19072014 printer presentation agfa .pdf

20
Direct Digital Imaging - DDI (Drystar®) Information about Drystar® Agfa HealthCare India Private Limited, 2 nd Floor, B – 14, Road No. 1, Wagle Industrial Estate, Thane (West) – 400 604

Upload: swapnil-solaskar

Post on 08-Nov-2015

33 views

Category:

Documents


10 download

TRANSCRIPT

  • Direct Digital Imaging - DDI(Drystar)

    Information about DrystarAgfa HealthCare India Private Limited,

    2ndFloor, B 14, Road No. 1,Wagle Industrial Estate,Thane (West) 400 604

  • 2 Company Confidential

    FFDM, CR, DR, CT, MR,DSA, digital R&F508 dpi (50 m px)DRYSTAR DT2 filmDmax > 3.2

    With mammography printing option DRYSTAR DT2 Mammo film, Dmax > 3.6

    DRYSTAR AXYS

    Agfa Hardcopy Imagers & Film Assortment

    De-centralized Printing

    Centralized Printing

    DRYSTAR 5302

    CR, DR, DSA, digital R&F320 dpi (79 m px)DRYSTAR DT2 filmDmax > 3.1

    DRYSTAR 5300

    CT, MR, DSA, digital R&F, CR, DR,320 dpi (79 m px)DRYSTAR DT2 filmDmax > 3.1

    DRYSTAR 5503

    FFDM, CT-MS, CT-SS, MR, CR, DR, US, NM508 dpi (50 m px)DRYSTAR DT2 filmDmax > 3.1

    With mammography printing option DRYSTAR DT2 Mammo film, Dmax > 3.7

    Software enhancement to bring DRYSTAR technology on wet laser sharpness level (Scopix LR 5200).

    Over 40,000 Medical

    Imagers installed

    worldwide

  • 3 Company Confidential

    Basic Principle of DDI:

    - Single Stage Process

    - Black / White Densities created by TPH

    - Film is sensitive to heat only

    Single Step Printing

  • 4 Company Confidential

    Thermal Print Head (TPH)

    Thermal Sensitive Film

    Drum to transport the film

    Direct Digital Imaging (DDI)

    DDI Technology - Direct Thermal Printing

    Resistors

    Thermal Print Head (TPH)

  • 5 Company Confidential

    A#Sharp Imaging Technology

    The main cause for the lower sharpness is heat spread and storage in the thermal print head TPH.

    Ideal TPH

    Requested image

    Real TPH

    Printed image

  • 6 Company Confidential

    Features of Dry Imager: Direct Digital Imaging1. DDI films can be loaded in Day Light, as they are sensitive to heat only

    2. DDI is a much newer technology

    3. DDI imagers require very less warm up time (5 minutes or less)

    4. DDI imagers consume amazingly low power, (to the tune of 700 W Max.)

    5. Footprint of Agfa DDI imagers, Drystar Series, is small, saving space

    6. DDI imagers produce a square type pixel, as Image on the film

    7. Complete range of Drystar to fulfill the need of radiology & imaging

    8. Ecologically friendly imager, without use of any harmful chemicals

    9. Installed base of over 2,350 imagers for Agfa alone in India

  • 7 Company Confidential

    Photo-Thermal Technology (Competition)

    Technically complex and delicate process

    2 STEPS

    1 Light exposure

    2 Heat processing laser exposure

    thermal processing on drum2

    1

    Light sensitive film

  • 8 Company Confidential

    Difference between DDI and Dry Laser:DDI

    Single step printingEco-friendly filmsFreehand daylight loadingLess Warm up timeLess Components / ServiceLess power consumptionHigh shelf life (pre & post print)Compact to occupy less spaceDirect DICOM ready

    Dry LaserTwo step printingFilms contain developing chemicals

    Cartridge daylight loadingHigher Warm up time (> 15 mins)More Components / ServiceHigher power consumptionEqual shelf life (pre & post print)Occupy larger spaceRequire interface for DICOM

  • 9 Company Confidential

    No Imager is pure LASER:

    Any Laser Exposure Thermal Development (LETD) Imager, popularly called as Dry Laser Imager, as the

    name suggests involves Thermal process within the two staged printing process

    :Stage I:Film (light sensitive) which contains embedded chemicals is exposed by a Laser beam. Embedded chemicals act to

    actually create the Gray shades:Stage II:

    Non reacted part of the emulsion (deemed white part) is removed by a thermal heater. Frequent replacements are required for the filters to keep this fixing process smooth.

