02 03 - (주)chc lab
TRANSCRIPT
05
www.chclab.com
04
Filtering efficiency rate is higher than 99.99% based on the minute particle of 0.3㎛, and this filtering efficiency is enhanced at the particles below and above 0.3㎛ by three principles; inertiar impaction, interception and diffusion.
The biological safety cabinet perfectly eliminates the possibility of contaminated air outflow from inside cabinet.It provides personnel, product and environment protections.
It prevents the direct inflow of the unpurified outside air into the inside of the chamber through the gap between the body and door sash when the front door is opened.
A blower supplying stable air flow assures the performance of low noise and non-vibration by optimizing parts and balance, and reduces the energy consumption.
The air current of the intake grille not to be disturbed by the researcher's arm, and perfectly blocks the UV light exposure when the door is closed.
Threshold
The sash alarm will be activated in case sash level is out of preset level, e.g. 8 inches height.
Width
Depth
Height
Frame height
Leveling caster height
Urethane caster height
<Top View> <Front View> <Side View>
1060/ 1385/ 1690/ 1995mm
810mm
1590mm
633mm
80mm
103mm
The work surface, which is easily disassembled from the equipment, is convenient for cleaning and maintenance.
Wide intake grille and reserved air hole perfectly prevent the cross contamination of the inside and outside of a work table. Wide armrest reduces the degree of researcher’s fatigue.
High performance blowerSash level
Intake grille & Armrest Non-filtered air protection hole
Biological Safety CabinetCHC LAB
Threshold structure &UV light leak free guard
Dished work surface
HEPA filter which eliminates the minute particles of 0.3㎛ by 99.99%Design
CHC-222A2 & CHC-777A2
biolus®
BIO SAFETY CABINET
TM
15
90
23
05
63
31
70
597
389 50
65
5
1060 / 1385 / 1690 / 1995
925 / 1250 / 1555 / 1860
06 07
www.chclab.com
Class II Type A2 | CHC-111A2 Class II Type A2 | CHC-222A2 Class II Type A2 | CHC-777A2 Class II Type B2 | CHC-888B2
PRODUCT LIST
Model No.CHC-111A2
04
Dimension(mm)
Width 1340
Depth 700
Height 2075
Color Gray
Door Tempered safety glass
Control panel LCD Display panel
Sash level 203 mm (8inch)
External material SPCC / AL with epoxy powder coated
Chamber material STS 304
Certification X
Avg. inflow velocity 0.3 ~ 0.4 m/s
Avg. Downflow velocity 0.25 ~ 0.35 m/s
Filter efficeincy HEPA filter > 99.995% at 0.3㎛
Noise level (dB) < 62
Nominal exhaust volume 312 CMH
Power supply 220V, 50/60Hz
FL lamp 36W X 1EA
UV lamp 30W X 1EA
Power receptacle 220V X 2EA
Utility valve 1EA (Vacuum)
Drain valve X
Differential pressure gauge O
Duct (Ø) X
Hard exhaust connection Option
Airthight damper X
Exhaust airflow sensor X
Model No.CHC-222A2
03 04 05 06
Dimension(mm)
Width 1060 1385 1690 1995
Depth 810
Height 2305
Color Gray
Door 6.38mm Laminated glass
Control panel LCD Display panel
Sash level 203 mm (8inch)
External material SPCC / AL with epoxy powder coated
Chamber material STS 304
Certification - EN, CE - EN, CE
Avg. inflow velocity 0.53 m/s (105fpm)
Avg. Downflow velocity 0.35 m/s (70fpm)
Filter efficeincy HEPA filter > 99.995% at 0.3㎛
Noise level (dB) < 64 < 65
Nominal exhaust volume 358 CMH 484 CMH 602 CMH 720 CMH
Power supply 220V, 50/60Hz
FL lamp 36W X 2EA
UV lamp 30W X 1EA
Power receptacle 220V X 2EA
Utility valve 1EA (Vacuum)
Drain valve Option
