7. temperature measurement

Upload: jonathan-da-costa

Post on 05-Apr-2018

231 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/1/2019 7. Temperature Measurement

    1/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    1

    Temperature measurement

    To most people, temperature is an intuitive concept which tells whether abody is hot or cold. In second law of thermodynamics temperature isrelated to heat, as it is observed in nature that heat flows only from high

    temperature to low temperature. In kinetic theory of gases and statistical thermodynamics it is shown that

    temperature is related to average kinetic energy of the molecules of anideal gas. A further extensions of statistical thermodynamics show therelationship between temperature and energy level in liquids and solids.

    Temperature is a fundamental quantity, similar to mass, length and time.However unlike other quantities direct comparison with standards is notpossible in case of temperature,

    The law that is used in temperature measurement is known as the Zerothlaw of thermodynamics. This states, if two bodies are in thermalequilibrium with a third body, then they are all in thermal equilibrium witheach other. In other words the three bodies have the same temperature.

    It is therefore required to have a third body (a Thermometer) which can beconfirmed to have a thermal equilibrium with other two bodies then the twobodies can be said to have same temperature.

    Since direct comparison of temperature with a standard is not possible it isdifficult to establish a scale of temperature as a standard for comparison

    with the same. A theoretical scale was proposed by Lord Kelvin on thebasis of second law of thermodynamics and concept of an ideal reversibleCarnot cycle. The reversible heat engine working on Carnot cycle takesan amount of heat Q2 from a reservoir of infinite capacity at temperatureT2 and supplies an amount of heat Q1 to another reservoir at temperatureT1 according to relation

    Q2 = T2Q1 T1

    Kelvins thermodynamic scale is absolute in the sense that it isindependent of any material properties. But it is not realisable because ofits dependence on an ideal cycle. Fortunately, it can be shown thattemperature scales based on constant volume or constant pressure gasthermometer using an ideal gas is identical to the thermodynamic scale.

  • 8/1/2019 7. Temperature Measurement

    2/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    2

    International Temperature Scale 1990 (ITS-90)

    It should be noted that Ideal gas is a theoretical concept and in practice the

    Ideal gas thermometers have shortcomings. According to ideal gas therory

    1. Gas molecules are point masses, occupying no space

    2. Collisions between them are elastic

    3. No intermolecular force acts between them.

    Unfortunately no real gas satisfies the above description. For all practicalpurposes, temperature measurement by gas thermometers is cumbersome.

    Therefore an International practical temperature scale (IPTS) is defined at

    which thermal transducers can be calibrated. (in 1927 for the first time)

  • 8/1/2019 7. Temperature Measurement

    3/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    3

    The latest standard ITS-90 defines a number of fixed referencetemperature points as shown in the table. Between these fixed points,elaborate interpolation equation are specified by ITS-90for use withvarious interpolation standards.

    From 0.65 K to 5.0 K standard is based on measurement of vapour

    pressure of helium and use of equation relating vapour pressure withtemperature

    3.0 K to 24.5561 K (Triple point of neon) a constant volume helium gasthermometer is used.

    Over a broad range of 13.8033 K (triple point of hydrogen) to the normalfreezing point of silver (1234.95 K) the standard is defined by means of aplatinum resistance as a function of temperature. Elaborate process ofcalibration at several fixed point temperatures for platinum resistancethermometer to cover various sub ranges is provided by the standard.

    Finally above the melting point of silver, temperatures are determined bymeasurement of the thermal radiation emitted by a black body cavity invacuum and the plank radiation law.

    The radiant energy is typically measured by optical pyrometry. Theunknown temperature is then calibrated by comparing the emission of asource at the unknown temperature to that from a source at the referencetemperature.

  • 8/1/2019 7. Temperature Measurement

    4/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    4

    Temperature scales

    Two temperature scales which are in common use are Celsius andFahrenheit scales. These are based on freezing point and boiling point ofwater at standard atmospheric pressure. The Celsius scale has 100 unitsbetween these points and Fahrenheit scale has 180 points. They are

    related by equation

    The absolute temperature scales based on the thermodynamic idealCarnot cycle has been correlated with Celsius and Fahrenheit scales asfollows

    The zero points on both the absolute scales represent the same physicalstate and the ratio of two values is the same regardless of which absolutescale is used i.e.

    Freezing point and boiling point of water at a pressure of std oneatmosphere (101.3 kN/m2) are taken as 0o and 100o on Celsius scale and32o and 212o on the Fahrenheit scale.

    The relation ship between Rankine and Kelvin scale is as follows.

  • 8/1/2019 7. Temperature Measurement

    5/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    5

    Measurement of Temperature

    Temperature cannot be measured directly but must be measured byobserving the effect that temperature variation causes on themeasuring device. The temperature measurement methods are

    broadly classified as1. Non electrical methods

    2. Electrical methods

    3. Radiation methods

    Non electrical methods are based on one of the following principles

    1. Change in Physical state - Temp. at which a pure substance changetheir physical state are used for calibration of temperature scales butthese devices give one particular value of unique temperature and

    hence are not suitable for measurement of temperatures over a range.2. Change in Chemical properties For using the substances which

    change chemical properties, we have to look for a reversible processin order to have a repeatable/reproducible scale with respect tochange in temp. However such processes are very few in number

    3. Change in physical properties Most non electrical methods arebased on change in physical properties e.g. thermal expansion. Sincethese devices have no electrical connections they can be used inexplosion risk areas such as petrol storage tanks.

    e.g. Bimetallic thermometers, Liquid in glass thermometers, Pressurethermometer mercury in steel, constant volume inert gasthermometer, vapour pressure thermometer.

