1 Formulating Technique Charts • Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation exposure to the patient. • Remember that over half of the repeated films are due to errors in the technical factors resulting in under or over exposed films.

Upload: makena-postlewait

Post on 29-Mar-2015

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

1

Formulating Technique Charts

• Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation exposure to the patient.

• Remember that over half of the repeated films are due to errors in the technical factors resulting in under or over exposed films.

Page 2: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

2

Formulating Technique Charts

• Used equipment generally do not come with technique charts.

• If you buy someone’s practice, you may not get accurate charts.

• The complete x-ray chain must be in proper working order in order to establish accurate charts. Acceptance testing is a must!!!

Page 3: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

3

Sources of Technique Charts

• If you have high frequency equipment, the Nolan Filter System and the screen speed matches those described in the text, baseline charts can be used based upon those in the text book.

Page 4: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

4

Sources of Technique Charts

• Major x-ray equipment manufacturers may provide help with making technique charts for their equipment.– Anatomically programmed generators have

charts built into the unit. Make sure that they can be easily modified.

– Smaller dealer sold companies may not have the resources to provide much help.

Page 5: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

5

Sources of Technique Charts

• Major film and screen manufacturers can be a source for help producing technique charts.– The screens and film must match in order to

get consistent results. The manufacturer can be a great help in this area.

– It is in the film manufacturer’s best interest that your films are of consistently high quality.

Page 6: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

6

Types of Charts

• Fixed kVp charts are recommended because:– Optimum kVp for the body part is used– Consistent contrast is maintained– Reduced exposure for the patient– Greater latitude of exposure.

• Fixed mAs charts lack these qualities.

Page 7: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

7

Getting Started

• Make sure that the x-ray machine is properly calibrated across all mA stations.

• Make sure the processor is working properly and has fresh chemistry.

• The x-ray machine type (single phase, three phase or high frequency) will have a significant impact on the charts.

Page 8: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

8

Getting Started

• The method for entering factors will affect the appearance of the chart.– Do you enter mAs or mA and time?– What focal spot and mA selections are

available?– Do you have power selection capabilities?– Does the unit have line voltage adjustment?

Page 9: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

9

Getting Started

– Do you have a non-Bucky film holder?– What is the grid ratio of the Bucky?– What is the Relative Speed Value of the

cassettes and film used?– What additional information do you want on

the charts?• SID, Tube Angles, Film Size or I.D. Location?

Page 10: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

10

mAs Chart

• Units where the mA and time are set independently need a mAs Chart.

• You can change the mA stations and time to help control motion and get the same mAs.

Page 11: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

11

Hand Written Charts

• This beats nothing.• If may be all you get

with a used unit.• It is a good starting

point.• This is all we had at

Palmer when I arrived and there were no extremity charts !

Page 12: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

12

Complete Technique Charts

• Chart should be located in the control booth and easy to read.

• This will help avoid technique errors.

• They will work fine for 85% of the patients.

Page 13: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

13

Adjustment of Factors

• Patient body habitus will greatly impact the technical factors.– Fit muscular patient will require higher kVp

and sometime increased mAs also.– Fatty tissue will need an increase in mAs and

no increase in kVp.– Frail or elderly patients need a reduction in

kVp.

Page 14: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

14

Supertech Technique Calculator

• The Supertech Technique Calculator can be used to compute the technical factors.

• It consists of the slide rule, a penetrometer and calibration film.

Page 15: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

15

Supertech Technique Calculator

• An exposure is made of the penetrometer for each mA station and cassette speed used.

• This can be used to check mA calibration and different factor established for each mA station.

Page 16: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

16

Supertech Technique Calculator

• The correction factor is used to adjust the calculator for the equipment type, calibration and film speed.

Page 17: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

17

Supertech Technique Calculator

• With the correction factors are set, just enter the patient measurements on one side.

• Turn it over and read the combinations of mAs and kVp that will produce a satisfactory image.

Page 18: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

18

Supertech Technique Calculator

• Addition correction factors are provided for patient body habitus and changes in SID or film speed.

• The Supertech Technique Calculator works !!

Page 19: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

19

Supertech Technique Program

• A computer program is also available. Supertech is now available for the Palm Pilot

• By entering the machine correction factor, technique charts can be printed.

