25hpplus

Upload: catweazle999

Post on 03-Jun-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/13/2019 25hpplus

    1/2

    The new Model 25HP PLUS is an advanced ultrasonic gage that makes thickness, material velocity, and timeof flight measurements on materials such as composites, fiberglass, plastic, rubber, and cast metals that aredifficult or impossible to measure using standard ultrasonic gages. Measuring sound velocity with this gageallows monitoring of changes in material properties such as the degree of nodularity in cast metals.

    This 25HP PLUS features a large LCD that displays both the selected type of measurement in bold numerals

    and the live ultrasound waveform. This unique gage features a host of standard measurement features and asophisticated alphanumeric file-based datalogger.

    A-Scan Display for Waveform Verification

    Measurement Types:- Thickness- Velocity- Time of Flight

    Pulser Configuration:- Pulse Echo- Pitch Catch- Thru-Transmission

    Internal Alphanumeric File-Based Datalogger Stores

    18,000 thickness or 1,750waveforms

    Application Auto-Recall withStored Standard and CustomSetups

    Selectable Waveform UpdateRate (4, 8, 12 or 20Hz)

    Instant In/Mm Conversion

    Two Year Limited Warranty

    25HP PLUS Ultrasonic Thickness,Velocity, and Time ofFlight Gage

    Ultrasonic Thickness,Velocity, and Time of

    Flight Gage 25HP PLUS For Fiberglass, Composites, Rubber, Cast Metals, Plastics

    Input Thickness Via Optional Digital Caliper

  • 8/13/2019 25hpplus

    2/2

    Measurement Modes:Mode 1: Time interval between excitation pulseand first backwall echo. Using contacttransducers.Mode 2: Time interval between the first interfaceecho after the excitation pulse and the firstbackwall echo. Using delay line or immersiontransducers.Mode 3: Time interval between successivebackwall echoes following the first interface echoafter the excitation pulse. Using delay line orimmersion transducers.

    Measurement Types:- Thickness- Velocity- Time of Flight

    Pulser Configuration:- Pulse Echo- Pitch Catch- Thru-Transmission

    Thickness Measurement Range*:- Pulse Echo and Pitch Catch

    steel: 0.050 - 25.000 inches(1.27 - 635.00mm)

    plastic: 0.020 - 3.00 inches(0.50 - 75.00mm)

    - Thru-Transmissionsteel: 0.075 - 50.000 inches

    (1.91 - 1270.0mm)plastic: 0.035 - 6.000 inches

    (0.89 - 152.00mm)

    Velocity Measurement Range:0.02000 - 0.55110 inch/S(0.5080 - 13.9979mm/S).

    Time of Flight Measurement Range:0.00 - 225.00S.

    Resolution:- Thickness, keypad selectable:

    low 0.01 0.1mmStandard 0.001 0.01mm

    - Velocitylow 0.001 inch/S (00.01mm/S)standard 0.0001 inch/S (00.001mm/S)

    - Time of Flightfixed 000.01S

    Battery: 6V Rechargeable NiCad battery pack, orfield-replaceable alkaline AA batteries.

    Battery Life: 25 hours in normal measurementmode with backlight off.

    Fast-Charger: Two-hour Fast Charger withuniversal voltage.

    Auto Power Off:On or off. Unit automaticallyshuts off after six minutes (keypad selectable).

    Transducer Frequency Range: 0.5-5.0MHz(-3dB).Operating Temperature: -10C to +50C.(+14 to 122F).

    Keypad: Sealed color-coded keypad with tactileand audible feedback.

    DisplayDisplay: Liquid Crystal Display with backlight.Contrast keypad adjustable. Display area 4.0 x 3.39inch (102 x 86mm).

    Backlight: Backlit electroluminescent display with

    Power Save feature.Zoom Mode: Expands the horizontal span of thewaveform to the minimum range associated with eachtransducer and automatically centers the measuredechoes.

