with diffusion imaging and was a

source of continuous motivation. Guita Banan was a wonderful physics mentor on whose support

Page 6: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

6

I relied heavily to understand the complicated electromagnetic interactions in MREIT. Magdoom

Kulam has been a brother away from home providing constant brainstorming sessions and

helping me with data acquisition at the 3 T magnet for the human imaging work. Manish Amin is

a wonderful person in general and was willing to share his personal equipment for my use in the

lab for MREIT experiments. I would also like to thank Aprinda Indahlastari, Chris Anderson,

Kevin Castellano, Andrew Girard and Kevin Montes for their help and assistance with modeling,

administration and imaging experiments.

Finally I would like to thank my friends and family for believing in me and providing

great support in times of need. Bhargava Kandala, Anuj Goyal and Vijay Pappu were

instrumental in motivating me with excellent discussions on a variety of topics. Sunantha

Sethuraman provided constant support and motivation in every way possible and took great care

of me during my recovery from surgery. My brother, Girija Kumar, showed extreme love and

patience constantly supporting and shielding me so that I could fulfil my responsibilities. My

parents, Major Shiva Prasad and Mangala Gowri sacrificed a lot for me and showed

unconditional love and support all my life. They played a vital role in making me the person I am

today. I am extremely thankful and will remain forever grateful to them.

Page 7: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

7

TABLE OF CONTENTS

page

ACKNOWLEDGMENTS ...............................................................................................................4

LIST OF TABLES .........................................................................................................................10

LIST OF FIGURES .......................................................................................................................11

LIST OF ABBREVIATIONS ........................................................................................................14

ABSTRACT ...................................................................................................................................16

CHAPTER

1 MOTIVATION AND SPECIFIC AIMS ................................................................................18

1.1 Motivation .........................................................................................................................18 1.2 Specific Aims ....................................................................................................................20

1.2.1 Specific Aim1: Develop MR methodology for facilitating measurement of

low amplitude current induced magnetic fields in the presence of temporal drift

at 3 T MRI ....................................................................................................................20

1.2.2 Specific Aim 2: Measure at 3 T, current induced magnetic field maps and

current density distributions due to tACS like stimulation in healthy volunteers .......21

1.2.3 Specific Aim 3: In a DBS like experimental setup, determine feasible

electrode orientation for reliable measurement of magnetic fields due to low

amplitude current injection ..........................................................................................21

2 BACKGROUND AND PREVIOUS STUDIES ....................................................................22

2.1 Neurophysiology ...............................................................................................................22 2.2 Transcranial Electrical Stimulation ..................................................................................27

2.3 Magnetic Resonance Imaging ...........................................................................................31 2.4 Magnetic Resonance Electrical Impedance Tomography ................................................37

2.4.1 Signal and Noise in MREIT ............................................................................41 2.4.2 Magnitude noise ..............................................................................................43 2.4.3 Phase noise ......................................................................................................44

2.4.4 Optimization of imaging parameters to improve sensitivity ...........................46 2.4.5 Projected Current Density ...............................................................................48

3 DETERMINATION OF THE EFFECT OF TEMPORAL VARIATIONS IN THE

STATIC FIELD ON MREIT ACQUISITIONS .....................................................................53

3.1 Introduction .......................................................................................................................53 3.2 Theory ...............................................................................................................................54 3.3 Methods ............................................................................................................................57

3.3.1 Pilot Experiments ...................................................................................................57

Page 8: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

8

3.3.1.1 Phantom preparation. ...................................................................................57

3.3.1.2 Image acquisition. ........................................................................................57 3.3.1.3 Data inconsistencies .....................................................................................58 3.3.1.4 Troubleshooting data inconsistencies ...........................................................58

3.3.2 Field Corrected Experiments ..................................................................................59 3.3.3 Electrical Stimulation .............................................................................................59

3.3.3.1 Switching circuit. .........................................................................................60 3.3.3.2 Circuit description. .......................................................................................61

3.3.4. Data Processing .....................................................................................................62

3.4 Results...............................................................................................................................63 3.5 Discussion and Conclusion ...............................................................................................64

4 MEASUREMENT OF ELECTRICAL CURRENT DENSITY DUE TO

TRANSCRANIAL ALTERNATING CURRENT STIMULATION OF THE HUMAN

BRAIN AT 3 T .......................................................................................................................75

4.1 Introduction .......................................................................................................................75

4.2 Methods ............................................................................................................................78 4.2.1 Phantom Experiments .............................................................................................78

4.2.2 Human Experiments ...............................................................................................79 4.2.3 MR Imaging ............................................................................................................79 4.2.4 MREIT Data Processing .........................................................................................80

4.2.5 Image Segmentation ...............................................................................................81 4.2.6 Computational Modeling ........................................................................................82

4.3 Results and Discussion .....................................................................................................82 4.3.1 Gel Phantom ...........................................................................................................82

4.3.1.1 Field measurements ......................................................................................82 4.3.1.2 Optimization of magnetic field maps ...........................................................83

4.3.2 Healthy Volunteers .................................................................................................84 4.3.3 Quality of field maps ..............................................................................................86 4.3.4 Projected Current Density maps .............................................................................87

4.4 Conclusions.......................................................................................................................89

5 MEASUREMENT OF MAGNETIC FIELDS DUE TO LOW AMPLITUDE

INJECTION CURRENTS USING IMPLANTED ELECTRODES ....................................104

5.1 Introduction .....................................................................................................................104

5.2 Theory .............................................................................................................................104 5.2.1 Effect of background magnetic field on frequency encoding ..............................106

5.2.2 MREIT-ICNE .......................................................................................................108 5.3 Methods ..........................................................................................................................109

5.3.1 Field Stability measurements ...............................................................................109 5.3.2 Pilot Studies ..........................................................................................................110

5.3.2.1 Fabrication of carbon electrodes ................................................................110

5.3.2.2 Experimental setup .....................................................................................110 5.3.2.3 MR Imaging ...............................................................................................111

5.3.3 Extended Studies ..................................................................................................111

Page 9: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

9

5.3.3.1 Phantom preparation ..................................................................................111

5.3.2.2 Electrical Stimulation .................................................................................112 5.3.3.3 MR Imaging ...............................................................................................112

5.4 Results and Discussion ...................................................................................................113

5.5 Conclusions.....................................................................................................................115

6 CONCLUSIONS AND FUTURE WORK ...........................................................................124

6.1 Summary .........................................................................................................................124 6.2 Future Work ....................................................................................................................125

LIST OF REFERENCES .............................................................................................................128

BIOGRAPHICAL SKETCH .......................................................................................................135

Page 10: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

10

LIST OF TABLES

Table page

4-1 SNR values for volunteer MRI data and noise in measured magnetic field maps at

3T. ....................................................................................................................................102

5-1 Noise standard deviations for current induced field maps acquired with conventional

spin echo MREIT and spin echo MREIT with ICNE sequences. ....................................122

Page 11: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

11

LIST OF FIGURES

Figure page

2-1 Graphic representation of low frequency and high frequency current flow in tissue. .......50

2-2 Schematic for experimental setup in an MREIT experiment while injecting current

into a cylindrical agarose gel. ............................................................................................50

2-3 Representation of magnetization vector and the effect of noise on the phase (ϕ) of the

magnetization vector. A represents the magnetization in the absence of noise, and M,

in the presence of noise. For simplicity the deviation in ϕ due to noise is shown to. ........51

2-4 Distribution of phase noise for varying levels of magnitude signal-to-noise ratio. The

distribution begins to approximate to a Gaussian distribution for SNR values greater

than 3. .................................................................................................................................51

2-5 Spin echo pulse sequence for MREIT acquisition. The current polarity is reversed

after the 180° pulse to account for the phase reversal of spins. .........................................52

2-6 Spin echo sequence with current injection during the sampling interval altering the

linear relationship between the spin position and frequency. ............................................52

3-1 Gel phantom used in MR experiments. A) Freshly prepared agarose gel phantom

after removal from refrigerator. B) Experimental setup showing the electrodes

secured on the gel with a rubber strap. C) Schematic showing the placement of the. .......66

3-2 Spoiled multi gradient echo sequence (2D) used for data acquisition. The repetition

time (TR) was 50 ms. Current pulses Ic = ±1.5 mA in amplitude were injected for a

duration of 32 ms. The bandwidth of acquisition was 550 Hz/pixel while 100. ...............67

3-3 Layout of the experimental setup used for imaging. The current injected was

carefully monitored using the oscilloscope to ensure the synchronization of RF

trigger and injection pulses as well as the injection amplitude. The components of . .......68

3-4 Schematic of the switchbox circuit and layout of the different components used in

coalition for data acquisition. R-test is the test resistance of 1 KΩ across which the

stimulation is checked using an oscilloscope.....................................................................69

3-5 Preliminary experimental results from two identical acquisitions performed with 1

mA current injection through a wire placed in a gel phantom.. .........................................70

3-6 Oscilloscope recording during pilot experiments for 1 mA current injection. Each

division on the X-axis corresponds to 50 ms. The TTL signal (green trace) generated

after every RF excitation (see Figure 3-2) during imaging and corresponding .................70

3-7 Phase variation in a 30x30 pixel ROI across each of the complex divided datasets. ........71

Page 12: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

12

3-8 Regression fit for the phase values within a 30 x 30 pixel ROI drawn in the center of

the image for the first echo time across the ten volumes collected as part of the first

acquisition in the reversed experiments. Acquisition time for 1 volume was ...................72

3-9 Variation of the mean phase in a 20 x 20 pixel ROI drawn in the center of the image.

Each of the twenty-three plots represents the phase obtained by complex dividing

data of every repetition by its preceding one. The X-axis in each plot represents the ......73

3-10 Oscilloscope recording of the voltage across the test resistor (see Figure 3-4) due to

1.5 mA current injection (blue) alternating polarity with every incoming TTL signal

(green). ...............................................................................................................................73

3-11 Magnetic field maps for the first slice location obtained from field corrected

experiments by acquiring the positive and negative current injection data very close

in time. Identical maps across multiple current injection runs show measurement...........74

4-1 MR magnitude images from all 32 channels reconstructed by carrying out the inverse

Fourier transform of complex k-space data. The coil combined magnitude image can

be generated using a sum of squares approach with or without appropriate weighting. ...91

4-2 Representative images showing electrode placement, image reconstruction and

decrease in SNR with increasing echo times.. ...................................................................92

4-3 MR magnitude and phase images for the center slice location collected at the first

echo time. The increased sensitivity at the bottom of the magnitude image was due to

the proximity of the gel phantom to the bottom set of RF coils. .......................................92

4-4 Phase maps produced from complex division of acquired data scaled from -0.1

radians to 0.1 radians. ........................................................................................................93

4-5 Magnetic field maps computed from the measured phase changes (Figure 4-4) scaled

from -10 nT to 10 nT. ........................................................................................................93

4-6 MR magnitude data for all slices and their corresponding T2* maps computed from

data acquired without any current injection. The T2* maps are displayed in seconds

giving an average T2* of about 48 ms in the gel phantom. Volume fitting for all the .......94

4-7 Comparison of measured and simulated current-induced magnetic field maps in the

gel phantom. .......................................................................................................................95

4-8 Experimental setup and stimulation paradigm for transcranial electrical stimulation

on healthy volunteers.. .......................................................................................................96

4-9 Current-induced phase and magnetic field images for healthy volunteers. Note the

steady increase in phase across echoes. Electrical stimulation was performed from

right to left in the image. The static field direction is into the page. .................................97

Page 13: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

13

4-10 Current-induced phase and magnetic field maps for the Fpz-Oz direction. The phase

maps are scaled from -0.1 to 0.1 radians while the magnetic field maps cover a range

of -10nT to 10 nT. ..............................................................................................................98

4-11 Representative images of segmented T1 images, computational model generated and

comparison of simulated and measured magnetic fields. ..................................................99

4-12 Comparison of measured and simulated magnetic fields due to T3-T4 stimulation for

the central slice location. Good agreement can be observed in subjects 009 and 012.

Results from subjects 010 and 011could have contributions from motion. Magnetic.....100

4-13 Comparison of measured and simulated magnetic fields, following electrical

stimulation with electrodes placed at Fpz and Oz, for the central slice location .............101

4-14 Comparison of simulated current density images and current density maps computed

from measured magnetic field maps. Shown are the maps for X and Y components of

the current density vector and the corresponding norm calculated from the two ............102

4-15 Current density (norm) images for healthy volunteers shown in relation to their high

resolution T1 images. The current density maps were scaled from 0 – 0.7 A/m2. Also

shown are the current density maps overlaid onto the T1 images with 50%. ..................103

5-1 Asymmetric spin echo pulse sequence where the slice select 180° RF pulse is moved

from its original position to create a net phase accumulation due to free precession. .....117

5-2 Field stability measurements using the asymmetric spin echo sequence with varying

free precession delay times. .............................................................................................118

5-3 Demonstration of the dominant magnetic fields due to current in the electrodes.The

dissimilarity in the horizontal line plot across electrodes, was due to the electrodes

not being perfectly parallel to eachother. The image is scaled from -10 nT to 10 nT.. ...119

5-4 Current-induced magnetic field measurements for varying current amplitudes using

conventional spin echo measurements. ............................................................................120

5-5 Current-induced magnetic field measurements (in Tesla) using the Injection Current

Non-Linear Encoding (ICNE) approach with varying levels of current amplitude. ........121

5-6 Magnetic field maps (in Tesla) due to 0.25 mA current injection in three phantoms

with different orientations of copper electrodes. .............................................................123

Page 14: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

14

LIST OF ABBREVIATIONS

AMRIS advanced magnetic resonance imaging and spectroscopy

BLM bilayer lipid membrane

BW bandwidth

CNS central nervous system

CSF cerebrospinal fluid

DBS deep brain stimulation

DMSO dimethyl sulfoxide

EPI echo planar imaging

EIT electrical impedance tomography

fMRI functional magnetic resonance imaging

GE gradient echo

Gfe frequency encoding gradient

GM gray matter

Gpe phase encoding gradient

GRAPPA generalized auto-calibrating partially parallel acquisition

HARDI high angular resolution diffusion imaging

ICNE injection current non-linear encoding

MREIT magnetic resonance electrical impedance tomography

MRI magnetic resonance imaging

NMR nuclear magnetic resonance

nT nano tesla

PCD projected current density

PLA poly lactic acid

PVDF polyvinylidene fluoride

Page 15: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

15

RF radio frequency

SE spin echo

SENSE sensitivity encoding

SNR signal-to-noise ratio

SPGR spoiled gradient echo

T1 longitudinal relaxation time

T2 transverse relaxation time

tACS transcranial alternating current stimulation

Tc current injection time

tDCS transcranial direct current stimulation

TE echo time

tES transcranial electrical stimulation

TR repetition time

Ts sampling time

TTL transistor transistor logic

VNS vagal nerve stimulation

WM white matter

Page 16: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

16

Abstract of Dissertation Presented to the Graduate School

of the University of Florida in Partial Fulfillment of the

Requirements for the Degree of Doctor of Philosophy

MAPPING CURRENT DENSITY DUE TO ELECTRICAL STIMULATION USING

MAGNETIC RESONANCE ELECTRICAL IMPEDANCE TOMOGRAPHY

By

Aditya Kumar Kasinadhuni

December 2016

Chair: Thomas Mareci

Major: Biomedical Engineering

Electrical stimulation therapies like Deep Brain Stimulation (DBS) and Vagal Nerve

Stimulation (VNS) have been widely used for the treatment of Parkinson’s Disease and Epilepsy.

Transcranial direct current stimulation (tDCS) has been shown to have the potential for being an

augmentative therapy and improve cognitive functions. However, the underlying mechanisms of

operation of these electrical stimulation therapies are still under debate and a sound working

hypothesis is still in want. Understanding the flow and distribution of injected currents can

provide information to improve the efficiency of these therapies. So far, computational models

using information from high resolution MR images have been used extensively for prediction

and estimation of current densities. Magnetic Resonance Electrical Impedance Tomography

(MREIT) can be used to visualize current distributions by tracking changes in magnetic fields

produced by the injected currents. The produced current density maps can help guide stimulation

therapies and validate existing modeling techniques. However, past implementations of MREIT

used high current amplitudes to reliably measure the induced magnetic fields and therefore were

not viable in in vivo studies due to safety considerations.

The magnetic fields generated by currents used in electrical stimulation therapies are

usually on the order of micro to nanoTesla which is a major deterrent to detecting the current

Page 17: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

17

induced magnetic fields. Reliable quantification of these fields is challenging requiring high

signal-to-noise ratio in the measurement that generally requires an increase in the imaging time.

Additionally, the miniscule nature of the fields makes the measurement susceptible to minor

variations in an otherwise static main magnetic field of the MR scanner. Appropriate imaging

methods were developed and necessary electronic circuitry was constructed to facilitate data

acquisition in a way that decreased the measurement sensitivity to temporal variations in the

static field.

Magnetic fields due to the current were measured in homogenous phantoms and healthy

individuals and compared against those generated from sample specific computational models.

The results show good correlation with discrepancies attributable to extraneous factors like

subject motion. Current density estimates from the measured magnetic fields and numerical

simulations were also compared. Finally, to understand the distribution of current density during

DBS-like stimulation, current induced magnetic fields while using implanted electrodes were

measured using different electrode orientations.

.

Page 18: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

18

CHAPTER 1

MOTIVATION AND SPECIFIC AIMS

1.1 Motivation

The brain is one of the most important and complex organs in the human body. Attempts

to understand brain anatomy and function date back to the prehistoric times as old as 4000 B.C.

when the Sumerians first documented the effects of poppy plant on the “mind”. In 300 B.C the

first dissection of a human body was performed by Alexandrian biologists who classified the

motor and sensory nerves. The basic anatomical classification of what we know to be the

“modern brain” was first established by Thomas Willis in 1664 A.D (in Cerebri Anatome),

known also for the discovery of the “Circle of Willis” which is a circular group of arteries on the

base of the brain.. Advances in experimentation and technology paved way for our understanding

of the brain anatomy and we now know it to comprise of billions of fundamental functional units

termed “neurons” that work together in coordination enabling us to operate in daily life. All our

experiences, emotions and thoughts are governed by electrochemical changes occurring in the

environment of these neurons. However, even to this day, in spite of the cornucopia of

knowledge and research on the brain, there is a lot we do not know about the brain’s function, its

connections and response to pathology.

Nerve and muscle tissue are collectively termed as “excitable” tissue and coordinate

activity in the human body through neuromuscular junctions (NMJ). The underlying

neurophysiology dictates that electrochemical impulses are responsible to bring about this

coordinated function which prompted early medical practitioners to use electrical and chemical

treatment strategies to restore function in cases of pathology. The idea was to use these treatment

strategies to alter the environment around the excitable tissue to promote or inhibit function. The

first form of electrical stimulation treatment was developed in the 1930s, called

Page 19: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

19

electroconvulsive therapy (ECT), for treatment of major depression and catatonia. The therapy

informally known as “shock treatment” was induced as a last option to control psychiatric

illnesses and is currently one among many techniques that are collectively termed

“neuromodulation” techniques. Today, neuromodulation is an extremely important field of

research that encompasses a gamut of interventional techniques aimed at long term activation,

inhibition, modification and/or regulation of neural activity, using electromagnetic, chemical and

optogenetic methodologies. Deep Brain Stimulation (DBS) which is a form of electrical

stimulation is an FDA approved technique for the treatment of medically intractable Parkinson’s

Disease. Over the past decade, Transcranial Direct Current Stimulation (tDCS) has been in the

spotlight promising the potential to improve working memory, stroke and a host of cognitive

functions (1,2). Transcranial Alternating Current Stimulation, one of the variants of tDCS, has

been shown to modulate the existing background neural activity promoting excitation or

inhibition of specific functions (3).

Despite the beneficial effects of these techniques, very little is understood with regards to

their mechanism of action. Treatments involving these techniques rely heavily on computational

models for guidance and selection of stimulation parameters. However generalized

computational models suffer as they are blind to variations in the head size and composition that

differ vastly across people. Subject specific models provide a more robust approach at accurate

current density predictions in different brain regions by utilizing information on the underlying

brain structure obtained through MRI or CT scans. However there is a dire need for the

validation of these computational models and MREIT or magnetic resonance electrical

impedance tomography is aptly placed to address this need.

Page 20: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

20

1.2 Specific Aims

The goal of this research is to develop an MR methodology to effectively map current

density for improved understanding of the mechanisms governing neuromodulation techniques

like tDCS and DBS. The developed technique can be used to validate existing models that aim to

describe current flow in the brain during electrical stimulation. This was achieved by tackling the

temporal drift in the static field of the MRI scanner by acquiring current injection datasets as

close in time as possible. Validation of magnetic field and current density maps from sample

specific computational models was performed. Finally, for characterization of current density in

DBS like setups, current injection using magnet wire to identify the most feasible electrode

orientation such that the magnetic field from current density in the electrodes does not mask that

in the sample will be presented.

