Bio-Potential Amplifiers
-
7/31/2019 69647_Origin of Bio-Potentials (1)
2/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
3/72
Basic Requirements for Amplifiers
Type of amplification: Voltage Amplification Current Amplification
High input impedance ( 10 MW) why? Isolation and protection circuits why? Low output impedance why?
High common mode rejection ration why? The appropriate frequency spectrum, SNR, gain, Calibration input to calibrate the amplifier
-
7/31/2019 69647_Origin of Bio-Potentials (1)
4/72
Example of Bio-Potential AmplifierECG Amplifier
Origins of the electrocardiogram Blood Cycle
The cardiac vector The ECG waveform Indicator of a good ECG
12-lead electrocardiography
-
7/31/2019 69647_Origin of Bio-Potentials (1)
5/72
Blood Cycle of the Heart
Objectives of the cycle: Provide Oxygenated blood to all body cells Remove Carbon-Dioxide accumulating in cells
Two Blood Cycles simultaneously Pulmonary cycle add oxygen and remove Co 2 Blood Cycle carry oxygen to body cells
-
7/31/2019 69647_Origin of Bio-Potentials (1)
6/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
7/72
The Heart Beat (Cardiac cycle)
Two Phases1. Two atriums contract and two ventricles relax2. Two ventricles contract and two atriums relax
3. Four chambers relax Terminology:
Systolic phase contraction
Diastolic phase relaxation Control is done via an independent nervous
system in the heart though electrical signals
-
7/31/2019 69647_Origin of Bio-Potentials (1)
8/72
Conduction mechanismof the heart
-
7/31/2019 69647_Origin of Bio-Potentials (1)
9/72
Sequence of Control Signals
SA node (heart pace maker) initiate signal toatriums causing it to contract blood flowsthrough valves to ventricles
Impulse flows till it reaches AV node AV node fires a signal along the bundle of his
and ventricles contract in top to bottom fashion
to push the blood out of the heart http://en.wikipedia.org/wiki/File:ECG_principle_slow.gif#file
http://en.wikipedia.org/wiki/File:ECG_principle_slow.gifhttp://en.wikipedia.org/wiki/File:ECG_principle_slow.gifhttp://en.wikipedia.org/wiki/File:ECG_principle_slow.gifhttp://en.wikipedia.org/wiki/File:ECG_principle_slow.gif -
7/31/2019 69647_Origin of Bio-Potentials (1)
10/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
11/72
Components of ECG
The ECG is drawn against time. It contains elements indicating
the temporal relationshipbetween different actions takingplace during a cardiac cycle.
The respective amplitudes of these elements and their
respective positions are studied
Picture from wikipedia.com
http://upload.wikimedia.org/wikipedia/commons/9/9e/SinusRhythmLabels.svg -
7/31/2019 69647_Origin of Bio-Potentials (1)
12/72
ECG Amplifiers
The electrical activity of theheart can be modeled as anElectric dipole in the thorax
This dipole is representedwith a vector that varies inamplitude and direction(cardiac vector)
Defined lead vectors: unitvectors with fixed orientation
Vector can be observed fromdifferent directions andangles
-
7/31/2019 69647_Origin of Bio-Potentials (1)
13/72
The electrocardiogram is measured by puttingelectrodes on specific locations on the body. The electrodes used, their respective wires and
resistors for measuring the ECG from aparticular direction is called the lead
Measurement in Frontal Plane Bipolar limb leads, I, II, III Unipolar limb leads, aV R, aV L, aV F
Precordial (chest) leads for transverse plane(V1-V6)
12-Lead Electrocardiography
-
7/31/2019 69647_Origin of Bio-Potentials (1)
14/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
15/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
16/72
Augmentedunipolar limb leads
33% increase in voltagemeasured by disconnectingthe measuring electrodefrom the Wilsons terminal
without affecting thedirection of the lead vector.Three new configurationsthus obtained as aVR, aVLand aVF. a stands foraugmented.A resistance of R/2 isadded to balance the inputimpedances seen by both
inputs of the diff. amplifier.
