rev. 02/2010 I N N O V A T I O N A N D T E C H N O L O G Y F O R B E T T E R P A T I E N T O U T C O M E S ORTHOPAEDIC DRAINAGE BLOOD MANAGEMENT Quick User’s Guideline This document is not intended to replace the ORTHO P.A.S. Dry Wash user’s manual, it is for quick reference only. For complete ORTHO P.A.S. Dry Wash set up and procedure please refer to its User’s manual. Close both clamps Innovation and Technology for Better Patient Outcomes www.eurosets.it Remove caps (male and female) from the bag Remove bag and filter. The system is now working as drainage only 4 2 3 disconnect the bag from the reservoir and close the connections END OF REINFUSION 1 Turn vacuum to max, turn lever to “pause” , disconnect vacuum pump REDON FUNCTIONING

Upload: vankhue

Post on 28-Apr-2018

215 views

Category:

Documents


2 download

TRANSCRIPT

rev.

02/

2010

I N N O V A T I O N A N D T E C H N O L O G Y F O R B E T T E R P A T I E N T O U T C O M E S

ORTHOPAEDIC DRAINAGE

BLOOD MANAGEMENT

ORTHO P.A.S.DRY WASH

Quick User’s Guideline

This document is not intended to replace the ORTHO P.A.S. Dry Wash user’s manual, it is for quick reference only.

For complete ORTHO P.A.S. Dry Wash set up and procedure please refer to its User’s manual.

Close both clamps

Innovation and Technology for Better Patient Outcomes

www.eurosets.it

Remove caps (male and female)

from the bag

Remove bag and filter. The system is now working

as drainage only

42 3

disconnect the bag from the reservoir and close the connections

END OF REINFUSION

1

Turn vacuum to max, turn lever to “pause”, disconnect vacuum pump

REDON FUNCTIONING

Remove the green pouchKeep the transparent pouch

1

A - Place the Quality Sensor to the reservoir.

B - Open the cover

1

Move Ortho P.A.S. Dry Wash out of the sterile field. Remove the tran-sparent pouch. Close the clamp.

4

Turn the lever to the left (“vent” position). The yellow

indicator becomes visible

4

Turn the pump on and adjust the vacuum Close the clamp

7 7

cut the fingers connector according to the

drainage size in use

2

A - Slide the bag out from the pouch at the back of the reservoir. B - Put the red tube inside the clamp

C - Hang the bag

2

Connect the vacuum pump: holding the black ring down,

slide the pump up

5

A - Turn on Quality SensorB - Open the clamp

5

The yellow indicator disappears

(vacuum ON)

8

Turn the lever back to “suction” position.

Yellow indicator disappears

8

write down the blood collection

starting date and time

6

Press Start Procedure. The blood moves into the bag. Supernatant and fat remain in the reservoir

6

Hang Ortho P.A.S. Dry Wash to the bed.

System working

Remove Quality Sensor at the end of the procedure

9

REINFUSIONSET UP

Connect the drainage to the fingers connector

3

Close the cover

Transfer Ortho P.A.S. Dry Wash

to the sterile field.

3

9

In case of multiple reinfusion, repeat the complete sequence without disconnecting the Goccia filter, neither from the bag, nor from the Patient

Take the bag andopen the connector

10

Spikez the Goccia 40mm filter

11

Prime the Goccia 40mm filter and the reinfusion line. Reinfuse the blood*

12

*Th

e b

loo

d s

ho

uld

be

rein

fuse

d a

cco

rdin

g to

yo

ur H

osp

ital

’s Pr

oto

col

1 Holding system (hook)

2 Open/close button

3 RBC’s sensor + automatic clamp

4 Hook for transfer bag

5 Procedure key-pad