3.1. introduction synthesis techniquesshodhganga.inflibnet.ac.in/bitstream/10603/40351/8/08... ·...
TRANSCRIPT
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 61
3.1. Introduction Synthesis Techniques
In the presented research work, the synthesis of NiO and GDC
nanopowders for anode material was carried out by using solution combustion
synthesis. A thorough introduction of combustion synthesis has been discussed
in Chapter I as the research work is fully dependent on this technique for the
synthesis of anode material for IT-SOFC. For synthesis of GDC10 thin film on
NiO-GDC substrates, Spray Pyrolysis Technique (SPT) was used. Also to form
the complete single cell assembly i.e. NiO-GDC/GDC/LSCF-GDC, screen
printing was used to form thick layer of LSCF-GDC cathode on NiO-
GDC/GDC structures. Hence both these techniques have been discussed in the
present chapter.
3.2. Solution Combustion Synthesis (SCS)
SCS of oxide materials was unexpectedly discovered during the reaction
between aluminum nitrate and urea. The schematic of the SCS process is
shown in Fig. 3.1. A mixture of Al (NO3)3.9H2O and urea solution, when
rapidly heated around 500 °C in a muffle furnace foamed and ignited to burn
with an incandescent flame yielding voluminous white product which was
identified as α-Al2O3 [1].
Figure 3.1 Schematic of the SCS process
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 62
To understand the highly exothermic nature of this reaction, concepts
used in propellant chemistry were employed [2]. A solid propellant contains an
oxidizer like ammonium perchlorate and a fuel like carboxy terminated
polybutadiene together with aluminum powder and some additives. The
specific impulse (Isp) of a propellant, which is a measure of energy released
during combustion is given by the ratio of thrust produced per pound of the
propellant.
It is expressed as
��� = �� ���� ��������������������� � (3.1)
The highest heat Tc is produced when the equivalence ratio (Ψ = oxidizer/fuel
ratio) is unity. The equivalence ratio of an oxidizer and fuel mixture is
expressed in terms of the elemental stoichiometric coefficient.
� = ∑��� !!"#" $%�!�&"'"("$) * + $%"$�� #"!"#!�,+-*./0�1.* $#2/�34/∑��� !!"#" $%�!, '-#"$) * + $%"$�� #"!"#!�,+-*./0�1.* $#2/
(3.2)
A mixture is said to be stoichiometric when Ψ = 1, fuel lean when Ψ >
1, and fuel rich when Ψ < 1. Stoichiometric mixtures produce maximum
energy. The oxidizer/fuel molar ratio (O/F) required for a stoichiometric
mixture (Ψ = 1) is determined by summing the total oxidizing and reducing
valences in the oxidizer compounds and dividing it by the sum of the total
oxidizing and reducing valences in the fuel compounds. In this type of
calculation oxygen is the only oxidizing element; carbon, hydrogen, and metal
cations are reducing elements and nitrogen is neutral. Oxidizing elements have
positive valances and reducing elements have negative valences [3].
In solution combustion calculations, the valences of the oxidizing
elements was modified and considered as negative, and the reducing elements
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 63
as positive, similar to the oxidation number concept familiar to chemists.
Accordingly, the elemental valency of C, Al, and H is +4, +3, and +1,
respectively, and oxidizing valency of oxygen is taken as −2. The valency of
nitrogen is considered to be zero [4].
3.2.1. Role of Fuels
Urea is familiar as an ideal fuel for the combustion synthesis of high
temperature oxides like alumina and alkaline earth aluminates. However, a
need to employ alternate fuels to prepare oxides which are unstable above 1000
°C such as transition metal aluminates was necessary. In this regard, hydrazine-
based fuels like carbohydrazide (CH), oxalyl dihydrazide (ODH) and malonic
dihydrazide (MDH) which have low ignition temperature and are combustible
due to the presence of N–N bond that decomposes exothermically to N2 (N≡N)
were found to be suitable.
These fuels serve the following purposes
1. They are the source of C and H, which on combustion form
simple gaseous molecules of CO2 and H2O and liberate heat.
2. They form complexes with the metal ions facilitating
homogenous mixing of cations in solution.
3. They break down into components from which they are formed.
These components in turn decompose to produce combustible
gases like HNCO, NH3 which ignite with NOx .
The simple compounds such as urea and glycine are recognized as
potential fuels. The compounds containing N–N bonds in their molecular
formulae are particularly found to assist the combustion better. Some of the
important criteria that qualify an ideal fuel should
� Be water soluble.
� Have low ignition temperature (< 500 °C).
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 64
� Be compatible with metal nitrates, i.e., the combustion
reaction should be controlled and smooth and not lead to
explosion.
� Evolve large amounts of gases that are of low molecular
weight and harmless during combustion.
� Yield no other residual mass except the oxide in question.
� Be readily available or easy to prepare.
A combustion synthesis reaction is influenced by the type of fuel and the
fuel-to-oxidizer ratio. The exothermic temperature of the redox reaction (Tad)
varies from 1000 °C to 1500 °C. Depending upon the fuel used and the type of
metal ion involved, the nature of combustion differs from flaming (gas phase)
to nonflaming (smoldering and heterogeneous) type. Flaming reactions could
be attributed to the generation of gaseous products like nitrogen oxides (NOx)
by metal nitrates and HNCO, NH3, CO, etc., generated by fuels like urea.
Interestingly, some of these fuels were found to be specific to a particular class
of oxides. Urea for example is specific for alumina and related oxides.
Similarly, CH is specific for zirconia and related oxides; ODH for Fe2O3 and
ferrites; TFTA for TiO2 and related oxides; glycine for chromium and related
oxides, etc. The fuel specificity appears to be dictated by the metal–ligand
complex formation, the thermodynamics of the reaction as well as thermal
stability of the desired oxide formed. Theoretically any redox mixture once
ignited undergoes combustion. All metal nitrates on pyrolysis yield
corresponding metal oxides. The decomposition temperature of the metal
nitrates is lowered by the addition of a fuel. So the choice of fuel is critical in
deciding the exothermicity of the redox reaction between the metal nitrate and
the fuel. Depending upon the exothermicity of the reaction, combustion is
smoldering, flaming, or explosive. For example, aluminum nitrate-urea reaction
is highly exothermic (Tad ∼ 1500 °C) but is not explosive, probably due to the
thermal insulating nature of the alumina formed. Whereas, the transition metal
nitrate-urea reaction is violent. By changing the fuel from urea to glycine, the
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 65
combustion is much more controlled due to the complex formation of the metal
ions with ligands like CH.
