1-s2.0-s0976120913000107-main

6
Original Article A rational approach for HPLC calibration using mass spectrometer as detector Paresh Varshney a, *, Atul Kaushik b , Alok Sharma c , Sajal Srivastava d , Parmesh Kumar Dwivedi d a Department of Pharmacology, Sai Nath University, Ranchi, Jharkhand 835217, India b Formulation and Development, Watson Pharma Pvt Ltd., Mumbai 400099, India c Analytical Division, Dabur Research & Development Centre, Dabur India Ltd., Sahibabad 201010, India d Amity Institute of Pharmacy, Amity University Uttar Pradesh, Lucknow Campus Gomti Nagar Extension, Lucknow 226010, Uttar Pradesh, India article info Article history: Received 25 February 2013 Accepted 2 April 2013 Available online 9 April 2013 Keywords: HPLC calibration LCeMS/MS Injection volume repeatability test Injector position check System suitability abstract Background/Objective: A direct, accurate, precise, reproducible and cost-time efficient approach for calibration of HPLC using mass spectrometer as detector has been described which simulates the routine analysis in bioanalytical laboratories. Methods: Injector position check and injection volume repeatability test for the calibration of autosampler; and pump flow rate accuracy test for the calibration of pumps were illustrated using direct weighing method. Carryover test, pump repeatability test, linearity and precision were carried out for holistic (overall system) calibration of the instrument using a sensitive high throughput LCeMS/MS pre-validated method. Results/Conclusion: The results of calibration experiments for individual modules suggested that pumps are yielding accurate flow rate, autosampler is aspirating the correct injection volume irrespective to position or volume, and column oven and autosampler chambers are maintaining steady temperatures. Holistic calibration verified the accuracy, precision, repeatability and reproducibility of the instrument with insignificant carryover. Copyright ª 2013, JPR Solutions; Published by Reed Elsevier India Pvt. Ltd. All rights reserved. 1. Introduction High-performance liquid chromatography coupled to mass spectrometry (LCeMS/MS) has been the mainstay of bio- analytical industry from last two decades. 1e4 It amalgamate the chromatographic resolution power of HPLC and the sensitivity and specificity of mass spectrometer to develop rapid and high throughput methods for the bioanalysis of small molecule drugs, metabolites and endogenous com- pounds in biological matrices. 4e7 The bioanalytical laboratories throughout the world rely on conventional procedures for HPLC calibration which use Photodiode Array (PDA) or variable UV/Visible spectropho- tometers as detectors. 8e14 However, due to continuous evo- lution of bioanalytical techniques and with the rising compliance expectations, an increased trend has been * Corresponding author. Bioanalytical Department, DX Clinical, 260, Zone Industrial, El Jadida 24000, Morocco. Tel.: þ212 653686403; fax: þ212 523373127. E-mail address: [email protected] (P. Varshney). Available online at www.sciencedirect.com journal homepage: www.elsevier.com/locate/ijcas international journal of chemical and analytical science 4 (2013) 102 e107 0976-1209/$ e see front matter Copyright ª 2013, JPR Solutions; Published by Reed Elsevier India Pvt. Ltd. All rights reserved. http://dx.doi.org/10.1016/j.ijcas.2013.04.001

Upload: giamillia

Post on 26-Sep-2015

213 views

Category:

Documents


0 download

DESCRIPTION

Qualification HPLC

TRANSCRIPT

  • Paresh Varshney a,*, Atul KauParmesh Kumar Dwivedi d

    a th Univ

    on Phar

    & Deve

    Univers

    ion of the instrument

    .

    al modules suggested

    g the correct injection

    utosampler chambers

    e accuracy, precision,

    t carryover.

    dia Pvt. Ltd. All rights

    reserved.

    spectrometry (LCeMS/MS) has been the mainstay of bio-

    analytical industry from last two decades.1e4 It amalgamate

    the chromatographic resolution power of HPLC and the

    sensitivity and specificity of mass spectrometer to develop

    rapid and high throughput methods for the bioanalysis of

    d endogenous com-

    pounds in biological matrices.

    conventional procedures for HPLC calibration which use

    Photodiode Array (PDA) or variable UV/Visible spectropho-

    tometers as detectors.8e14 However, due to continuous evo-

    lution of bioanalytical techniques and with the rising

    compliance expectations, an increased trend has been

    * Corresponding author. Bioanalytical Department, DX Clinical, 260, Zone Industrial, El Jadida 24000, Morocco. Tel.: 212 653686403;fax: 212 523373127.

