In this type of instruments no calibration or comparison with other
instruments is necessary.
They are mostly used as means of standard measurements and are
maintained lay national laboratories and similar institutions.
Some of the examples of absolute
instruments are:
* Tangent galvanometer
* Raleigh current balance
* Absolute electrometer.
-
8/12/2019 1 Measuring Instruments
4/34
(B) Secondary Instrument :
The instrument which measures an electrical
quantity on a calibrated scale is termed as secondary instrument. Most of
the measuring instruments which are generally used are secondary
instruments
They are often calibrated by comparing them with either some absolute
instruments or with those which have already been calibrated.
-
8/12/2019 1 Measuring Instruments
5/34
-
8/12/2019 1 Measuring Instruments
6/34
-
8/12/2019 1 Measuring Instruments
7/34
According to Working Principle:1.Electromagnetic Instrument: Moving coil and moving iron instruments belong to this category.
2.Electro dynamic Instrument: Dynamo type ammeter, voltmeter and wattmeter belong to this category
3.Electro thermal Instrument: Hot wire ammeter and thermocouple meters belong to this category.
4.Electro induction Instrument: Watt meters and energy meters belong to this category.
5.Electrochemical Instrument: D.C. ampere-hour meter and watt-hour meters belong to this category.
6.Electrostatic Instrument: Certain voltmeters belong to this category.
FORCES ACTING IN AN INSTRUMENT
Following forces are necessary for working of an instrument1.Deflecting force
2.Controlling force
3.Damping force.
-
8/12/2019 1 Measuring Instruments
8/34
Deflecting force:
The force acting on the moving system of an instrument
produce a torque which moves the moving system, is called deflecting force
or deflecting torque.
The deflection of the pointer is proportional to the deflecting
torque, and which itself is proportional to the current passed through the
instrument.
Controlling Force:This force opposes the action of deflecting force, and
the pointer will rest in a position where the two force will neutralize each
other.
The controlling force stops the pointer to move
beyond the actual value of electrical quantity under measurement. It alsobrings the pointer back to zero position after disconnecting the meter
from the circuit. Thus no indicating instrument can work without a
controlling force.
-
8/12/2019 1 Measuring Instruments
9/34
The controlling force can be developed in the following two ways:
1.Spring control :
2.Gravity control :1.Spring Control :
The spindle of the moving system rests on two pivots.
A hair spring made of phosphor bronze wire is attached to the spindle.
-
8/12/2019 1 Measuring Instruments
10/34
-
8/12/2019 1 Measuring Instruments
11/34
Gravity Control:In this system, the pointer should rotate in a vertical plane.
The control is obtained by attaching a control weight to the pointer in
such a way that it produces a controlling torque because of change in its
position,
A balance weightis also attached to the pointer. Both the weights should
be adjustable.
The controlling torque is proportional to the sine of the angle of
deflection and hence the scale will not be a uniform scale.
-
8/12/2019 1 Measuring Instruments
12/34
-
8/12/2019 1 Measuring Instruments
13/34
The various methods adopted for damping are:
1.Air friction damping
2.Fluid friction damping
3.Eddy current damping
-
8/12/2019 1 Measuring Instruments
14/34
In air friction damping a light aluminum piston is attached to the moving
part of the instrument.
The piston moves in an air chamber which is closed at its one end. It will
have a small clearance with the air chamber.
The system works on the compression and suction of the air. Now days,
two aluminum vanes are attached to the moving part, which rotate in
sector shaped air-chambers.
1.Air Friction Damping:
-
8/12/2019 1 Measuring Instruments
15/34
Fluid Friction Damping:
This system is similar to the air friction damping, but it employs a liquid in place of
air.
The liquid should have a high viscosity so as to produce sufficient damping force.
The system is not suitable for portable instruments, hence it is out dated now.
-
8/12/2019 1 Measuring Instruments
16/34
Eddy Current Damping:
In this system, a circular thin disc of copper or aluminum is mounted on
the spindle. A damping magnet is placed on side of the disc.
The magnet is place in such a way that the rotating disc will cut the
magnetic lines of force of the magnet.
Thus, eddy currents will be induced in the disc which will oppose its
motion, and thus a damping force is produced in the instrument.
-
8/12/2019 1 Measuring Instruments
17/34
Permanent Magnet Moving Coil (PMMC) Meter
A coil of fine wire is suspended in a magnetic field produced by permanent
magnet.
