1450-99030901007j

Upload: christopher-robles

Post on 03-Apr-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 1450-99030901007J

    1/6

    JOURNAL OF AUTOMATIC CONTROL,UNIVERSITY OF BELGRADE,VOL.19:7-12,2009

    Abstract We present the design, simulation and test

    results of a new AC amplifier for electrophysiologicalmeasurements based on a three op-amp instrumentation

    amplifier (IA). The design target was to increase the common

    mode rejection ratio (CMRR), thereby improving the quality

    of the recorded physiological signals in a noisy environment.

    The new amplifier actively suppresses the DC component of

    the differential signal and actively reduces the common mode

    signal in the first stage of the IA. These functions increase the

    dynamic range of the amplifiers first stage of the differential

    signal. The next step was the realization of the amplifier in a

    single chip technology. The design and tests of the new AC

    amplifier with a differential gain of 79.2 dB, a CMRR of 130

    dB at 50 Hz, a high-pass cutoff frequency at 0.01 Hz and

    common mode reduction in the first stage of the 49.8 dB are

    presented in this paper.

    Index Terms AC-amplifier; electrophysiological signals;CMRR; DC offset suppression

    I. INTRODUCTION

    HE non-invasive recordings of electrophysiological

    signals are of great importance for assessing the

    function of peripheral nerves, muscles, cortical activity and

    the heart. The signals of interest are very small when

    compared to the noise (e.g., power line interference), the

    base line drift and instability, different artifacts and

    electrode impedance variability [1], [2], [3] and [4]. Thestandard DC three op-amp instrumentation amplifier (IA)

    circuit is one of the most popular basic configurations used

    for electrophysiological amplifiers. This topology provides

    high input impedance and a high common mode rejection

    ratio (CMRR). High CMRR can be obtained by high gain

    at the first stage of the IA and well-matched resistors in the

    second stage of the IA [5]. The gain at the first stage of the

    DC IA needs to be limited to prevent saturation caused by

    offset potentials of electrodes.

    The use of an AC amplifier instead of the DC IA

    eliminates the offset. An AC amplifier based on well-

    known modifications to the classic three op-amp IAtopology is achievable by various techniques [6], [7], [8],

    [9] and [10]. However, the techniques suggested reduce

    the performance of the DC IA.

    N. Jorgovanovi is with University of Novi Sad, Faculty ofEngineering, Novi Sad, Serbia, Fax: +381 21 458873, Phone: +381 21

    4852452, e-mail: [email protected]

    D. Bojani is with University of Novi Sad, Faculty of Engineering,Novi Sad, Serbia Fax: +381 21 458873, Phone: +381 21 4852446, e-

    mail:[email protected]

    V. Ili is with University of Novi Sad, Faculty of Engineering, NoviSad, Serbia, Fax: +381 21 458873 , Phone: +381 21 4852446, e-mail:

    [email protected]

    D. Stanii is with University of Novi Sad, Faculty of Engineering,Novi Sad, Serbia, Fax: +381 21 458873, Phone: +381 21 4852452 , e-

    mail:[email protected]

    DOI: 10.2298/JAC0901007J

    Most new electrophysiological amplifier designs use

    active DC suppression. This technique relates tosubtraction of the integrated IA output (amplified DC

    component of the input signal) from the original input

    signal (sum of the signal of interest and undesirable DC

    component). Subtraction of these two signals should not

    degrade the amplifiers balanced structure or its CMRR.

    The optimal solution to using this concept needs to

    addresses how and when to perform the subtraction. Some

    solutions for using the subtraction method are suggested

    elsewhere, [11], [12], [13], [14], [15] and [16], indicating

    excellent DC component suppression.

    A rarely discussed factor that limits the first stage gain

    of the IA and global CMRR is the dynamic range occupied

    by the common mode signal. An available dynamic rangeis extremely important for battery-powered devices with

    low voltage power supplies. The useful dynamic range of

    the first stage could be increased by additional attenuation

    of the common mode signal.

    We suggest a new subtraction method for the input and

    the integrated output signals. Subtraction is performed in

    the first stage of the IA to preserve high input impedance

    and to ensure high first stage gain and high CMRR. This

    new technique reduces the common mode signal in the first

    stage of the amplifier to an insignificant level, thereby

    practically providing full dynamic range of the amplifiers

    first stage. Moreover, common mode reduction isperformed locally in the first stage of the amplifier without

    involvement of the third electrode. Therefore, the new

    design can be applied with only two electrodes in the

    configuration [17].

