6998312603_rev0 (system integration guide)
TRANSCRIPT
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
1/74
Release 7 April 2011 (Rev. 0)
APOLLO EZ
System Integration Guide
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
2/74
SYSTEM INTEGRATION GUIDEContents
(Rev. 0) APOLLO EZi
Contents
1 INTRODUCTION 1
2 CONFIGURATION WITH APOLLO EZ/ CCD N23 / G100R/F-A /CHEST STAND 3
2.1 Power supplies connection instruction.................................................... 4
2.2 Hardware interface instruction ............................................................... 5
2.2.1 APOLLO EZ – Generator interface ........................................................5
2.2.2 Generator – I.I. component plate interface............................................ 6
2.2.3
APOLLO EZ – I.I. component plate interface ......................................... 7 2.2.4 Monitor interface.................................................................................8
2.2.5 Generator – Three field solid state AEC device 1 ................................... 9
2.2.6 Generator - X-ray tube interface ........................................................ 10
2.2.7 Chest Stand VD3 35x43 – Generator interface ...................................11
2.2.8 Generator – Three field solid state AEC device 2 ................................. 12
2.2.9 Generator jumpers and wiring configuration instruction .................... 13
2.2.9.1 Room interface board jumpers configuration.......................13
2.2.9.2 Room interface board jumper wiring...................................13
2.2.9.3 Generator interface PWB jumpers configuration..................13
2.2.9.4 Accessories connection.......................................................14
2.3 Generator set-up programming instruction........................................... 15
2.3.1 Tube selection................................................................................... 15
2.3.1.1 X-ray tube present in library ..............................................15
2.3.1.2 Loading X-ray tube data from GenWare® utility software ....16
2.3.1.3 X-ray tube not present neither in library nor in
GenWare® utility software..................................................16
2.3.2 AEC set-up ....................................................................................... 17
2.3.3 Fluoro set-up ....................................................................................17
2.3.4 Generator Image receptor programming worksheet ............................19
2.3.5 Generator Receptor 1 I/O configuration worksheet............................. 20
2.3.6
Generator Receptor 2 I/O configuration worksheet............................. 21
2.3.7 Generator Receptor 3 I/O configuration worksheet............................. 22
3 CONFIGURATION WITH APOLLO EZ/ CCD N33 / G100R/F-A /CHEST STAND 23
3.1 Power supplies connection instruction.................................................. 24
3.2 Hardware interface instruction ............................................................. 25
3.2.1 APOLLO EZ – G 100 R/F-A interface.................................................. 25
3.2.2 G 100 R/F-A – I.I. component plate interface .....................................27
3.2.3
APOLLO EZ – I.I. component plate interface ....................................... 28
3.2.4 Monitor interface............................................................................... 29
3.2.5 G 100 R/F-A – Three field solid state AEC device 1............................. 30
3.2.6 G 100 R/F-A - X-ray tube interface.................................................... 31
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
3/74
SYSTEM INTEGRATION GUIDEContents
APOLLO EZ (Rev. 0)ii
3.2.7 Chest Stand VD3 35x43 – G 100 R/F-A interface............................... 32
3.2.8
G 100 R/F-A – Three field solid state AEC device 2 ............................ 33
3.2.9 G 100 R/F-A jumpers and wiring configuration instruction................ 34
3.2.9.1 Room interface board jumpers configuration ......................34
3.2.9.2 Room interface board jumper wiring................................... 34
3.2.9.3 Generator interface PWB jumpers configuration ................. 34
3.2.9.4 T.D.I. board SW1 Dip-switches setting ............................... 35
3.2.9.5 Accessories connection ......................................................35
3.3 G 100 R/F-A set-up programming instruction ...................................... 36
3.3.1 Tube selection................................................................................... 36
3.3.1.1 X-ray tube present in library.............................................. 36
3.3.1.2
Loading X-ray tube data from GenWare® utility software .... 37 3.3.1.3 X-ray tube not present neither in library nor in
GenWare® utility software..................................................37
3.3.2 AEC set-up ....................................................................................... 38
3.3.3 Fluoro set-up .................................................................................... 38
3.3.4 G 100 R/F-A Image receptor programming worksheet........................ 41
3.3.5 G 100 R/F-A Receptor 1 I/O configuration worksheet ........................ 42
3.3.6 G 100 R/F-A Receptor 2 I/O configuration worksheet ........................ 43
3.3.7 G 100 R/F-A Receptor 3 I/O configuration worksheet ........................ 44
4 CONFIGURATION WITH APOLLO EZ/ DIVA-D / G100R/F-A /
CHEST STAND VD3 45
4.1 Power supplies connection instruction.................................................. 46
4.2 Hardware interface instruction ............................................................. 47
4.2.1 APOLLO EZ – G 100 R/F-A interface ................................................. 47
4.2.2 APOLLO EZ – DIVA-D interface.......................................................... 49
4.2.3 G 100 R/F-A – DIVA-D interface........................................................ 50
4.2.4 G 100 R/F-A – DIVA-D serial connection ........................................... 50
4.2.5 APOLLO EZ – I.I. component plate interface....................................... 51
4.2.6 DIVA-D – I.I. component plate interface ............................................. 52
4.2.7
Monitor interface............................................................................... 52 4.2.8 G 100 R/F-A – Three field solid state AEC device 1 ............................ 53
4.2.9 G 100 R/F-A - X-ray tube interface.................................................... 54
4.2.10 Chest Stand VD3 35x43 – G 100 R/F-A interface............................... 55
4.2.11 G 100 R/F-A – Three field solid state AEC device 2 ............................ 56
4.2.12 G 100 R/F-A jumpers and wiring configuration instruction................ 57
4.2.12.1 Room interface board jumpers configuration ......................57
4.2.12.2 Room interface board jumper wiring................................... 57
4.2.12.3 Generator interface PWB jumpers configuration ................. 57
4.2.12.4 T.D.I. board SW1 Dip-switches setting ............................... 58
4.2.12.5 Accessories connection ......................................................58
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
4/74
SYSTEM INTEGRATION GUIDEContents
(Rev. 0) APOLLO EZiii
4.3 G 100 R/F-A set-up programming instruction ...................................... 59
4.3.1
Tube selection................................................................................... 59
4.3.1.1 X-ray tube present in library ..............................................59
4.3.1.2 Loading X-ray tube data from GenWare® utility software ....60
4.3.1.3 X-ray tube not present neither in library nor inGenWare® utility software..................................................60
4.3.2 AEC set-up ....................................................................................... 61
4.3.3 Fluoro set-up ....................................................................................61
4.3.4 G 100 R/F-A Image receptor programming worksheet........................ 64
4.3.5 G 100 R/F-A Receptor 1 I/O configuration worksheet ........................ 65
4.3.6 G 100 R/F-A Receptor 2 I/O configuration worksheet ........................ 66
4.3.7 G 100 R/F-A Receptor 3 I/O configuration worksheet ........................ 67
This publication can only be reproduced, transmitted, transcribed or translated intoany human or computer language with the written consent of VILLA SISTEMIMEDICALI S.p.A.
This Manual in English is the original version.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
5/74
SYSTEM INTEGRATION GUIDEContents
APOLLO EZ (Rev. 0)iv
THIS PAGE IS INTENTIONALLY LEFT BLANK
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
6/74
SYSTEM INTEGRATION GUIDEIntroduction
(Rev. 0) APOLLO EZ1
1 INTRODUCTION This Manual includes the interconnections list, the hardware andsoftware configuration for a typical radiological room using a generatorG 100 R/F-A.For the use of this generator with different configurations of room and forX-ray tube and AEC calibration, please make reference to theG 100 R/F-A Service Manual and equipment you have to connect with.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
7/74
SYSTEM INTEGRATION GUIDEIntroduction
APOLLO EZ (Rev. 0)2
THIS PAGE IS INTENTIONALLY LEFT BLANK
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
8/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ3
2 CONFIGURATION WITH APOLLO EZ/CCD N23 / G100R/F-A / CHEST STAND
Room configuration
• Apollo EZ
– I.A.E. X-ray tube type
– Three field solid state AEC
–
Standard I.I. – CCD N23.• G 100 R/F-A
• Chest Stand VD3 35x43 with motorised foot rest (See Note )
– Three field solid state AEC.