1.2.1 Specific Aim1: Develop MR methodology for facilitating measurement of low

amplitude current induced magnetic fields in the presence of temporal drift at 3 Tesla

using MRI

Temporal drift in the static field of MRI scanners can adversely affect the measurement

of low amplitude current-induced magnetic fields by completely masking them. Data acquired

using a conventional MREIT approach and another approach to decrease the sensitivity to

background field drifts are presented. The field drift is characterized in the absence of current

injection and compared against manufacturer specifications. Necessary electronic circuitry to

synchronize electrical stimulation with image acquisition and enable data acquisition using the

second approach was designed and constructed. The reliability of measurements using the

second approach is presented.

Page 21: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

21

1.2.2 Specific Aim 2: Measure the current induced magnetic field maps and current density

distributions at 3 Tesla due to tACS like stimulation in healthy volunteers

Calculating current density distributions within the human head due to transcranial

electrical stimulation techniques requires reliable measurement of current-induced magnetic

fields. Using the method developed in specific aim 1, reliable determination of the minute fields

was performed in gel phantoms and extended to healthy volunteers who experienced electrical

stimulation in two orthogonal directions. Computational head models for the different volunteers

were generated from MR data and current-induced magnetic fields were calculated against which

the experimental data compared. Signal-to-noise ratio (SNR) was measured in the magnetic field

maps during the gel experiments to check the fidelity of MREIT measurements. Finally, current

density distributions generated from the measured magnetic fields were also compared against

their numerical simulations. The methods developed in this chapter can be used to validate

computational modeling techniques that are currently the guiding tools for electrical stimulation

therapies. Reliable characterization of induced current densities due to transcranial electrical

stimulation (tES) could bolster the efficiency of these therapies while simultaneously providing

insights into their mechanisms of operation.

1.2.3 Specific Aim 3: Using implanted electrodes, determine feasible electrode orientation

for reliable measurement of magnetic fields due to low amplitude current injection

Current injection using electrodes placed perpendicular to the static magnetic field of the

MR scanner can be detrimental to the observation of magnetic fields produced in tissue. Pilot

experiments elucidating this shortcoming are presented and current injection using different

electrode orientations was performed. Though challenging from a surgical standpoint, the

methods provide a compromise against tackling the problem of magnetic fields produced in

tissue being masked due to the strong magnetic fields from current in the electrodes.

Page 22: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

22

CHAPTER 2

BACKGROUND AND PREVIOUS STUDIES

2.1 Neurophysiology

The human brain that forms an integral part of the central nervous system (CNS) can be

divided into gray matter (GM), white matter (WM) and cerebrospinal fluid (CSF). The GM gets

its name from the grayish hue of the neuronal cell bodies while their axons, which are much

paler, constitute the WM. The brain is surrounded by CSF which provides it mechanical

protection and plays a vital role in cerebral blood flow and autoregulation. Knowledge of the

neuroanatomy and physiology is vital to our understanding of brain function and can provide

insights into the mechanisms of exogenous/endogenous current flow. Before understanding the

effect of induced current or electric field on the brain tissue, it is necessary to understand the

constituents of brain tissue and the environment surrounding neurons that make up the brain.

Neurons and glial cells make up the primary composition of the nervous system in the

human body. Neurons interact with each other through synaptic junctions while physical and

metabolic support is provided by the glial cells. Neurons can be classified structurally as

unipolar, bipolar, multipolar neurons. Functionally, they can be sensory (afferent) or motor

(efferent) neurons depending on whether they transmit signals from or to the target cells.

Interneurons are specialized neurons that connect other neurons to specific parts of the CNS. The

structure of a typical neuron consists of the cell body or “soma”, “dendrites” or fan like

projections that form synaptic connections with other neurons, and a long tail or “axon” which is

responsible for the transmission of signals. A neuron communicates with another neuron at the

synapse by releasing neurotransmitters (eg. gamma amino butyric acid or GABA and glutamate)

that adhere to special chemical receptor proteins present in the post-synaptic neuron. Neuronal

axons can sometimes be “myelinated” due to the glial cells that wrap around the neuron

Page 23: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

23

providing an insulating layer and improving its conduction velocity. Myelination in the

peripheral nervous system (PNS) is due to the outgrowth of Schwann cells, while that of

oligodendrocytes forms the myelin in CNS. Nerve and muscle cells are termed “excitable” due to

the existence of a constant voltage gradient across their cell membranes at rest. The interior of

these cells is maintained at a constant negative potential with respect to its exterior due to the

balance of net electrical and diffusional forces. This is termed as the resting membrane potential

(RMP) and is about -70 mV in a typical human cell (4). To understand the effect of induced

electrical fields on brain tissue, it is necessary to gain insight into the different processes

involved in maintenance of the voltage gradient.

The intracellular and extracellular spaces around a cell are separated from one another

through a very thin membrane (7-15 nm) consisting of a lipoprotein complex. This is called the

bilayer lipid membrane (BLM) and is impermeable to most ions as the fatty acid tails of lipids

exert hindrance to the polar molecules. A special layer of carbohydrates that covalently link with

the lipids and proteins of the BLM forms the “glycocalyx” and provides functions of structural

integrity, cellular identification and interaction. Many cells are connected to each other through

their membranes forming specialized junctions known as desmosomes, tight junctions and gap

junctions. Integral proteins embedded in the membrane extend into the cytoplasm and function as

transport channels for water and various ions. The phospholipids in the bilayer and the channel

proteins are not linked or bonded to each other and hence are free to move. This results in

random lateral movement of both the proteins and lipids parallel to surface of the bilayer. The

flow of ions across the cell membrane occurs due to passive or active transport. Passive transport

is due to the difference in concentration of ions across the cell membrane through integral

protein channels and the channels’ permeability to these ions. The concentration of sodium ions

Page 24: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

24

is very high in the extracellular space while that of potassium is low. The reverse is true for the

interior of the cell. In addition, the permeability (number of open protein channels that facilitate

transport) for potassium ions is extremely high compared to sodium which creates a negative

potential inside the cell. Active transport of ions across the cell membrane is facilitated by the

presence of specific proteins in the lipid bilayer which utilize energy in the form of adenosine

triphosphate (ATP), to move the charged molecules against their concentration gradient. One

such example of proteins is the sodium-potassium pump or Na+-K+ ATPase which moves sodium

out of the cell and potassium into the cell, thus, establishing homeostasis by maintaining the

concentration of these ions in the intra and extracellular spaces. Thus the net membrane potential

in the resting state, measured as the difference of intra and extracellular potentials, is always

negative and equal to -70 mV. The overall expression for membrane potential therefore depends

on the concentration and membrane permeability (or conductance) to different ionic species.

This was first developed by Goldman (5) and later modified by Hodgkin and Katz (6) giving rise

to the famed Goldman-Hodgkin-Katz equation,

K Na Clo o i

K Na Cli i o

P K P Na P ClRTE ln

F P K P Na P Cl

(2-1)

In Equation 2-1, Pi refers to the permeability coefficient of the membrane to particular ionic

species ‘i’, E represents the equilibrium resting membrane potential (RMP), F is the Faraday

constant, R is the universal gas constant and T is the temperature of the medium. Even though, of

the diffusible ions, potassium, sodium and chloride are extensively studied due to their extremely

high concentrations, it is important to remember that there exist other charged particles like

bicarbonate, calcium, magnesium, phosphates, sulphates, amino acids and proteins in both the

Page 25: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

25

intra and extracellular compartments which are vital to the maintenance of homeostasis of the

cell.

Bioimpedance. The concepts of bioimpedance and bioelectricity arise from the existence

of charged particles found abundantly in cells and their environment, and from the analysis of

changes in the resting membrane potential due to an external stimulus. The effect of electric

fields on biological systems can be quite complicated even while considering a simple system

like a biological fluid. Current flow in a biological fluid is given by the sum of drift and diffusion

currents of each component making up the fluid (water, proteins, lipids etc). The applied electric

fields can change the orientation of polar molecules and induce dipole moments in bound

molecules by distorting either their electron clouds or by causing displacement of ions relative to

each other. The latter is termed ionic polarization and occurs as a result of the shift in the

hydration layer surrounding an ion due to the application of the field. The influence of an electric

field on membranes can vary with the direction of application (7). This is because, parallel to the

membrane, the induced field will exert a force on the charged particles in the bilayer and cause

them to move freely along the surface whereas in the transverse direction, these particles are

bound within the bilayer. However, the transverse field can modulate the movement of ions

through the channels by altering the membrane conductance. A further complication arises when

considering the interactions of biological fluids and cell membranes as the electric field can alter

or modulate the way new molecules are bound to the membrane surface. Furthermore, the

exchange of ions or molecules between the membrane and the fluid can be affected (8). The

result is an alteration of the transport properties of the membrane thereby altering function of the

cell which directly affects the organ system of which the cell is a part. The first equivalent circuit

of an unmyelinated nerve fiber was given in the seminal paper by Hodgkin and Huxley (4). With

Page 26: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

26

the resting membrane potential around -70 mV, and the approximate thickness of the membrane

around 7-15 nm, typical transverse electric fields in the order of tens of millions of volts/meter

exist in a resting cell. The resistivities of the intra and extracellular fluids are in the range of 0.5 -

2 Ωm with a relative dielectric constant of approximately 50-80. Cells can therefore be modeled

as a group of electronic components where the extracellular and intracellular spaces can be

represented as resistors while the BLM can be modeled as a capacitor. The first of such models

that is still considered a good approximation for the passive electrical nature of the cell was

given by Fricke in 1925 (9) where each infinitesimal portion of the extracellular and intracellular

space was modeled as a resistance and every infinitesimal portion of the membrane was modeled

as a capacitance. The capacitive membrane shields the interior of the cell from applied electric

fields that have a frequency less than 1 kHz and conducts current in the presence of high

frequency electric fields (MHz regime of the electromagnetic spectrum) (Figure 2-1). When

considering the effect of a low frequency electric field on a large cohort of cells, the overall

conductance of the system can be represented as the sum of individual conductance of all cells

while the effects of membrane capacitance and permittivity can be ignored (since the capacitors

act as open circuits at low frequencies). The current and voltage characteristics can then be

understood using Ohm’s law. At a macroscopic level, most predictions of electrical current flow

in biological tissue for diagnosis or therapy (10-14) are usually based on this simplified

representation of an otherwise complex, intricate and dynamic system.

Bidomain Model. The expressions for current density in the intra and extracellular

spaces can be described using Ohm’s law in a volume conductor. Assuming that these spaces are

isotropic, the conductivity tensors in these spaces can be represented by their scalar counterparts

(σi and σe) and the equations can be given by

Page 27: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

27

i i

e e e

i U

J

J

U

(2-2)

where Ui and Ue are the electric potentials inside and outside the cell. If β represents the surface

to volume ratio of neurons in the tissue under consideration, then current entering and leaving the

intra and extracellular domains must be equal and opposite. Representing the transmembrane

current as Im, this can be written as,

i e mJ J I (2-3)

Defining the membrane potential as m i eU U U , the equations governing the extracellular and

intracellular potentials can be given as,

i m i e e

mi i m

U U

UU C

t

(2-4)

In Equation 2-4, Cm depicts the membrane capacitance and the membrane current, Im, is assumed

to be time independent. To understand the effect of injected currents on tissue at a macroscopic

level we assume that the difference in conductivities of the intra and extracellular spaces is

negligible. Thus for a tissue at rest, the solution to Equation 2-3 is given by mI 0 and

i eJ J 0 . This implies that there exists a constant potential difference between the two

domains and no activation of tissue occurs due to the externally applied current. This forms the

basis for understanding electrical current flow or distribution in tissue following stimulation.

2.2 Transcranial Electrical Stimulation

Over the past decade, electrical stimulation techniques like transcranial direct/alternating

current stimulation (tDCS/tACS) have shown great promise in the improvement of cognition and

working memory (15) and their success has been attributed to the modulation of excitability of

Page 28: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

28

cell membranes (16,17) or by means of modulating the existing background activity (18).

Liebetanz et al. (19) pharmacologically investigated the mechanisms of action of tDCS that

contribute to neuroplasticity and concluded that changes in membrane excitability and NMDA

receptor activation played a major role in the after-effects of tDCS. Transcranial AC stimulation

has been researched extensively in an attempt to understand the modulating effect of applied

electric fields on underlying brain activity (18,20,21). Typically, these transcranial stimulation

techniques employ a pair of electrodes housed in a conducting medium (saline soaked sponges,

conductive electrode gel etc.) that facilitates the flow of current. The electrodes and conducting

medium are securely placed on the scalp of the subject for current injection. Until now, we have

discussed the presence of sodium and chloride ions in the intra and extracellular spaces.

However, the independent existence of sodium and chloride ions (without combining to form

sodium chloride) is because of the hydration ability of water due to its large dipole moment. The

hydration causes an effective increase in the size of the charged particles and hence the resulting

conductivity in tissue at low frequencies (or any ionic conductor, liquid or solid) which is due to

the motion of these hydrated ions, is governed greatly by their ionic mobilities and not so much

due to the number of charge carriers.

Electrode-Electrolyte interface. When an electrode is placed in an electrolyte, the

electrode metal being a source/sink of electrons exchanges charges with the arriving ions or

ionizes neutral substances. An electric double layer or an “interphase region” is formed as soon

as the electrode is placed in the electrolyte and transformation of electronic to ionic conduction

occurs at the electrode-electrolyte interface. In all such interphases, non-uniform distribution of

charge results in an electric potential across the interphase. Two kinds of charge transfer

mechanisms take place at the interface. One is a non-Faradaic charge transfer where there is no

Page 29: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

29

electron exchange between the electrode and electrolyte. Charges in the bulk of the electrolyte

redistribute as the electrons in the metal attract counterions in the electrolyte. The second

mechanism is one which involves transfer of electrons from the electrode to the electrolyte

causing reduction or oxidation of chemical species in the electrolyte.

Capacitive charge transfer. The reason for the double layer formation is due to several

phenomena that include a transient transfer of charge carriers between the electrode and

electrolyte resulting in two oppositely charged layers (counterion phenomenon) in the

electrolyte. Some ionic species can also adsorb to the surface of the electrode thereby creating

charge separation. Polar molecules like water can exhibit preferential orientation at the electrode

that separates charge. When the charge on the “working electrode” (defined as the electrode one

is interested in studying) is reversed, redistribution of charges in the double layer and bulk of the

electrolyte occur. However, electroneutrality at the electrode interface and in the bulk of the

electrolyte is preserved. At the “counter electrode” (electrode necessary for circuit completion)

the opposite processes occur and even here the net charge in the double layer is zero. Thus when

the total delivered charge is small only charge redistribution occurs and when the polarity of the

voltage is reversed, the charge stored in the double layer capacitance can be recovered

effectively.

Faradaic charge transfer. Charge transfer can also occur due to the transfer of electrons

from the electrode to the electrolyte through processes of oxidation and reduction. This

phenomenon of charge transfer forms products in the solution which may or may not be

recovered by reversing the current direction. Faradaic charge transfer being chemical reactions

can also be reversible to some extent if the reaction rate is under kinetic control. This means that

if the reaction is governed by fast kinetics then the product formed does not move away from the

Page 30: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

30

surface quickly (relative to the kinetic rate of the reaction) and hence when the current is

reversed, some of it can be converted to its initial form. However in a reaction with slow

kinetics, the reactants are able to diffuse to the surface of the electrode to support the kinetic rate

of the reaction and the formed products diffuse away quickly in comparison. These products that

have diffused away cannot be recovered and could be damaging to the electrode or the tissue

undergoing stimulation.

At equilibrium, the net electrochemical potential across the electrodes and the electrolytes

is zero and charge transfer is ceased. Now if current is injected using the electrodes, the net

current during electrical stimulation is defined by the sum of charge transfer following the above

two methods where the capacitive transfer depends on the rate of change of the equilibrium

potential while the Faradaic transfer is affected by shift of the potential from electrochemical

equilibrium. In most electrical stimulation applications, the stimulation is delivered using a

current controlled approach in the form of pulses. Stimulation using biphasic pulses causes the

second pulse to reverse any electrochemical processes occurring as a result of the first pulse. A

detailed understanding of the underlying processes and effect of different stimulation approaches

can be found in Merill et al. (22).

In spite of the simplified representation of the underlying processes that govern current

flow in the brain, there still remains ample complexity arising due to its heterogeneous nature

making it extremely challenging to design rigorous experimental paradigms that can faithfully

testify and corroborate the pros and cons of electrical stimulation techniques. Variations in the

site of electrode placement, intensity of stimulation and even duration of stimulation have been

shown to produce different effects and prompted researchers to establish specific scientific

guidelines for performing non-invasive brain stimulation (23). Computational models are widely

Page 31: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

31

used as guiding tools for targeting areas of interest in the brain but these models are developed

on a general atlas of the head which may decrease the accuracy of prediction. For instance, an

atlas based model is blind to the variation in skull thicknesses or ventricle sizes across subjects

and hence attributing any observations solely to electrical stimulation would be inaccurate.

Subject specific computational models would serve as better estimates but this requires some sort

of a priori information about the subject’s head that can be obtained with the aid of Computed

Tomography (CT) or Magnetic Resonance Imaging (MRI). Even with this information, there

could be variation in the amount of hair across participants, information about which cannot be

obtained by these imaging methods. This could also affect current distribution. The need for

accurate segmentation of scalp and bone tissues, and taking into account the different anisotropy

of biological tissues make it extremely challenging for computational models to accurately

predict the injected current distribution, calling for new approaches that can accurately quantify

the localization of current density in different brain regions following electrical stimulation.

Magnetic Induction Tomography (MIT), Magneto Acoustic Tomography with Magnetic

Induction (MAT-MI), Magnetic Resonance Electrical Properties Tomography (MREPT) and

Electrical Impedance Tomography (EIT) are some of the techniques developed in the past to map

the conductivity of the imaging object and thereby understand the current distribution. Each of

the above techniques has its own merits and demerits in terms of portability and invasiveness of

measurement. Over the last decade, MRI coupled with EIT, called MREIT (24), has come into

prominence showing promise with its ability to reliably detect injected current distributions at

high spatial resolutions.

2.3 Magnetic Resonance Imaging

Magnetic Resonance Imaging, commonly referred to as MRI, is a diagnostic imaging

approach that facilitates the generation of non-invasive high resolution images of an object and is

Page 32: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

32

currently the gold standard for imaging soft tissue in the human body. MRI in its various forms

(diffusion MRI, functional MRI, quantitative MRI, etc.) has proven to be extremely useful in

diagnosing a wide gamut of disorders in their incipient stages.

At its root, MRI stems from a thorough understanding of Nuclear Magnetic Resonance or

NMR which was measured and described for the first time by Isador Rabi (25) in 1938, which

won him the Nobel Prize in Physics. Felix Bloch (26,27) and Edward Purcell (28,29) extended

this work to liquids and solids and were jointly awarded the Nobel prize in 1952. In 1948, the

BPP theory (30), named after its formulators and renowned physicists Nicolaas Bloembergen,

Edward Purcell and Robert Pound, explained and demonstrated the interaction between nuclei

immersed in an external field with their surrounding lattice and neighboring nuclei based on

which modern day diagnostic MRI is understood. NMR received a major thrust and its

diagnostic potential came to light with the work of Raymond Damadian (31) who showed that it

was possible to identify cancerous lesions by their sufficiently long relaxation times. With the

advent of computers and computerized tomography, Damadian attempted to build a prototype of

the modern MR scanner. However, it was the pioneering work of Paul Lauterbur in 1973 (32)

and Sir Peter Mansfield (33-35) who realized that a spatially varying magnetic field in the form

of magnetic field gradients produced a distribution of spatially varying nuclear spins and that

helped create the first proton density images which paved the way for modern day Magnetic

Resonance Imaging.

Spin is an intrinsic quantum property of atomic nuclei. The overall spin of the nucleus is

determined by the spin quantum number ‘s’ and can be ±1/2 (or their integral multiples), referred

to as the spin-up and spin-down states. When the total number of neutrons and protons in a

nucleus are equal, the overall spin can be considered to be zero. Hydrogen ion (H+) has a total

Page 33: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

33

positive charge due to the presence of just the one proton in the nucleus giving H+ a total spin of

+1/2. The human body comprises of an abundance of protons whose spins can be detected and

localized using MRI making proton MRI the most viable methodology for diagnostic imaging. In

the remainder of this dissertation, for simplicity, proton MRI will be explained using a classical

approach and referred to as simply MRI.