    II. CIRCUIT DESIGN AND SIMULATION OF

    ELECTROPHYSIOLOGICAL AMPLIFIER

    A. The Novelty of the New AC AmplifierActive suppression of the common mode signal and the

    DC component of the differential signal in the first stage of

    the three op-amp DC IA are obtained by two feedback

    loops, shown in Figure 1.The first feedback loop returns the IA output signal via

    the integrator K1 and controlled voltage sources V1 and V2.

    The transfer function for the differential signal is given by

    ( )0

    1 0

    ( )( )

    ( ) 2

    o DD

    d D

    V s A sA s

    V s s K A= =

    +, (1)

    where AD0 is the band pass gain of the designed AC

    amplifier,

    2 40

    1 3

    21D

    R RA

    R R

    = +

    , (2)

    and a lower cut-off frequency is given by

    10c D

    Kf A

    = . (3)

    An Improved AC-amplifier for Electrophysiology

    Nikola Jorgovanovi, Dubravka Bojani, Vojin Ili and Darko Stanii

    T

    mailto:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]
  • 7/28/2019 1450-99030901007J

    2/6

    8 JORGOVANOVIN, ET AL.AN IMPROVED AC-AMPLIFIER FORELECTROPHYSIOLOGY

    The second feedback loop returns the extracted common

    mode signal from the output of the first stage of the IA VC1

    via amplifier K2 and the controlled voltage source V3. If

    the value of the voltage V3 is given by

    13

    2

    12

    cm

    RV V

    R

    = +

    , (4)

    then the common mode voltage signal is completely

    removed from the outputs of the first stage. In the

    proposed circuit voltage source, V3 is controlled by the

    second feedback loop (with amplifier K2). Therefore, the

    expression for common mode suppression in the first stage

    is

    1 1 22 1

    1 2 2 2

    2 1

    2

    c

    cm

    V R Rfor R R

    V R K R K

    += >>

    +. (5)

    Figure 1: Block diagram of the designed IA

    As shown, the voltage sources V1 and V2 in the IA input

    loop do not need to be "truly" floating. Instead of truefloating sources, it is satisfactory to provide two controlled

    voltage sources referenced to the amplifiers with voltages

    V3+V1 and V3-V2; this can be accomplished by using

    ordinary op-amps, which will be described later.

    B. Circuit Design of the AC AmplifierFor testing purposes, an AC amplifier, shown in Figure

    2, was designed. The characteristics of the designed

    amplifier are the following: 1) a differential gain of 79.2

    dB and 2) a low cut-off frequency (0.01 Hz) and common

    mode attenuation at the first stage of the IA (49.8 dB at 50

    Hz). To achieve a CMRR that is as high as possible, a gain

    of 59.2 dB is concentrated at the first stage. Therefore, thegain of the first stage and the gain-bandwidth product of

    the op-amps define the upper cut-off frequency. The

    design was based on and tested with a frequently used op-

    amp (OP27) so that an upper cut-off frequency of 8.8 kHz

    is achieved. The higher cut-off frequency, if required,

    could be achieved with higher gain-bandwidth op-amps.

    A standard three op-amp IA is based on op-amps U1, U2

    and U3 and resistors R1, R2, R3, R4, R7, R8, R9, R10 and R19.

    Feedback signals, linear combinations of the voltages

    V3+V1 and V3-V2, are defined by two adders: U6 (R11, R13

    and R15)and U7 (R12, R14 and R16). The first feedback loop

    returns output voltages from the IA via two branches: 1)

    attenuator R20-R21; integrator U4-R22-C1;and adder U7 (R12,

    R14 and R16) (voltage V3+V1) and 2) attenuator R20-R21;

    integrator U4-R22-C1; inverter U5 (R23 and R24); and adder

    U6 (R11, R13 and R15) (voltage V3-V2). The second

    feedback loop returns the common mode signal (Vc1) from

    the output of the first stage of the IA, formed at the middle

    point of the series connection of the two equal resistors R5

    and R6. The Vc1 signal is buffered by op-amp U8 and

    amplified by an inverting amplifier based on op-amp U9,

    resistors R25 and R26 and compensation capacitor C2(voltage -V3). Gain of the inverting amplifier is set to 60

    dB at DC. Due to frequency compensation, the low pass

    cut-off frequency of the inverting amplifier is 15.9 Hz.