Receptor configuration
Receptor 1: Apollo EZ Spot Film Device – Tube 1
Receptor 2: Direct X-ray from console – Tube 1
Receptor 3: Chest Stand VD3 35x43 – Tube 1 (See Note )
NOTE:When the Generator is supplied with "Falling load" feature, the GalKeyU29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the"SERIAL" exposure is enable (paragraph 2.3.4 - Image ReceptorProgramming Worksheet: "SERIAL=YES"); whenever the Falling Load hasto be activated on the Spot Film Device receptor, the SERIAL parameterhas to be set to "NO".
In case Falling Load feature and SERIAL exposure are required will benecessary to create for each a dedicated receptor.
NOTE:Reference to the Chest Stand VD3 35x43 with motorized foot rest is madein order to allows the installer to record the necessary information for thehardware interface among Generator and Potter Bucky.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
9/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)4
2.1 Power supplies connection instruction
( * ) If this remote controlled switch is not available use a remote controlledswitch with the following characteristics:
–Coil: 220-230 Vac 50 Hz
230-240 Vac 60 Hz
– Contact: 4 N.O. contacts, 16 A.
(**) Made this connection only when a motorised foot rest chest stand is present.
MainPower Supply
GeneratorTerminals
R
S
T
N
R
S
T
N
R S T N
CHEST STANDTerminal X1
R S T
APOLLO EZTerminals R-S-T (**)
G 100 R/FROOM INTERFACE BOARD TB7.1 (0 Vac)
G 100 R/FROOM INTERFACE BOARD TB10.1 (220 Vac)
* INSTALLATION KIT230Vac AUXILIARY CONTACTORV.S.M. cod.4291515200
(The auxiliary contactor is already mountedinside the G 100 R/F-A Generator)
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
10/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ5
2.2 Hardware interface instruction
2.2.1 APOLLO EZ – Generator interface
APOLLO EZ Generator
Description Terminal Terminal Description
X27.15 TB6.9
Remote exposure
(Room Interface Board)
X27.3 TB6.7Remote prep
(Room Interface Board)
X-rayrequest(2° step)
X27.16 TB2.11Bucky 1 select
(Room Interface Board)
Common0VS
X0.20 TB6.1Ale output
(Room Interface Board)
TB6.11Spare output
(Room Interface Board)X-ray ON
fromgenerator
X0.13 TB6.2Ale output
(Room Interface Board)
TB3.6 Tomo exposure
(Room Interface Board)
Generator
ready X0.12 TB2.12
Bucky 1 select
(Room Interface Board)
Tomo speed
1-2-3-4
X27.4
X27.1
X27.14
X27.17
TB3.4Remote tomo select
(Room Interface Board)
SFDselection
X0.15 TB6.12Spare output
(Room Interface Board)
Safetyswitch
X0.9 J17.1Safety switch
(Generator Interface PWB)
Safetyswitch
X0.11 J17.2Safety switch
(Generator Interface PWB)
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
11/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)6
2.2.2 Generator – I.I. component plate interface
Generator N23 I.I. component plate
Description Terminal Wire n° Pcb A2 Description
Remote Fluoro Exposure
(Room Interface board)TB6.5 X2.3 Fluoro
0Vdc(Generator Interface PWB)
J7.7 X2.1 0Vdc
Proportional dc(Generator Interface PWB)
J7.12 X2.2 Analog kV auto
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
12/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ7
2.2.3 APOLLO EZ – I.I. component plate interface
APOLLO EZ I.I. component plate
Pcb A1
Description Terminal Wire n°4 fields 3 fields
Description
Common X28.1 X4.4 X4.4 Common
Medium field X28.2 X4.2 X4.1 Medium zoom
Minimum field X28.14 X4.3 X4.2
Maximum
zoom
Maximum field X28.15 X4.1 N.C. Minimum
zoom
Power supply N X0.2 Power supply
Power supply S3 X0.1 Power supply
Ground X0.3 Ground
APOLLO EZ I.I. component plate
Description Terminal Wire n° Pcb A2 Description
X28.18 X3.3 Fluoro request
X28.4 X3.4 Inv
X28.16 X3.5 Inv
0 Vs common X28.3 X3.2 0 Vs common
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
13/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)8
2.2.4 Monitor interface
I.I. component plate Table and wall version
Description Pcb A2 Wire n° Plug Description
0 Vs common X4.1 Green 0 V 0 Vs common
Monitor L.I.H.visualization
X4.5 White L IMonitor L.I.H.visualization
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
14/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ9
2.2.5 Generator – Three field solid state AEC device 1
NOTE:If the AEC interface PWB it is not supplied with LEMO connectors thefollowing hardware interface have to be done. The installer should cut theAEC cable by the PWB side connector and connect the colored wires tothe terminals on the Universal AEC board making reference to the nexttable.
Generator
AEC board(solid state chamber)
type 737992
AEC device 1connecting cable
Description Terminal Wire Description
Anode (L) J11.1 Pink Anode Left
Cath (L) J11.2 Grey Cathode Left
Anode (M) J11.3 White Anode Middle
Cath (M) J11.4 Brown Cathode Middle
Anode (R) J11.5 Green Anode Right
Cath (R) J11.6 Yellow Cathode Right
Ground J11.7 Shield Ground
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
15/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)10
2.2.6 Generator - X-ray tube interface
Generator X-ray tube
Description Terminal Wire n° Terminal Description
Main Tube 1 P Principal
Shift Tube 1 A Auxiliary
Common Tube 1 C Common
Ground Tube 1 Ground
Thermal switch Tube 1 Thermal switch
Thermal switch Tube 1 Thermal switch
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
16/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ11
2.2.7 Chest Stand VD3 35x43 – Generator interface
NOTE: This hardware interface has to be made only if the Potter Bucky ispresent.
Chest stand VD3 35x43 Generator
Description Terminal Wire n° Terminal Description
X-ray request X1.B1 TB1.11 Bucky 3 select
X-ray request X1.B2 TB1.12 Bucky 3 select
Potter reply X1.X TB2.5 Bucky contacts
Potter reply X1.X TB2.4 Bucky contacts
WARNING: The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the ChestStand VD3 35x43 Technical and User's Manual for the Bucky startvoltage setting.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
17/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)12
2.2.8 Generator – Three field solid state AEC device 2
NOTE: This hardware interface has to be made only if the Potter Bucky withAEC device is present.
NOTE:If the AEC interface PWB it is not supplied with LEMO connectors thefollowing hardware interface have to be done. The installer should cut theAEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the nexttable.
GeneratorAEC board
(solid state chamber)type 737992
AEC device 2connecting cable
Description Terminal Wire Description
Anode (L) J12.1 Pink Anode Left
Cath (L) J12.2 Grey Cathode Left
Anode (M) J12.3 White Anode Middle
Cath (M) J12.4 Brown Cathode Middle
Anode (R) J12.5 Green Anode Right
Cath (R) J12.6 Yellow Cathode Right
Ground J12.7 Shield Ground
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
18/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ13
2.2.9 Generator jumpers and wiring configuration instruction
2.2.9.1 Room interface board jumpers configuration
– JW2: Live contact (this jumper configuration must be made only ifthe Potter Bucky is present on Receptor 3)
– JW4: Dry contact
– JW7: Dry contact (2-3)
– JW8: Dry contact.
