7940010r1d4

Upload: hamilton

Post on 10-Mar-2016

4 views

Category:

Documents


0 download

DESCRIPTION

manual

TRANSCRIPT

  • PYLON MODEL

    AB6A

    RADIATION MONITOR MANUAL

    MANUAL NUMBER: 7940010

    Revision 1 Draft 4

    Copyright 2014 Pylon Electronics Inc. (ALL RIGHTS RESERVED)

    Registered Trademark Canada and U.S.A.

  • Page 2 of 82

    Document Number: 7940010 Rev 1 Draft 4

    WARNING 1) Voltages up to 1250 volts DC are present on the Photomultiplier Tube (PMT) when a

    method is being run. There are no User serviceable parts inside the unit. Do not open the unit.

    2) If the equipment is used in a manner that is not specified in this manual, the protection

    provided by this equipment may be impaired.

  • Page 3 of 82

    Document Number: 7940010 Rev 1 Draft 4

    CAUTION 1) The AB6A PMT assembly contains a photomultiplier tube which may be damaged by

    ambient light: Before starting a measurement run, make sure that the photomultiplier tube is

    protected from light by a detector or cap. Make sure that the measurement run is stopped before exposing the photomultiplier

    tube to ambient light. 2) Do not drop the AB6A or any of its components as serious damage may result. 3) DO NOT attempt to thread any older Pylon products such as the ABA, CPC, 200P, and

    200A onto the AB6A PMT assembly. The threads are incompatible. Forcing any of these older detector products onto an AB6A PMT assembly will damage the threads of the older Pylon products and the AB6A PMT assembly threads.

    4) Do not force any AB6A detector or accessory onto the AB6A PMT assembly since this may

    damage the cell and / or AB6A PMT assembly threads. A small amount of lubricating grease may be applied to the threads to assist with installation. This may be renewed when necessary.

  • Page 4 of 82

    Document Number: 7940010 Rev 1 Draft 4

    IMPORTANT 1) The AB6A is designed for use with a range of detectors and accessories. Consult the

    instruction manual which accompanied the detector or accessory in order to use the AB6A for making measurements with the detector or accessory.

    2) The nominal detector Sensitivity and Efficiency parameter values may be used for most

    calculations. The calibrated detector Sensitivity and Efficiency parameters values will provide a more accurate result in the calculations. If maximum measurement accuracy is required, the detector should be calibrated with the AB6A monitor with which it will used. Consult the instruction manual that accompanied the detector.

    3) As photomultiplier tubes and scintillation cells are light sensitive, the AB6A will require a

    wait period of 30 seconds before starting a method, if the photomultiplier tube and / or scintillation cell has been exposed to light.

    4) Many of the terms and abbreviations used in this manual are described in Pylon manual

    number 7940018 - Glossary of Selected Pylon Terms. 5) The following symbols may be used throughout this manual:

    WARNING The Warning symbol is use to identify notes that are used to warn of potential hazards that could cause injury or death to personnel as well as damage to the equipment.

    CAUTION The Caution symbol is use to identify notes that are used to warn of potential hazards that could damage the equipment.

  • Page 5 of 82

    Document Number: 7940010 Rev 1 Draft 4

    WARRANTY PYLON ELECTRONICS INC. products are warranted against defects in material and workmanship for a period of one year from the date of shipment. Our warranty obligation is limited to the repair of products or parts thereof which, on our examination, prove to be defective during the warranty period. The warranty shall not apply to any equipment which has been subject to accident, incorrect wiring not of our own or operation not in accordance with manufacturer's written instructions. PYLON ELECTRONICS INC. repairs are warranted against defects in material and workmanship for a period of 90 days from the date of shipment. Our warranty obligation is limited to the repair of the unit returned prepaid, to our factory and which, on our examination, prove to be defective during the warranty period. Equipment which has work performed under Warranty will be returned to the Distributor freight prepaid. The warranty shall not apply to any equipment which has been subject to accident, incorrect wiring not of our own, or operation not in accordance with manufacturer's written instructions. This warranty is given by PYLON in lieu of all other warranties arising in law or otherwise in respect of the goods, and this Company shall not be liable under any circumstances for consequential damage. Note: All product(s) must be returned prepaid, to our factory. All warranty claims shall be addressed to: Pylon Electronics Inc. 147 Colonnade Road, Ottawa, ON K2E 7L9 CANADA Phone: 613-226-7920 Fax: 613-226-8195 Email: [email protected]

  • Page 6 of 82

    Document Number: 7940010 Rev 1 Draft 4

    DISCLAIMER Pylon Electronics Inc. has attempted to ensure that this manual is complete and accurate. However, Pylon will not be held responsible for any errors and omissions in this manual including consequential damage due to the information contained herein. Please forward all comments and suggestions for improvements on the product or this manual to: Pylon Electronics Inc. 147 Colonnade Road, Ottawa, ON K2E 7L9 CANADA Phone: 613-226-7920 Fax: 613-226-8195 Email: [email protected] For technical support, contact us at the above address. Please visit us on our website at www.pylonelectronics.com. Click on the Products icon.

  • Page 7 of 82

    Document Number: 7940010 Rev 1 Draft 4

    EC DECLARATION OF CONFORMITY We, the undersigned,

    Manufacturer: Pylon Electronics Inc.Address, City: 147 Colonnade Road, Ottawa, ON K2E 7L9 Country: Canada

    Declare the following apparatus:

    Product name : Portable Radiation Monitor Model name : AB6A

    Conforms with the essential requirements of the following directives:

    2004/108/EC-EMC Directive Using following standard(s): IEC 61326-1 Ed 2.0 2012-07 EN 61000-3-2: 2006+A2 2009 EN 61000-3-3: 2013 CISPR 11

    2006/95/EC-LVD Directive

    Using following standard(s): IEC 61010-1 Edition 3.0 2010-06

    CE Marking

    (EC conformity marking)

    (http://ec.europa.eu/enterprise/faq/ce-mark.htm) Declaration Date: 10 June 2014 Position: Vice President Year to begin affixing CE marking: 2014 Name: James S. Mullins

  • Page 8 of 82

    Document Number: 7940010 Rev 1 Draft 4

    ENVIRONMENTAL

    The battery used in this product contains lead. Dispose of the battery and product

    in accordance with current local regulations.

  • Page 9 of 82

    Document Number: 7940010 Rev 1 Draft 4

    TABLE OF CONTENTS WARNING ................................................................................................................................................................... 2CAUTION .................................................................................................................................................................... 3IMPORTANT .............................................................................................................................................................. 4WARRANTY ............................................................................................................................................................... 5DISCLAIMER ............................................................................................................................................................. 6EC DECLARATION OF CONFORMITY ............................................................................................................... 7ENVIRONMENTAL ................................................................................................................................................... 81.0 INTRODUCTION ........................................................................................................................................... 13

    1.1 SCOPE OF MANUAL ......................................................................................................................................... 131.2 APPLICATIONS ................................................................................................................................................ 131.3 FEATURES ....................................................................................................................................................... 131.4 SPECIFICATIONS .............................................................................................................................................. 141.5 ANCILLARY PRODUCTS AND SERVICES ........................................................................................................... 14

    1.5.1 AB6A Detectors ..................................................................................................................................... 151.5.1.1 Active Lucas Type Scintillation Cells ............................................................................................................. 151.5.1.2 Passive Lucas Type Scintillation Cells ............................................................................................................ 151.5.1.3 Trace Environmental Level Detector .............................................................................................................. 151.5.1.4 Filter Detector ................................................................................................................................................. 161.5.1.5 Low Efficiency Lucas Type Cells ................................................................................................................... 16

    1.5.2 AB6A Accessories .................................................................................................................................. 161.5.2.1 Pumps .............................................................................................................................................................. 16