-
7/31/2019 69647_Origin of Bio-Potentials (1)
17/72
Homework 1Bio-Potential Amplifiers and ECG leads
1. Explain why Bio-Potential Amplifiers need to have the followingrequirements:
High Input Impedance Low output impedance Frequency response appropriate to the signal Isolation Circuit
2. Show that the voltage at wilson central point (figure 6.4) is the averageof the voltages at each node
3. Show that voltage at the augmented lead shown in figure 6.5 increasesthe output voltage and calculate this increase
Notes:
Homework due at beginning of Tue Oct 20th
class Late submission are subject to 10% decrease for everyday after the class No late submissions would be accepted after Sun Oct 24 th class
-
7/31/2019 69647_Origin of Bio-Potentials (1)
18/72
Bio-Potential Amplifier II
-
7/31/2019 69647_Origin of Bio-Potentials (1)
19/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
20/72
Electrode connections in uni-polar limbleads
-
7/31/2019 69647_Origin of Bio-Potentials (1)
21/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
22/72
Electrode connections in unipolar chestleads
-
7/31/2019 69647_Origin of Bio-Potentials (1)
23/72
Voltage and frequency ranges ofsome common bio-potential signals
-
7/31/2019 69647_Origin of Bio-Potentials (1)
24/72
l k d f l f
-
7/31/2019 69647_Origin of Bio-Potentials (1)
25/72
Block diagram of an earlier version ofan electrocardiograph
-
7/31/2019 69647_Origin of Bio-Potentials (1)
26/72
F
-
7/31/2019 69647_Origin of Bio-Potentials (1)
27/72
Frequencydistortion
The high f components of the signal are attenuatedyielding a rounding of sharp edges and droppingin the R-wave magnitude.This is a distortion in thesignal due to high f limitation.
A distortion due to low f limitation. The signal lookslike a differentiated one andthe stable baseline neededfor the clinical ECG is lost
-
7/31/2019 69647_Origin of Bio-Potentials (1)
28/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
29/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
30/72
Artifacts from large electrical transients
Example: Defibrillator
S l i T i
-
7/31/2019 69647_Origin of Bio-Potentials (1)
31/72
Solution: Transientprotection
A voltage protection scheme at the input of an ECG amplifier
-
7/31/2019 69647_Origin of Bio-Potentials (1)
32/72
SolutionVoltage-limiting devices used for input
protection
Gas dischargetube (neonlight)
Back-to-back silicon zenerdiodes
Parallelsilicon diodes
Current-voltagecharacteristic
2-20 V 50-90 V
How do voltage limiting devices
protect the input of the amplifier?
-
7/31/2019 69647_Origin of Bio-Potentials (1)
33/72
60-Hz power line andelectromyography interference
60-Hz power-line interference on the ECG
Electromyography interference on the ECGSolution : Proper filter design
-
7/31/2019 69647_Origin of Bio-Potentials (1)
34/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
35/72
Effect of Frequency
Let go current Higher frequency,
less dangerous Why dont we
generateelectricity at high
frequencies?
-
7/31/2019 69647_Origin of Bio-Potentials (1)
36/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
37/72
2-Generation
and effects ofmagneticfield
Protection through alternative path for the current
C Pi k d i h B d
-
7/31/2019 69647_Origin of Bio-Potentials (1)
38/72
Currents Picked up in the Body1-Magnetic field pickup
SolutionTwisting lead wires togetherand keeping them close tobody minimizes interference
Lead wires for lead I making aclosed loop with patient and
ECG machine
Currents Picked up in the Body
-
7/31/2019 69647_Origin of Bio-Potentials (1)
39/72
Currents Picked up in the Body2-Pickup due to displacement current flowing through the patient
GdbcmZ iv =
=in
cm B A Z Z Z
vvv 12
Z 1 and Z 2
-
7/31/2019 69647_Origin of Bio-Potentials (1)
40/72
C t Pi k d i th B d
-
7/31/2019 69647_Origin of Bio-Potentials (1)
41/72
Currents Picked up in the Body4-Electromagnetic interference
EM waves generated by Radar facilities X-ray machines Nearby transformers Radio waves
EM waves picked-up by patient and lead wires Demodulated by p-n junctions of transistors and/or
electrode-electrolyte interfaces Modulating audio signal appears as interference on
top of the ECG signal Solution: Can be eliminated by shunting the input
terminals of the ECG amplifier with a small capacitor(around 200pF)
-
7/31/2019 69647_Origin of Bio-Potentials (1)
42/72
Solution: Electrostatic shielding
-
7/31/2019 69647_Origin of Bio-Potentials (1)
43/72
Driven-right-leg circuit
Right leg connected to the output of an OPAMP instead of ground Why?1. Displacement current flows through output resistance instead of
body2. Patient un-grounded when high voltage appears between
patient and ground (R f and R o large values)
-
7/31/2019 69647_Origin of Bio-Potentials (1)
44/72
Analysis of the driven-right-leg circuit
-
7/31/2019 69647_Origin of Bio-Potentials (1)
45/72
Electrical and ComputerEngineering (Biomedical
45
Bio-Potential Amplifiers 3
-
7/31/2019 69647_Origin of Bio-Potentials (1)
46/72
Electrical and ComputerEngineering (Biomedical
46
2-Generation
and effects ofmagneticfield
Protection achieved through alternative low resistancepaths for the current
Currents Picked up in the Body
-
7/31/2019 69647_Origin of Bio-Potentials (1)
47/72
Electrical and ComputerEngineering (Biomedical
47
Currents Picked up in the Body1-Magnetic field pickup
SolutionTwisting lead wires togetherand keeping them close tobody minimizes interference
Lead wires for lead I making aclosed loop with patient and
ECG machine
Currents Picked up in the Body
-
7/31/2019 69647_Origin of Bio-Potentials (1)
48/72
Electrical and ComputerEngineering (Biomedical
48
Currents Picked up in the Body2-Pickup due to displacement current flowing through the patient
GdbcmZ iv =
=in
cm B A Z Z Z
vvv 12
Z 1 and Z 2
-
7/31/2019 69647_Origin of Bio-Potentials (1)
49/72
Electrical and ComputerEngineering (Biomedical
49
Currents Picked up in the Body3-Electrical field pickup by connecting wires and instrument
What is a Capacitor?Coupling between hot sideof the power line and leadwires
2211 Z i Z ivv d d B A =With i d1 i d2
)( 211 Z Z ivv d B A =
How to minimize the current picked up by lead wires and instrument?
