2-daithermy fni$

Upload: chanlal

Post on 14-Apr-2018

230 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/30/2019 2-DAITHERMY fni$

    1/56

    High Frequency Heat Therapy

    *physical stimulus employed of physiotherapyis in the form of heat, either by simple heatradiation or by the application of high frequencyenergy obtained from special generators.

    *The use of high frequency energy inthermotherapy has the advantage ofconsiderable penetration as compared with

    simple heat application.

  • 7/30/2019 2-DAITHERMY fni$

    2/56

    High Frequency Heat Therapy

    *Thus with high frequency energy ,dipper

    lying e.g., muscles, bones, internal organs,ect.Can provide heat.

    *High frequency heat therapy is based on thebased on the fact that the dipole molecules ofthe body are normally placed randomly under theinfluence an electrical field

  • 7/30/2019 2-DAITHERMY fni$

    3/56

    *they rotate according to the polarityof their charge

    *in the direction of the field theyrotate according to the polarity of theircharge in the direction of the field lines

  • 7/30/2019 2-DAITHERMY fni$

    4/56

    *The positively charge end of thedipole then orients itself to the minus

    pole and the negatively end to the pluspole.*Since the polarity of the electric

    field alternates, a micro-heatingeffect results from the continuous re-alignment of the molecules. Highfrequency energy for heating isobtained by various ways.

  • 7/30/2019 2-DAITHERMY fni$

    5/56

    *It may be from the short-wave therapyunit making use of either the condenser fieldor the inductor field method..*Microwaves and ultrasound waves are alsoused for heating purposes in special cases.

  • 7/30/2019 2-DAITHERMY fni$

    6/56

  • 7/30/2019 2-DAITHERMY fni$

    7/56

    WHAT IS ELECTROSURGICAL DIATHERMY

    ?

    *Electrosurgery defined as the use of radiofrequency electric curent to sever tissue or

    achieve hemostasis.*A high frequency is used because a lowfrequency (below 100,000 Hz) will stimulatenerves or in other words can electrocute thepatient .

  • 7/30/2019 2-DAITHERMY fni$

    8/56

    * Medical diathermy is similar toElectrosurgery in that a radio frequency

    current is passed through the patient body.Entry and exit electrodes are chosen sothat tissue heating is never high enough tocause necrosis.

    *Typically electrosurgical generators canoperate at frequencies up to 4 MHz, atthese frequencies reactive phenomenon

    (capacitance and inductance) become quiteprominent. Therefore some famous modelslike VALLEY LAB SSE2K

    operate at 500 kHz.

  • 7/30/2019 2-DAITHERMY fni$

    9/56

  • 7/30/2019 2-DAITHERMY fni$

    10/56

  • 7/30/2019 2-DAITHERMY fni$

    11/56

    There are three surgical effects which can beachieved by Electrosurgery :

    *ELECTROSURGICAL DESSICATION

    (low power coagulation with out sparking).

    * ELECTROSURGICAL CUTTING

    (electric sparking to tissues with cutting effect).

    * ELECTROSURGICAL FULGURATION

    (electric sparking to tissues without cuttingeffect).

    SURGICAL EFFECTS OF DIATHERMY:-

  • 7/30/2019 2-DAITHERMY fni$

    12/56

    CONTROLPANEL

    FOOT

    SWITCHPOWER

    SUPPLY

    POWER

    OUTPUT BOARD

    AUDIO TONE

    GENERATOR

    LOGIC

    BOARD

    ISOLATIO

    N SWITCHBOARD

    BLOCK DIAGRAM OF SURGICAL DIATHERMY

  • 7/30/2019 2-DAITHERMY fni$

    13/56

    ELECTROSURGICAL DESICCATION:-

    *In desiccation electric current is passedthrough the electrical resistance of tissueand heat arises in the tissue.

    *when the tissue becomes hot water is drivenout of the tissue Visually during desiccationtissue becomes brown then it steams as thewater is driven out.*SSE 2K can perform desiccation with

    either BLEND , CUT or COAG.

  • 7/30/2019 2-DAITHERMY fni$

    14/56

    ELECTROSURGICAL CUTTING:-

    *In electrosurgical cutting objective is to heat thetissue so rapidly that cell explodes into steam,leaving a cavity in the cell matrix .