    1.5.2.1.1 Internal Pump .......................................................................................................................................... 161.5.2.1.2 Hand Vacuum Pump ............................................................................................................................... 161.5.2.1.3 120V Electric Vacuum Pump .................................................................................................................. 161.5.2.1.4 240V Electric Vacuum Pump .................................................................................................................. 16

    1.5.2.2 External Photomultiplier Tube System............................................................................................................ 171.5.2.3 Water Degassing System ................................................................................................................................. 171.5.2.4 Soil Probe ........................................................................................................................................................ 171.5.2.5 Sealed Check Source ....................................................................................................................................... 171.5.2.6 AB-5 Detector Adapter ................................................................................................................................... 171.5.2.7 Filter Kits ........................................................................................................................................................ 171.5.2.8 Drying Column Kits ........................................................................................................................................ 181.5.2.9 Vacuum Gauge Kits ........................................................................................................................................ 181.5.2.10 Flow Meter Kits ............................................................................................................................................ 18

    1.5.3 AB6A Consumables and Parts .............................................................................................................. 191.5.4 AB6A Services ....................................................................................................................................... 19

    1.5.4.1 Operation Point Determination ....................................................................................................................... 191.5.4.2 Repair Services ............................................................................................................................................... 19

    2.0 AB6A DESCRIPTION .................................................................................................................................... 192.1 DESCRIPTION .................................................................................................................................................. 19

    2.1.1 Overview ............................................................................................................................................... 192.1.2 Front Panel ........................................................................................................................................... 19

    2.1.2.1 DC Power Jack (J1) ......................................................................................................................................... 212.1.2.2 Host USB Connector (J2) ................................................................................................................................ 212.1.2.3 Client USB Connector (J3) ............................................................................................................................. 212.1.2.4 Pulse Output Connector (J4) ........................................................................................................................... 222.1.2.5 Audio Jack (J5) ............................................................................................................................................... 222.1.2.6 External PMT Connector (J6) ......................................................................................................................... 232.1.2.7 Internal PMT Interface .................................................................................................................................... 232.1.2.8 Optional Internal Pump Bay ............................................................................................................................ 24

  • Page 10 of 82

    Document Number: 7940010 Rev 1 Draft 4

    2.1.2.9 Fuse Holder ..................................................................................................................................................... 242.1.2.10 Temperature / Relative Humidity Sensor ...................................................................................................... 242.1.2.11 Power Switch ................................................................................................................................................ 252.1.2.12 Display .......................................................................................................................................................... 25

    2.1.3 Power .................................................................................................................................................... 252.1.4 Environmental Barometric Pressure Sensor ......................................................................................... 26

    2.2 THEORY OF OPERATION .................................................................................................................................. 263.0 OPERATION ................................................................................................................................................... 26

    3.1 OVERVIEW ...................................................................................................................................................... 263.2 POWERING ON ................................................................................................................................................ 27

    3.2.1 Powering Off ......................................................................................................................................... 283.3 TOUCHSCREEN OPERATION ............................................................................................................................ 283.4 AB6A GRAPHICAL USER INTERFACE (GUI) OVERVIEW ................................................................................. 29

    3.4.1 General.................................................................................................................................................. 293.4.1.1 Title Bar .......................................................................................................................................................... 303.4.1.2 Screen Display ................................................................................................................................................ 303.4.1.3 Status Bar ........................................................................................................................................................ 30

    3.4.1.3.1 Battery Charge Status.............................................................................................................................. 303.4.1.3.2 USB Connection ..................................................................................................................................... 303.4.1.3.3 System Locked ........................................................................................................................................ 303.4.1.3.4 Status LED .............................................................................................................................................. 313.4.1.3.5 Selection Box .......................................................................................................................................... 313.4.1.3.6 Time (Date) ............................................................................................................................................. 31

    3.4.2 Startup Form ......................................................................................................................................... 313.4.2.1 Methods ........................................................................................................................................................... 323.4.2.2 Files ................................................................................................................................................................. 323.4.2.3 User Setup ....................................................................................................................................................... 333.4.2.4 Touchscreen Calibration ................................................................................................................................. 333.4.2.5 About .............................................................................................................................................................. 33

    3.4.3 Files ....................................................................................................................................................... 333.4.4 User Setup ............................................................................................................................................. 35

    3.4.4.1 Monitor Setup ................................................................................................................................................. 353.4.4.2 Pump Setup ..................................................................................................................................................... 363.4.4.3 Detector Setup ................................................................................................................................................. 363.4.4.4 Source Setup ................................................................................................................................................... 363.4.4.5 Calibration ....................................................................................................................................................... 36

    3.4.5 Monitor Setup ........................................................................................................................................ 373.4.5.1 Operating Point ............................................................................................................................................... 373.4.5.2 Units of Measure ............................................................................................................................................. 393.4.5.3 Date and Time ................................................................................................................................................. 393.4.5.4 Password ......................................................................................................................................................... 403.4.5.5 Owner .............................................................................................................................................................. 41

    3.4.6 Pump Setup ........................................................................................................................................... 423.4.7 Detector Setup ....................................................................................................................................... 44

    3.4.7.1 Detector Selection ........................................................................................................................................... 443.4.7.2 Detector Editing .............................................................................................................................................. 45

    3.4.8 Source Setup .......................................................................................................................................... 463.4.8.1 Sealed Source Selection .................................................................................................................................. 463.4.8.2 Sealed Source Editing ..................................................................................................................................... 47

    3.5 AB6A METHODS OVERVIEW .......................................................................................................................... 483.5.1 Run Tab ................................................................................................................................................. 483.5.2 Setup Tab ............................................................................................................................................... 503.5.3 Detector Tab ......................................................................................................................................... 573.5.4 Source Tab ............................................................................................................................................ 59

    3.6 SETUP INSTRUCTIONS ..................................................................................................................................... 603.6.1 Administrative ....................................................................................................................................... 603.6.2 Detectors ............................................................................................................................................... 60

  • Page 11 of 82

    Document Number: 7940010 Rev 1 Draft 4

    3.6.3 Sources .................................................................................................................................................. 603.7 MEASUREMENTS ............................................................................................................................................. 61

    3.7.1 Methods Overview ................................................................................................................................. 613.7.1.1 Standard Continuous Method .......................................................................................................................... 613.7.1.2 Standard Grab Method .................................................................................................................................... 633.7.1.3 System Background Method ........................................................................................................................... 663.7.1.4 Quasi Method .................................................................................................................................................. 673.7.1.5 Monitor Check Method ................................................................................................................................... 68

    3.7.2 Methods Setup ....................................................................................................................................... 683.7.2.1 Standard Continuous Method .......................................................................................................................... 69

    3.7.2.1.1 Setup Tab ................................................................................................................................................ 693.7.2.1.2 Detector Tab ........................................................................................................................................... 70

    3.7.2.2 Standard Grab Method .................................................................................................................................... 703.7.2.2.1 Setup Tab ................................................................................................................................................ 703.7.2.2.2 Detector Tab ........................................................................................................................................... 71

    3.7.2.3 System Background Method ........................................................................................................................... 713.7.2.3.1 Setup Tab ................................................................................................................................................ 713.7.2.3.2 Detector Tab ........................................................................................................................................... 72

    3.7.2.4 Quasi Method .................................................................................................................................................. 723.7.2.4.1 Setup Tab ................................................................................................................................................ 723.7.2.4.2 Detector Tab ........................................................................................................................................... 73

    3.7.2.5 Monitor Check Method ................................................................................................................................... 733.7.2.5.1 Setup Tab ................................................................................................................................................ 733.7.2.5.2 Source Tab .............................................................................................................................................. 74

    3.7.3 Measurements ....................................................................................................................................... 743.8 PC SOFTWARE ................................................................................................................................................ 74

    3.8.1 PC Software Operation Instructions ........................................................................................................ 744.0 MAINTENANCE ............................................................................................................................................. 74

    4.1 CLEANING....................................................................................................................................................... 755.0 CALIBRATION ............................................................................................................................................... 756.0 REPAIR & TROUBLESHOOTING.............................................................................................................. 76

