Contents Intelligent Pressure Transmitter ....................................................................................... - 1 -
Brief Introduction ........................................................................................................................................... - 3 - Features ......................................................................................................................................................... - 3 - Working Principle .......................................................................................................................................... - 4 - EJC - DP Differential Pressure Transmitter ............................................................................................ - 6 - 1 Performance Specification ....................................................................................................................... - 7 - 2 Functional Specification ............................................................................................................................ - 8 - 3 Installation ................................................................................................................................................... - 9 - 4 Physical Specification ............................................................................................................................. - 10 - 5 Electric connection chart ........................................................................................................................ - 12 - 6 Process connection instruction .............................................................................................................. - 12 - 7 Model and Specification Code Table .................................................................................................... - 12 -
-GP Pressure Transmitter ................................................................................................................ - 14 - -AP Differential Pressure Transmitter ............................................................................................. - 14 -
1 Performance Specification ..................................................................................................................... - 14 - 2 Functional Specification .......................................................................................................................... - 15 - 3 Installation ................................................................................................................................................. - 17 - 4 Physical Specification ............................................................................................................................. - 17 - 5 Electric connection chart ........................................................................................................................ - 19 - 6 Process connection instruction .............................................................................................................. - 19 - 7 Model and Specification Code Table .................................................................................................... - 21 -
-DGP Bracket Installation Gauge Pressure Transmitter ................................................................ - 22 - -DAP Bracket Installation Absolute Pressure Transmitter ........................................................... - 22 -
1 Performance Specification ..................................................................................................................... - 22 - 2 Functional Specification .......................................................................................................................... - 23 - 3 Installation ................................................................................................................................................. - 25 - 4 Physical Specification ............................................................................................................................. - 25 - 5 Electric connection chart ........................................................................................................................ - 27 - 6 Process connection instruction .............................................................................................................. - 28 - 7 Model and Specification Code Table .................................................................................................... - 28 -
EJC Series
EJCEJC
EJCEJC
- 3 -
1 Differential Pressure Transmitter (DP)
Measured media: gas, steam, liquid
Measured range(with no shift): 0Pa~100Pa…3MPa
Accuracy: ±0.075%
Diaphragm contacting with liquid: Stainless Steel 316L, Hast-alloy C
2 Gauge Pressure Transmitter (GP)
Absolute Pressure Transmitter (AP)
Measured media: gas, steam, liquid
Measured range(with no shift): 0Pa~600Pa…40MPa
Accuracy: ±0.075%
Diaphragm contacting with liquid: Stainless Steel 316L, Hast-alloy
3 Bracket Installation Gauge Pressure Transmitter (DGP)
Bracket Installation Absolute Pressure Transmitter (DAP)
Measured media: gas, steam, liquid
Measured range(with no shift): 0Pa~600Pa…40MPa
Accuracy: ±0.075%
Diaphragm contacting with liquid: Stainless Steel 316L, Hast-alloy
Features
※ The central sensing element of transmitter uses the world’s leading high-accuracy silicon sensor technology,
accuracy is ±0.075%;
※ Working pressure of transmitter has three levels--16MPa, 25MPa and 40MPa, the highest one-way
overpressure is 40MPa;
※ Excellent static pressure performance, optimal static pressure error ≤±0.1%/10MPa;
EJC-GP/AP Gauge Pressure Transmitter /Absolute Pressure Transmitter
EJC-DGP/DAP Bracket Installation Gauge/Absolute Pressure Transmitter
EJC-DP Differential Pressure Transmitter
- 4 -
※ The inner of pressure sensor integrates high sensitive temperature sensor.
※ Excellent temperature performance, optimum≤± (0.20 ×TD+0.10)%×Span /-20 ℃~65 ℃;
※ All stainless steel 316L, silicone oil filling with welded sealing construction;
※ Low gauge /absolute pressure transmitter uses the world’s leading die resistance to high overload protection.
6kPa: one-way overpressure and protection pressure is 300kPa; 40kPa: one-way overpressure and protection
pressure is 1MPa. The max. Pressure is 50 times of full span pressure.