The spin-up and spin-down states have the same energy and hence the number of protons

(or simply spins) occupying these states are equal in conditions of thermal equilibrium. When

these spins are placed in an external magnetic field B , the densities of spins in the two states are

altered (described by Boltzmann statistics) such that there is an excess of spins occupying the

lower energy state (parallel to the external field). Though this value is extremely small, the sheer

number of protons available in biological tissue makes up for this smallness giving rise to a net

magnetic moment (magnetization) that precesses around the external field with an angular

frequency that is dependent on the gyromagnetic ratio γ of H+ and the strength of the external

field, B . The coordinate frame used to describe this precession is a right handed system and so

we can write the precession frequency (ω) as,

(2-5)

A key point of significance is that the frequency of precession, also called the Larmor frequency,

happens to be proportional to the difference in the energy of the two spin states of the proton.

This frequency is in the MHz range and therefore, in order to excite the spins and create a

resonance condition, it is imperative that the energy deposited falls in the radio frequency range

of the electromagnetic spectrum. Therefore, an RF coil can be used to generate sufficient energy

to tip the spins away from the external field (along z direction) at any arbitrary angle θ thereby

even causing them to precess in the transverse plane (xy plane). The net magnetization of these

Page 34: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

34

precessing spins will induce a small voltage signal in a nearby coil following Faraday’s laws of

electromagnetic induction. This induced signal is then mixed with a reference frequency equal to

the Larmor frequency of the spins to enable recording (this is equivalent to looking at the spins

from a frame of reference rotating at their precession frequency) and then digitized for storage.

The precessing magnetization will return to its equilibrium value Mo along the z direction, upon

the removal of the RF field due to the interactions of the spin with its surrounding lattice

(longitudinal relaxation) and due to its interaction with the neighboring spins (transverse

relaxation). These two mechanisms can be described as exponential regrowth and decay of the

magnetization with time constants T1 and T2. The time evolution of the precessing magnetization

can be understood from the Bloch equation, Equation 2-6, that captures the effect of these

relaxation parameters and is given as,

x y o z

2 1

ˆ ˆ ˆM x M y M M zdMM B

dt T T

(2-6)

where Mo is the magnitude of the equilibrium magnetization, x y zˆ ˆ ˆM M x M y M z and

ˆB B z .

Image formation. In MRI we use three orthogonal magnetic field gradients that produce

a variation in the z-component of the main magnetic field along the three coordinate axes, in

order to create a distribution of spins that precess with different Larmor frequencies. This can be

written as,

z z zˆ ˆ ˆˆr B z G.r where GB B

i jB

x y z

k

(2-7)

Page 35: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

35

where r is a position vector in space, Bz represents the magnitude of the net magnetic field along

the z direction and G is the gradient of Bz. The complex time-dependent signal detected at the

end of the receiver chain of an MRI system, ignoring relaxation effects, can be represented by,

i r, t

r des t r

(2-8)

In Equation 2-8, ρ refers to the spin density, ϕ is the phase accumulated by the magnetization

vector in the transverse plane over the time interval t. The phase ϕ can be written as 2 k.r

where t

0

k G t dt2

and is called the wave vector which represents the spatial frequencies

present in the image. A very basic explanation (without much math) for generation of a two-

dimensional (2D) image along the z direction in the sample is as follows: the magnetic field

gradient applied along the z direction (Gz) creates a spatial variation of the precession

frequencies along the z direction described by the following the equation

zz B G z (2-9)

In reality it is not possible to create a perfect 2D slice, and the applied RF pulses excite a band of

frequencies (band-limited) thereby providing a finite thickness to the “slice”. Therefore as the

bandwidth of the RF pulse increases, we obtain thicker slices as we now excite a bigger region of

precessing spins. The slice thickness can be related to the RF excitation bandwidth as,

o zB G z (2-10)

If the RF excitation pulse (with a frequency content Δω) is applied along x, it will tip the spins in

the region Δz oriented along z, towards the –y direction causing them to precess in the xy plane,

while those that are off-resonance i.e spins that are not defined by the precession frequency range

Δω, remain largely unaffected by the pulse. Ignoring relaxation effects, this effectively produces

Page 36: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

36

a slab of spins precessing in the xy plane. A subsequent field gradient turned on for a brief period

of time τ along the x-direction will cause the slab of spins to precess with varying frequencies

along the gradient direction effectively creating rows of spins along the x direction that are

precessing with the same Larmor frequency. However, since the gradient is turned on only for a

brief period of time, the spins experience this additional field gradient only for the period τ,

accumulating a net phase, γGxτ. Upon removal of the gradient, the spins begin to precess with

their initial Larmor frequencies but can now be distinguished due to their differences in phase.

This is termed as “phase encoding”. A third gradient now applied perpendicular to the initial two

gradients, causes the spins along each row in the slab to vary in frequency thereby creating a slab

of spins where each spin is unique in terms of its frequency and phase. This is called “frequency

encoding”. A receive coil used to detect the signal from the entire slab at each frequency

encoding step. The frequency content of the detected signal can be analyzed using a 2D inverse

Fourier transformation which enables us to localize the various detected frequencies in 2D space.

Mathematically, the entire process can be summarized as,

x yi2 k x k y

x yS k ,k x, y e dxdy

(2-11)

The inverse Fourier transformation operation yielding the desired spatial distribution of spins

(proton density), can then be written as,

i2 k rr S k e dk

(2-12)

Note that although the slice, phase and frequency encoding directions were assumed to be along

z, y and x directions, it is completely reasonable to switch around the directions of encoding. For

this reason it is always convenient in MRI to refer to the data acquisition in terms of the axes of

encoding (frequency, phase and slice) as opposed to the laboratory coordinate axes.

Page 37: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

37

2.4 Magnetic Resonance Electrical Impedance Tomography

In Electrical Impedance Tomography (EIT), conductivity of the underlying tissue is

extracted by measuring potentials at the boundary of the object. This represents an ill-posed

problem which can be solved using additional information obtained from the MRI scanner.

Therefore the combination of the two techniques gave birth to MREIT (Magnetic Resonance

Electrical Impedance Tomography). The approach bears roots in the ability to produce current

induced magnetic field maps using a conventional MRI scanner, first shown possible in the

1990s by Scott et al. (36,37). The authors computed the injected current density J, by measuring

the magnetic field map Bc, produced by the current. To estimate all three components of the

magnetic field vector Bc, they rotated the object in three orthogonal directions. The idea behind

MREIT is that the z component of Bc interacts with Boz of the scanner and produces a change in

the spatial distribution of main magnetic field that can be measured as a variation in the phase of

the measured MR signal. The measured 2D spatial distribution of spins in MRI is described by

Equation 2-11, where the term x yi2 k x k y

e

refers to the phase of the measured MR signal

resulting from the phase and frequency encoding gradients. The current induced magnetic field

imparts an additional phase term to the precessing spins, described by c cT , where Tc is

the total duration of current injection. Equation 2-11 can then be modified as,

x y c ci2 k x k y i T

x yS k ,k x, y e e dxdy

(2-13)

The MR acquisition is repeated with an alternating current polarity, and the resulting datasets are

complex divided to extract the magnetic field image (38). The following equations describe the

process of magnetic field extraction from MRI data where So represents the MR signal in the

absence of current injection, ρ is the 2D spin density, kx & ky are the wave vectors in the phase

and frequency encoding directions. S+ and S- represent the MR signal in the presence of positive

Page 38: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

38

and negative current injections applied for time Tc within one repetition time TR of the pulse

sequence. ϕc represents the additional phase imparted by Bc to the MRI signal.

x y

c c

c c

i2 k x k y

o

i

c c c

T

o

i T

o

S x, y e dxdy

S S e

S S e

Sarctan B2

ST

(2-14)

From Equation 2-14, one component of the current induced magnetic field can be measured as,

cc

c

B2 T

(2-15)

Figure 2-2 shows a typical arrangement of an MREIT experiment using a cylindrical agarose gel

phantom doped with sodium chloride and copper sulphate, and placed in the bore of an MRI

scanner. The idea is to inject current from left to right thereby creating the magnetic field Bc

which interacts with Bo. Seo et al. (24) analyzed the distribution of current density within the

phantom and formulated the following equations,

J = 0 inside the phantom (Ω

ˆJ.n = g on the boundary ( Ω)

)

(2-16)

The first equation in Equation 2-16 describes the conservation of current owing to preservation

of electro neutrality of the phantom. In the second equation, ‘g’ is the Neumann boundary

condition due to injected current describing the presence of current near only the electrodes and

is zero everywhere else on the boundary of the phantom (since no current is radiating out of the

phantom, ˆJ n 0 , except at the electrodes). The magnetic field produced due to this distribution

of current density J, can be described using the Biot-Savart equation as.

3

r rB r J r dr

4 r r

(2-17)

Page 39: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

39

By treating the phantom as a volume conductor, the current density can be linearly related to the

electric potential (U) as J U and thus the problem in MREIT becomes that of estimation

of σ from the measured component of B(r) (39). This is a very complicated problem due to the

unavailability of the full B vector and also because σ is generally a macroscopic parameter that

can be anisotropic, complex (has real and imaginary parts) and depends on both temporal and

spatial frequencies. Making an assumption that the underlying conductivity is isotropic, using the

steady-state Maxwell’s equations, it can be shown that,

1

n JJ J l

(2-18)

From Equation 2-18, the nature of the component of ln that is parallel to J is unclear. It is

for this reason that in traditional MREIT acquisitions, current is injected in two non-collinear

directions. It is important to note that if the conductivity is believed to be anisotropic, then the

conductivity tensor can be estimated by injecting current in seven different directions (40). There

exist numerous reconstruction algorithms developed to compute σ or J, from the measured field

maps. Seo et al., Oh et al. and Park et al. (39,41) used the harmonic Bz algorithm which

computes the conductivity images based on the relation between 2B U . Nam et al.

proposed the J substitution algorithm (42) while Birgul et al. used the current constrained voltage

scaled reconstruction (CCVSR) (43). The harmonic Bz algorithm is advantageous in that no

rotations of the imaging object are necessary for the computation of J. Park et al. analyzed the

projected current density approach (PCD) (44) to reconstruct two dimensional current density

with some approximations. One advantage of using the harmonic Bz algorithm and the PCD

approach is that they work by incorporating the Laplacian of the current induced magnetic field

which remains unaffected by the magnetic field produced due to surface currents on the

Page 40: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

40

electrodes and the current in lead wires. This can be proved by using the Biot-Savart equation

described in Equation 2-17. For a current density J, within a domain Ω, the magnetic field due to

that current density can be calculated using the Biot-Savart equation. As the gradient of 1/r can

be written as –r/|r|3, Equation 2-17 can be written as,

o

J r 'B r dr '

4 r r

(2-19)

where J is the current density inside the domain Ω. From the expansion of curl of curl of a vector

field, given that J 0 and fg f g f g , we can write Equation 2-19 as

21 1B J r dr J r dr r

4 r r 4 r r

(2-20)

Using the identities, 1 1

r r r r

and 2 14 r r

r r

, gives us,

oo

o

1B J r dr J r , and integrating by parts results in,

4 r r

J rBJ r dr , r

4 r r

(2-21)

Using the divergence theorem, the above can be written as,

o

ˆJ r n rBJ r dS

4 r r

(2-22)

Since the current J in the domain Ω can be expressed aso

B

, the second term on the right hand

side of Equation 2-22 can be written as,

l

o

B r r ˆJ r n rdS

4 r r

(2-23)

Page 41: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

41

where Bϵ and Bl represent the magnetic field due to surface currents on the electrodes and the

magnetic field due to the current in the lead wires, such that, lB B B B . The influence of

Bϵ and Bl can be reflected to the boundary condition ˆJ.n g , implying,

l

2

2

B r B r 0 , r as,

10 r r

r r

(2-24)

Since the reconstructed conductivity and current density images rely on the fidelity of the

Laplacian of induced magnetic field maps, it is imperative that the measured field maps are

ideally noise-free.

2.4.1 Signal and Noise in MREIT

In MREIT, the reliability of the reconstructed conductivity or current density map is

governed by the noise characteristics of the measured induced magnetic field which is a scaled

form of the measured phase of the MR signal. Therefore we rely on the optimization of pulse

sequences and imaging parameters to generate phase changes that are above the image noise

floor. In general, noise in MR images can arise due to a multitude of sources making it

problematic to observe the tissues of interest. The precessing transverse magnetization is

detected by a receiver coil and the induced voltage is manipulated through a series of electronic

circuitry collectively called the receiver chain. As with any measurements, systematic noise in

the measuring apparatus, in this case the electronics, affects the signal-to-noise ratio (SNR) that

can be achieved. In MRI, SNR is dependent on various factors not limited to imaging parameters

like voxel size, number of averages, bandwidth of the low pass filter present during signal

readout (BWread), thermal fluctuations in the sample being imaged and the impedance of the

Page 42: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

42

receive coil dictated by the amount of loading due to the sample. The measured k-space MR

signal for 2D imaging represented by Equation 2-11, can be extended to 3D and re-written as,

2 k rk dS r e r (2-25)

where the term ρ(r) represents the effective spin density. Equation 2-25 represents the continuous

sampling of the signal while in reality, the MR signal is sampled as a discrete set of points in 3D

k-space. If the total image comprises of Nf, Np and Ns data points each separated by Δkf, Δkp and

Δks in k-space, then the signal in a voxel of the reconstructed image can be written as,

f p s

mm' nn ' pp 'i2

N N N

f p s

m',n ',p 'f p s

1m f ,n p,p s S m ' k ,n ' k ,p ' k e

N N N

(2-26)

Equation 2-26 shows the relation between the reconstructed signal in a particular voxel and its

corresponding k-space counterpart (45). This reconstructed signal is proportional to the volume

Δf.Δp.Δs and as the volume of voxels, which is controllable in an MR experiment, is increased,

the net signal is also increased thereby increasing the SNR. As the Fourier transform is a linear

operation and the recorded signal comprises of the true signal along with a noise component, the

variance of the measured signal due to noise also depends on the volume of the voxel. The

variance of the induced voltage in the receiver chain can be written as 4kT R BW where R

represents the effective impedance due to the sample, coil and the electronics. This noise is

considered to be “white” as it is equally distributed in power across all frequencies within the

readout bandwidth. As white noise can be characterized as a Gaussian distribution with zero

mean and variance, σm2, the variance of noise (white) in the image domain (σo

2) can be written

as,

2

m

f p sN N N

(2-27)

Page 43: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

43

i.e noise in the image domain is spread equally over all the voxels and is considerably lower.

Repeated measurements of the imaging experiment can further decrease the noise level and the

variance of the measured noise is decreased proportional to the number of measurements (Navgs).

The overall SNR in a voxel can then be written as,

avgs

f p s

f p s N

BW

N N

N

N

AS R

(2-28)

To better understand the signal and noise characteristics in an MR image, it is necessary to pay

attention to the two components that make up the image, namely the magnitude and phase.

2.4.2 Magnitude noise

The magnitude images are constructed from the real and imaginary images produced

from the mixing stage of the receiver chain as the square root of sum of their squares. This is a

non-linear mapping and therefore the noise in magnitude images no longer follows a Gaussian

distribution but can be represented using a Rician distribution. The Rice density (46,47) can be

represented as,

2

2

2M A

2M 2 2o

M A MP (M) e I

(2-29)

where, ‘PM’ is the probability distribution for the measured pixel intensity ‘M’. ‘A’ is the image

pixel intensity in the absence of noise. Io is the zeroth order Bessel function of the first kind and

‘σ’ is the standard deviation of Gaussian noise in the real and imaginary images. The Rician

distribution starts to approximate to Gaussian distribution for SNR values that are greater than

three. In the regions outside the sample (background in the image), since there is no “true” MR

signal, A in Equation 2-29 becomes zero and we get,

Page 44: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

44

2

2

2

M

2M

MP (M) e

(2-30)

Equation 2-30 describes a Rayleigh distribution. The standard deviation of such a distribution

can be related to the Gaussian distributed white noise, σo, as 0.655σo.

2.4.3 Phase noise

The phase images are generated by calculating the four quadrant arctangent of the ratio of

imaginary and real images which, like the magnitude calculation, is also a non-linear function,

and hence the distribution of noise is no longer expected to be Gaussian. The distribution of

phase noise (Δθ) can be given by,

2 2 22

2 2

A

2

Acos

A cos x

2 21 A

P e 1 2 c1

e e do x2

s2

(2-31)

Though the above expression seems extremely complicated, in the limits A = 0 and A ≫ σ the

distribution can be understood much more easily. In regions with only noise, A=0 and the above

equation converges to 1/(2π) as long as -π < Δθ < π. This is equivalent to a circle in the complex

plane since the noise vector is equally likely to point in any direction. The phase noise

distribution as a function of Δθ in Equation 2-31 is shown in Figure 2-4. When A ≫ σ, the

integral in Equation 2-31 reaches 1 and the distribution can be written as,

2

2

2

2 2A 1 cos2

2

2 AA c

2

A

os 1P e e

2

(2-32)

which represents a Gaussian distribution. This result corroborates the idea that if one is to

represent the magnetization as a vector in the complex plane, the variance in the magnetization

due to noise can be decomposed into parallel and perpendicular vectors along the magnetization,

Page 45: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

45

and given that the magnitude of the magnetization is very large, the parallel component can be

ignored. The perpendicular component is then responsible for the deviation of the magnetization

from the real axis thereby causing an uncertainty in the phase of the magnetization (Figure 2-3).

Using the approximation that tan for a small angle δ, the uncertainty can be linearized as

σp/A where σp represents the perpendicular component of noise. The standard deviations for the

phase noise can then be given as 1

SNR

when A

A (45) and

22 if 0

3

A .Therefore, in

MREIT, the noise standard deviation of the magnetic field calculated as in Equation 2-15 can be

written as,

c

cB

c c

1

2 T 2 T SNR

(2-33)

From Equation 2-33, it is clear that the noise in the induced field maps can be decreased by

increasing the signal-to-noise ratio in magnitude MR images. Another factor determining the

noise levels is the current injection time. As the current injection time increases, greater phase

accumulation due to the current’s magnetic field occurs increasing our ability to characterize

magnetic fields produced by low amplitude currents. For currents in the order of hundreds of

microampere the accumulated phase is restricted to the interval [-π,π] and phase unwrapping

algorithms are not necessary. However, the increase in Tc is limited by the T2 of the sample

being imaged which controls the degree of magnetization available thereby controlling the SNR

in the magnitude image. Thus there exists a tradeoff between increasing the current injection

time (Tc) and achieving the best possible SNR that dictates the reliability of measured phase

change for a given MREIT acquisition.

Page 46: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

46

2.4.4 Optimization of imaging parameters to improve sensitivity

A typical spin echo MREIT sequence is shown in Figure 2-5. The magnetization signal

detected in MRI depends on the longitudinal and transverse relaxation times, T1 and T2, of the

precessing spins and is given by the solution to Equation 2-6 at echo time as,

1 2

TR TE

T TM x, y x, y 1 e e

(2-34)

Expressing the echo time (TE) of the imaging sequence in terms of RF pulse width duration

(assuming the 90° and 180° pulse have the same pulse width, τ), sampling time interval and the

current injection time we have,

srfc

TT

2T

2

3E

(2-35)

Since the SNR of the experiment is dependent on T2 and is directly proportional to the square

root of sampling time, Ts, the noise standard deviation in Equation 2-33 can be rewritten as,

2

c

TE

T

B

rfc c

e

3T 2 TE T

2

(2-36)

Differentiating Equation 2-36 with respect to Tc to obtain the optimal pulse width for

minimizing the noise standard deviation, we obtain the optimal value as,

rc f

2T TE 3

3 (2-37)

This gives an optimal sampling time that corresponds to the current injection time as,

rs cf

2T TE 3 T

3 (2-38)

Injecting current into the sampling interval was performed by Kwon et al. in 2007 (Figure 2-6).

Though the injected currents inside the sampling interval distort the linear relation between spin

Page 47: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

47

position and precession frequency, the magnetic fields due to injected currents are extremely

small when considering traditional image resolutions (millimeter) and their effects can be

ignored. However, taking into account the non-linear mapping could potentially improve the

sensitivity of measurement considerably (as high as 40 %). For such an experiment, the authors

predicted the optimal current injection time and sampling times to be,

rfc

fs r

2

33 3T TE

3

2 3T TE 3

3

(2-39)

The above equations describe the optimization of the injection current pulse width to obtain the

best possible noise characteristics in the induced magnetic field map. Assuming perfect

electronics, increasing the sampling time to decrease the bandwidth of the image acquisition

could lead to unwanted off-resonance effects like chemical shift artifacts. Decreasing the noise

standard deviation in the induced field map can also be achieved by improving the SNR of the

magnitude image. One possible way for achieving this is to use a multi-echo sequence as

opposed to a single echo and using the relaxation time envelope governing the signal over the

multiple echoes to obtain an average map that has the best SNR characteristic. However with

multiple RF spin echoes, imperfect 180° pulses pose a problem and could cause an inherent

signal loss affecting the SNR of the image. Recently Oh et al. (48) employed a multiple gradient-

echo sequence to decrease the noise standard deviation of the current induced field maps that

showed great promise for the measurement of small magnetic fields due to low amplitude current

injections.