3.2.2. Purpose of Glycine
The glycine plays a dual role in the combustion process. Glycine has the
ability of forming complexes with the metal cations in the solution. The
complexation improves the solubility of the metal cations and prevents
selective precipitation during the removal of water by boiling. The metal ions
are kept in solution and randomly distributed until the point of ignition and
oxide formation. This gives an oxide with great compositional uniformity. In
addition to the complexation, glycine serves as a fuel for the combustion
reaction. The combustion is driven by the oxidation of glycine by the nitrate
ions. Hence, in this study we have used glycine as a fuel.
3.2.3. Advantages of the Solution Combustion Synthesis
SCS has emerged as a novel technique for the synthesis wide of
spectrum of the materials like advanced ceramics, phosphors, catalysts,
intermetallics, composites, and nanomaterials [5]. This technique has various
advantages as given below,
a. It is a simple, low cost and fast process of material synthesis and does
not need special instrument as in other SHS methods.
b. It offers control over structure, homogeneity, and stoichiometry of the
products.
c. It is suitable for the formation of high-purity products.
d. Desired size (micron to nano) and shape (spherical to hexagonal) of the
products can be achieved by this process.
e. For industrial application, it is economically attractive and easy to scale
up. It permits incorporation of desired quantity of dopants in the oxide
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 66
hosts to prepare industrially useful materials such as phosphors as well
as high Tc cuprates and SOFC (solid oxide fuel cell) materials.
The typical set up used for combustion synthesis in the present study is
as shown in Fig. 3.2. It consists of glass dome having 30 cm diameter and 50
cm height, fixed in the well designed chamber. It prevents the nano powder
from splitting openly into air. It also helps for collecting the powder after
combustion. In this set up, hot plate is fixed such that the distance between the
hot plate and dome can be varied accordingly with size of beaker used for
combustion.
Glass Dome
Hot Plate
Exhaust
Figure 3.2 Typical combustion set up used for synthesis of nanopowders
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 67
3.3. Spray Pyrolysis Technique
Thin film synthesis techniques can be broadly classified as physical and
chemical techniques. The physical methods include sputtering, physical vapor
deposition, laser ablation and molecular beam epitaxy. The chemical methods
involve gas-phase deposition method and solution techniques. Chemical vapor
deposition (CVD) and atomic layer epitaxy (ALE) are gas-phase deposition
methods. However, spray pyrolysis, sol-gel, spin and dip coating, selective ion
rinsing etc employ precursor solutions. In the present studies, we have used
spray pyrolysis technique for the deposition of the GDC10 electrolyte as
chemical methods have given freedom to tailor microstructure in turn tune the
property [6]. Spray pyrolysis is simple, versatile, cost effective technique used
for the deposition of oxide thin films. It offers capability to deposit continuous
and uniform films. Also as solution precursor is used in this process, one can
deposit thin films with desired stoichiometry.
3.3.1. Basic Principle
Spray pyrolysis is nothing but the pyrolytic decomposition of precursor
salts of desired compounds to be deposited on a preheated substrate.
In this technique, fine aerosols of a precursor solution are directed
towards preheated substrates. Depending on the method of the atomization,
spray pyrolysis technique is classified into electrostatic spray deposition, Flame
spray deposition; Pressurized spray deposition, Ultrasonic spray pyrolysis and
Mist spray pyrolysis. Out of all these pressurized spray pyrolysis is very simple
and uses pressurized air as carrier gas. Spray parameters can be easily
controlled. Hence in the present case this technique is employed to deposit
GDC10 electrolyte films.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 68
3.3.2. Components of Spray Pyrolysis Unit
Fig. 3.3 shows the schematic diagram of the spray pyrolysis unit. It
consists of various parts like spray nozzle, rotor for nozzle motion, liquid level
monitor, gas regulator valve, hot plate, and air tight chamber. The details of the
above listed parts are given below.
a) Spray Nozzle
It is made up of glass and consists of the solution tube surrounded by the
specially designed glass bulb. It works on Bernoulli principle. With application
of the pressure to the carrier gas, pressure decreases at the tip of the spray
nozzle. Due to this pressure decrease, the solution is automatically sucked in
the solution tube. It facilitates the breaking of precursor solution into droplets.
Figure 3.3 Schematic diagram of the spray pyrolysis unit
Thermo-
couple
Heating Plate
Compressed air
Mechanical assembly for
nozzle motion
Substrates
Heater
Temperature
Controller
Electrical Supply
Compressor
Nozzle
To Exhaust
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 69
b) Rotor for nozzle motion
The to and fro motion of the nozzle and its speed is controlled by using
stepper motor controlled with microprocessor based controller. With this
arrangement one can monitor the movement of the nozzle over the required
length of hot plate. During to and fro motion of the nozzle provides sufficient
time for the nucleation which helps to form adherent deposit.
c) Liquid level monitor
The arrangement for changing the height of the precursor solution forms
the liquid level monitor. Height of the solution is measured from the upper
level of the solution to the tip of the nozzle. At constant carrier gas pressure,
spray rate of the solution depends on the height of the solution. Proper
monitoring of the spray rate is very important in the process of the spray
pyrolysis
d) Gas regulator valve
Gas regulator valve is used to regulate and control the gas flowing
through the nozzle. It is attached to the gas flow meter designed from the glass
tube of length 25 cm and diameter 1.5 cm. Since pressure depends on the size
of the air flow meter, the air flow meter should be calibrated from nozzle to
nozzle.
e) Hot plate
Hot plate is made up of rectangular iron plate of size 5 cm x 18 cm x 0.7
cm. It is heated with 2000 W heater coil. The temperature of the hot plate is
controlled by using temperature controller. The chromel- alumel thermocouple
is attached at the front side of the iron plate.
f) Air tight chamber
In spray pyrolysis technique fine aerosols are formed from the precursor
solutions. Some time precursor solutions contains toxic elements or number of
toxic gases are evolved during the thermal decomposition process. So it is
necessary to fix the spray pyrolysis system inside air tight chamber. The outlet
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 70
of the chamber is provided with the exhaust fan to remove the gases evolved
during the thermal decomposition process.