Differential pressure gauge O
Duct (Ø) X
Hard exhaust connection Option
Airthight damper X
Exhaust airflow sensor X
Model No.CHC-888B2
04 06
Dimension(mm)
Width 1385 1995
Depth 810
Height Leveling caster - 2540 / Fixed foot - 2503
Color Gray
Door 6.38mm Laminated glass
Control panel LCD Display panel
Sash level 203 mm (8inch)
External material SPCC / AL with epoxy powder coated
Chamber material STS 304
Certification NSF, UL
Avg. inflow velocity 0.53 m/s (105fpm)
Avg. Downflow velocity 0.30 m/s (60fpm)
Filter efficeincy HEPA filter > 99.995% at 0.3㎛
Noise level (dB) < 64 < 65
Nominal exhaust volume 1390 CMH 2092 CMH
Power supply 220V, 50/60Hz
FL lamp 36W X 2EA
UV lamp 30W X 1EA
Power receptacle 220V X 2EA
Utility valve 1EA (Vacuum)
Drain valve Option
Differential pressure gauge O
Duct (Ø) 200 or 250
Hard exhaust connection O
Airthight damper O
Exhaust airflow sensor O
Model No.CHC-777A2
03 04 05 06
Dimension(mm)
Width 1060 1385 1690 1995
Depth 810
Height 2305
Color Gray
Door 6.38mm Laminated glass
Control panel LCD Display panel
Sash level 203 mm (8inch)
External material SPCC / AL with epoxy powder coated
Chamber material STS 304
Certification NSF, UL, KS UL, KS NSF, UL, KS
Avg. inflow velocity 0.53 m/s (105fpm)
Avg. Downflow velocity 0.35 m/s (70fpm)
Filter efficeincy HEPA filter > 99.995% at 0.3㎛
Noise level (dB) < 64 < 65
Nominal exhaust volume 358 CMH 484 CMH 602 CMH 720 CMH
Power supply 220V, 50/60Hz
FL lamp 36W X 2EA
UV lamp 30W X 1EA
Power receptacle 220V X 2EA
Utility valve 1EA (Vacuum)
Drain valve Option
Differential pressure gauge O
Duct (Ø) X
Hard exhaust connection Option
Airthight damper X
Exhaust airflow sensor O
Biological Safety CabinetCHC LAB
● Room air● HEPA filtered air● Contaminated air under negative pressure● Contaminated air under positive pressure
● Room air● HEPA filtered air● Contaminated air under negative pressure● Contaminated air under positive pressure
● Room air● HEPA filtered air● Contaminated air under negative pressure● Contaminated air under positive pressure
● Room air● HEPA filtered air● Contaminated air under negative pressure
Airflow Airflow Airflow Airflow
09
www.chclab.com
08
Laminar flow cabinet provides only product protection for researcher. Biological safety cabinet provides personnel, product and environment.
Laminar flow cabinetClean air through HEPA filter flows inside chamber and protect product from contaminated air
Biological safety cabinet70% of the airflow at work space is re-circulated through the HEPA Filter, and 30% of it is exhausted to the outside through HEPA filter, and the polluted airflow is not leaked as perfectly surrounded by the negative pressure.
The product classified as BSL 1, harmless to human will be treated in laminar flow cabinet, and the product classified as BSL 2 or 3, high risk to human will be treated in biological safety cabinet.
Laminar flow cabinet provides primarily product protection from contamination. It is suitable when the specimen with comparatively less danger of contamination is treated, or suitable for the experiment of DNA, RNA, and PCR.
Laminar Flow CabinetCHC LAB
Laminar flow cabinet and Biological safety cabinet differences
Des
ign
Ergo
nom
ic
biolus®
LAMINAR FLOW CABINET
All the operations can be controlled with one controller.
■ Polluted air■ Clean air■ Particle
It can be easily moved to any desired places as the caster of stopper function is attached.