※ Stable and reliable, optimal long-term drift performance:±0.1%/year, 5-year maintenance-free;
※ Wide measured range: 100Pa~60MPa;
※ Max. 100:1 pressure range proportion adjustable;
※ EMC conforms to GB/T 18268.1-2008 standard;
Working Principle
1 Differential Pressure Transmitter(DP)
Chart 1 Differential Pressure Transmitter (DP)
Differential pressure transmitter includes two functional units:
※ Main unit
※ Auxiliary unit
Main unit includes sensor and process connection, working principle as followed:
The sensor module uses whole welded technology, in which has a compact overload diaphragm, a differential
pressure sensor and a temperature sensor. The temperature is taken as a reference for temperature compensation.
- 5 -
The positive end of the differential pressure sensor is connected with high pressure chamber of sensor capsule; the
negative end is connected with low pressure chamber of sensor capsule. Through the isolated diaphragm and filling
liquid, the differential pressure is transmitted to silicon die in the inner of differential pressure sensor, which makes the
resistor of sensor die change. So the detection system outputs different voltage. The output voltage is in proportion to
the pressure variation, and then it is transmitted to standard output by adapter and amplifier.
2 Gauge Pressure Transmitter (EJCGP )
Absolute Pressure Transmitter (EJCAP )
Chart 2 Gauge (GP)pressure transmitter/Absolute(AP)pressure transmitter
Pressure transmitter includes two functional units:
※ Main unit
※ Auxiliary unit
Main unit includes sensor and process connection, working principle as followed:
Through corrugated, isolated diaphragm and filling oil, process media is pressurized to diaphragm of pressure
sensor. The other end of pressure sensor diaphragm is connected to the air (for gauge measurement) or vacuum(for
absolute measurement). In such way, it makes the resistor of sensor die change so that the detection system outputs
different voltage. The output voltage is in proportion to the pressure variation, and then it is transmitted to standard
output by adapter and amplifier.
3 Bracket Installation Gauge Pressure Transmitter (EJCDGP )
Bracket Installation Absolute Pressure Transmitter (EJCDAP )
- 6 -
Chart 3 Differential Gauge Pressure Transmitter(DGP)/Differential Absolute Pressure Transmitter(DAP)
Differential pressure transmitter includes two functional units:
※ Main unit
※ Auxiliary unit
Main unit includes sensor and process connection, working principle as followed:
The sensor module uses whole welded technology, in which has a pressure sensor and a temperature sensor.
The pressure sensor is used for gauge or absolute pressure measurement. The temperature is taken as a reference
for temperature compensation. The measured end of pressure sensor is connected to working chamber of sensor
capsule; the negative end is connected to reference chamber of sensor capsule and open to air. In such way, the
pressure difference is transmitted to silicon die of pressure sensor, which makes the resistor of sensor die change. So
the detection system outputs different voltage. The output voltage is in proportion to the pressure variation, and then it
is transmitted to standard output by adapter and amplifier.
-DP Differential Pressure Transmitter
-DP Differential Pressure Transmitter is used for level, density, pressure and
flow measurement of liquid, gas and steam. Then it will outputs 4mA~20mA DC HART signal
and also it could be connected to RST375 hand communicator or RSM100 Modem to do the
specification setting and process control.
Standard Specification
(Standard zero as the reference calibration range, Stainless steel 316L diaphragm, filling liquid
is silicone oil)
EJC
EJC
- 7 -
1 Performance Specification
Reference accuracy for range calibration(including linearity, hystersis and repeatability from zero): ± 0.075%
If TD>10 (TD=Max. Pressure range/calibration range), the accuracy is ±(0.0075×TD)%
The accuracy of square root output is 1.5 times of above reference accuracy.
Environmental Temperature Effect
Range Code -20℃~65℃ Total effect value Range Code -40℃~-20℃和 65℃~85℃Total effect value
A ±(0.45×TD+0.25)%×Span A ±(0.45×TD+0.25)%×Span
B ±(0.30×TD+0.20)%×Span B ±(0.30×TD+0.20)%×Span
C/D/F ±(0.20×TD+0.10)%×Span C/D/F ±(0.20×TD+0.10)%×Span
Over range effect: ±0.075%×Span
Static pressure effect
Overpressure effect
Range Code Effect value
A ±0.5%×Span/4MPa
B ±0.2%×Span/16MPa
C/D/F ±0.1%×Span/16MPa
Long-term stability
Range Code Effect value
A ±0.5%×Span/1 year
B ±0.2%×Span/1 year
C/D/F ±0.1%×Span/1 year
Power effect ±0.001% /10V (12V~42V DC), negligible.