2.2.9.2 Room interface board jumper wiring
– TB8.1 – TB11.5: (This wiring must be made only if the 24VdcPotter Bucky is present on Receptor 3)
– TB4.8 – TB4.9: Thermal switch.
2.2.9.3 Generator interface PWB jumpers configuration
- JW1: 1 – 2- JW2: 1 – 2 / 3 – 4
- JW3: 1 – 2 / 3 – 4
- JW10: 1 – 2 / 3 – 4
- JW11: 2 – 3
- JW12: Open
- JW13: Open
- JW14: 1 – 2 / 3 – 4
- JW15: 1 – 2 / 3 – 4
- JW21: 2 - 3
NOTE: The configuration of the Jumpers not listed in the previous tables is notrelevant for this system integration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
19/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)14
2.2.9.4 Accessories connection
Generator Accessory
Description Terminal Wire n° Description
Room door interlock(Room Interface board)
TB4.4 Door contact
Room door interlock(Room Interface board)
TB4.5 Door contact
Room light 0V(Room Interface board)
TB7Generator ON lamp
(white lamp)
Room light power supply (Room Interface board)
(*)Generator ON lamp
(white lamp)
Room light contact(Room Interface board)
TB4.11X-ray ON lamp
(red lamp)
Room light contact(Room Interface board)
TB4.12X-ray ON lamp
(red lamp)
(*) TB10 for 220V ac / TB9 for 110V ac / TB8 for 24V dc
NOTE:Remove jumper JW13 on Room Inteface Board.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
20/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ15
2.3 Generator set-up programming instruction
After the hardware connections indicated in the previously tables it isneeded to proceed to the software configuration of the Generator.Refer to the Generator Service Manual, chapter 3, where are describedthe interfacing and the programming information to the generator. Thischapter also allows the installer to record the necessary information tocomplete the installation, as well as to record the programming values.Refer to the tables in paragraphs 2.3.2, 2.3.3, 2.3.4, 2.3.5, 2.3.6, 2.3.7 ofthis Installation Guide for the programming of the generator set-upsubmenu.
2.3.1 Tube selection
Prior to beginning tube calibration, the tube used in the installation mustbe properly selected, and generator limits should be programmed.
2.3.1.1 X-ray tube present in library
If the x-ray tube used in the installation is present in the library of the
generator refer to the Generator Service Manual, chapter 3C -Programming the generator in order to record all the necessaryinformation to program the generator set-up.
The configuration of the starter board it is requested too. Refer to theGenerator Service Manual, chapter 2 - Programming the low speedstarter-Programming the dual speed starter to record the necessaryinformation to complete the configuration.
NOTE: The tube library, concerning the I.A.E. X-ray tubes, contains only highspeed tube types. When a low speed I.A.E. X-ray tube has to be installedan high speed tube must be selected informing the generator about thelow speed tube. The generator and the tube limits must be set inaccording with the tube type installed.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
21/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)16
2.3.1.2 Loading X-ray tube data from GenWare® utility software
If the X-ray tube used in the installation is not present in the library ofthe generator should be present in the utility software. The datacommunication with a computer it is allowed using the GenWare® utilitysoftware. Refer to the Generator Service Manual, chapter 3C - DATA LINKto record the necessary information.
2.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in thegenerator library nor in the GenWare® utility software a manual settingof the tube and generator parameters is required. Before changing theseparameters is strictly recommended to consult the data sheet of the x-raytube that is going to be installed. An X-ray tube present in the library orin the GenWare® utility software, as similar as possible to the one isgoing to be installed, can be selected. The tube and the generatorparameter required in the generator set-up must be set in according withthe information recorded from the data sheet of the x-ray tube. Refer tothe Generator Service Manual, chapter 3C - Tube selection andGenerator limits in order to record all the necessary information to
program the generator set-up. The configuration of the starter board it is requested too. Refer to theGenerator Service Manual, chapter 2 - Programming the low speedstarter - Programming the dual speed starter to record the necessaryinformation to complete the configuration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
22/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ17
2.3.2 AEC set-up
The installation of the AEC devices and their calibration are covered inthe Generator Service Manual, chapter 3E - Automatic exposure control(AEC) calibration.
2.3.3 Fluoro set-up
NOTE:
The Fluoro set-up data must be set in according with the X-ray tubetechnical specification.
The following tables meets the Fluoro set-up request in according withmost of the X-ray tube supplied by Villa Sistemi Medicali; during theinstallation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in theGenerator Service Manual, chapter 3 - Automatic brightness stabilisation(ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 1.00 VABS LEVEL 1: 1.60 V
ABS LEVEL 2: 1.10 V
ABS LEVEL 3: 1.60 V
SIGNAL GAIN 10
ABS GAIN 20
ABS DEADBAND 3%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
23/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)18
FL ABS CURVE FL ABS CURVE 1: ON
40KV: 0.5 mA50 KV: 1.0 mA
60 KV: 1.6 mA
70 KV: 2.1 mA
80 KV: 2.7 mA
90 KV: 3.2 mA
100 KV: 3.8 mA
110 KV: 4.4 mA
120 KV: 5.0 mA
125 KV: 5.0mA
FL DOSE LIMIT * FL MAN DL *
40KV: 0.5 mA
50 KV: 1.0 mA60 KV: 1.6 mA
70 KV: 2.1 mA
80 KV: 2.7 mA
90 KV: 3.2 mA
100 KV: 3.8 mA
110 KV: 4.4 mA
120KV: 5.0 mA
125 KV: 5.0 mA
* FL ABS DL *
40KV: 5.0 mA
50 KV: 5.0 mA 60 KV: 5.0 mA
70 KV: 5.0 mA
80 KV: 5.0 mA
90 KV: 5.0 mA
100 KV: 5.0 mA
110 KV: 5.0 mA
120KV: 5.0 mA
Note: The fluoro dose limit is set at 5 mA in order tomeet the code of federal regulations 21 paragraph1020. If it is not required to meet these regulationsa higher value can be set in according with theloading factors of the x-ray tube.
125 KV: 5.0 mA
FL II COMP Not UsedHLF SETUP Not Used
PF SETUP PF: OFF
NOTE: The Pulsed Fluoro set-up is not relevant for this system integrationbecause the function is not available with the N23 CCD camera.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
24/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ19
2.3.4 Generator Image receptor programming worksheet
IMAGE RECEPTOR PROGRAMMING WORKSHEET
DATE:
SERIAL#:
FUNCTION RECEPTOR1
RECEPTOR2
RECEPTOR3
RECEPTOR4
RECEPTOR5
RECEPTOR6
RECEPTORNAME
TUBE 1 1 1 NONE NONE NONE
TOMO YES NO NO NO NO NO
FLUORO YES NO NO NO NO NO
SERIAL YES NO NO NO NO NO
INTERFACEOPTS
0 0 0 0 0 0
FUNCTIONALOPTS
0 0 0 0 0 0
RECEPTORSYM
FLUOROHANG
30 SEC 30 SEC 30 SEC 30 SEC 30 SEC 30 SEC
RAD HANG 0 SEC 0 SEC 0 SEC 0 SEC 0 SEC 0 SEC
LAST IMAGEHOLD
40 Ms 40 Ms 40 Ms 40 Ms 40 Ms 40 Ms
MEMORY NO NO NO NO NO NO
REM TOMOBUT
2500 Ms 2500 Ms 2500 Ms 2500 Ms 2500 Ms 2500 Ms
SF/LFSWITCH
AUTO AUTO AUTO AUTO AUTO AUTO
AEC BACKUP FIXED FIXED FIXED FIXED FIXED FIXED
AEC BACKUPMAS
500 500 500 500 500 500
AEC BACKUPMS
3200 3200 3200 3200 3200 3200
AECCHANNEL
1 0 2 0 0 0
AEC LOCKEXPS
3 3 3 3 3 3
NOTE:Receptor 3 has to be programmed only if Chest Stand with Potter Buckyand AEC device is present.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
25/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)20
2.3.5 Generator Receptor 1 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- ON -----
REM. TOMO SEL ON ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF ON ON OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF ON ON
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE ON ON ON ON ON
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
26/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /G100R/F-A / Chest Stand
(Rev. 0) APOLLO EZ21
2.3.6 Generator Receptor 2 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF ON ON OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF ON OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
27/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand
APOLLO EZ (Rev. 0)22
2.3.7 Generator Receptor 3 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF ON OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE: The configuration of the Receptor 3 I/O has to be made only if the ChestStand with Potter Bucky is present.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
28/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ23
3 CONFIGURATION WITH APOLLO EZ/CCD N33 / G100R/F-A / CHEST STAND
Room configuration
• Apollo EZ
– I.A.E. X-ray tube type
– Three field solid state AEC
–
I.I. H.R. – CCD N33.• G 100 R/F-A (T.D.I. Interfaced)
• Chest Stand VD3 35x43 with motorised foot rest (See Note )
– Three field solid state AEC.