Electrical receptacle with safety cover is for the convenience of the experiment in work space. utility valves are available to be installed, gas, air, water and vacuum.
Differential pressure gaugeHEPA filter
Electrical receptacle &Utility valve LCD Display panelLeveling caster
Width
Depth
Height
Frame height
Leveling caster height
<Top View> <Front View> <Side View>
1040/ 1340/ 1650/ 1960mm
742mm
1231mm
633mm
82mm
Design
CLB-201 & CLB-201S By usage purose
By airflow
By product
It is available to check filter replacement time.
High efficiency particulate air filter
Particles of 0.3㎛ (dusts, bacteria, virus etc.) are removed by more than 99.99%
10
PRODUCT LIST
Model No.CLB-201
03 04 05 06
Dimension(mm)
Width 1040 1340 1650 1960
Depth 742
Height 1946
Color Gray
Door Tempered safety glass (5mm)
Control panel LCD Display panel
External material SPCC / AL with epoxy powder coated
Worktop STS 304
Support frame SPP
Certification CE
Avg. Downflow velocity 0.3~0.4 m/s (Variable set speed)
Filter efficeincy HEPA filter > 99.995% at 0.3㎛
Pre-filter Nylon filter
Noise level (dB) ≤ 65 dB(A)
Safety alarm UV timer
Power supply 220V, 50/60Hz
FL lamp 36W X 1EA 36W X 2EA
UV lamp 20W X 1EA 30W X 1EA
Power receptacle 220V X 1EA 220V X 2EA
Utility valve Option
Utility hole Option
Differential pressure gauge O
Model No.CLB-201S
03 04 05 06
Dimension(mm)
Width 1040 1340 1650 1960
Depth 742
Height 1946
Color Gray
Door Tempered safety glass (5mm)
Control panel LCD Display panel
External material SPCC / AL with epoxy powder coated
Worktop STS 304
Support frame SPP
Certification CE
Avg. Downflow velocity 0.3~0.4 m/s (Variable set speed)
Filter efficeincy HEPA filter > 99.995% at 0.3㎛
Pre-filter Nylon filter
Noise level (dB) ≤ 65 dB(A)
Safety alarm UV timer
Power supply 220V, 50/60Hz
FL lamp 36W X 1EA 36W X 2EA
UV lamp 20W X 1EA 30W X 1EA
Power receptacle 220V X 1EA 220V X 2EA
Utility valve Option
Utility hole Option
Differential pressure gauge O
Model No.CLB-202D
03 04 05 06
Dimension(mm)
Width 1040 1340 1650 1960
Depth 777
Height 2015
Color Gray
Door Basic - Tempered glass / Option - Acrylic panel door
Control panel LCD Display panel
External material SPCC / AL with epoxy powder coated
Worktop STS 304
Support frame SPP
Certification CE
Avg. Downflow velocity 0.3~0.4 m/s (Variable set speed)
Filter efficeincy HEPA filter > 99.995% at 0.3㎛
Pre-filter Nylon filter
Noise level (dB) ≤ 65 dB(A)
Safety alarm UV timer
Power supply 220V, 50/60Hz
FL lamp 36W X 1EA 36W X 2EA
UV lamp 20W X 1EA 30W X 1EA
Power receptacle 220V X 2EA
Utility valve Option
Utility hole Option
Differential pressure gauge O
Model No.CLB-060F
03 04
Dimension(mm)
Width 906 1206
Depth 660
Height 1876
Color Gray
Door Front folding sash, (Acrylic 8mm)
Control panel LCD Display panel
External material Steel plate with epoxy powder coated
Worktop STS 304
Support frame SPP
Certification CE
Avg. Downflow velocity 0.3~0.4 m/s (Variable set speed)
Filter efficeincy HEPA filter > 99.995% at 0.3㎛
Pre-filter Nylon filter
Noise level (dB) ≤ 65 dB(A)
Safety alarm UV timer
Power supply 220V, 50/60Hz
FL lamp 36W X 1EA 36W X 2EA
UV lamp 10W X 1EA 10W X 2EA
Power receptacle 220V X 1EA
Utility valve X
Utility hole X
Differential pressure gauge X
● Room air● HEPA filtered air
● Room air● HEPA filtered air
● Room air● HEPA filtered air
● Room air● HEPA filtered air
11
www.chclab.com
Vertical Laminar Flow Cabinet Vertical Laminar Flow Cabinet Horizontal Laminar Flow Cabinet Laminar Flow Cabinet
Airflow Airflow Airflow Airflow
Laminar Flow CabinetCHC LAB
WorkstationCHC LAB
A-STWS-900 A-STWS-1100
A-STWS-900F A-STWS-1100F
It is mainly used in animal laboratory to remove harmful gas or odor in order to protect the researcher.
