1-s2.0-s1872497316300631-main

12
Research paper Human DNA quantication and sample quality assessment: Developmental validation of the PowerQuant 1 system Margaret M. Ewing a, *, Jonelle M. Thompson a , Robert S. McLaren a , Vincent M. Purpero b , Kelli J. Thomas b , Patricia A. Dobrowski c , Gretchen A. DeGroot c , Erica L. Romsos d , Douglas R. Storts a a Promega Corporation, 2800 Woods Hollow Rd., Madison, WI 53711, USA b Wisconsin State Crime Laboratory Madison, 4626 University Ave., Madison, WI 53705, USA c Wisconsin State Crime Laboratory Milwaukee, 1578 S 11th St., Milwaukee, WI 53204-2860, USA d U.S. National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899-8314, USA A R T I C L E I N F O Article history: Received 5 November 2015 Received in revised form 18 March 2016 Accepted 12 April 2016 Available online 16 April 2016 Keywords: Forensic science DNA analysis Short tandem repeat (STR) qPCR Quantication Validation PowerQuant 1 PowerPlex 1 A B S T R A C T Quantication of the total amount of human DNA isolated from a forensic evidence item is crucial for DNA normalization prior to short tandem repeat (STR) DNA analysis and a federal quality assurance standard requirement. Previous commercial quantication methods determine the total human DNA and total human male DNA concentrations, but provide limited information about the condition of the DNA sample. The PowerQuant 1 System includes targets for quantication of total human and total human male DNA as well as targets for evaluating whether the human DNA is degraded and/or PCR inhibitors are present in the sample. A developmental validation of the PowerQuant 1 System was completed, following SWGDAM Validation Guidelines, to evaluate the assays specicity, sensitivity, precision and accuracy, as well as the ability to detect degraded DNA or PCR inhibitors. In addition to the total human DNA and total human male DNA concentrations in a sample, data from the degradation target and internal PCR control (IPC) provide a forensic DNA analyst meaningful information about the quality of the isolated human DNA and the presence of PCR inhibitors in the sample that can be used to determine the most effective workow and assist downstream interpretation. ã 2016 The Author(s). Published by Elsevier Ireland Ltd. All rights reserved. 1. Introduction An evidentiary item collected from a crime scene may contain non-human sources of DNA. After DNA is isolated from evidence, a measurement of the amount of recovered human DNA is necessary for normalization to the optimal mass for typing with a short tandem repeat (STR) multiplex and compliance with FBI Quality Assurance Standards [1]. Quantitative PCR (qPCR) is a sensitive method for DNA quantication and is compatible with multiplexing targets. qPCR is routinely used to measure the concentration of total human and total human male DNA [24]. The inclusion of an additional quantication target to assess DNA degradation in forensic samples has also been used [59]. The PowerQuant 1 System (Promega Corporation, Madison, WI) combines multicopy quantication for each target and an internal PCR control (IPC) in one 5-dye qPCR assay to quantify total human and human male DNA present in a sample and ascertain whether PCR inhibitors and/or degraded human DNA are present. Hydroly- sis probes are used for detection with an Applied Biosystems 1 7500 Real-Time PCR System instrument [10]. The PowerQuant 1 20X Primer/Probe/IPC Mix provided in the system contains the primers, probes, internal PCR control (IPC) template and passive reference dye. Also included in the system are the PowerQuant 1 2X Master Mix and the PowerQuant 1 Male gDNA Standard. The PowerQuant 1 2X Master Mix uses a hot-start chemistry for room temperature reaction set-up and is designed for performance comparable to newer STR systems [11,12]. The PowerQuant 1 Male gDNA Standard consists of pooled genomic human male DNA and is used to generate the standard curve for determining DNA concentration in the samples. * Corresponding author. E-mail addresses: [email protected] (M.M. Ewing), [email protected] (J.M. Thompson), [email protected] (R.S. McLaren), [email protected] (V.M. Purpero), [email protected] (K.J. Thomas), [email protected] (P.A. Dobrowski), [email protected] (G.A. DeGroot), [email protected] (E.L. Romsos), [email protected] (D.R. Storts). http://dx.doi.org/10.1016/j.fsigen.2016.04.007 1872-4973/ã 2016 The Author(s). Published by Elsevier Ireland Ltd. All rights reserved. This is an open access article under the CC BY-NC-ND license (http://creativecommons. org/licenses/by-nc-nd/4.0/). Forensic Science International: Genetics 23 (2016) 166177 Contents lists available at ScienceDirect Forensic Science International: Genetics journal homepage: www.else vie r.com/locate /fsig

Upload: alexandra-galan

Post on 08-Jul-2016

5 views

Category:

Documents


0 download

DESCRIPTION

k

TRANSCRIPT

Page 1: 1-s2.0-S1872497316300631-main

Forensic Science International: Genetics 23 (2016) 166–177

Research paper

Human DNA quantification and sample quality assessment:Developmental validation of the PowerQuant1 system

Margaret M. Ewinga,*, Jonelle M. Thompsona, Robert S. McLarena, Vincent M. Purperob,Kelli J. Thomasb, Patricia A. Dobrowskic, Gretchen A. DeGrootc, Erica L. Romsosd,Douglas R. Stortsa

a Promega Corporation, 2800 Woods Hollow Rd., Madison, WI 53711, USAbWisconsin State Crime Laboratory � Madison, 4626 University Ave., Madison, WI 53705, USAcWisconsin State Crime Laboratory � Milwaukee, 1578 S 11th St., Milwaukee, WI 53204-2860, USAdU.S. National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899-8314, USA

A R T I C L E I N F O

Article history:Received 5 November 2015Received in revised form 18 March 2016Accepted 12 April 2016Available online 16 April 2016

Keywords:Forensic scienceDNA analysisShort tandem repeat (STR)qPCRQuantificationValidationPowerQuant1

PowerPlex1

A B S T R A C T

Quantification of the total amount of human DNA isolated from a forensic evidence item is crucial forDNA normalization prior to short tandem repeat (STR) DNA analysis and a federal quality assurancestandard requirement. Previous commercial quantification methods determine the total human DNA andtotal human male DNA concentrations, but provide limited information about the condition of the DNAsample. The PowerQuant1 System includes targets for quantification of total human and total humanmale DNA as well as targets for evaluating whether the human DNA is degraded and/or PCR inhibitors arepresent in the sample. A developmental validation of the PowerQuant1 System was completed, followingSWGDAM Validation Guidelines, to evaluate the assay’s specificity, sensitivity, precision and accuracy, aswell as the ability to detect degraded DNA or PCR inhibitors. In addition to the total human DNA and totalhuman male DNA concentrations in a sample, data from the degradation target and internal PCR control(IPC) provide a forensic DNA analyst meaningful information about the quality of the isolated humanDNA and the presence of PCR inhibitors in the sample that can be used to determine the most effectiveworkflow and assist downstream interpretation.

ã 2016 The Author(s). Published by Elsevier Ireland Ltd. All rights reserved.

Contents lists available at ScienceDirect

Forensic Science International: Genetics

journal homepage: www.else vie r .com/locate / fs ig

1. Introduction

An evidentiary item collected from a crime scene may containnon-human sources of DNA. After DNA is isolated from evidence, ameasurement of the amount of recovered human DNA is necessaryfor normalization to the optimal mass for typing with a shorttandem repeat (STR) multiplex and compliance with FBI QualityAssurance Standards [1].

Quantitative PCR (qPCR) is a sensitive method for DNAquantification and is compatible with multiplexing targets. qPCRis routinely used to measure the concentration of total human andtotal human male DNA [2–4]. The inclusion of an additional

* Corresponding author.E-mail addresses: [email protected] (M.M. Ewing),

[email protected] (J.M. Thompson), [email protected](R.S. McLaren), [email protected] (V.M. Purpero),[email protected] (K.J. Thomas), [email protected](P.A. Dobrowski), [email protected] (G.A. DeGroot), [email protected](E.L. Romsos), [email protected] (D.R. Storts).

http://dx.doi.org/10.1016/j.fsigen.2016.04.0071872-4973/ã 2016 The Author(s). Published by Elsevier Ireland Ltd. All rights reserved. Thorg/licenses/by-nc-nd/4.0/).

quantification target to assess DNA degradation in forensicsamples has also been used [5–9].

The PowerQuant1 System (Promega Corporation, Madison, WI)combines multicopy quantification for each target and an internalPCR control (IPC) in one 5-dye qPCR assay to quantify total humanand human male DNA present in a sample and ascertain whetherPCR inhibitors and/or degraded human DNA are present. Hydroly-sis probes are used for detection with an Applied Biosystems1

7500 Real-Time PCR System instrument [10]. The PowerQuant1

20X Primer/Probe/IPC Mix provided in the system contains theprimers, probes, internal PCR control (IPC) template and passivereference dye. Also included in the system are the PowerQuant1

2X Master Mix and the PowerQuant1 Male gDNA Standard. ThePowerQuant1 2X Master Mix uses a hot-start chemistry for roomtemperature reaction set-up and is designed for performancecomparable to newer STR systems [11,12]. The PowerQuant1 MalegDNA Standard consists of pooled genomic human male DNA andis used to generate the standard curve for determining DNAconcentration in the samples.

is is an open access article under the CC BY-NC-ND license (http://creativecommons.