Page 48: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

48

2.4.5 Projected Current Density

In MREIT, reconstruction of conductivity distribution requires the use of at least two

independent currents to tackle the problem of uniqueness as described by Equation 2-18.

However in 2007, Park et al. (44) developed the projected current density approach that

facilitates the reconstruction of current density using just one injection current, under

assumptions of a transversal two-dimensional internal current distribution (for example in the

XY plane).

For a cylindrical imaging domain Ω that can be represented by the union of ‘m’ thin

slices, Ωt where t 1,m , Jp is a projection of the internal current density J inside the domain Ωt.

Jp can be uniquely determined and decomposed into curl free and divergence free components Jo

and J*. Under the assumption that the component of current density perpendicular to the slice Ωt

is the same for both J and Jp, it can be shown that J =Jp can be estimated from measuring Bz in

just one direction. The curl free component of Jp, Jo, is a two-dimensional current density due to

a homogenous potential α and can be represented by . J* (divergence free component) can

be represented by the curl of a two dimensional field β. Using Equation 2-16, α and β then satisfy

the following equations,

2

22 c

o

0 in

ˆ ˆn J.n on

BJ in

0 in

n 0 on

(2-40)

where Bc is the calculated current induced magnetic field from Equation 2-15 and n is the

outward normal vector on the boundary ∂Ω. Park et al. (44) showed that the measured projection

of J, Jp, is unique and independent of the conductivity of the sample being imaged as long as the

Page 49: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

49

internal current is predominantly planar in Ωt. For a three-dimensional distribution of J, the

difference vector pJ J is calculated and the authors show that the norm of the difference vector

pJ J up to some tolerance is dependent on the norm of the difference of z-components of J

and Jp. The reconstruction of projected current density is carried out using the following steps.

1. Assuming a uniform conductivity, σo = 1 S/m, calculate the internal current density oJ

and the magnetic field o

cB due to oJ , using a computational model.

2. Using the calculated o

cB and the measured Bc, calculate the projected current density given

by Park et al.(44) as

o

m o m o

c c c c

o

B B B B1J

xJ 0

y, ,

(2-41)

3. Iterate the reconstruction procedure to improve the reconstruction of J using the

measured Bc by reducing the error between J and oJ till desired tolerance is achieved.

Page 50: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

50

Figure 2-1. Graphic representation of low frequency and high frequency current flow in tissue.

Figure 2-2. Schematic for experimental setup in an MREIT experiment while injecting current

into a cylindrical agarose gel.

Page 51: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

51

Figure 2-3. Representation of magnetization vector and the effect of noise on the phase (ϕ) of the

magnetization vector. A represents the magnetization in the absence of noise, and M,

in the presence of noise. For simplicity the deviation in ϕ due to noise is shown to

increase ϕ though in reality it could decrease ϕ as well.

Figure 2-4. Distribution of phase noise for varying levels of magnitude signal-to-noise ratio. The

distribution begins to approximate to a Gaussian distribution for SNR values greater

than 3.

Page 52: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

52

Figure 2-5. Spin echo pulse sequence for MREIT acquisition. The current polarity is reversed

after the 180° pulse to account for the phase reversal of spins.

Figure 2-6. Spin echo sequence with current injection during the sampling interval altering the

linear relationship between the spin position and frequency.

Page 53: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

53

CHAPTER 3

DETERMINATION OF THE EFFECT OF TEMPORAL VARIATIONS IN THE STATIC

FIELD ON MREIT ACQUISITIONS

3.1 Introduction

The field of neuromodulation has gained immense popularity over the past decade. In

particular, transcranial electrical stimulation techniques have been in the spotlight as a result of

the concerted effort from several research groups attempting to understand transcranial

direct/alternating current stimulation (tDCS/tACS). tDCS and tACS have been shown to be able

to modulate cortical activity and hence might serve as augmentative therapies to promote

rehabilitation if carefully controlled. tDCS is hypothesized to operate by modulating the resting

membrane potential of neurons making them either hyper or hypo excitable (17), whereas tACS

works by modulating the background neuronal activity occurring in the brain (18). Effects of

these types of electrical stimulations have been predominantly studied by understanding the

behavior of participants undergoing electrical stimulation and few research groups have

attempted to quantify current induced changes on the brain tissue (49). There exists ample

literature showing the use of tDCS in the MRI scanner (23) where researchers have attempted to

capture changes in brain connectivity during electrical stimulation. However, given the

complexity of brain connections and the concept of functional connectivity which addresses the

relation between two different regions of the brain involved in the same functional task (50),

understanding the distribution of induced electric currents becomes extremely challenging. (51).

Scott et al., in the late 80s, imaged current flowing in a gel phantom using the concept of MRI,

thus laying the foundation for current density imaging. It was later improved in the early 2000s

and came to be called as Magnetic Resonance Electrical Impedance Tomography (MREIT) (36).

MREIT relies on changes produced in the phase of the detected MR signal caused by injected

currents. These phase changes are reflective of the local magnetic field from which the current

Page 54: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

54

density can be calculated. Further, computational models to predict the distribution of current

inside the human head have been developed (52). Given the conductivity of brain tissue and

applied current amplitude, currents that reach the cortex and subcortical structures are extremely

small and so are the magnetic fields due to such small currents. Therefore, measurements of

these fields are easily corrupted by the presence of noise and any variations in the field

homogeneity or stability. Moreover, there exists an inherent drift in the static magnetic field of

superconducting magnets owing to imperfect construction joints which is usually quoted in the

magnet’s specifications. Fast imaging sequences could cause heating of resistive magnet

structures such as existing passive shims and alter their calibration for generating a homogenous

field. For high amplitude current injections, shifts in the proton resonance frequency due to

current-induced local heating could also pose a serious problem. In MREIT, the phase change in

the MR signal due to injected current is calculated as the four quadrant inverse tangent of the

complex division of positive and negative current injected datasets (36).

This chapter addresses the effect of variations in the static magnetic field on MREIT

measurements and the development of necessary electronics for facilitating acquisitions that are

relatively insensitive to the magnetic field inconsistencies. Experimental data following data

collection using two methods is presented and the usefulness of the second method in

circumventing sources of static field inhomogeneity is demonstrated.

3.2 Theory

Variations in the static magnetic field of an MR scanner can be calculated directly from

measured phase in the MR image, which is advantageous to MREIT. By measuring phase

changes relative to a baseline condition, the net change in image phase can be calculated which

is reflective of the applied current-induced inhomogeneity in the static field. The MRI signal

equation for a two dimensional image acquired without any current injection can be written as,

Page 55: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

55

oi

o

x,yS x, y S x, y e

(3-1)

where ϕo represents the background phase occurring due to various sources including the RF

pulse used to create the image, sample loading on the detection coil, variations in electronics etc.

In gradient echo imaging, the measured signal is affected by two important parameters. One is

the T2 effect due to spin-spin relaxation and the other is the presence of any local field

inhomogeneity that could cause a dephasing of spins in the affected voxel. The final

reconstructed image at the echo time (TE) can then be written as,

ooi , TE

o

i B x yS x, y S x, y e e

(3-2)

The background inhomogeneity can be extracted by performing repeated measurements

at varying time intervals and estimating the variation via the image phases (53). Since the

temporal variation of the main magnetic field is unknown, the time dependent phase evolution of

spins can be written as,

1

2

t

1 1 o o

0

t

2 2 o o

0

o

x, y, t x, y B x, y, t dt

x, y, t x, y B x, y, t dt

x, y, tB x, y, t

t

(3-3)

In Equation 3-3, Δϕ refers to the net phase change in the time interval 2 1t t t . The temporal

drift of the main magnetic field can be assumed to be linear in the absence of variations

occurring due to heating of resistive structures. This allows us to drop the integral representation

for describing the phase of spins and write the net change in magnetic field as the last equation in

Equation 3-3. Repeated measurements can be made at different intervals either by using a fixed

echo time and creating fresh transverse magnetization for each measurement (Equation 3-3) or

Page 56: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

56

by following the transverse magnetization phase at multiple echo times for a single excitation as

in a multi-echo acquisition,

o

x, y,TEB x, y,TE

TE

(3-4)

A shift in the phase of an MR image can also be generated by shifting the 180° RF pulse of a

spin echo sequence. Such a shift causes a difference in the durations of precession before and

after the 180° pulse resulting in the accumulation of phase due to existing background field

inhomogeneities that can be specifically measured according to the equation,

1,2 1,2

o o fp

o fp

x, y, t x, y 2 B x, y, t

x, y, t 2 B x, y, t

(3-5)

where, τfp is called the free-precession delay term (54). In MREIT, in addition to any existing

background field inhomogeneity, we impress an additional current-induced magnetic field in

order to extract it (which is an inhomogeneity in the static field) as a change in the phase of the

signal. The equations for the complex divided image in MREIT and the current induced field can

be given as,

c co o

c c

i B x,y Ti B TE

B x,y

i

o

i2

c

T

c

S x, y S x, y e

S x, ye

S x, y

S x, yarctan

S x, yB x, y

2

e e

T

(3-6)

where Bc is the magnitude of the current induced magnetic field and Tc is the duration for which

the induced magnetic field persists. Due to the complex division to extract the current induced

magnetic field, static field inhomogeneities that remain unchanged during the acquisition of the

positive and negative current injection scans do not affect MREIT measurements. The work

Page 57: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

57

detailed in this chapter focuses on the effect of temporal changes in the static magnetic field on

MREIT acquisitions and the development and use of an acquisition protocol that helps mitigate

these effects in current injection experiments.

3.3 Methods

3.3.1 Pilot Experiments

3.3.1.1 Phantom preparation

An agar gel phantom was prepared by mixing 25 g of agar in 1200 ml distilled water. 3 g

of NaCl and 0.3 g of CuSO4 were added to the solution to increase its conductivity and facilitate

conduction of injected current. The mixture was heated in a microwave for uniform heating.

Dissolution of the agar resulted in some frothing causing the liquid to rise. At this time the

microwave was turned off and the mixture was allowed to drop back to its initial level. This

process of heating and cooling was repeated until all of agar (seen as a precipitate in the bottom

of the beaker) was completely dissolved. Once the solution no longer rose but boiled in place, it

was taken out, cleared of froth, and was poured into a 3 liter beaker with an approximate

diameter of 16 cm. The beaker was covered to trap the heat and maintain the liquid state of the

mixture while another identical solution of 1200 ml was prepared and combined with it to

achieve a total volume of 2400 ml. The mixture was left to gel for over 6 hours at room

temperature following which it was refrigerated until it was hard facilitated easy removal from

the beaker. The phantom was allowed to remain at room temperature for at least a couple of

hours prior to use in the imaging experiments.

3.3.1.2 Image acquisition

The phantom was placed in the bore of a 3 Tesla Philips Achieva magnet and a spoiled

gradient echo sequence with ten echoes (Figure 3-2) was used to acquire 2D axial images using a

32 channel head coil. The imaging resolution was 2.24 x 2.24 mm using a matrix size of 100 x

Page 58: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

58

100. The bandwidth used was 550 Hz/pixel. A repetition time (TR) of 50 ms was used for

acquisition of each phase encode step where the ten echoes were collected.

3.3.1.3 Data inconsistencies

Electrode cables were attached to the exposed ends of magnet wire that was driven into

the gel phantom and current was injected using the experimental setup shown in Figure 3-3.

Current of one polarity was injected throughout the acquisition for every phase encoding step

and the 2D axial images were acquired at ten different slice locations. The time taken for

acquiring one such volume of data was approximately 50 seconds. To improve the signal-to-

noise ratio, 4 such 3D volumes (dynamics) were acquired amounting to a total acquisition time

of approximately 3½ minutes. This entire acquisition was repeated for the opposite current

polarity and magnetic field maps were calculated from these two current injection datasets using

the Equation 3-6. The contribution of current-induced fields during the sampling interval was

predicted to be extremely small and therefore not accounted for in the data processing. The

resulting field maps are shown in Figure 3-5. When repeated measurements were performed with

identical imaging and electrical stimulation parameters, very different magnetic field maps were

obtained (Figure 3-5) which indicated a temporally unstable static magnetic field. Variations in

the static magnetic field could alter the reference against which the current-induced field maps

were computed thereby producing inconsistent results.

3.3.1.4 Troubleshooting data inconsistencies

To understand the inconsistencies observed, experiments using the same acquisition

method as in the pilot experiments were performed without any current injection to map the

temporal variation of the static magnetic field. To characterize the temporal variation, each

volume (3D) of acquisition was repeatedly acquired (n = 40) with the first ten acquisitions

collecting two volumes each while the next two acquisitions collected ten volumes each. In a

Page 59: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

59

separate experiment, these scans were performed in a reverse order and an additional acquisition

with ten volumes was performed at the end (total n = 50) to check if the observed inconsistencies

were dependent on the order of acquisitions. The total time for each experiment was

approximately 30 minutes.

3.3.2 Field Corrected Experiments

To circumvent the sources of inconsistencies observed during pilot experiments a new

acquisition method was used where the same spoiled gradient echo sequence with ten echoes was

employed to collect 2D axial images at three different slice locations, while collecting repeated

measurements (n = 24) of each phase encoded step as opposed to the full 3D volume. This

ensured that each repetition in the data was interlaced to be closest in time to the preceding one

thereby greatly minimizing the time window for any substantial background inhomogeneity to

influence measurements. The rationale behind this approach was that injection of currents of

opposing polarity for each repeated measurement would maintain the fidelity of the computed

current-induced magnetic field maps. First, the stability of the static magnetic field over time was

measured using the acquisition method employed in pilot experiments and compared against that

specified by the manufacturer. Next, current injection experiments using the two acquisition

methods were compared.

3.3.3 Electrical Stimulation

Electrical stimulation was delivered through conductive rubber electrodes housed in

saline soaked sponges using an MRI compatible, battery driven, constant current source

(NeuroConn DCMC, Ilmenau, Germany). The stimulator was triggered using a TTL signal

generated by the MRI scanner following the installation of a custom patch and the current was

injected at 1.5 mA using pulse widths of 32 ms. The stimulator was positioned outside the

magnetic field in the operator room. Current was administered via two RF filter boxes, one

Page 60: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

60

placed closed to the filter panel connecting the scanner room with the operator room and the

other was placed inside the bore of the MRI scanner to which electrode leads were attached. A

switching circuit (developed in-house) was also used between the stimulator and the outer filter

box in order to deliver fixed-duration current pulses during the time interval following RF

excitations. The complete experimental setup is as depicted in Figure 3-4. For the experiments

using approach two, the stimulation polarity was alternated at each RF excitation with the help of

an onboard programmable microcontroller which was a part of the switching circuit. No such

alternation was performed until the completion of the scan during the pilot experiments. Anode

and cathode leads coming out of the inner filter box housed two 5 kΩ resistors to prevent

induction of high voltages from RF pulses produced during imaging. It was ensured that the

electrode cable emanating from the inner filter box ran parallel to the static field by securing it in

place using sand bags.

3.3.3.1 Switching circuit

The NeuroConn DC MC stimulator posed two major challenges. First, it was not possible

for the stimulator to be triggered by an external device (MRI scanner in this case) and hence it

was not possible to precisely control the delivery of current. Second, it did not allow for sub-

second stimulation durations which meant that the minimum TR usable in the MR acquisitions

was at least one second which drastically increased the scanning time. To address these

shortcomings, a switching circuit was developed in-house which would enabled us to capture the

5V TTL signal from the MRI scanner and gate the electrical stimulation appropriately. With the

ability to gate the stimulation, controlling the duration of stimulation meant controlling the

gating interval which was easily achieved with the use of a microcontroller. The switching

circuit also played an integral role in allowing us to control the polarity of the applied

stimulation.

Page 61: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

61

3.3.3.2 Circuit description

The circuit schematic is shown in Figure 3-4. All components are represented as boxes

with gray shadings and where possible, anode cables are shown in red and cathode cables in

blue. LCB110 and LH1540 are optocouplers that depend on the bias of an available light

emitting diode and sensor pair for break/make type contacts. In the default state, LCB110 is a

normally closed switch (NC) and LH1540 is a normally open switch (NO). LCC110 is an

integrated chip that possesses both NC and NO connections whose states can be toggled by

forward/reverse biasing its light emitting diode. The microcontroller used is a commercially

available MSP430G2553 microcontroller housed on a MSP-EXP430G2 launch pad built by

Texas Instruments. The circuit also includes a test resistance (R-test) in order to monitor the

delivered electrical stimulation on an oscilloscope (also shown in figure). Since the

MSP430G2553 microcontroller has a maximum input operating voltage of 3V, the 5V TTL

signal was reduced using a voltage dividing circuit.

Working. The constant-current stimulator was set at the appropriate current rating and

configured to deliver a continuous DC current of the desired amplitude. The working of the

circuit can be divided into three states depending on the TTL input from the magnet: no TTL,

odd numbered TTL and even numbered TTL

No TTL. This is the default or initial state of the system in which no current injection

was desired and the output of the stimulator was fed to LCB110 which operates as a “normally

closed” switch, thereby completing the current path from anode to cathode of the stimulator.

Odd numbered TTL. For all odd-numbered TTL pulses, the microcontroller was

programmed to generate a “HIGH” (3 V signal) on pin 17 in response to the received TTL signal

on pin 25. This signal generated sufficient voltage to forward bias the diode in the optocoupler

LH1540 (normally open switch) closing the switch and creating a route for the current from

Page 62: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

62

anode to the outer box. After accounting for the switching time to complete this routing, the

microcontroller generated a high signal to change the state of LCB110 to open circuit. This

ensured that the current from the stimulator now traversed along the path; anode-LCC110 (top)-

LH1540-outerbox-subject-outerbox-Rtest-LCC110 (bottom)-cathode. This status was maintained

for the duration of current injection time (Tc) following which the states of LH1540 and LCB110

were restored to their default condition. This ensured that the phantom received current injection

for the desired time Tc.

Even numbered TTL. For the even numbered TTLs, in addition to the process described

above, the microcontroller was programmed to also toggle the states of LCC110s by sending a

high signal on pin 23 that was input to the LCC110s. Toggling the states of LCC110

optocouplers altered the route of input stimulation so as to make it traverse along the path;

anode-LCC110 (bottom)-Rtest-outerbox-subject-outerbox-LH1540-LC110 (top)-anode. This

effectively switched the direction of the current injection perceived by the imaging object and

enabled us to achieve a bipolar stimulation during the imaging session.

Apart from enabling triggering, the circuit also provided the facility for polarity

switching of the stimulator output on the timescales of the repetition time of MR acquisitions.

The TTL trigger and current waveforms used for stimulation in both acquisition methods are

shown in Figure 3-6.

3.3.4. Data Processing

For the measurements obtained without current injection, in both the pilot and corrected

experiments, the acquired raw k-space multi-echo data was first read into IDL 8.1 as an 8D array

of complex floating point values. The k-space lines along the frequency encoding direction were

flipped to account for the negative readout gradient and inverse Fourier transformed to calculate

the complex MR image data. The pilot and corrected measurement data were then partitioned

Page 63: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

63

into 40 and 24 volumes respectively to obtain a time-series of multi-channel complex images.

Each of those volumes was complex divided by its preceding volume in the time series and the

multi-coil data was combined following the method described by Bernstein et al. (55). Complex

dividing the multichannel data prior to coil combination removed any fluctuations in the coil

phase and ensured that the coil sensitivity maps were not required for the reconstruction (56).

Following coil combination, the phase of the coil combined dataset was computed as the four

quadrant arc tangent for every slice location and echo collected. The magnetic field

inhomogeneity was measured from the computed phase maps at different echo times using

Equation 3-4.

3.4 Results

To address data inconsistencies, each of the 40 volumes of data collected by repeated

acquisitions without current injection was complex divided by its preceding volume to generate

39 datasets. Phase maps for these 39 datasets were computed and a region of interest (ROI) 30

pixels by 30 pixels in size was drawn in each of these datasets. Figure 3-7 shows a plot of the

mean and standard deviation of phase in the region of interest for the center slice and first and

last echo times of the original and reversed experiments. The haphazard variation observed in the

mean phase within the region of interest followed acquisition order when the experiments were

reversed, indicating that the inconsistency was directly related to the acquisition itself and not

due to heating of gradients or tissue relaxation between acquisitions. In addition, the low inter-

scan variability as evident from the first part of the plots suggested that it would be better to

acquire the desired data within a single acquisition. Phase values in the region of interest when

plotted as a function of echo number (as opposed to divided datasets) revealed a linear variation

of phase across echo times indicating the presence of a static field inhomogeneity in the

background. The precession frequency shift (γΔBo) due to the presence of the static field

Page 64: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

64

inhomogeneity was calculated as the slope of the linear fit across repetitions using the first echo

time according to Equation 3-3 to be 0.57 Hz/hr (Figure 3-8). However, in the field-corrected

experiments, the variation of the mean phase over the repetitions was quite stable (Figure 3-9)

indicating that these measurements would be less sensitive to the changing static magnetic field

and MREIT data acquired using this style would be more accurate.