    *The heat is dissipated in the steam and ,

    therefore it does not conduct through the tissue todry out adjacent cells.

    *When the electrode is moved and fresh tissue iscontacted , new cells are exploded and the incision

    is made .*When the radio frequency current jumps across

    the air gap to tissue .

  • 7/30/2019 2-DAITHERMY fni$

    15/56

    *The bright light in the air gap is

    technically calledspark

    .*An ARC is a similar phenomenon which

    requires longer intervals to becomeestablished .

    Electrosurgical cutting involves sparkingto tissue .*A hot cattery wire ( a heated wirewith no electric current ) can also cut

    the tissue by the mechanismexplained above)

  • 7/30/2019 2-DAITHERMY fni$

    16/56

    *The essential characteristic of CUT is that

    they are continuos sine waves . That is if thevoltage is plotted over time a pure CUT waveform is a continuos sine wave at 500 kHz

  • 7/30/2019 2-DAITHERMY fni$

    17/56

    ELECTROSURGICAL FULGURATION:-

    *COAG wave form generally consists of shortbursts of radio frequency sine wave.*The frequency is 450 kHz and the bursts occur at20,000 times/second .

    *the average power delivered in a COAG wave is lessbecause it is turned off most of the time .

    *A good COAG wave form without significant cuttingeffect because the heat is more widely spread by

    the long sparks and because the heating isintermittent.

  • 7/30/2019 2-DAITHERMY fni$

    18/56

    *The temperature of water in the cellsdoes not go high enough to flash into

    steam, in this way cells are dehydratedslowly and are not torn apart to form anincision .*Some times desiccation and fulguration arethought to be same , Fulguration can becontrasted with desiccation in several ways

    ..first, sparking to tissue always producesnecrosis, whereas desiccation may notproduce necrosis.

  • 7/30/2019 2-DAITHERMY fni$

    19/56

    Typical current = 0.5 Amp. Electrode in good contact

    with the tissue.

    Deep coagulation.

    Spreads rapidly. Relatively soft,light brown

    eschar.

  • 7/30/2019 2-DAITHERMY fni$

    20/56

    Typical current 0.1 Amp.

    Electrode free from tissue.

    Long sparks result insuperficial coagulation

    Deeper necrosis asfulguration continues.

    Eschar is hard and black.

  • 7/30/2019 2-DAITHERMY fni$

    21/56

    Typical current 0.1 Amp.

    Electrode is separated fromtissue by a thin layer ofsteam.

    Small intense sparks flashcells into steam.

    Hemostasis can be minimal.

  • 7/30/2019 2-DAITHERMY fni$

    22/56

    *When using monopolar electrodes, a patientreturn is always required because it provides amean for the current to leave the patients bodysafely and return to the generator.

    *The patient return electrode must be

    considerably larger than active electrode so thatthe current density is kept low and nosignificant heating occurs.

    * This return electrode is referred to as patientreturn electrode and not ground plate

    because there is a confusion which might makepeople to assume that it is convenient

    goring for ECGs or other purposes.

    PATIENT RETURN ELECTRODE:-

  • 7/30/2019 2-DAITHERMY fni$

    23/56

  • 7/30/2019 2-DAITHERMY fni$

    24/56

    FOOT SWITCH

  • 7/30/2019 2-DAITHERMY fni$

    25/56

    Short -Wave Diathermy

    *The term of *diathermy* means throughheating or producing deep heating directly inthe tissues of the body.

    *Externally applied sources of the heat likehot towels, infrared lamps and electricheating pads ,often produce discomfort andskin burns long before adequate heat haspenetrate to the deeper tissue.

  • 7/30/2019 2-DAITHERMY fni$

    26/56

    *But with the diathermy technique, thesubjects body becomes a part of the

    electrical circuit and the heat is producedwithin the body and not transferredthrough the skin. , Another advantage ofdiathermy is that treatment can becontrolled precisely.

    *Careful placement of the electrodes

    permits location of the heat to the regionto

  • 7/30/2019 2-DAITHERMY fni$

    27/56

    ,*that has to be treated the amount of heat

    can be closely adjusted by means of circuit

    parameters.*the heating of tissues is carried out byhigh frequency alternating current whichgenerally has a frequency of 27.12KHz and

    the wavelength of 11m.