Range Code Effect value
A ±(0.5%Span)/4MPa
B ±(0.3%Span)/10MPa
C/D/F ±(0.1%Span)/10MPa
- 8 -
2 Functional Specification
Pressure range and limits
range/limits kPa mbar
A range 0.1~1 1~10
limits -1~1 -10~10
B range 0.2~6 2~60
limits -6~6 -60~60
C range 0.4~40 4~400
limits -40~40 -400~400
D range 2.5~250 25~2500
limits -250~250 -2500~2500
F range 30~3000 0.3~30 bar
limits -500~3000 -5~30 bar
Pressure range limit
The pressure is adjustable within the upper and lower limit.
It is recommended to choose the range code with the lowest pressure range proportion to optimize the
performance specification.
Zero setting
The zero and pressure range could be adjusted to any value within the measured rang in the table, only the
calibrated range≥Min. Range is valid.
Mounting position effect
The change of mounting position parallel to diaphragm could not influence the zero drift. If the angle between
mounting position and diaphragm is over 90° , the zero drift is <0.4kPa which could be calibrated by zero setting. No
other effect on pressure range.
Output
2- wire, 4mA~20mA DC, HART communication protocol, linearity or square root output optional.
Output signal limit: Imin=3.9mA,Imax=20.5mA
Alarm Current
Low alarm mode(min.): 3.7mA
High alarm mode(max.): 21mA
No alarm mode(holding): holding the active current value before failure
Standard alarm current mode: High alarm mode
- 9 -
Response time
The damping constant of amplifier parts is 0.1s, time constant of sensor is 0.1s~1.6s, which is decided by the
pressure range and pressure range ratio. The additional adjustable time constant is 0.1s~60s. The non-linearity
output(eg. Square root output) is influenced by this function and could be calculated by it.
Warm-up time
< 15s
Environmental temperature
-40 ℃~ 85℃
With LCD display and viton sealing ring, the temperature is -20℃~65 ℃.
Storage temperature/ transportation temperature
-50 ℃~ 85 ℃;with LCD display:-40℃~ 85 ℃
Working pressure
Rated working pressure: 16MPa, 25MPa, 40MPa
Static pressure limit
From 3.5kPa absolute pressure to rated pressure, protection pressure can be pressurized to both high and low
side of transmitter; and it can be higher than 1.5 times of rated pressure.
One-way overpressure limit
One-way overpressure could reach the rated pressure
EMC
Please refer to next page ”EMC table”
3 Installation
Power and load condition
Power supply: 24V DC, R≤(Us-12V)/Imax (kΩ)
Imax=23mA
Max. Voltage supply: 42V DC
Min. Voltage supply: 12V DC, 15V DC(Backlit LCD display)
Digital communication load resistance range: 250Ω~600Ω
Electric Connection
- 10 -
M20×1.5 cable sealing buckle, terminals are suitable for (0.5~2.5)mm2 wire.
Process connection
NPT 1/4 and UNF 7/16″ female at both sides of process connection flange.
4 Physical Specification
Material
Measuring capsule: Stainless Steel 316L
Diaphragm: Stainless Steel 316L, Hast-alloy C
Process flange: Stainless steel 304
Nut and bolt: Stainless steel(A4)
Filling liquid: silicone oil
Sealing ring: NBR, FKM, PTFE
Transmitter housing: Aluminum alloy material, epoxy resin glue sprays on the surface
Housing sealing ring: NBR
Nameplate: Stainless steel 304
Weight
3.3kg(not including LCD display, mounting support and process connection)
Housing protection
IP67
EMC Table
code Test terms Standard Test condition Performance degree
1 Radiated
interference(housing) GB/T 9254-2008 table5 30MHz~1000MHz qualified
2 Transmission interference
(DC power port) GB/T 9254-2008 table1 0.15MHz~30MHz qualified
3 ESD immunity GB/T 17626.2-2006 4kV(contact)
8kV(air) B
4 Radiofrequency
ectromagnetic field immunity GB/T 17626.3-2006
10V/m
(80MHz~1GHz) A
5 Power frequency magnetic
field immunity GB/T 17626.8-2006 30A/m A
6 EFT immunity GB/T 17626.4-2008 2kV(5/50ns,5kHz) B
- 11 -
Note: (1) A degree:performance is normal within the technical standard range during testing.