Page 2: 1-s2.0-S1872497316300631-main

M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177 167

The primers and hydrolysis probe (labeled with FAM/BHQ1-1)for the 84 base pair autosomal target quantify the total amount ofhuman DNA in a sample. This proprietary target is present at highcopy number for sensitivity, but exhibits low variation in thehuman population for consistent total human DNA to total humanmale DNA ratios ([Auto]/[Y]) [13].

The primers and hydrolysis probe (labeled with Quasar1 670/BHQ1-2) for the degradation target are used to amplify and detecta longer portion (294 bp) of the same proprietary locus as theautosomal target. To eliminate any potential for amplificationinterference of one target by the other, the autosomal anddegradation amplicons are separated by several kilobases. Thelarger degradation target is more likely to be affected bydegradation of the DNA template than the shorter autosomaltarget. When DNA is degraded, the concentration of DNA detectedwith the degradation target is less than the concentration detectedwith the autosomal target. The ratio of the human DNAconcentration detected with the autosomal target relative to thedegradation target ([Auto]/[D]) may be used to evaluate DNAdegradation.

The Y primers and probes (labeled with Cal Fluor1 Gold 540/BHQ1-1) target two different proprietary multicopy loci (81 bpand 136 bp amplicons) specific to the Y chromosome and measurethe total amount of human male DNA in a sample. Targeting twomulticopy loci increases the sensitivity while minimizing the effectthat variation in copy number of one Y locus would have on theoverall Y quantification result.

The primers and hydrolysis probe (labeled with TMR/BHQ1-2)for the internal PCR control (IPC) are used to amplify and detect anovel DNA template. The IPC’s quantification cycle (Cq) providesinformation about the presence of PCR inhibitors in a DNA sample[2–9,14]. To increase susceptibility to PCR inhibitors relative to theother targets in the PowerQuant1 System, including the 294 bpdegradation amplicon, the IPC is the longest amplicon (435 bp).

Developmental validation of the of the PowerQuant1 Systemincludes all of the studies required by the FBI Quality AssuranceStandards for Forensic DNA Testing Laboratories and the SWGDAMValidation Guidelines for DNA Analysis Methods [1,15]. The resultsdemonstrate that the PowerQuant1 System is a sensitive andreliable primate specific quantification method capable of detect-ing degraded DNA, as well as concentrations of commonlyencountered PCR inhibitors such as hematin, humic acid andtannic acid that would have an effect on an STR profile.Concentrations of human DNA as low as 0.5 pg/mL are consistentlydetected. Data from the PowerQuant1 System provide actionableinformation related to the DNA quality and subsequent analysis ofa forensic DNA sample.

2. Materials and methods

2.1. Spectral calibration

The concentrations of the PowerQuant1 Calibration Standards(i.e. system dyes) provided in the PowerQuant1 Calibration Kit are100X. The PowerQuant1 Calibration Standards were diluted to 1Xor 2X with the PowerQuant1 Calibration Buffer to prepare thesystem dye calibration plates used to calibrate the AppliedBiosystems1 7500 Real-Time PCR System (Thermo Fisher Scien-tific, Waltham, MA) [16].

2.2. DNA standard curves

For all studies, serial 25-fold dilutions of the PowerQuant1

Male gDNA Standard were amplified in duplicate to create four-point standard curves for the autosomal, Yand degradation targets.The standard curves were used to estimate a sample’s DNA

concentration based on the results of each target as described inthe PowerQuant1 System Technical Manual [16]. The DNA concen-trations in the standard curves were 50 ng/mL, 2 ng/mL, 0.08 ng/mLand 0.0032 ng/mL.

2.3. PowerQuant1 Amplification Reactions

The PowerQuant1 20X Primer/Probe/IPC Mix and the Power-Quant1 2X Master Mix were optimized for the best possibleperformance. For the majority of the studies, the PowerQuant1

System reactions contained 7 mL of Water, Amplification Grade,10 mL of PowerQuant1 2X Master Mix, 1 mL of PowerQuant1 20XPrimer/Probe/IPC Mix and 2 mL of DNA template (or water fornegative controls) for a final reaction volume of 20 mL. Studies thatused a different amplification reaction mix were noted in thissection.

As part of the studies examining PCR-based procedures,experiments were conducted to evaluate the effect fluctuationsin the concentration of primer mix or master mix had onamplification performance (e.g. pipetting error). Variable amountsof the PowerQuant1 20X Primer/Probe/IPC Mix or the Power-Quant1 2X Master Mix were added to the amplification reaction toevaluate the effects on amplification and detection. The finalconcentrations of the PowerQuant1 20X Primer/Probe/IPC Mix orthe PowerQuant1 2X Master Mix for these studies were 0.8X, 0.9X,1.0X, 1.1X and 1.2X. The volume of Water, Amplification Grade wasadjusted to accommodate changes in the volume of the primer mixor master mix in the amplification reaction.

The SwabSolutionTM Kit (Promega Corporation) was designedfor rapid DNA extraction from buccal swabs prior to amplificationwith a PowerPlex1 System. As part of the stability study, thecompatibility of the SwabSolutionTM Kit with the PowerQuant1

System was tested at one laboratory with eight replicates of 1 ng/mL male genomic DNA with different volumes of heat inactivatedSwabSolutionTM Reagent. The SwabSolutionTM Reagent wasincubated at 70 �C for 30 min as described in the SwabSolutionTM

Kit Technical Manual [17]. The volumes of SwabSolutionTM Reagent(0, 1, 2, 3, 4, or 5 mL) replaced the equivalent volume of Water,Amplification Grade in the PowerQuant1 System amplificationreaction.

2.4. DNA samples

The DNA concentrations (based on A260) and the number ofreplicates for each study were indicated in Section 3. For themajority of the studies, human DNA samples were purified withphenol:chloroform; the exceptions were noted in this section [18].

Biological material recovered from evidence items in forensiccases may include DNA from non-human sources. Speciesspecificity studies were conducted using purified DNA from higherprimates, non-primate animals, and common microorganisms.Species tested included: chimpanzee, gorilla, orangutan, pig, horse,deer, cat, rabbit, chicken, dog, cow, mouse, rat, Escherichia coli,Enterococcus faecalis, Saccharomyces cerevisiae, Fusobacteriumnucleatum, Micrococcus luteus, Streptococcus salivarius, Candidaalbicans, Streptococcus mitis, Acinetobacter lwoffii, Streptococcusmutans, and Lactobacillus acidophilus. The DNA concentrations ofthe primate species samples (private collection) ranged from0.25 ng/ml to 10 ng/mL. The other non-human DNA samples wereobtained from ATCC (Manassas, VA), Novagen (Madison, WI),Promega Corporation, and Zyagen (San Diego, CA). The DNAconcentration of the samples from all other animals and micro-organisms was 10 ng/mL. Each of the species samples was tested induplicate.

DNA isolated from oral swabs collected from three differentfemale individuals and isolated using the DNA IQTM System

Page 3: 1-s2.0-S1872497316300631-main

168 M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177

(Promega Corporation) were among the samples used to evaluatemale specificity [19]. These samples were tested undiluted anddiluted 1:10 in quadruplicate.

The human DNA samples used for the population studies wereisolated from blood samples using a modified salt out procedure[20].

Along with human DNA, PCR inhibitors may be present in aforensic evidentiary sample and could affect STR analysis. Toevaluate the effect of PCR inhibitors on the PowerQuant1 System,the following common inhibitors were tested: hematin (Sigma,cat.# H3281; Sigma-Aldrich, St. Louis, MO), humic acid (Sigma–Aldrich, cat.# 53680), tannic acid (Sigma-Aldrich, cat.# 403040),and DNA IQTM Lysis Buffer (Promega Corporation). Differentconcentrations of the inhibitors were added to 33.3 pg/mL (basedon A260) of a male genomic DNA. The same male DNA sample withno inhibitor was used as a control. This DNA concentration wasselected to maximize the amount of template with inhibitor(15 mL) added to a PowerPlex1 Fusion System (PromegaCorporation) amplification reaction to achieve the recommendedtemplate mass for a PowerPlex1 Fusion reaction (0.5 ng);resulting in a final inhibitor concentration six times higher inthe PowerPlex1 Fusion reaction than the PowerQuant1 reaction[21]. The concentrations of tannic acid in the DNA samples were:0, 333, 668 or 1000 ng/mL. The concentrations of hematin in theDNA samples were: 0, 1000, 1333 or 1668 mM. The concentrationsof humic acid in the DNA samples were: 0, 250, 500 or 750 ng/mL.The concentrations of DNA IQTM Lysis Buffer in the DNA sampleswere: 0, 0.02X, 0.04X and 0.08X. Each DNA sample was tested inquadruplicate with the PowerQuant1 System and the Power-Plex1 Fusion System (N = 4).