According to the fundamental law of electromagnetic force, the coil will rotatein the magnetic field when it carries an electric current by electromagnetic (EM)
torque effect.
A pointer which attached the movable coil will deflect according to the amount
of current to be measured which applied to the coil.
-
8/12/2019 1 Measuring Instruments
18/34
-
8/12/2019 1 Measuring Instruments
19/34
-
8/12/2019 1 Measuring Instruments
20/34
-
8/12/2019 1 Measuring Instruments
21/34
Uniform scale.
Eddy current damping, so very effective & reliable.
No hysteresis loss.
Low power consumption as the driving power is small.
No effect of stray magnetic field.
High torque/weight ratio, require small operating current.
Very accurate & reliable.
The scale covered upto 270 degree.
-
8/12/2019 1 Measuring Instruments
22/34
Can not be used for AC measurement.
Costlier in comparison to moving iron instruments.
Friction & temperature might introduce some error.
Errors due to ageing of control springs & permanentmagnets.
-
8/12/2019 1 Measuring Instruments
23/34
E i f A R
-
8/12/2019 1 Measuring Instruments
24/34
Extension of Ammeter Range:
Since the coil winding in PMMC meter issmall and light, they can carry only small
currents (A-1mA). Measurement of large current requires a shunt external r esistor to
connect with the meter movement, so only a fraction of the total current will passes
through the meter.
Vsh = Vm
IshRsh= ImRm
But Ish = IT Im
Ish
RmIm
shR
mT
RmImsh
II
R
-
8/12/2019 1 Measuring Instruments
25/34
Example:If PMMC meter have internal resistance of 10 and full scale range of 1mA.
Assume we wish to increase the meter range to 1A.
we must connect shunt resistance with the PMMC meter of Rsh
ImTI
RmImshR
-
8/12/2019 1 Measuring Instruments
26/34
-
8/12/2019 1 Measuring Instruments
27/34
Example 1:- A moving coil ammeter has a full
scale deflection of 50 Amp and a coil resistance
of 1000 . What will be the value of the shuntresistance required for the instrument to be
converted to read a full scale reading of 1 Amp.
0.05
Example 2 :- The full scale deflection current of
an ammeter is 1 mA and its internal resistance is
100 . If this meter is to have scale deflection at5 A, what is the value of shunt resistance to be
used.
R sh = 0.020004 ohm
-
8/12/2019 1 Measuring Instruments
28/34
VOLTMETER
-
8/12/2019 1 Measuring Instruments
29/34
ExampleConvert a 50 micro A, 3kPMMC into a 0-10VDC voltmeter.
The voltmeter looks like this:
A PMMC meter with a coil resistance 100 and a full scale deflection currentof 100A is to be used in the voltmeter circuit as shown in Fig.42.6. The
voltmeter ranges are to be 50. Determine the required value of resistances
for each range.
(i) For 50V - 0.. 4999M
(ii) For 100V - 0.9999M
(iii)For 150V - 1.4999M
-
8/12/2019 1 Measuring Instruments
30/34
A moving coil ammeter gives full scale deflection with 15 mA and has aresistance of 5 . Calculate the resistance to be connected in (a)Parallel to enable the instrument to read upto 1 A (b) Series to enableit to read up to 10 V.
Rsh = 0.07614 Rse = 661.6667
-
8/12/2019 1 Measuring Instruments
31/34
Moving-iron InstrumentsThe deflecting torque in any moving-iron instrument is due to forces on a small
piece of magnetically soft iron that is magnetized by a coil carrying the
operating current.(i) Attraction Type (ii) Repulsion Type
-
8/12/2019 1 Measuring Instruments
32/34
REPULSION TYPE
Repulsive forces will act when two similarly magnetized iron pieces are
placed near to each other.
-
8/12/2019 1 Measuring Instruments
33/34
Sensitivity: It is defined by the change in the output or response of the
instrument for a unit change of input or measured variable.
Resolution: Resolution is the smallest change in a measured variable (or
measurand ) to which the instrument will respond.
Terms used :
Error: It is the deviation of the measured (or indicated) value from the
true (or expected) value of a quantity.
In other words, error is the difference between the measured value and
the true value of the unknown quantity. It is also called absolute error are
maximum possible error.
A = Am At
whereAm = measured value of the quantity
At = true value of the quantity
-