3.3.3. Steps Involved in Thin Film Formation
Viguie and Spitz [7, 8] classified chemical spray deposition process
according to the type of reaction. In all, variation in parameters like, solution
concentration, solution flow rate, carrier gas flow rate , droplet radius , nozzle
to substrate distance, substrate temperature were considered. When substrate
temperature is more than the decomposition temperature of the precursors, thin
film formation takes place in the following steps.
A) In the first step, an aqueous precursor solution is converted into
aerosols (droplets) by spray nozzle and the solvent evaporation takes
place.
B) In this step, vaporization of the solvent leads to the formation of
precipitate as the droplets approaches the substrates.
C) Pyrolysis of the precipitate occurs in succession before the
precipitate reaches the substrates.
D) When the precipitate reaches the substrate, nucleation and the growth
of the metal oxide thin films on the substrate takes place.
E) Finally, the growth of the nuclei leads to the formation of the
continuous thin layer of metal oxide.
3.4. Characterization Techniques
3.4.1. Introduction
Characterization plays very important role in optimizing as well as
monitoring the desired properties of the materials. Day by day instrumentation
is improving and modern experimental techniques are helping scientist to
understand more about the samples under study. These techniques provide data
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 71
that can be analyzed for efficient use of the materials. Generally structural and
morphological characterizations are carried out at room temperatures.
However, for application of the materials, properties are tested as per the
requirement of the temperatures, air or gas atmospheres. Present chapter
describes the working principle, specifications of the setup, advantages, and
disadvantages of the instruments used for the characterization of NiO-GDC
samples and investigating their properties. The various instrumental techniques
used are listed below.
A. Thermogravimetric Analysis (TGA) and Differential Thermal
Analysis (DTA)
B. X-ray Diffraction (XRD)
C. Scanning Electron Microscopy (SEM) and Field Emission Scanning
Electron Microscopy (FE-SEM)
D. Transmission Electron Microscopy (TEM)
E. Fourier Transform Infra Red Spectroscopy ( FTIR)
F. Fourier Transform Raman Spectroscopy ( FT-RAMAN)
G. DC electrical Measurement- Two probe and Four probe
H. AC impedance spectroscopy
3.5. Thermogravimetric and Differential thermal analysis (TGA-DTA)
The classical meaning of the analysis is dissolution, separation; break up
into constituent elements. In more comprehensive sense, it is nothing but
quantitative and qualitative determination of a physical quantity with
presentation of the measuring method, processing and interpretation of the
measured values within the framework of the model concept. More
specifically, thermal analysis is nothing but a group of techniques in which
property of the sample is monitored against time or temperature in a specific
gaseous atmosphere [9]. In thermogravimetric analysis, change in the weight of
the sample is continuously measured as a function of time or temperature.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 72
However, in differential thermal analysis (DTA), change of the difference in
temperature of the sample under study and reference sample is measured.
3.5.1. Thermogravimetric Analysis (TGA)
Principle of Operation
The schematic of the instrument is shown in Fig. 3.4. The system is
based on dual beam horizontal design that supports both TGA and DTA
measurements. A bifilar-wound furnace provides uniform controlled heating
upto 1500 °C. The sample contained in a pan and the reference (typically an
empty pan) sits on platinum sensor at the end of the each balance arm.
The TGA weight is measured by a tautband meter movement located at
the rear of each balance arm. An optically active servo arm loop maintains the
balance arm in the horizontal (null) position by regulating the amount of the
current flow through the transducer coil. An infrared LED light source and pair
of photo-sensitive diode detect movement of the arm. A flag at the end of the
balanced arm controls the amount of light reaching each photo sensor. As
Figure 3.4 Dual Beam electrobalance
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 73
weight is lost or gained, the beam becomes unbalanced, causing unequal light
to strike the photo-diodes. A restoring current is generated to eliminate this
imbalance and return to the null point. The amount of restoring current is a
direct measure of the weight change. The sample balance monitors actual
sample weight, while reference balance is used to correct the TGA
measurement for temperature effects like beam growth via a proprietary
algorithm.
3.5.2. Differential Thermal Analysis (DTA)
It is a technique in which the change of the difference in the temperature
between sample and a reference sample is monitored against time while they
are exposed to a temperature alteration. In this analysis, DTA curve is
obtained by plotting the temperature difference (∆T) on the ordinate with
endothermic reactions downwards (exothermic reactions upwords) and time t
or temperature T increasing from left to right. The shape and size of a typical
DTA curve is determined by the environment surrounding the sample and
reference materials , mechanism controlling the reaction and the material
characteristics. Practically one can carry out DTA of samples like minerals,
polymers, organic and inorganic materials. However, care must be taken that
reference material used in these experiments should be inert over the
temperature range studied, unreactive to the crucible or thermocouple and
preferably, similar in thermal properties to the sample. The reference
materials most frequently used in DTA of minerals and inorganics are α-
alumina, Al2O3 (which must be heated to around 1500 °C to remove
adsorbed water), titania, TiO2, or carborundum, SiC. These have thermal
conductivities in the range 3 to 40 W m-1 K-1 in the temperature range up
to 1000 °C. In order to make the sample and reference more similar, the
sample may be diluted with the reference, although this will decrease the
sample signal. Care must be taken that no reaction takes place between
sample and reference diluent. The use of inert gases such as nitrogen,
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 74
argon or helium, and reactive gases, particularly air and oxygen, is
possible with most of DTA instruments. However, their flow rates should be
adjusted properly as the thermal conductivity of the gases used will affect the
response of the sensors [10].