  • 10 Company Confidential

    Perception of Resolution:DDI

    - Perceived resolution is better, even at the same addressability

    Dry Laser- Perceived resolution is less, even at the same addressability

    Conversion for commonly used interchangeable technical terms

  • 11 Company Confidential

    Significance of correct Resolution:DDI

    - Modality CR Mammo(Input Resolution of 10 lp/mm)

    - Imager(Output Resolution of 10 lp/mm)

    - True Size Printing

    Dry Laser- Modality CR Mammo

    (Input Resolution of 10 lp/mm)- Imager

    (Output Resolution of 12.5 lp/mm)- True Size Printing

    ?

    Higher resolution of Imager is not always beneficial

  • 12 Company Confidential

    Other technical parameters:Access Time: Better

    DDI has much lesser access time contributing to meaningful throughput. One prints many different films on printer; rarely the same film many times.

    Gray Scale Resolution: =

    Throughput: ~More or less is very close to each other in a practical environment

    Number of Input trays: =

    Number of supported film sizes: BetterUse of 10 X 12 or 11 X 14 allows maximum utilization of space for images from the modality, typically with the aspect ratios used in medical diagnosis

    Maximum Density: Slightly Lesser

    Black Border: Not Available

  • 13 Company Confidential

    Imager Model: Drystar 5300

    Ideal for CT / MRI Printing

    Mounting Table TopOn-Line Tray 1

    Foot Print 27.5 X 31. 5 InchesWeight 55 Kg

    Resolution 320 Pixels / "GrayScale 12 Bits / PixelThroughput 70 / Hour

    Sizes14 X 17 "14 X 14 "11 X 14 "

    Applications CTMRI

  • 14 Company Confidential

    Imager Model: Drystar 5302

    Ideal Centralized Printing

    Mounting Table TopOn-Line Tray 2

    Foot Print 28.7 X 28.1 InchesWeight 90 Kg

    Resolution 320 Pixels / "GrayScale 12 Bits / PixelThroughput 75 / Hour

    Sizes

    14 X 17 "14 X 14 "11 X 14 "10 X 12 "8 X 10 "

    Applications

    CTMRICRUSNM

  • 15 Company Confidential

    Imager Model: Drystar AxysMounting Table Top

    On-Line Tray 2Foot Print 28.7 X 28.1 Inches

    Weight 90 KgResolution 508 Pixels / "GrayScale 14 Bits / PixelThroughput 75 / Hour

    Sizes

    14 X 17 "14 X 14 "11 X 14 "10 X 12 "8 X 10 "

    Applications

    CTMRICRUSNM

    MammoMammography ready Imager

  • 16 Company Confidential

    Imager Model: Drystar 5503Mounting Floor

    On-Line Tray 3Foot Print 28.3 X 28.1 Inches

    Weight 193 KgResolution 508 Pixels / "Throughput 100 / Hour

    Sizes

    14 X 17 "14 X 14 "11 X 14 "10 X 12 "8 X 10 "

    Applications

    CTMRICRUSNM

    Mammo

    Centralized Imager for all Applications

  • 17 Confidential

    DRYSTAR DT1 Film

    Media for the DRYSTAR 2000, 3000, 4500 and 4500 M :

    DT1B (blue base) DT1C (clear base) in media sizes

    8 x 10 (20 x 25 cm), 10 x 12 (25 x 30 cm), 11 x 14 (28 x 35 cm), 14 x 17 (35 x 43 cm),

    First generation Direct Digital grayscale medium

    Max. optical density >3.0 Secure archiving up to 20 years (when

    stored under proper temperature conditions)

    DYSTAR 2000

    DRYSTAR 3000

    DRYSTAR 4500DRYSTAR 4500 M

  • 18 Confidential

    DRYSTAR DT2 Film

    Media dedicated for DRYSTAR 5503, 5500, AXYS, 5302 and 5300

    DT2B (blue base) and DT2C (Clear base) in all media sizes

    8 x 10 (20 x 25 cm), 10 x 12 (25 x 30 cm), 11 x 14 (28 x 35 cm), and 14 x 17 (35 x 43 cm)

    Max. optical density >3.1 Designed to

    withstand high printing power resist high temperature peaks to have good image hue even at higher printing

    speed. Secure archiving up to 20 years (when stored under

    proper temperature conditions)

    DYSTAR 5500/5503

    DYSTAR AXYS

    DYSTAR 5302

    DYSTAR 5300

  • 19 Confidential

    Media dedicated for DRYSTAR 5500, 5503 and AXYS imagers with mammography printing option Blue base Media sizes

    8 x 10 (20 x 25 cm) 10 x 12 (25 x 30 cm) 11 x 14 (28 x 35 cm)

    Max. optical density > 3.79 Extended term storage: minimum 20

    years (when stored under proper temperature conditions)

    DRYSTAR DT2 Mammo film

    DYSTAR 5500/5503

    DYSTAR AXYS

  • 20 Confidential

    Thank you

    r r