Receptor configuration
Receptor 1: Apollo EZ Spot Film Device – Tube 1
Receptor 2: Direct X-ray from console – Tube 1
Receptor 3: Chest Stand VD3 35x43 – Tube 1 (See Note )
NOTE:When the Generator is supplied with "Falling load" feature, the GalKeyU29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the"SERIAL" exposure is enable (paragraph 3.3.4 - Image ReceptorProgramming Worksheet: "SERIAL=YES"); whenever the Falling Load hasto be activated on the Spot Film Device receptor, the SERIAL parameterhas to be set to "NO".
In case Falling Load feature and SERIAL exposure are required will benecessary to create for each a dedicated receptor.
NOTE:Reference to the Chest Stand VD3 35x43 with motorized foot rest is madein order to allows the installer to record the necessary information for thehardware interface among G 100 R/F-A and Potter Bucky.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
29/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)24
3.1 Power supplies connection instruction
( * ) If this remote controlled switch is not available use a remote controlledswitch with the following characteristics:
–Coil: 220-230 Vac 50 Hz
230-240 Vac 60 Hz
– Contact: 4 N.O. contacts, 16 A.
(**) Made this connection only when a motorised foot rest chest stand is present.
MainPower Supply
GeneratorTerminals
R
S
T
N
R
S
T
N
R S T N
CHEST STANDTerminal X1
R S T
APOLLO EZTerminals R-S-T (**)
G 100 R/FROOM INTERFACE BOARD TB7.1 (0 Vac)
G 100 R/FROOM INTERFACE BOARD TB10.1 (220 Vac)
* INSTALLATION KIT230Vac AUXILIARY CONTACTORV.S.M. cod.4291515200
(The auxiliary contactor is already mountedinside the G 100 R/F-A Generator)
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
30/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ25
3.2 Hardware interface instruction
3.2.1 APOLLO EZ – G 100 R/F-A interface
APOLLO EZ G 100 R/F-A
Description Terminal Wire n° Terminal Description
X27.15 TB6.9Remote Exposure
(Room Interface Board)
X27.3 TB6.7Preparation radiography(Room Interface Board)
Common0Vdc
X0.20 TB6.11Spare Output
(Room Interface Board)
X-ray ONfrom
Generator
X0.13 J6.21X-ray ON
(Table Digital Interface board)
Generator
ready X0.12 J6.22
Generator ready
(Table Digital Interface board)
Spot FilmDevice
selection
X0.15 TB6.12Spare Output
(Room Interface Board)
Safety switch X0.9 J17.1Safety switch
(Generator Interface PWB)
Safety switch X0.11 J17.2Safety switch
(Generator Interface PWB)
AlarmX-ray ON X27.7(see Note) J7.6 X-ray OFF(Table Digital Interface board)
FluoroMan/Auto
X27.20 J6.1Fluoro Man/Auto
(Table Digital Interface board)
NOTE:During the auto calibration of the X-ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).It is important to remove the jumper at the end of the X-ray tubeauto calibration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
31/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)26
APOLLO EZ G 100 R/F-A
Description Terminal Wire n° Terminal Description
Vref. Anal.Fluoro
X27.11 J6.7Analogue Vref Fluoro Manual kV
(Table Digital Interface board)
TomoSpeed 1
X27.4
TomoSpeed 2
X27.1
TomoSpeed 3
X27.14
J6.3 Tomo Speed 1
(Table Digital Interface board)
TomoSpeed 4
X27.17 J6.4 Tomo Speed 2
(Table Digital Interface board)
Tomo 7° X27.5 J7.1 Tomo Angle 7°
(Table Digital Interface board)
Tomo 20° X27.18 J7.3 Tomo Angle 20°
(Table Digital Interface board)
Tomo 30° X27.6 J7.4 Tomo Angle 30°(Table Digital Interface board)
Tomo 45° X27.19 J7.5 Tomo Angle 45°
(Table Digital Interface board)
Common 0V X27.22 J6.2Signal Ground
(Table Digital Interface board)
Fluoro MankV analog 0V
X27.23 J6.23Signal Ground
(Table Digital Interface board)
J7.2 Signal Ground(Table Digital Interface board)
J7.7Chassis Ground
(Table Digital Interface board)
J7.20Chassis Ground
(Table Digital Interface board)
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
32/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ27
3.2.2 G 100 R/F-A – I.I. component plate interface
G 100 R/F-A I.I. component plate
Description Terminal Wire n° Pcb A2 Description
Remote Fluoro Exposure
(Room Interface board)TB6.5 X2.3 Fluoro
L.I.H.(Table Digital Interface
board)
J6.17 X2.4 Last Image Hold
Proportional dc
(Table Digital Interfaceboard)
J7.21 X2.1 Analog kV auto
0Vdc(Table Digital Interface
board)
J7.2 X2.2 0Vdc
SYNC(Table Digital Interface
board)
J7.24 X4.5 SYNC
P.F. request
(Table Digital Interfaceboard)
J6.22 X4.6 Pulsed Fluoro
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
33/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)28
3.2.3 APOLLO EZ – I.I. component plate interface
APOLLO EZ I.I. component plate
Pcb A1
Description Terminal Wire n°4 fields 3 fields
Description
Common X28.1 X4.4 X4.4 Common
Medium field X28.2 X4.2 X4.1 Medium zoom
Minimum field X28.14 X4.3 X4.2
Maximum
zoom
Maximum field X28.15 X4.1 N.C. Minimum
zoom
Power supply N X0.2 Power supply
Power supply S3 X0.1 Power supply
Ground X0.3 Ground
APOLLO EZ I.I. component plate
Description Terminal Wire n° Pcb A2 Description
X2.15 R.O.I.R.O.I.
X28.8(see Note) X2.11 Recursive Filter X0
Fluoro call X28.18 X2.5 Fluoro call
Vertical imageinversion
X28.16 X2.8Vertical image
inversion
Horizontal imageinversion
X28.4 X2.9Horizontal image
inversion
Recursive Filter X4 X28.21(see Note)
X2.12 Recursive Filter X4
Recursive Filter X8 X28.9
(see Note) X2.13 Recursive Filter X8
Common 0V X28.3 X4.10 Common 0V
Common 0V X28.22
NOTE:
Depending by selection of the Apollo EZ set-up, cell 246 – X28.8 outputcan drive R.O.I. or Recursive Filter. In the first case X28.21 and X28.9are disabled, in the second case are enabled.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
34/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ29
3.2.4 Monitor interface
I.I. component plate Table and wall version
Description Pcb A2 Wire n° Plug Description
Monitor L.I.H.
visualization X4.2 White L I
Monitor L.I.H.
visualization
0 Vs common X4.10 Green 0 V 0 Vs common
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
35/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)30
3.2.5 G 100 R/F-A – Three field solid state AEC device 1
NOTE:If the AEC interface PWB it is not supplied with LEMO connectors thefollowing hardware interface have to be done. The installer should cut theAEC cable by the PWB side connector and connect the colored wires tothe terminals on the Universal AEC board making reference to the nexttable.