    Therefore, amplification of the inverting amplifier is

    reduced to 49.8 dB at 50 Hz. The output voltages of

    amplifiers U6 and U7 are fed into the IA input loop via the

    resistive network R17 - R18 - R19. This ensures attenuation

    of the voltage difference V1-V2 by 53.5 dB and no

    attenuation of voltage V3. The resistive network R17 - R18 -

    R19 affects the low cutoff frequency through the constant

    K1 from expression (3). The constant K1 also depends on

    the integrators constant and the voltage divider R20-R21(6):

    19211

    20 21 22 1 17 18 19

    1 RRKR R R C R R R

    =+ + +

    . (6)

    For the values of the components shown in Figure 2, the

    constant K1 is equal to 6.4910-6

    , which ensures a low

    cutoff frequency of 0.01 Hz. The maximum DC signal

    value that can be suppressed is limited by the maximum

    output voltage of the integratorVimax (defined by the power

    supply voltage and op-amp output characteristics) and the

    attenuation of the resistive network R17 - R18 - R19.

    Additionally, Vimaxis transformed into a voltage difference

    between the outputs of the two adders, attenuated through

    the resistive networks and subtracted from the input

    differential voltage. Maximum DC suppression can bedescribed as

    19( ) max

    17 18 19

    in DC i

    RV V

    R R R=

    + +. (7)

    For the components shown in Figure 2, with a power

    supply of 12 V, DC component suppression is 25 mV,

    which is sufficient for the recordings. The suppression can

    be increased with lower attenuation in the resistive

    network R17 - R18 - R19.

    As we have already stated, the resistive network R17 - R18 -

    R19 ensures attenuation of the differential signal, but should

    have no influence on the common mode. Expression (8)

    shows a relationship between the voltage on resistor R19(V19) and the common mode voltage Vc1:

    11 1219 2 1

    15 16

    ( )DS cR R

    V K K V R R

    = , (8)

    whereK2 is the amplification of the common mode voltage

    Vc1 by amplifier U9,

    252

    26 2 25

    1

    1

    RK

    R sC R=

    +. (9)

    KDS is the attenuation of the resistive network R17 - R18 -

    R19:

    19

    17 18 19

    DS

    RK

    R R R

    =+ +

    . (10)

    If the ratio R11/R15 is equal to R12/R16,the voltage on R19is zero and the common mode signal will not be

  • 7/28/2019 1450-99030901007J

    3/6

  • 7/28/2019 1450-99030901007J

    4/6

    10 JORGOVANOVIN, ET AL.AN IMPROVED AC-AMPLIFIER FORELECTROPHYSIOLOGY

    III. TESTING OF THE ACAMPLIFIER

    To test the new topology with two feedback loops, we

    built a prototype without optimizing the power

    consumption, the size of the prototypes PCB area and the

    circuit complexity. The reasons for skipping optimization

    are the plans for integration of the amplifier into the single

    chip.

    Two types of prototype verification were performed: 1)verification of the amplifiers characteristics based on test

    signals and 2) clinical verification by EMG recording in a

    noisy environment. Data acquisition included a Tektronix

    TDS3012 digital scope with a Tektronix differential

    voltage probe ADA400A. A test signal was generated by

    an Agilent 33220A signal generator.

    A DC signal suppression test was performed by using the

    sine wave input differential signal with an amplitude of 0.4

    mVp-p, a frequency of 1 kHz and a DC offset of 0.8 mV.

    The common mode signal on the input was set to zero in

    this test. The result of the test is shown in Figure 5. The

    upper waveform (Ch1) shows the differential signal at the

    input of the amplifier. The bottom waveform (Ch2)represents the output of the amplifier with a sine wave

    signal with an amplitude of 4 Vp-p, a frequency of 1 kHz

    and a DC offset equal to zero. Test results confirm

    calculations and simulations and they were similar to the

    other amplifiers with active DC signal suppression.

    Testing of the common mode signal reduction in the

    first stage of the amplifier was performed with a high

    amplitude common mode input signal. We used a sine

    wave test signal with an amplitude of 15.8 Vp-p and a

    frequency of 50 Hz; the differential mode signal at the

    input was set to zero. The results of the CMRR test are

    presented in Figure 6a. The upper waveform (Ch1) shows

    the common mode signal at the input of the amplifier. The

    bottom waveform (Ch2) represents the common mode

    signal at the output of the first stage of the amplifier.

    According to the measured results, the common mode

    signal attenuation in the first stage of the amplifier is 49.8

    dB at 50 Hz. This result is a significant improvement

    compared to the other amplifiers that had no common

    mode reduction in the first stage.