Simple washing is available during the experiment as cupsink and water tap are installed.
Applying the three-sided acrylic observation, it is so easy to obtain the best possible view inside.
Cup Sink & Water Tap Three-sided Acrylic
Unit : mm
12
airos® Safrty Trimming
Workstation
Airflow
● Room Air● Fume (gas, vapor)● Clean Air
13
Lab Furniture CHC LAB
Finter Box (CLE-030)
Dimension500 X 500 X 865 (WxDxH)FilterCarbon / HEPA / Pre-filter
Safety Trimming Workstation
Width Depth Height
906 / 1106 560 1405
Safety Trimming Workstationwith Filter Box
Width Depth Height
906 / 1106 560 1405
Ergonomic
Lab ChairBimos, laboratory chair with European design provides comfort for a long time using ergonomic design.
Seat height adjustment
Auto-motion technology Seat height adjustment Backrest height adjustment
Seat swivel range of 20ㅇ
to the left and right
Functions
Lab Stool (Backrest)
Labster Fin
Lab Stool
Cleanroom Plus
Black
BlackBlack
BlackGrayWhiteGray
Order No.White : 9103-6907Gray : 9103-6911
Model No.Black : CLF-932
Order No.Black : 9144-2000Gray : 9144-2002
Model No.Black : CLF-925
Order No.Black : 9181-2571
Features
FRP (Clean Room Only)
White
Model No.White : Decois-8170
CHC LAB Co.,Ltd. Product List
Lab Furniture hulab® - LEVEL PRO® Lab table
- HARD PRO® Lab table
- COZY PRO® Lab table
Ventilation System airos® - Bypass air fume hood - Air curtain fume hood - Auxiliary air fume hood - Low work-in fume hood - Walk-in fume hood - Center tabletop fume hood - Side tabletop fume hood - Arm hood
Filtering System cleair® - Ductless fume hood - Chemical storage cabinet (2in1 filter type) - Chemical storage cabinet (Ion cluster type)
Bio Equipment biolus® - Biological safety cabinet Class I - Biological safety cabinet Class II Type A2 - Biological safety cabinet Class II Type B2 - Vertical laminar flow cabinet - Horizontal laminar flow cabinet - Laminar flow cabinet - PCR Workstation
Workstation airos® - Safety balance workstation - Safety trimming workstation
CHC LAB Co.,Ltd. Head office & factory139, Techno 2-ro, Yuseong-gu, Daejeon, KoreaTel.+82-42-933-0036 Fax.+82-42-933-0039
Chungwon factory336-8, Biryong-ri, Nami-myeon, Seowon-gu, Cheongju-si, Chungbu, KoreaTel.+82-43-277-8475 Fax.+82-43-277-8476
Seoul Branch office#207, 140, Beolmal-ro, Dongan-gu, Anyang-si, Gyeonggi-do, KoreaTel.+82-31-423-5530 Fax.+82-31-423-5542
Web Sitehttp://www.chclab.com
Social Networkhttp://www.facebook.com/chclab | http://www.twitter.com/chchulab
Bloghttp://blog.naver.com/chclabblog