Specifications:
Model: Trans Analytical Instruments USA: SDT-2960
Temperature range: Ambient to 1500 °C
Sample capacity (max): 200 mg (350 mg including sample holder system)
Balance sensitivity: 0.1 µg
Heating rate: 0.1 to 100 ˚C/min 0.01 °C/min increments (ambient to 1000 °C)
0.1 to 25˚C/min 0.01 °C/min increments (ambient to 1500 °C)
3.6. X-Ray Diffraction (XRD)
X-ray diffraction is an indispensible tool for crystalline materials [11]. It
is found to be most powerful in determination of the crystal structure of the
materials. In general, characterizations of material can be classified into phase
(structural and compositional), physical and chemical properties. However
more emphasis should be paid on characterization of phase of the materials.
3.6.1. Principle of Diffraction
Basic principle of the X-ray diffraction can be found in text books by
various authors e.g. Buerger [11], Klug and Alexander [12], Cullity [13],
Tayler [14], Guinier [15], Barrett and Massalski [16].
The phenomenon of bending of radiation towards the geometrical
shadow of an aperture is called diffraction. Phenomenon of the diffraction can
be explained by the Huygen’s wave theory of radiation. According to this
theory, diffracted beams can interfere with each other giving rise to bright and
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 75
dark fringes in the shadow region depending upon their phase difference. The
diffraction grating shows such effect in visible radiations. In 1912, Max von
Laue demonstrated such effect with X-rays. Equivalent to the diffraction
grating, crystal lattice consist of the parallel lines of the atoms and inter planar
spacing between them could be successfully determined from the separation of
the fringes of the diffraction pattern. According to the Bragg’s diffraction law,
constructive interference of the diffracted beams takes place when it satisfies
Bragg’s condition,
2d sin θ = nλ (3.3)
where,
λ = wavelength of X-ray
θ = glancing angle (called Bragg angle)
d = inter planer spacing
n = order of diffraction
Further, successful efforts on the development of theory and
experimental procedure made the X-ray diffraction a unique tool for the study
of the solids.
3.6.2. Phenomenon of Diffraction
In 1895, German scientist Wilhelm Conrad Rontgen produced X- rays.
X-ray being electromagnetic wave, they interact with the electron cloud of the
atom. Thus, the fluctuating electric field of the X-rays makes the electrons of
the atom to vibrate with same frequency as that of X-rays. Hence, they emit
radiations of same frequency as that of incident X-rays which made them to
vibrate. Hence incident X-rays appears to be scattered by the atom. This X-ray
scattering takes place in all directions. Intensity of these scattered X-rays
depends upon the number of electrons in the atom. More number of electrons
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 76
more intense will be the diffraction. This scattering of X-rays by the electron is
given by the Thomson equation and called as the scattering power of an
electron �= /. Thus scattering power of the atom at rest is usually expressed as
scattering factor or form factor and is equals to the atomic number (Z) of the
atom. It is well known that at any temperature atoms in the unit cell vibrates at
their mean position. The scattering factor decreases with their amplitude of
vibration and also with the increase in the diffraction angle.
Thus scattering factor of an atom is given by the relation,
f = f?eABCDEFG
HF (3.4)
Where,
=? = Scattering factor of an atom when it is at rest and at 0°
λ = Wavelength of x-ray
θ = Angle of diffraction
B = A constant (called isotropic temperature factor). This factor is given by
B =8JKLK, where, LK= mean of square displacement of the atom from its mean
position. The exponential term in equation 3.2 is called as Debye- Waller
factor)
3.6.3. X-ray Diffraction Setup
X-ray diffraction experimental setup consists of X-ray source, sample
under investigation and the detector to pick up the diffracted X-rays. The
radiation source can be monochromatic or polychromatic and X-ray
diffractions can be carried out on single crystal or polycrystalline sample.
Single crystal samples are studied by Laue method, Weissenberg photograph
methods or most commonly with automated four circle diffractometer. The
powder samples are studied by Debye Scherrer photographic methods by the
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 77
diffractometer and radiation detection can be made with the help of
photographic film or radiation counter.
However for fast and most efficient detection, modern diffractometer use
image plates or CCD. Fig. 3.5(a) shows a typical block diagram of XRD unit.
The X-ray tubes use different targets like Cu, Co, Fe, Cr, Mo, Ag and
W. However, the selection of the X-ray tube depends upon the requirements,
sample character (absorbance and unit cell parameters), resolution required and
angle of coverage. Higher wavelength will give more resolved data but
sometimes it is difficult to collect full data. The wavelength of X-rays from
copper target is 0.1542 nm and is generally used for standard applications. This
target produces X-rays with two wavelengths Kα and Kβ. These β radiations
can be filtered with the help of the filter made up of element having atomic
number Z-1 of the target. i.e. in case of Cu target Ni is used as filter.
The sample in the form of powder either in form of smear or compact
flat pack or a capillary is exposed to monochromatic beam of X-ray and the
diffracted beam intensity is collected in the range of two theta angles with
Figure 3.5 (a) A typical block diagram of a powder XRD unit
Voltage Stabilizer
V High
voltage X-Ray
Tube
I
D
Monochromato
Sample
Collimato
Slits
Current Stabilizer
Amplifier Computer Printer
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 78
respect to the incident beam. The intensity corresponding to a constructive
interference of the diffracted beam from a crystallographic plane is observed as
a peak. A typical diffraction pattern is shown in Fig. 3.5 (b)
Each peak in the pattern can be assigned miller indices (h k l) which
equate the inter-planer spacing with the unit cell parameters. This typical
assignment of the (h k l) values to diffraction peaks is called as indexing. One
can determine crystallite size using the broadening of the diffraction peaks.
This broadening of the peaks takes place due to incomplete destructive
interference. Thus according to Debye Scherer formula,
D = ?.OPQ �R (3.5)
where, D = crystallite size, λ = wavelength, β = FWHM, θ = Bragg’s angle
Figure 3.5 (b) A typical XRD pattern
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 79
3.6.4. Applications of XRD
There are various applications of XRD data. Such as study of the nature
of the sample (crystalline or amorphous), phase identification, phase transition,
monitoring the solid state reactions, unit cell determination, crystallite size
determination, structure and strain determination, study of defects. Since the
XRD pattern is the finger print of materials, it can be used to identify the
materials.