(2) B degree: During, the function or performance is lowered or lost temporarily, but it could be recovered by itself. Actual
operation state, storage and data will keep the same.
Outline Dimension
Unit: mm
Horizontal Piping Installation (side view) Horizontal Piping Installation (front view)
Wall Installation Vertical Piping Installation
06
- 12 -
5 Electric connection chart
Note: the function of shortcut interface is equal to signal terminal.
6 Process connection instruction
7 Model and Specification Code Table
-DP Intelligent Differential Pressure Transmitter
Code Output
H 4mA~20mA DC with HART
N 4mA~20mA DC analog output
Code Pressure range
A 0Pa…100Pa~1kPa(0mm…10mm~100 mm H2O)/(0mbar…1mbar ~10mbar)
B 0Pa…200Pa~6kPa(0mm…20mm~600 mm H2O)/(0mbar…2mbar ~60mbar)
C 0Pa…400Pa~40kPa(0mm…40mm~4000 mm H2O)/(0mbar…20mbar ~400mbar)
D 0kPa…2.5kPa~250kPa(0m…0.25m~25 m H2O)/(0mbar…25mbar ~2500mbar)
F 0kPa…30kPa~3MPa(0m…3m~300 m H2O)/ (0bar…0.3bar~30bar)
Code Diaphragm material Filling
A Stainless steel316L silicone oil
C Hastelloy C silicone oil
Code Rated working pressure 0 0.2MPa (only for range A)
EJC
- 13 -
7 7MPa (only for range A) 1 16MPa 2 25MPa 3 40MPa
Code Process connection
N 1/4 NPT and 7/16 UNF thread hole without release valve
B 1/4 NPT and 7/16 UNF thread hole, release valve mounting in the end-face of flange back
U 1/4 NPT and 7/16 UNF thread hole, release valve mounting in upper flange side
D 1/4 NPT and 7/16 UNF thread hole, release valve mounting in lower flange side
Code Sealing materials contacting with liquid N NBR F FKM
P PTFE
Code Additional function N None F Square root output
O No oil processing ( For oxygen measurement: fluorocarbon oil filling, viton sealing ring 、<6MPa、<60℃)
Code Mounting bracket N None 1 Stainless steel 2 Galvanized Carbon Steel
Code Process connection parts N None
1 1/2 NPT Female with stainless steel oval flange
2 M20X1.5 male with stainless steel T joint
Code Display N No display 1 LCD 2 LCD with back-light
Code Others N None A Intrinsic safe D Exd
E Exd version with Explosion-proof cable joint
S Stainless steel 316 plate
-DP H C A 1 B N F 1 1 1 N The whole spec EJC
- 14 -
-GP Pressure Transmitter
-AP Differential Pressure Transmitter
series gauge/absolute pressure transmitter is used for level, density,
pressure and flow measurement of liquid, gas and steam. Then it will outputs
4mA~20mA DC HART signal and also it could be connected to RST375 handing
communicator or RSM100Modem to do the specification setting and process control.
Standard Specification
(Standard zero as the reference calibration range, Stainless steel 316L diaphragm,
filling liquid is silicone oil)
1 Performance Specification
Reference accuracy for range calibration (including linearity, hystersis and repeatability from zero): ± 0.075%
If TD>10 (TD=Max. Pressure range/calibration range), the accuracy is ±(0.0075×TD)%
The accuracy of square root output is 1.5 times of above reference accuracy.
Environmental Temperature Effect
Range code -20℃ ~ 65℃ total effect value
B/L ±(0.30×TD+0.20)%×Span
others ±(0.20×TD+0.10)%×Span
Range code -40℃~-20℃ and 65℃ ~85℃ total effect value
B/L ±(0.30×TD+0.20)%×Span
others ±(0.20×TD+0.10)%×Span
Over pressure effect: ±0.075%×Span
Long-term stability
Range Code Effect value
B/L ±0.2%×Span/1 year
other ±0.1%×Span/1 year
Power effect
±0.001% /10V (12V~42V DC),negligible.