    6.1 REPAIR ........................................................................................................................................................... 766.2 TROUBLESHOOTING ........................................................................................................................................ 76

    6.2.1 Power On Problem ................................................................................................................................ 766.2.2 Status LED Problem .............................................................................................................................. 77

    7.0 INITIAL SETUP .............................................................................................................................................. 777.1 OVERVIEW ...................................................................................................................................................... 777.2 UNPACKING .................................................................................................................................................... 78

    7.2.1 Packing List ........................................................................................................................................... 787.2.2 Unpacking Instructions ......................................................................................................................... 78

    7.3 INITIAL OPERATION AND TESTING .................................................................................................................. 787.3.1 Power Connection ................................................................................................................................. 787.3.2 Functional Test ..................................................................................................................................... 79

    7.3.2.1 Power On ........................................................................................................................................................ 797.3.2.2 Date and Time Check ...................................................................................................................................... 807.3.2.3 Units of Measure Check .................................................................................................................................. 80

    7.3.3 Initial Operation Conclusion ................................................................................................................ 817.4 USER SUPPLIED COMPONENTS ........................................................................................................................ 81

    7.4.1 Pulse Output Cable ............................................................................................................................... 817.4.2 Headphones ........................................................................................................................................... 81

    7.5 PC SOFTWARE INSTALLATION ........................................................................................................................ 817.5.1 PC Requirements .................................................................................................................................... 827.5.2 PC Software Installation Instructions ...................................................................................................... 82

  • Page 12 of 82

    Document Number: 7940010 Rev 1 Draft 4

    LIST OF FIGURES

    NOTE The illustrations are for reference only. Some illustrations may

    not exactly match the delivered unit. FIGURE 1 - FRONT PANEL LAYOUT ............................................................................................................................... 20FIGURE 2 - STARTUP FORM ........................................................................................................................................... 32FIGURE 3 - FILES POPUP ............................................................................................................................................... 34FIGURE 4 - USER SETUP FORM ..................................................................................................................................... 35FIGURE 5 - MONITOR SETUP FORM ............................................................................................................................... 37FIGURE 6 - OPERATING POINT FORM ............................................................................................................................ 38FIGURE 7 - UNITS OF MEASURE FORM .......................................................................................................................... 39FIGURE 8 - PASSWORD FORM ....................................................................................................................................... 41FIGURE 9 - OWNER FORM ............................................................................................................................................. 42FIGURE 10 - PUMP SETUP FORM ................................................................................................................................... 43FIGURE 11 - DETECTOR SETUP SELECTION FORM......................................................................................................... 44FIGURE 12 - EDIT DETECTOR FORM ............................................................................................................................. 46FIGURE 13 - SOURCE SETUP SELECTION FORM ............................................................................................................. 46FIGURE 14 - EDIT SOURCE FORM .................................................................................................................................. 48FIGURE 15 - MEASURE RUN TAB FORM ........................................................................................................................ 50FIGURE 16 - MEASURE SETUP TAB FORM ..................................................................................................................... 51FIGURE 17 - MEASURE DETECTOR TAB FORM .............................................................................................................. 58FIGURE 18 - MEASURE SOURCE TAB FORM .................................................................................................................. 59FIGURE 19 - STARTUP FORM ......................................................................................................................................... 79

  • Page 13 of 82

    Document Number: 7940010 Rev 1 Draft 4

    1.0 INTRODUCTION 1.1 Scope of Manual This manual is designed to provide the User with information on the AB6A monitor

    (hereinafter referred to as the AB6A). Besides providing a detailed description of the AB6A, this manual also provides instructions covering operation, maintenance, calibration instructions, repair and troubleshooting, and installation of the monitor.

    In order to measure radiation, the User must attach the appropriate detector to the AB6A.

    These detectors must be purchased separately and include active and passive Lucas style scintillation cells.

    Although some information is provided in this manual, the detailed use and operation of

    the detectors is beyond the scope of this manual. The User is directed to the manuals for the individual detectors for the use and operation of the detector.

    1.2 Applications With the appropriate detector and accessory or accessories, the AB6A may be used to

    measure radon or thoron concentration in any of the following applications: Homes. Buildings. Laboratories. Mines. Soil. Water. Environment. 1.3 Features The AB6A contains the following items and features: Single Board Computer (SBC). Touchscreen Liquid Crystal Display (LCD). Windows operating system. Graphical User Interface (GUI) Built in Photomultiplier Tube (PMT). Connectors for: External PMT. Pulse output signal. Headphones. USB ports (Host and Client). High sensitivity. Battery backup.

  • Page 14 of 82

    Document Number: 7940010 Rev 1 Draft 4

    User selectable concentration Unit of Measure (UOM) [Bq/m3 or pCi/l] User programmable measurement interval. User programmable alarm level. Calibration parameters stored in non-volatile memory. PC software. 1.4 Specifications The following table provides the specifications for the AB6A: Mode of Operation: Multiple Sample and Count Periods: User Programmable Maximum Count Rate: 10,000 counts per second Electronic Background: < 0.4 cpm Detector: Various

    Display / Data Entry: Back-Lit High Contrast LCD complete with

    Touchscreen Memory: 16 GB (Minimum) SD Memory Card Data Ports: 1) Host USB 2) Client USB 3) 5V TTL Pulse via SMA Connector Alarm Level: User Programmable

    Environmental Measurements:

    Temperature, Relative Humidity, Barometric Pressure, Dew/Frost Point Temperature.

    Operating Power: 12 to 14.7 volts, direct current, 1.2 amps Battery Type: 12 Volt Rechargeable Lead-Acid Gel Cell Re-Charge Duration: 12 Hours AC Battery Charger Power: 100 to 240 VAC 50 ~ 60 Hz 1.2A Calibration: Factory or User Calibrated

    Operating Temperature Range:

    0 to + 40 C (+32 to +104 F)

    Storage Temperature Range: -20 to +60 C (-4 to +140 F) Humidity: 0 - 90% Non Condensing Dimensions: 31 cm x 23 cm x 20 cm (12" x 9" x 7.75") [W x D x H] Weight: 6 kg (13 lbs.)

    All specifications are nominal and subject to change without notice. 1.5 Ancillary Products and Services Pylon offers a variety of ancillary products and services including detectors, accessories,

    etc. to supplement, enhance, or customize the AB6A for specific applications. The following paragraphs provide a brief overview of these options. These options are sold separately. Please contact Pylon for more information and current pricing.

  • Page 15 of 82

    Document Number: 7940010 Rev 1 Draft 4

    1.5.1 AB6A Detectors Pylon offers the following detectors for the AB6A: NOTE The User must purchase or already own a minimum of one

    appropriate detector in order to use the AB6A. 1.5.1.1 Active Lucas Type Scintillation Cells Active Lucas type scintillation cells are cells which require the Operator to actively

    acquire the gas sample. Active Lucas type scintillation cells may be used for both continuous or grab sampling radon measurements.

    Grab sample measurements may be achieved by placing the cell under a vacuum and then

    releasing the vacuum to draw the gas sample into the cell or by pumping the gas sample through the cell. When pumping the gas sample through the cell, it is important to ensure that the flow rate and pumping time are such that the total volume of gas flowed through the cell is at least 3 times the volume of the cell.

    Continuous measurements may be achieved by continuously pumping the gas sample

    through the cell. Pylon offers 2 models of active Lucas type cells for the AB6A. The model 600A has a

    nominal volume of 0.272 l and the model 610A has a nominal volume of 0.154 l. 1.5.1.2 Passive Lucas Type Scintillation Cells Passive Lucas type scintillation cells are cells which passively allow the gas sample to

    diffuse into the cell. Passive Lucas type scintillation cells are primarily used for continuous radon measurements.