To allow for a controlled assessment of the detection ofdegraded DNA, a human male genomic DNA sample was exposedto different amounts of UV-C energy using a Stratalinker 1800(Stratagene, La Jolla, CA) to produce degraded DNA (no exposure,50 mJ,100 mJ and 300 mJ). The UV-C degraded samples were testedin quadruplicate on two different instruments at two differentlaboratories (N = 8).

The reproducibility of the PowerQuant1 System was testedwith the National Institute of Standards and Technology (NIST)Standard Reference Material1 (SRM) 2372 Human DNA Quantita-tion Standard (Gaithersburg, MD). Three component genomic DNAsamples were provided in SRM 2372. Component A was derivedfrom a single male donor. Component B was derived from multiplefemale donors. Component C was derived from multiple male andfemale donors. The conventional single-stranded DNA concen-trations (ng/mL) reported in the Certificate of Analysis for StandardReference Material1 2372 for components A, B and C were 57,61 and 59 ng/mL respectively [22]. Each component of the NISTSRM 2372 was quantified in duplicate at the concentrationprovided and a 1:10 dilution. The samples and dilutions weretested on four different amplification plates on three differentinstruments at two different laboratories (N = 8).

DNA was isolated from the majority of the case-type sampleswith a phenol:chloroform:isoamyl alcohol extraction method,including two non-sperm fractions recovered after differentialextraction [23]. DNA was isolated from a baby tooth and a stainfrom a piece of cloth with the DNA IQTM System [24].

2.5. Real-time PCR

The developmental validation studies with the PowerQuant1

System were performed using the Applied Biosystems1 7500 Real-Time PCR System with either Applied Biosystems1 7500 Software,Version 2.0.6 or HID Real-Time PCR Analysis Software, Version1.2 as described in the PowerQuant1 System Technical Manual [16].A two-stage thermal cycling method was used. The first stage was a

98 �C hold for 2 min. The second stage was 39 cycles of 98 �C for 15 sfollowed by 62 �C for 35 s.

The cycling protocol for the PowerQuant1 System wasdeveloped with an annealing temperature of 62 �C. Since temper-atures may vary slightly from cycler to cycler, the annealingtemperature was increased and decreased 2 �C from the recom-mended temperature as part of the studies evaluating PCR-basedprocedures to determine the effect this variation had on the overallperformance of the PowerQuant1 System.

2.6. Analysis of PowerQuant1 Data

Data were analyzed with the Applied Biosystems1 7500 Analy-sis Software 2.0.6 or HID Real-Time PCR Analysis Software, Version1.2 using analysis settings described in the PowerQuant1 SystemTechnical Manual. The results were imported into the Power-Quant1 System Analysis Tool as described in the PowerQuant1

System Technical Manual for additional calculations and analysis[16].

The calculations for the [Auto]/[Y] ratios and the [Auto]/[D]ratios were performed with the PowerQuant1 Analysis Tool. Thetool also calculated the difference in IPC Cq values of the sampleand the closest DNA standard; these values were reported as theIPC Cq Shift [16]. The IPC Cq shift may be used to evaluate thepresence of PCR inhibitors in a DNA sample. Inhibition of thePowerQuant1 System was defined as a shift �0.3 Cq. This valuewas determined experimentally with different concentrations ofinhibitors during product development by evaluating the correla-tion of amplification efficiency between the PowerPlex1 Systemsand the PowerQuant1 IPC Cq.

2.7. Normalization and STR amplification

Amplification and analysis with the PowerPlex1 Fusion Systemwas performed with samples from some studies to correlate thePowerQuant1 System results to STR DNA profiles. The maximumtemplate volume (15 mL) of the DNA samples was added to thePowerPlex1 Fusion amplification reactions for the specificity andsensitivity studies, as well as the inhibited DNA samples from thestability studies. The DNA concentrations determined with theautosomal target of the PowerQuant1 System were used tonormalize the degraded DNA samples to 0.5 ng per PowerPlex1

Fusion amplification reaction. Two microliters of each of the0.25 ng/mL DNA samples containing a mixture of female and maleDNA were added to PowerPlex1 Fusion amplification reactions fora final template mass of 0.5 ng. Based on the concentration of DNAdetected with the autosomal target of the PowerQuant1 System,the majority of the case-type samples were normalized to a finaltemplate mass in the range of 0.24–0.76 ng per PowerPlex1 Fusionamplification reaction. The amplification reactions were preparedas described in the PowerPlex1 Fusion System Technical Manualand amplified for 30 cycles [21].

A mixed DNA sample with 200 ng/mL female genomic DNA with1 ng/mL of male genomic DNA based on A260 was amplified andanalyzed with the PowerPlex1 Y23 System (Promega Corporation).The average detected concentration of male DNA based on fourreplicates tested with the PowerQuant1 System was 1.18 � 0.05 ng/mL. Based on the average concentration determined with thePowerQuant1 System, the male DNA was normalized to 0.029 ng/mL in order to achieve a 0.5 ng final DNA mass in a 25 mLPowerPlex1 Y23 amplification reaction with addition of 17.5 mLDNA template. The sample was amplified in quadruplicate asdescribed in the PowerPlex1 Y23 System Technical Manual andamplified for 30 cycles [25].

All samples were amplified using a GeneAmp1 9700 (ThermoFisher Scientific). One microliter of the PowerPlex1 Fusion System

Page 4: 1-s2.0-S1872497316300631-main

M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177 169

amplification product was added to 1 mL of CC5 Internal LaneStandard 500 (Part # DG152A) and 10 mL of HiDiTM Formamide(Thermo Fisher Scientific). One microliter of the PowerPlex1

Y23 System amplification product was added to 1 mL ofCC5 Internal Lane Standard 500 Y23 (Part # DG380A) and 10 mLof HiDiTM Formamide (Thermo Fisher Scientific). Electrophoresisand detection were performed using an Applied Biosystems1

3500xL instrument (Thermo Fisher Scientific) with a 1.2 kV, 24 sinjection or an Applied Biosystems1 3130xl instrument (ThermoFisher Scientific) with a 3 kV, 5 s injection. Except where noted,electrophoresis data from an Applied Biosystems1 3500xLinstrument were analyzed with Applied Biosystems1 Gen-eMapper1 ID-X using a 175rfu analysis threshold. Electrophoresisdata from an Applied Biosystems1 3130xl instrument wereanalyzed with Applied Biosystems1 GeneMapper1 ID-X (ThermoFisher Scientific) using a 50rfu analysis threshold.

3. Results and discussion

3.1. Species specificity

DNA was detected in all of the higher primate DNA samples(Supplemental Table 1). The higher primate DNA samples were notamplified with an STR system because cross-reactivity wasexpected. Amplification was not detected with any of thequantification targets for all of the non-primate samples, exceptfor the chicken (0.4 pg/mL with the autosomal target and 0.3 pg/mLwith the degradation target) and the S. salivarius sample (0.1 pg/mLwith the autosomal target). The chicken and the S. salivarius DNAsamples were amplified in quadruplicate with the PowerPlex1

Fusion System to determine whether human DNA was present inthese samples. Both samples generated amplification products inall replicates consistent with the low-level human DNA contami-nation detected with the PowerQuant1 System (data not shown).New sources of the chicken and S. salivarius DNA were purchasedand tested with the PowerQuant1 System. No amplification wasdetected in the replicates of the new samples confirming theoriginal samples were contaminated with human DNA rather thanspecies cross-reactivity in the PowerQuant1 System (data notshown).

3.2. Male specificity

Mixtures of male and female DNA are commonly encounteredin forensic DNA analysis. To ensure that the PowerQuant1 Systemreliably and specifically detects male DNA, the system was testedwith different concentrations of single source female genomicDNA. This sample was tested with DNA concentrations determined

Fig. 1. A total of 253 human DNA samples from three different populations were testenumber and grouped by population: African American, Caucasian and Hispanic.

by A260 of 200 ng/mL, 150 ng/mL, 100 ng/mL and 50 ng/mL; noreactivity with the Y chromosomal target was detected in any offour replicates at each DNA concentration (data not shown).Autosomal DNA concentrations detected from the samplescollected from female individuals with DNA isolated using theDNA IQTM System ranged from �26 ng/mL to �1 ng/mL (data notshown) and no male DNA was detected in any of the replicatesdemonstrating male specificity of the Y chromosomal target.