The specifications of XRD unit used for the characterization of NiO-
GDC samples are given below.
Specifications:
Model: PW1710/PW3710 PHILIPHS Holland
Angle: 2θ = 10° to 100°
Target: Cu
3.7. Scanning Electron Microscopy (SEM)
Interaction of the electrons with elements is well understood and has
been extensively used for characterizing the materials. Interesting fact is that
electrons can be focused to micro to submicron size, hence are well suited for
analyzing submicron sized areas or features. When electron strikes the atom,
variety of interactions takes place. Scattering of the incident electrons from the
electron of the atom gives rise to the backscattered electrons and secondary
electrons. Some electrons may get transmitted if the sample is thin. However
primary electrons with sufficient energy may knock out the electron from the
inner shells of atom and excited atom may relax with the liberation of Auger
electron or X-ray photons. All these interactions of an electron carry
information about the sample. Scanning electron microscopy uses secondary
electrons for studying the topography of the sample.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 80
3.7.1. Principle of Operation
SEM is an instrument, which is used to observe the morphology of the
sample at higher magnification, higher resolution and depth of the focus as
compared to the optical microscope. Herein, an accelerated mono-energetic
beam of electrons is incident on the surface of the sample and scanned over
small area of the sample. Several signals are generated and are collected as per
the mode of operation. Detected signals are amplified and made to form the
synchronous image on the cathode ray tube (CRT). The camera is used to
photograph the images or it may be digitized and processed on a computer.
Magnetic
Lens
Scanningc
oil (X,Y)
Magnetic
Lens
Sample Current
Sample
Detector BE
Scanning
Generator
Image C.R.T
Magnification Generator
Video
Amplifier SE
Cathode
Whenlt
Anode
Beam
6V-100V 5-50KV
Figure 3.6 A typical line diagram of scanning electron microscope.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 81
The general outline of the SEM is shown in Fig. 3.6. The essential
features are a) electron gun b) electron optics c) electron signal detectors d)
specimen stage and e) vacuum system [17]
a) Electron gun
The electron gun used to get electrons is a triode gun as shown in
Fig.3.7. Tungsten filament works as a cathode. In order to emit the electrons
from the tip it is bent in ‘V’ shape. The diameter of the filament wire is of the
order of 0.1 mm. The filament is heated up by passing current through it. Thus
electrons are emitted by the thermionic emission.
The emission current is controlled by the wehnelt cylinder around the
filament and kept at negative voltage V. The electron beam is accelerated due
to anode having potential V. However for the safety of the operator it is
grounded. Like tungsten, lanthanum hexaboride (LaB6) is used as a cathode for
the emission of the electrons due to its lower work function (2.7 eV) as
compared to the tungsten. As emission of electrons from the LaB6 filament
takes place a lower temperature than the tungsten filament, its life span in
longer.
R
_V
Anode
Filament
Wehnelt
Figure 3.7 Ray diagram of electron gun
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 82
b) Electron optics
The electron beam is produced from the filament is focused using
electromagnetic lenses.
c) Electromagnetic lenses
The electron lenses help to focus the electron beam on the sample. The
size of the probe on the sample depends on the reduction factor of the lenses
used. The initial size of the beam at the aperture of the anode is 50µm is
reduced to 5 nm to 1 µm on the sample by using 2 or 3 lenses.
d) Electron Detectors
Backscattered electrons and the secondary electrons are detected using
electron detectors. Generally, Everhart-Thornly detectors are used for the
detection of the back scattered electrons. The grid near the sample is kept at -50
V to avoid secondary electrons from being collected. Back scattering
coefficient increases with atomic number Z. Two types of contrasts are
obtained from the backscattered electrons namely i) chemical contrast and ii)
topological contrast. Secondary electrons are detected with the scintillators
coupled with photomultiplier. These detectors are kept near the sample. The
topographical contrast due to secondary electrons results from the collected
intensity, increase with the angle of inclination of the sample. So the samples in
SEM are inclined.
3.7.2. Sample Preparation
Charging of the sample surface affects the imaging process. So it can be
avoided by coating platinum over the surface of the sample. Carbon or copper
tapes are used to ground the sample surface. Biological sample are specially
treated to immobilize them. Proper vacuum and the low acceleration voltages
are used to avoid damages to the sample.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 83
3.7.3. Applications
SEM is well suited for the morphological characterization of the
samples in varies fields like material science, biology, agriculture etc.
Specifications:
The instrument is of JEOL, Japan make JSM-6360. It uses an
acceleration voltage of 0.5 KV to 30 KV, a maximum magnification upto X
1,00,000 is possible. It offers a resolution of 3 nm. The instrument accepts a
specimen upto 150 mm in diameter.
3.7.4. Field Emission Scanning Electron Microscope (FE-SEM)
(a) Principle
Principle of the FE-SEM is same as common SEM. Only difference is
that it uses electrons librated by the field emission source. The object is
scanned by the electrons according to the zig-zag pattern.
(b) Source of electron in FE-SEM
In FE-SEM no heating but so called “cold” source is employed. In FE-
SEM, very thin and sharp tungsten (tip diameter 10-7 to 10-8 m) is used as a
filament. It works as a cathode in front of primary and secondary anodes. The
voltage of the order of 0.5 to 30 KV is applied across the cathode and anode.
This filament produces electron beam 1000 times smaller than the conventional
device. Hence resolution is better. One more advantage is that current flash
regularly decontaminates the electron source and life time of the filament is
more than conventional filaments.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 84
(c) Image formation
Image formation process is same as in a conventional SEM. However
difference is that it detects secondary electrons from the surface of the object
under investigation and uses those for the image formation.
(d) Applications
Researchers in material science, biology and chemistry are using FE-
SEM for observing the features of small size of the order of 1 nm.