    Pylon offers a single model of the passive Lucas type cell for the AB6A. The model 600P

    has a nominal volume of 0.272 l. 1.5.1.3 Trace Environmental Level Detector The trace environmental level detector is a radon detector that detects very low levels of

    radon.

    NOTE The Operator must purchase an external photomultiplier tube

    system (sold separately) in order to use the Trace Environmental Level Detector with the AB6A.

  • Page 16 of 82

    Document Number: 7940010 Rev 1 Draft 4

    The model 6500 is the trace environmental level detector for the AB6A. 1.5.1.4 Filter Detector Pylon offers a filter detector kit that is used to measure the amount of radon daughters

    that have been collected on a filter by means of scintillation disks. The model 6400 is the filter detector kit for the AB6A. 1.5.1.5 Low Efficiency Lucas Type Cells Pylon offers special low efficiency cells to allow the AB6A system to measure very high

    radon concentrations. Efficiencies as low as 2% are offered, which allow for the measurement of radon concentrations in the 740 million Bq/m3 (20 million pCi/l) range.

    1.5.2 AB6A Accessories Pylon offers the following accessories for the AB6A: 1.5.2.1 Pumps 1.5.2.1.1 Internal Pump The model 6800 is a user-installable pump for the AB6A. This pump allows the Operator

    to use active Lucas type scintillation cells for both continuous and grab sample measurements.

    1.5.2.1.2 Hand Vacuum Pump The model 6660 hand vacuum pump allows the Operator to evacuate the air from active

    Lucas type scintillation cells for use for grab sample measurements. 1.5.2.1.3 120V Electric Vacuum Pump The electric vacuum pump allows the Operator to evacuate the air from active Lucas type

    scintillation cells for use for grab sample measurements. The 120V version operates using 120V AC, 60 Hz power.

    1.5.2.1.4 240V Electric Vacuum Pump The electric vacuum pump allows the Operator to evacuate the air from active Lucas type

    scintillation cells for use for grab sample measurements. The 240V version operates using 240V AC, 50 Hz power.

  • Page 17 of 82

    Document Number: 7940010 Rev 1 Draft 4

    1.5.2.2 External Photomultiplier Tube System Some detectors such as the trace environmental detector cannot be attached to the internal

    PMT on the AB6A due to their size, mode of operation, etc. In addition, there may be measurements with, for example, Lucas type scintillation cells, where the detector will be placed in one area and the AB6A place in another. (e.g., Chamber and lab, respectively.) The model 6200 external photomultiplier tube system allows the Operator to use these detectors in the desired manner.

    1.5.2.3 Water Degassing System The water degassing system allows the Operator to extract radon gas samples from water

    and place the sample in an active Lucas type scintillation cell for measurement. 1.5.2.4 Soil Probe The soil probe allows the Operator to extract radon gas samples from the soil and place

    the sample in an active Lucas type scintillation cell for measurement. 1.5.2.5 Sealed Check Source The model 6000 sealed check source allows the Operator to easily determine or check the

    AB6A operating point.

    WARNING THESE SOURCES CONTAIN RADIOACTIVE MATERIALS.

    ONLY KNOWLEDGEABLE USERS SHOULD HANDLE THESE SOURCES. MOST COUNTRIES HAVE REGULATIONS COVERING THE OWNERSHIP OF RADIOACTIVE SOURCES. PLEASE REFER TO THE LOCAL AUTHORITIES FOR MORE INFORMATION.

    1.5.2.6 AB-5 Detector Adapter The model 6300 AB-5 Detector Adapter allows the Operator to attach AB-5 Lucas type

    scintillation cell detectors such as the models 110A, 300A, and CPRD to the AB6A. In addition, the adapter may also be used to attach the AB-5 check source (model 3150A) to the AB6A.

    1.5.2.7 Filter Kits Normally, gas samples are filtered before they enter the detector in order to remove

    particulates that may contain radioactive radon daughters which could provide an inaccurate measurement of the radon concentration. In addition, filters are sometimes

  • Page 18 of 82

    Document Number: 7940010 Rev 1 Draft 4

    used to gather these particulates in order to make Working Level measurements. Pylon offers two types of filter kits to assist with these tasks.

    KIT DESCRIPTION MODEL Product; In-Line Filter Kit; Active Cell 6600 Product; Open Face Filter Kit; Active Cell 6610

    1.5.2.8 Drying Column Kits Condensing humidity can eventually damage Lucas type cells with corrosion. Pylon

    offers two sizes of drying column kits to remove moisture from the gas sample prior to the detector so that condensing moisture will not eventually lead to damage to the detector.

    KIT DESCRIPTION MODEL Product; Drying Column Kit; Small; Active Cell 6620 Product; Drying Column Kit; Large; Active Cell 6630

    1.5.2.9 Vacuum Gauge Kits For some sampling techniques, a specific vacuum must be applied to an active detector.

    Pylon offers two types of vacuum gauge kits to make these measurements. KIT DESCRIPTION MODEL Product; Vacuum Gauge; Digital; Active Cell 6640 Product; Vacuum Gauge; Analog; Active Cell 6650

    1.5.2.10 Flow Meter Kits For some sampling techniques and measurements, the flow rate of the sample must be

    known. Pylon offers several types of flow meter kits for these measurements. KIT DESCRIPTION MODEL Product; Flow Meter Kit; Adjustable; 0.15-1 L/Min; Active Cell 6700 Product; Flow Meter Kit; Adjustable; 0.2-4 L/Min; Active Cell 6710 Product; Flow Meter Kit; Adjustable; 1-10 L/Min; Active Cell 6720 Product; Flow Meter Kit; 0.15-1 L/Min; Active Cell 6730 Product; Flow Meter Kit; 0.2-4 L/Min; Active Cell 6740 Product; Flow Meter Kit; 1-10 L/Min; Active Cell 6750

  • Page 19 of 82

    Document Number: 7940010 Rev 1 Draft 4

    1.5.3 AB6A Consumables and Parts Pylon offers a wide variety of consumables and parts for use with the AB6A including

    but not limited to: Filters Drierite Tubing Mating Connectors Filter Holders 1.5.4 AB6A Services Pylon offers the following services for the AB6A: 1.5.4.1 Operating Point Determination Pylon can determine the optimum operating point for the AB6A. 1.5.4.2 Repair Services Pylon can perform all repairs on the AB6A. 2.0 AB6A DESCRIPTION 2.1 Description 2.1.1 Overview The AB6A is a monitor that is designed to measure radon and thoron when coupled with an

    appropriate detector. The unit is compact and portable and designed to be simple to set up and use.

    2.1.2 Front Panel Refer to Figure 1 - Front Panel Layout for the front panel layout of the AB6A. The

    following electrical connections and items are provided on the front panel: DC Power Jack (J1) Host USB Connector (J2) Client USB Connector (J3) Pulse Output Connector (J4) Audio Jack (J5) External PMT Connector (J6) Internal PMT Interface Optional Pump Bay

  • Page 20 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Fuse Holder Temperature / Relative Humidity Sensor Power Switch Display The following sections provide a description of each of the above items.

    Figure 1 - Front Panel Layout NOTE

    The symbol indicates that the User must use caution when using this equipment as there is a risk of danger. Read and follow the instructions in this manual to operate this equipment.

  • Page 21 of 82

    Document Number: 7940010 Rev 1 Draft 4

    2.1.2.1 DC Power Jack (J1) This jack provides the connection to the 12 volt direct current charger for the AB6A unit.

    This connection is used to operate the unit using AC power and to charge the integrated battery.

    The symbol on the front panel for this connector is:

    The following provides the connector pin out information: PIN DESCRIPTION NOTE(S) Center +12 - 14.7 VDC Input Power 1 Outside Common (Ground) 1 The input power to the AB6A monitor is fused for 4 Amps. Please refer to Section

    2.1.2.9. 2.1.2.2 Host USB Connector (J2) This connector provides a standard USB 2.0 host interface. For the AB6A, this port is

    primarily used for USB memory stick devices. This allows the User to provide additional memory for the AB6A and to transfer files between the AB6A and computer.