Current injection scans acquired using 1.5 mA and a bipolar approach as shown in Figure

3-10 provided more stable measurements for the current-induced magnetic field. Figure 3-11

shows the consistency in the current induced field map for the center slice across three repeated

acquisitions of the 24-volume dataset.

3.5 Discussion and Conclusion

The results presented in this chapter demonstrate the influence of temporal drift in the

static field on MREIT acquisitions. MREIT or Current Density Imaging (CDI) acquisitions

typically involve the complex division of two datasets that differ only in the polarity of current

injection (37). Therefore, static field inhomogeneities are less concerning in MREIT than

temporal fluctuations. As per the specification sheet of the Philips Achieva scanner, the field

drift was approximately 1.28 Hz/hr and that measured in this work was close to 0.57 Hz/hr or

approximately 0.01 Hz/min. Magnetic fields produced due to milliampere current injections are

in the nanoTesla range which corresponds to frequency shifts of about 0.043 Hz. Therefore, for

the successful characterization of the current-induced fields, it is necessary that the MREIT data

acquired for complex division is collected over very short timescales. In the work presented here,

this effect is well established as the temporal changes in the static field have been shown to be

strong enough to mask the effects of injected currents.

It has been shown that the noise in the current-induced magnetic field is inversely

proportional to the current injection time and SNR of the magnitude MR image (57). Repeated

Page 65: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

65

measurements, if stable, can be averaged for each acquisition, thereby improving the SNR and

hence the reliability of detecting minute fields. All data acquisitions were performed after

allowing for at least two hours of rest for the MRI scanner. This duration was hypothesized to be

a sufficient time for any field inhomogeneities from eddy currents generated during prior use to

subside. Observed results support this hypothesis and indicate that the major contributor which

masks the current-induced magnetic field is the temporal drift of the scanner. Another approach

to mitigate the effect of temporal inhomogeneity would be to use the field fluctuation

information from a distant region that is known to be unaffected by the applied current (58). The

best approach to compensate for the field variation would be to quickly characterize the Bo map

in between the MREIT acquisitions. This was however not feasible due to limitations on

scanning time. By acquiring the two datasets that need to be complex divided extremely close in

time (averaging over each individual phase encoding step), even though the static field varies to

a large degree over the length of the full k-space acquisition, we believe that this effect is

diffused similarly in both the datasets and is cancelled more efficiently. This is apparent from

Figure 3-9 which shows considerable stability to the temporal variations in the static field.

Page 66: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

66

Figure 3-1. Gel phantom used in MR experiments. A) Freshly prepared agarose gel phantom

after removal from refrigerator. B) Experimental setup showing the electrodes

secured on the gel with a rubber strap. C) Schematic showing the placement of the gel

phantom in the MR scanner.

A B C

Page 67: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

67

Figure 3-2. Spoiled multi gradient echo sequence (2D) used for data acquisition. The repetition

time (TR) was 50 ms. Current pulses Ic = ±1.5 mA in amplitude were injected for a

duration of 32 ms. The bandwidth of acquisition was 550 Hz/pixel while 100 samples

were collected during the frequency encoding. The echo spacing was 3 ms.

(Method 1): 1.5 mA was injected till the completion of all frequency and phase encoding steps

for all slice locations to generate a 3D volume. This 3D volume was repeated for the

required number of averages before -1.5 mA was injected to collect data in a similar

fashion. The first echo time (TE1) was 6 ms.

(Method 2) The required averages were collected for each phase encoding step with +1.5 mA

and -1.5 mA alternatingly injected for each average. A delay time (Td) of 4 ms was

used to allow polarity switching. The phase encoding step size was incremented after

collecting all averages until a full slice acquisition was completed. The first out of ten

echo times (TE1) was 7 ms.

Page 68: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

68

Figure 3-3. Layout of the experimental setup used for imaging. The current injected was

carefully monitored using the oscilloscope to ensure the synchronization of RF trigger

and injection pulses as well as the injection amplitude. The components of the

stimulator commercially available are shown in orange while the in house

components added to the circuit are shown in gray.

Page 69: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

69

Figure 3-4. Schematic of the switchbox circuit and layout of the different components used in coalition for data acquisition. R-test is

the test resistance of 1 KΩ across which the stimulation is checked using an oscilloscope.

Page 70: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

70

Figure 3-5. Preliminary experimental results from two identical acquisitions performed with 1

mA current injection through a wire placed in a gel phantom. A) Images acquired

during acquisition 1. B) Images acquired using identical acquisition parameters as in

A, showing data inconsistency due to magnetic field instability.

Figure 3-6. Oscilloscope recording during pilot experiments for 1 mA current injection. Each

division on the X-axis corresponds to 50 ms. The TTL signal (green trace) generated

after every RF excitation (see Figure 3-2) during imaging and corresponding

stimulation voltage (blue trace) recorded across the test resistor due to injected

current (Ic+ in Figure 3-2) are shown.

A

B

Page 71: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

71

Figure 3-7. Phase variation in a 30x30 pixel ROI across each of the complex divided datasets. A)

The first 19 datasets corresponding to data acquired as two volumes per acquisition,

and the next 20 to data collected as 10 volumes per acquisition. B) Data from similar

experiments as in A, acquired in reverse order along with an additional 9 datasets

from the repeated 10 volume acquisition.

A

B

Page 72: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

72

Figure 3-8. Regression fit for the phase values within a 30 x 30 pixel ROI drawn in the center of

the image for the first echo time across the ten volumes collected as part of the first

acquisition in the reversed experiments. Acquisition time for 1 volume was

approximately 52 seconds. The measured frequency drift from the regression fit was

0.57 Hz/hr.

Page 73: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

73

Figure 3-9. Variation of the mean phase in a 20 x 20 pixel ROI drawn in the center of the image.

Each of the twenty-three plots represents the phase obtained by complex dividing data

of every repetition by its preceding one. The X-axis in each plot represents the echo

number. A total of 24 repetitions/averages were acquired during the field corrected

experiments.

Figure 3-10. Oscilloscope recording of the voltage across the test resistor (see Figure 3-4) due to

1.5 mA current injection (blue) alternating polarity with every incoming TTL signal

(green).

Page 74: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

74

Figure 3-11. Magnetic field maps for the first slice location obtained from field corrected experiments by acquiring the positive and

negative current injection data very close in time. Identical maps across multiple current injection runs show measurement

reliability and insensitivity to temporal drift of the static magnetic field.

Page 75: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

75

CHAPTER 4

MEASUREMENT OF ELECTRICAL CURRENT DENSITY DUE TO TRANSCRANIAL

ALTERNATING CURRENT STIMULATION OF THE HUMAN BRAIN AT 3 T

4.1 Introduction

The effect of electrical stimulation of the human brain has been a hot topic of research for

quite some time and there have been numerous efforts in understanding the spatial distribution

and temporal variation of the applied electric field. Given that the brain is a complex and

intricate network of neurons, neuronal activation thresholds and integration characteristics of

neurons to enable firing also play a major role in the observed effects of electrical stimulation.

Knowledge of the applied electric field provides insights into the different regions affected due

to stimulation, the extent to which they are influenced spatially and how strongly the affected

regions are modulated by it. The applied electric field causes free ions in tissue to move,

constituting a current density which is defined as the charge flowing through a cross sectional

area per second and is related to the applied field via Ohm’s law EJ .

Electrical conductivity or σ is a property of the underlying tissue and is defined as the

amount of resistance faced by the moving ions under the influence of applied field. Under the

assumptions that there are no sources or sinks present in the volume of tissue, total current

entering the volume should be equal to the net current exiting the volume, i.e. J 0 . The

presence of free and bound charges in the tissue gives rise to conduction and displacement

currents that can be described by the conductivity and permittivity of brain tissue. In the range of

excitation frequencies spanning 0 – 10 kHz, the conductivity of brain tissue is largely determined

by the extracellular conductivity. A comprehensive review of the dielectric properties of

biological tissue was given by Gabriel et al. (59-61).The human head comprises of different

tissue types (fat, skin, muscle, white matter and gray matter etc.) and transcranial stimulation

Page 76: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

76

applied using surface electrodes gives rise to different current densities in these tissues. For

instance, muscle tissue comprises of a collection of muscle spindles or fibers that have varying

conductivities in parallel (0.624 S/m) and perpendicular directions (0.131 S/m) (60,62,63).

Therefore muscles under the electrode area will strongly shunt the electrical stimulation

decreasing the amount of current entering the brain. In addition to the muscle, fat, which has a

low conductivity (0.025 S/m), will also decrease the applied electric field in the brain. Skin is a

multilayered tissue where the outer layers are dry and resistive (2x10-5S/m) while the deeper

layers are conductive (0.3 S/m) (64). The skull is probably the greatest barrier that hinders the

applied stimulation from reaching the brain. It comprises of three different layers of bone tissue,

a spongy cancellous layer enclosed between two hard cortical layers. Moreover the thickness of

the skull varies in different parts of the head giving rise to inhomogeneous conductivity. The

skull also comprises of immovable joints called sutures that provide low resistance paths for

applied current. Akhtari et al. measured the conductivity of the three skull layers in preserved

(65) and live tissue (66) and concluded that live tissue was more conductive by at least a factor

of 1.5. Baumann et al. measured the electrical conductivity of cerebrospinal fluid at room

temperature (25° C, 1.45 S/m) and body temperature (37° C, 1.79 S/m). They reported that these

values differed by approximately 25% and that these values were constant over the range of

frequencies from 10 Hz – 10 kHz. Brain tissue, categorized as gray and white matter, is

heterogeneous and importantly anisotropic like the skull. The gray matter can be assumed to be

isotropic and homogeneous with an average conductivity value of approximately 0.4 S/m (67).

Nicholson (68) measured the impedance of white matter in cats’ internal capsules, and reported

that the conductivity along the length of the track was approximately 10 times larger than that

perpendicular to the track. When compared to gray matter, the conductivity of white matter is

Page 77: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

77

greater along the length of the fiber (1.13 S/m) but the transverse conductivity is lower.(0.13

S/m). In addition to anisotropy, the heterogeneity of tissue has a significant effect on the induced

electric field distribution. Since the electric field across different tissues is constant but their

conductivities vary, charge accumulation occurs at the interface of different tissue types that

generates a secondary field which opposes the flow of charge in the high conductivity region. As

a result, the current density normal to the interface in the high conductivity region decreases such

that the current density on either side of the interface becomes the same. At this juncture, the

amount of charge entering and leaving the interface is the same. This process in biological tissue

occurs in the time span of microseconds and can be considered instantaneous, thus validating

assumptions that the tissue is purely resistive. Miranda et al. (69) showed that the secondary field

generated at these interfaces is a significant fraction of the applied field and cannot be neglected.

Plonsey and Heppner (70) presented several approximations to compute the electric field

distribution at low frequency (DC – 10 kHz). They neglected propagation effects, i.e. the electric

field in the brain was assumed to vary without any significant phase differences or delays in

response to the stimulation and the tissue was considered to be purely resistive. Under these

assumptions, the spatial and temporal variations of induced electric field can be represented by

separate functions with the temporal variation characterized by injected current from the constant

current source.

Mapping the spatial distribution of the electric field or current density however is much

more challenging. Using MRI, Scott and colleagues (36), developed current density imaging, a

methodology to reconstruct injected current density by extracting the three components of the

magnetic flux density vector following rotations of sample in the MR scanner. Assuming sample

conductivity to be isotropic, Seo et al. (39) extended the idea of current density imaging by

Page 78: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

78

reconstructing maps of injected current using only one component of the magnetic flux density

vector eliminating the need for subject rotation. Kwon et al. (71) extracted tissue properties by

exploiting the relation between conductivity and magnetic field. Recently, Jog et al. (72)

quantified the magnetic field induced due to injected currents in the human brain by tracking

changes in phase of the MRI signal in the presence and absence of current injection and

analyzing them using a generalized linear model.

The work detailed in this chapter is based on the experimental design from chapter 3 and

extends the experiments from phantoms to healthy volunteers. Current-induced magnetic fields

from the experiments are qualitatively compared against those simulated from realistic head

models. In addition the inverse problem of reconstructing electrical current density distribution

from measured magnetic fields in the human brain is solved using the projected current density

approach introduced by Park et al.(44). Aprinda Indahlastari performed the image segmentation

and generation of computational models against which the experimental data compared. Munish

Chauhan reconstructed the projected current density maps. My role was to recruit the

participants, handle the electrical stimulation, perform MR imaging and develop the necessary

data processing software for analysis of the acquired data.

4.2 Methods

4.2.1 Phantom Experiments

A homogeneous agar gel phantom of uniform conductivity was prepared following the

protocol outlined in subsection 3.3.1.1. Electrode cables with carbon rubber electrodes housed in

0.9 % saline soaked sponges were attached diametrically opposite to each other. The entire setup

was then placed in the bore of a 3 T Philips Achieva magnet and MREIT data with 1.5 mA

current injection was collected following the installation of a custom-built patch that enabled

triggering of the constant current source.

Page 79: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

79

4.2.2 Human Experiments

Healthy volunteers (n = 6) between the ages 18-30 years, without any history of mental

health issues, predominantly right handed, had no history of prior seizures and did not have any

metallic implants that could interfere with the MR imaging, were recruited for participation after

they were duly consented. Handedness was determined by the Edinburgh handedness inventory

test. A Mini Mental examination was performed to assess the IQ of participants. The volunteers

also completed an MRI screening form to ascertain their compliance with the MR imaging

procedure and answered a questionnaire to assess their mood both prior to and following the

imaging session. All participants received electrical stimulation at 1.5 mA via conductive rubber

electrodes enclosed in saline soaked sponges that were placed on predetermined locations

marked on their scalp following the 10-20 system of electrode placement. Current injection

during the imaging was performed along two directions, Fpz – Oz and T7 – T8. The current

polarity was alternated with every RF excitation used in the MR acquisition which corresponded

to a frequency of stimulation of approximately 10 Hz. Subject recruitment, preparation and

imaging were performed in accordance with the experimental protocol (IRB201600695)

approved by the University of Florida Institutional Review Board.

4.2.3 MR Imaging

MREIT measurements were performed using a Philips Achieva 3 T MRI scanner. A

custom patch was installed on the scanner to provide a TTL pulse following each RF pulse

generated in the pulse sequence. A 32 channel head coil was used for signal reception. High

resolution T1 weighted 3D FLASH acquisition, at 1 mm isotropic resolution, was acquired to

facilitate image segmentation and generation of computational models. MREIT measurements

were made using a 2D spoiled gradient echo sequence (SPGR) at a resolution of 2.24 x 2.24

mm2. Three slices at a thickness of 5 mm each were acquired without any gap such that the

Page 80: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

80

imaged region contained all the four electrodes. The data was collected as multi-echo single slice

acquisitions with a repetition time of 50 ms within which 10 echoes were acquired. The first

echo time was 7 ms and the echo spacing was set to 3 ms. The first echo time was set so as to

allow the switching of current polarity, as needed, following every RF excitation. The k-space

data was collected so as to acquire 24 averages of the same phase encoding step, while altering

the current injection polarity for every average. This ensured that the positive and negative

current injection datasets were acquired as closely in time as possible, thereby minimizing the

effect of the static magnetic field drift. A hundred phase encoded steps were acquired so that the

scan time for each slice was two minutes. This amounted to a total scan time of six minutes per

acquisition.

For each current injection direction, two such acquisitions were performed which were

averaged in the data-processing for better signal-to-noise ratio. One acquisition with the above

parameters was performed without any current injection to serve as the control dataset. For the

experiments with healthy volunteers, the total imaging time for MREIT scans with and without

current injection was approximately 30 minutes. High angular resolution diffusion imaging

(HARDI) data with b values of 100 (in 6 directions) and 1000 (in 64 directions) was also

collected so that the principal Eigen vector of the processed diffusion tensors from this data

could be used in computational modeling.

4.2.4 MREIT Data Processing

Raw k-space data (.DATA and .LIST) files were exported from the Philips scanner and

processed using software developed in-house in IDL. The raw data was read as a

multidimensional dataset comprising of frequency, phase, slice, echoes, averages, and coils as

the dimensional parameters. First, the raw data pertaining to every other echo in the multiple

echo acquisition was flipped in k-space to account for the negative gradient encoding. Spatial

Page 81: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

81

domain data was then computed by performing the inverse Fourier transform in the frequency

and phase encoding directions. The current induced magnetic field within the sampling window

during data readout was assumed to be negligible in comparison to the magnitude of readout

gradient and therefore, the distortion caused by it was not accounted for during image

reconstruction. Positive and negative current information was encoded in the repetitions as the

polarity of current injection alternated for every repetition of the phase encoding step. To

account for this, the complex reconstructed dataset was split into two datasets comprising of 12

averages each of positive and negative current encoding followed by complex division of

positive data by the negative data. Complex division of the datasets prior to combining the

multiple coil data ensured that the individual coil sensitivities need not be taken into account (56)

and any differences in the phase across coils did not affect the measured data. Magnitude data

was then reconstructed using the sum of squares reconstruction as suggested by Roemer et al.

(73). To calculate the phase of the combined dataset, the individual coil phases were summed

assuming that noise variances in the different channels were identical. The four quadrant

arctangent of this complex divided dataset was calculated to generate phase images. Magnetic

field maps were computed by taking into account the appropriate current injection times for each

of the measured ten echoes.

4.2.5 Image Segmentation

A combination of automatic and manual processes was used to segment a single human

head into ten tissue types. The high resolution T1 dataset was imported into Freesurfer image

processing software to resample the T1 data into an isotropic resolution of 1mm3 and a matrix

size of 256 x 256 x 256. Freesurfer was also used to auto-segment the human head into white and

gray matter. SPM was used to auto-segment bone, skin and air. The remaining tissue types

comprising of cerebrospinal fluid (CSF), muscle, eyes, fat and blood, along with the attached

Page 82: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

82

electrodes, were segmented manually with Scan IP, Simpleware Ltd, using an atlas of the human

brain as reference. All automatic segmentation outputs were manually corrected in ScanIP

according to the human atlas.

4.2.6 Computational Modeling

The 3D segmented head model derived from high resolution T1 MR images was meshed

in ScanFE, Simpleware Ltd. and exported into the modeling software COMSOL 5.0. Literature

referenced conductivity values for the different tissue types were then assigned to the segmented

3D head model (52) prior to simulation. Finite element calculations, to solve the Laplace

equation with mixed boundary values at the surface of the head model was performed via

COMSOL-MATLAB Livelink Interface. Current density results from finite element simulation

were interpolated into a fine stencil (1001x1001x91, 0.5mm resolution) and were used to

calculate Bz based on the Biot-Savart equation. The Jx and Jy data were used to compute the z-

component of magnetic field having the same dimensions such that,

x y

z 3

y y J r x x J rB r dr

4 r r

(4-1)

Finally, the simulated magnetic field was averaged over each voxel volume and resampled into

2.24 x 2.24 x 5 mm3 resolution to register them to the experimentally measured field maps.

Corresponding slices in the final registered simulated maps were extracted and compared against

the measured magnetic field values obtained from experiments.

4.3 Results and Discussion

4.3.1 Gel Phantom

4.3.1.1 Field measurements

Multichannel MR magnitude images of the gel phantom for the first echo and center slice

location obtained from the inverse Fourier transformation of the complex k-space raw data are

Page 83: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

83

shown in Figure 4-1. The reconstructed magnitude and phase images following the sum of

squares reconstruction are shown in Figure 4-3. shows the binary mask generated from the high

resolution T1 data and the different slices for the first echo in the multiple echo dataset. Also

shown in are the different echoes for the center slice location. Figure 4-4 and Figure 4-5 show

phase and magnetic field maps generated from the complex division of data acquired with and

without current injection. The increased sensitivity to induced phase and magnetic field, across

echoes, due to increased current injection duration is clear from these images. Patterns of the

magnetic field observed with a darkish hue on one side and a brighter pattern on the other, are

consistent with those reported in literature. This is to be expected because when the current

travels from left to right (or right to left) the magnetic field perpendicular to the slice adds or

subtracts from the static field thus producing the dark and light patterns.

4.3.1.2 Optimization of magnetic field maps

The multiple echo dataset acquired without any current injection was used to generate T2*

maps of the different slices. These maps are shown in Figure 4-6 along with their corresponding

magnitude slices. Each of the magnetic field maps for every echo was multiplied by a weighting

factor (48) that took into account the signal decay due to relaxation and field inhomogeneities.

