1-s2.0-s0976120913000107-main
DESCRIPTION
Qualification HPLCTRANSCRIPT
-
Paresh Varshney a,*, Atul KauParmesh Kumar Dwivedi d
a th Univ
on Phar
& Deve
Univers
ion of the instrument
.
al modules suggested
g the correct injection
utosampler chambers
e accuracy, precision,
t carryover.
dia Pvt. Ltd. All rights
reserved.
spectrometry (LCeMS/MS) has been the mainstay of bio-
analytical industry from last two decades.1e4 It amalgamate
the chromatographic resolution power of HPLC and the
sensitivity and specificity of mass spectrometer to develop
rapid and high throughput methods for the bioanalysis of
d endogenous com-
pounds in biological matrices.
conventional procedures for HPLC calibration which use
Photodiode Array (PDA) or variable UV/Visible spectropho-
tometers as detectors.8e14 However, due to continuous evo-
lution of bioanalytical techniques and with the rising
compliance expectations, an increased trend has been
* Corresponding author. Bioanalytical Department, DX Clinical, 260, Zone Industrial, El Jadida 24000, Morocco. Tel.: 212 653686403;fax: 212 523373127.
Available online at www.sciencedirect.com
.e
i n t e rn a t i o n a l j o u rn a l o f c h em i c a l a n d an a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7E-mail address: [email protected] (P. Varshney).High-performance liquid chromatography coupled to mass The bioanalytical laboratories throughout theworld rely on1. Introduction small molecule drugs, metabolites an4e7HPLC calibration
LCeMS/MS
Injection volume repeatability test
Injector position check
System suitability
and precision were carried out for holistic (overall system) calibrat
using a sensitive high throughput LCeMS/MS pre-validated method
Results/Conclusion: The results of calibration experiments for individu
that pumps are yielding accurate flow rate, autosampler is aspiratin
volume irrespective to position or volume, and column oven and a
are maintaining steady temperatures. Holistic calibration verified th
repeatability and reproducibility of the instrument with insignifican
Copyright 2013, JPR Solutions; Published by Reed Elsevier InKeywords: illustrated using direct weighing method. Carryover test, pump repeatability test, linearityDepartment of Pharmacology, Sai Nab Formulation and Development, WatscAnalytical Division, Dabur ResearchdAmity Institute of Pharmacy, Amity
Lucknow 226010, Uttar Pradesh, India
a r t i c l e i n f o
Article history:
Received 25 February 2013
Accepted 2 April 2013
Available online 9 April 20130976-1209/$ e see front matter Copyright http://dx.doi.org/10.1016/j.ijcas.2013.04.001shik b, Alok Sharma c, Sajal Srivastava d,
ersity, Ranchi, Jharkhand 835217, India
ma Pvt Ltd., Mumbai 400099, India
lopment Centre, Dabur India Ltd., Sahibabad 201010, India
ity Uttar Pradesh, Lucknow Campus Gomti Nagar Extension,
a b s t r a c t
Background/Objective: A direct, accurate, precise, reproducible and cost-time efficient
approach for calibration of HPLC using mass spectrometer as detector has been described
which simulates the routine analysis in bioanalytical laboratories.
Methods: Injector position check and injection volume repeatability test for the calibration
of autosampler; and pump flow rate accuracy test for the calibration of pumps weremass spectrometer as detector
A rational approach for HPLC calibration usingOriginal Article
journal homepage: www2013, JPR Solutions; Publilsevier .com/locate/ i jcasshed by Reed Elsevier India Pvt. Ltd. All rights reserved.
-
observed for the number of FDA 483 observations for
HPLC calibration while using hyphenated LCeMS/MS in-
struments.15,16 The limitations to the conventional ap-
proaches for HPLC calibration are summarized below.
The assessment of injector position check (IPC) and in-
jection volume repeatability test (IVRT) is done by statistical
comparison between peak area response and the concen-
tration of the injected samples from different vial positions
(for IPC) or with different volume injections (for IVRT).