3.8. Transmission Electron Microscopy (TEM)
TEM is a tool which can provide information about the microstructural
characterization broadly means ascertaining the morphology of phases, number
of phases, structure of phases, identification of the crystallographic defects and
composition of the phases.
3.8.1. Principle
In TEM, specimen under investigation is illuminated with the electrons
emitted from the electron guns. These electrons are passed through the sample
and transmitted electrons from the sample are recorded.
3.8.2. Working of the TEM
Basically TEM consists of electron gun, electromagnetic lenses, vacuum
system, high voltage generator, recording devices and the associated
electronics.
Fig. 3.8 shows the ray diagram of the TEM. TEM operates on principles
similar to that of the optical microscope [18, 19].
The accelerated ray of electrons passes a drill-hole at the bottom of the
anode. The lens-systems consist of electronic coils generating an
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 85
electromagnetic field. The ray is first focused by a condenser. It then passes
through the object, where it is partially deflected. The degree of deflection
depends on the electron density of the object. The greater the mass of the
atoms, the greater is the degree of deflection. After passing the object the
scattered electrons are collected by an objective lense. Thereby an image is
formed, that is subsequently enlarged by an additional lens-system (called
projective with electron microscopes). Thus the formed image is made visible
on a fluorescent screen or it is documented on photographic material. Photos
taken with electron microscopes are always black and white.
Figure 3.8 Ray diagram of Transmission Electron Microscope
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 86
The lanthanum hexaboride (LaB6) and tungsten filaments are found to
be the best source of electrons. The electrons produced by the gun are
accelerated by the chosen voltages. This acceleration voltage determines the
wavelength of the electron beam produced.
The wavelength and acceleration voltage is related by an equation, λ=
(1.5)1/2
/ (V)1/2
where, λ is the wavelength in nm and V is the acceleration
voltage in volts. For higher accelerating voltages relativistic correction
becomes very important and thus should be taken into account.
The resolution of the electron microscope depends not only on the
wavelength of the electron beam but also governed by the aberration of the
image forming lenses. The spherical aberration coefficient of objective lens is
very important and it is related to the wavelength of the electron beam and the
resolution (ropt ) of the TEM by the equation, ropt= 1.21 λ3/4 (Cs)1/4.
In modern TEM, developments are going on to reduce the spherical
aberration to increase the resolution. Further improvement in the resolution is
done by adding field emission gun (FEG). This FEG reduces the instrumental
contribution to the chromic aberration. Hence information from the planes of
lower interplanar spacing is also transferred to the image.
Since the limit of resolution is in the order of a few angstroms, it is a
very useful and powerful tool for nanoparticles characterization. Low
resolution TEM can generally provide information regarding the size and
overall shape of the sample and is routinely used to elucidate such information.
3.8.3. Applications of TEM
TEM is useful for determining size, shape and arrangement of the
particles which make up the specimen. TEM is typically used for high
resolution imaging of thin films of a solid sample for nanostructural and
compositional analysis. The topographic information obtained by TEM in the
vicinity of atomic resolution can be utilized for structural characterization and
identification of various phases of mesoporous materials, viz., hexagonal, cubic
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 87
or lamellar [20]. TEM also provides real space image on the atomic distribution
in the bulk and surface of a nanocrystal [21].
3.9. Fourier Transform Infrared (FTIR) Spectroscopy
A variety of IR techniques have been used in order to get information on
the surface chemistry of different solids. With respect to the characterization of
metal oxides two techniques largely used, namely the transmission/absorption
and the diffuse reflection techniques. The infrared spectroscopy is the method
of qualitative analysis of organic material and it has broad application in
inorganic substances as well. The molecular vibrations in the sample can be
detected by the vibration spectroscopy. The roots of the vibrational
spectrometry are strengthened by the discovery of inelastic scattering of the
photons also known as “Raman effect” in 1928 by Raman and Krishnan [22].
It is well known that, IR encompasses a spectral region from red end of
visible spectrum (12,500cm-1
, 0.8 mm) to the microwave (10 cm-1
, 1000 mm)
in the electromagnetic spectrum.
Figure 3.9 Block diagram of the FTIR
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 88
However, based upon applications and instrumentation involved, it is
divided into near (12,500 to 4000 cm-1
), mid-IR (4000 to 400 cm-1
) and far-IR
(400 to 10 cm-1). It is found that most of the fundamental molecular vibrations
occur in mid-IR region.
Modern Fourier transform IR spectrometers are superior to that of
dispersive IR spectrometers on several counts. Due to these advantages the
measurements of transmission, reflection or even emission spectra has become
significantly faster and even with higher sensitivity than ever before. FTIR
spectrometers are based upon Michelson interferometer.
Fig. 3.9 shows schematic of the spectrometer. A typical spectrometer
mainly consists of components like (a) radiation source, which is always Nernst
filament (ZrO2 + Y2O3) or Globar (SiC). (b) optical path and monochromator,
in which the beam is guided and focused by the mirrors aluminized or
silverized on their surfaces. (c) detectors, which detects the heat radiations .
To obtain an IR absorption spectrum, one mirror of the interferometer
moves to generate interference in the radiations reaching the detector. However
as all wavelengths are passing through the interferometer, interferogram is a
complex pattern. In these experiments the absorption/transmission spectrum as
function of wavenumber (cm-1
) is obtained from the Fourier transform of the
interferogram, which is the function of the mirror movement. Interestingly, this
design doesn’t have the reference cell which is generally used in the dispersive
IR instrument, so a reference spectrum is recorded and stored in the memory to
subtract from the sample spectrum.
3.9.1. Sample Preparation
Solid samples under test are mixed with KBr, palletized and used for
measurement. Liquid samples were taken in KBr/CaF2 window cells and gases
in gas cells.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 89
3.9.2. Applications
It is used for the determination of functional group attached to atom or
molecule and compositional characterization of materials both organic and
inorganic and it is widely used in areas like material science, catalysis,
medicine, biochemistry, forensic science etc.