Moreover, the weighting function was also adapted to take into account the increase in current

induced field which is dependent on the current injection time Tc. Optimized field maps for the

three different slices are shown in Figure 4-7. Figure 4-7 also shows the comparison of measured

magnetic field maps against numerical simulations from a computational model. The

experimental and simulated magnetic field maps show similar range of variation but the

experimental value seemed to be affected by a global offset of 1nT. It is possible that this offset

existed in MREIT experiments by other research groups mentioned in the literature (41,49) but

was not appreciable given the high amplitudes of current used in those studies. Possible sources

Page 84: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

84

of this offset are the surface currents on electrodes pads or magnetic fields due to currents in the

lead wires.

4.3.2 Healthy Volunteers

The data from two participants (MREIT-005 and MREIT-006) who had electrodes placed

at T3, T4, Fpz, Oz locations showed large noise levels in the computed magnetic field maps and

were hence not further analyzed. This variability was believed to be due to poor electrode

contact with the scalp owing to much denser hair in these participants. For clinical studies, it

would be beneficial to shave the hair of patients at the electrode locations in order to ensure good

contact. This could eliminate confounds/variability due to improper electrode contact area.

Figure 4-8 shows the electrode locations on a healthy volunteer along with the stimulation

waveform used. The frequency of stimulation as revealed by the Fourier transform was

approximately 10 Hz. Experimentally measured current-induced phase changes and magnetic

field maps at the central slice location for four different volunteers along T3 – T4 direction is

shown in Figure 4-9. Phase and field maps due to current injection along the Fpz-Oz direction are

shown in Figure 4-10. Optimized magnetic field maps were computed using T2* maps similar to

phantom experiments. A realistic head model along with electrode locations generated by

segmenting the high resolution T1 dataset into multiple regions is shown in Figure 4-11. The

head model comprised of ten different tissue types which are GM, WM, CSF, skin, air, bone,

muscle, fat eyes and blood. The current-induced magnetic fields calculated from numerical

simulations performed on the generated head model compared against the optimized magnetic

field maps are also shown in Figure 4-11. The simulated magnetic field values for T3-T4

stimulation varied from -1 nT to 1.5 nT while the experimental values varied from -1 nT to 3 nT.

For the current injection along Fpz-Oz the simulations predicted field changed from -2 nT to 2

Page 85: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

85

nT while the experiment recorded -0.5 nT to 0.5 nT. Qualitative comparisons of the experimental

and simulated magnetic field maps in the two current injection directions for four volunteers are

shown in Figure 4-12 and Figure 4-13 respectively.

The experimental and simulated field maps were in good agreement for subjects 009 and

012, but not so much for subjects 010 and 011. This can be attributed to the subjects’ experience

during electrical stimulation especially along the Fpz-Oz direction. Subjects 010 and 011, when

questioned, reported discomfort during Fpz-Oz stimulation which was reflected in their ratings

of stimulation perception. Retinal phosphenes, or flashes of light in the field of vision, consistent

with those reported in tACS literature were observed in all participants in addition to a

tingling/pricking sensation under the electrodes. Subject 009 however did not report strong

perception for current injection in either direction but 010, 011 and 012 reported feeling the

phosphenes strongly. Subject 010 and 011 claimed that they moved their head during image

acquisition to reduce discomfort from the electrical stimulation. This would have generated

motion artifacts that can easily compromise the fidelity of the measured MR signal phase. The

improvement in the comparisons for subject 012 is believed to be due to strongly persuading the

participant not to move during the imaging and rigorously securing that participant in the head

coil. In addition to subject movement, cardiac pulsing and changes in size of the thoracic cavity

due to respiration are known to affect the precession frequency of spins due to changes in

susceptibility.

Since the majority of the MR signal resides in the center of k-space, the influence of

respiratory and cardiac fluctuations can be understood by analyzing the variation of phase of the

k-space central point over time. Fourier analysis was performed to extract the different

frequencies present in the time series of k-space center following twenty-four repetitions and

Page 86: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

86

contained a large peak only at the zero frequency. This revealed that the physiological noise due

to cardiac fluctuations usually observed in fMRI acquisitions did not influence the MREIT

measurement process. However, typical frequency of respiration, which is around 0.2 – 0.5 Hz,

was smaller than the best frequency resolution possible with the employed acquisition technique

and therefore the effect of respiration could not be quantitatively addressed.

4.3.3 Quality of field maps

As reviewed in Section 2.4, signal-to-noise ratio is critical in MREIT acquisitions for the

reliable measurement of current induced magnetic fields. Twenty-four repetitions of each phase

encoding step of the MREIT scan were acquired to improve signal-to-noise ratio. SNR is

typically calculated using a two-region approach and is evaluated using the statistics of the signal

and noise in these regions. This approach is however based on two conditions. First, the noise is

spatially homogeneous over the entire image and second, the distribution of noise is known so

that statistics measured from a background region can be used to calculate its behavior inside the

imaging object. The noise distribution in magnitude images for single channel receivers was

shown to follow a Rician distribution (Section 2.4). Constantinides et al. (74) showed that for

phased array coils that use ‘n’ receiver channels, noise is no longer Rician distributed but follows

a noncentral chi distribution inside the imaging object. In regions without any signal

(background) the noise statistics change and can be described by a central chi distribution.

Dietrich and colleagues (75) compared four different SNR measurement methods and showed

that SNR calculated from two identical acquisitions provided the best results. In this approach

SNR was calculated as the ratio of mean and standard deviation in a region containing the object,

where the mean was computed in a combined map of two identical acquisitions and the standard

deviation was measured in a difference map of the two acquisitions. The inverse relation

between the SNR in magnitude images and the noise in phase images is still valid and a

Page 87: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

87

theoretical estimate for the noise in phase maps obtained during current injection can be

calculated. Magnitude SNR for the phantom and human datasets was measured using the

difference method (75) for datasets with and without current injection as shown in Table 4.1.

Sadleir et al. (57) developed an analytical approach to test the fidelity of MREIT

experiments using a method that relates the Laplacian of current induced magnetic field to the

Laplacian of noise for homogeneous phantoms. The noise standard deviation of the magnetic

field map can be calculated as,

2

c

c

B

B

4 4

s

20 6

(4-2)

In Equation 4-2, Bc refers to the current induced magnetic field without noise and Bc is the

current-induced magnetic field in the presence of noise. Δ represents the pixel width along the

frequency and phase directions (assuming they are equal) and Δs refers to the slice thickness.

Theoretical and experimental values of the noise standard deviations of the averaged current

induced field for the first and last echo were 1.07 nT, 0.7 nT and 0.16 nT, 0.15 nT respectively.

4.3.4 Projected current density maps

The projected current density map computed as gradient of the magnetic field map

following the description in subsection 2.4.5 for the phantom experiment is shown in Figure 4-

14. Since the electrodes used for injecting current spanned the entire field of view used for

MREIT current injections, the projection of true current density, Jp, was expected to match the

true current density J in the imaging field of view. Moreover, as the current density was

calculated using a Laplacian operation on the measured magnetic field, any global offset

affecting the measured magnetic field did not cause problems in the estimation of current

density. This is evident from the similarity between current patterns in gel phantom despite the

Page 88: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

88

1nT offset that was observed when comparing numerical simulations to measured magnetic

fields. Figure 4-15 shows the projected current densities for healthy volunteers. This figure also

shows an overlay of current densities on T1 images to help understand/visualize the regions

affected by the applied current.

Since the electric and magnetic fields corresponding to the applied current are always

orthogonal to each other, judging from the measured magnetic field maps, the projected current

density maps seem to provide reasonable estimates of current density distributions. In all four

volunteers the results are consistent with the hypothesis that the largest current densities in the

brain are obtained closest to the electrodes (76). Sadleir et al. (63) reported current densities as

high as 0.8 A/m2 in a realistic head model when using injection current densities of 0.45 A/m2.

Reconstructed current densities in the work presented here also vary within a comparable range

(0-0.7 A/m2). Recently, some researchers performed invasive current density measurements on

cadavers and reported that the majority of current (90%) during tACS was shunted through the

scalp. However, the conductivity of living tissue is higher than that of dead tissue and the work

detailed here shows that a significant amount of applied current distributes within the brain.

Nevertheless the current density was still small and incapable of eliciting action potentials. The

apparent variation in current density across the volunteers could be due to the difference in

electrode placement or leakage of saline from the sponges increasing the effective area of the

electrode. Gel or conductive paste as the interface medium between the electrode and tissue has

been widely used in the literature but the work presented here has not yet taken into account the

effect of electrolytic medium. It is possible that gel electrodes could present a better alternative

for saline soaked sponges for localizing the area of contact for the electrodes during MRI.

Page 89: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

89

Bone has a very short T2 relaxation time and the signal from bone tissue is completely

lost for conventional echo times used in MR imaging. Therefore, extracting the distribution

within the bone tissue presents a great challenge and novel pulse sequences like ultrashort echo

time, UTE (77), could be tailored for use with MREIT to enable extracting the current density

information in bone tissue. The work presented in this chapter utilizes bipolar current injection

pulses that correspond to roughly 10 Hz stimulation. However extrapolating this work to

mapping current density during tDCS will require careful selection of stimulation parameters and

experimental design. Jog et al. (72) mapped magnetic fields due to varying amplitudes of tDCS

stimulation by injecting current while simultaneously mapping the static magnetic field Bo.

However, the interaction of current-induced magnetic fields with RF fields used for slice

selection in MR imaging was not addressed which is crucial for high resolution imaging.

An interesting observation in the projected current density maps was the high current

density localization in the ventricles during Fpz - Oz stimulation for subjects 009 and 011. This

could be explained by the high conductivity of CSF in comparison with other tissues. However

this was not immediately appreciable in the density maps due to T3 – T4 stimulation indicating a

complex interplay between the current injection direction and tissue conductivity. In addition,

large current densities were observed for subject 009 in the posterior part of the brain between

the two hemispheres. Whether this was the result of magnetic field measurements close to the

noise floor or due to tissue interfaces dictating current density localization remains to be

determined.

4.4 Conclusions

In this chapter the methodology for mapping current density distribution in the in vivo

human brain following a tACS like stimulation was presented. While researchers in the past have

measured current density in inanimate objects, and more recently magnetic fields due to in vivo

Page 90: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

90

tDCS in the human brain was performed by (72), the results in this chapter are the first

measurements of current density in the human brain performed in vivo. Projected current density

was computed for a two dimensional dominant current distribution in an agarose gel phantom

and healthy volunteers. The projected current density maps calculated from measured magnetic

fields were in good agreement with current densities simulated using computational models for

the phantom data. Though the modeling enlisted heterogeneous conductivity values presented in

literature for different regions of the head, the anisotropy of certain tissue types (muscle, white

matter etc) was not taken into consideration. Improving the computational models by

incorporating diffusion tensor information to account for the underlying anisotropies would

result in more precise comparisons of the measured and simulated current densities. The

variation of measured current distributions within subjects underlines the importance of carefully

controlling various experimental parameters like electrode location, contact area and subject

motion.

Page 91: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

91

Figure 4-1. MR magnitude images from all 32 channels reconstructed by carrying out the inverse Fourier transform of complex k-

space data. The coil combined magnitude image can be generated using a sum of squares approach with or without

appropriate weighting.

Page 92: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

92

Figure 4-2. Representative images showing electrode placement, image reconstruction and

decrease in SNR with increasing echo times. A) Binary mask generated from the high

resolution T1 data showing the location of electrodes. B) Sum of squares

reconstructed magnitude MR images for each of the three slice locations. C)

Magnitude MR images for the center slice showing all echoes in the acquisition. Note

the decrease in SNR in the images going from left to right.

Figure 4-3. MR magnitude and phase images for the center slice location collected at the first

echo time. The increased sensitivity at the bottom of the magnitude image was due to

the proximity of the gel phantom to the bottom set of RF coils.

A

C

B

Page 93: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

93

Figure 4-4. Phase maps produced from complex division of acquired data scaled from -0.1

radians to 0.1 radians. A) Phase maps when no current was injected into the phantom.

Figures B, C, and D show the phase maps computed when 1.5 mA current with

alternating polarity was injected into the phantom. Note the consistency of results

across B, C and D.

Figure 4-5. Magnetic field maps computed from the measured phase changes (Figure 4-4) scaled

from -10 nT to 10 nT. A) Magnetic field maps without any current injection. B, C and

D show magnetic field maps computed when 1.5 mA current was injected with

alternating polarity. The sensitivity to current induced magnetic field is higher for the

later echoes due to greater current injection duration (Tc).

Page 94: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

94

Figure 4-6. MR magnitude data for all slices and their corresponding T2* maps computed from

data acquired without any current injection. The T2* maps are displayed in seconds

giving an average T2* of about 48 ms in the gel phantom. Volume fitting for all the

voxels was performed using the Levenberg Marquardt algorithm.

Page 95: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

95

Figure 4-7. Comparison of measured and simulated current-induced magnetic field maps in the

gel phantom. A) Magnetic field images generated following optimal combination of

the multi-echo field maps for all slice locations. B) Comparison of measured and

simulated magnetic fields, in the central slice, along the lines drawn at the same

regions in both images. The field values are shown in Tesla.

Page 96: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

96

Figure 4-8. Experimental setup and stimulation paradigm for transcranial electrical stimulation

on healthy volunteers. A) Photographs depicting the electrode placement at T3, T4,

Fpz, Oz locations following the 10-20 EEG system. B) Oscilloscope recording

showing the trigger pulse (green) and the stimulation waveform (blue). C) Fourier

transform of the waveform shown in B. Note that maximum power in the frequency

spectrum corresponded with ±10 Hz stimulation.

A

B

C

Page 97: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

97

Figure 4-9. Current-induced phase and magnetic field images for healthy volunteers. Note the

steady increase in phase across echoes. Electrical stimulation was performed from

right to left in the image. The static field direction is into the page.

Page 98: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

98

Figure 4-10. Current-induced phase and magnetic field maps for the Fpz-Oz direction. The phase

maps are scaled from -0.1 to 0.1 radians while the magnetic field maps cover a range

of -10nT to 10 nT.

Page 99: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

99

Figure 4-11. Representative images of segmented T1 images, computational model generated and comparison of simulated and

measured magnetic fields. A) Computational model generated from manually segmented tissue types obtained from the

high resolution T1 data. B) T1 data resampled to facilitate easy segmentation and segmented T1 images are shown. C)

Experimental and simulated magnetic field maps scaled from -5nT to 5 nT and profile plots along the black lines for both

the current injection directions. (Model and segmentation images generated by Aprinda Indahlastari.)

A

B

C

Page 100: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

100

Figure 4-12. Comparison of measured and simulated magnetic fields due to T3-T4 stimulation

for the central slice location. Good agreement can be observed in subjects 009 and

012. Results from subjects 010 and 011could have contributions from motion.

Magnetic field maps are scaled from -10 nT to 10 nT.

Page 101: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

101

Figure 4-13. Comparison of measured and simulated magnetic fields, following electrical

stimulation with electrodes placed at Fpz and Oz, for the central slice location

Page 102: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

102

Table 4-1. SNR values for volunteer MRI data and noise in measured magnetic field maps at 3T.

SNR in magnitude data Theoretical σBc Measured σBc

No Current T3-T4 Fpz-Oz T3-T4 Fpz-Oz T3-T4

Gel phantom 454 503 - 0.15 nT - 0.18 nT

MREIT 009 348 330 395 0.26 nT 0.22 nT -

MREIT 010 432 406 307 0.21 nT 0.29 nT -

MREIT 011 380 388 436 0.22 nT 0.20 nT -

MREIT 012 521 525 495 0.17 nT 0.18 nT -

Figure 4-14. Comparison of simulated current density images and current density maps

computed from measured magnetic field maps. Shown are the maps for X and Y

components of the current density vector and the corresponding norm calculated from

the two components. The units of scaling are in ampere per square meter or A/m2.

Page 103: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

103

Figure 4-15. Current density (norm) images for healthy volunteers shown in relation to their high

resolution T1 images. The current density maps were scaled from 0 – 0.7 A/m2. Also

shown are the current density maps overlaid onto the T1 images with 50%

transparency. The highest current densities were observed underneath the electrode

contact regions (Figures generated by Munish Chauhan).

Page 104: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

104

CHAPTER 5

MEASUREMENT OF MAGNETIC FIELDS DUE TO LOW AMPLITUDE INJECTION

CURRENTS USING IMPLANTED ELECTRODES

5.1 Introduction

This chapter details with the work involved in the measurement of magnetic fields due to

very low amplitude injection currents using implanted electrodes. The measurements were

performed using a 4.7 T Agilent magnet system housed in the McKnight Brain Institute,

University of Florida. Magnetic field measurements due to varying levels of current injection in

gel phantoms, using spin echo techniques are presented. Data acquisition using copper wires and

fabricated carbon electrodes are presented. Effects of large current densities in the electrodes on

measured field distributions inside the gel phantom are discussed. Extended studies using current

injections with three different electrode orientations are discussed.

5.2 Theory

Current injection due to implanted electrodes must typically be maintained under a

certain threshold in order to avoid damage to the local tissue as a result of Joule heating. More

importantly, the implanted electrode should not react with the surrounding tissue and should

possess a magnetic susceptibility that is close to water in order to avoid artifacts in MR imaging

(since brain tissue is predominantly composed of water). The magnetic field due to current

passing through an infinitely long wire at a distance ‘a’ away from the wire can be written using

Ampere’s law as,

ocB

ar

I

2

(5-1)

For wires with finite length, the magnetic field associated due to current in the wires is

given by the Biot-Savart equation.

Page 105: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

105

or r

B r J r drc 34 r r

(5-2)

Stimulation using implanted electrodes results in current travelling through a well-

defined path (through the electrode) and then into the tissue while stimulation using surface

electrodes results in a more distributed current. The result is that in the electrodes the current

density is large, resulting in a large magnetic field associated with the current injection.

Depending on the orientation of the electrode during stimulation, this large magnetic field can

mask the distribution of magnetic field due to the current flow in the tissue. The effect can be

pronounced depending on the spacing between the electrodes which can be tailored in order to

decrease this effect. As the current-induced magnetic field is measured as a change in the main

magnetic field of the MRI scanner, interaction of the current induced field with the static field

can be controlled by varying the orientation of electrodes with respect to each other and the static

magnetic field so as to minimize the effect of field due to current inside the electrodes.

The magnetic field changes in the sample due to injected currents can be quantified with

a conventional spin echo sequence imaging sequence as described by Scott et al.(36). Recall

from Section 2.4 that as the current induced magnetic field imparts an additional phase to the

spins without altering the encoding generated by the imaging gradients, the resulting datasets can

be complex divided to extract the field due to current injection. A more sensitive approach to

measuring the induced magnetic fields was developed by Kwon et al., termed as the injection

current non-linear encoding method (78). The “non-linear” in the name stems from injection of

current during the data acquisition period thereby changing the one-one mapping of spin

precession frequency with position. To understand the ICNE approach, a detailed understanding

Page 106: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

106

of the effect of background gradients on frequency and phase encoding of precessing spins is

necessary.

5.2.1 Effect Of Background Magnetic Field On Frequency Encoding

The effect of a background magnetic field can be assumed to be caused due to a

stationary gradient applied in conjunction with the frequency encoding gradient. For simplicity,

let us consider that the inhomogeneity exists only along the frequency encoding direction.

Without loss of generality we can assume that the frequency encoding is performed along x and

the phase encoding is performed in the y direction. Then the precession frequency of spins in the

rotating frame can be described by the equation

x= G .x (5-3)

In the presence of the magnetic field inhomogeneity Equation 5-3 can be written as,

x x= G .x G .x (5-4)

The phase of precessing spins can be given by the relation 2 k.rt . Using the wave

vector t

0

k t2

G dt

to describe the spatial frequencies in the image, the measured signal at a

particular frequency encoding step, kx, in the presence of the background field can be represented

by,

x

xx

Gk xi2

x o

1G

S k S e

(5-5)

This causes a local distortion in the image since the spins that are at position ‘x’ are now

misappropriated to be present at x′ due to the additional gradient present during the sampling

interval. As a result the original field of view, L, given by x

1 2

k G tL

changes to L′ such

Page 107: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

107

that the local effective field of view is now scaled by inverse ratio of the term involving

gradients in Equation 5-5. The scaling factor can be expressed as,

x

x x

G

G G

(5-6)

Depending on the magnitude and sign of the background field gradient, signal enhancement

could occur as a result of the mispositioning of spins upon image reconstruction. This effect is

particularly strong in gradient echo imaging where the background gradient is not refocused at

echo time. In the case of spin echo sequences, the additional 180° RF pulse ensures that the

background gradient is refocused at the gradient echo time.

