31p-mrsi of the brain with b1-shimmed noe enhancement v6
TRANSCRIPT
31P-MRSI of the brain with B1-shimmed NOE enhancement
B.L. van de Bank, S. Orzada, F. Smits, M.W. Lagemaat, A.K. Bitz and T.W.J. Scheenen
31P-MRSI of the brain with B1-shimmed NOE enhancement
Why spectroscopy (31P)• Understanding metabolic processes
• Energy metabolism and/or phospholipid metabolism
• Advantages of 31P over 1H• Comprehensible spectrum
• No overlapping resonances• Direct insight in forementioned processes• Lower Larmor frequency
• Longer wavelengths, less destructive interferences• Homogeneous excitation patterns
October 8th, 2014Page 2
0.511.522.533.54
31P-MRSI of the brain with B1-shimmed NOE enhancement
Our previous work
October 8th, 2014Page 3
Tacq: 16 minVoxel: 6.3cc
Wijnen et al. ISMRM 18 (2010) 377
31P-MRSI of the brain with B1-shimmed NOE enhancement
How did we do it at that time?
October 8th, 2014Page 4
Wijnen et al. ISMRM 18 (2010) 377
31P• Surface coil
• Big transceive element• Quadrature driven
•Inhomogeneous B1
• Spatially dependent field• Adiabatic pulses• SAR demanding
• High SNR
1H• Birdcage coil
• Inhomogeneous field distribution (cp mode)
• No optimization possible
31P-MRSI of the brain with B1-shimmed NOE enhancement
What do we want?
October 8th, 2014Page 5
31P• Surface coil
• Big transceive element• Quadrature driven
•Inhomogeneous B1
• Spatially dependent field• Adiabatic pulses• SAR demanding
• High SNR
1H• Birdcage coil
• Inhomogeneous field distribution (cp mode)
• No optimization possible
31P• Volume Tx coil, quadrature driven
• Array of receive coils• Phased signal addition
•Homogeneous B1
• Spatially independent field• Short Hard pulses• Lower SAR constraints
• High SNR
1H• Volume coil
• Independent acting channels
• Field optimization1
• B1-shimming
Metzger et al. MRM. 59:396-409 (2008)
31P-MRSI of the brain with B1-shimmed NOE enhancement
Aim• To design, develop and contruct a coil setup:
• Signal handling: 120.3MHz & 297.2MHz
October 8th, 2014Page 6
1H• Anatomical background imaging• B0- & B1-shimming• NOE Signal enhancement
31P• Homogeneous excitation• Spectroscopic imaging• Local receive SNR Boost
31P-MRSI of the brain with B1-shimmed NOE enhancement
Coil design – 1H
October 8th, 2014Page 7
1H
• Adapted existing design1
• 8 individual channels• Transceive elements• With meander
• Octogonal shaped• Ø |26cm (/28cm)
1Orzada et al. Proc. ISMRM 17 (2009) # 3010
31P-MRSI of the brain with B1-shimmed NOE enhancement
Coil design – 31P
October 8th, 2014Page 8
31P• Birdcage coil• Insert for 8CH headcoil• TXRX & TX-only • Quadrature driven• Plexiglass tube:
• Øout = 25cm, length = 18cm
• thk = 0,5cm, #legs = 8, HP
• Local receive array• 7 Channels• Receive only
• Ø = 5cm
• decoupled by overlap & pre-amp
31P-MRSI of the brain with B1-shimmed NOE enhancement
Coil design – 1H & 31P
October 8th, 2014Page 9
1H 31P
31P-MRSI of the brain with B1-shimmed NOE enhancement
Coil design – 1H & 31P
October 8th, 2014Page 10
31P-MRSI of the brain with B1-shimmed NOE enhancement
Coil design – 1H & 31P
October 8th, 2014Page 11
31P-MRSI of the brain with B1-shimmed NOE enhancement
Safety check – |E|- & |H|-fields
October 8th, 2014Page 12
|E|-field|H|-field
8CH
8CH + BC8CH + BC + RX
2
• Robot measurement• Probes1:
• H3DV7 - |H|
• ES3DV2 - |E|
• Tissue simulating Fluid:• 120MHz: ε = 76,5 & σ = 0,78 S/m
• 296MHz: ε = 56,3 & σ = 0,98 S/m
1Schmid & Partner Engineering AG (SPEAG), Zürich, Switzerland2Bitz et al. Proc. ISMRM 19 (2011) #490
31P-MRSI of the brain with B1-shimmed NOE enhancement
Transceive vs. Separate TX & RX
Boost SNR locally > 600%
SNR Birdcage
Bird
cage
RX-a
rray
SNR RX-array SNR RX-array
SNR Birdcage
SNR RX-array
SNR Birdcage
SNR RX-array
SNR Birdcage
SNR RX-array
SNR Birdcage
SNR RX-array
SNR Birdcage
SNR RX-array
SNR Birdcage
SNR
incr
ease
% SNR increase ( (RX-BC)/BC )
0
500
1000% SNR increase ( (RX-BC)/BC )
0
500
1000% SNR increase ( (RX-BC)/BC )
0
500
1000% SNR increase ( (RX-BC)/BC )
0
500
1000% SNR increase ( (RX-BC)/BC )
0
500
1000% SNR increase ( (RX-BC)/BC )
0
500
1000% SNR increase ( (RX-BC)/BC )
0
500
1000
0
100
200
300
400
500
600
700
800
900
1000 %
October 8th, 2014Pagina 13
31P-MRSI of the brain with B1-shimmed NOE enhancement
Enhancement?no NOE NOE
October 8th, 2014Pagina 14
31P-MRSI of the brain with B1-shimmed NOE enhancement
0
20
40
60
80
100
120
140
160
180
Whole brain• 3D MRSI
• Pulse acquire
0
20
40
60
80
100
120
140
160
180
0
180
160
140
120
100
80
60
40
20
Flip angle (°)
NO
E enhancement (%
)PCr signal (a.u.)
no NOE NOE
B1-map
October 8th, 2014Pagina 15 30 ± 9%
TR/TE: 1500/0.10ms
FA: 45°Block pulse: 0.3ms
FOV: 240x240x240mm3
Matrix: 12x12x8
Actual voxel size: 35cc
Tacq: 7:48
31P-MRSI of the brain with B1-shimmed NOE enhancement
All together• 3D MRSI
• NOE-enhancement• Receive array
October 8th, 2014Pagina 16
TR/TE: 1500/0.10ms
FA: 45°Block pulse: 0.3ms
FOV: 160x160x140
Matrix: 12x12x10
Actual voxel size: 7.3cc
TA: 15:55
31P-MRSI of the brain with B1-shimmed NOE enhancement
Summary• Development and functionality 7T coil setup
• 1H B1-shimming• Separating TX&RX
• SNR boost (up to 700%)• Signal enhancement
• ~30% • Nuclear Overhauser Effect
October 8th, 2014Pagina 17
Start exploiting all exciting features
Thank you very much for your attention
31P-MRSI of the brain with B1-shimmed NOE enhancement
Noise correlation
October 8th, 2014Page 19
• Check decoupling between individual coil elements
• Average noise correlation: 10 ± 7%• Maximum: 23% (CH 3&6)
31P-MRSI of the brain with B1-shimmed NOE enhancement
Field distribution altered?
October 8th, 2014Page 20
• Assess 1H field distribution• Is it altered by birdcage?
• Simulated1 & measured• With & without birdcage
+BC-BC
1CST Studio Suite, CST, Darmstadt, Germany