07/27/2004 XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

Upload: lorena-chapman

Post on 31-Dec-2015

215 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery

Vadim Sajaev

Advanced Photon Source

Argonne National Laboratory

Page 2: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Theory

• Each particle in the bunch radiates an electromagnetic pulse e(t)

• Total radiated field is

• Fourier transform of the field is

• Power spectrum of the radiation is

N

kkttetE

1

)()(

N

k

ti keeE1

)(ˆ)(ˆ

N

mk

ttiN

m

tiN

k

ti mkmk eeeeeeP1,

)(2

11

* )(ˆ)(ˆ)(ˆ)(

Page 3: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Average power

Average power is

222

)(ˆ)(ˆ)( fNNeP

The coherent radiation term carries information about the distribution of the beam at low frequencies of the order of -1

coherent radiation

incoherent radiation

Page 4: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Difference between coherent and incoherent power is huge

• consider a Gaussian beam with t=1 ps and total charge of

1 nC (approximately 1010 electrons)

2

2

2

2

1)( t

t

t

etf

2

22

)(ˆt

ef

and

• Coherent to incoherent ratio2

)(ˆ)( fNR

At 1 THz: R1010

At 10 THz: R10-

34

Page 5: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

High frequencies still contain information

• Power spectrum before averaging:

• Each separate term of the summation oscillates with the period =2/(tk-tm)~2/t

• Because of the random distribution of particles in the bunch, the summation fluctuates randomly as a function of frequency .

N

mk

tti mkeeP1,

)(2)(ˆ)(

Page 6: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Single electron Electron bunch

N

kkttetE

1

)()(e(t)

Spectrometer

N

k

ti keeE1

)()(

~10 fs ~1 ps

Example: incoherent radiation in a wiggler

Spike width is inversely proportional to the bunch length

Page 7: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Bunch profile measurements using fluctuations of incoherent radiation

• The method was proposed by M. Zolotorev and G. Stupakov and also by E. Saldin, E. Schneidmiller and M. Yurkov

• Emission can be produced by any kind of incoherent radiation: synchrotron radiation in a bend or wiggler, transition, Cerenkov, etc.

• The method does not set any conditions on the bandwidth of the radiation

Page 8: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Quantitative analysis

N

qpmk

ttitti qpmkeeePP1,,,

)()(22)(ˆ)(ˆ)()(

2222 )(ˆ1)(ˆ)(ˆ)()( feeNPP

2

)(ˆ12

1)( fg

We can calculate autocorrelation of the spectrum:

Page 9: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Variance of the Fourier transform of the spectrum

Fourier transform of the spectrum:

Its variance:

It can be shown, that the variance is related to the convolution function of the particle distribution:

dePG i)()(

2)()()( GGD

dttftfAD )()()(

Page 10: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Some of the limitations

• Bandwidth of the radiation has to be larger than the spike width

• In order to neglect quantum fluctuations, number of photons has to be large

• Transverse bunch size – radiation has to be fully coherent to observe 100% intensity fluctuations

2rad

eph Nn2

1

Page 11: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

LEUTL at APS

Page 12: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Spectrometer

 Grating

Grooves/mm 600

Curv. radius [mm] 1000

Blaze wavelength[nm] 482

CCD camera

Number of pixels 1100330

Pixel size [m] 24

Concave mirror curv. radius [mm] 4000

Spectral resolution [Å] 0.4

Bandpass [nm] 44

Resolving power at 530 nm 10000

Wavelength range [nm] 250 – 1100

Page 13: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Single shot spectrumTypical single-shot spectrum

Average spectrum

Page 14: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Spectrum for different bunch length

Long 2-ps rms bunch Short 0.4-ps rms bunch

Note: Total spectrum width (defined by the number of poles in the wiggler) is barely enough for the short bunch.

Page 15: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Spectrum correlation

i

ii

niin PPPC 2)(/)()(

Page 16: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Bunch length

From the plot the correlation width is 2 pixels.

Frequency step corresponding to one pixel is 2.4·1011 rad/s.

Assuming the beam to be Gaussian, from equation

2

)(ˆ12

1)( fg

we get

psnb 2

1

Page 17: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Convolution of the bunch profile

ch

ch

N

m

Nimknmnk ePG

1

/2,,

p pN

m

N

nnk

pmkk G

NGD

1

2

1,,

1

2 0 20

1

Step functionGaussian

Bunch profile

Page 18: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Convolution recovered from the measurements

t 2

Convolution of the Gaussian is also a Gaussian with

psb 8.1

The Gaussian fit gives us

Page 19: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Phase retrieval

022

)(/)(ln2)(

x

xdxPm

0

)(cos)(1

)(c

zd

czS m

The amplitude and the phase information of the radiation source can be recovered by applying a Kramers-Kronig relation to the convolution function in combination with the minimal phase approach.

Page 20: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Phase retrieval example

Calculation of longitudinal distribution for different bunches

Time (arb. units)

Calculated shape

Exact shape

Page 21: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Bunch profile

Two different measurements (two sets of 100 single-shot spectra)

FWHM4ps

Page 22: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Conclusions

• Measurements of incoherent radiation spectrum showing intensity fluctuations were done.

• A technique for recovering a longitudinal bunch profile from spectral fluctuations of incoherent radiation has been implemented. Although we used synchrotron radiation, the nature of the radiation is not important.

• Typically, analysis of many single shots is required, however one can perform statistical analysis over wide spectral intervals in a single pulse

Page 23: 07/27/2004XFEL 2004 Measurement of Incoherent Radiation Fluctuations and Bunch Profile Recovery Vadim Sajaev Advanced Photon Source Argonne National Laboratory

07/27/2004 XFEL 2004

Conclusions

• An important feature of the method is that it can be used for bunches with lengths varying from a centimeter to tens of microns (30 ps – 30 fs)

• There are several important conditions for this technique. In order to be able to measure a bunch of length t, the

spectral resolution of the spectrometer should be comparable with 1/t. Also, the spectral width of the

radiation and the spectrometer must be larger than the inverse bunch length