    Electrical Methods are mainly of two types.

    1. Thermo resistive variable resistance transducers e.g. mettalicresistance, semiconductor (thermistor)

    2. Thermo electric type emf generating type. E.g. thermocouples, solidstate temp. sensors, Quartz thermometer

    Radiation methods (Pyrometry)

    1. Total radiation pyrometer2. Selective radiation pyrometer

  • 8/1/2019 7. Temperature Measurement

    6/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    6

    Bimetallic Thermometer

    This type of thermometer employs the principle of solid expansion andconsist of a bimetal strip usually in the form of a cantilever beam. Thiscomprises strips of two metals, having different coefficients of thermalexpansion, welded or riveted together so that relative motion between

    them is prevented. An increase in temperature causes the deflection ofthe free end of strip.

    When a bimetal strip is formed by firmly bonding two metal strips A & Bhaving different thermal coefficient of expansion a & b. If this bimetalstrip is a straight line at temperature T1, then at elevated temperature T2the strip will form a uniform circular arc of radius of curvature r such that

    Where t is the thickness of the strip

    Different forms of bimetal such as

    cantilever, U-Tube, spiral etc. are

    used

  • 8/1/2019 7. Temperature Measurement

    7/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    7

    The deflection with temperature is nearly linear, depending mainly on coeff. of

    linear expansion. Invar is generally used as the low expansion metal. This is an

    iron / nickel (36%) alloy. Its coefficient of expansion is around 1/20th of ordinary

    metals (nearly zero). Brass is used as high expansion metal for measurement

    of low temperatures, whereas nickel alloys are used when high temperatures

    are to be measured.

    A plain bimetallic strip has less sensitivity, however the sensitivity can be

    improved by using a longer strip in a helical form as shown above. One end of

    helix is anchored to the casing and the other end which is free is connected to

    the pointer which sweeps over a circular dial graduated in degrees of

    temperature. A long stem can be provided between pointer and the free end of

    bimetal element so that, the element enclosed in protective sheath, can be

    submerged in the hot area and dial and pointer can be located at safe distance.

    Bimetallic thermometers are usually employed in the range of -30 to 550oC,

    with inaccuracies of the order of +0.5 to +1% of FSD in highly accurate

    thermometers. In addition to temperature indication, they are also used in

    application for combined sensing and control purpose. Control action is usually

    on-off type (thermostats), where the strip has sufficient force to actuate a control

    switch in electric ovens, irons and refrigerators. These can also be used as

    compensating element for ambient temperature is pressure thermometers,

    aneroid barometers and watches. The bimetallic strip has advantage of being self-generating type instrument

    with low cost, practically no maintenance expenses and stable operation over

    extended period of time. Main disadvantage is its inability to measure rapidly

    changing temperatures due to relatively high thermal inertia.

  • 8/1/2019 7. Temperature Measurement

    8/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    8

    Liquid in glass Thermometers

    The liquid in glass thermometer is one of the most common temperature

    measuring devices. Both liquid and glass expand on heating and their

    differential expansion is used to indicate the temperature. The lower limit for

    mercury is -37.8oC while -130oC for pentane. The higher temperature range is

    340oC (boiling point of mercury is 357oC) This may be extended to 560oC byfilling the space above mercury with CO2 or N2 at high pressure to increase

    the boiling point and the range.

    The precision of these thermometers depend on the care used in

    calibration. The calibration should also be checked regularly to take into

    account aging effects. Accuracy of these thermometers does not exceed

    0.1oC. However when increased accuracy is required Beckmann range

    thermometer can be used. It has big bulb attached to a very fine capillary. The

    range of this thermometer is limited to 5-6oC with an accuracy of 0.005oC

    Liquid in glass thermometers have qualities like low cost, simplicity in useportability and convenient visual indication without use of any externalpower. However their use is limited to laboratory applications.

    It is not preferred in industrial applications because of its fragility and its

    adaptability to remote indication. Also it introduces a time lag in dynamicmeasurements due to thermal inertia of the bulb.

  • 8/1/2019 7. Temperature Measurement

    9/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    9

    Filled System Thermometers (Filled Thermometers)

    Filled system thermometer works on pressure or volume change of a fluid,gas or changes in vapour pressure of a liquid. It consists of four parts

    1. Bulb

    2. Capillary Tube3. Pressure or volume sensitive element

    4. Indicating Device.

    The capillary tube connects the bulb containing a fluid that is sensitive to

    temperature changes to the element that is sensitive to pressure or volume

    changes. The pressure sensitive or volume sensitive element may be a

    Bourdon tube, a helix of bourbon tube, a diaphragm or bellows.

    The gas filled systems are sometimes called gas thermometers or pressure .

    thermometers.