    The symbol on the front panel for this connector is:

    The following provides the connector pin out information: PIN DESCRIPTION NOTE(S) 1 USB Port "+" Power 2 USB Port "D+" Data Signal 3 USB Port "D-" Data Signal 4 USB Port "-" Power 2.1.2.3 Client USB Connector (J3) This connector provides a standard USB 2.0 client interface to allow connection with a

    computer. Please refer to Section 3.8 for more information.

  • Page 22 of 82

    Document Number: 7940010 Rev 1 Draft 4

    The symbol on the front panel for this connector is:

    The following provides the connector pin out information: PIN DESCRIPTION NOTE(S) 1 USB Port "+" Power 2 USB Port "D+" Data Signal 3 USB Port "D-" Data Signal 4 USB Port "-" Power 2.1.2.4 Pulse Output Connector (J4) This interface provides the connection to a constant 0 VDC level with +5V pulses for each

    pulse detected by the system. This may be used for a User supplied counting system such as a datalogger. The User is responsible for providing the cabling to connect to this connector. Information on the signal and wiring requirements is provided in Section 7.4.1.

    The symbol for this connector on the front panel is:

    The following provides the connector pin out information: PIN DESCRIPTION NOTE(S) Center Pulse Output Signal Shield Common (Ground) 2.1.2.5 Audio Jack (J5) This jack provides the connection for stereo headphones. The symbol on the front panel for this connector is:

    The following provides the connector pin out information: PIN DESCRIPTION NOTE(S) Ring Right Channel Sleeve Common (Ground) Tip Left Channel

  • Page 23 of 82

    Document Number: 7940010 Rev 1 Draft 4

    2.1.2.6 External PMT Connector (J6) This connector provides the connection for the external PMT which is required to operate

    some detectors. The symbol on the front panel for this connector is:

    The following provides the connector pin out information: PIN DESCRIPTION NOTE(S) Center External PMT Signal Pulse Shield Common (Ground) 1 +5 VDC Output Power 1 2 Common (Ground) 3 External PMT High Voltage Control 2 4 +12 - 14.7 VDC Output Power 3 1 The +5VDC output power to the optional external PMT assembly is current limited by the internal

    circuitry to 5 Amps. 2 There is no high voltage on this connection. This is a low voltage signal of less than +5 VDC that controls

    the level of high voltage that is applied to the PMT in the external PMT assembly. 3 The +12 - 14.7 VDC output power to the optional external PMT assembly is fused internally for 0.5

    Amps. 2.1.2.7 Internal PMT Interface The internal PMT interface provides the mechanical connection for detectors that are

    attached directly to the AB6A. The face of the detector is mounted against the face of the internal PMT.

  • Page 24 of 82

    Document Number: 7940010 Rev 1 Draft 4

    CAUTION THE PHOTOMULTIPLIER TUBE IS A DELICATE AND

    EXPENSIVE COMPONENT. DO NOT SUBJECT THE AB6A TO SUDDEN SHOCKS OR EXCESSIVE VIBRATION AS THIS MAY RESULT IN IRREPARABLE DAMAGE.

    IN ADDITION, MINIMIZE THE EXPOSURE OF THE PMT

    DETECTION SURFACE TO AMBIENT LIGHT WHEN THE AB6A IS POWERED OFF OR A METHOD IS NOT BEING RUN. DO NOT EXPOSE THE DETECTION SURFACE TO AMBIENT LIGHT WHEN A METHOD IS BEING RUN.

    2.1.2.8 Optional Internal Pump Bay This bay provides the opening to install the optional internal pump. The internal pump

    may be purchased at a later date and installed by the User. 2.1.2.9 Fuse Holder The fuse holder contains a 250 VAC / 60 VDC 4A fuse that provides a safety fuse for the

    12 - 14.7 V DC internal power for the AB6A. The symbol on the front panel for this device is:

    2.1.2.10 Temperature / Relative Humidity Sensor The temperature / relative humidity sensor is exposed to the ambient conditions through the

    front panel. The temperature sensor has an accuracy of 1 C (1.8 F). The relative humidity sensor has an accuracy of 5%.

  • Page 25 of 82

    Document Number: 7940010 Rev 1 Draft 4

    CAUTION THE BLACK CAP COVERING THE TEMPERATURE /

    RELATIVE HUMIDITY SENSOR IS POROUS. THIS CAP SHOULD NOT BE COVERED WHEN MAKING MEASUREMENTS.

    IN ADDITION, CARE SHOULD BE TAKEN WHEN

    CLEANING THE FRONT PANEL TO ENSURE WATER AND CLEANING PRODUCTS ARE NOT USED ON THE SURFACE OF THE BLACK CAP. OTHERWISE THE TEMPERATURE / RELATIVE HUMIDITY SENSOR COULD BE DAMAGED.

    2.1.2.11 Power Switch The power switch is a lighted push-on, push-off switch that turns the AB6A on and off. The symbol on the front panel for this switch is:

    2.1.2.12 Display The Display provides the local User interface for the AB6A monitor. 2.1.3 Power The power to the AB6A is provided by the included AC adapter / charger. When required for a portable application, the internal battery may be used to temporarily

    power the AB6A. A new, fully charged battery will provide power for up to 11 hours if the optional internal pump is not used. If the operational internal pump is used, a new, fully charged battery will provide power for up to 7 hours.

    The internal battery also provides power so that the AB6A will remain operational in the

    event of a temporary power failure. The battery will be recharged automatically once AC power is restored. Allow 12 hours for full charging the battery.

  • Page 26 of 82

    Document Number: 7940010 Rev 1 Draft 4

    NOTE If the power failure is prolonged, the gel cell battery will

    eventually run down. At this time, the AB6A will no longer be operational. However, the data that was collected prior to the battery failure is stored on the SD memory card and is available once the power is restored.

    2.1.4 Environmental Barometric Pressure Sensor The environmental barometric pressure sensor is located on the interface board inside the

    AB6A. The environmental barometric pressure sensor has an accuracy of 2 kPa (0.6 inches Hg

    (0C)) 2.2 Theory of Operation The detectors use proven scintillation technology to detect radon in a gas sample. When

    radon or thoron gases decay, they give off an alpha particle. The alpha particle strikes the scintillation material in the detector which emits a photon of light in response. The photon of light is detected, converted to an electrical pulse, and amplified by the photomultiplier tube (PMT) in the PMT Assembly. This pulse is sent to the interface board electronics for further amplification, waveform shaping, discrimination, and, if appropriate, counting.

    The AB6A uses the count along with various programmed formulae and constants to

    provide data on or calculate the radon levels within the sample. The AB6A also performs many peripheral tasks such as local and remote communications, semi-automated calibration functions, temperature monitoring, etc.

    3.0 OPERATION 3.1 Overview After the Operator has set up the administrative, detector, and method parameters via the

    AB6A LCD, the AB6A system is ready to start taking measurements. Radon and thoron give off alpha particles as they decay. The number of alpha particles

    given off in a specified time period provides information on the radon or thoron activity. Most of Pylon's detectors for the AB6A use a standard, industry proven alpha particle

    detection material - zinc sulphide scintillator. The zinc sulphide alpha scintillator in the detectors scintillate (i.e., Give off a photon of light) when struck by an alpha particle.

  • Page 27 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Information on the various detectors and the sampling methods is provided in the selected detector manual and is beyond the scope of this manual. The description of the process starts with the assumption that the desired detector complete with sample is already attached to the AB6A.

    The photons that are given off by the scintillator in the detector are detected, converted to

    electrical pulses, and amplified by a PMT. The PMT is driven by a high voltage (HV) signal to obtain the optimum detection and amplification of the pulses. The pulses from the PMT are transmitted to the electronic circuit.