3.3. Population study

Multicopy targets present in high copy number improve assaysensitivity. However, the number of copies of these targets mayvary from individual to individual [2,26]. To evaluate the variationof the multicopy targets in the PowerQuant1 System, individualhuman DNA samples from three different male populations:African American (N = 85), Caucasian (N = 83), and Hispanic(N = 85) were tested at the National Institute of Standards andTechnology to evaluate the variation in the DNA concentrationdetected with the autosomal target relative to the Y chromosomaltarget ([Auto]/[Y] ratio) as well as the degradation target ([Auto]/[D] ratio).

[Auto]/[Y] ratios were calculated for all of the samples. Datafrom all three populations were combined and plotted (Fig. 1). Theaverage [Auto]/[Y] ratio was 1.06 � 0.14 (N = 253). The coefficient ofvariation was 13.3%. The lowest [Auto]/[Y] ratio was 0.53 and thehighest ratio was 1.83.

[Auto]/[D] ratios were calculated for all of the samples. Datafrom all three populations were combined and plotted (Supple-mental Fig. 1). The average [Auto]/[D] ratio was 1.16 � 0.15(N = 253). The coefficient of variation was 12.8%. The lowest[Auto]/[D] ratio was 0.70 and the highest ratio was 1.60.

3.4. Dynamic range and sensitivity

A male genomic DNA sample was serially diluted to differentconcentrations based on A260 (100, 50, 10, 2, 0.4, 0.08, 0.016, and0.0032 ng/mL) to evaluate the dynamic range of the PowerQuant1

System. Each DNA concentration was tested in quadruplicate attwo different laboratories on three different instruments (N = 12)(Table 1). The highest percentage difference for all three targetswas observed with the 100 ng/mL DNA sample.

Additionally, the sensitivity of the assay was tested with twodifferent male genomic DNA samples serially diluted to thefollowing concentrations: 1 pg/mL, 0.5 pg/mL, 0.25 pg/mL, 0.125 pg/mL, and 0.0625 pg/mL. Each DNA concentration was tested inquadruplicate at two different laboratories on three differentinstruments (N = 12). The number of replicates with DNA detected

d with the PowerQuant1 System. The [Auto]/[Y] ratio was plotted against sample

Page 5: 1-s2.0-S1872497316300631-main

Table 1The average DNA concentration and standard deviations detected with the autosomal, Y, and degradation targets (N = 12), as well as the percentage difference between theaverage DNA concentration detected with each PowerQuant1 System target and the DNA concentration based on A260.

Sample Concentration Based onA260 (ng/mL)

[Autosomal]Average (ng/mL)

[Autosomal]Percentage Difference

[Y]Average (ng/mL)

[Y] PercentageDifference

[Degradation]Average (ng/mL)

[Degradation] PercentageDifference

0.0032 0.0026 � 0.0004 19% 0.0032 � 0.0009 1.4% 0.0024 � 0.0005 28%0.016 0.014 � 0.002 14% 0.016 � 0.003 2.0% 0.012 � 0.002 29%0.08 0.066 � 0.009 18% 0.083 � 0.008 3.9% 0.066 � 0.003 19%0.4 0.35 � 0.04 13% 0.42 � 0.04 5.0% 0.34 � 0.03 16%2 1.7 � 0.2 16% 2.1 � 0.1 4.9% 1.7 � 0.1 16%10 7.7 � 0.6 26% 9.6 � 0.5 4.3% 7.9 � 0.8 24%50 37 � 4 31% 48 � 6 4.4% 40 � 5 22%100 70 � 4 35% 88 � 4 13% 73 � 3 31%

Table 2The number of replicates with DNA detected are displayed for the PowerQuant1 System targets and DNA concentrations tested. An asterisk (*) was used to denote only threereplicates were included for the 1 pg/mL Male 1 sample because at two loci, three alleles were detected; one allele at each of these loci was not consistent with the knownprofile.

Male 1 DNA concentration PowerQuant1 System PowerPlex1 Fusion System

Autosomal (N = 12) Degradation (N = 12) Y (N = 12) Total template mass (pg) Average% alleles detected (N = 4)

1 pg/mL 12 12 12 15 75 � 9*0.5 pg/mL 12 11 11 7.5 41 � 40.25 pg/mL 8 6 8 3.8 22 � 70.125 pg/mL 3 2 2 1.9 4.0 � 30.0625 pg/mL 3 0 3 0.94 3.0 � 3

Male 2 DNA concentration PowerQuant1 System PowerPlex1 Fusion System

Autosomal (N = 12) Degradation (N = 12) Y (N = 12) Total template mass (pg) Average% alleles detected (N = 4)

1 pg/mL 12 12 12 15 68 � 10.5 pg/mL 12 10 12 7.5 36 � 40.25 pg/mL 11 6 7 3.8 23 � 90.125 pg/mL 3 3 3 1.9 7.6 � 30.0625 pg/mL 2 0 0 0.94 7.0 � 2

No template control PowerQuant1 System PowerPlex1 Fusion System

Autosomal (N = 24) Degradation (N = 24) Y (N = 24) Total template mass (pg) Average% alleles detected (N = 4)

1 1 1 0 0

170 M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177

for every concentration of each sample are shown in Table 2. ThePowerQuant1 System consistently detected 0.5 pg/mL with theautosomal target. DNA was detected in one of the twenty-fournegative control replicates. The DNA concentration determinedwith the autosomal target for that replicate was 0.5 pg/mL,

Fig. 2. Titration of four different concentrations of tannic acid in a DNA sample with 33replicates are displayed. Samples with an IPC Cq shift >0.3 Cq or [Auto]/[D] ratios >2 aradded to a PowerPlex1 Fusion System reaction. Amplification reactions were performedinstrument using a 1.2 kV, 24 s injection. Representative electropherograms for each co

suggesting low-level DNA contamination. At the DNA concen-trations tested, the number of alleles detected varied betweenreplicates. A partial STR profile was detected for each of thereplicates of the 0.5 pg/mL male DNA samples (Supplemental

.3 pg/mL of male DNA. The average IPC Cq shift and the standard deviation for foure highlighted. Fifteen microliters of each DNA sample mixed with tannic acid were using a GeneAmp1 9700 and 1 mL of each sample was electrophoresed on a 3500xLncentration of inhibitor are displayed.

Page 6: 1-s2.0-S1872497316300631-main

Fig. 3. PowerPlex1 Fusion System amplification reactions with corresponding UV-C exposure and the average [Auto]/[D] ratios determined with the PowerQuant1 Systemand standard deviations (N = 8). Samples with an [Auto]/[D]ratio >2 or an [Auto]/[Y] ratio � 2 are highlighted. The PowerPlex1 Fusion amplification reactions contained 0.5 ngof DNA exposed to increasing amounts of UV-C radiation and were performed using the GeneAmp1 PCR System 9700. One microliter of each amplified sample waselectrophoresed on an Applied Biosystems1 3130xl instrument using a 3 kV, 5 s injection.

Table 3The average DNA concentration (ng/mL) and coefficient of variation (CV) detected with each of the targets in the PowerQuant1 System for three different DNAconcentrations tested on three different instruments. The number of replicates tested on each instrument are indicated in the table.

DNA Concentration and Run # Average [Autosomal](ng/mL)

[Autosomal] CV Average [Degradation](ng/mL)

[Degradation] CV Average [Y](ng/mL)

[Y] CV

40 ng/mL DD302 Instrument #1 (N = 8) 39 � 2 6.2% 40 � 1 3.2% 38 � 1 2.6%40 ng/mL DD302 Instrument #2 (N = 8) 39 � 2 4.4% 42 � 2 5.1% 40 � 2 4.8%40 ng/mL DD302 Instrument #3 (N = 3) 43 � 3 7.2% 38 � 4 10% 35 � 3 9.4%2 ng/mL DD302 Instrument #1 (N = 7) 2.1 � 0.2 7.5% 2.1 � 0.1 4.4% 1.8 � 0.1 3.2%2 ng/mL DD302 Instrument #2 (N = 8) 2.0 � 0.1 6.0% 2.0 � 0.1 7.0% 1.9 � 0.1 4.5%2 ng/mL DD302 Instrument #3 (N = 3) 2.1 � 0.4 18% 1.9 � 0.4 21% 1.9 � 0.3 15%0.1 ng/mL DD302 Instrument #1 (N = 8) 0.13 � 0.01 11% 0.11 � 0.01 7.2% 0.092� 0.007 7.2%0.1 ng/mL DD302 Instrument #2 (N = 8) 0.11 � 0.01 5.5% 0.10 � 0.01 7.8% 0.10 � 0.00 5.2%0.1 ng/mL DD302 Instrument #3 (N = 3) 0.098 � 0.011 11% 0.092 � 0.013 14% 0.096 � 0.017 18%

M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177 171

Fig. 2). A few alleles were detected from each of the replicates ofthe 0.0625 pg/mL male DNA samples (Supplemental Fig. 3).