Instrument Specifications:
Model: PERKIN ELMER SPECTRUM 1
Range: 400 –4000 cm-1
3.10. Raman Spectroscopy
In 1928, Prof. C.V.Raman discovered that when a sample is irradiated
with an intense monochromatic light source ( Laser), most of the radiation is
scattered by the sample at the same wavelength as that of the incoming laser
radiation in a process known as Rayleigh scattering. However, a small
proportion of the incoming light – approximately one photon out of million – is
scattered at the wavelength that is shifted from the original laser wavelength. Is
known as the Raman effect.
Vibrational spectroscopy deals with the study of the vibrational energy
states in materials. These states involve vibration of atoms around their
equilibrium position. Vibrational spectroscopy can be classified as Raman and
infra red spectroscopy and used for characterization of samples in solid, liquid
ands gaseous form. In solids, characteristic vibrations come from the
individual atoms as well as vibration of any rigid or semi rigid molecular unit
present in the unit cell. However in gases or liquids, characteristic vibrations
belong to the atomic vibration in individual molecules. Total number of
independent vibrations depends on the number of atoms and given by 3N-5 and
3N-6 for linear and non-linear molecules respectively. In this case number 5 or
6 represents the transitional and rotational motion of the molecules. On the
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 90
other hand corresponding numbers for the solid are 3N-3, where 3 correspond
to the three acoustic modes. The vibrational states are the characteristic features
of these materials and hence can be used to identify the materials.
In solids, Raman scattering is different than liquids and gases because in
solids atoms are bound and hence they lose transitional and rotational degrees
of freedom. The atoms and the molecules in the unit cell are close enough to
interact with each other. This gives rise to the changing of the number of
Raman scattered lines and their intensity distribution. In solids, vibrational
states are governed by the phonon dispersion relation which represent the
phonon frequency or energy with the phonon wave vector [23].
3.10.1. Principle of Operation
Prof. C.V. Raman received Nobel Prize for the discovery of Raman
effect. In Raman spectroscopy, a monochromatic incident radiation is scattered
inelastically from the different vibrational states of the sample and spectrum of
the scattered radiation contains the information of the vibrational energy levels
in terms of energy differences between the incident and scattered radiation.
Raman spectroscopy is two photon second order process and now a days
Raman spectrometers can record almost all kinds of as prepared samples.
3.10.2. Experimental Setup of Fourier Transform Raman Spectrometer
In place of visible excitation lasers, an FT-Raman spectrometer uses a
laser in the near infrared – usually at 1064 nm. At this wavelength fluorescence
is almost completely absent, however because of the 1/λ4 relationship between
Raman scattering intensity and wavelength, the Raman signal is weak. In
addition, silicon CCD detectors cannot be used in this region of the spectrum.
FT-Raman uses sensitive, single-element, near-infrared detectors such as
indium gallium arsenide (InGaAs) or liquid nitrogen-cooled germanium (Ge)
detectors. An interferometer converts the Raman signal into an interferogram,
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 91
permitting the detector to collect the entire Raman spectrum simultaneously.
Application of the Fourier transform algorithm to the interferogram converts
the results into a conventional Raman spectrum.
FT-Raman spectroscopy is particularly well-suited for bulk sample
analysis and can be configured to accept samples in most common formats,
including vials, cuvettes, tubes, plastic bags, bottles, powders, films and solids.
A typical Raman scattering setup consists of (a) an excitation source, (b)
a spectrometer/spectrograph, (c) a detector and (d) detection electronics with
output device as shown in Fig. 3.10. The excitation sources are usually in the
visible region and principle requirement is they should be highly
monochromatic (width < 0.1 cm-1
), power density a few KW/cm2 and
directional (beam divergence < 1 m rad). However, in compact and portable
Raman spectrometers, semiconductor and diode pumped solid state lasers are
used. The most common sources are 514.5, 532.0, 488.0 and 789.0 mm laser
lines in the visible region. The Laser light is incident on the sample and
Figure 3.10 Schematic diagram of the FT-Raman
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 92
scattered radiation is collected over a large solid angle by a collector lens and is
focused on the entrance of the spectrometer. This is collimated by a concave
mirror to fall on the grating which spatially separates the Raman lines. These
are then focused on another concave mirror on the exit slit and different Raman
lines are recorded by rotating the grating so that they are transmitted by the exit
slit in sequential manner. In a spectrograph, the exit slit is replaced by a broad
window with a multichannel detector, which can record full spectrum over a
finite range. Principal requirement of the Raman spectrometer is its stray light
rejection capacity. If more spectrometers are used in series one can reduce the
stray light but due to more monochromators in series will result in weak signal
due to finite reflectivity of mirrors and gratings. So for the practical purpose
compromise is done by using two monochromators.
The resolving power of the spectrometer is very important and is usually
described by linear dispersion at the exit slit and this depends on the number of
lines on the grating, focal length of the monochromator, number of stages of
the monochromator, and the wavelength.
3.10.3. Detector Assembly
Raman signals are weak hence very sensitive and low noise detectors are
needed. While choosing the detectors, one should consider (a) quantum
efficiency i.e. probability that incident photon on the detector generates the
measurable signal. (b) dark signal i.e the signal generated by the detector in
dark and (c) the response curve representing the quantum efficiency variation
with respect to wavelength.
Most commonly photomultiplier tubes (PMT) are used as a detector.
Raman spectrum is represented by plot of the counts/sec as a function of
Raman shift (cm-1
). In modern multi channel Raman spectrometers Charge
Coupled Device detectors are used. A CCD is a two dimensional array of
potential wells which stores the generated photo electrons and read out in a
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 93
very complicated fashion. Each well is known as a “pixel” and distance
between two consecutive pixels is ~ 25 mm.
3.10.4. Applications
Along with the identification of the materials, this technique is found to
be very much useful in studying the optical properties of the thin films,
superconductors, nano-crystals etc. Also it is widely used to study the phase
transition in materials under different external conditions like temperature,
pressure, electric and magnetic fields.
Instrument Specifications:
Model: MultiRam spectrophotometer
Source: Nd:YAG laser.
Resolution: 0.1cm−1.