  • 8/1/2019 7. Temperature Measurement

    10/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    10

    Operating principle

    The volume expansion of a liquid is given by the equation

    Where VT and Vo are the volumes of the liquid at temperatures T and 0oC

    respectively. and is the coefficient of volume expansion.

    The bulb which contains the liquid also expands or contracts with the

    variation of temperature. Those these expansion / contraction are much

    smaller than those of the liquids, they introduce some degree on non linearity

    in the relation.

    For the gas filled system, Boyles law is guiding principle. Hence

    In this case the absolute (Kelvin) temperature will show linear relationshipwith the pressure. The non linearity introduced by volume change of bulbwith temperature is negligible.

    Classification of filled system thermometers is made by SAMA(Scientific Apparatus Makers Association, USA. Into four classes as perthe table below.

  • 8/1/2019 7. Temperature Measurement

    11/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    11

    Class I (Liquid filled thermometers Other than mercury)

    The liquids generally used in these thermometers are alcohol, pentane and

    toluene. The minimum temperature that can be measured by such system

    depends on the freezing point of the organic liquid used. This lies usually

    between -200oC and -75oC depending on the liquid used.

    The maximum temperature depends, not only on the boiling points ofliquids but also on the linearity of their volume expansion at higher temperatures.

    For organic liquids usually 300oC is the upper limit.

    Class V (Mercury or Mercury-Thallium eutectic amalgam filled)

    The mercury in steel thermometer has a near linear characteristic. Sufficient

    power is available to operate a recording pen if required. The range of operation

    is from -38oC to 650oC

    Class II Vapour Filled Thermometer (vapour pressure thermometer)

    The system in the vapour pressure thermometer is partly filled with liquid andpartly with vapour of the same liquid, so that there is a liquid vapour interface in

    the bulb. The liquid vapour system does not have any error as long as a free

    liquid surface exists in the sensing bulb, since such system follows Daltons law

    of partial pressure which states that if both liquid and vapour are present, there

    is only one saturation pressure corresponding to the given temperature. The

    general usefulness of this thermometer is restricted due to limited number of

    liquids providing suitable saturation vapour pressure ranges. Commonly used

    liquids are ethane, ethyl alcohol, ethyl chloride, methyl chloride, chlorobenzene.

    toluene, pentane etc. Scale range is of the order of 100oC and accuracy up to

    +1%. The pressure is roughly logarithmic function of temperature

    log P = a b/T

    Thus the scale is non linear.

    Class III Gas Filled Thermometer (Constant volume thermometer)

    This uses an inert gas (usually N2), the principle of working is pressure of the

    gas increases as temperature is increased while operating at constant volume.

    In reality the volumes of bulb, capillary and Bourdon tube increase with pressureand temperature increase.

    Operating range -130oC to 540oC mostly linear and accuracy of the order of +1%

    at lower range and +2% above 300oC

  • 8/1/2019 7. Temperature Measurement

    12/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    12

    In full compensation both Bourdon tube and Capillary are covered. The fullcompensation is necessary if the capillary length is greater than 3 m.

    Despite compensation, to obtain sufficiently accurate data from sucharrangements, the volume of the temperature sensing bulb has to besufficiently larger than the Bourdon tube.

    Another potential cause of error is the change in pressure head which is

    introduced by change in relative levels of the bulb and display. If the bulb is

    raised by height h there is an increase in pressure equal to gh. This error is

    constant for specific setup and may be removed by means of zero adjustment.

    Errors in measurement with filled systems :The filled system thermometers are vulnerable to errors arising out of thedifferential expansion of fluid contained in the bulb and also in the capillarytubeand the Bourdon tube, which are subjected to ambient temperatures. These

    errors are minimised through case compensation and full compensationmethods.

    In case compensation onlythe error arising out of thedifferential expansion of theFluid in the Bourdon tube isCompensated for.

  • 8/1/2019 7. Temperature Measurement

    13/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    13

  • 8/1/2019 7. Temperature Measurement

    14/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    14

    Temperature measurement using changein electrical properties

    Electrical methods in general are preferred for the measurement oftemperature as they furnish a signal which can be easily detected, amplifiedor used for control purposes

    The two main electrical methods are

    Thermo resistive i.e. variable resistance transducer

    Thermo electric i.e. emf generating transducers

    Electric Resistance Thermometers

    In resistance thermometers, the change

    in resistance of various materials, which

    varies in reproducible manner with

    temperature, forms the basis of this

    technique. The material in actual use

    Fall in two classes namely, conductors

    (metals) and semiconductors.

    In general, the resistance of the highly

    conducting materials (metals) increases

    With increase in temperature. Whereas

    The resistance of semiconductor materialsgenerally (not always) decreases with

    increase in temperature. These are called

    negative temperature characteristic (NTC)

    Thermistors.

    The fig shows typical variation of specific resistance of the metals (e.g.

    Platinum) and the NTC Thermistor.