    The electronic circuit discriminates the amplitude of the pulses and counts those that are

    above a minimum noise threshold. This information is provided for all measurements. For some methods, the AB6A can calculate the concentration if all of the appropriate

    information such as the detector information is provided. The microprocessor uses these counts, the measurement interval, the detector information (e.g., sensitivity or efficiency), and the background parameters to calculate the level of radioactivity in the units of measure (Bq/m3 or pCi/l) chosen by the Operator.

    NOTE The concentration data provided in the data file will be based on

    the units of measure chosen for the measurement. If the User changes the units of measure for a future measurement, the existing data's unit of measure does not change. In other words, if a measurement was made using Bq/m3, the data file provides the results in Bq/m3. If the User changes the units of measure to pCi/l, the data file for the original measurement still provides the results in Bq/m3.

    It should be noted that the high voltage signal level and the discriminator settings are

    determined during calibration and are chosen to minimize the background noise while providing an optimum signal for detection. The operating point is specific to detector types (e.g., active and passive Lucas type detector cells) and may need to be changed when the detector type is changed. The Operator should not change these values for a given detector type.

    The measurement data may be transferred from the AB6A via the standard USB interface.

    Alternatively, the data may be reviewed via the AB6A. In addition, the Pulse Output signal may be used to provide raw count data to a User supplied counting device such as a data logger.

    3.2 Powering On To turn the AB6A on, press the green on/off switch on the front panel. The switch will light

    to show that the unit is powered on.

  • Page 28 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Several displays including a blank screen will appear during the boot up phase. In addition,

    the piezo buzzer will sound for a period of time during the boot up phase to indicate that the unit is powering on. Should the internal pump be installed, it will run for a short period of time during the boot up phase as part of the self test. Once the initialization is complete, the startup form will appear.

    NOTE The complete boot up process will take approximately 30

    seconds. NOTE The internal clock for the unit will be set at the factory. This time

    may not be valid for your time zone. The internal clock should be reset to the correct time prior to using the system for measurements. The instructions are provided below.

    NOTE It is intended that the AB6A remain on at all times when in use.

    There is no need to turn the unit off when installing or replacing detectors as the software only turns on the PMT when a method is being run. The AB6A AC battery charger should be used when the AB6A will be left on for over 6 hours.

    3.2.1 Powering Off To shutdown and turn off the AB6A, press the power on/off switch to turn the unit off. 3.3 Touchscreen Operation All data entry, commands, etc. that are required to operate the AB6A locally are performed

    using the LCD touchscreen. Select fields, tabs, buttons, etc. by touching them on the LCD with the supplied touchscreen stylus.

    CAUTION IT IS RECOMMENDED THAT THE SUPPLIED

    TOUCHSCREEN STYLUS OR EQUIVALENT NON-MARKING DEVICE BE USED ON THE LCD. SHARP DEVICES CAN DAMAGE THE LCD, WRITING INSTRUMENTS CAN LEAVE A RESIDUE ON THE TOUCHSCREEN, ETC.

  • Page 29 of 82

    Document Number: 7940010 Rev 1 Draft 4

    NOTE A User-supplied standard mouse may be used instead of the

    stylus to operate the AB6A. It would be plugged into the Host USB port on the front panel.

    3.4 AB6A Graphical User Interface (GUI) Overview When used as a stand-alone system, all system programming and operation is performed via

    the LCD on the AB6A. The AB6A uses the Windows Compact 7 operating system as the base program. The AB6A

    program is Graphical User Interface (GUI) driven and uses icons, tabs, forms, popups, etc. to select functions, enter data, provide information, etc.

    The following provides an overview of the GUI. Due to the complexity of the GUI, not all

    information is provided in this overview. For example, most popups are not described. However, there will be popups for data entry when the field is selected. There are also confirmation popups and information popups. These popups are self-explanatory.

    NOTE The "enter" key for the keyboard popups uses the symbol to

    designate this key. 3.4.1 General The Startup form appears upon power up. It consists of two sections. These are: A screen display. A status bar at the bottom of the form. The measurement and calibration forms are divided into 3 sections. These are: A title bar. A screen display. A status bar at the bottom of the form. These sections are described in more detail below. Most of the data entry forms in the AB6A GUI are popups.

  • Page 30 of 82

    Document Number: 7940010 Rev 1 Draft 4

    3.4.1.1 Title Bar The title bar is displayed at the top of the form and besides providing the title of the module

    (Measure or Calibration), provides the minimize and exit buttons as well as 3 different size keyboard icons to allow the User to select the desired size of keyboard for data entry.

    3.4.1.2 Screen Display The screen display is displayed provides the access to the various functions of the AB6A.

    These are discussed in more detail in the following sections. 3.4.1.3 Status Bar The status bar is displayed at the bottom of the form and provides status information, from

    left to right, on the following items: Battery Charge Status. USB Connection. System Locked. Status "LED". A Selection Box to Open Minimized Methods. Time (Date). 3.4.1.3.1 Battery Charge Status This display provides a rough indication of the power remaining in the battery if the AC

    charger is not connected or an indication that the AC charger is connected to the unit and providing power.

    This status is always on display when the status bar is displayed. 3.4.1.3.2 USB Connection A USB icon will be displayed if a USB connection is detected. When both USB connections

    are detected, the icon will also display a "2". This icon will only be displayed if a USB connection is detected. Otherwise, the field will be

    blank. 3.4.1.3.3 System Locked When the monitor operation is password protected, a lock icon will appear to indicate that

    password protection is available. To lock the monitor, click on the lock icon. The screen will go black with a Pylon logo.

    Touching the screen will bring up a popup requesting the password.

  • Page 31 of 82

    Document Number: 7940010 Rev 1 Draft 4

    The lock icon will only be displayed if the monitor operation is password protected.

    Otherwise, the field will be blank. 3.4.1.3.4 Status "LED" The status "LED" icon provides information on the internal self-test functions of the

    monitor. A green colour indicates that there are no self-test problems detected with the AB6A. An orange colour indicates that a non-critical fault has been detected by the self-test. This

    means that radon measurements can still be made but some peripheral functions and measurements may not be accurate or available. For example, if the temperature / relative humidity sensor fails, radon measurements can still be taken. However, the temperature and relative humidity readings are not available or not accurate.

    A red colour indicates that a critical fault has been detected by the self-test. This means that

    radon measurements cannot be made until the fault is corrected. The status "LED" icon may be clicked at any time to obtain more information on the status

    and / or detected fault. Please refer to Section 6.2 - Troubleshooting for more information. This status is always on display when the status bar is displayed. 3.4.1.3.5 Selection Box The selection box allows the User to re-expand the Measure or Calibration forms if they

    have been minimized. This selection box only appears when the Measure or Calibration form is displayed. 3.4.1.3.6 Time (Date) The current time is displayed on the bottom bar. The current date can be obtained by

    clicking on the time. The time is always on display when the status bar is displayed. 3.4.2 Startup Form The startup form is the top level form for the AB6A monitor. It is used to locally access

    and operate all functions and modules in the AB6A software.

  • Page 32 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 2 - Startup Form shows the main display (AB6A) form. The following icons are displayed:

    Methods Files User Setup Touchscreen Calibration About

    Figure 2 - Startup Form 3.4.2.1 Methods This icon is used to access the measure module. The measure module allows the user to

    select a measurement method and define the applicable parameters for the measurement method.

    Please refer to Section 3.5 for more information. 3.4.2.2 Files This icon is used to access the AB6A file manipulation module. The file manipulation

    module allows the user to view, copy, and delete files on the AB6A monitor.