Another important effect that occurs due to the presence of background inhomogeneity is

a shift in position of the echo during readout. This effectively causes the echo to no longer be

centered in the sampling window and depending on the strength of the background gradient, the

echo could be completely pushed out of the readout interval leading to a complete loss of

measured signal. If the background field is not sufficient to move the echo out of the sampling

interval, the time shift in the echo upon image reconstruction can be observed as a shift in the

image phase. For a fixed echo time, the time shift due to the background gradient creates phase

dispersion over the length of the imaging voxel leading to an increased transverse relaxation rate

that directly translates to a decrease in signal. The phase dispersion produced across the voxel as

a result of echo shift in k-space can be written as,

s

2 t 2 t

T N t

(5-7)

where δt corresponds to time shift of the echo and Ts (sampling time) is the total time required to

collect N data points along the frequency encoding direction evenly spaced at Δt interval. Since

the readout is sampled as a discrete set of points, the background field induced shift of the echo

Page 108: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

108

can be represented by the number of points by which the echo shifted. This can be written

mathematically as,

x

x

G Tn

G t

E

G

(5-8)

The important thing to note here is the shift in the echo occurs to the right resulting in an

effective echo time TE′ where TE < TE′ when the background gradient has a negative sign with

respect to the readout gradient. The reverse is true when the polarity of the background gradient

is positive with respect to the readout gradient. In low amplitude MREIT current injections, the

magnetic field produced by the current (nT – μT regime) is orders of magnitude smaller than the

frequency encoding gradient (mT/m) and therefore the image distortion and phase dispersion are

minimal and can be ignored during the reconstruction process.

5.2.2 MREIT-ICNE

The current-induced magnetic field that exists as a background gradient during the image

acquisition, if extracted, can provide additional sensitivity to the MREIT experiment enabling the

decrease in the injected current amplitude that is crucial when delivering current directly to the

brain tissue. This is achieved using the ICNE method and the formulation is as described below.

Since the one-one linear relationship between the precession frequency of spins and their

position is no longer preserved, a linear operation like the Fourier transform can no longer be

applied for image reconstruction. The k-space signal from the ICNE approach where the current

is applied during the sampling interval (Ts) can be described by,

csx yc c

x

B x,yT i2 k x k yi T B x,yGi x,y 2

x y x, y e e e dS k xdy,k

(5-9)

Page 109: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

109

where c c sT T T .Taking the difference of the positive and negative k-space signals, defining

s xc

x

T kT 2

2 G

and finally dividing by the term 2iα, we can obtain,

x yx y x y i2 k x k yci x,y

c

c

sin B x, yx, y e B x, y e

S k ,k S k ,k

2idxdy

B x, y

(5-10)

Using the expansion of sin(θ)/θ, and since αBc is extremely small, we can write the final relation

for Bc(x,y) as,

1 x

,

y x y

c i x y

S k ,k S kB x, y

,k1FT

2ix, y e

(5-11)

where, s xc

x

T kT 2

2 G

. Compared to the conventional method of measuring Bc(x,y),

the ICNE method has been shown to decrease the noise in the measured magnetic field map by

approximately 42% (78).

5.3 Methods

5.3.1 Field Stability Measurements

In tubes of 50 ml capacity, 2 % agar gel doped with sodium chloride and copper sulphate

were prepared using the same procedure described in Section 3.3.1. The tube was left at room

temperature for a whole day and in the evening, T1, T2 relaxation times were quantified using a

4.7 T Agilent magnet system. Background magnetic fields were characterized using an

asymmetric spin echo sequence (Figure 5-1) with free precession delay times (Tfp) of 0, 0.5, 1,

1.5, 2 ms. The repetition and echo times for the experiment were 1000 and 14 ms. The in-plane

resolution was 250 x 250 μm2 and the slice thickness was 1 mm. A total of 10 slices were

acquired. The entire experiment set was repeated two more times and differences in background

field maps were computed to ensure the temporal stability of the static magnetic field.

Page 110: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

110

5.3.2 Pilot Studies

5.3.2.1 Fabrication of carbon electrodes

Carbon has a magnetic susceptibility close to that of water. In the form of graphite, is

highly conductive due to its layered structure thus making it a viable option for use in building

electrodes for electrical stimulation. Carbon fiber tows consisting of 12000 individual filaments

(Formosa Plastics, Taiwan) were used to make the electrodes using a procedure similar to that of

Dunn et al. (79). Polyvinylidene fluoride (PVDF), a biocompatible polymer (solid at room

temperature) was used to coat the carbon fibers to provide insulation. Pellets of PVDF were

dissolved in 99% dimethyl sulfoxide (DMSO) by heating at approximately 70° C and stirring

constantly at 340 rpm. Complete dissolution of the polymer occurred over a period of 14 hours,

yielding a clear dark yellow solution. The solution was continually stirred to prevent

precipitation or clumping of polymer. Groups of carbon fiber strands (12000 strands) were

manually isolated and soaked in 0.9 % saline solution to prevent splaying. The thickness of the

groups was qualitatively estimated under a magnifying glass and those roughly 200 μm

(measured against a scale) in thickness were selected. The soaking also ensured that the

interstitial spaces between the filaments were filled with conductive medium. Soaked carbon

fibers were dipped into the PVDF solution for approximately 30 seconds to allow adequate

coating. The fibers were then heated in a conventional oven at 200° C for 15 minutes to

evaporate the DMSO and concentrate the fibers with PVDF plastic. The entire process was

repeated two more times to ensure thorough coating of PVDF on the carbon fibers.

5.3.2.2 Experimental setup

Carbon fiber electrodes fabricated using the procedure described above, were crushed at

one end to expose the fibers that were attached to a brass screw using a conductive silver epoxy.

A 40 x 40 x 40 mm box with a cap attachment was printed in 3D using PLA (poly lactic acid)

Page 111: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

111

with a Makerbot Replicator 2. Two holes matching the diameter of the screws were drilled on the

cap approximately 1 cm apart. The electrodes along with screws were inserted into the box

through the holes such that the screw heads were flushed against the cap surface. The screws

were epoxied in place and the exposed ends of insulated copper wires were soldered to the screw

heads. The box was filled with 2 % agar gel doped with NaCl and CuSO4. The entire

experimental setup was placed in the bore of a 4.7 T Agilent magnet system at the AMRIS

facility in University of Florida. The cores of coaxial cables were soldered to the other ends of

insulated wires and were attached to the front of the magnet’s bore using BNC connectors.

Coaxial cables carrying current from a constant current source developed at Kyung Hee

University, South Korea were attached to the other end of the BNC connectors.

5.3.2.3 MR imaging

Spin echo MREIT measurements using the sequence shown in Figure 5-1 were

performed using a current injection of 0.2 mA. A TR of 2000 ms was used along with a TE of 50

ms to facilitate current injection pulse widths of 20 ms each, prior to and following the 180° RF

pulse. The total current injection time was 40 ms. Twelve slices 1 mm thick were acquired along

the axial direction with an in-plane resolution of approximately 0.35 x 0.35 mm2.

5.3.3 Extended Studies

5.3.3.1 Phantom preparation

0.5 mm holes were drilled into a 50 ml cylindrical tube (Fisherbrand) in diametrically

opposite positions and insulated magnet wire (AWG 25) with only the tips exposed were inserted

through them until the wires were at 1 cm distance from each other. Epoxy was applied at the

holes to secure the wires in place and the setup was left overnight to dry. Two percent agar

solution doped with NaCl and CuSO4 was prepared and the mixture was poured into the

cylindrical tube to form a gel. Using similar approach, two more gel phantoms, one with the

Page 112: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

112

wires inserted parallel to the cylinder’s length and another where the wires were at right angles to

each other, were also constructed. Electrical stimulation was performed and current-induced

magnetic fields were mapped.

5.3.2.2 Electrical stimulation

Electrical stimulation was performed using a commercially available stimulator system

built by Tucker Davis Technologies, Alachua, Florida. The system consisted of an IZ2H

electrical stimulator that was synchronized with the magnet for stimulation using a RZ5D control

station. The control station was configured to accept a TTL input from the Agilent magnet

system and stimulation programmed to switch polarities for every incoming TTL signal. Initial

testing of experimental setup was performed by injecting current at 0.1, 0.2, 0.3, 0.5 and 0.75

mA into the phantom with wire ends placed diametrically opposite. Current of 0.25 mA was

injected into all three phantoms while simultaneously monitoring the stimulation waveforms.

5.3.3.3 MR imaging

Using the phantom with diametrically opposite wire ends, MREIT measurements using

the approach developed by Scott and colleagues (36) were performed at five different current

ratings (0.1, 0.2, 0.3, 0.5, 0.75 mA). MREIT-ICNE measurements were also performed at three

different current ratings (0.1, 0.2, 0.3 mA) to compare against standard MREIT measurements.

An echo time of 50 ms was used to inject current for a duration of 40 ms. Multi-slice acquisitions

with ten slices at 2 mm slice thickness were performed. A TR of 2 seconds was used to collect

data with an in-plane resolution of 0.33 x 0.33 mm2. The total acquisition time for each

measurement was 3 minutes. In all three phantoms, a current of 0.25 mA was injected and a total

of 5 averages were acquired to compare the current induced field maps across multiple averages.

In all experiments, a readout bandwidth of 16 KHz was used to improve the signal-to-noise ratio.

Page 113: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

113

5.4 Results and Discussion

T1 and T2 relaxation times for the 2% Agar gel phantom at 4.7 T ranged as 717 ± 19 ms

and 44.6 ± 0.8 ms. Maps of the static field inhomogeneity are presented in Figure 5-2 along with

their difference maps. The absence of any temporal drift over the time scale of MR acquisitions

was apparent from the difference maps. Magnetic field maps due to current injection using

fabricated carbon fiber electrodes are shown in Figure 5-3. The overwhelming effect of large

magnetic fields from current in the electrodes is evident in the region towards the top of the

phantom. Current induced magnetic field maps acquired using the conventional spin echo

sequence and the ICNE approach are shown in Figure 5-4 and Figure 5-5 for different ratings of

current. The increase in magnetic field intensity with current is clearly depicted. The results

indicated that for a given SNR during image acquisition and a set current injection time at 40 ms,

reliable measurements of magnetic field can be made using 0.3 mA of current injection. Current

injection times typically used in electrical stimulation techniques like DBS are orders of

magnitude lower than those used here. However, though electrical conductivity in tissue is

frequency dependent, for the frequencies used in DBS (60-150 Hz) or that used in this study (5 -

10Hz), the difference in conductivity values is extremely small and can be ignored. Therefore,

current injection using varying pulse lengths should produce similar magnetic fields and ergo

current distribution patterns. Therefore though it may not yet be possible to use MREIT for real

time measurements of current distributions during DBS yet, it is definitely possible to obtain a

sense of the volume of tissue activated during the procedure thereby working towards the

improvement of currently used computational models. Using very low amplitude injected

currents requires excellent image quality to ensure reliability which usually means longer

imaging times. The ICNE approach (Section 5.2.2) is one way to increase sensitivity in the

magnetic field measurement without the need for increased imaging times. This is apparent from

Page 114: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

114

the noise standard deviations within computed current-induced magnetic field maps given in

Table 5-1. However this approach relies on the use of increased sampling times and hence is

highly susceptible to off-resonance effects like chemical shift effects. In addition, since the one-

one mapping of spin position and precession frequency no longer holds, image reconstruction

using a linear operator like the Fourier transform is no longer viable and approximations about

the injected field magnitude become necessary to facilitate extraction of the current-induced

fields (Section 5.2.2). Faster pulse sequences adapted to perform MREIT (48,80) or parallel

imaging approaches like sensitivity encoding (SENSE) (81) or generalized auto-calibrating

partially parallel acquisition (GRAPPA) (82) can also be employed to circumvent the problem of

longer imaging times. Custom made array coils can be developed to further increase the image

quality while employing parallel imaging (83).

The greatest challenge for extending the current methodologies for in vivo imaging in

rodents is, however, the effect of large current densities in the electrode leads that can easily

mask the magnetic fields due to currents in tissue thereby affecting our calculation of current

densities. Figure 5-6 shows the magnetic field characterization following a current injection of

0.25 mA using different orientations of electrodes with respect to the static field of the MRI

scanner. Magnetic fields due to the electrodes angled in plane at 90° to each other seem to

produce fairly unbiased results. The diametrically opposite electrode orientation though optimal

for measuring current-induced magnetic fields within the region between the electrodes, it

presents complex options for surgery making it extremely challenging. Interestingly, for

electrodes oriented parallel to the main magnetic field, the current-induced magnetic fields inside

the gel phantom are completely masked by the noise levels in the image. Magnetic field changes

are observable only at the tip of the electrode where the current density is significantly high.

Page 115: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

115

Characterizing field changes in this kind of a stimulation setup is quite challenging as the tip of

the electrode is masked by image distortions due to susceptibility differences between the

electrode and surrounding medium. Electrodes fabricated from materials that are similar in

magnetic susceptibility to surrounding tissues could provide a better understanding of current

distributions and magnetic field changes close to the tip of the electrodes (84). Though MREIT

measurement using electrodes parallel to the static field of an MRI scanner is highly impractical

in in vivo imaging due to the requirement of inserting the electrodes through the eyes/olfactory

bulb of the rodent, the approach poses intriguing questions about the distribution of current

induced magnetic fields and the sensitivity of currently performed MREIT measurements.

5.5 Conclusions

Current induced magnetic fields due to diametrically opposite electrodes were mapped at

different current ratings to understand the sensitivity of spin echo and spin echo ICNE MREIT

approaches. Theoretical noise standard deviations, computed using the relation given by Sadleir

et al. (57), agree well with the experimentally measured noise standard deviation values. Current

induced magnetic fields due to varying electrode orientations were investigated and targeting

regions of the brain using electrodes placed at large angles with respect to each other seems to be

the best approach. Future studies should focus on improvement of the sensitivity of MREIT

measurements while employing smaller current pulse widths and faster imaging times. Parallel

imaging techniques should be explored thoroughly in order to improve the measurement

reliability and enable the feasibility of real-time MREIT measurements. Though distant, in the

future, these techniques would also provide the ability to integrate MREIT within the framework

of fMRI acquisitions to track conductivity changes due to neural activity. As the greatest changes

in current density can be expected close to the electrode tips, fabrication of electrodes that are

Page 116: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

116

susceptibility matched to the surrounding in combination with zoom MRI (85) can provide better

understanding of magnetic field and current density distributions.

Page 117: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

117

Figure 5-1. Asymmetric spin echo pulse sequence where the slice select 180° RF pulse is moved

from its original position to create a net phase accumulation due to free precession.

Page 118: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

118

Figure 5-2. Field stability measurements using the asymmetric spin echo sequence with varying

free precession delay times. A) The static field inhomogeneity fits for the acquired

data following three sets of repetition using Equation 3-5. B) Difference maps

computed from the ΔBo measurements showing good stability of Bo field over

experiment timescales of 30 minutes. The units are in Tesla.

A

B

Page 119: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

119

Figure 5-3. Demonstration of the dominant magnetic fields due to current in the electrodes.The

dissimilarity in the horizontal line plot across electrodes, was due to the electrodes not

being perfectly parallel to eachother. The image is scaled from -10 nT to 10 nT. The

Y-axis in the line plots represents magnetic field in Tesla while the X-axis indicates

position along the lines shown in the figures. The lines on the figure are drawn for the

entire length and width of the phantom. The discrepancies in the axes of the

experimental and simulated lineplots are due to the variation in the size of the

experimental phantom and the generated model.

Page 120: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

120

Figure 5-4. Current-induced magnetic field measurements for varying current amplitudes using conventional spin echo measurements.

A) Magnetic field maps showing that field patterns remain the same irrespective of current strengths. B) Magnetic field

values along the lines shown in A. Using fixed durations of current (40 ms) and identical imaging parameters, these results

show the reliability in magnetic field measurements with varying current amplitudes. The magnetic field values are shown

in Tesla.

A

B

Page 121: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

121

Figure 5-5. Current-induced magnetic field measurements (in Tesla) using the Injection Current Non-Linear Encoding (ICNE)

approach with varying levels of current amplitude. A) Magnetic field maps for different current ratings. B) Corresponding

magnetic field values plotted along the line shown in A. The difference in field maps at 0.3 mA obtained with conventional

spin echo (Figure 5-4) and the spin echo ICNE sequence can be attributed to non-linear reconstruction approach used in

ICNE.

B

A

Page 122: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

122

Table 5-1. Noise standard deviations for current induced field maps acquired with conventional spin echo MREIT and spin echo

MREIT with ICNE sequences.

Theoretical σBc (nT) Experimental σBc (nT)

0.1mA 0.2mA 0.3mA 0.5mA 0.75mA 0.1mA 0.2mA 0.3mA 0.5mA 0.75mA

Conventional 1.02 0.96 0.92 0.81 0.97 1.54 1.58 1.72 1.82 1.64

ICNE 1.03 0.89 0.86 - - 1.27 1.29 1.21 - -

Page 123: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

123

Figure 5-6. Magnetic field maps (in Tesla) due to 0.25 mA current injection in three phantoms

with different orientations of copper electrodes. A) Magnetic field maps for different

electrode orientations. B) Magnetic field values plotted along the lines shown in A.

The direction of static magnetic field is into the page in this figure. All data were

collected with five averages and the improvement in SNR for the angled case was due

to better positioning of the sample within the RF coil.

A

B

Page 124: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

124

CHAPTER 6

CONCLUSIONS AND FUTURE WORK

6.1 Summary

The work detailed in this dissertation addresses the dire need for quantifying current

density distribution in the human brain following transcranial electrical stimulation therapies. An

experimental methodology to reliably detect the minute magnetic fields due to low current

amplitudes using MREIT was developed. Additionally current-induced magnetic fields were

detected in phantoms and healthy volunteers to validate computational modeling techniques and

contribute to their development. Current density estimates from the measured magnetic fields

were obtained and compared against those produced from numerical simulations. Finally the

approach to understand magnetic field distributions was extended to electrical stimulation

techniques utilizing implanted electrodes like DBS. The best orientation of the stimulating

electrodes with respect to the static magnetic field of MRI machine was explored using a set of

three possible orientations.

In Chapter 3, discrepancies in MREIT acquisitions in pilot studies were addressed by

analyzing repeated acquisitions without any current injection. The analysis revealed the presence

of a magnetic field inhomogeneity that varied with time affecting the determination of current-

induced magnetic fields. The effect of static magnetic field drift with time was estimated and a

different acquisition style to collect imaging data that was insensitive to temporal drift was

employed. Necessary electronic circuitry was adapted to facilitate image acquisition and the

results indicated reliable measurement of current-induced magnetic fields.

In Chapter 4, the work done in Chapter 3 was extended to measure the low magnetic

fields and injected current densities in gel phantoms and healthy volunteers. This is the first

study in literature presenting injected current density maps in healthy volunteers. Current

Page 125: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

125

induced magnetic fields characterized in the gel phantom revealed an offset in the measurement

which could be due to extremely small imbalances in the amplitudes of the different current

polarities. The extreme sensitivity of these fields to electrode location, contact area and subject

motion was apparent from the results. Reproducible results were obtained in two participants

while quite different field patterns were obtained in two others during current injection along one

of the two directions. All the volunteers showed expected patterns (with very minor

discrepancies) for stimulation using T3, T4 electrode locations. Though the measurements

depicted the detectability of current induced magnetic fields using diametrically opposite

electrode locations in the head, the methods can easily be extended to study magnetic field or

current density patterns due to stimulation from arbitrary electrode locations.

In Chapter 5 characterization of magnetic fields was extended to implanted electrodes

with the aim to understand stimulation techniques like DBS. Different electrode orientations with

respect to the main magnetic field in MRI were explored to avoid the problem of magnetic fields

due to high current densities in electrodes masking the magnetic fields due to current flow within

the brain. Electrode orientation parallel to the field would produce least susceptibility image

distortion while those place diametrically opposite give best estimates of magnetic field due to

current injection but are both impractical for in vivo animal imaging. Angled electrodes provide

the best of both worlds and successful mapping of magnetic fields in such an orientation paves

way for improving our understanding of brain response due to cortical stimulation

6.2 Future Work

Quantifying current densities within different regions of the brain following electrical

stimulation has been a long standing goal in the field of neuroscience. Electrical stimulation

therapies rely on realistic computational models and validating these models is a giant leap that

could challenge/alter longstanding approaches and assumptions. The primary focus of this

Page 126: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

126

dissertation has been to provide evidence of small but measurable currents localized in the brain

due to transcranial stimulation. Successful characterization of currents and their magnetic fields

opens up innumerable prospects for the improvement of existing neuromodulation techniques.

