LECO’s Citius High Res LC/MS:
Faster results fewer runs
Dave SchlegelLECO Separation Science
2
LECO Citius
• Today’s researchers are under pressure to provide accurate results to complex problems in less time than ever before. To meet these challenges, many researchers rely on high resolution mass spec systems to obtain qualitative results and triple quads to obtain quantitative results.
• However, there's a more efficient way to reach your goals--through the use of high performance time-of-flight mass spectrometry. This technology allows you to collect qualitatively and quantitatively useful data not just on the same instrument, but also at the same time. The different qualitative and quantitative functions are accomplished post hoc with data analysis software, not by performing serial experiments or by compromising duty cycle in a single run.
• The Citius provides researchers with a single instrument that can produce excellent results in a single run.
3
Mass Spectrometry with Comprehensive CID (MSc2)
CHARACTERISTICS• Same high resolution and mass
accuracy in precursor and product ion spectra
– 1 ppm; 50k or 100k– Accurate Isotope Information
• Available ALL of the time• High acquisition speed
– Up to 100 product and 100 precursor spectra/sec
• Signal Deconvolution• Independent acquisition
channels for precursor and product ions
• MRM-like selectivity amplification
Ionization
+++
+
++++++++
++
++
Transport
+++
+Separat
eDetect
Ionization
+++
+Fragmen
tSeparat
eDetect
• High Speed Switching(100 s/s precursor/product)
• Controlled pressure• Refocused during
fragmentation and before MS
High ResolutionHigh Mass Accuracy
High Integrity Isotopes
Precursor
Product
Citius™ LC-HRT
4
Mass Analyzer Design
Citius™ LC-HRT
5
Citius™ Ion Interface
Sample Nozzle
Multipole Ion Guide #1
SkimmerMultipole Ion Guide #2
Ion Optics Stack
Orthogonal Accelerator
To mass analyzerCID
region
6
MSc2 in Drug Metabolism - DMPK
50 55 60 65 700
2000
4000
6000
8000
10000
12000
Time (s)
243.149190±0.000486 (Parent Ions)257.164840±0.000514 (Parent Ions)271.180490±0.000542 (Parent Ions)287.175404±0.000577 (Parent Ions)182.096426±0.000546 (Fragment Ions)
DIdesmethyldoxylamineDesmethyldoxylamine
DoxylamineDoxylamine + O
Common Fragment;loss of (substituted) ethanolamine
Assignment Expected m/z Observed m/zRelative m/z Error
/ ppm
Didesmethyldoxylamine 243.149190 243.14925 0.2
Desmethyldoxylamine 257.164840 257.16497 0.5
Doxylamine 271.180490 271.18049 0.0
Doxylamine + O 287.175404 287.17543 -0.1
100 150 200 250 300 350 400 450 5000
2500
5000
7500
10000
12500
15000
17500
20000
Inte
nsity
(cou
nts)
M/Z
271.18049
Peak True - sample"3001KS2011w07-025 AM Sample", Peak 18, at 57.3423 s
100 150 200 250 300 350 400 450 5000
2500
5000
7500
10000
12500
15000
17500
20000
Inte
nsity
(cou
nts)
M/Z
182.09672
271.
1810
5
90.0
9145
Peak True(Cid) - sample"3001KS2011w07-025 AM Sample", Peak 18, at 57.3423 s
Parent Ion Channel
Product Ion Channel
Product Ion Channel
AssignmentObserved
m/zR (FWHM)
m/z Error /ppm
Relative Isotopic
Accuracy
[C4H12NO]+ 90.09145 35232 1.2
[C13H12NO]+ 182.09672 45503 1.6
M+1 183.10023 43879 2.1%
M+2 184.10332 42877 5.8%
[C17H23N2O]+ 271.18105 48606 2.1
M+1 272.18446 49421 -4.6%
M+2 273.18732 49147 29.2%
Parent Ion Channel
AssignmentObserved
m/zR (FWHM)
m/z Error /ppm
Relative Isotopic
Accuracy
[C17H23N2O]+ 271.18049 49409 0.0
M+1 272.18398 49044 -4.5%
M+2 273.18671 41602 -3.8%
7
Mass Accuracy in Fragment Ion Analysis
Target m/z Observed m/zPrecursor or
Fragment 1 2 3 4 5 Avg. Error (ppm)
124.9998 F 124.99976 124.99972 124.99971 124.99984 124.99971 -0.416
104.107 F 104.10691 104.107 104.10701 104.10694 104.10697 -0.327
86.09643 F 86.09633 86.09635 86.09636 86.09637 86.09637 -0.860
60.08078 F 60.08073 60.08067 60.08067 60.08074 60.08075 -1.132
184.07332 F 184.07306 184.07311 184.07313 184.07316 184.07319 -1.032
782.56943 P (PC 36:4) 782.56826 782.56875 782.56861 782.56868 782.56789 -1.268
806.56943 P (PC 38:5) 806.56837 806.56835 806.56728 806.56652 806.56643 -2.529