    *Currents of such high frequencies do not

    stimulate motor or sensory nerves, nordo they produce any muscle contraction.Thus

  • 7/30/2019 2-DAITHERMY fni$

    28/56

    *when a such a current is passed throughthe body, no discomfort is caused to

    the subject.

    *The current being it is possible to pass

    through the tissue currents of a muchgreater intensity to produce directheating in the tissues similar to any otherelectrical conductor.

  • 7/30/2019 2-DAITHERMY fni$

    29/56

    *The method consist in applying the out

    put of a radio (RF) oscillator to a pair

    electrodes which are positioned on the bodyover the region to be treated.

    *The RF energy heats the tissues andpromotes heating of injured tissues andinflammations.

  • 7/30/2019 2-DAITHERMY fni$

    30/56

  • 7/30/2019 2-DAITHERMY fni$

    31/56

    The short wave diathermy machineconsists of two main circuits

    *Oscillating circuit: which produced highfrequency current.

    *Patient circuit: which is connected tothe oscillating circuit and through whichthe electrical energy is transferred tothe patient.

  • 7/30/2019 2-DAITHERMY fni$

    32/56

  • 7/30/2019 2-DAITHERMY fni$

    33/56

    Application technique of short-wavetherapy:

    *The pattern of tissue heating is generallyaffected by the method of short-wave

    diathermy delivery.

    *The two most common forms of applicationinclude the capacitor plate method and the

    inductive method.

  • 7/30/2019 2-DAITHERMY fni$

    34/56

    *In the capacitor plate method, the out

    put of the short-wave diathermy machine is

    connected to metal electrodes which arepositioned on the body over the region to betreated.

    * This electrodes called *PADS* in theterminology of the diathermy.

    Application technique of short wave

  • 7/30/2019 2-DAITHERMY fni$

    35/56

    Application technique of short-wavetherapy:

    *These pads or electrodes do not directlyconnect with skin. Usually layers of towelsare interposed between the metal and the

    surface of the body.*The pads are placed to that portion of thebody to be treated is sandwiched betweenthem. This arrangement is called the

  • 7/30/2019 2-DAITHERMY fni$

    36/56

    *Condenser Method* wherein the metalpads act as two plates while the body

    tissues between the pads as*dielectric* ofthe capacitor

    .

  • 7/30/2019 2-DAITHERMY fni$

    37/56

    *When the radio frequency out put isapplied to the pads, the dielectric

    losses of the capacitor manifestthemselves as heat in the interveningtissues.

    *The dielectric losses may be due tovibration of ions and rotation of manifestin the tissue fluids (electrolytes) andmolecular distortion in tissues which are

    virtually insulators like fat

  • 7/30/2019 2-DAITHERMY fni$

    38/56

    *Alternatively, the output of thediathermy machine may be connected to

    a flexible cable instead of pads. Thiscable is coiled around the arm or knee orany other portion of the of the bodywhere plate electrodes are inconvenient

    to use

  • 7/30/2019 2-DAITHERMY fni$

    39/56

    *When RF current is passed throughsuch a cable ,an electro static field isset up between its ends and a magneticfield around its centre.

    *Deep heating in the tissue results from

    electrostatic action whereas the heatingof the superficial tissues is obtained byeddy currents set by a magnetic effect.

    This technique is inductothermy.

  • 7/30/2019 2-DAITHERMY fni$

    40/56

    *Although most of the short-wave

    diathermy machines have an out putpowerControl, yet there is no indication of theamount of converted and absorbed heat

    within the body tissues.*Therefore ,the intensity of treatment isdependent on the sensation of warmthfelt by the patient.

  • 7/30/2019 2-DAITHERMY fni$

    41/56

    Microwave diathermy

    *This type consist of in irradiating the

    tissues of the patients body with veryshort wireless waves having frequency inthe microwave region.

    *Microwaves are a form of electrodesmagnetic radiation with a frequency rang of300-30,000MHz and wave length varying

    from 10mm to 1m.

