2025 Partisol Sequential Sampler. Deborah Bowe Thermo Electron Corporation. 2025 Verification & Routine Maintenance. Verification Ambient air temperature Filter temperature Ambient pressure External leak check Internal leak check Flow rate. 2025 Verification & Routine Maintenance. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: 2025 Partisol  Sequential Sampler

2025 Partisol Sequential Sampler

Deborah Bowe

Thermo Electron Corporation

Page 2: 2025 Partisol  Sequential Sampler

2

2025 Verification & Routine Maintenance

Verification• Ambient air temperature• Filter temperature• Ambient pressure• External leak check• Internal leak check• Flow rate

Page 3: 2025 Partisol  Sequential Sampler

3

2025 Verification & Routine Maintenance

Routine Maintenance• Filter cassettes: inspect damage or contamination

• Filter cassette magazines: Generally not necessary to clean, but if damaged certain parts can be replaced

• WINS impactor: In accordance with USEPA guidelines for PM2.5

• VSCC: In accordance with U.S. EPA guidelines

• External leak check: Every 4 weeks of use

• PM10 Inlet: Every 14 days of use

Page 4: 2025 Partisol  Sequential Sampler

4

2025 Verification & Routine Maintenance

Routine Maintenance• Particle trap filter: inspect 6 months, replace as

needed• V seals: Check and clean as needed• Rainhoods: Clean air screens 6 months• Batteries: Check every 6 months replace as

needed• Pump: Check every 6 months and rebuild; needed

average lifetime is 12 to 18 months

Page 5: 2025 Partisol  Sequential Sampler

5

2025 Plumbing Diagram

Page 6: 2025 Partisol  Sequential Sampler

6

2025 Backup Battery Check

Measure across 188_VBAT & PGND

Drops below 4.2VDC replace batteries

Page 7: 2025 Partisol  Sequential Sampler

7

2025 Pump CheckPressure Check: Quality Pump should give 40 psi on the exhaust

Page 8: 2025 Partisol  Sequential Sampler

8

Partisol 2000 Single Channel FRM Sampler

Deborah Bowe

Thermo Electron Corporation

Page 9: 2025 Partisol  Sequential Sampler

9

2000 FRM Audit Screen

Ambient temperature + 2 °C

Filter temperature + 2 °C

Ambient pressure + 10 mmHg

Page 10: 2025 Partisol  Sequential Sampler

10

2000 FRM External Leak CheckManual method

Record reading on gauge

30 seconds after pump

valve is closed

Should not drop more

than 8.5 inHg in 30

seconds. Corresponds to

80ml/min rate

Page 11: 2025 Partisol  Sequential Sampler

11

2000 FRM External Leak CheckSemi Automatic

Tolerance is 5” in 60 seconds

Page 12: 2025 Partisol  Sequential Sampler

12

2000 FRM Internal Leak Check

Manual: 8.5”Hg in 30 seconds

Semi-automatic: 9.3”Hg in 30 seconds

Page 13: 2025 Partisol  Sequential Sampler

13

Flow Schematic

Page 14: 2025 Partisol  Sequential Sampler

14

2000 FRM Flow Audit

Streamline Pro, FTS or Bios flow meter

+ 5% tolerance