  • 8/1/2019 7. Temperature Measurement

    15/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    15

    Resistance Temperature Detectors (RTDs) Metallic resistance

    thermometers are commonly known as RTD. The relation between electrical

    resistance of a metal RT and the correspondingtemperature T is generally

    given as

    RT = Ro (1 + C1T + C2 T2 + .. + CnT

    n)

    Where the Cs are constants and Ro is the resistance at temperature T =0oC

    Although the above relationship is non linear it can be seen from the fig that

    the curve is nearly linear for Copper and Platinum over a long range.

    The copper being easily susceptible to chemical reactions in process industry

    platinum is preferred for the RTDs. The platinum resistance thermometers are

    also referred as PRT, hence the terms RTD and PRT are used interchangeably.

    Platinum is very widely used sensor and its operating range is from 4 K to

    1064oC. Because it provides extremely reproducible output, it is used in

    establishing International Practical Temperature Scale from 13.8K to 961.93o

    C.RTDs are very suitable for both laboratory and industrial applications because of

    high degree of accuracy as well as long term stability, wide operating range and

    linear characteristics. Limitations are low sensitivity, higher cost compared to

    other sensors and errors due to contact resistance, shock & accelerations.

  • 8/1/2019 7. Temperature Measurement

    16/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    16

    Construction :- Metallic resistance Thermometers (RTDs) are constructed

    in many forms , but the temperature sensitive element is usually in the form

    of a coil or fine wire supported in stress free manner. A typical construction is

    shown below.

    The wire of metal is wound on a grooved ceramic former and is

    covered with protective cement. The ends of the coils are welded to stiff

    copper leads that are taken out, to be connected in a Wheatstone bridgecircuit. Most of the times there is a protective metal sheath above the

    cemented portion to provide rigidity and mechanical strength.

    Alternatively RTD sensors may be fabricated by depositing thin films of

    platinum, nickel or copper on a ceramic substrate. These thin film sensors

    have the advantage of extremely low mass and consequently more rapid

    thermal response.

    Platinum, in spite of its low sensitivity and high cost compared to

    nickel and copper, is the most widely used material for RTDs because of

    following

    1. The temperature resistance characteristics of pure platinum are welldefined and stable over a wide range of temperatures

    2. It has high resistance to chemical attack and contamination.

    3. It has very good accuracy and reproducibility. +0.01oC upto 500oC and+0.1oC up to 1200oC.

    The variation of resistance of sensing element is normally measured using

    some form of electrical bridge circuit which may employ either deflection mode

    or null mode of operations. Lead length and their resistance are sometimescritical hence should be kept to minimum. In addition some modifications may

    be required for lead compensation. It is also essential that thermoelectric emfs

    do not affect the system and hence techniques like ac excitation or manually

    varying dc polarity are used.

  • 8/1/2019 7. Temperature Measurement

    17/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    17

    Cable Compensation arrangements for RTD:- One of the primary sources

    of error in the electrical resistance thermometer is the effect of the resistance

    of the leads which connect the element to the bridge circuit. Some of the

    arrangements that may be used to correct for this effect are shown in the fig.

    above. The Siemens three lead arrangement is the simplest type. At balance

    condition the central lead carries no current and the effect of resistance of the

    other two leads is cancelled out. The Callenders four lead arrangement solves

    the problem by inserting two additional lead wires in the adjustable leg of thebridge so that the effect of the lead wires on the resistance thermometer is

    cancelled out.

    Practical Problems with RTDs: are due to lead error and relatively bulky size

    which is sometimes the reason for poor transient response and conduction error.

    RTD also has a relatively fragile construction. As current has to be fed to the

    RTD for bridge measurement, there is the possibility of self heating (=i2R) which

    may alter the temperature of the element. This is especially prominent in air

    temperature measurement where self heating effect is more prominent.

  • 8/1/2019 7. Temperature Measurement

    18/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    18

    Semiconductor Resistance Sensors (Thermistors)Thermistor (shortform of thermal resistor) is a thermally sensitive variable

    resistor made of ceramic like semi conducting material. They are available in a

    greater variety of shapes and sizes having cold resistance ranging from a few

    ohms to mega ohms. The size range can range from extremely small bead, a

    thin disc, thin chip of wafer to a large sized rod as shown in fig. Most thermistorsunlike metals respond negatively to temperature and their coefficient of

    resistance is of the order of 10 times higher than of platinum and copper.

    If R is the change in resistance corresponding to a temperature change ofT, approximately they are related as

    R=TTWhere T is the temperature coefficient of resistivity. Thermisters can bedivided into two categories depending on whether T is positive or negative.Those with positive temperature coeff. are called PTC and those with negative

    temperature coeff. are called NTC thermistors. Thermistors are tiny or small in

    size and have a small temperature range with a high response and sensitivity,

    hence they are quite different from RTDs.

    Thermistors are fabricated from the semiconducting materials which

    include the oxides of copper, manganese, nickel, cobalt, lithium and titanium.These are blended in a suitable proportion and compressed into desired shape

    from powders and recrystalised them to form a dense ceramic body with

    required characteristics.