Currents Picked up in the Body
-
7/31/2019 69647_Origin of Bio-Potentials (1)
50/72
Electrical and ComputerEngineering (Biomedical
50
Currents Picked up in the Body4-Electromagnetic interference
EM waves generated by
Radar facilities X-ray machines Nearby transformers Radio waves
EM waves picked-up by patient and lead wires Demodulated by p-n junctions of transistors and/or
electrode-electrolyte interfaces Modulating audio signal appears as interference on
top of the ECG signal Solution: Can be eliminated by shunting the inputterminals of the ECG amplifier with a small capacitor(around 200pF)
-
7/31/2019 69647_Origin of Bio-Potentials (1)
51/72
Electrical and ComputerEngineering (Biomedical
51
Solution: Electrostatic shielding
-
7/31/2019 69647_Origin of Bio-Potentials (1)
52/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
53/72
Electrical and ComputerEngineering (Biomedical
53
Analysis of the driven-right-leg circuit
-
7/31/2019 69647_Origin of Bio-Potentials (1)
54/72
Electrical and ComputerEngineering (Biomedical
54
0
2
=+ f o
a
cm
R
v
R
v
cma
f o v R
Rv
2=
od RLcm vi Rv +=
KCL at point x
i.e.
But
Therefore
d
a
f
RL
cm i R
R
R
v 21 +=
What is the effective resistance betweenright leg and ground?
Large transients Saturation choselarge R f and R o =~ 5 M
Regular operation want v cm as small aspossible large R f and small R a
Driven Right Leg Circuit
-
7/31/2019 69647_Origin of Bio-Potentials (1)
55/72
Bio-Potential Amplifiers -4
-
7/31/2019 69647_Origin of Bio-Potentials (1)
56/72
Example of a simple ECG amplifier
-
7/31/2019 69647_Origin of Bio-Potentials (1)
57/72
Examples of Bio-Potential
Amlifiers
Bi di l Si l P
-
7/31/2019 69647_Origin of Bio-Potentials (1)
58/72
Biomedical Signal ProcessorExamples
Cardiac Tachometers Electromyogram integrators Fetal electrocardiography Cardiac monitors Biotelemetry
-
7/31/2019 69647_Origin of Bio-Potentials (1)
59/72
TachometersAveraging type
-
7/31/2019 69647_Origin of Bio-Potentials (1)
60/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
61/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
62/72
EMG integrators
Switch
Absolute-valuecircuit
Monostablemultivibrator
Comparator
C
EMG
Integrator
+
R
P 1
v t
1
Counter
2
3
-
7/31/2019 69647_Origin of Bio-Potentials (1)
63/72
EMG integrators
-
7/31/2019 69647_Origin of Bio-Potentials (1)
64/72
Fetal electrocardiology
A t h i f i l ti f t l ECG f
-
7/31/2019 69647_Origin of Bio-Potentials (1)
65/72
A technique for isolating fetal ECG frommaternal
-
7/31/2019 69647_Origin of Bio-Potentials (1)
66/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
67/72
TelemetryFrequency modulation
-
7/31/2019 69647_Origin of Bio-Potentials (1)
68/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
69/72
Digital landline telemetry system
ree-c anne me- v s onl i l d di l
-
7/31/2019 69647_Origin of Bio-Potentials (1)
70/72
multiplexed radiotelemetrytransmitter
Example of outputwaveform from commutator
-
7/31/2019 69647_Origin of Bio-Potentials (1)
71/72
-
7/31/2019 69647_Origin of Bio-Potentials (1)
72/72
Three-channelfrequency-divisionmultiplexed
radiotelemetrysystem