3.5. Stability studies

3.5.1. InhibitorsThe same non-degraded DNA sample was used for all inhibited

samples; the condition of the DNA was confirmed with the [Auto]/[D] ratios and the STR profiles obtained from the non-inhibitedsample. A gradual increase in the IPC Cq shift was observed withincreasing tannic acid concentration (Fig. 2). The [Auto]/[D] ratioalso increased gradually with the tannic acid concentration.However, the IPC Cq shift indicated inhibition at a lowerconcentration of tannic acid than the concentration of tannicacid that increased the [Auto]/[D] ratio greater than the value of

Table 4The average percentage error determined with the PowerQuant1 System based on the 50different DNA concentrations tested on three different instruments. The number of rep

DNA Concentration and Run # Percentage Error [Autosomal]

40 ng/mL DD302 Run #1 (N = 8) 2.8%

40 ng/mL DD302 Run #2 (N = 8) 1.6%

40 ng/mL DD302 Run #3 (N = 3) 6.7%

2 ng/mL DD302 Run #1 (N = 7) 3.7%

2 ng/mL DD302 Run #2 (N = 8) 0.31%

2 ng/mL DD302 Run #3 (N = 3) 3.8%

0.1 ng/mL DD302 Run #1 (N = 8) 27%

0.1 ng/mL DD302 Run #2 (N = 8) 7.9%

0.1 ng/mL DD302 Run #3 (N = 3) 2.2%

2.0 typically associated with degraded DNA (Fig. 3). Theseobservations suggested the high [Auto]/[D] ratio observed with100 ng/mL of tannic acid was a function of the inhibitorconcentration affecting amplification performance of the degra-dation target rather than the integrity of the DNA. All of theexpected alleles (N = 43) were detected in the PowerPlex1 FusionSTR profiles with no tannic acid or with a final concentration of200 ng/mL tannic acid in the amplification reaction. With 400 ng/mL and 600 ng/mL tannic acid, an average of 42 � 2 and38 � 1 alleles, respectively were detected. However, the signalintensity of these alleles was low, consistent with an inhibitedSTR amplification reaction. An IPC Cq shift >0.3 Cq was observedin the PowerQuant1 System data for all of the concentrations oftannic acid that affected the PowerPlex1 Fusion System STRprofile.

ng/mL concentration supplied for the PowerQuant1 Male gDNA Standard for threelicates tested on each instrument are indicated in the table.

Percentage Error [Degradation] Percentage Error [Y]

0.065% 4.2%4.2% 0.59%5.2% 12%3.8% 9.1%1.3% 6.3%4.1% 3.5%12% 8.4%0.52% 4.9%7.5% 4.1%

Page 7: 1-s2.0-S1872497316300631-main

Table 6The average DNA concentration and IPC Cq results for eight replicates of each component of SRM 2372 at the concentration provided and a 1:10 dilution of each component.The single-stranded DNA concentrations (ng/mL) for components A, B and C stated in the Certificate of Analysis for Standard Reference Material1 (SRM) were 57, 61 and 59 ng/mL respectively. N/A indicates not applicable.

Sample Average [Autosomal](ng/mL)

[Autosomal] CV Average [Degradation](ng/mL)

[Degradation] CV Average [Y](ng/mL)

[Y] CV AverageIPC Cq

IPC CV

NIST SRM 2372 A 51 � 12 23% 45 � 9 21% 55 � 11 20% 21.01 � 0.29 1.4%NIST SRM 2372 A diluted 1:10 4.9 � 1.1 22% 4.6 � 0.8 17% 5.2 � 0.8 15% 20.84 � 0.26 1.3%NIST SRM 2372 B 57 � 11 19% 45 � 10 21% N/A N/A 20.69 � 0.30 1.4%NIST SRM 2372 B diluted 1:10 5.5 � 1.1 20% 4.7 � 0.8 17% N/A N/A 20.67 � 0.28 1.4%NIST SRM 2372 C 46 � 6 12% 38 � 3 6.7% 31 � 2 5.3% 20.86 � 0.31 1.5%NIST SRM 2372C diluted 1:10 4.6 + 0.9 19% 4.0 � 0.4 10% 3.1 � 0.3 11% 20.74 � 0.25 1.2%

Table 5The intra-run and inter-run average DNA concentration (ng/mL) and coefficient of variation (CV) detected with each of the targets in the PowerQuant1 System for a 1 ng/mLhuman male DNA sample. The number of replicates for each run are indicated in the table. Runs #1 through #3 were performed on one instrument; runs #4 and #5 wereperformed on a different instrument.

Run Average [Autosomal](ng/mL)

[Autosomal] CV Average [Degradation](ng/mL)

[Degradation] CV Average [Y](ng/mL)

[Y] CV AverageIPC Cq

IPC Cq CV

Run #1 (N = 8) 0.99 � 0.13 14% 0.88 � 0.04 5.0% 1.0 � 0.0 3.3% 20.48 � 0.11 0.53%Run #2 (N = 8) 1.0 + 0.1 11% 1.0 � 0.1 5.2% 1.1 � 0.1 7.2% 20.93 � 0.20 0.93%Run #3 (N = 8) 0.74 + 0.08 11% 0.72 � 0.03 4.3% 0.83 � 0.04 4.9% 20.70 � 0.09 0.42%Run #4 (N = 8) 0.80 + 0.06 7.4% 0.82 � 0.06 6.9% 0.99 � 0.03 2.6% 21.07 � 0.11 0.51%Run #5 (N = 8) 0.93 + 0.03 3.4% 0.81 � 0.05 6.7% 1.0 � 0.0 4.7% 20.85 � 0.10 0.50%Inter-run (N = 40) 0.90 � 0.14 16% 0.85 � 0.11 12% 0.98 � 0.09 10% 20.80 � 0.24 1.1%

172 M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177

A gradual increase in the IPC Cq shift was observed withincreasing hematin concentration (Supplemental Fig. 4). None ofthe [Auto]/[D] ratios exceeded the value 2.0. All of the expectedalleles (N = 43) were detected in the STR profiles with no hematinor a final concentration of 600 mM hematin in the PowerPlex1

Fusion amplification reaction; although the average peak heightdetected was lower in the STR profiles with 600 mM hematin. With800 mM hematin in the PowerPlex1 Fusion reaction, an average of37 � 7 alleles were detected with allelic dropout observed at

Fig. 4. The ratio of female DNA to male genomic DNA based on concentrations determiSystem (secondary vertical axis). The percentage of alleles unique to the male contribueach ratio are displayed on the primary vertical axis. The PowerPlex1 Fusion amplificaSystem 9700. One microliter of each amplified sample was electrophoresed on an Ap

Amelogenin, Penta D and Penta E. An average of 16 � 7 alleles weredetected with the PowerPlex1 Fusion reaction with 1000 mMhematin. An IPC Cq shift >0.3 Cq was observed in the PowerQuant1

System data for all of the concentrations of hematin that affectedthe PowerPlex1 Fusion System STR profile.

A gradual increase in the IPC Cq shift was observed withincreasing humic acid concentration (Supplemental Fig. 5). All ofthe [Auto]/[D] ratios were less than the value 2.0. All of theexpected alleles (N = 43) were detected in the STR profiles with no

ned with A260 was plotted against the [Auto]/[Y] ratio detected with the PowerQuant1

tor detected after amplification and analysis with the PowerPlex1 Fusion System fortion reactions contained 0.5 ng of DNA and were performed using the GeneAmp1 PCRplied Biosystems1 3500xL instrument using a 1.2 kV, 24 s injection.

Page 8: 1-s2.0-S1872497316300631-main

Fig. 5. The average DNA concentration detected with the autosomal target for the PowerQuant1 System was plotted on the primary vertical axis against the [Auto]/[Y] ratiobased on A260. The average DNA concentration detected with the Y chromosomal target for the PowerQuant1 System was plotted on the secondary vertical axis.

M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177 173

humic acid or a final concentration of 150 ng/mL humic acid in thePowerPlex1 Fusion amplification reaction. With 300 ng/mL humicacid in the PowerPlex1 Fusion reaction, an average of 18 � 0 alleleswere detected. With 450 ng/mL humic acid in the PowerPlex1

Fusion reaction, one allele was detected at the TH01 locus in one

Table 7Summary of the [Auto]/[Y], [Auto]/[D], IPC and STR data for thirteen case-type sample

replicate. An IPC Cq shift >0.3 Cq was observed in the PowerQuant1

System data for all of the concentrations of humic acid that affectedthe PowerPlex1 Fusion System STR profile.

An IPC Cq shift was observed with 0.004X DNA IQTM Lysis Bufferin the PowerQuant1 amplification reaction (Supplemental Fig. 6).

s. Samples with an [Auto]/[Y] ratio > 2 or an [Auto]/[D] ratio >2 are highlighted.