  • 8/1/2019 7. Temperature Measurement

    19/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    19

    Positive temperature coefficient thermistors (PTC) : The commercial PTCs

    are of two major categories. The first category is of thermally sensitive silicon

    resistors (also known as silisters). They exhibit fairly uniform positive temp.

    coeff. (about +0.77%/oC) over most of their operating range but may exhibit

    negative coeff. at temperatures above 150 oC. These devices are used for

    temperature compensation of silicon semiconducting devices in the range of

    66 oC to +150 oC

    The other category of PTC are made up of doped polycrystalline ceramic

    Materials. These devices have a resistance-temperature characteristics that

    exhibits a very small negative temp coeff until the device reaches a critical

    temperature Tc, referred as Curie temperature. Below Curie point, it shows a

    low resistance and a small negative temp coeff.. At curie point the resistance

    increases sharply.

    Other group of PTC

    thermistors are made

    with polymers and

    carbon.grains. They are

    called polyswitch, They

    also exhibit similar sharp

    rise in resistance with

    Increase in temperature.

    Thus the PTC type thermistors are mostly used for switching in temperature

    controllers rather than proportional temperature measurement.

    Negative temperature coefficient thermistors (NTC) : Owing to their larger

    resistance change with temperature, NTC devices are usually more suitable

    for precision temperature measurement, even though the relevant

    characteristic curve is non linear.

  • 8/1/2019 7. Temperature Measurement

    20/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    20

    Construction: Oxides of Manganese, nickel or cobalt are milled & mixed in

    proper proportion with binders, pressed into desired shapes and then sintered

    to form thermistors in the form of rods, discs, flakes or beads.

    Lastly, the wire leads are attached, and the combination is coated with

    glass or epoxy. This coating provides mechanical strength and electrical

    Insulation thereby reducing signal noise. The thermistor element and the

    connecting wires along with the glass insulation are put inside a metal sheath

    to form a thermistor probe as shown below.

    By varying the mixture of oxides, it is possible to produce thermistors having

    range of resistances from 30 to 20M (at 25oC). Speed of response is faster

    in case of small thermistors. Beads are made as small as 0.3mm to 2mm.

  • 8/1/2019 7. Temperature Measurement

    21/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    21

    Thermistor resistances are usually specified at room temperature (i..e. 25oC).

    The resistances of commercially available thermistors are usually 2252 ohms,

    5000 ohms and 10000ohms. The standard temperature ranges are -55oC to

    150oC. Longer range -100oC to 400oC are also available.

    No special leads are required to connect thermistors because the deviceoperates at very high resistance compared to leads. Another aspect which has

    to be considered is that the source of error is the temperature rise caused by self

    heating.

    Thermistors have following advantages for temperature measurement,

    1. A large temperature coeff. Which makes thermistor extremely sensitivedevice, thus measuring accuracy up to +0.01oC is achievable.

    2. Ability to withstand electrical and mechanical stresses

    3. Fairly good operating range -100oC to 300oC4. Low cost and easy adaptability to the available resistance bridge circuit.

    5. High sensitivity and availability in extremely small size (size of pin head)enable fast speed of thermal response.

    6. They are very useful for dynamic measurements.

    Disadvantages

    1. Highly non linear resistance-temperature characteristics and

    2. Problems of self heating which necessitate the use of much lower

    current compared to metallic sensors.

  • 8/1/2019 7. Temperature Measurement

    22/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    22

    Thermoelectric Temperature MeasurementThe most common method of measuring and controlling temperature uses

    thermocouple. A thermocouple consists of two electrical conductors that are

    made of dissimilar metals and have at least one electrical connection. This

    electrical connection is referred to as a junction.

    A thermocouple junction can be created by welding, soldering, or by anyother method that provides good electrical contact between the twoconductors, such as twisting the wires around one another.

    The output of a thermocouple circuit is a voltage, and there is a definiterelationship between the voltage and the temperatures of the junctions thatconstitute the circuit.

    Thermoelectricity

    Consider the basic thermocouple circuit shown above. The junction 1 is attemperature T1 and the junction 2 is at temperature T2. If T1 and T2 are notequal, a finite open circuit potential, (emf1), will be measured. Themagnitude of potential difference will depend on the difference in the

    temperatures and the particular metals that form the thermocouplejunctions.

    Thus the open circuit emf in the circuit is the basis for temperaturemeasurement. In the above circuit emf will measure the difference betweenT1 and T2 .

    Thermoelectric phenomena is a result of simultaneous flow of heat and

    electricity in an electrical conductor. In an electrical conductor subject to

    temperature gradient, there will be both flow of thermal energy and electricity.

    Both these phenomena are closely linked with free electrons in metal. The

    behavior of the free electrons in an electrical circuit composed of dissimilar

    metals results in useful relationship between temperature & emf. There are

    three basic phenomena that can occur in the thermocouple circuit.

  • 8/1/2019 7. Temperature Measurement

    23/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    23

    There are three basic phenomena that can occur in thermocouple circuit

    (1) The Seebeck effect

    (2) The Peltier effect

    (3) The Thomson effect.

    The Seebeck Effect : refers to the generation of an emf in an open

    thermocouple circuit caused by a difference in temperature between junctions in

    the circuit. The Seebeck emf can be measured when there is no current flow in

    the circuit. Micro-voltmeter of very high input impedance may be included in the

    circuit to measure the resulting emf. There is a fixed reproducible relationship

    between the emf and the junction temperatures T1 and T2. This relation ship is

    expressed by the Seebeck coefficient, AB defined as

    For any pair of materials, the relative Seebeck emf can be measured and

    tabulated as a function of temperature.