  • Page 33 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Please refer to Section 3.4.3 for more information. 3.4.2.3 User Setup This icon is used to access the user setup module. The user setup form is designed to

    access modules and functions that setup the AB6A including: Monitor. Pump(s). Detector(s). Sealed check source(s). Calibration Module Please refer to Section 3.4.4 for more information. 3.4.2.4 Touchscreen Calibration This icon is used to access the touchscreen calibration module. The touchscreen

    calibration module allows the user calibrate the touchscreen. NOTE The touchscreen requires periodic calibration due to normal

    drift. The touchscreen should be calibrated every 6 months. The Operator should be able to use touchscreen calibration icon on the AB6A to perform the calibration. However, if the touchscreen has drifted out of calibration too far, the Operator may not be able to access the touchscreen calibration module from the AB6A monitor. In these cases, the touchscreen calibration must be initiated by the PC software. Please refer to Section 3.8.

    3.4.2.5 About This icon is used to access the About popup. The popup provides information of the

    software revision and the AB6A serial number. 3.4.3 Files Figure 3 - Files Popup shows the files popup. This popup allows the Operator to: Select the desired measurement file from a drop down menu box. View the measurement file on the monitor screen. Delete the measurement file from the AB6A memory. Delete all of the measurement files from the AB6A memory.

  • Page 34 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Copy the measurement file to a USB memory stick for transfer to a computer for further manipulation and / or storage.

    NOTE

    Measurement files are stored as comma separated value (.csv)

    files to allow them to be easily imported in most spreadsheet programs. Electronic formatting, manipulation, calculation, etc. of the data may then be performed. A text (.txt) version is also available.

    NOTE

    The program requires a specific destination folder on the USB

    memory stick in order to copy the measurement files to the USB memory stick. The folder is named "MeasurementFiles". If this folder does not exist on the USB memory stick, the program will automatically create the folder prior to transferring the file.

    Figure 3 - Files Popup

  • Page 35 of 82

    Document Number: 7940010 Rev 1 Draft 4

    3.4.4 User Setup Figure 4 - User Setup Form shows the user setup form. The user setup form is designed to

    access modules and functions that setup the AB6A. This form provides icons to access the following modules:

    Monitor Setup Pump Setup Detector Setup Source Setup Calibration

    Figure 4 - User Setup Form 3.4.4.1 Monitor Setup This icon is used to access the AB6A monitor setup module. The monitor setup module

    allows the Operator to access the modules that: Set and edit the monitor operating point. Set the desired units of measure (UOM). Set and edit the date and time. Set, change, delete passwords.

  • Page 36 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Enter AB6A monitor owner information. Please refer to Section 3.4.5 for more information. 3.4.4.2 Pump Setup This icon is used the access the AB6A pump setup module. The AB6A pump setup

    module allows the Operator to select a pump option and set the flow rate value for the measurement reports.

    Please refer to Section 3.4.6 for more information. 3.4.4.3 Detector Setup This icon is used the access the detector setup module. The detector setup module allows

    the Operator to add and edit detector information in the AB6A. Please refer to Section 3.4.7 for more information. 3.4.4.4 Source Setup This icon is used the access the sealed check source setup module. The sealed check

    source setup module allows the Operator to add and edit sealed source check source information in the AB6A.

    Please refer to Section 3.4.8 for more information. 3.4.4.5 Calibration This icon is used the access the calibration module. The calibration module allows the

    Operator to access the calibration methods that assist the user for determining or calibrating:

    Active Cell Detectors. Passive Cell Detectors. Trace Environmental Level (TEL) Detectors. Monitor Plateaus. Monitor Efficiencies. Monitor Electronic Backgrounds. Sealed Sources. NOTE The calibration module is password protected to prevent access

    by unauthorized personnel.

  • Page 37 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Please refer to Section 5.0 for more information. 3.4.5 Monitor Setup Figure 5 - Monitor Setup Form shows the monitor setup form. The monitor setup form is

    designed to access modules and functions that setup the AB6A monitor. This form provides icons to access the following modules:

    Operating Point. Units on Measure. Date and Time. Password. Owner.

    Figure 5 - Monitor Setup Form 3.4.5.1 Operating Point This icon is used the access the operating point setup module. The operating point setup

    module allows the Operator to enter the AB6A PMT operating point information that is used during measurements.

  • Page 38 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 6 - Operating Point Form shows the operating point form popup. After the AB6A operating point has been determined (Please refer to Section 5.0), this form allows the Operator to:

    Select either the internal PMT or external PMT via a drop down menu box. Enter / change the high voltage value. The number will be between 0 and 255. The nominal voltage represented by the entered number is displayed

    beside the data entry field. Enter / change the discriminator value. The number will be between 0 and 255. The nominal voltage represented by the entered number is displayed

    beside the data entry field. Enter / change the electronic background value in counts per minute (cpm). Enter / change the AB6A calibration date. Enter / change the AB6A calibration due date. Select to be reminded that the AB6A calibration is due. Save the changes.

    Figure 6 - Operating Point Form

  • Page 39 of 82

    Document Number: 7940010 Rev 1 Draft 4

    3.4.5.2 Units of Measure This icon is used the access the units of measure module. The units of measure module

    allows the Operator to set the units of measure for the following parameters: Concentration. Temperature. Pressure. Flow Rate. Figure 7 - Units of Measure Form shows the units of measure form popup. Select the

    appropriate radio buttons for the desired units of measure.

    Figure 7 - Units of Measure Form 3.4.5.3 Date and Time This icon is used the access the date and time module. The date and time module allows

    the Operator to access the operating system module that: Sets the current date and time. Selects the appropriate time zone.

  • Page 40 of 82

    Document Number: 7940010 Rev 1 Draft 4

    NOTE Setting the date and time requires a reboot of the unit. This is

    performed automatically. Follow the instructions on the popups when setting the date and time.

    3.4.5.4 Password This icon is used the access the password module. The password module allows the

    Operator to: Set, change, and delete the Monitor password in the AB6A. If available and authorized, change the Calibration password in the AB6A. The Monitor password restricts general access to the monitor to: Prevent unauthorized changes to measurement settings. Ensure that measurement run data is nor visible to unauthorized personnel. The Calibration password restricts access to the calibration module to prevent

    unauthorized use of the calibration methods. Figure 8 - Password Form shows the password form popup when the Operator accesses

    the password module using the Calibration password. If the Monitor password is entered, only the monitor password fields are shown.

    If a Monitor password has been entered previously, a popup requesting the current

    password will appear. NOTE A monitor password must be entered in order to allow the

    Operator to enter the calibration password in order to allow changes to the calibration password.

    Enter the desired password in the first field and repeat it in the second field to verify the

    entry. Then click the save button. The Monitor password may be deleted by making the password fields blank and then

    clicking the save button.

  • Page 41 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 8 - Password Form 3.4.5.5 Owner The Owner icon is used the access the owner module. The owner module allows the

    Operator to enter owner information. Figure 9 - Owner Form shows the owner form popup. The Operator may enter

    information on the owner of the AB6A such as the company name in the field.

  • Page 42 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 9 - Owner Form 3.4.6 Pump Setup The pump setup icon allows the Operator to access the pump setup module. This modules

    allows the Operator to set up the desired pump setting for the measurement run. Figure 10 - Pump Setup Form shows the pump setup form popup. The Operator can

    select one of three pump selection options, as follows: Internal Pump (If installed). External Pump. No Pump Selected. It also allows the Operator to record the flow rate that was used in the measurement.

  • Page 43 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 10 - Pump Setup Form The setup information is used by the program to set up options and constraints for

    methods. E.g., The internal pump must be installed and selected in order to run the QUASI method.

    This information is also provided in the reports to provide a record of the measurement setup.

    NOTE During the initialization phase, the AB6A will automatically

    detect if the internal pump is installed. If the internal pump is not installed, the Internal Pump selection will be greyed out.

    NOTE The flow rate must be manually setup for both the internal and

    external pumps. The AB6A does not perform any automated measurement of the flow rate.

  • Page 44 of 82

    Document Number: 7940010 Rev 1 Draft 4

    NOTE The Pump Setup module is also accessible by selecting the Edit

    button in the Measure - Setup form for those measurement methods that allow the use of a pump.