G100 R/F-A
AEC board(solid state chamber)
type 737992
AEC device 1connecting cable
Description Terminal Wire Description
Anode (L) J11.1 Pink Anode Left
Cath (L) J11.2 Grey Cathode Left
Anode (M) J11.3 White Anode Middle
Cath (M) J11.4 Brown Cathode Middle
Anode (R) J11.5 Green Anode Right
Cath (R) J11.6 Yellow Cathode Right
Ground J11.7 Shield Ground
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
36/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ31
3.2.6 G 100 R/F-A - X-ray tube interface
G 100 R/F-A X-ray tube
Description Terminal Wire n° Terminal Description
Main Tube 1 P Principal
Shift Tube 1 A Auxiliary
Common Tube 1 C Common
Ground Tube 1 Ground
Thermal switch Tube 1 Thermal switch
Thermal switch Tube 1 Thermal switch
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
37/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)32
3.2.7 Chest Stand VD3 35x43 – G 100 R/F-A interface
NOTE: This hardware interface has to be made only if the Potter Bucky ispresent.
Chest stand VD3 35x43 G 100 R/F-A
Description Terminal Wire n° Terminal Description
X-ray request X1.B1 TB1.11 Bucky 3 select
X-ray request X1.B2 TB1.12 Bucky 3 select
Potter reply X1.X TB2.5 Bucky contacts
Potter reply X1.X TB2.4 Bucky contacts
WARNING: The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the ChestStand VD3 35x43 Technical and User's Manual for the Bucky startvoltage setting.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
38/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ33
3.2.8 G 100 R/F-A – Three field solid state AEC device 2
NOTE: This hardware interface has to be made only if the Potter Bucky withAEC device is present.
NOTE:If the AEC interface PWB it is not supplied with LEMO connectors thefollowing hardware interface have to be done. The installer should cut theAEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the nexttable.
G100 R/F-AAEC board
(solid state chamber)type 737992
AEC device 2connecting cable
Description Terminal Wire Description
Anode (L) J12.1 Pink Anode Left
Cath (L) J12.2 Grey Cathode Left
Anode (M) J12.3 White Anode Middle
Cath (M) J12.4 Brown Cathode Middle
Anode (R) J12.5 Green Anode Right
Cath (R) J12.6 Yellow Cathode Right
Ground J12.7 Shield Ground
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
39/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)34
3.2.9 G 100 R/F-A jumpers and wiring configuration instruction
3.2.9.1 Room interface board jumpers configuration
– JW2: Live contact (this jumper configuration must be made only ifthe Potter Bucky is present on Receptor 3)
– JW3: Live contact
– JW4: Live contact
– JW8: Dry contact (Spot Film Device selection).
3.2.9.2 Room interface board jumper wiring
– TB8.1 – TB11.5 (This wiring must be made only if the 24VdcPotter Bucky is present on Receptor 3)
– TB4.8 – TB4.9: Thermal switch.
3.2.9.3 Generator interface PWB jumpers configuration
- JW1: 1 – 2 - JW13: Open- JW2: 1 – 2 / 3 – 4 - JW14: 1 –2 / 3 – 4
- JW3: 1 – 2 / 3 – 4 - JW15: 1 – 2 / 3 – 4
- JW4: Open - JW19: 2 – 3
- JW5: Close - JW20: Close
- JW6: 1 – 2 / 3 – 4 - JW21: 1 – 2
- JW7: 1 – 2 / 3 – 4 - JW22: 2 – 3
- JW8: 1 – 2 / 3 – 4 - JW24: Close
- JW9: 1 – 2 / 3 – 4 - JW25: Close
- JW10: 1 – 2 / 3 – 4 - JW26: Close
- JW11: 2 – 3 - JW27: Close- JW12: Open - JW28: Open
NOTE: The configuration of the Jumpers not listed in the previous tables is notrelevant for this system integration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
40/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ35
3.2.9.4 T.D.I. board SW1 Dip-switches setting
- SW1-1: OFF - SW1-5: ON
- SW1-2: OFF - SW1-6: OFF
- SW1-3: ON - SW1-7: OFF
- SW1-4: OFF - SW1-8: OFF
3.2.9.5 Accessories connection
Generator Accessory
Description Terminal Wire n° Description
Room door interlock(Room Interface board)
TB4.4 Door contact
Room door interlock(Room Interface board)
TB4.5 Door contact
Room light 0V
(Room Interface board)TB7
Generator ON lamp
(white lamp)
Room light power supply(Room Interface board)
(*)Generator ON lamp
(white lamp)
Room light contact(Room Interface board)
TB4.11X-ray ON lamp
(red lamp)
Room light contact(Room Interface board)
TB4.12X-ray ON lamp
(red lamp)
(*) TB10 for 220V ac / TB9 for 110V ac / TB8 for 24V dc
NOTE:Remove jumper JW13 on Room Inteface Board.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
41/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)36
3.3 G 100 R/F-A set-up programming instruction
After the hardware connections indicated in the previously tables it isneeded to proceed to the software configuration of the G 100 R/F-A.Refer to the G 100 R/F-A Service Manual, chapter 3, where are describedthe interfacing and the programming information to the generator. Thischapter also allows the installer to record the necessary information tocomplete the installation, as well as to record the programming values.Refer to the tables in paragraphs 3.3.2, 3.3.3, 3.3.4, 3.3.5, 3.3.6, 3.3.7 ofthis Installation Guide for the programming of the generator set-upsubmenu.
3.3.1 Tube selection
Prior to beginning tube calibration, the tube used in the installation mustbe properly selected, and generator limits should be programmed.
3.3.1.1 X-ray tube present in library
If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C -Programming the generator in order to record all the necessaryinformation to program the generator set-up.
The configuration of the starter board it is requested too. Refer to theG 100 R/F-A Service Manual, chapter 2 - Programming the low speedstarter-Programming the dual speed starter to record the necessaryinformation to complete the configuration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
42/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ37
3.3.1.2 Loading X-ray tube data from GenWare® utility software
If the X-ray tube used in the installation is not present in the library ofthe generator should be present in the utility software. The datacommunication with a computer it is allowed using the GenWare® utilitysoftware. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATALINK to record the necessary information.
3.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in thegenerator library nor in the GenWare® utility software a manual settingof the tube and generator parameters is required. Before changing theseparameters is strictly recommended to consult the data sheet of the x-raytube that is going to be installed. An X-ray tube present in the library orin the GenWare® utility software, as similar as possible to the one isgoing to be installed, can be selected. The tube and the generatorparameter required in the generator set-up must be set in according withthe information recorded from the data sheet of the x-ray tube. Refer tothe G 100 R/F-A Service Manual, chapter 3C - Tube selection andGenerator limits in order to record all the necessary information to
program the generator set-up. The configuration of the starter board it is requested too. Refer to theG 100 R/F-A Service Manual, chapter 2 - Programming the low speedstarter - Programming the dual speed starter to record the necessaryinformation to complete the configuration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
43/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)38
3.3.2 AEC set-up
The installation of the AEC devices and their calibration are covered inthe G 100 R/F-A Service Manual, chapter 3E - Automatic exposurecontrol (AEC) calibration.
3.3.3 Fluoro set-up
NOTE:
The Fluoro set-up data must be set in according with the X-ray tubetechnical specification.