    The Peltier Effect: If a current flows through

    a thermocouple, heat is absorbed at one

    junction and liberated at other. Unlike Jouleheating (i2R) which liberates heat irrespective

    of the direction of current flow, the reversal of

    current direction in thermocouple circuit will

    result into reversal of heat absorption and rejection ends in the circuit.

    2

    1

    T

    TABdT

    12 TTAB

  • 8/1/2019 7. Temperature Measurement

    24/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    24

    The Peltier effect is due to the thermodynamically reversible conservationof energy as current flows across the junction, in contrast to theirreversible dissipation of energy associated with i2R losses.

    The Peltier heat is the quantity of heat in addition to the quantity i2R thatmust be removed from the junction to maintain the junction at a constanttemperature. This amount of energy is proportional to the current flowingthrough the junction. The proportionality constant is the Peltier coefficient

    AB

    and the heat transfer required to maintain a constant temperature is

    QAB.i caused by Peltier effect alone. This principle is used in thermoelectric

    refrigerator which works without any moving parts.

    The Thomson Effect : In addition to theSeebeck effect & Peltier effect, there is athird phenomenon that occurs inthermoelectric circuits. Consider aConductor as shown in the figure which

    Is subjected to a longitudinal temperatureGradient and also a potential difference,Such that there is a flow of current andHeat in the conductor. To maintain aConstant temperature in the conductor it isfound that quantity of energy differentthan i2R must be removed from the conductor.This energy is expressed in terms of the Thomson Coefficient as follows

    Q = i (T1 -T2. )

    It may be experimentally observed that if currentis allowed to flow through a conductor maintained

    in a temperature Gradient (ref. fig.), left half sidebecomes cooler than the right half. This is called the

    positive Thomson effect. For Cadmium, antimony,silver and zinc exactly reverse happened which is

    called the negative Thomson effect.

    For a thermocouple circuit, all three effects maybe present and may contributte to the overall emfof the circuit.

  • 8/1/2019 7. Temperature Measurement

    25/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    25

    Thermo-emf measurement

    The magnitude of the thermo emf depends on the materials and thetemperature difference between the junctions.

    The process converts heat energy to electrical energy and theconversion is reversible.

    Measurement of thermo emf is linked to temperature measurementunless the current flow within thermocouple is inhibited. Because if thecurrrent is allowed to flow, it will change the temperature of the junction(by Peltier effect)

    Thermocouple laws

    The behavior of thermocouples can be summarised into two laws.

    (1) Law of intermediate temperatures

    (2) Law of intermediate metals

    Law of intermediate temperatures : If a thermocouple is made of two metals

    A & B, then

    (1) If the junctions of the thermocouple are maintained at temperatures T1and T2., then it will produce the same emf whatever be the value of thetemperatures at other parts of the thermocouple.

    (2) If E12 is the emf generated for junctions at T1 and T2. , E23 is that for

    Junctions at T2 and T3. then the emf E13 for junctions at T1 and T3 isE12 + E23

    The law allows a thermocouple calibrated for one reference temperature to be

    used with another reference temperature.

  • 8/1/2019 7. Temperature Measurement

    26/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    26

    Law of intermediate Metals : If a thermocouple junctions are maintained at

    T1 and T2., then this law has the following implications.

    1) If we have a thermocouple AB made up of metals (1) & (2) and anotherthermocouple also made up of metals (1) & (2) but another metal (3) isincorporated somewhere in between, then the net emf will be same in

    both the cases

    2) If EAC is the emf generated by the thermocouple AC and EBC is that bythe thermocouple BC then the emf EAB generated by the thermocoupleAB will be EAC+ EBC (= EAB)

    In the following circuit the measured emf will be same as the open circuit emf,

    which corresponds to temperatures T1 & T2 as long as the temperatures of

    junctions AC & BC are same such as in the case of terminal box of the

    thermocouple from where copper wires are taken to voltmeter.

    Thus copper wires or the copper terminals in the terminal box are like thematerial C shown above but that will not alter the emf generated by

    thermocouple AB. Hence copper extension wires can be used to transmit

    thermocouple emf as long as all terminals and wires are at same temperature.

  • 8/1/2019 7. Temperature Measurement

    27/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    27

    The basic thermocouple circuit

    shown above can be used

    to measure the difference

    between the two temperatures

    T1 and T2 . For practicalTemperature measurements,

    One of these junctions becomes

    a reference junction and is

    maintained at some known constant temperature reference point (such as ice

    bath shown). The other junction then becomes the measuring junction.

    It may be noted that extension wires are needed when the measuring

    instrument is to be placed at a considerable distance from the reference

    junction. Maximum accuracy is obtained when the leads are of the same

    material as the thermocouple element as shown in fig.