    To confirm the high total CMRR of the amplifier, the

    output signal of the amplifier was recorded and the results

    are shown in Figure 6b. The upper waveform (Ch1) shows

    the common mode signal at the input of the amplifier. The

    bottom waveform (Ch2) represents the common mode

    signal at the output of the amplifier. According to the

    measured results, the CMRR is 130 dB at 50 Hz.

    Figure 5: Verification of the differential gain and DC-offset suppression.

    Ch1 shows the input signal waveform to the amplifier. Ch2 shows the

    output signal waveform.

    The experimental verification of the amplifier on real

    signals is demonstrated by the EMG recording in an

    intentionally made, extremely noisy environment. The

    tested amplifier was located very close to three switch

    mode power supplies and GSM phones, also it was tested

    without EMI filters, shielding, or a driven right leg circuit.

    We recorded EMG signals by using two Neuroline

    electrodes positioned over the Flexor Digitorum m.,

    following the positioning instructions defined in the

    SENIAM project [18]. The neutral (i.e., ground) electrode

    over the Ulnar nerve was connected directly to the

    common electrode of the amplifier. The recordings from

    the contracted muscle of an able-bodied individual are

    presented in Figure 7a with the time scale set to 10 ms per

    division. Figure 7b shows three consecutive contractions

    (weak, medium and strong) from the same muscle. The

    time scale (horizontal axes) is 1 second per division.

    Figure 6: Verification of the common-mode rejection in the first stage of the amplifier (left). Ch1: common-mode signal at the input of the amplifier;

    Ch2: common-mode signal at the output of the first stage of the amplifier; Verification of the total common-mode rejection (right). Ch1: common-

    mode si nal at the in ut of the am lifier; Ch2: common-mode si nal at the out ut of the am lifier

  • 7/28/2019 1450-99030901007J

    5/6

    JOURNAL OF AUTOMATIC CONTROL,UNIVERSITY OF BELGRADE,VOL 19,2009. 11

    The upper waveform (Ch1) in Figure 8 shows the same

    EMG signal with a different time scale setting (i.e., 400 ms

    per division). The bottom waveform represents the

    frequency spectrum of the signal.

    Figure 8: EMG signal of the Flexor Digitorum Muscle and its frequency

    spectrum

    The tests verify that the realized amplifier matches the

    calculations and simulation results.

    IV. CONCLUSION

    The development of a new amplifiers topology that has

    characteristics necessary for electrophysiological

    recordings in a noisy environment was presented in this

    paper. The novel topology is the active suppression of both

    the DC component of the differential signal and the

    common mode signal in the first stage of the three op-amp

    IA. Active suppression of DC-offset was obtained by

    feeding back an integrated form of the amplifier's output

    that is equal to the input DC-offset, however with the

    opposite sign. The active common mode signal

    suppression at the first stage of the IA was obtained by

    feeding back an extracted common mode signal from the

    output of the first stage and presents an originalimprovement in the electrophysiological amplifiers

    design.

    As a result, almost the entire dynamic range of the

    amplifiers first stage is available for a useful component

    of the input signal. The active DC offset suppression

    allows DC coupling between the electrodes and the

    amplifier, providing extremely high input impedance for

    the designed amplifier. The entire design of the novel AC

    amplifier, with simulations and test results, were presented

    in this paper. A high CMRR of 130 dB at 50 Hz, a

    differential gain of 79.2 dB, a cutoff frequency at 0.01 Hz,

    a bandwidth of 8.8 KHz and, in particular, a common

    mode reduction of 49.8 dB in the first stage of the

    amplifier were achieved. Major characteristics of the

    amplifier were confirmed by the test results. With these

    characteristics, this amplifier can be used to record EMG

    and other electrophysiological signals. The describedcircuit is now under clinical testing as part of our virtual

    EMG system [19] and [20].

    REFERENCES

    [1] B. B. Winter and J. G. Webster, "Reduction of interference due to

    common mode voltage in biopotential amplifiers," IEEE Trans

    Biomed Eng, vol. BME-30(1), pp. 58-62, 1983.

    [2] N. V. Thakor and J. G. Webster, "Ground-free ECG recording with

    two electrodes, " IEEE Trans Biomed Eng, vol. BME-27(12), pp.

    699-704, 1980.

    [3] J. C. Hunta and J. G. Webster, 60-Hz interference in

    electrocardiography, IEEE Trans Biomed Eng, vol. BME-20,

    pp.91-101, Mar. 1973.

    [4] A. F. Pacela, Collecting the bodys signals, Electronics, 40, (14),pp. 103-112, 1967.