Usually the precision (for IPC and IVRT) and coefficient of
correlation (for IVRT) are calculated for results interpreta-
sensitivity UV detector for holistic (overall) calibration. As
bioanalytical methods involve analysis of drugs at very low
2.1. Chemicals and reagents
Pioglitazone hydrochloride (purity: 98.90%) and metaxalone
(purity: 99.89%) were sourced from Varda biotech (P) limited
(Mumbai, India). HPLC-grade acetonitrile was sourced from
Rankem (India) andmethanol was sourced fromS.D. fine chem
limited (Mumbai, India). HPLC grade water was obtained from
aMilli-Q gradient A10 purifier. Ammonium formate (analytical
grade) was sourced from Panreac (Barcelona, Spain).
i n t e r n a t i o n a l j o u r n a l o f c h em i c a l a n d a n a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7 103concentrations, it is essential to ensure that even the slight
variations in the performance of HPLC are noticed by the de-
tector during calibration which may not be possible by UV
detector because of its low sensitivity.
Moreover, in the straightforward procedures like
measuring the effluent volume for pump flow rate accuracy
(PFRA) or temperatures for column oven or autosampler no in-
process evidence documents are generated.
The proposed research work is aimed to demonstrate
methodologies for the calibration of HPLC using highly sen-
sitive mass spectrometer as detector to overcome the limita-
tions of the conventional procedures. All the modules of HPLC
viz. pumps, column oven and autosampler were individually
calibrated without using any detector and then holistic cali-
bration was performed coupling HPLC in-tandem with mass
spectrometer.
2. Materials and methods
A selective, sensitive and validated LCeMS/MS bioanalytical
method for the estimation of Pioglitazone (PIO) was used as a
model to carry out HPLC calibration. The method was vali-
dated over the concentration range of 24.00e4003.30 ng/ml for
PIO using Metaxalone (MET) as an Internal Standard (IS).
Table 1 e Optimized parameters for mass spectrometry.
Source and gas parameters
Curtain gas (psi) 25
Ion spray voltage (volts) 5000
Heater gas (psi) 50
Source temperature (C) 550Nebulizer gas (psi) 50tion. However, the procedures count on the indirect mea-
surement of peak area response which is the function of
detector and not the autosampler. This indirect approach
for autosampler calibration cannot express the results
accurately especially when the autosampler is compro-
mised due to unnoticed hardware alterations or aging
components.
Second limitation is the irrational coupling of HPLC to lowCollision associated dissociation gas (psi) 6Compound parameters
PIO MET
Declustering potential (volts) 80 32
Entrance potential (volts) 5 10
Collision energy (volts) 25 14
Collision cell exit potential (volts) 3 32.4. Calibration parameters
The calibration of HPLC was performed in the following
order: Pumps {PFRA test} / Column Oven {COTP
check}/ Autosampler {ASTP check, IPC (for precision), IVRT
(for precision and linearity)} / holistic calibration {system
suitability, precision, linearity}. Before calibration, each
module was shutdown and powered-up to evoke automated
internal initialization procedure and built-in diagnostics for
detecting problem situations.2.3. LCeMS/MS conditions during holistic calibration
An isocratic mobile phase {ammonium formate buffer 5 mM
(pH 6.2): acetonitrile (10:90, v/v)} was run onWaters XTerraMS
C18 5 mm (4.6mm 150mm) column at a constant flow rate of0.5 ml/min. The temperatures of the column oven and auto-
sampler were maintained at 40 C 1.0 C and 10 C 1.0 C,respectively.
Mobile phase was introduced into mass spectrometer
through an ESI source operating in the positivemode at an ion
spray voltage of 5000 V. The transitions of PIO and MET were
monitored at 357.10> 134.10 and 222.20> 161.20, respectively.