    Available online at www.sciencedirect.com

    .e

    i n t e rn a t i o n a l j o u rn a l o f c h em i c a l a n d an a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7E-mail address: [email protected] (P. Varshney).High-performance liquid chromatography coupled to mass The bioanalytical laboratories throughout theworld rely on1. Introduction small molecule drugs, metabolites an4e7HPLC calibration

    LCeMS/MS

    Injection volume repeatability test

    Injector position check

    System suitability

    and precision were carried out for holistic (overall system) calibrat

    using a sensitive high throughput LCeMS/MS pre-validated method

    Results/Conclusion: The results of calibration experiments for individu

    that pumps are yielding accurate flow rate, autosampler is aspiratin

    volume irrespective to position or volume, and column oven and a

    are maintaining steady temperatures. Holistic calibration verified th

    repeatability and reproducibility of the instrument with insignifican

    Copyright 2013, JPR Solutions; Published by Reed Elsevier InKeywords: illustrated using direct weighing method. Carryover test, pump repeatability test, linearityDepartment of Pharmacology, Sai Nab Formulation and Development, WatscAnalytical Division, Dabur ResearchdAmity Institute of Pharmacy, Amity

    Lucknow 226010, Uttar Pradesh, India

    a r t i c l e i n f o

    Article history:

    Received 25 February 2013

    Accepted 2 April 2013

    Available online 9 April 20130976-1209/$ e see front matter Copyright http://dx.doi.org/10.1016/j.ijcas.2013.04.001shik b, Alok Sharma c, Sajal Srivastava d,

    ersity, Ranchi, Jharkhand 835217, India

    ma Pvt Ltd., Mumbai 400099, India

    lopment Centre, Dabur India Ltd., Sahibabad 201010, India

    ity Uttar Pradesh, Lucknow Campus Gomti Nagar Extension,

    a b s t r a c t

    Background/Objective: A direct, accurate, precise, reproducible and cost-time efficient

    approach for calibration of HPLC using mass spectrometer as detector has been described

    which simulates the routine analysis in bioanalytical laboratories.

    Methods: Injector position check and injection volume repeatability test for the calibration

    of autosampler; and pump flow rate accuracy test for the calibration of pumps weremass spectrometer as detector

    A rational approach for HPLC calibration usingOriginal Article

    journal homepage: www2013, JPR Solutions; Publilsevier .com/locate/ i jcasshed by Reed Elsevier India Pvt. Ltd. All rights reserved.

  • observed for the number of FDA 483 observations for

    HPLC calibration while using hyphenated LCeMS/MS in-

    struments.15,16 The limitations to the conventional ap-

    proaches for HPLC calibration are summarized below.

    The assessment of injector position check (IPC) and in-

    jection volume repeatability test (IVRT) is done by statistical

    comparison between peak area response and the concen-

    tration of the injected samples from different vial positions

    (for IPC) or with different volume injections (for IVRT).