    However this approach is not economical while using the expensive

    Thermocouple materials. Further a small inaccuracy is still possible if the binding

    post of the instrument is made of say copper and the two binding posts are at

    different temperature. Therefore it is preferable to employ the system shown in

    fig below to keep the copper iron and copper constantan junctions in the

    thermos flask at 0oC and provide binding posts of copper this ensures

    maximum accuracy in the thermocouple operation. The law of intermediate

    materials ensures that neither the potentiometer nor the lead wires will change

    emf in the circuit., as long as the connecting junctions are at same temperature.

  • 8/1/2019 7. Temperature Measurement

    28/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    28

    To be able to measure temperature with this circuit it is necessary to know

    the relationship between the output emf and the temperature of the measuring

    Junction, for the specific reference temperature.

    One method of determining the relationship is to calibrate the thermocouple.

    However for reasonable level of uncertainty for temperature measurement,

    standard materials and procedures allow thermocouples to be accurate devices,without the necessity of calibration.

    As illustrated earlier, a very common reference junction temperature is

    provided by ice point, 0oC, which is created by filling a vacuum flask, or Dewar

    flask. With finely crushed ice and adding a small quantity of water.

    Electronic reference junctions provide a convenient means of themeasurement of measurement of temperature without the necessity to construct

    an ice bath. Numerous manufacturers produce commercial temperature

    measuring device with built in reference junction compensation. Electronics

    generally rely on a thermistor to determine the local ambient temperature.

    Thermocouple materials : The choice of materials for thermocouples isgoverned by following factors

    (1) ability to withstand the temperature at which they are used(2) Immunity from contamination / oxidation etc. which ensures maintenance of

    the precise thermo electric properties with continuous use and

    (3) Linearity characteristics

    It may be noted that the relectionship between thermo electric emf and the

    difference between hot and cold junction temperatures is approximately of the

    parabolic form

    E = aT = bT2

  • 8/1/2019 7. Temperature Measurement

    29/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    29

    Thermocouples are broadly classified as1. Base metal thermocouples &2. Rear metal thermocouples.

    Base metal thermocouples use combination of pure metals and alloys of iron,copper and nickel and are used for temperature up to 1450 K. These are mostcommonly used in practice as they are more sensitive cheaper and have nearlylinear characteristics.Their main limitation is the lower operating range because of their low meltingpoint and vulnerability to oxidation.

    Rare metal thermocouples use a combination of pure metals and alloys of

    platinum for temperatures up to 1600o

    C and Tungsten, Rhodium andMolybdenum for temperatures up tp 3000oC

    Thermopile : For special purposes where high sensitivity is needed,thermocouple may be attached in series. The out put is then the numericalsum of the voltages expected from each single thermocouple. Thiscombination is known as thermopile.

    When connected in parallel, a group of thermocouples will give a readingthat is numerical average of the individual once provided the resistance ofeach individual thermocouple is the same.

  • 8/1/2019 7. Temperature Measurement

    30/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    30

    Typical thermocouples with their temperature ranges and other salient

    operating characteristics are given in the table below.

    NIST provides specifications for materials and construction of standard

    thermocouple circuits for measurements. Many material combinations exist

    for thermocouple, these are identified by thermocouple type and denoted

    by letters E , J,K,S,T. For each standard thermocouple charts of

    temperature and emf are available with reference junction at 0oC.

  • 8/1/2019 7. Temperature Measurement

    31/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    31

    Type Material Combination Application

    Positive (+) Negative (-)

    E Chromel Constantan Highestsensitivity(

  • 8/1/2019 7. Temperature Measurement

    32/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    32

    The measurement of temperature through the detection of thermal

    radiation presents some unique advantages. All the temperature measuring

    devices viz. filled system thermometers, thermister or thermocouples require themeasuring sensor to be brought in physical contact with the body whose

    temperature is to be measured. This means that the sensor must be capable of

    withstanding this temperature. In the case of very hot bodies, the thermometer

    may melt at high temperature.

    Secondly for bodies that are moving, a non contacting device for

    measuring temperature is more convenient. Thirdly if estimation of temperature

    over the surface of an object is required then a non contact type device can

    readily scan the surface.

    For temperatures above 650oC, the heat radiations emitted from the

    body are of sufficient intensity to be used for measuring temperature. The sensor

    of thermal radiation need not be in contact with the surface to be measured,

    which makes this method attractive for a wide variety of applications.

    Instruments that employ radiation principles fall into three general classes(a) Total radiation pyrometer

    (b) Selective (Partial) radiation pyrometer

    (c) Infrared (IR) pyrometer

    The first one is sensitive to all radiation that enters the instrument while second

    one is sensitive to only a particular wavelength. IR pyrometer employ only the

    infrared portion of the spectrum for measurement.

    Radiative Temperature Measurements

  • 8/1/2019 7. Temperature Measurement

    33/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    33

    Radiation Fundamentals

    Radiation refers to electromagnetic waves from the surface of an object.The radiation has characteristic of both waves and particles i.e. the radiationis composed of photons. Photons usually travel in straight line from point ofemission to another surface where they may get absorbed, reflected or

    transmitted. Radiation exists over a large range of wavelengths thatincludes x-rays, Ultra violet radiation, visible light and thermal radiation.

    The thermal radiation emitted from an object is related to its temperature. Ithas wavelengths ranging from 10-7 to 10-3 m

    Radiation emitted by an object is proportional to 4th power of temperature.

