8/18/06 Gxxxxxx
Introduction to CalibrationIntroduction to CalibrationIntroduction to CalibrationIntroduction to Calibration
Brian O’Reilly
SciMon Camp 2006
Brian O’Reilly
SciMon Camp 2006
8/18/06 Gxxxxxx
Frequency Domain CalibrationFrequency Domain Calibration
We model the DARM loop in MATLAB Compare this model to measurements of the open-
loop gain, electronics in the Actuation and Sensing chains and DC value of the Actuation.
Optical and loop gain are tracked by time-dependent coefficients which are generated on minute or second time scales.
These coefficients are used to propagate measurements at t0 to other times.
We model the DARM loop in MATLAB Compare this model to measurements of the open-
loop gain, electronics in the Actuation and Sensing chains and DC value of the Actuation.
Optical and loop gain are tracked by time-dependent coefficients which are generated on minute or second time scales.
These coefficients are used to propagate measurements at t0 to other times.
8/18/06 Gxxxxxx
A(f)
DD(f)
(t)C D0(f)
DARM_ERR
+
s=h(t)+n(t)
DARM_CTRL
sres=(Lx-Ly)/L
+
DARM_CTRL_EXC
8/18/06 Gxxxxxx
ActuationActuationActuationActuation
Ax(f) Ay(f)
+
++
EXCx(t)
kykx
• DARM feeds back to the ETMs.• Measuring the actuation has typically been the least accurate and most angst-ridden part of the calibration.
8/18/06 Gxxxxxx
Actuation FunctionActuation FunctionCalibrate the ASQ signal for a simple Michelson This establishes the length scale in AS_Q counts.
Use it to calibrate ITMs:
Use single arms to calibrate ETMs with ITMs
8/18/06 Gxxxxxx
Actuation FunctionActuation Function
Treat mass as a simple pendulum.
Knowing the DC value we can set the scale for the transfer function.
Methods for measuring DC value explicitly have also been tried: sneaky poles
Treat mass as a simple pendulum.
Knowing the DC value we can set the scale for the transfer function.
Methods for measuring DC value explicitly have also been tried: sneaky poles
8/18/06 Gxxxxxx
Actuation FunctionActuation Function
?
8/18/06 Gxxxxxx
Compensate the ElectronicsCompensate the Electronics
8/18/06 Gxxxxxx
The Payoff… Small ErrorsThe Payoff… Small Errors
8/18/06 Gxxxxxx
Digital FiltersDigital Filters
Know them perfectly?
8/18/06 Gxxxxxx
The Input MatrixThe Input Matrix
8/18/06 Gxxxxxx
Sensing FunctionSensing Function
Model as a cavity pole Have to understand the
sensing electronics chain Photodiode, Whitening,
Demodulation, Anti-Aliasing etc.
How well do we know the cavity pole?
How well do we know C(f)? Not directly measured.
Model as a cavity pole Have to understand the
sensing electronics chain Photodiode, Whitening,
Demodulation, Anti-Aliasing etc.
How well do we know the cavity pole?
How well do we know C(f)? Not directly measured.
8/18/06 Gxxxxxx
Open Loop Gain DiscrepancyOpen Loop Gain DiscrepancyL1 H1
5-10% error on response at 2 kHz
8/18/06 Gxxxxxx
Model InputsModel Inputs
8/18/06 Gxxxxxx
Frequency Domain CalibrationFrequency Domain Calibration
Measure Open-Loop Gain at a reference time t0
G0(f) = A(f)CD0(f)DD(f)
h(f,t) = RDERR(f,t)DERR(f,t)
Similar equations for AS_Q
Measure Open-Loop Gain at a reference time t0
G0(f) = A(f)CD0(f)DD(f)
h(f,t) = RDERR(f,t)DERR(f,t)
Similar equations for AS_Q€
DERRR ( f , t) =1+ G( f , t)
DC ( f , t)=
1+ γ (t)0G ( f )
D0γ(t)C ( f )
8/18/06 Gxxxxxx
PropagatePropagate
This value stays within ~5% of unity, barring any problems with the code.
This value stays within ~5% of unity, barring any problems with the code.
8/18/06 Gxxxxxx
8/18/06 Gxxxxxx
8/18/06 Gxxxxxx
Errors on the ResponseErrors on the Response
By breaking the error down into these components we identify problem areas.
By breaking the error down into these components we identify problem areas.
8/18/06 Gxxxxxx
8/18/06 Gxxxxxx
8/18/06 Gxxxxxx
FinallyFinally
After diligent work we feel we can control calibration errors to the level of 5-10%.
Doing better than this is hard, but: 15Mpc/10 = 1.5Mpc = range of L1 during S2!!
Other ways to calibrate: HEPI or Tidal Actuators VCO Photon Calibrator time-domain h(t)
Calibrating eLIGO or advLIGO will present a new set of challenges.
After diligent work we feel we can control calibration errors to the level of 5-10%.
Doing better than this is hard, but: 15Mpc/10 = 1.5Mpc = range of L1 during S2!!
Other ways to calibrate: HEPI or Tidal Actuators VCO Photon Calibrator time-domain h(t)
Calibrating eLIGO or advLIGO will present a new set of challenges.