    [5] R. Pallas-Areny and J. G. Webster,Common Mode Rejectrion

    Ratio in Differenctial Amplifiers, IEEE Trans Biomed Eng, vol.

    BME-40, pp. 669-676, 1991.

    [6] R. Pallas-Areny, J. Colominas and J. Rosell, "An improved buffer

    for bioelectric signals,"IEEE Trans Biomed Eng, vol. BME-36(4),

    pp. 490-493, 1989.

    [7] R. Palls-Areny and J. G. Webster, "Analog Signal Processing,"

    New York: Wiley, 1999.

    [8] M. J. Burke and D. T. Gleeson, "A micropower dry-electrode ECG

    preamplifier,"IEEE Trans Biomed Eng, Vol. BME-47(2), pp. 155-

    162, 2000.

    [9] R. Pallas-Areny and J. G. Webster, " AC Instrumentation Amplifier

    for Bioimpedance Measurements," IEEE Trans Biomed Eng, vol.

    BME-40(8), pp. 830-833, 1993.

    [10] Z. Nikolic, D. B. Popovic, R. B. Stein and Z. Kenwell Z."Instrumentation for ENG and EMG recordings in FES systems,"

    IEEE Trans Biomed Eng, vol. BME-41. pp. 703-706, 1994.

    [11] Yue-Der Lin, Chang-Da Tsai, Hui-Hsun Huang, Dah-Chuan Chiou

    and Chien-Ping Wu, Preamplifier with a Second-Order High-Pass

    Figure 7 Short-time contraction of the Flexor Digitorum Muscle (left). Amplification of the amplifier is 79.2dB. Amplitude of the recorded signal is

    2.2mVp-p; EMG signal of Flexor Digitorum Muscle (right). Amplification of the amplifier is 79.2dB. Amplitude of the low contraction is 0.6mVp-p,

    amplitude of the medium contraction is 1mVp-p and amplitude of the high contraction is 2mVp-p

  • 7/28/2019 1450-99030901007J

    6/6

    12 JORGOVANOVIN, ET AL.AN IMPROVED AC-AMPLIFIER FORELECTROPHYSIOLOGY

    Filtering Characteristic, IEEE Trans Biomed. Eng, vol. BME-46,

    pp. 609-612, 1999.

    [12] H. W. Smit, K. Verton and C. A. Grimbergen, "A low-cost

    multichannel preamplifier for physiological signals," IEEE Trans

    Biomed Eng, vol. BME-34(4), pp. 307-310, 1987.

    [13] E. M. Spinelli and M. A. Mayosky, "AC coupled three op-amp

    biopotential with active DC suppression," IEEE Trans Biomed

    Eng, vol. BME-47(12), pp. 1616-1619, 2000.

    [14] E. M. Spinelli, N. Martinez, M. A. Mayosky and R. Pallas-Areny,

    "A Novel Fully Differential Biopotential Amplifier With DC

    Suppression," IEEE Trans Biomed Eng, vol.51, No.8, pp. 1444-

    1448, August 2004.

    [15] T. Degen and H. Jckel, "A Pseudodifferential Amplifier for

    Bioelectric Events With DC-Offset Compensation Using Two-

    Wired Amplifying Electrodes," IEEE Trans Biomed Eng, vol.53,

    No.2, pp. 300-310, February 2006.

    [16] B.Gosselin, M. Sawan and C.A. Chapman, "A Low-Power

    Integrated Bioamplifier With Active Low-Frequency Suppression,"

    IEEE Trans Biomedical Circuits and Systems, vol.1, No.3, pp.

    184-192, September 2007.

    [17] E. M. Spinelli and M. A. Mayosky, Two-electrode Biopotential

    Measurements: Power Line Interference Analysis, IEEE Trans

    Biomedical Enginering, vol. 52, No. 8, pp. 14361442, Aug. 2005.[18] H. J. Hermens, B. Fredriks, R. Merlett, D. Stegeman, J. Block, G.

    Rau, et al., SENIAM European Recommendations for Surface

    electromyography, Roessingh Research and Development BV,

    1999.

    [19]N. Stojanova, M. Petrovic and D. B. Popovic, "New approach to

    laboratory measurements virtual instruments," J Applied

    Measurem, vol. 12, pp. 17-24, 1999.

    [20] M. Piperski, D. B. Popovi, Virtual instrument for measuringevoked potentials of peripheral nerves: electroneurography and

    EMG reflexology, in Proc. 8th Information Technology Conf,

    abljak, February 22-26 2004, pp. 21-24.