The following tables meets the Fluoro set-up request in according withmost of the X-ray tube supplied by Villa Sistemi Medicali; during theinstallation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in theG 100 R/F-A Service Manual, chapter 3 - Automatic brightnessstabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 1.00 VABS LEVEL 1: 4.00 V
ABS LEVEL 2: 1.10 V
ABS LEVEL 3: 1.60 V
SIGNAL GAIN 3
ABS GAIN 10
ABS DEADBAND 30%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
44/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ39
FL ABS CURVE FL ABS CURVE 1: ON
40KV: 0.5 mA50 KV: 1.0 mA
60 KV: 1.6 mA
70 KV: 2.1 mA
80 KV: 2.7 mA
90 KV: 3.2 mA
100 KV: 3.8 mA
110 KV: 4.4 mA
120 KV: 5.0 mA
125 KV: 5.0mA
FL DOSE LIMIT * FL MAN DL *
40KV: 0.5 mA
50 KV: 1.0 mA60 KV: 1.6 mA
70 KV: 2.1 mA
80 KV: 2.7 mA
90 KV: 3.2 mA
100 KV: 3.8 mA
110 KV: 4.4 mA
120KV: 5.0 mA
125 KV: 5.0mA
* FL ABS DL *
40KV: 5.0 mA
50 KV: 5.0 mA
60 KV: 5.0 mA
70 KV: 5.0 mA
80 KV: 5.0 mA
90 KV: 5.0 mA
100 KV: 5.0 mA
110 KV: 5.0 mA
120KV: 5.0 mA
Note: The fluoro dose limit is set at 5 mA in order tomeet the code of federal regulations 21 paragraph1020. If it is not required to meet these regulationsa higher value can be set in according with theloading factors of the x-ray tube.
125 KV: 5.0 mA
FL II COMP Not UsedHLF SETUP Not Enable
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
45/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)40
PF SETUP PF: ON
PF DEFAULT: NONEPPS DEFAULT: NONE
PF DEFAULT MA: 5=40mA
6=60mA
7=80mA
8=99mA
Note: Four options for the settings of the Pulsed Fluoro default mA are possible. It issuggested to set the value 5 = 40mA as default parameter. The others settings are
possible on line by means the +/- mA keys. If, once set at 40 mA the default mA value for the pulsed fluoro, is necessary to reduce the dose is possible to reduce the lengthof the pulses by means the settings of the jumpers on the A2 board of the I.I. system(make reference to the I.I. system service manual)
PF DEFAULT Ms: 2
PF SYNC FREQ: 50 HZ
PF ABS SETUP ABS LEVEL 0: 4.0 V
ABS LEVEL 1: 1.10 V
ABS LEVEL 2: 1.20 V
ABS LEVEL 3: 1.90 V
SIGNAL GAIN: 3
ABS GAIN: 5
ABS DEADBAND: 40%
ABS DELAY: 100 mS
ABS START KV: 75
PF INTEGRATOR: OFF
1PPS COMP: 0%3PPS COMP: 0%
6/7.5PPS COMP: 0%
12/15PPS COMP: 0%
25/30PPS COMP: 0%
PF ABS CURVE DEF: OFF
PF ABS CURVES Not Used
PF DOSE LIMITS Not Used
PF II COMP Not Used
PF PPS SETUP 1 PPS: ON
3 PPS: ON
6/7.5 PPS: ON
12.5/15 PPS: ON28/30 PPS: ON
HPF SETUP Not Enable
HLF SELECT: CONSOLE
PF SELECT: CONSOLE
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
46/74
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
47/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)42
3.3.5 G 100 R/F-A Receptor 1 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE ON ON ON ON ON
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
48/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ43
3.3.6 G 100 R/F-A Receptor 2 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON OFF
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
49/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)44
3.3.7 G 100 R/F-A Receptor 3 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF ON OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE: The configuration of the Receptor 3 I/O has to be made only if the ChestStand with Potter Bucky is present.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
50/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ45
4 CONFIGURATION WITH APOLLO EZ/DIVA-D / G100R/F-A / CHEST STAND VD3
Room configuration
• Apollo EZ
– I.A.E. X-ray tube type
– Three field solid state AEC
–
Diva-D.• G 100 R/F-A (T.D.I. Interfaced)
• Chest Stand VD3 35x43 with motorised foot rest (See Note )
– Three field solid state AEC.
Receptor configuration
Receptor 1: Apollo EZ Spot Film Device – Tube 1
Receptor 2: Digital Video Acquisition System – Tube 1
Receptor 3: Chest Stand VD3 35x43 – Tube 1 (See Note )
NOTE:When the Generator is supplied with "Falling load" feature, the GalKeyU29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the"SERIAL" exposure is enable (paragraph 4.3.4 - Image ReceptorProgramming Worksheet: "SERIAL=YES"); whenever the Falling Load hasto be activated on the Spot Film Device receptor, the SERIAL parameterhas to be set to "NO".
In case Falling Load feature and SERIAL exposure are required will benecessary to create for each a dedicated receptor.
NOTE:Reference to the Chest Stand VD3 35x43 with motorized foot rest is madein order to allows the installer to record the necessary information for thehardware interface among G 100 R/F-A and Potter Bucky.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
51/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)46
4.1 Power supplies connection instruction
( * ) If this remote controlled switch is not available use a remote controlledswitch with the following characteristics:
–Coil: 220-230 Vac 50 Hz
230-240 Vac 60 Hz
– Contact: 4 N.O. contacts, 16 A.
(**) Made this connection only when a motorised foot rest chest stand is present.
MainPower Supply
GeneratorTerminals
R
S
T
N
R
S
T
N
R S T N
CHEST STANDTerminal X1
R S T
APOLLO EZTerminals R-S-T (**)
G 100 R/FROOM INTERFACE BOARD TB7.1 (0 Vac)
G 100 R/FROOM INTERFACE BOARD TB10.1 (220 Vac)
* INSTALLATION KIT230Vac AUXILIARY CONTACTORV.S.M. cod.4291515200
(The auxiliary contactor is already mountedinside the G 100 R/F-A Generator)
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
52/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ47
4.2 Hardware interface instruction
4.2.1 APOLLO EZ – G 100 R/F-A interface
APOLLO EZ G 100 R/F-A
Description Terminal Wire n° Terminal Description
X27.15 TB6.9Remote Exposure
(Room Interface Board)
X27.3 TB6.7Preparation radiography(Room Interface Board)
Common0Vdc
X0.20 J6.20V
(Table Digital Interface board)
Common 0VS X27.23 J6.230V analog
(Table Digital Interface board)
Manual kV
ref.
X27.11 J6.7Manual kV fluoro input
(Table Digital Interface board)
Fluoro kVmanual/auto
selection
X27.20 J6.1Man/Auto fluoro
(Table Digital Interface board)
Tomospeed 1
X27.4
Tomo
speed 2 X27.1
Tomo
speed 3 X27.14
J6.3 Tomo speed 1
(Table Digital Interface board)
Tomospeed 4
X27.17 J6.4 Tomo speed 2
(Table Digital Interface board)
Tomo 7° X27.5 J7.1 Tomo 7°
(Table Digital Interface board)
Tomo 20° X27.18 J7.3 Tomo 20°
(Table Digital Interface board)
Tomo 30° X27.6 J7.4
Tomo 30°
(Table Digital Interface board)
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
53/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)48
APOLLO EZ G 100 R/F-A
Description Terminal Wire n° Terminal Description
Tomo 45° X27.19 J7.5 Tomo 45°
(Table Digital Interface board)
AlarmX-ray ON
X27.7
(see Note)J7.6
X-ray OFF(Table Digital Interface board)
SFD selection X0.15 TB2.12Bucky 1 select
(Room Interface Board)
DSI selection X0.14 TB2.1 Bucky 2 select(Room Interface Board)
X-ray ONfrom
generator
X0.13 J6.21X-Ray on
(Table Digital Interface board)
Generatorready
X0.12 J6.22Generator Ready
(Table Digital Interface board)
Safety switch X0.9 J17.1Safety switch
(Generator Interface PWB)
Safety switch X0.11 J17.2Safety switch
(Generator Interface PWB)
NOTE:During the auto calibration of the X-ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).It is important to remove the jumper at the end of the X-ray tubeauto calibration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
54/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ49
4.2.2 APOLLO EZ – DIVA-D interface
APOLLO EZ DIVA-D
Description Terminal Wire n° POD pcb Description
X28.18 20 Fluoroscopy request
X28.6 15 Generator ready
D.S.I.