Page 9: 1-s2.0-S1872497316300631-main

ble

8su

mmaryof

thestan

dardcu

rves

generated

withea

chfinal

conce

ntrationof

thePo

werQuan

t120

XPrim

er/Probe

/IPC

Mix

intheqP

CRreaction

.Thestan

dardcu

rves

werecrea

tedwithduplicate

amplification

ofthe25

-foldserial

lution

ofthePo

werQuan

t1MalegD

NAStan

dard.

Final

PowerQuan

t120

XPrim

er/Probe

/IPCMix

Con

centration

0.8X

0.9X

1.0X

1.1X

1.2X

Stan

dardCurvePa

rameters

Autoso

mal

YDeg

radation

Autoso

mal

YDeg

radation

Autoso

mal

YDeg

radation

Autoso

mal

YDeg

radation

Autoso

mal

YDeg

radation

R2

0.99

81.00

1.00

1.00

0.99

90.99

91.00

0.99

91.00

0.99

80.99

71.00

0.99

91.00

0.99

9Slop

e�3

.364

�3.268

�3.361

�3.320

�3.302

�3.438

�3.392

�3.326

�3.390

�3.322

�3.387

�3.469

�3.368

�3.367

�3.412

Y-intercep

t26

.01

26.10

26.31

26.06

26.29

26.43

26.33

26.51

26.62

26.49

26.73

26.73

26.48

26.83

26.85

Amplifi

cation

Efficien

cy98

.27%

102.31

%98

.40%

100.08

%10

0.84

%95

.37%

97.16%

99.82%

97.22%

100.00%

97.34%

94.21%

98.10%

98.15%

96.38%

174 M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177

No Cq was detected for the IPC or the degradation target with0.008X DNA IQTM Lysis Buffer in the PowerQuant1 amplificationreaction. The IPC Cq shift indicated inhibition at a lowerconcentration of DNA IQTM Lysis Buffer than the concentrationthat caused amplification failure of the degradation target. Allexpected alleles (N = 43) were detected in the STR profiles with noDNA IQTM Lysis Buffer in the PowerPlex1 Fusion amplificationreaction. With 0.012X DNA IQTM Lysis Buffer in the PowerPlex1

Fusion reaction, an average of 40 � 0.5 alleles were detected withdropout occurring at the Penta E and Penta D loci. No alleles weredetected with 0.024X or 0.048X DNA IQTM Lysis Buffer added to thePowerPlex1 Fusion amplification reaction.

3.5.2. Degraded DNAThe [Auto]/[Y] and [Auto]/[D] ratios for each amount of UV-C

exposure were calculated. These DNA samples were also amplifiedwith the PowerPlex1 Fusion System in order to correlate the[Auto]/[D] ratio to an STR profile. As the UV-C exposure increased,the [Auto]/[D] ratios also increased. The [Auto]/[Y] ratio graduallyincreased with UV-C exposure, demonstrating the [Auto]/[Y] ratiowas minimally affected until the DNA was severely degraded. DNAexposure to UV-C did not affect the IPC Cq. The higher [Auto]/[D]ratios corresponded to STR profiles exhibiting characteristics ofdegraded DNA (Fig. 3).

3.5.3. SwabSolutionTM KitNo significant effects on the IPC or the quantification results for

each target were observed with any of the volumes of SwabSo-lutionTM Reagent tested (data not shown).

3.6. Precision and accuracy

3.6.1. PrecisionTo evaluate the precision of the PowerQuant1 System, the

PowerQuant1 Male gDNA Standard (Part # DD302) was seriallydiluted to concentrations within the linear range of the standardcurve based on the 50 ng/mL concentration provided. Theconcentrations tested were 40, 2 and 0.1 ng/mL (based on thestarting concentration). Replicates of each concentration werequantified on three different instruments at two differentlaboratories. The number of replicates for each DNA concentration,the average DNA concentration determined for each target, and thecoefficient of variation (CV) are displayed in Table 3.

3.6.2. AccuracyThe samples used for the precision study were also used to

evaluate the accuracy of the PowerQuant1 System. The percentageerror was determined based on the expected concentrations of thePowerQuant1 Male gDNA Standards and the measured concen-trations with the PowerQuant1 System. The number of replicatesfor each DNA concentration and the percentage error for eachtarget are displayed in Table 4. In general, the percentage error wasfairly low. Higher percentage error was observed in some replicatesof the 0.1 ng/mL samples which was consistent with expectationsfor samples with lower DNA concentration.

3.6.3. RepeatabilityTo evaluate the repeatability of the PowerQuant1 System, a

human male genomic DNA sample was diluted to 1 ng/mL (A260)and eight replicates of the sample were quantified with fivedifferent PowerQuant1 runs on two different instruments. Theintra-run and inter-run average DNA concentration detected andthe CV for each target are shown in Table 5. The average IPC Cq andthe coefficient of variation are also shown. In general, the intra-runCV was less than 15% and the inter-run CV was 16% or less.

Ta A di

Page 10: 1-s2.0-S1872497316300631-main

M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177 175

3.6.4. ReproducibilityThe NIST SRM 2372 components were tested on four different

amplification plates on three different instruments at two differentlaboratories (N = 8). The inter-run average DNA concentrationdetected and coefficient of variation for each target are displayed inTable 6. The average IPC Cq and the coefficient of variation are alsoshown. In general, the CV was less than 23% for the quantificationtargets.

3.7. Mixture studies

The ability to detect mixtures of male and female DNA with aquantification system, reliably providing a ratio of total human tototal human male DNA, allows a forensic DNA analyst to use thatinformation to guide the selection of an STR system that will resultin the most useful DNA profile for that sample (e.g. autosomal STRsor Y-STRs). Detection of mixtures of male and female DNA with thePowerQuant1 System was tested with three different mixture setsprepared from male and female genomic DNA.

The total DNA concentration of the first mixture set was 0.25 ng/mL with female DNA to male DNA ratios of 30:1, 20:1, 10:1, 5:1,2:1 and 1:1. All of the samples in this mixture set were tested inquadruplicate on three different instruments at two differentlaboratories (N = 12) and the [Auto]/[Y] ratio was plotted againstthe female DNA to male DNA ratio (secondary vertical axis inFig. 4). To correlate the mixture ratios with STR profiles, thesamples were amplified and analyzed with the PowerPlex1 FusionSystem. The number of alleles unique to the male contributor werecounted and the percentage of alleles unique to the malecontributor was determined (primary vertical axis in Fig. 4).

The second mixture set included a constant male DNAconcentration of 1 pg/mL (based on A260) with increasing concen-trations of female DNA (0, 50, 100 or 200 ng/mL based on A260).Data shown in Fig. 5 were from four replicates run on oneinstrument. The autosomal DNA concentration detected with thePowerQuant1 System increased as the ratio of female DNA to maleDNA increased; however, the amount of male DNA detected withthe Y chromosomal target of the PowerQuant1 System remainedconstant at approximately 1 pg/mL. This mixture set demonstratedthat 1 pg/mL of male DNA was reliably detected in a background of200,000 fold excess of female genomic DNA.

The third mixture set included samples with a constant femaleDNA to male DNA ratio of 200:1 while decreasing the total DNAconcentration. This mixture set was prepared with a mixed DNAsample with 200 ng/mL female genomic DNA and 1 ng/mL malegenomic DNA based on A260. This mixed DNA sample was seriallydiluted 10-fold for two additional concentrations: 20 ng/mL femaleDNA with 0.1 ng/mL male DNA and 2 ng/mL female DNA with0.01 ng/mL male DNA. Data shown in Supplemental Fig. 7 werefrom four replicates tested on one instrument. This mixture setdemonstrated that the [Auto]/[Y] ratio detected with the Power-Quant1 System for the mixture sample remained constant as thetotal concentration of DNA decreased.

For a sample with a 200:1 ratio of female to male genomic DNA,Y-STR testing is required for analysis of the male DNA. The average

Table 9A summary of the average DNA concentration and IPC Cq and standard deviation foconcentration of the PowerQuant1 20X Primer/Probe/IPC Mix in the amplification reac

Final PowerQuant1 20X Primer/Probe/IPC Mix Concentration [Autosom

0.8X 0.92 � 00.9X 0.84 � 01.0X 0.97 � 01.1X 1.1 � 01.2X 1.0 � 0

concentration of male DNA detected with the Y target of the mixedDNA sample with 200 ng/ml female genomic DNA and 1 ng/mLmale genomic DNA was 1.18 � 0.05 ng/mL. This concentration wasused to normalize the male DNA to 0.029 ng/mL for amplification of500 pg of template (based on 17.5 mL addition to the amplificationreaction) with the PowerPlex1 Y23 System. A complete Y-STRprofile was detected with peak heights in the expected rfu rangedemonstrating the concentration of male DNA detected is reliablefor normalizing male DNA prior to Y-STR analysis (SupplementalFig. 8).