EJC
EJC
EJC
- 15 -
2 Functional Specification
Pressure range and limits(EJC-GP Gauge Pressure)
range/limits kPa bar
B range 0.6~6 (6~60)mbar
limits -6~6 (-60~60)mbar
C range 2~40 0.02~0.4
limits -40~40 -0.4~0.4
D range 2.5~250 0.025~2.5
limits -100~250 -1~2.5
F range 30~3000 0.3~30
limits -100~3000 -1~30
G range (0.1~10)MPa 1~100
limits (-0.1~10)MPa -1~100
H range (0.21~21)MPa 2.1~210
limits (-0.1~21)MPa -1~210
I range (0.4~40)MPa 4~400
limits (-0.1~40)Mpa -1~400
J range (0.6~60)MPa 6~600
limits (-0.1~60)MPa -1~600
Pressure range and limits (MDM3051S-AP Absolute Pressure)
range/limits kPa bar
L range 2~40 0.02~0.4
limits 0~40 0~0.4
M range 2.5~250 0.025~2.5
limits 0~250 0~2.5
O range 30~3000 0.3~30
limits 0~3000 0~30
Pressure range limit
The pressure is adjustable within the upper and lower limit.
It is recommended to choose the range code with the lowest pressure range proportion to optimize the
performance specification.
Zero setting
The zero and pressure range could be adjust to any value within the measured rang in the table, only the
calibrated range≥Min. Range is valid.
- 16 -
Mounting position effect
The change of mounting position parallel to diaphragm could not influence the zero drift. If the angle between
mounting position and diaphragm is over 90° , the zero drift is <0.25kPa for range C and <0.15kPa for other ranges
which could be calibrated by zero setting. No other effect on pressure range.
Output
2- wire, 4mA~20mA DC, HART communication protocol, linearity or square root output optional.
Output signal limit: Imin=3.9mA,Imax=20.5mA
Alarm Current
Low alarm mode(min.): 3.7mA
High alarm mode(max.): 21mA
No alarm mode(holding): holding the active current value before failure
Standard alarm current mode: High alarm mode
Response time
The damping constant of amplifier parts is 0.1s, time constant of sensor is 0.1s~1.6s, which is depended on the
pressure range and pressure range proportion.
The additional adjustable time constant is 0.1s~60s.
Warm-up time
< 15s
Environmental temperature
-40℃~ 85 ℃
With LCD display and viton sealing ring, the temperature is -20℃~65 ℃.
Storage temperature/ transportation temperature
-50℃~ 85℃;with LCD display:-40℃~85 ℃
Pressure limit
It is from vacuum to Max. Pressure range.
Overpressure Limit
Pressure
range
6kPa
(B)
40kPa
(C/L)
250kPa
(D/M)
3MPa
(F/O)
Overpressure
limit 0.2MPa 1MPa 4MPa 16MPa
Overpressure 10MPa 21MPa 40MPa 60MPa
- 17 -
range (G) (H) (I) (K)
Overpressure
limit 20MPa 50MPa 50MPa 70MPa
EMC
Please refer to next page ”EMC table”
3 Installation
Power and load condition
Power supply: 24V DC, R≤(Us-12V)/Imax (kΩ)
Imax=23mA
Max. Voltage supply: 42V DC
Min. Voltage supply: 12V DC, 15V DC(Backlit LCD display)
Digital communication load resistance range: 230Ω~600Ω
Electric Connection
M20×1.5 cable sealing buckle, terminals are suitable for (0.5~2.5)mm2 wire.
Process connection
Standard process connection: NPT 1/2 female, which can be transferred to NPT 1/2 male, G1/2 male, M20×1.5
male and KF16 vacuum port etc.
4 Physical Specification
Material
Diaphragm: Stainless Steel 316L, Hast-alloy C
Process Connection: Stainless steel 316L
Filling liquid: silicone oil
Sealing ring: NBR, FKM, PTFE
Transmitter housing: Aluminum alloy material, epoxy resin glue spraying on the surface
Housing sealing ring: NBR
Nameplate: Stainless steel 304
- 18 -
Weight
1.6kg (not including LCD display, mounting support and process connection)
Housing protection
IP67
EMC Table
code Test terms Standard Test condition Performance degree
1 Radiated
interference(housing) GB/T 9254-2008 table5 30MHz~1000MHz qualified
2 Transmission interference
(DC power port) GB/T 9254-2008 table1 0.15MHz~30MHz qualified
3 ESD immunity GB/T 17626.2-2006 4kV(contact)
8kV(air) B
4 Radio frequency
ectromagnetic field immunity GB/T 17626.3-2006
10V/m
(80MHz~1GHz) A
5 Power frequency magnetic
field immunity GB/T 17626.8-2006 30A/m A
6 EFT immunity GB/T 17626.4-2008 2kV(5/50ns,5kHz) B
Note: (1) A degree:performance is normal within the technical standard range during testing.