X28.19 21 Exposure request
Common0Vdc
X28.22 11 0 V digital common
Common X28.3 12 0 V digital common
X28.4 18 Vertical image revers
X28.16 19 Horizontal image revers
X-ray ONfrom
generator
X0.13 14 X-ray ON
1313
X-ray command commonFluoroscopy command common
0 V digital common
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
55/74
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
56/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ51
4.2.5 APOLLO EZ – I.I. component plate interface
APOLLO EZ I.I. component plate
Pcb A1
Description Terminal Wire n°4 fields 3 fields
Description
Common X28.1 X4.4 X4.4 Common
Medium field X28.2 X4.2 X4.1 Medium zoom
Minimum field X28.14 X4.3 X4.2
Maximumzoom
Maximum field X28.15 X4.1 N.C. Minimum
zoom
Power supply N X0.2 Power supply
Power supply S3 X0.1 Power supply
Ground X0.3 Ground
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
57/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)52
4.2.6 DIVA-D – I.I. component plate interface
DIVA-D I.I. component plate
Description POD Pcb Wire n° Pcb A1 Description
Analog intensifierfield selection
25 X5.1V Ref. zoom I.I.
(6 - 12 V)
0 V analog 29 X5.2 0 V analog
4.2.7 Monitor interface
I.I. component plate Table and wall version
Description Pcb A2 Wire n° Plug Description
Monitor L.I.H.
visualization X4.2 L I
Monitor L.I.H.
visualization
0 Vs common X4.10 0 V 0 Vs common
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
58/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ53
4.2.8 G 100 R/F-A – Three field solid state AEC device 1
NOTE:If the AEC interface PWB it is not supplied with LEMO connectors thefollowing hardware interface have to be done. The installer should cut theAEC cable by the PWB side connector and connect the colored wires tothe terminals on the Universal AEC board making reference to the nexttable.
G100 R/F-A
AEC board(solid state chamber)
type 737992
AEC device 1connecting cable
Description Terminal Wire Description
Anode (L) J11.1 Pink Anode Left
Cath (L) J11.2 Grey Cathode Left
Anode (M) J11.3 White Anode Middle
Cath (M) J11.4 Brown Cathode Middle
Anode (R) J11.5 Green Anode Right
Cath (R) J11.6 Yellow Cathode Right
Ground J11.7 Shield Ground
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
59/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)54
4.2.9 G 100 R/F-A - X-ray tube interface
G 100 R/F-A X-ray tube
Description Terminal Wire n° Terminal Description
Main Tube 1 P Principal
Shift Tube 1 A Auxiliary
Common Tube 1 C Common
Ground Tube 1 Ground
Thermal switch Tube 1 Thermal switch
Thermal switch Tube 1 Thermal switch
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
60/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ55
4.2.10 Chest Stand VD3 35x43 – G 100 R/F-A interface
NOTE: This hardware interface has to be made only if the Potter Bucky ispresent.
Chest stand VD3 35x43 G 100 R/F-A
Description Terminal Wire n° Terminal Description
X-ray request X1.B1 TB1.11 Bucky 3 select
X-ray request X1.B2 TB1.12 Bucky 3 select
Potter reply X1.X TB2.5 Bucky contacts
Potter reply X1.X TB2.4 Bucky contacts
WARNING: The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the ChestStand VD3 35x43 Technical and User's Manual for the Bucky startvoltage setting.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
61/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)56
4.2.11 G 100 R/F-A – Three field solid state AEC device 2
NOTE: This hardware interface has to be made only if the Potter Bucky withAEC device is present.
NOTE:If the AEC interface PWB it is not supplied with LEMO connectors thefollowing hardware interface have to be done. The installer should cut theAEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the nexttable.
G100 R/F-AAEC board
(solid state chamber)type 737992
AEC device 2connecting cable
Description Terminal Wire Description
Anode (L) J12.1 Pink Anode Left
Cath (L) J12.2 Grey Cathode Left
Anode (M) J12.3 White Anode Middle
Cath (M) J12.4 Brown Cathode Middle
Anode (R) J12.5 Green Anode Right
Cath (R) J12.6 Yellow Cathode Right
Ground J12.7 Shield Ground
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
62/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ57
4.2.12 G 100 R/F-A jumpers and wiring configuration instruction
4.2.12.1 Room interface board jumpers configuration
– JW2: Live contact (this jumper configuration must be made only ifthe Potter Bucky is present on Receptor 3)
– JW3: Live contact
– JW4: Live contact.
4.2.12.2 Room interface board jumper wiring
– TB8.1 – TB11.5 (This wiring must be made only if the 24VdcPotter Bucky is present on Receptor 3)
– TB4.8 – TB4.9: Thermal switch.
4.2.12.3 Generator interface PWB jumpers configuration
- JW1: 1 – 2 - JW13: Open
- JW2: 1 – 2 / 3 – 4 - JW14: 1 –2 / 3 – 4- JW3: 1 – 2 / 3 – 4 - JW15: 1 – 2 / 3 – 4
- JW4: Open - JW19: 2 – 3
- JW5: Close - JW20: Close
- JW6: 1 – 2 / 3 – 4 - JW21: 1 – 2
- JW7: 1 – 2 / 3 – 4 - JW22: 2 – 3
- JW8: 1 – 2 / 3 – 4 - JW24: Close
- JW9: 1 – 2 / 3 – 4 - JW25: Close
- JW10: 1 – 2 / 3 – 4 - JW26: Close
- JW11: 2 – 3 - JW27: Close
- JW12: Open - JW28: Open
NOTE: The configuration of the Jumpers not listed in the previous tables is notrelevant for this system integration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
63/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)58
4.2.12.4 T.D.I. board SW1 Dip-switches setting
- SW1-1: OFF - SW1-5: ON
- SW1-2: OFF - SW1-6: OFF
- SW1-3: ON - SW1-7: OFF
- SW1-4: OFF - SW1-8: OFF
4.2.12.5 Accessories connection
Generator Accessory
Description Terminal Wire n° Description
Room door interlock(Room Interface board)
TB4.4 Door contact
Room door interlock(Room Interface board)
TB4.5 Door contact
Room light 0V
(Room Interface board)TB7
Generator ON lamp
(white lamp)
Room light power supply (Room Interface board)
(*)Generator ON lamp
(white lamp)
Room light contact(Room Interface board)
TB4.11X-ray ON lamp
(red lamp)
Room light contact(Room Interface board)
TB4.12X-ray ON lamp
(red lamp)
(*) TB10 for 220V ac / TB9 for 110V ac / TB8 for 24V dc
NOTE:Remove jumper JW13 on Room Inteface Board.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
64/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ59
4.3 G 100 R/F-A set-up programming instruction
After the hardware connections indicated in the previously tables it isneeded to proceed to the software configuration of the G 100 R/F-A.Refer to the G 100 R/F-A Service Manual, chapter 3, where are describedthe interfacing and the programming information to the generator. Thischapter also allows the installer to record the necessary information tocomplete the installation, as well as to record the programming values.Refer to the tables in paragraphs 4.3.2, 4.3.3, 4.3.4, 4.3.5, 4.3.6, 4.3.7 ofthis Installation Guide for the programming of the generator set-upsubmenu.
4.3.1 Tube selection
Prior to beginning tube calibration, the tube used in the installation mustbe properly selected, and generator limits should be programmed.
4.3.1.1 X-ray tube present in library
If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C -Programming the generator in order to record all the necessaryinformation to program the generator set-up.