The source and compound parameters optimized for the
detection are summarized in Table 1.2.2. Instrumentation
Holistic calibration was carried out on Perkin Elmer HPLC (se-
ries 200 LC pumps, series 200 column oven, series 200 auto-
sampler) connected in-tandem with calibrated API-3200 triple
quadrupole mass spectrometer (from MDS Sciex, Applied Bio-
systems, Canada). Gas generator (from Peak Scientific, Scot-
land), Model NM20ZAwas used to supply nitrogen gas and zero
grade air. The instrumentation was monitored using Analyst
software version 1.5 for data acquisition and processing.Dwell time (msec) 200 200
-
ference between the mean of actual temperature and set
Global %RSD was also calculated for the injection volume. %
pump repeatability and overall system verification using
UV detector is a time consuming procedure and typically take
i n t e rn a t i o n a l j o u rn a l o f c h em i c a l a n d an a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7104RSD should be
-
Table 2 e Results for PFRA, IPC and IVRT.
PFRA Results
Set flow rate (mL/min) Mean flow rate (mL/min) % Deviation Mean flow rate (mL/min) % Deviation
Pump 1 Pump 2
0.2 0.200954 0.48 0.201691 0.85
0.5 0.50324 0.65 0.504815 0.96
1.0 1.005985 0.60 0.991259 0.872.0 1.983424 0.83 1.988246 0.59
IPC Results IVRT Results
Vial Position Mean injectionvolume (mL)
%RSD Set injectionvolume (mL)
Actual injectionvolume (mL)
Injection volumeaccuracy
1 9.88 0.04 0.44 5 5.01 100.2010 9.98 0.11 1.06 10 10.02 100.1755 9.95 0.03 0.25 15 15.05 100.3691 10.00 0.11 1.05 20 20.00 99.98100 9.96 0.09 0.86 25 24.97 99.89Mean global (mL) 9.95 0.08 Mean injection volume accuracy 100.12 0.18%RSD global 0.80 Coefficient of correlation (R2) 1
i n t e r n a t i o n a l j o u r n a l o f c h em i c a l a n d a n a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7 105laboratory, the holistic calibration of HPLCwas completed in a
day using in-tandemmass spectrometer as detector with a set
of meaningful acceptance criteria that conform to the in-
dustry norms.
For PFRA test, % deviation found to be less than 1% for both
the pumps at the set flow rates of 0.2e2.0 ml/min using direct
weighing method (Table 2). The method proposed less mar-
gins of error as high sensitivity analytical balance was used to
mark the volume measurement. The process was time effi-
cient. In-process data was also generated on the analytical
balance.
For COTP check the deviation in temperatures ranged from
0.07 to 0.13 C while for ASTP check the deviation was0.00e0.03 C from the set temperatures. Datalogger was used
Fig. 1 e Representative LCeMS/MS chromatograms of BLK1to record the temperature readings as a means of generating
in-process document.
To calibrate autosampler, IPC and IVRT were performed
using direct weighing method. %RSD ranged from 0.25 to 1.06
at individual positions while %RSD for injection volume at all
the positions (global statistics) found to be 0.80 during
IPC (Table 2). The calibration plot for IVRT was generated be-
tween actual injection volume and set injection volume at
different injection levels and yielded a linear equation
(y 0.9981x0.038) with a unit coefficient of correlation. Theprecision for IVRT was evaluated in terms of %RSD for injec-
tion volume accuracy between different injection levels and
found to be 0.18 (Table 2). As IPC and IVRT were executedwithout any detector, completed with ease in few hours, and
, AQMIX and BLK2 of PIO and MET for carryover test.
-
Conflicts of interest
r e f e r e n c e s
Mean peak area response
in AQMIX
557125.7 455491.2
i n t e rn a t i o n a l j o u rn a l o f c h em i c a l a n d an a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7106direct weighing ensured that the injection volume aspirated
by the autosampler was accurate and precise.
