04 - artifacts 2011

Upload: jose

Post on 04-Jun-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/14/2019 04 - Artifacts 2011

    1/15

    Radiographic Artifacts

    NOT ESPECIALLY

    OLD ORVALUABLE

    Chris Ober, DVM,PhD, DACVR

    14 February 2011

    Artifacts Defined

    Any opacity on the radiograph that does

    not correspond to an actual anatomicstructure

    Any misrepresentation of an actual

    anatomic structure

    Anything decreasing radiographic quality

    Quick CR primer

    Cassette exposed by x-

    rays; storage phosphors

    excited

    Quick CR primer

    Cassette exposed by x-

    rays; storage phosphors

    excited

    Cassette put into reader

    Quick CR primer

    Laser shines onto

    detectors

    Quick CR primer

    Laser shines onto

    detectors

    Excited phosphors give

    off light photons

  • 8/14/2019 04 - Artifacts 2011

    2/15

    Quick CR primer

    Laser shines onto

    detectors

    Excited phosphors give

    off light photons

    Light photons travel

    through light guide

    Quick CR primer

    Laser shines onto

    detectors

    Excited phosphors give

    off light photons

    Light photons travel

    through light guide

    Image produced by

    computer

    Quick DR primer

    Panel exposed by x-

    rays, detectors

    excited

    Data sent directly to

    computer via cord

    Image created

    Sequence: Film-Screen

    Film storage &

    handling

    Positioning

    Exposure

    Film handling

    Film processing

    Manual

    Automatic

    Image archiving

    Sequence: CR (With Cassette)

    Cassette storage

    Positioning

    Exposure Postexposure

    Cassette reading

    Workstation

    Sequence: DR (No Cassette)

    Panel storage

    Positioning

    Exposure Image reading &

    transfer

    Workstation

  • 8/14/2019 04 - Artifacts 2011

    3/15

    Another Classification System

    Color

    Black White

    Distribution

    Focal

    Regional or Global

    Storage & Pre-Exposure Handling

    Artifacts

    Screen-Film

    Pressure Abrasions & Scratches

    Fingerprints

    Static electricity

    Material in cassette

    Fog

    Light leak

    CR & DR

    Storage scatter (CR)

    Pressure Artifacts

    Film Crease Crescents (SF)

    Black & focal

    Cause: rough

    handling

    Pressure from crease

    causes activation of

    crystals

    NOT caused by

    fingernail (usually)

    Pressure Artifacts

    Film Crease Crescents (SF)

    Abrasions & Scratches (SF)

    Black or white, focal

    Cause: rough handling

    Black: pressure from

    abrasion activatescrystals

    White: emulsion has

    been removed (scratch)

    Can feel texture in

    emulsion

    Fingerprints (SF)

    Focal

    White: most common

    Oil on fingertips blocksdeveloper

    OR

    Fixer on fingertips

  • 8/14/2019 04 - Artifacts 2011

    4/15

    Fingerprints (SF)

    Focal

    White: most common

    Oil on fingertips blocks

    developer

    OR

    Fixer on fingertips

    Fingerprints (SF)

    Black: less common

    Developer onfingertips

    Prevent:

    Keep fingers dry and

    clean

    Touch only film edges

    Hogge JP, et al.

    RadioGraphics 1999.

    Static Electricity (SF)

    Black & focal

    Two patterns

    Branching

    Smudge / Spot

    Cause: electrical

    discharge

    Static Electricity (SF)

    Prevent:

    Moderate humidity

    Avoid friction while

    handling film

    Antistatic cassette

    cleaners

    Static Electricity (SF)

    Hogge JP, et al.

    RadioGraphics 1999.

    Material in Cassette /

    Dirty Screen (SF) White & focal

    Sharp margins (close

    to film no

    penumbra) Cause: Visible light

    photons from screen

    cant reach and

    expose film emulsion

  • 8/14/2019 04 - Artifacts 2011

    5/15

    Material in Cassette /

    Dirty Screen (SF) THUS radiolucent

    material can stillcause this artifact

    Common causes

    Hair

    Dust

    Fog (SF)

    Increased blackness,

    usually regional orglobal

    Definition: any

    unwanted film

    exposure /

    development

    Decreased image

    contrast and detail

    Fog (SF)

    Causes:

    Light exposure

    Heat

    (store at < 68 F)

    Humidity

    (keep at 30-50%)

    Radiation (including

    scatter)

    Chemicals

    Old film

    Fog (SF)

    Hogge JP, et al.