(2) B degree: During testing, the function or performance is lowered or lost temporarily, but it could be recovered by itself. Actual
operation state, storage and data will keep the same.
Outline Dimension
Unit: mm
Horizontal Piping Installation (side view) Horizontal Piping Installation (front view)
06
- 19 -
Wall Installation Vertical Piping Installation
5 Electric connection chart
Note: the function of shortcut interface is equal to signal terminal.
6 Process connection instruction
6.1 Standard version(code 1)
Pressure port for range D/M/F/G/H/I/K/O Pressure port for range B/C/L
- 20 -
6.2 Derived threads
1/2 NPT male (code 2) M20×1.5 male (code 3)
G 1/2 male (code 4) Vacuum port DIN 28403 KF16 / ISO 2861(code 5)
1/4NPT male (code 6)
- 21 -
7 Model and Specification Code Table
-GP Intelligent Pressure Transmitter
-AP Intelligent Absolute Pressure Transmitter
Code Output
H 4mA~20mADC with HART
N 4mA~20mADC analog output
Code Pressure range
Gauge
B 0kPa…0.6kPa~6kPa(0mm…60mm~600 mm H2O)/(0mbar…6mbar~60mbar)
C 0kPa…2kPa~40kPa /(0mm…200mm~4000 mm H2O)/(0mbar…20mbar~400mbar)
D 0kPa…2.5kPa~250kPa /(0m…0.25m~25 m H2O)/(0mbar…25mbar~2500mbar)
F 0kPa…30kPa~3MPa /(0m…3m~300 m H2O)/(0bar…0.3bar~30bar)
G 0MPa…0.1MPa~10MPa /(0bar…1bar~100bar)
H 0MPa…0.21MPa~21MPa /(0bar…2.1bar~210bar)
I 0MPa…0.4MPa~40MPa /(0bar…4bar~400bar)
J 0MPa…0.6MPa~60MPa /(0bar…6bar~600bar)
Absolute
L 0kPa…2kPa~40kPa /(0mm…200mm~4000mm H2O)/(0mbar…20mbar~400mbar)
M 0kPa…2.5kPa~250kPa /(0mbar…25mbar~2500mbar)
O 0kPa…30kPa~3MPa /(0bar…0.3bar~30bar)
Code Diaphragm material Filling
A Stainless steel316L silicone oil
C Hastelloy C silicone oil
Code Process connection 1 1/2 NPT female 2 1/2 NPT male(with 1/4 NPT inside) 3 M20x1.5 male 4 G1/2 male 5 Vacuum connector DIN 28403 KF16 / ISO 2861[2] 6 1/4 NPT male
Code Process connection parts
N None
O No oil processing (For oxygen measurement: fluorocarbon oil filling, viton sealing ring )
Code Mounting bracket N None 1 Stainless steel 2 Galvanized Carbon Steel
Code Display
EJC
EJC
- 22 -
N No display 1 LCD 2 LCD with back-light
Code Others N None A Intrinsic safe D Exd
E Exd version with Explosion-proof cable joint
-DP H L A 1 N N 1 N The whole spec
Note: Vacuum port DIN 28403 KF16 / ISO 2861 is only suitable for pressure range within 2.5bar.
-DGP Bracket Installation Gauge Pressure Transmitter
-DAP Bracket Installation Absolute Pressure Transmitter
MDM3051S series Bracket Installation Gauge/Absolute Pressure Transmitter is used for
level, density and pressure measurement of liquid, gas and steam. Then it will outputs
4mA~20mA DC HART signal and also it could be connected to RST375 hand communicator or
RSM100Modem to do the specification setting and process control.