System suitability accounted for the carryover, pump
repeatability and overall system verification. Both carryover
blanks (BLK1 and BLK2) found to be free of any significant
interference at the RT of PIO or MET (5% of the mean peakarea response of PIO (or MET) in AQMIX) (Fig. 1) and the
carryover found to be 0.02% for PIO and 0.01% for MET. %RSD for retention times of PIO and MET in 6 AQMIX injections
were 0.03% and 0.04%, respectively, which suggested that both
pumps were producing a uniform and constant flow rate. For
overall system verification, peak area response ratio of PIO to
METwas calculated for 6 AQMIX injections and%RSD found to
be 1.39%which suggested HPLC is performing with acceptable
variations using mass spectrometer as detector. The results
for pump repeatability, carryover and system suitability are
summarized in Table 3.
The linearity and precision were performed by plotting
Response in BLK2 36 33
% Carryover 0.02 0.01
System suitability
Mean area ratio (PIO/MET) %RSD for area ratio
1.223 0.017 1.39Table 3 e Results for holistic calibration.
Pump repeatability
Analyte Mean RT %RSD for RT
PIO 3.159 0.001 0.03
MET 3.076 0.001 0.04
Carryover test
PIO MET
Response in BLK1 168 60calibration curve of PIO between eight concentration levels,
i.e. 24.00, 51.00, 106.30, 221.40, 461.20, 960.80, 2001.60 and
4003.30 ng/ml and their respective area response ratios to
MET. The precision was calculated at each concentration level
for the three calibration curves and found to be ranging be-
tween 0.80 and 2.84% (Table 3). The results indicated that data
generated for linearity and precision for overall system veri-
fication was precise, accurate and reproducible.
4. Conclusion
In the proposed research work, the efforts have been made to
elevate the standards of bioanalytical industry practices for
HPLC calibration and to relieve the actual burden of the process
by considerably reducing the time and resource expenditures
involved. The methodologies have simple procedures and for-
mulas, well defined acceptance criteria as per industry norms
and accurate and precise demonstration of calibration experi-
ments for individual modules and overall system. Moreover, as
the procedure for calibration utilized mass spectrometer soft-
ware hence undermined the necessity of any extra HPLC1. Xu RN, Fan L, Rieser MJ, El-Shourbagy TA. Recent advances inhigh-throughput quantitative bioanalytical analysis byLCeMS/MS. J Pharm Biomed Anal. 2007;44:342e355.All authors have none to declare.software, its purchase and validation cost. Any software
compatibility issuewith themass spectrometer software is also
out of question. Additionally, calibration of HPLC serves as the
cross-calibration of mass spectrometer as well.
Linearity and precision results
Nominal conc.
(ng/mL)
Mean calculated conc.
(ng/mL)
%RSD
24.0 23.67 0.35 1.4851.0 52.33 1.23 2.36106.3 105.73 0.85 0.80221.4 225.23 2.63 1.17461.2 462.67 12.17 2.63960.8 958.07 15.16 1.58
2001.6 1999.63 39.62 1.984003.3 3877.57 110.05 2.84
Slope Intercept Coefficient of
correlation
Injection 1 0.000361 0.000307 0.9995
Injection 2 0.000369 0.00114 0.9999
Injection 3 0.000663 0.00145 0.99942. Wells DA. High Throughput Bioanalytical Sample Preparation:Methods and Automation Strategies. 1st ed. Amsterdam:Elsevier; 2003.
3. Shen JX, Tama CI, Hayes RN. Evaluation of automated microsolid phase extraction tips (m-SPE) for the validation of aLCeMS/MS bioanalytical method. J Chromatogr B.2006;843:275e282.
4. Lee MS, Kerns ED. LC/MS applications in drug development.Mass Spectrom Rev. 1999;18:187e279.
5. Eeckhaut AV, Lanckmans K, Sarre S, Smolders I, Michotte Y.Validation of bioanalytical LCeMS/MS assays: evaluation ofmatrix effects. J Chromatogr B Anal Technol Biomed Life Sci.2009;877:2198e2207.