    Usually the precision (for IPC and IVRT) and coefficient of

    correlation (for IVRT) are calculated for results interpreta-

    sensitivity UV detector for holistic (overall) calibration. As

    bioanalytical methods involve analysis of drugs at very low

    2.1. Chemicals and reagents

    Pioglitazone hydrochloride (purity: 98.90%) and metaxalone

    (purity: 99.89%) were sourced from Varda biotech (P) limited

    (Mumbai, India). HPLC-grade acetonitrile was sourced from

    Rankem (India) andmethanol was sourced fromS.D. fine chem

    limited (Mumbai, India). HPLC grade water was obtained from

    aMilli-Q gradient A10 purifier. Ammonium formate (analytical

    grade) was sourced from Panreac (Barcelona, Spain).

    i n t e r n a t i o n a l j o u r n a l o f c h em i c a l a n d a n a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7 103concentrations, it is essential to ensure that even the slight

    variations in the performance of HPLC are noticed by the de-

    tector during calibration which may not be possible by UV

    detector because of its low sensitivity.

    Moreover, in the straightforward procedures like

    measuring the effluent volume for pump flow rate accuracy

    (PFRA) or temperatures for column oven or autosampler no in-

    process evidence documents are generated.

    The proposed research work is aimed to demonstrate

    methodologies for the calibration of HPLC using highly sen-

    sitive mass spectrometer as detector to overcome the limita-

    tions of the conventional procedures. All the modules of HPLC

    viz. pumps, column oven and autosampler were individually

    calibrated without using any detector and then holistic cali-

    bration was performed coupling HPLC in-tandem with mass

    spectrometer.

    2. Materials and methods

    A selective, sensitive and validated LCeMS/MS bioanalytical

    method for the estimation of Pioglitazone (PIO) was used as a

    model to carry out HPLC calibration. The method was vali-

    dated over the concentration range of 24.00e4003.30 ng/ml for

    PIO using Metaxalone (MET) as an Internal Standard (IS).

    Table 1 e Optimized parameters for mass spectrometry.

    Source and gas parameters

    Curtain gas (psi) 25

    Ion spray voltage (volts) 5000

    Heater gas (psi) 50

    Source temperature (C) 550Nebulizer gas (psi) 50tion. However, the procedures count on the indirect mea-

    surement of peak area response which is the function of

    detector and not the autosampler. This indirect approach

    for autosampler calibration cannot express the results

    accurately especially when the autosampler is compro-

    mised due to unnoticed hardware alterations or aging

    components.

    Second limitation is the irrational coupling of HPLC to lowCollision associated dissociation gas (psi) 6Compound parameters

    PIO MET

    Declustering potential (volts) 80 32

    Entrance potential (volts) 5 10

    Collision energy (volts) 25 14

    Collision cell exit potential (volts) 3 32.4. Calibration parameters

    The calibration of HPLC was performed in the following

    order: Pumps {PFRA test} / Column Oven {COTP

    check}/ Autosampler {ASTP check, IPC (for precision), IVRT

    (for precision and linearity)} / holistic calibration {system

    suitability, precision, linearity}. Before calibration, each

    module was shutdown and powered-up to evoke automated

    internal initialization procedure and built-in diagnostics for

    detecting problem situations.2.3. LCeMS/MS conditions during holistic calibration

    An isocratic mobile phase {ammonium formate buffer 5 mM

    (pH 6.2): acetonitrile (10:90, v/v)} was run onWaters XTerraMS

    C18 5 mm (4.6mm 150mm) column at a constant flow rate of0.5 ml/min. The temperatures of the column oven and auto-

    sampler were maintained at 40 C 1.0 C and 10 C 1.0 C,respectively.

    Mobile phase was introduced into mass spectrometer

    through an ESI source operating in the positivemode at an ion

    spray voltage of 5000 V. The transitions of PIO and MET were

    monitored at 357.10> 134.10 and 222.20> 161.20, respectively.