Standard Specification
(Standard zero as the reference calibration range, Stainless steel 316L diaphragm, filling liquid is silicone oil)
1 Performance Specification
Reference accuracy for range calibration Reference accuracy for range calibration(including linearity, hystersis and repeatability from zero): ± 0.075%
If TD>10 (TD=Max. Pressure range/calibration range), the accuracy is ±(0.0075×TD)%
Environmental Temperature Effect
Range Code -20 ℃~65 ℃ Total effect value
1B/1L ±(0.30×TD+0.20)%×Span
other ±(0.20×TD+0.10)%×Span
Range Code -40 ℃~-20 ℃和 65 ℃~85 ℃ Total effect value
1B/1L ±(0.30×TD+0.20)%×Span
other ±(0.20×TD+0.10)%×Span
Over range effect: ±0.075%×Span
EJC
EJC
EJC
- 23 -
Long-term stability
Range Code Effect value
1B/1L ±0.2%×Span/1 year
other ±0.1%×Span/1 year
Power effect ±0.001% /10V (12V~42V DC), negligible.
2 Functional Specification
Pressure range and limits(MDM3051S-DGP gauge pressure transmitter)
range/limits kPa bar
1B range 0.6~6 (6~60)mbar
limits -6~6 (-60~60)mbar
1C range 2~40 0.02~0.4
limits -40~40 -0.4~0.4
1D range 2.5~250 0.025~2.5
limits -100~250 -1~2.5
1F range 30~3000 0.3~30
limits -100~3000 -1~30
1G range (0.1~10)MPa 1~100
limits (-0.1~10)MPa -1~100
1H range (0.21~21)MPa 2.1~210
limits (-0.1~21)MPa -1~210
1I range (0.4~40)MPa 4~400
limits (-0.1~40)Mpa -1~400
Pressure range and limits(MDM3051S-DAP Absolute Pressure Transmitter)
range/limits kPa bar
1L range 2~40 0.02~0.4
limits 0~40 0~0.4
1M range 2.5~250 0.025~0.25
limits 0~250 0~0.25
1O range 30~3000 0.3~30
limits 0~3000 0~30
Pressure range limit
The pressure is adjustable within the upper and lower limit.
It is recommended to choose the range code with the lowest pressure range proportion to optimize the
performance specification.
- 24 -
Zero setting
The zero and pressure range could be adjust to any value within the measured rang in the table, only the
calibrated range≥Min. Range is valid.
Mounting position effect
The change of mounting position parallel to diaphragm could not influence the zero drift. If the angle between
mounting position and diaphragm is over 90° , the zero drift is <0.4kPa which could be calibrated by zero setting. No
effect on pressure range.
Output
2- wire, 4mA~20mA DC, HART communication protocol, linearity or square root output optional.
Output signal limit: Imin=3.9mA,Imax=20.5mA
Alarm Current
Low alarm mode(min.): 3.7mA
High alarm mode(max.): 21mA
No alarm mode(holding): holding the active current value before failure
Standard alarm current mode: High alarm mode
Response time
The damping constant of amplifier parts is 0.1s, time constant of sensor is 0.1s~1.6s, which is depended on the
pressure range and pressure range proportion. The additional adjustable time constant is 0.1s~60s.
Warm-up time
< 15s
Environmental temperature
-40 ℃~85 ℃
With LCD display and viton sealing ring, the temperature is -20 ℃~65 ℃.
Storage temperature/ transportation temperature
-50 ℃~85 ℃;with LCD display:-40 ℃~85 ℃
Pressure limit
It is from vacuum to Max. Pressure range.
Overpressure Limit
- 25 -
EMC
Please refer to next page “EMC table”
3 Installation
Power and load condition
Power supply: 24V DC, R≤(Us-12V)/Imax (kΩ)
Imax=23mA
Max. Voltage supply: 42V DC
Min. Voltage supply: 12V DC, 15V DC(Backlit LCD display)
Digital communication load resistance range: 230Ω~600Ω
Electric Connection
M20×1.5 cable sealing buckle, terminals are suitable for (0.5~2.5)mm2 wire.
Process connection
NPT 1/4 and UNF 7/16″ female at both sides of process connection flange.