6. Lim CK, Lord G. Current developments in LCeMS forpharmaceutical analysis. Biol Pharm Bull. 2002;25:547e557.
7. Pucci V, Palma SD, Alfieri A, Bonelli F, Monteagudo E. A novelstrategy for reducing phospholipids-based matrix effect inLCESI-MS bioanalysis by means of HybridSPE. J PharmBiomed Anal. 2009;50:867e871.
8. Horvath CG, Lipsky SR. Use of liquid ion exchangechromatography for the separation of organic compounds.Nature. 1966;211:748e749.
9. Rao RN, Nagaraju V. An overview of the recent trends indevelopment of HPLC methods for determination ofimpurities in drugs. J Pharm Biomed Anal. 2003;33:335e377.
-
10. Riordon JR. Diode array detectors for HPLCdhighperformance across the spectrum. Anal Chem.2000;72:483Ae487A.
11. Tracqui A, Kintz P, Mangin P. Systematic toxicologicalanalysis using HPLC/Dad. J Forensic Sci. 1995;40:254e262.
12. Lambert WE, Van Bocxlaer JF, De Leenheer AP. Potential ofhigh-performance liquid chromatography with photodiodearray detection in forensic toxicology. J Chromatogr B BiomedSci Appl. 1997;689:45e53.
13. Chan HK, Carr GP. Evaluation of a photodiode array detectorfor the verification of peak-homogeneity in high-performanceliquid chromatography. J Pharm Biomed Anal. 1990;8:271e277.
14. Nicoletti I, De Rossi A, Giovinazzo G, Corradini D.Identification and quantification of stilbenes in fruits oftransgenic tomato plants (Lycopersicon esculentum Mill.) byreversed phase HPLC with photodiode array and massspectrometry detection. J Agric Food Chem. 2007;55:3304e3311.
15. Chow F, Lum S, Ocampo A, Vogel P. Current challenges forFDA-regulated bioanalytical laboratories for human (BA/BE)studies, Part II: recent FDA Inspection trends for bioanalytical
laboratories using LC/MS/MS methods and FDA Inspectionreadiness preparation. Qual Assur J. 2007;11:111e122.
16. Brendelberger G. GMP News: HPLC in FDAWarning Letters. 10January 2003. http://www.gmp-compliance.org/eca_news_277.html; last Accessed 09.12.
17. Crowther J, Dowling J, Hartwick R, Ciccone B. PerformanceQualification of HPLC Instrumentation in Regulated Laboratories.LC-GC North America e Chromatography Online; May 1, 2008.
18. Lam H. Performance verification of HPLC. In: Chan CC, Lam H,Lee YC, Zhang XM, eds. Analytical Method Validation andInstrument Performance Verification. 1st ed. New Jersey: JohnWiley & Sons; 2004:173e186.
19. Ahuja S, Dong MW. Handbook of Pharmaceutical Analysis byHPLC. 1st ed. Amsterdam: Elsevier; 2005.
20. Dong M, Paul R. Committeeing to Calibrate HPLC, a casehistory of one companys effort to expedite regulatorycompliance. Todays Chem Sep Sci. 2001;10:42e48.
21. Ebel S, Kuhnert H, Muck W. Limits of determinationand calibration in HPLC. Chromatographia.1987;23:934e938.
i n t e r n a t i o n a l j o u r n a l o f c h em i c a l a n d a n a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7 107
A rational approach for HPLC calibration using mass spectrometer as detector1. Introduction2. Materials and methods2.1. Chemicals and reagents2.2. Instrumentation2.3. LCMS/MS conditions during holistic calibration2.4. Calibration parameters2.5. PFRA test2.6. COTP and ASTP check2.7. Injector position check2.8. Injection volume repeatability test2.9. Holistic calibration
3. Results and discussion4. ConclusionConflicts of interestReferences