    Freeze Mode: Freeze and Un-Freeze live A-Scan withthickness.

    Hold/Blank (keypad selectable): Display blankswaveform and thickness, or holds waveform andthickness on LOS (Loss of Signal).

    Rectification: RF, half wave positive or negative, andfull wave.

    Waveform Display Range and Delay Control: Thehorizontal span of the waveform is set at selectedintervals with a fully adjustable delay.

    Ordering InformationModel 25HP PLUS Digital UltrasonicThickness/Velocity/Time of Flight Gage, AC or BatterOperation, 50-60Hz, with Internal AlphanumericDatalogger. Including:Universal Quick Charger/AC AdapterTransducer CableTest BlockCouplantRS-232 I/O CableCarrying CaseInstruction ManualTwo Year Limited Warranty

    Optional AccessoriesHPV/C Digital Caliper for Thickness InputPLUS/RPC Protective Rubber Boot w/Stand &

    Straps36DLP/SPC/KIT Protective Pouch with Neck StrapWin25DLPLUS Interface Program25DLP/EW Extended Limited Warranty, 3rd Yea2214E 5-Step Test Block, 1018 steel,

    EnglishUnits: .100", .200", .300", .400", .500

    2213E 5-Step Test Block, Aluminum,EnglishUnits: .100", .200", .300", .400", .500

    26DLPLUS/HDC Heavy Duty Shipping Case

    Test Blocks are available in metric units.

    Please contact Panametrics for other accessoriesavailable with the Model 25HP PLUS

    Case: Water-resistant, gasketed Lexan case withsealed connectors.

    Metric/English Units: (keypad selectable) Allowsinstant conversion between English and metricunits.

    Min/Max Mode:Works in conjunction with FASTMode for rapid location of minimum or maximumthickness reading and associated waveform.

    Two Alarm Modes: Programmable Hi-Low Alarmwith audible and visual indicators.1) Standard Hi-Low2) Previous thickness

    -Absolute-Percentage

    Two Differential Modes:1) Thickness difference between actualmeasurement and reference value.2) Percent difference between actualmeasurement and reference value.

    Application Auto-Recall: Automatically adjustsinternal parameters and zero offset for a widevariety of transducers.

    Stored Standard Setups: 25 stored transducersetups to allow fast, easy calibration forPanametrics standard transducers.

    Stored Custom Setups: Up to 35 stored customtransducer setups for best performance in specialapplications.

    Size: 9.375 x 5.45 x 1.5" (238 x 138 x 38mm).

    Weight: 2.1 lbs. (0.95Kg).

    Internal DataloggerDatalogger and RS-232: The 25HP PLUSidentifies, stores, recalls, clears, and transmitsthickness readings, waveform images, and gagesetup information via the RS-232 Serial Port.Baud Rate, Word Length, Stop Bits, and Parityare adjustable from the keypad.

    Maximum # of Stored Values:Standard: 18,000 thickness readings or 1,700waveforms with thickness.Upgrade 1: 36,000 thickness readings or 3,400waveforms with thickness.Upgrade 2: 54,000 thickness readings or 5,100waveforms with thickness.

    Location Codes: 8-character file name plus 16-character alphanumeric location code input.Multiple Comments per location.

    File Structures: Data can be stored in 7 standardor custom application-specific file structures.Reports: On-gage reporting of: Summary withstatistics, Min/Max with locations, and FileComparison. On-screen Comparison of currentand previous readings.

    25HP PLUS Specifications 25HP PLUS Specifications

    *Thickness range depends on material, transducer, surface condition, temperature, and SETUP selected.All specifications are subject to change without notice.

    ISO 9001 QualitySystem Registered

    48 Woerd Ave Waltham, MA 02453 USA TEL 781-899-2740 Toll free 800-225-8330FAX 781-899-1552 www.panametrics-ndt .com e-mail:info @panametrics-ndt .com