    RadioGraphics 1999.

    Light Leak (SF)

    Black, focal to global

    A form of fog

    Sources:

    Storage bin (along 1edge)

    Cassette not closed (1

    edge or corner)

    Overhead light

    Safelight filter crack

    Storage Scatter (CR)

    Black & global

    Like fog unwanted

    exposure of CR plate

    Causes: Background radiation

    Scatter from imaging

    procedure

    Jimnez DA, et al.

    Vet Radiol Ultrasound 2008.

  • 8/14/2019 04 - Artifacts 2011

    6/15

    Positioning Artifacts

    Screen-Film

    Magnification Foreshortening /

    distortion

    Patient rotation

    Upside-down cassette

    Grid cutoff

    CR & DR

    Magnification Foreshortening /

    distortion

    Patient rotation

    Upside-down cassette

    (CR) or panel (DR)

    Grid cutoff

    Magnification

    Object closer to film:

    Closer to normal size Sharper margins

    Object farther from

    film:

    Magnified

    Less distinct margins

    Foreshortening / Distortion

    X-ray beam does not

    pass perpendicular to

    long axis of structure

    Differential

    magnification of

    structure

    Foreshortening / Distortion

    Patient Rotation

    A form of distortion

    Lesions may bemasked by atypical

    superimposition onnormal anatomy

    Unusual projection ofnormal anatomy maybe incorrectlydiagnosed asabnormal

    Upside-down Cassette (SF)

    White & focal

    Springs & latches onback of cassette block

    x-ray beam

    Global lightness

    Lead backing absorbssome x-ray photons(film appearsunderexposed)

  • 8/14/2019 04 - Artifacts 2011

    7/15

    Upside-down Cassette (SF) Upside-down Cassette (SF)

    Upside-down Cassette (CR)

    White & multifocal

    Construction of

    cassette back

    absorbs x-rays

    Pattern depends on

    manufacturer

    Upside-down Cassette (CR)

    Upside-down Panel (DR)

    White & multifocal

    Electronics in the

    back of the panel

    absorb x-rays

    Grid Cutoff (SF, CR, DR)

    White, regional to

    global

    Orientation of lead

    strips in grid requiresappropriate geometry

    relative to x-ray beam

    Wrong geometry

    lead strips attenuate

    more x-rays

  • 8/14/2019 04 - Artifacts 2011

    8/15

    Laterally Decentered

    Focused Grid

    Result: Too white across

    entire image

    Off-Level Focused Grid

    Result: Too white across

    entire image

    Upside-Down Focused Grid

    Result: OK in the center,Severely too white at periphery

    Distance Decentered

    Focused Grid Similar to upside-

    down grid

    OK in the center,

    Markedly too white at

    periphery

    Often not as severe

    as upside-down grid

    Combined Lateral and Distance

    Decentering of Grid

    Result: OK on one side,

    Gradually whiter toward

    other side

    Exposure Artifacts

    Screen-Film

    Motion

    Double exposure

    Overexposure

    Underexposure

    Material obstructing x-ray

    beam

    CR & DR

    Motion

    Double exposure (CR)

    Overexposure(Saturation)

    Underexposure

    Material obstructing x-ray

    beam

  • 8/14/2019 04 - Artifacts 2011

    9/15

    Motion (SF, CR, DR)

    Blurring of moving

    parts Exposure does have

    a finite interval

    Reduce with:

    Better restraint

    Sedation

    Decreased exposure

    time

    Motion (SF, CR, DR)

    Double Exposure (SF, CR)

    Appearance of 2

    summated images

    Cause: Cassette

    inadvertently exposed

    more than once

    Press exposure trigger

    twice

    Forget to change

    cassette after first view

    Double Exposure (SF, CR)

    SF: Black, global

    2 exposures =

    overexposure artifact

    CR: Normal grayscale

    Wide latitude can

    usually accommodate

    the overexposure

    DR: Cant happen

    Image is transferred to

    computer immediately

    Double Exposure (SF, CR) Overexposure (SF)

    Black & global

    Patient info label

    normal

    Cause: Too muchactivation of emulsion

    crystals

    High mAs or kVp

    Line surge

    Double exposure, Fog

  • 8/14/2019 04 - Artifacts 2011

    10/15

    Saturation (CR, DR)