    The source and compound parameters optimized for the

    detection are summarized in Table 1.2.2. Instrumentation

    Holistic calibration was carried out on Perkin Elmer HPLC (se-

    ries 200 LC pumps, series 200 column oven, series 200 auto-

    sampler) connected in-tandem with calibrated API-3200 triple

    quadrupole mass spectrometer (from MDS Sciex, Applied Bio-

    systems, Canada). Gas generator (from Peak Scientific, Scot-

    land), Model NM20ZAwas used to supply nitrogen gas and zero

    grade air. The instrumentation was monitored using Analyst

    software version 1.5 for data acquisition and processing.Dwell time (msec) 200 200

  • ference between the mean of actual temperature and set

    Global %RSD was also calculated for the injection volume. %

    pump repeatability and overall system verification using

    UV detector is a time consuming procedure and typically take

    i n t e rn a t i o n a l j o u rn a l o f c h em i c a l a n d an a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7104RSD should be

  • Table 2 e Results for PFRA, IPC and IVRT.

    PFRA Results

    Set flow rate (mL/min) Mean flow rate (mL/min) % Deviation Mean flow rate (mL/min) % Deviation

    Pump 1 Pump 2

    0.2 0.200954 0.48 0.201691 0.85

    0.5 0.50324 0.65 0.504815 0.96

    1.0 1.005985 0.60 0.991259 0.872.0 1.983424 0.83 1.988246 0.59

    IPC Results IVRT Results

    Vial Position Mean injectionvolume (mL)

    %RSD Set injectionvolume (mL)

    Actual injectionvolume (mL)

    Injection volumeaccuracy

    1 9.88 0.04 0.44 5 5.01 100.2010 9.98 0.11 1.06 10 10.02 100.1755 9.95 0.03 0.25 15 15.05 100.3691 10.00 0.11 1.05 20 20.00 99.98100 9.96 0.09 0.86 25 24.97 99.89Mean global (mL) 9.95 0.08 Mean injection volume accuracy 100.12 0.18%RSD global 0.80 Coefficient of correlation (R2) 1

    i n t e r n a t i o n a l j o u r n a l o f c h em i c a l a n d a n a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7 105laboratory, the holistic calibration of HPLCwas completed in a

    day using in-tandemmass spectrometer as detector with a set

    of meaningful acceptance criteria that conform to the in-

    dustry norms.

    For PFRA test, % deviation found to be less than 1% for both

    the pumps at the set flow rates of 0.2e2.0 ml/min using direct

    weighing method (Table 2). The method proposed less mar-

    gins of error as high sensitivity analytical balance was used to

    mark the volume measurement. The process was time effi-

    cient. In-process data was also generated on the analytical

    balance.

    For COTP check the deviation in temperatures ranged from

    0.07 to 0.13 C while for ASTP check the deviation was0.00e0.03 C from the set temperatures. Datalogger was used

    Fig. 1 e Representative LCeMS/MS chromatograms of BLK1to record the temperature readings as a means of generating

    in-process document.

    To calibrate autosampler, IPC and IVRT were performed

    using direct weighing method. %RSD ranged from 0.25 to 1.06

    at individual positions while %RSD for injection volume at all

    the positions (global statistics) found to be 0.80 during

    IPC (Table 2). The calibration plot for IVRT was generated be-

    tween actual injection volume and set injection volume at

    different injection levels and yielded a linear equation

    (y 0.9981x0.038) with a unit coefficient of correlation. Theprecision for IVRT was evaluated in terms of %RSD for injec-

    tion volume accuracy between different injection levels and

    found to be 0.18 (Table 2). As IPC and IVRT were executedwithout any detector, completed with ease in few hours, and

    , AQMIX and BLK2 of PIO and MET for carryover test.

  • Conflicts of interest

    r e f e r e n c e s

    Mean peak area response

    in AQMIX

    557125.7 455491.2

    i n t e rn a t i o n a l j o u rn a l o f c h em i c a l a n d an a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7106direct weighing ensured that the injection volume aspirated

    by the autosampler was accurate and precise.