3.8. Case-type samples

A total of thirteen non-probative and mock casework sampleswere tested with the PowerQuant1 System and the PowerPlex1

Fusion System in the Madison and Milwaukee laboratories in theWisconsin State Crime Laboratory Bureau. Four touched items,three bloodstains, a reference sample collected in 2001, a cigarettebutt, one baby tooth, and three male/female mixed DNA sampleswere quantified with the PowerQuant1 System and amplified forautosomal STRs with the PowerPlex1 Fusion System. An [Auto]/[Y]ratio greater than two was used to indicate that a sample was amixture of female and male DNA. An [Auto]/[D] ratio greater thantwo was used to indicate a DNA sample was degraded. The sampleswere also evaluated for the presence of PCR inhibitors with thePowerQuant1 System IPC. An IPC shift greater than 0.3 Cq was usedto indicate inhibition; however, differential amplification of lociconsistent with inhibition was not observed in any of the STRprofiles. An STR profile was considered a mixture if three or morealleles were detected at more than one locus. An STR profile wasconsidered degraded if the rfu values of the smaller loci were atleast two times greater than the rfu values at the larger loci(sloped). The correlation of the [Auto]/[Y] ratios, [Auto]/[D] ratios,and IPC Cq shift to the STR profiles were evaluated and aredisplayed in Table 7.

The STR profile detected from the cigarette butt was consistentwith a single source female contributor (data not shown). A malecontributor (0.008 ng/mL) was detected with the Y chromosomaltarget of the PowerQuant1 System with an [Auto]/[Y] ratio of287.13 (Table 7). Failure to detect a male contributor with anautosomal STR system was consistent with the [Auto]/[Y] ratio.

The concentration of DNA detected from the cutting of a leafwith a bloodstain was 1.8 pg/mL. Addition of the maximum amountof this DNA sample (27 pg) to a PowerPlex1 Fusion amplificationreaction resulted in a partial profile (locus drop-out at Penta E andDYS391) with low peak heights (data not shown). Three alleleswere detected at the D16S539 locus. The PowerQuant1 System didnot detect a mixture of male and female DNA (Table 7).

The PowerPlex1 Fusion System generated a profile consistentwith a least two contributors (major female, minor male) from thenon-sperm fraction of a semen-positive bloodstain in a pair ofunderwear. The only indications of male DNA in the STR profilewere the Y allele at Amelogenin and the allele detected at DYS391(Supplemental Fig. 9). The [Auto]/[Y] ratio detected with thePowerQuant1 System was 22.11 while the [Auto]/[D] ratio was

r four replicates of a 1 ng/mL (A260) male DNA sample detected with each finaltion.

al] [Degradation] [Y] IPC Cq

.05 0.81 � 0.05 0.97 � 0.05 20.39 � 0.06

.05 0.75 � 0.07 0.93 � 0.05 20.39 � 0.22

.10 0.88 � 0.06 1.1 � 0.0 20.52 � 0.04

.0 0.93 � 0.07 1.2 � 0.0 20.73 � 0.05

.0 1.0 � 0.1 1.2 � 0.0 20.93 � 0.05

Page 11: 1-s2.0-S1872497316300631-main

Table

10Asu

mmaryof

thestan

dardcu

rves

generated

withea

chfinal

conce

ntrationof

thePo

werQuan

t12X

MasterMix

intheam

plification

reaction

.Thestan

dardcu

rves

werecrea

tedwithduplicate

amplification

ofthe25

-foldserial

dilution

ofthePo

werQuan

t1MalegD

NAStan

dard.

PowerQuan

t12X

MasterMix

0.8X

0.9X

1.0X

1.1X

1.2X

Stan

dardCurvePa

rameters

Autoso

mal

YDeg

radation

Autoso

mal

YDeg

radation

Autoso

mal

YDeg

radation

Autoso

mal

YDeg

radation

Autoso

mal

YDeg

radation

R2

0.99

71.00

0.99

90.99

91.00

0.99

81.00

0.99

90.99

90.99

90.99

90.99

90.99

91.00

0.99

9Slop

e�3

.286

�3.220

�3.301

�3.281

�3.318

�3.390

�3.271

�3.445

�3.346

�3.373

�3.377

�3.492

�3.358

�3.443

�3.502

Y-intercep

t26

.48

26.49

26.79

26.24

26.45

26.64

26.30

26.59

26.55

26.32

26.55

26.51

26.13

26.63

26.65

Amplification

Efficien

cy10

1.51

%10

4.42

%10

0.87

%10

1.72

%10

0.17

%97

.24%

102.18

%95

.10%

99.02%

97.90%

97.75%

93.35%

98.52%

95.17%

92.99%

176 M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177

2.01, which is on the low end of ratios that may indicatedegradation (Table 7).

3.9. PCR-based procedures

3.9.1. Annealing temperaturePerformance with different annealing temperatures was tested

with duplicate standard curves, quadruplicate negative controls,and duplicates of the following extracted DNA samples on twodifferent instruments at one laboratory: female DNA (205 and50 ng/mL); male DNA (1 and 0.001 ng/mL); and a mixed DNAsample with 1 pg/mL male contributor and 100 ng/mL femalecontributor. At all annealing temperatures, the standard curveparameters were within the expected ranges described in thePowerQuant1 System Technical Manual (Supplemental Table 2)[16]. For all annealing temperatures tested, no quantificationvalues were detected with any of the targets in the negativecontrols. No changes in specificity of the Y target were observed inany of the female DNA samples with either 60 �C or 64 �C annealingtemperature. No differences in sensitivity of any of the quantifica-tion targets were observed when the annealing temperature wasaltered by 2 �C in either direction (data not shown).

3.9.2. PowerQuant1 20X Primer/Probe/IPC Mix concentrationA male DNA sample (1 ng/mL determined with A260) was

quantified in quadruplicate with each reaction mix. Changes in thefinal concentration of the PowerQuant1 20X Primer/Probe/IPC Mixdid not have any significant effect on the standard curves or thequantification of the male DNA sample (Tables 8 and 9).

3.9.3. PowerQuant1 2X master Mix concentrationA male DNA sample (1 ng/mL determined with A260) was

quantified in quadruplicate with each reaction mix. Changes in thefinal concentration of the PowerQuant1 2X Master Mix did nothave any significant effect on the standard curves or thequantification of the male DNA sample, although a later IPC Cqvalue was observed with 1.2X final concentration of the Power-Quant1 2X Master Mix (Tables 10 and 11). This corresponds to a12 mL addition of 2X Master Mix per 20 mL amplification reactionrather than the recommended 10 mL. The observed effect on theIPC can likely be attributed to the higher cation concentrationreducing amplification efficiency of the 435 bp IPC amplicon. Thiseffect was consistent with data generated during productdevelopment with titration of individual master mix components.

4. Conclusions

The PowerQuant1 System was designed not only to providetotal human and total human male DNA concentrations in a DNAsample extracted from a forensic evidence item, but also dataabout the condition of the DNA sample. The additional dataobtained at quantification may be used by the analyst to determinethe most effective procedure for processing the sample for STRanalysis. The experiments presented in this paper verify thePowerQuant1 System is a robust method producing reliable andreproducible results and address the SWGDM Validation Guide-lines for evaluating a new methodology. The results of theexperiments completed at the collaborating laboratories demon-strate that the DNA concentrations determined with the Power-Quant1 System may be used to normalize DNA samples for STRamplification. The system routinely detects DNA at concentrationstoo low to yield interpretable STR profiles. Low levels of male DNAwere detected in samples with high concentrations of female DNA.Since the IPC amplicon has greater sensitivity to PCR inhibitorsthan the degradation amplicon, a sample with an [Auto]/[D] ratiogreater than two and no observed IPC Cq shift likely contains

Page 12: 1-s2.0-S1872497316300631-main

Table 11A summary of the average DNA concentration and IPC Cq and standard deviation for four replicates of a 1 ng/mL (A260) male DNA sample detected with each finalconcentration of the PowerQuant1 2X Master Mix in the amplification reaction. One replicate of the IPC failed to cross the detection threshold with a 1.2X final concentration.Data for that sample are indicated with an * (N = 3).

Final PowerQuant1 2X Master Mix Concentration [Autosomal] [Degradation] [Y] IPC Cq

0.8X 0.84 � 0.10 0.78 � 0.03 0.94 � 0.04 20.29 � 0.070.9X 0.81 � 0.08 0.78 � 0.02 0.94 � 0.04 20.38 � 0.131.0X 0.94 � 0.05 0.82 � 0.07 1.1 � 0.1 20.64 � 0.151.1X 1.1 � 0.1 0.92 � 0.10 1.2 � 0.1 21.02 � 0.211.2X 1.0 � 0.1 0.86 � 0.10 1.2 � 0.1 31.47 � 16.71*

M.M. Ewing et al. / Forensic Science International: Genetics 23 (2016) 166–177 177

degraded DNA rather than PCR inhibitors. With PCR inhibitors inthe PowerQuant1 reaction, no instances of an [Auto]/[D] ratiogreater than two were observed without a shift in the IPC Cq,preventing the false identification of a degraded DNA sample dueto PCR inhibition. Additionally, the PowerQuant1 System providesan analyst data to assess the presence of PCR inhibitors and DNAdegradation that could affect the corresponding STR profilesmaking it a useful tool for evaluating the quality of an extractedDNA sample.