The configuration of the starter board it is requested too. Refer to theG 100 R/F-A Service Manual, chapter 2 - Programming the low speedstarter-Programming the dual speed starter to record the necessaryinformation to complete the configuration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
65/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)60
4.3.1.2 Loading X-ray tube data from GenWare® utility software
If the X-ray tube used in the installation is not present in the library ofthe generator should be present in the utility software. The datacommunication with a computer it is allowed using the GenWare® utilitysoftware. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATALINK to record the necessary information.
4.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in thegenerator library nor in the GenWare® utility software a manual settingof the tube and generator parameters is required. Before changing theseparameters is strictly recommended to consult the data sheet of the x-raytube that is going to be installed. An X-ray tube present in the library orin the GenWare® utility software, as similar as possible to the one isgoing to be installed, can be selected. The tube and the generatorparameter required in the generator set-up must be set in according withthe information recorded from the data sheet of the x-ray tube. Refer tothe G 100 R/F-A Service Manual, chapter 3C - Tube selection andGenerator limits in order to record all the necessary information to
program the generator set-up. The configuration of the starter board it is requested too. Refer to theG 100 R/F-A Service Manual, chapter 2 - Programming the low speedstarter - Programming the dual speed starter to record the necessaryinformation to complete the configuration.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
66/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ61
4.3.2 AEC set-up
The installation of the AEC devices and their calibration are covered inthe G 100 R/F-A Service Manual, chapter 3E - Automatic exposurecontrol (AEC) calibration.
4.3.3 Fluoro set-up
NOTE:
The Fluoro set-up data must be set in according with the X-ray tubetechnical specification.
The following tables meets the Fluoro set-up request in according withmost of the X-ray tube supplied by Villa Sistemi Medicali; during theinstallation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in theG 100 R/F-A Service Manual, chapter 3 - Automatic brightnessstabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 5.00 VABS LEVEL 1: 5.00 V
ABS LEVEL 2: 5.00 V
ABS LEVEL 3: 5.00 V
SIGNAL GAIN 10
ABS GAIN 10
ABS DEADBAND 3%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
67/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)62
FL ABS CURVE FL ABS CURVE 1: ON
40KV: 0.5 mA50 KV : 1.0 mA
60 KV : 2.0 mA
70 KV : 3.0 mA
80 KV : 4.0 mA
90 KV : 5.0 mA
100 KV : 6.0 mA
110 KV : 7.0 mA
120 KV: 8.0 mA
125 KV : 8.0mA
FL DOSE LIMIT * FL MAN DL *
40KV: 0.5 mA
50 KV: 1.0 mA60 KV: 1.6 mA
70 KV: 2.1 mA
80 KV: 2.7 mA
90 KV: 3.2 mA
100 KV: 3.8 mA
110 KV: 4.4 mA
120KV: 5.0 mA
125 KV: 5.0mA
* FL ABS DL *
40KV: 5.0 mA
50 KV: 5.0 mA
60 KV: 5.0 mA
70 KV: 5.0 mA
80 KV: 5.0 mA
90 KV: 5.0 mA
100 KV: 5.0 mA
110 KV: 5.0 mA
120KV: 5.0 mA
Note: The fluoro dose limit is set at 5 mA in order tomeet the code of federal regulations 21 paragraph1020. If it is not required to meet theseregulations a higher value can be set in accordingwith the loading factors of the x-ray tube.
125 KV: 5.0 mA
FL II COMP Not UsedHLF SETUP Not Used
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
68/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ63
PF SETUP PF: ON
PF DEFAULT: NONEPPS DEFAULT: NONE
PF DEFAULT MA: 5=40mA
6=60mA
7=80mA
8=99mA
Note: Four options for the settings of the Pulsed Fluoro default mA are possible. It issuggested to set the value 5 = 40mA as default parameter. The others settings are
possible on line by means the +/- mA keys. If, once set at 40 mA the default mA value for the pulsed fluoro, is necessary to reduce the dose is possible to reduce the lengthof the pulses by means the settings of the jumpers on the A2 board of the I.I. system(make reference to the I.I. system service manual)
PF DEFAULT Ms: NotUsed
PF SYNC FREQ: 50 HZ
PF ABS SETUP ABS LEVEL 0: 5.0 V
ABS LEVEL 1: 5.0 V
ABS LEVEL 2: 5.0 V
ABS LEVEL 3: 5.0 V
SIGNAL GAIN: 10
ABS GAIN: 8
ABS DEADBAND: 10%
ABS DELAY: 150 mS
ABS START KV: 75
PF INTEGRATOR: OFF1PPS COMP: 0%
3PPS COMP: 0%
6/7.5PPS COMP: 0%
12/15PPS COMP: 0%
25/30PPS COMP: 0%
PF ABS CURVE DEF: NONE
PF ABS CURVES Not Used
PF DOSE LIMITS Not Used
PF II COMP Not Used
PF PPS SETUP 1 PPS: ON
3 PPS: ON
6/7.5 PPS: ON12.5/15 PPS: ON
28/30 PPS: ON
HPF SETUP Not Enable
HLF SELECT: CONSOLE
PF SELECT: CONSOLE
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
69/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)64
4.3.4 G 100 R/F-A Image receptor programming worksheet
IMAGE RECEPTOR PROGRAMMING WORKSHEET
DATE:
SERIAL#:
FUNCTION RECEPTOR1
RECEPTOR2
RECEPTOR3
RECEPTOR4
RECEPTOR5
RECEPTOR6
RECEPTORNAME
TUBE 1 1 1 NONE NONE NONE
TOMO NO NO NO NO NO NO
FLUORO YES YES NO NO NO NO
SERIAL YES YES NO NO NO NO
INTERFACEOPTS
20 21 0 0 0 0
FUNCTIONALOPTS
0 0 0 0 0 0
RECEPTORSYM
FLUOROHANG
30 SEC 30 SEC 30 SEC 30 SEC 30 SEC 30 SEC
RAD HANG 0 SEC 0 SEC 0 SEC 0 SEC 0 SEC 0 SEC
LAST IMAGEHOLD
40 Ms 40 Ms 40 Ms 40 Ms 40 Ms 40 Ms
MEMORY NO NO NO NO NO NO
REM TOMOBUT
2500 Ms 2500 Ms 2500 Ms 2500 Ms 2500 Ms 2500 Ms
SF/LFSWITCH
AUTO AUTO AUTO AUTO AUTO AUTO
AEC BACKUP FIXED FIXED FIXED FIXED FIXED FIXED
AEC BACKUPMAS
500 500 500 500 500 500
AEC BACKUPMS
3200 3200 3200 3200 3200 3200
AECCHANNEL
1 1 2 0 0 0
AEC LOCKEXPS
3 3 3 3 3 3
NOTE:Receptor 3 has to be programmed only if Chest Stand with Potter Buckyand AEC device is present.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
70/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ65
4.3.5 G 100 R/F-A Receptor 1 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT ON ON ON ON ON
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF ON ON
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
71/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)66
4.3.6 G 100 R/F-A Receptor 2 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT ON ON ON ON ON
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF ON OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
72/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /G 100 R/F-A / Chest Stand
(Rev. 0) APOLLO EZ67
4.3.7 G 100 R/F-A Receptor 3 I/O configuration worksheet
FUNCTIONS STANDBY PREP GEN RDY RAD EXP FLUORO EXP
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF ON OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE: The configuration of the Receptor 3 I/O has to be made only if the ChestStand with Potter Bucky is present.
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
73/74
SYSTEM INTEGRATION GUIDEConfiguration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand
APOLLO EZ (Rev. 0)68
THIS PAGE IS INTENTIONALLY LEFT BLANK
-
8/15/2019 6998312603_Rev0 (System Integration Guide)
74/74
Cod. 6998312603_Rev.0