    System suitability accounted for the carryover, pump

    repeatability and overall system verification. Both carryover

    blanks (BLK1 and BLK2) found to be free of any significant

    interference at the RT of PIO or MET (5% of the mean peakarea response of PIO (or MET) in AQMIX) (Fig. 1) and the

    carryover found to be 0.02% for PIO and 0.01% for MET. %RSD for retention times of PIO and MET in 6 AQMIX injections

    were 0.03% and 0.04%, respectively, which suggested that both

    pumps were producing a uniform and constant flow rate. For

    overall system verification, peak area response ratio of PIO to

    METwas calculated for 6 AQMIX injections and%RSD found to

    be 1.39%which suggested HPLC is performing with acceptable

    variations using mass spectrometer as detector. The results

    for pump repeatability, carryover and system suitability are

    summarized in Table 3.

    The linearity and precision were performed by plotting

    Response in BLK2 36 33

    % Carryover 0.02 0.01

    System suitability

    Mean area ratio (PIO/MET) %RSD for area ratio

    1.223 0.017 1.39Table 3 e Results for holistic calibration.

    Pump repeatability

    Analyte Mean RT %RSD for RT

    PIO 3.159 0.001 0.03

    MET 3.076 0.001 0.04

    Carryover test

    PIO MET

    Response in BLK1 168 60calibration curve of PIO between eight concentration levels,

    i.e. 24.00, 51.00, 106.30, 221.40, 461.20, 960.80, 2001.60 and

    4003.30 ng/ml and their respective area response ratios to

    MET. The precision was calculated at each concentration level

    for the three calibration curves and found to be ranging be-

    tween 0.80 and 2.84% (Table 3). The results indicated that data

    generated for linearity and precision for overall system veri-

    fication was precise, accurate and reproducible.

    4. Conclusion

    In the proposed research work, the efforts have been made to

    elevate the standards of bioanalytical industry practices for

    HPLC calibration and to relieve the actual burden of the process

    by considerably reducing the time and resource expenditures

    involved. The methodologies have simple procedures and for-

    mulas, well defined acceptance criteria as per industry norms

    and accurate and precise demonstration of calibration experi-

    ments for individual modules and overall system. Moreover, as

    the procedure for calibration utilized mass spectrometer soft-

    ware hence undermined the necessity of any extra HPLC1. Xu RN, Fan L, Rieser MJ, El-Shourbagy TA. Recent advances inhigh-throughput quantitative bioanalytical analysis byLCeMS/MS. J Pharm Biomed Anal. 2007;44:342e355.All authors have none to declare.software, its purchase and validation cost. Any software

    compatibility issuewith themass spectrometer software is also

    out of question. Additionally, calibration of HPLC serves as the

    cross-calibration of mass spectrometer as well.

    Linearity and precision results

    Nominal conc.

    (ng/mL)

    Mean calculated conc.

    (ng/mL)

    %RSD

    24.0 23.67 0.35 1.4851.0 52.33 1.23 2.36106.3 105.73 0.85 0.80221.4 225.23 2.63 1.17461.2 462.67 12.17 2.63960.8 958.07 15.16 1.58

    2001.6 1999.63 39.62 1.984003.3 3877.57 110.05 2.84

    Slope Intercept Coefficient of

    correlation

    Injection 1 0.000361 0.000307 0.9995

    Injection 2 0.000369 0.00114 0.9999

    Injection 3 0.000663 0.00145 0.99942. Wells DA. High Throughput Bioanalytical Sample Preparation:Methods and Automation Strategies. 1st ed. Amsterdam:Elsevier; 2003.

    3. Shen JX, Tama CI, Hayes RN. Evaluation of automated microsolid phase extraction tips (m-SPE) for the validation of aLCeMS/MS bioanalytical method. J Chromatogr B.2006;843:275e282.

    4. Lee MS, Kerns ED. LC/MS applications in drug development.Mass Spectrom Rev. 1999;18:187e279.

    5. Eeckhaut AV, Lanckmans K, Sarre S, Smolders I, Michotte Y.Validation of bioanalytical LCeMS/MS assays: evaluation ofmatrix effects. J Chromatogr B Anal Technol Biomed Life Sci.2009;877:2198e2207.