Page 20: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

20

Supertech®: RadiographicTechnique Calculator

Gary M. Guebert, D.C., DACBR™

Page 21: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

21

The Calculator

Page 22: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

22

“Houston, we have a problem”

• FDA – CRCPD– NEXT studies

• Missouri MORSELS study

Page 23: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

23

The body parts

Page 24: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

24

Choose a body part

Page 25: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

25

Enter the appropriate TCF

Page 26: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

26

What is a TCF?

• It takes into account all of the variables involved:– X-ray generating system

• Single phase vs high frequency

– Film/screen speed• 100, 400/800, 600/1200

– Grid ratio• Non grid, 8:1, 10:1, 12:1

– Focal film distance changes• 40” or 72”

– Processing variables• Temperature, chemical concentration

Page 27: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

27

Turn the Supertech® over

Page 28: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

28

Select kVp, mA, time

Page 29: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

29

mA stations

Page 30: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

30

How to establish your TCF

• Work backwards from an existing good radiograph

• Use the penetrometer

Page 31: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

31

Work backwards from an existing good radiograph

• From the exposure log find the part size– e.g. 21 cm lumbar spine, AP

• You cannot set the TCF at this point• Turn the Supertech® over• Find the time & mA (or mAs used)• Line up the kVp next to that mAs value• Turn the Supertech® over again and there

is your TCF for THAT EXPOSURE

Page 32: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

32

Making changes

• Focal film distance

• Grid ratio

• Body habitus

• Casts

Page 33: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

33

Changing Focal Film Distance

Page 34: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

34

Changing Grid Ratio

Page 35: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

35

Calculate TCF @ 72”, nongrid

• TCF @ 40 was calculated as –4

• From 40 to 72 add 3 ½

• Going from 10:1 grid to nongrid subtract 3 ½

• The new TCF is –4– The inverse square law change is balanced

by removing the 10:1 grid from the beam

Page 36: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

36

Adjusting technique for body habitus

Page 37: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

37

Adjusting technique for casts

Page 38: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

38

The 15% rule is built in

• Supertech permits you to use kVp changes other than 15% to change contrast– There is a lower limit to kVp– Must be enough to penetrate the body part

Page 39: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

39

How to account for children

• Children under the age of 12 have less calcium in their bones, compared to adults

• Use the red boxes– The difference in the position of the red/black

boxes constitutes a 30% decrease in mAs for children

Page 40: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

40

Change TCF for non normal patient

• For patients with a muscular body habitus add a MINIMUM of +1

• For patients with a diminished bone density subtract a MINIMUM of -1

Page 41: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

41

Exposure through casts

• For plaster casts– If still wet, add +3– If dry, add +2

• For fiberglass casts no adjustment is required

Page 42: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

42

But I have a new x-ray machine with programmed techniques…

• These programs often only anticipate one film/screen combination at a fixed FFD

• You can use the Supertech® to adjust the programmed technique for your circumstance

Page 43: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

43

Other help

• More contrast• Less contrast

Page 44: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

44

Other help• Positioning aids

Page 45: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

45

Page 46: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

46

Relationships

Page 47: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

47

Penetrometer

Page 48: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

48

Use the penetrometer

• Expose the penetrometer to 60 kVp and 3.3 mAs at each mA station for high frequency generators– 100 mA @ 0.033 seconds– 200 mA @ 0.015 seconds– 300 mA @ 0.01 seconds

• Process the film• Use the master film to determine TCF

– The penetrometer image should be virtually identical at each mA station

Page 49: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

49

TCF test

Page 50: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

50

TCF test

Page 51: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

51

TCF test

Page 52: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

52

TCF test

Page 53: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

53

Page 54: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

54

Bad test, 200 mA out of calibration

Page 55: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

55

TCF test film

Page 56: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

56

Master film

Page 57: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

57

Using Master Film

Page 58: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

58

The End

Page 59: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

59

Nolan Accu-rad Calipers

• Along with the filter system, Dr. Nolan designed the Accu-Rad Calipers.

• Once set up by setting film speed, grid ratio and machine type, the calipers will:

Page 60: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

60

Nolan Accu-rad Calipers

– Provide the technical factors

– CM thickness– Points for additional

filters needed for the view.

• The caliper system uses high kVp to reduce patient exposure.

Page 61: 1 Formulating Technique Charts Accurate measurements and the used of technique charts will have the greatest impact on image quality and minimizing radiation

61

End of Lecture