  • 7/30/2019 2-DAITHERMY fni$

    42/56

    *The most commonly used microwavefrequency for therapeutic heating is

    2450Mhz corresponding to a wavelengthof 12.25 cm.

    *microwavediathermy provides one ofmost valuable sources of therapeutic

    heat available to the physician

  • 7/30/2019 2-DAITHERMY fni$

    43/56

    Microwave diathermy

    *Effect of microwave diathermy are similar

    to short-wave yet in other better resultare obtained by using microwave.

    *The technique application of microwavediathermy is very simple. Not likeshortwave

  • 7/30/2019 2-DAITHERMY fni$

    44/56

    * the pads are used tobring in thepatient as a part of the circuit, themicrowave are transmitted from anemitter, and are directed towards theportion of the body to be treat.

    *Thus no tuning is necessary forindividual treatments These waves passthrough the intervening air space and areabsorbed by the surface of the body

    producing the heating effect.

  • 7/30/2019 2-DAITHERMY fni$

    45/56

    *such currents produced by special typeof device called*magnetron* is used for

    the production of high frequency currentsof high power.

    *The magnetron consist of cylindricalcathode surrounded by an anodestructure that contains cavities openinginto the cathode-anode space by mean

    slots.

  • 7/30/2019 2-DAITHERMY fni$

    46/56

  • 7/30/2019 2-DAITHERMY fni$

    47/56

    *the out put energy is derived from theresonator system by means of a coupling loop

    which is forced into one of the cavities.*The energy picked up on the coupling loop iscarried out of the magnetron on the central

    conductor of a co-axial out put tube througha glass seal to a director which consist ofradiating element of antenna and a reflectorwhich directs the energy for application to

    the patient

  • 7/30/2019 2-DAITHERMY fni$

    48/56

    *The electrical current is transformedinto electromagnetic radiation on passing

    through the antenna.

    *The reflector then focuses thiselectromagnetic energy and beams intothe where subsequence absorbed,reflected or refracted, according to theelectrical properties of the tissue

  • 7/30/2019 2-DAITHERMY fni$

    49/56

    *Tissues of lower water content arepenetrated to greater depth by this

    little is absorbed, whereas tissues ofhigh water content absorbedelectromagnetic energy but allow littlepenetration.

    *the out put powermagnetrondependsuponanode voltage, magnetic field andthe magnitude and phase of the loadimpedance to which the magnetron output power is delivered.

  • 7/30/2019 2-DAITHERMY fni$

    50/56

    *Therefore the cable used to carry theenergy from the magnetron to thedirector is always of a define length for aparticular frequency.

    *A part of the energy fed to the

    magnetron is also converted into heat inthe anode on account of the collision ofthe electron with the anode so that theout put energy is considerably less thanthe input energy.

  • 7/30/2019 2-DAITHERMY fni$

    51/56

  • 7/30/2019 2-DAITHERMY fni$

    52/56

    Repair Tips

    Problem*No Output in any mode.*Output Coagulation OK No Cut.

    *Cut OK no Coagulation.

    *Hand control OK Foot switch notworking.

    *Machine at no response.

    *Low output in Cutting OR CoagulationMode.

  • 7/30/2019 2-DAITHERMY fni$

    53/56

    Causes of these Problems

    No Output in any Mode:Pencil Error.Power Driver may be faulty.Fault in RF board .

    Output Coagulation OK No Cut:Fault in Pencil/foot paddle.Iso-block board must be checked.Multi-vibrator board may have some error.

    Output relay in the main board may notworking properly.

  • 7/30/2019 2-DAITHERMY fni$

    54/56

    Causes of these problems

    :Cut OK No CoagulationFault in Pencil/foot paddle.Iso-block board must be checked.Multi-vibrator board may have some error.

    Output relay in the main board may notworking properly.Hand control OK Foot switch not working.

    Foot switch connector damage.In continuity in the wires of paddle.

  • 7/30/2019 2-DAITHERMY fni$

    55/56

    Causes of these repairs

    Machine at No response.

    Error in Power switch.Problem in power board.

    Problem Cooler assembly.

    Low Output in Cutting or Coagulation mode.Fuse may be open.

    May be fault of Power Transistor.

    Rust on Patient Plate.

  • 7/30/2019 2-DAITHERMY fni$

    56/56