However to begin with, the reliability and reproducibility of this work across individuals

should be established using a larger sample space. Carefully controlled experiments with

minimal variation in parameters such as electrode location, contact area and minimal subject

motion during imaging can improve the consistency of results. The beneficial effects of electrical

stimulation therapies are known to vary with stimulation strength, area of stimulation and

stimulus duration. All these factors should be meticulously addressed and controlled across

subjects to provide a comprehensive comparison against published literature. The work presented

in this dissertation does not provide rigorous statistical comparisons while validating the

computational models. Models taking into account the anisotropic nature of different regions in

the brain should be developed to better compare against measurement values. Neurovascular

coupling can potentially confound the measurement of current-induced magnetic fields

indicating the need for carefully designed experiments that control for such effects. As the

magnetic fields due to current injection are extremely small, it is possible that they produce

negligible distortion of the B1 field used in MR imaging. Imaging slice distortions with varying

amplitudes and durations of current injection should be quantified. Such characterizations can

improve the measurement of current-induced magnetic fields providing capabilities of extending

this work to direct current applications. Fast imaging sequences like echo planar imaging or fast

spin echo sequences coupled with parallel imaging techniques can significantly improve the

measurement process by increasing the volume of imaging. Currents measured in the work

presented in this dissertation depict a highly distributed path across the brain. Validated

Page 127: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

127

computational models can be employed to simulate current densities using an array of electrodes

to increase current density in deeper brain regions of brain tissue. The data from such

experiments can then once again be validated using the methods developed in this dissertation.

To obtain absolute validations of the current density imaging techniques presented in this work,

invasive measurements of underlying sample conductivity and ergo current density could be

made in freshly dead tissue. However, these measurements would have to take into account the

time for which the tissue was dead and the resulting change in tissue conductivity. MR

thermometry techniques can be applied to measure the change in temperature as a result of

decreased perfusion to the tissue and help augment the accuracy of conductivity/current density

measurements.

Current density characterization could be extended to stimulation using implanted

electrodes. Electrodes constructed from materials whose susceptibility is close to that of brain

tissue can provide excellent characterization of current-induced magnetic fields following the

details outlined in Chapter 5. The advantage of implanted electrodes is that, though invasive,

they can provide greater control over the region of brain stimulated and can therefore be used to

probe specific tissue properties while simultaneously providing deeper insights into the

mechanisms behind current distribution in the brain. Meticulously designed experiments can

improve our understanding by providing quantitative estimates of the degree of blood-brain

barrier disruption, information that can be utilized in drug delivery techniques. MREIT, by virtue

of its ability to generate high resolution conductivity maps, can also be used for cancer diagnosis.

Many studies have looked at feasibility and the promise of MREIT in successfully detecting

breast or prostate cancers (86,87). Current density imaging can also be used in non-biological

applications, like measuring the currents in a printed circuit board (88).

Page 128: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

128

LIST OF REFERENCES

1. Nitsche MA, Seeber A, Frommann K, Klein CC, Rochford C, Nitsche MS, Fricke K,

Liebetanz D, Lang N, Antal A, Paulus W, Tergau F. Modulating parameters of

excitability during and after transcranial direct current stimulation of the human motor

cortex. J Physiol 2005;568(Pt 1):291-303.

2. Clark VP, Coffman BA, Mayer AR, Weisend MP, Lane TD, Calhoun VD, Raybourn EM,

Garcia CM, Wassermann EM. TDCS guided using fMRI significantly accelerates

learning to identify concealed objects. Neuroimage 2012;59(1):117-128.

3. Antal A, Paulus W. Transcranial alternating current stimulation (tACS). Front Hum

Neurosci 2013;7(317):1-4.

4. Hodgkin AL, Huxley AF. A quantitative description of membrane current and its

application to conduction and excitation in nerve. J Physiol 1952;117(4):500-544.

5. Goldman DE. Potential, impedance and rectification in membranes. J Gen Physiol

1943;27(1):37-60.

6. Hodgkin AL, Katz B. The effect of sodium ions on the electrical activity of giant axon of

the squid. J Physiol 1949;108(1):37-77.

7. Singer SJ, Nicolson GL. The fluid mosaic model of the structure of cell membranes.

Science 1972;175(4023):720-731.

8. Poo M. In situ electrophoresis of membrane components. Annu Rev Biophys Bioeng

1981;10:245-276.

9. Fricke H. A mathematical treatment of the electric conductivity and capacity of disperse

systems ii. The capacity of a suspension of conducting spheroids surrounded by a non-

conducting membrane for a current of low frequency. Physical Review 1925;26(5):678-

681.

10. Datta A, Bikson M, Fregni F. Transcranial direct current stimulation in patients with skull

defects and skull plates: high-resolution computational FEM study of factors altering

cortical current flow. Neuroimage 2010;52(4):1268-1278.

11. Dmochowski JP, Datta A, Bikson M, Su Y, Parra LC. Optimized multi-electrode

stimulation increases focality and intensity at target. J Neural Eng 2011;8(4):1-16.

12. Datta A, Baker JM, Bikson M, Fridriksson J. Individualized model predicts brain current

flow during transcranial direct-current stimulation treatment in responsive stroke patient.

Brain Stimul 2011;4(3):169-174.

13. Bikson M, Rahman A, Datta A. Computational models of transcranial direct current

stimulation. Clin EEG Neurosci 2012;43(3):176-183.

Page 129: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

129

14. Truong DQ, Huber M, Xie X, Datta A, Rahman A, Parra LC, Dmochowski JP, Bikson

M. Clinician accessible tools for GUI computational models of transcranial electrical

stimulation: BONSAI and SPHERES. Brain Stimul 2014;7(4):521-524.

15. Fregni F, Boggio PS, Nitsche M, Bermpohl F, Antal A, Feredoes E, Marcolin MA,

Rigonatti SP, Silva MT, Paulus W, Pascual-Leone A. Anodal transcranial direct current

stimulation of prefrontal cortex enhances working memory. Exp Brain Res

2005;166(1):23-30.

16. Nitsche MA, Paulus W. Sustained excitability elevations induced by transcranial DC

motor cortex stimulation in humans. Neurology 2001;57(10):1899-1901.

17. Nitsche MA, Paulus W. Excitability changes induced in the human motor cortex by weak

transcranial direct current stimulation. J Physiol 2000;527(3):633-639.

18. Antal A, Boros K, Poreisz C, Chaieb L, Terney D, Paulus W. Comparatively weak after-

effects of transcranial alternating current stimulation (tACS) on cortical excitability in

humans. Brain Stimul 2008;1(2):97-105.

19. Liebetanz D, Nitsche MA, Tergau F, Paulus W. Pharmacological approach to the

mechanisms of transcranial DC-stimulation-induced after-effects of human motor cortex

excitability. Brain 2002;125(10):2238-2247.

20. Kanai R, Chaieb L, Antal A, Walsh V, Paulus W. Frequency-dependent electrical

stimulation of the visual cortex. Curr Biol 2008;18(23):1839-1843.

21. Zaghi S, de Freitas Rezende L, de Oliveira LM, El-Nazer R, Menning S, Tadini L, Fregni

F. Inhibition of motor cortex excitability with 15Hz transcranial alternating current

stimulation (tACS). Neurosci Lett 2010;479(3):211-214.

22. Merrill DR, Bikson M, Jefferys JG. Electrical stimulation of excitable tissue: design of

efficacious and safe protocols. J Neurosci Methods 2005;141(2):171-198.

23. Woods AJ, Antal A, Bikson M, Boggio PS, Brunoni AR, Celnik P, Cohen LG, Fregni F,

Herrmann CS, Kappenman ES, Knotkova H, Liebetanz D, Miniussi C, Miranda PC,

Paulus W, Priori A, Reato D, Stagg C, Wenderoth N, Nitsche MA. A technical guide to

tDCS, and related non-invasive brain stimulation tools. Clin Neurophysiol

2016;127(2):1031-1048.

24. Seo JK, Kwon O, Woo EJ. Magnetic resonance electrical impedance tomography

(MREIT): conductivity and current density imaging. Journal of Physics: Conference

Series 2005;12(1):140-155.

25. Rabi II, Zacharias JR, Millman S, Kusch P. A new method of measuring nuclear

magnetic moment. Physical Review 1938;53(4):318.

26. Bloch F, Hansen W, Packard M. The nuclear induction experiment. Physical Review

1946;70(7,8):474-485.

Page 130: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

130

27. Bloch F. Nuclear induction. Physical Review 1946;70(7,8):460-474.

28. Purcell EM, Torrey HC, Pound RV. Resonance absorption by nuclear magnetic moments

in a solid. Physical Review 1946;69:37-38.

29. Purcell EM. Spontaneous emission probabilities at radio frequencies. Physical Review

1946;69:681.

30. Bloembergen N, Purcell EM, Pound RV. Relaxation effects in nuclear magnetic

resonance absorption. Physical Review 1948;73(7):679-712.

31. Damadian R. Tumor detection by nuclear magnetic resonance. Science

1971;171(976):1151-1153.

32. Lauterbur PC. Image formation by induced local interactions. Examples employing

nuclear magnetic resonance. Clin Orthop Relat Res 1973;244:3-6.

33. Mansfield P, Maudsley AA, Baines T. Fast scan proton density imaging by NMR.

Journal of Physics E-Scientific Instruments 1976;9(4):271-278.

34. Garroway AN, Grannell PK, Mansfield P. Image formation using nuclear magnetic

resonance. US Patents 1977;610436(4021726):1-12.

35. Mansfield P, Pykett IL. Biological and medical imaging by NMR. Journal of Magnetic

Resonance 1978;29:355-373.

36. Scott GC, Joy MLG, Armstrong RL, Henkelman RM. Measurement of nonuniform

current density by magnetic resonance. IEEE Transactions on Medical Imaging

1991;10(3):362-374.

37. Scott GC, Joy MLG, Armstrong RL, Henkelman RM. Sensitivity of magnetic-resonance

current-density imaging. Journal of Magnetic Resonance 1992;97(2):235-254.

38. Lee BI, Park C, Pyo HC, Kwon O, Woo EJ. Optimization of current injection pulse width

in MREIT. Physiol Meas 2007;28(1):N1-7.

39. Seo JK, Yoon JR, Woo EJ, Kwon O. Reconstruction of conductivity and current density

images using only one component of magnetic field measurements. IEEE Trans Biomed

Eng 2003;50(9):1121-1124.

40. Seo JK, Pyo HC, Park C, Kwon O, Woo EJ. Image reconstruction of anisotropic

conductivity tensor distribution in MREIT: computer simulation study. Phys Med Biol

2004;49(18):4371-4382.

41. Park C, Kwon OI. Current density imaging using directly measured harmonic Bz data in

MREIT. Comput Math Methods Med 2013;381507:1-9.

Page 131: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

131

42. Nam HS, Lee BI, Choi J, Park C, Kwon OI. Conductivity imaging with low level current

injection using transversal J-substitution algorithm in MREIT. Phys Med Biol

2007;52(22):6717-6730.

43. Birgul O, Eyuboglu BM, Ider YZ. Current constrained voltage scaled reconstruction

(CCVSR) algorithm for MR-EIT and its performance with different probing current

patterns. Phys Med Biol 2003;48(5):653-671.

44. Park C, Lee BI, Kwon OI. Analysis of recoverable current from one component of

magnetic flux density in MREIT and MRCDI. Phys Med Biol 2007;52(11):3001-3013.

45. Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic Resonance Imaging

Physical Principles and Sequence Design. New York: John Wiley & Sons; 1999.

46. Andersen AH. On the Rician distribution of noisy MRI data. Magn Reson Med

1996;36(2):331-333.

47. Gudbjartsson H, Patz S. The Rician distribution of noisy MRI data. Magn Reson Med

1995;34(6):910-914.

48. Oh TI, Jeong WC, Kim JE, Sajib SZ, Kim HJ, Kwon OI, Woo EJ. Noise analysis in fast

magnetic resonance electrical impedance tomography (MREIT) based on spoiled multi

gradient echo (SPMGE) pulse sequence. Phys Med Biol 2014;59(16):4723-4738.

49. Woo EJ, Seo JK. Magnetic resonance electrical impedance tomography (MREIT) for

high-resolution conductivity imaging. Physiological Measurement 2008;29(10):R1-26.

50. Polania R, Paulus W, Antal A, Nitsche MA. Introducing graph theory to track for

neuroplastic alterations in the resting human brain: a transcranial direct current

stimulation study. Neuroimage 2011;54(3):2287-2296.

51. Antal A, Bikson M, Datta A, Lafon B, Dechent P, Parra LC, Paulus W. Imaging artifacts

induced by electrical stimulation during conventional fMRI of the brain. Neuroimage

2014;85:1040-1047.

52. Sadleir RJ, Vannorsdall TD, Schretlen DJ, Gordon B. Target optimization in transcranial

direct current stimulation. Front Psychiatry 2012;3(90):1-13.

53. El-Sharkawy A, Schär M, Bottomley P, Atalar E. Monitoring and correcting spatio-

temporal variations of the MR scanner’s static magnetic field. Magnetic Resonance

Materials in Physics, Biology and Medicine 2006;19(5):223-236.

54. Astary GW, Peprah MK, Fisher CR, Stewart RL, Carney PR, Sarntinoranont M, Meisel

MW, Manuel MV, Mareci TH. MR measurement of alloy magnetic susceptibility:

Towards developing tissue-susceptibility matched metals. Journal of Magnetic

Resonance 2013;233:49-55.

Page 132: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

132

55. Bernstein MA, King KE, Zhou XJ. Handbook of MRI Pulse Sequences: Academic Press;

2004.

56. Parker DL, Payne A, Todd N, Hadley JR. Phase reconstruction from multiple coil data

using a virtual reference coil. Magn Reson Med 2014;72(2):563-569.

57. Sadleir R, Grant S, Zhang SU, Lee BI, Pyo HC, Oh SH, Park C, Woo EJ, Lee SY, Kwon

O, Seo JK. Noise analysis in magnetic resonance electrical impedance tomography at 3

and 11 T field strengths. Physiol Meas 2005;26(5):875-884.

58. Rieke V, Vigen KK, Sommer G, Daniel BL, Pauly JM, Butts K. Referenceless PRF shift

thermometry. Magn Reson Med 2004;51(6):1223-1231.

59. Gabriel C, Gabriel S, Corthout E. The dielectric properties of biological tissues: I.

Literature survey. Phys Med Biol 1996;41(11):2231-2249.

60. Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: II.

Measurements in the frequency range 10 Hz to 20 GHz. Phys Med Biol

1996;41(11):2251-2269.

61. Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: III.

Parametric models for the dielectric spectrum of tissues. Phys Med Biol

1996;41(11):2271-2293.

62. van Dongen R, Burger HC. Specific electric resistance of body tissues. Physics in

Medicine and Biology 1961;5(4):431-449.

63. Sadleir RJ, Vannorsdall TD, Schretlen DJ, Gordon B. Transcranial direct current

stimulation (tDCS) in a realistic head model. Neuroimage 2010;51(4):1310-1318.

64. Heethaar RM, Faes TJC, Meij HA, de Munck JC. The electric resistivity of human tissues

(100 Hz-10 MHz): a meta-analysis of review studies. Physiological Measurement

1999;20(4):R1-10.

65. Akhtari M, Bryant HC, Mamelak AN, Heller L, Shih JJ, Mandelkern M, Matlachov A,

Ranken DM, Best ED, Sutherling WW. Conductivities of three-layer human skull. Brain

Topogr 2000;13(1):29-42.

66. Akhtari M, Bryant HC, Mamelak AN, Flynn ER, Heller L, Shih JJ, Mandelkern M,

Matlachov A, Ranken DM, Best ED, DiMauro MA, Lee RR, Sutherling WW.

Conductivities of three-layer live human skull. Brain Topogr 2002;14(3):151-167.

67. Logothetis NK, Kayser C, Oeltermann A. In vivo measurement of cortical impedance

spectrum in monkeys: implications for signal propagation. Neuron 2007;55(5):809-823.

68. Nicholson PW. Specific impedance of cerebral white matter. Experimental Neurology

1965;13(4):386-401.

Page 133: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

133

69. Miranda PC, Hallett M, Basser PJ. The electric field induced in the brain by magnetic

stimulation: a 3-D finite-element analysis of the effect of tissue heterogeneity and

anisotropy. IEEE Trans Biomed Eng 2003;50(9):1074-1085.

70. Plonsey R, Heppner DB. Considerations of quasi-stationarity in electrophysiological

systems. Bull Math Biophys 1967;29(4):657-664.

71. Kwon OI, Jeong WC, Sajib SZ, Kim HJ, Woo EJ, Oh TI. Reconstruction of dual-

frequency conductivity by optimization of phase map in MREIT and MREPT. Biomed

Eng Online 2014;13(24):1-15.

72. Jog MV, Smith RX, Jann K, Dunn W, Lafon B, Truong D, Wu A, Parra L, Bikson M,

Wang DJJ. In vivo imaging of magnetic fields induced by transcranial direct current

stimulation (tDCS) in human brain using MRI. Scientific Reports 2016;6(34385):1-10.

73. Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array.

Magn Reson Med 1990;16(2):192-225.

74. Constantinides CD, Atalar E, McVeigh ER. Signal-to-noise measurements in magnitude

images from NMR phased arrays. Magn Reson Med 1997;38(5):852-857.

75. Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal-to-

noise ratios in MR images: influence of multichannel coils, parallel imaging, and

reconstruction filters. J Magn Reson Imaging 2007;26(2):375-385.

76. Miranda PC, Lomarev M, Hallett M. Modeling the current distribution during

transcranial direct current stimulation. Clinical Neurophysiology 2006;117(7):1623-1629.

77. Du J, Carl M, Bydder M, Takahashi A, Chung CB, Bydder GM. Qualitative and

quantitative ultrashort echo time (UTE) imaging of cortical bone. J Magn Reson

2010;207(2):304-311.

78. Kwon OI, Lee BI, Nam HS, Park C. Noise analysis and MR pulse sequence optimization

in MREIT using an injected current nonlinear encoding (ICNE) method. Physiol Meas

2007;28(11):1391-1404.

79. Dunn JF, Tuor UI, Kmech J, Young NA, Henderson AK, Jackson JC, Valentine PA,

Teskey GC. Functional brain mapping at 9.4T using a new MRI-compatible electrode

chronically implanted in rats. Magn Reson Med 2009;61(1):222-228.

80. Hamamura MJ, Muftuler LT. Fast imaging for magnetic resonance electrical impedance

tomography. Magnetic Resonance Imaging 2008;26(6):739-745.

81. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: Sensitivity encoding

for fast MRI. Magn Reson Med 1999;42(5):952-962.

82. Sodickson DK, McKenzie CA. A generalized approach to parallel magnetic resonance

imaging. Med Phys 2001;28(8):1629-1643.

Page 134: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

134

83. Muftuler LT, Chen G, Hamamura MJ, Ha SH. MREIT with SENSE acceleration using a

dedicated RF coil design. Physiol Meas 2009;30(9):913-929.

84. Astary GW, Peprah MK, Fisher CR, Stewart RL, Carney PR, Sarntinoranont M, Meisel

MW, Manuel MV, Mareci TH. MR measurement of alloy magnetic susceptibility:

towards developing tissue-susceptibility matched metals. J Magn Reson 2013;233:49-55.

85. Setser RM, Kotys M, Bolen MA, Muthupillai R, Flamm SD. High resolution imaging of

the right ventricle using ZOOM MRI. Journal of Cardiovascular Magnetic Resonance

2010;12(1):1-2.

86. Li G, Chen R, Hao L, Lin L. Three dimensional MREIT for breast cancer detection on

open MRI scanners. IEEE International Conference on Information and Automation

2012:446-450.

87. Liu Y, Zhang Y. A feasibility study of magnetic resonance electrical impedance

tomography for prostate cancer detection. Physiol Meas 2014;35(4):567-581.

88. Pesikan P, Joy MLG, Scott GC, Henkelman RM. Two-dimensional current density

imaging. IEEE Transactions on Instrumentation and Measurement 1990;39(6):1048-

1053.

Page 135: © 2016 Aditya Kumar Kasinadhuni - University of Floridaufdcimages.uflib.ufl.edu/UF/E0/05/06/90/00001/KASINADHUNI_A.pdf · ADITYA KUMAR KASINADHUNI A DISSERTATION PRESENTED TO THE

135

BIOGRAPHICAL SKETCH

Aditya Kumar Kasinadhuni was born in Hyderabad, a metropolitan city in South India.

He completed his schooling from the Hyderabad Public School, Begumpet in 2003 and earned

his Bachelor of Engineering degree from Osmania University (O.U) in the field of Biomedical

Engineering in 2009.While at O.U, he gained hands on expertize in Biomedical Instrumentation

during his internship with CARE Hospitals and developed image processing skills as part of his

undergraduate research in Apollo Gleneagles PET-CT center under the guidance of Dr. Jyotsna

Rao. He moved to the United States of America in 2009 to pursue his Master of Science degree

in Biomedical Engineering and achieved successful completion in 2011 from the University of

Florida, Gainesville. In the same year he embarked on his doctoral degree under the mentorship

of Dr. Thomas Mareci to study the development and applications of Magnetic Resonance

Imaging for mapping electrical current density in the brain. He received his Ph.D. in Biomedical

Engineering from the University of Florida in the fall of 2016.