    6. Lim CK, Lord G. Current developments in LCeMS forpharmaceutical analysis. Biol Pharm Bull. 2002;25:547e557.

    7. Pucci V, Palma SD, Alfieri A, Bonelli F, Monteagudo E. A novelstrategy for reducing phospholipids-based matrix effect inLCESI-MS bioanalysis by means of HybridSPE. J PharmBiomed Anal. 2009;50:867e871.

    8. Horvath CG, Lipsky SR. Use of liquid ion exchangechromatography for the separation of organic compounds.Nature. 1966;211:748e749.

    9. Rao RN, Nagaraju V. An overview of the recent trends indevelopment of HPLC methods for determination ofimpurities in drugs. J Pharm Biomed Anal. 2003;33:335e377.

  • 10. Riordon JR. Diode array detectors for HPLCdhighperformance across the spectrum. Anal Chem.2000;72:483Ae487A.

    11. Tracqui A, Kintz P, Mangin P. Systematic toxicologicalanalysis using HPLC/Dad. J Forensic Sci. 1995;40:254e262.

    12. Lambert WE, Van Bocxlaer JF, De Leenheer AP. Potential ofhigh-performance liquid chromatography with photodiodearray detection in forensic toxicology. J Chromatogr B BiomedSci Appl. 1997;689:45e53.

    13. Chan HK, Carr GP. Evaluation of a photodiode array detectorfor the verification of peak-homogeneity in high-performanceliquid chromatography. J Pharm Biomed Anal. 1990;8:271e277.

    14. Nicoletti I, De Rossi A, Giovinazzo G, Corradini D.Identification and quantification of stilbenes in fruits oftransgenic tomato plants (Lycopersicon esculentum Mill.) byreversed phase HPLC with photodiode array and massspectrometry detection. J Agric Food Chem. 2007;55:3304e3311.

    15. Chow F, Lum S, Ocampo A, Vogel P. Current challenges forFDA-regulated bioanalytical laboratories for human (BA/BE)studies, Part II: recent FDA Inspection trends for bioanalytical

    laboratories using LC/MS/MS methods and FDA Inspectionreadiness preparation. Qual Assur J. 2007;11:111e122.

    16. Brendelberger G. GMP News: HPLC in FDAWarning Letters. 10January 2003. http://www.gmp-compliance.org/eca_news_277.html; last Accessed 09.12.

    17. Crowther J, Dowling J, Hartwick R, Ciccone B. PerformanceQualification of HPLC Instrumentation in Regulated Laboratories.LC-GC North America e Chromatography Online; May 1, 2008.

    18. Lam H. Performance verification of HPLC. In: Chan CC, Lam H,Lee YC, Zhang XM, eds. Analytical Method Validation andInstrument Performance Verification. 1st ed. New Jersey: JohnWiley & Sons; 2004:173e186.

    19. Ahuja S, Dong MW. Handbook of Pharmaceutical Analysis byHPLC. 1st ed. Amsterdam: Elsevier; 2005.

    20. Dong M, Paul R. Committeeing to Calibrate HPLC, a casehistory of one companys effort to expedite regulatorycompliance. Todays Chem Sep Sci. 2001;10:42e48.

    21. Ebel S, Kuhnert H, Muck W. Limits of determinationand calibration in HPLC. Chromatographia.1987;23:934e938.

    i n t e r n a t i o n a l j o u r n a l o f c h em i c a l a n d a n a l y t i c a l s c i e n c e 4 ( 2 0 1 3 ) 1 0 2e1 0 7 107

    A rational approach for HPLC calibration using mass spectrometer as detector1. Introduction2. Materials and methods2.1. Chemicals and reagents2.2. Instrumentation2.3. LCMS/MS conditions during holistic calibration2.4. Calibration parameters2.5. PFRA test2.6. COTP and ASTP check2.7. Injector position check2.8. Injection volume repeatability test2.9. Holistic calibration

    3. Results and discussion4. ConclusionConflicts of interestReferences