3.11. DC Conductivity Measurement
There are various methods of measurement of the electrical conductivity
of the materials. It can be measured in terms of a dc or ac electrical conduction
in materials. The sample is biased by the dc power supply and variation in
current passing through the sample is measured. The resistivity of the sample is
given by the simple relation,
ρ = ST� (3.6)
where, U= resistance of the sample
V= area of cross section through which current is passing and
W= length of conductor.
One can easily determine the conductivity, as the reciprocal of the resistivity,
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 94
σ = 4ρ
(3.7)
The dc conductivity can also be represented as Arrhenius relation,
σ = σ?exp�− [\]�/ (3.8)
where, ^.is the activation energy for conduction
_ is the absolute temperature
K is the Boltzmann constant
σ? is the pre-exponential factor.
However, for the oxides, ionic conduction relation (3.8) can be
modified as,
σ = `σa� b exp�−[\]�/ (3.9)
In this case, conductivity is observed as a function of temperature. Depending
upon the resistance of the sample, dc conductivity can be measured by using
two probe and four probe technique.
3.11.1. Two Probe Technique
Two probe technique is simple to use but ohmic contacts are needed if
one wants to measure bulk resistance of the sample. Generally, this technique is
employed for the characterization of the samples with very high electrical
resistance. Figure 3.11 shows the schematic of the two probe technique.
The conductivity of the material is obtained by measuring the
conductance and the physical dimensions of the material. The material is cut in
to the shape known shape like rectangle having length�W/, width �c/ and
height�ℎ/.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 95
Since two copper wires are used as two contacts to the sample, it is known as
the two probe technique.
3.11.2 Four Probe Technique
This technique uses four contacts to the sample and hence termed as
four probe technique of conductivity. This technique is often used for the
measurement of the conductivity of the samples with low electrical resistance
[24]. The schematic of the four probe technique is shown in figure 3.12
Ammeter
I
Current source
Figure 3.12 Schematic of four probe technique for measurement of conductivity
e
Voltmeter
f
V
g
V I
Voltage
source Ammeter
W c
ℎ
Figure 3.11 Schematic of two probe technique for measurement of conductivity
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 96
In this technique, known current I is passed through the outer probes and
voltage V developed across the middle probes is measured. In this case
separation between the middle probes is considered as a length W and used to
measure the area of the sample.
3.12. AC Impedance Spectroscopy
Theoretical background of this technique has been discussed in Chapter
II. For the present study the impedance spectroscopy was carried out using a
Solartron 1260 FRA (frequency response analyser) at frequencies between 5Hz
to 10 MHz. The current amplitude was chosen in such a way that the voltage
response never exceeded a value of 12 mV rms.
These measurements were performed at an interval of 25 °C, each
temperature being kept constant with an accuracy of ± 5 °C. Sufficient
stabilization time was ensured at each particular temperature. The data analysis
was done using Z-view frequency response analyzer.
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 97
References
[1] J. J. Moore, H. J. Feng, Prog. Materials Sci, 39(1995) 275.
[2] A. G. Merzhanov, Int. J. Self-Propag High-Temp Synthesis, 2(1993)113
[3] J. Subrahmanyam, M. Vijayakumar, J. Materials Sci, 27 (1992) 6249.
[4] K. C. Patil, S. T. Aruna, S. Ekambaram, Current Opinion in Solid State
and Materials Science ,(2) (1997) 158.
[5] S. T. Aruna, A. S. Mukasyan, Current Opinion in Solid State and Materials
Science 12 (2008) 44–50.
[6] D. Beckel, A. Bieberle, A. Hutter, A. Harvey, A. Infortuna, U.P. Muecke,
M. Prestat, J. L. M. Rupp, L. J. Gauckler, Journal of Power Sources 173 (2007)
325.
[7] J.C. Viguie, J. Spitz, J. Electrochem. Soc. 122 (1975) 585.
[8] P.Bohac, L.J. Gauckler, Solid State Ionics, 119 (1-4) (1999) 317.
[9] Larry D Hanke, Handbook of Analytical methods for materials, materials
evolution and engineering, Inc. (2001).
[10] M. E. Brown, Hanbiik of Thermal Annalyasis and calorimetry, Elsevier
vol.1 (1998).
[11] M. J. Buerger, “X-ray Crystallography”, Wiley, New York, (1942).
[12] H. P. Klug, L. E. Alexander, “X-ray Diffraction Procedures”, Wiley, New
York, (1954).
[13] B. D. Cullity, “Elements of X-ray Diffraction”, Addison Wesley,
Massachusetts (1956).
[14] A. Taylor, “X-ray Metallography”, Wiley, New York, (1942).
[15] A. Guinier, “X-ray Diffraction “, Freeman, San Francisco, (1963).
[16] C. S. Barrett, T.B. Massalski, “Structure of metals”, McGraw-Hill, New
York, (1966).
[17] A. K. Dua, National Workshop on Advanced Methods for Materials
Characterizations, (2004) 51-58.
[18] M. De Graef, Introduction to Conventional Transmission Electron
Microscopy; Cambridge University Press: Melbourne, (2003).
Chapter III: Experimental Techniques
DEPARTMENT OF PHYSICS, SHIVAJI UNIVERSITY, KOLHAPUR Page 98
[19] V. Alfredsson, M. eung, A. Monnier, G. D. Stucky, K. K. Unger, F.
Schuth, J.Chem. Soc. Chem. Commun., 44 (1994) 921.
[20] R. Fryer, Chemical Applications of Transmission Electron Microscopy,
Academic Press, San Diego, 1979.
[21] Z. L. Wang, Characterization of Nanophase Materials, Ed: Z. L. Wang,
Wiley VCH, Weinheim, (2000) 37.
[22] C.V Raman, K.S. Krishnan, Nature, 121 (1928) 501.
[23] H. Poulet, J.P. Mathieu, Vibrational Spectra and Symmetry of Crystals,
Gordon and Breach, New York, (1976).
[24] R.V. Wandekar thesis ‘Synthesis And Physicochemical Studies Of
Property Measurements of Ceramic Materials And Their Applications’ BARC,
(2007) 53.