    Black, may only beapparent in parts ofimage

    Just like SFoverexposure

    High exposure maxesout sensitivity rangeof detector

    Anatomy (esp. thinnerparts) may be burnedthrough

    Saturation (CR, DR)

    Black, may only beapparent in parts ofimage

    Just like SFoverexposure

    High exposure maxesout sensitivity rangeof detector

    Anatomy (esp. thinnerparts) may be burnedthrough

    Paradoxical Overexposure,

    Saturation, & Planking (DR)

    Paradoxical Overexposure,

    Saturation, & Planking (DR)

    Underexposure (SF)

    White & global

    Patient info label

    normal

    Cause: Too littleactivation of emulsion

    crystals

    Low mAs or kVp

    General x-ray

    obstruction

    X-ray tube troubles

    Underexposure:

    Quantum Mottle (CR, DR) Noisy & global

    Computer algorithms

    adjust image to

    remain gray Low signal level

    causes grainy

    appearance and poor

    contrast & sharpness

    Cause: As with SF

  • 8/14/2019 04 - Artifacts 2011

    11/15

    Underexposure:

    Quantum Mottle (CR, DR)Material Obstructing

    X-Ray Beam (SF, CR, DR)

    White, focal

    Cause: Anyradiopaque structure

    in the path of the x-

    ray beam

    On patient

    In patient

    On table

    On cassette

    Material Obstructing

    X-Ray Beam (SF, CR, DR)

    OMG!

    Its a fracture fragment!

    Material Obstructing

    X-Ray Beam (SF, CR, DR)

    or gravel on the table.

    Material Obstructing

    X-Ray Beam (SF, CR, DR)

    Material Obstructing

    X-Ray Beam (SF, CR, DR)

  • 8/14/2019 04 - Artifacts 2011

    12/15

  • 8/14/2019 04 - Artifacts 2011

    13/15

    Insufficient Mixing (SF) Inadequate Fixation (SF)

    Initially film is cloudy

    and milky Over time appears

    yellow/brown

    Dichroic stain

    Incomplete Washing / Rinsing (SF)

    Fixer remains

    Cloudy, sticky residue

    Yellow-brown stain

    Sulfur smell

    Kissing Defects (SF)

    Dirty Light Guide (CR)

    White & focal

    Sharp white line

    Cause:

    Light emitted fromplate reaches lightguide

    Dirt blocks path for anentire line of data

    No light along that line= assumption thatthere was no exposure

    Dirty Light Guide (CR)

    White & focal

    Sharp white line

    Cause:

    Light emitted fromplate reaches lightguide

    Dirt blocks path for anentire line of data

    No light along that line= assumption thatthere was no exposure

  • 8/14/2019 04 - Artifacts 2011

    14/15

    Dirty Light Guide (CR)

    White & focal

    Sharp white line Cause:

    Light emitted fromplate reaches lightguide

    Dirt blocks path for anentire line of data

    No light along that line= assumption thatthere was no exposure

    Cassette Debris (CR)

    White & focal

    Sharp margins Essentially the same

    as FS dirty cassette

    EXCEPT

    Cassette Debris (CR)

    Problem is that

    photons emitted from

    plate during reading

    dont reach

    photomultiplier tube

    (Contrast with FS,

    where light photons

    from screen dont

    reach film)

    Workstation Artifacts

    Screen-Film

    Not Applicable

    CR & DR

    Faulty Transfer (DR)

    Clipping

    Planking

    Uberschwinger

    Faulty Transfer (CR, DR)

    Appearance: Anything

    goes, generally

    distorted anatomy

    Cause: Problems indata transfer to

    workstation

    Memory problems

    Communication errors

    (e.g. loose cables)

    Power fluctuation

    Faulty Transfer (DR)

  • 8/14/2019 04 - Artifacts 2011

    15/15

    Uberschwinger (CR, DR)

    Black & focal

    Black lines occuraround dense

    structures

    Can be mistaken for

    loosening or infection

    of implants

    Cause: Certain

    processing algorithms

    (edge-enhancement)

    Uberschwinger (CR, DR)

    What Have We Learned?

    People can manage to screw up any stage

    of the radiographic process.

    Recognizing artifacts and when they occur

    is the only way to make good radiographs.

    Look up Mr. Yuk on Wikipedia for a video

    that scarred my childhood.