4 Physical Specification
Material
Diaphragm: Stainless Steel 316L, Hast-alloy C
Process Connection: Stainless steel 316L
Filling liquid: silicone oil
Transmitter housing: Aluminum alloy material, epoxy resin glue sprays on the surface
Housing sealing ring: NBR
Pressure range 6kPa 40kPa 250kPa
(1B) (1C/1L) (1D/1M)
Overpressure
limit 16MPa 16MPa 16MPa
Pressure range 3MPa 10MPa 21MPa 40MPa
(1F/1O) (1G) (1H) (1I)
Overpressure
limit 16MPa 20MPa 25MPa 45MPa
- 26 -
Nameplate: Stainless steel 304
Weight
3.3kg(not including LCD display, mounting support and process connection)
Housing protection
IP67
EMC Table
code Test terms Standard Test condition Performance degree
1 Radiated
interference(housing) GB/T 9254-2008 table5 30MHz~1000MHz qualified
2 Transmission interference
(DC power port) GB/T 9254-2008 table1 0.15MHz~30MHz qualified
3 ESD immunity GB/T 17626.2-2006 4kV(contact)
8kV(air) B
4 Radio frequency
ectromagnetic field immunity GB/T 17626.3-2006
10V/m
(80MHz~1GHz) A
5 Power frequency magnetic
field immunity GB/T 17626.8-2006 30A/m A
6 EFT immunity GB/T 17626.4-2008 2kV(5/50ns,5kHz) B
Note: (1) A degree: performance is normal within the technical standard range during testing.
(2) B degree: During testing, the function or performance is lowered or lost temporarily, but it could be recovered by itself. Actual
operation state, storage and data will keep the same.
Outline Dimension
Unit:mm
Horizontal Piping Installation (side view) Horizontal Piping Installation (front view)
- 27 -
Wall Installation Vertical Piping Installation
5 Electric connection chart
Note: the function of shortcut interface is equal to signal terminal.
- 28 -
6 Process connection instruction
7 Model and Specification Code Table
-DGP Bracket Installation Intelligent Pressure Transmitter
-DAP Bracket Installation Intelligent Absolute Pressure Transmitter
Code Output
H 4mA~20mA DC with HART
N 4mA~20mA DC analog output
Code Pressure range
Gauge
1B 0kPa…0.6kPa~6kPa(0mm…60mm~600mm H2O)/(0mbar…6mbar~60mbar)
1C 0kPa…2kPa~40kPa /(0mm…200mm~4000mm H2O)/(0mbar…20mbar~400mbar)
1D 0kPa…2.5kPa~250kPa /(0m…0.25m~25m H2O)/(0mbar…25mbar~2500mbar)
1F 0kPa…30kPa~3MPa /(0m…3m~300m H2O)/(0bar…0.3bar~30bar)
1G 0MPa…0.1MPa~10MPa /(0bar…1bar~100bar)
1H 0MPa…0.21MPa~21MPa /(0bar…2.1bar~210bar)
1I 0MPa…0.4MPa~40MPa /(0bar…4bar~400bar)
Absolute
1L 0kPa…2kPa~40kPa /(0mm…200mm~4000mm H2O)/(0mbar…20mbar~400mbar)
1M 0kPa…2.5kPa~250kPa /(0mbar…25mbar~2500mbar)
1O 0kPa…30kPa~3MPa /(0bar…0.3bar~30bar)
Code Diaphragm material Filling
A Stainless steel316L silicone oil
C Hastelloy C silicone oil
Code Process connection N 1/4 NPT and 7/16 UNF thread hole,without release valve
B 1/4 NPT and 7/16 UNF thread hole, release valve mounting in the
end-face of flange back
EJC
EJC
- 29 -
U 1/4 NPT and 7/16 UNF thread hole, release valve mounting in upper
flange side
D 1/4 NPT and 7/16 UNF thread hole, release valve mounting in lower
flange side
Code Process connection parts N None
O No oil processing(For oxygen measurement: fluorocarbon oil filling, viton sealing ring )
Code Mounting bracket N None 1 Stainless steel 2 Galvanized Carbon Steel
Code Display N No display 1 LCD 2 LCD with back-light
Code Process connection parts N None
1 1/2 NPT Female with stainless steel oval flange
2 M20X1.5 male with stainless steel T joint
Code Others N None A Intrinsic safe D Exd
E Exd version with Explosion-proof cable joint
S stainless steel 316 plate
-DP H 1C A N N N 1 N N The whole spec
EJC