    Please refer to Pylon Document # 7940011 - Manual, Operator's, Internal Pump; AB6A

    for more information on the installation, setup, and operation of the optional internal AB6A pump.

    3.4.7 Detector Setup 3.4.7.1 Detector Selection Figure 11 - Detector Setup Selection Form shows the detector setup selection form

    popup. This form allows the Operator to: Select the desired detector from a drop down menu box if the detector has already

    been added to the file. Add a new detector to the file. Edit the information of the selected detector. Delete a detector from the file.

    Figure 11 - Detector Setup Selection Form

  • Page 45 of 82

    Document Number: 7940010 Rev 1 Draft 4

    3.4.7.2 Detector Editing Figure 12 - Edit Detector Form shows the detector editing form popup. When adding a

    new detector to the file, this form allows the Operator to enter the following information: The detector model number. The detector serial or asset number. The detector efficiency. The detector background. The detector volume. The detector sensitivity. The detector sensitivity units of measure. The system background. The detector calibration date. The detector calibration due date. When editing a detector file, the Operator can change any of the above parameters except

    for the model and serial/asset numbers. NOTE Some fields can updated automatically. A popup will ask the

    Operator if the field can be updated automatically.

  • Page 46 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 12 - Edit Detector Form Depending on the measurement method, some of this information is required to perform the

    automated calculations. Any available information is also included in the reports if the detector is selected for a measurement.

    3.4.8 Source Setup 3.4.8.1 Sealed Source Selection Figure 13 - Source Setup Selection Form shows the source selection form. This form

    allows the Operator to: Select the desired sealed check source from a drop down menu box if the check

    source has already been added to the file. Add a new sealed check source to the file. Edit the Activity of the selected sealed check source. Delete a sealed check source from the file.

    Figure 13 - Source Setup Selection Form

  • Page 47 of 82

    Document Number: 7940010 Rev 1 Draft 4

    3.4.8.2 Sealed Source Editing Figure 14 - Edit Source Form shows the edit source form. When adding a new sealed

    check source to the file, this form allows the Operator to enter the following information: The sealed check source model number. The sealed check source serial or asset number. The Activity of the sealed check source. The Efficiency of the sealed check source. When editing the information for an existing source, only the Activity and Efficiency fields

    may be modified.

    CAUTION THE ACTIVITY OF THE SEALED CHECK SOURCE MUST

    BE ENTERED IN DISINTEGRATIONS PER MINUTE (DPM). IF THE ACTIVITY OF THE SEALED CHECK SOURCE IS NOT IN DPM, THE ACTIVITY UNIT OF MEASURE MUST BE CONVERTED TO DPM TO BE ENTERED IN THE FILE. OTHERWISE ALL AUTOMATED CALCULATIONS WILL BE INCORRECT.

  • Page 48 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 14 - Edit Source Form This information is required to perform the automated calculations. Any available

    information is also included in the reports if the source is selected for a method. 3.5 AB6A Methods Overview In order to facilitate the ease of use of the AB6A, maintain flexibility, and meet

    measurement criteria, Pylon has developed a number of different Operator modifiable, pre-programmed methods that the Operator can select and use to make a wide variety of measurements.

    A common GUI template has been developed for the methods. Accessible parameters are

    defined for each method. This section provides an overview of this common GUI template for all of the parameters. More details on the parameters that are accessible for each method are provided in Sections 3.7 and 5.0 as well as in the individual detector manuals as follows:

    7940002 Manual; Operator's; Active Cells; AB6A 7940003 Manual; Operator's; Passive Cell; AB6A 7940005 Manual; Operator's; Trace Environmental Level System 7940006 Manual; Operator's; Kit; Filter Detector; AB6A Each method form will contain the following tabs: Run Setup In addition, depending on the type of method, one of the following tabs may be available: Detector Source 3.5.1 Run Tab Figure 15 - Measure Run Tab Form shows the run tab form. The Operator starts and stop

    the measurement in this form by clicking the Start / Stop button.

  • Page 49 of 82

    Document Number: 7940010 Rev 1 Draft 4

    CAUTION THE PMT IS SENSITIVE TO LIGHT AND MAY BE

    IRREVOCABLY DAMAGED IF IT IS EXPOSED TO LIGHT WHEN THE HIGH VOLTAGE HAS BEEN APPLIED TO THE PMT. THE HIGH VOLTAGE IS APPLIED TO THE PMT WHEN THE START BUTTON IS CLICKED (OR WHEN THE MEASUREMENT IS STARTED WHEN A DELAYED START IS SELECTED)! ENSURE THAT A DETECTOR HAS BEEN ATTACHED TO THE PMT OR THE PMT CAP HAS BEEN INSTALLED PRIOR TO CLICKING THE START BUTTON!

    Depending on the selected method and Operator programmed parameters for the method,

    the form may also provide information on the measurement such as: Specified measurement start. (Start / Stop button indicates "Delay" and turns green.) Delayed measurement start. (Message with count down.) The filename of the measurement data file for this run. The current number of counts detected for the current interval. The concentration calculated for the last interval. The current interval number. The total number of intervals. The method name. The interval length. The time remaining for the measurement. The pump flow rate. The pump source (internal or external). The temperature. The relative humidity. The barometric pressure. The dew / frost point. During the current measurement run, the Run tab form also displays the measurement data

    for each interval. This allows the Operator to review the data during the measurement run. A maximum of 500 intervals can be displayed although the file can contain many more intervals.

    The display is updated every second.

  • Page 50 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 15 - Measure Run Tab Form 3.5.2 Setup Tab Figure 16 - Measure Setup Tab Form shows the setup tab form. The Operator uses this

    form to select, when available, the desired method and to program the various parameters for the method.

  • Page 51 of 82

    Document Number: 7940010 Rev 1 Draft 4

    Figure 16 - Measure Setup Tab Form The Operator selects the desired measurement method from the drop down menu box. The Operator may then select and modify the selected method parameters. The following

    provides generic information for each of the parameters. NOTE As indicated in section 3.5 above, a generic template was

    developed for the method parameters. The template has been tailored for each method so that only those parameters that are applicable to a method are available. Any parameters that are not available or required for a method will be greyed out and inaccessible to the Operator.

    NOTE The applicable parameter information is recorded in the

    measurement report.

  • Page 52 of 82

    Document Number: 7940010 Rev 1 Draft 4

    PARAMETER DESCRIPTION / PURPOSE / COMMENTS Save (Button) This button saves any changes to the method.

    To Defaults

    (Button) This button restores all of the parameters in the method to the defaults.

    Reload (Button) In rare instances, when a method selection is changed, the Setup form does not load the appropriate accessible parameters for the selected method. The Reload button may be used to load the appropriate accessible parameters for the selected method.

    File The Operator may enter their own filename in this field. If left blank, the filename will default to one based on the date and time that the measurement was started.

    Operator The Operator may enter their name or identification code in this field to identify who performed the measurement.

    PMT The Operator may select the internal or, if available, the separately purchased external PMT for the measurements. The AB6A operating point HV and DISC values will automatically change to the programmed values for the selected PMT. (Please refer to Section 3.4.5.1.)

    Delay Should the Operator wish to delay the start of a measurement (e.g., For 3.5 hours after a grab sample has been obtained), the amount of delay may be entered in the delay field and appropriate radio buttons may be selected for the time units of measure. The start button in the run tab initiates the run. Once the start button is clicked and the delay is set, as indicated in Section 3.5.1, there will be a message indicating that there is a delay in starting and a countdown clock.

    Interval Length The Operator shall enter the desired interval length in this field and select the appropriate time units of measure.

  • Page 53 of 82

    Document Number: 7940010 Rev 1 Draft 4

    PARAMETER DESCRIPTION / PURPOSE / COMMENTS Number of Intervals

    to Discard In a number of measurements, it is recommended that at least the data in the first interval be discarded and not used in any calculations. This reduces the possibility of fluorescence of the sc