Appendix A. Supplementary data

Supplementary data associated with this article can befound, in the online version, at http://dx.doi.org/10.1016/j.fsigen.2016.04.007.

References

[1] Quality Assurance Standards for Forensic DNA Testing Laboratories. Availableat: https://www.fbi.gov/about-us/lab/biometric-analysis/codis/qas-standards-for-forensic-dna-testing-laboratories-effective-9-1-2011 (accessed02.05.16).

[2] Benjamin E. Krenke, Nadine Nassif, Cynthia J. Sprecher, Curtis Knox, MelissaSchwandt, Douglas R. Storts, Developmental validation of a real-time PCRassay for the simultaneous quantification of total human and male DNA,Forensic Sci. Int. Genet. 3 (2008) 14–21.

[3] Robert L. Green, Ines C. Roinestad, Cherisse Boland, Lori K. Hennessy,Developmental validation of the QuantifilerTM Real-Time PCR Kits for thequantification of human nuclear DNA samples, J. Forensic Sci. 50 (2005) 809–825.

[4] Maura Barbisin, Rixun Fang, Cristin E. O’Shea, Lisa M. Calandro, Monohar R.Furtado, J.G. Shewale, Developmental validation of the Quantifiler1 Duo DNAQuantification Kit for simultaneous quantification of total human and humanmale DNA and detection of PCR inhibitors in biological samples, J. Forensic Sci.54 (2009) 305–319 (correction available in Journal of Forensic Sciences. 57(2012) 282.).

[5] Katie L. Swango, Mark D. Timken, Mavis Date Chong, Martin R. Buoncristiani, Aquantitative PCR assay for the assessment of DNA degradation in forensicsamples, Forensic Science Int. 158 (2006) 14–26.

[6] Katie L. Swango, William R. Hudlow, Mark D. Timken, Martin R. Buoncristiani,Developmental validation of a multiplex qPCR assay for assessing the quantityand quality of nuclear DNA in forensic samples, Forensic Sci. Int. 170 (2007)35–45.

[7] William R. Hudlow, Mavis Date Chong, Katie L. Swango, Mark D. Timken,Martin R. Buoncristiani, A quadruplex real-time qPCR assay for thesimultaneous assessment of total human DNA, human male DNA, DNAdegradation and the presence of PCR inhibitors in forensic samples: adiagnostic tool for STR typing, Forensic Sci. Int. Genet. 2 (2008) 108–125.

[8] Quantifiler1 HP and Trio DNA Quantification Kits User Guide, Revision E.Available at: http://tools.thermofisher.com/content/sfs/manuals/4485354.pdf(accessed 03.17.16).

[9] Gina M. Pineda, Anne H. Montgomery, Robyn Thompson, Brooke Indest,Marion Carroll, Sudhir K. Sinha, Development and validation of InnoQuantTM, asensitive human DNA quantitation and degradation assessment method forforensic samples using high copy number mobile elements Alu and SVA,Forensic Sci. Int. Genet. 13 (2014) 224–235.

[10] Pamela M. Holland, Richard D. Abramson, Robert Watson, David H. Gelfand,Detection of specific polymerase chain reaction product by utilizing the 50 ! 30

exonuclease activity of Thermus aquaticus DNA polymerase, Proc. Natl. Acad.Sci. 88 (1991) 7276–7280.

[11] Kathryn Oostdik, Kristy Lenz, Jeffrey Nye, Kristin Schelling, Donald Yet, ScottBruski, Joshua Strong, Clint Buchanan, Joel Sutton, Jessica Linner, NicoleFrazier, Hays Young, Learden Matthies, Amber Sage, Jeff Hahn, Regina Wells,Natasha Williams, Monica Price, Jody Koehler, Melisa Staples, Katie L. Swango,Carolyn Hill, Karen Oyerly, Wendy Duke, Lesley Katzilierakis, Martin G.Ensenberger, Jeanne M. Bourdeau, Cynthia J. Sprecher, Benjamin Krenke,Douglas R. Storts, Developmental validation of the PowerPlex1 Fusion Systemfor analysis of casework and reference samples: a 24-locus multiplex for newdatabase standards, Forensic Sci. Int. Genet. 12 (2014) 69–76.

[12] Jonelle M. Thompson, Margaret M. Ewing, William E. Frank, Jill J. Pogemiller,Craig A. Nolde, D. Jody Koehler, Alyssandra M. Shaffer, Dawn R. Rabbach,Patricia M. Fulmer, Cynthia J. Sprecher, Douglas R. Storts, Developmentalvalidation of the PowerPlex1 Y23 System: a single multiplex Y-STR analysissystem for casework and database samples, Forensic Sci. Int. Genet. 7 (2013)240–250.

[13] Peter H. Sudmant, J.O. Kitzman, F. Antonacci, C. Alkan, M. Malig, A. Tsalenko, N.Sampas, L. Bruhn, J. Shendure, 1000 Genomes Project, E. E. Eichler, Diversity ofhuman copy number variation and multicopy genes, Science 330 (2010) 641–646.

[14] Stephen A. Bustin, Vladimir Benes, Jeremy A. Garson, Jan Hellemans, JimHuggett, Mikael Kubista, Reinhold Mueller, Tania Nolan, Michael W. Pfaffl,Gregory L Shipley, Jo Vandesompele, Carl T. Wittwer, The MIQE Guidelines:minimum information for publication of quantitative real-time PCRexperiments, Clin. Chem. 55 (2009) 611–622.

[15] Scientific Working Group on DNA Analysis Methods: Validation Guidelines forDNA Analysis Methods (SWGDAM). Available at: http://media.wix.com/ugd/4344b0_cbc27d16dcb64fd88cb36ab2a2a25e4c.pdf (accessed 11.05.15).

[16] PowerQuantTM System Technical Manual, 3/15. Available at http://www.promega.com/resources/protocols/technical-manuals/101/powerquant-system-protocol/ (accessed 05.21.15).

[17] SwabSolutionTM Kit Technical Manual, Revised 1/15. Available at https://www.promega.com/resources/protocols/technical-manuals/101/swabsolution-kit-protocol/ (accessed 05.22.15).

[18] Joseph Sambrook, David W. Russell, Molecular Cloning: A Laboratory Manual,Cold Spring Harbor Laboratory Press, Cold Spring Harbor, NY, 2001.

[19] DNA IQTM Casework Pro Kit for Maxwell1 16 Technical Manual, Revised 1/14.Available at https://www.promega.com/resources/protocols/technical-manuals/101/dna-iq-casework-pro-kit-for-maxwell-16-protocol/ (accessed05.21.15).

[20] S.A. Miller, D.D. Dykes, H.F. Polesky, A simple salting out procedure forextracting DNA from human nucleated cells, Nucleic Acids Res. 16 (1988) 1215.

[21] PowerPlex1 Fusion System Technical Manual, Revised 3/15. Available athttp://www.promega.com/resources/protocols/technical-manuals/101/powerplex-fusion-system-protocol/ (accessed 05.21.15).

[22] National Institute of Standards & Technology Certificate of Analysis StandardReference Material1 2372 Human DNA Quantitation Standard, Issued 01/13.Available at https://www-s.nist.gov/srmors/certificates/2372.pdf?CFID=13655200&CFTOKEN=3b92984e29ba83b7-16188008-C0AA-4399-73AA146AB89D9CC4(accessed 05.28.15).

[23] Catherine Theisen Comey, Barbara W. Koons, Kathryn W. Presley, Jill B.Smerick, Carl A. Sobieralski, Donna M. Stanley, Samuel Baechtel, DNAextraction strategies for amplified fragment length polymorphism analysis, J.Forensic Sci. 39 (1994) 1254–1269.

[24] Christopher Cowan, The DNA IQTM System on the Tecan Freedom EVO1 100,Profiles DNA 9 (2006) 8–10.

[25] PowerPlex1 Y23 System Technical Manual, Revised 3/15. Available at http://www.promega.com/resources/protocols/technical-manuals/101/powerplex-y23-system-protocol/ (accessed 02.02.16).

[26] Heather E. LaSalle, George Duncan, Bruce McCord, An analysis of single andmulti-copy methods for DNA quantitation by real-time polymerase chainreaction, Forensic Sci. Int. Genet. 5 (2011) 185–193.