    Eb=T4

    Where Ebis the flux of energy radiating from an ideal surface or the black body

    emissive power. For any non black finite surface, the radiant energy received by

    body B from body A is given by

    EA/B= C [TA4 - TB4 ] {W/m2}

    Where c is shape factor of B relative to A

    is emissivity of the body B &

    Is the Stefan Botzmann constant = 56.7x10-12 [kW/m2.K]

    The emissive power is a direct measure of the total radiation emitted by an

    object. However it is emitted over a range of wavelengths. At any temperature

    the distribution of energy emitted as a function of wavelength is unique. Total

    emissive power at given wavelength is given by Planks law.

    Eb= . 2 hp c2

    5[ exp(hpc/kBT) -1]Where is wavelength,c is speed of light,

    hp is planks constant = 6.6256x10-34 Js

    KB is Boltzmanns constant =1.3805x10-23 j/K

  • 8/1/2019 7. Temperature Measurement

    34/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    34

  • 8/1/2019 7. Temperature Measurement

    35/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    35

  • 8/1/2019 7. Temperature Measurement

    36/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    36

    Total Radiation Pyrometer The total radiation pyrometer receives a controlled sample of total radiation

    of a hot body (say a furnace) and focuses it on a temperature sensitivetransducer. The term total radiation includes both visible (light) and invisible(infrared) radiations.

    Wavelength of light in the visible range is 0.3 to 0.72 m, whereas infrared

    radiations are associated with relatively large wavelengths of 0.72 to 1000m. They require special optical material for focusing. As ordinary glassabsorbs infrared radiation it is unsuitable.

    Above fig shows a schematic diagram of Ferys total radiation pyrometer. It

    consists of blackened tube T open at one end to receive the radiations from the

    object, temperature of which is to be measured. The other end of tube has a

    sighting aperture in which an adjustable eyepiece is usually fitted. The thermal

    radiation impinge on the concave mirror. Its position can be adjusted suitably bya rack and pinion arrangement to get proper focusing of the thermal radiation on

    the detector disc S. The detector disc is made up off blackened platinum sheet

    and is connected to a thermocouple / thermopile junctions or to a resistance

    thermometer bridge circuit.

    The energy received by pyrometer is given by Stefan Boltzmann law as

    follows

    Where EA/B is energy received by the pyrometer in W/m2

    C is shape factor of B relative to A

    is emissivity of the body B & Is the Stefan Botzmann constant = 56.7x10-12 [kW/m2.K].TA

    & TBare steady state absolute temperature of the source and detector disc

    EA/B=C ( TA4 -TB4)

  • 8/1/2019 7. Temperature Measurement

    37/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    37

    These pyrometers are usually calibrated against known temperatures in

    the range of 700 to 2000oC, where thermocouples and resistance thermometers

    cannot be employed.

    Errors in measurement arise from two sources in actual use

    1. Any filtering material such as smoke, dust, gases, etc which werenot present during calibration will reduce the energy received andcause error.

    2. Error may be caused due to surface having changed emissivity aftercalibration due to oxidation etc.

    To reduce such uncertainties, pyrometers are calibrated from time to time in

    actual use.

    In view of the errors due to filtering and emissivity, the total radiation pyrometer

    is not very accurate temperature indicator. However it can be used to good

    advantage in fixed locations where emissivity and optical paths are well known

    and constant. A typical use is a large furnace in metal industries.

    The advantage is that the signal is electrical and therefore can be used for

    control applications.

    The principle of this instrument is based on Planks law which states that

    the energy level in the radiations from hot body are distributed in different

    wavelengths. As the temperature increases the emissive power shifts to shorter

    wavelengths. The Planks distribution equation is simplified as.

    Selective (Partial) Radiation Pyrometer

  • 8/1/2019 7. Temperature Measurement

    38/39

    DBIT Pvt. circulationNotes by Prof.S.P.Sabnis

    38

    The classical form of this optical pyrometer is the disappearing filamentoptical pyrometer. It is most accurate of all radiation pyrometers. Howeverits use is limited to temperatures greater than 700oC since it requiresvisual brightness match by human operator. This instrument is used byInternational Practical Temperature Scale above 1064oC.

    From Planks equation it is evident that for a given wavelength, the radiantintensity (Brightness) varies with temperature. In the disappearing filamentinstrument shown in the fig. an image of target is superimposed on theheated filament.

    The tungsten lamp, which is very stable, is previously calibrated so thatthe current through the filament is known, The brightness temperature ofthe filament is also known. A red filter that allows only narrow band of

    wavelengths around 0.65m is placed between the observer eye and thetungsten lamp and the target image.

    The observer controls the lamp current until the filament disappears in thesuperimposed target image.

  • 8/1/2019 7. Temperature Measurement

    39/39

    The temperature calibration is made in terms of the lamp heating current.

    Because of the manual null balancing principle, the optical pyrometer is notuseful for continuous reading or automatic control applications. However it ismore accurate and less subject to large errors compared to Total radiationpyrometer

    Typical accuracy for range 850oC-1200oC is +5%and for range 1100oC-1950oC is +10%