Signet pH/ORP Transmitter 10.20 pH 25.0C ENTER Signet 8750-1 pH/ORP Transmitter 2. Specifications General Compatible electrodes: Signet 3-2720 pH/ORP Preamplifier/3-271X Electrodes Accuracy: ± 0.03 pH, ± 2 mV ORP Enclosure: Rating: NEMA 4X/IP65 front Case: PBT Panel case gasket: Neoprene Window: Polyurethane coated polycarbonate Keypad: Sealed 4-key silicone rubber Weight: Approx. 325g (12 oz.) Display: Alphanumeric 2 x 16 LCD Contrast: User selected, 5 levels Update rate: 1 second Electrical Power: 12 to 24 VDC ±5%, regulated, 21 mA max. Sensor input range: pH: 0.00 to 14.00 pH temp. (pH only) 3K Balco, -25 to 120°C (-13 to 248°F) ORP: -1000 to +2000 mV, isolated (10KI.D. resistance T+, T-) Current output: 4 to 20 mA, isolated, fully adjustable and reversible CAUTION! Remove power to unit before wiring input and out- put connections. Follow instructions carefully to avoid personal injury. 1. Installation ProcessPro transmitters are available in two styles: panel mount and field mount. The panel mount is supplied with the necessary hardware to install the transmitter. This manual includes complete panel mounting instructions. Field mounting requires a separate mounting kit. The 3-8050 Universal kit enables the transmitter to be installed virtually anywhere. Detailed instructions for field installation options are included with the 3-8050 Universal kit. 1.1 Panel Installation 1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch. 2. Place gasket on instrument, and install in panel. 3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument. 4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE. Press quick-clips outward and remove. 106 mm (4.2 in.) 42 mm (1.7 in.) 64 mm (2.5 in.) SIDE VIEW 92 mm (3.6 in.) 97 mm (3.8 in.) 56 mm (2.2 in.) 41 mm (1.6 in.) Optional Rear Cover Field Mount Panel Mount FRONT VIEW 96 mm (3.8 in.) 96 mm (3.8 in.) quick-clips gasket panel terminals mounting bracket latch Output - Output + System Pwr Loop - System Pwr Loop + 2 1 4 3 Sensr Gnd (SHIELD) Sensr IN (RED) Sensr V+ (BLACK) 7 6 5 SIDE VIEW Field Mount & Panel Mount Panel Mount Installation Detail 82 mm (3.23 in.) 8050 Max loop impedance: 50 max. @ 12 V 325 max. @ 18 V 600 max. @ 24 V Update rate: 0.5 seconds Accuracy: ±0.03 mA @ 25°C, 24 V Open-collector output: Isolated, 50 mA sink or source, 30 VDC max. pull-up voltage Programmable for: • Hi or Lo w/adjustable hysteresis • Proportional pulse (400 pulses per minute maximum) Environmental Operating temperature: -10 to 70°C (14 to 158°F) Storage temperature: -15 to 80°C (5 to 176°F) Relative humidity: 0 to 95%, non-condensing Maximum altitude: 2000 m (6562 ft) Insulation category: II Pollution degree: 2 Standards and Approvals CE, UL listed Immunity: EN50082-2 Emissions: EN55011 Class B Manufactured under ISO 9001 and ISO 14001 *3-8750.090-1* 3-8750.090-1 Rev. F 3/06 English English

Upload: others

Post on 30-Jun-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

Signet pH/ORPTransmitter

10.20 pH25.0C

ENTER

Signet 8750-1 pH/ORP Transmitter

2. Specifi cationsGeneralCompatible electrodes: Signet 3-2720 pH/ORP Preamplifi er/3-271X ElectrodesAccuracy: ± 0.03 pH, ± 2 mV ORPEnclosure:• Rating: NEMA 4X/IP65 front• Case: PBT• Panel case gasket: Neoprene• Window: Polyurethane coated polycarbonate• Keypad: Sealed 4-key silicone rubber• Weight: Approx. 325g (12 oz.)Display:• Alphanumeric 2 x 16 LCD• Contrast: User selected, 5 levels• Update rate: 1 second

Electrical• Power: 12 to 24 VDC ±5%, regulated, 21 mA max.Sensor input range: • pH: 0.00 to 14.00 pH• temp. (pH only) 3K Balco, -25 to 120°C (-13 to 248°F)• ORP: -1000 to +2000 mV, isolated (10KΩ I.D. resistance T+, T-)Current output: • 4 to 20 mA, isolated, fully adjustable and reversible

CAUTION!• Remove power to unit before wiring input and out-

put connections.• Follow instructions carefully to avoid personal injury.

1. Installation ProcessPro transmitters are available in two styles: panel mount and fi eld mount. The panel mount is supplied with the necessary hardware to install the transmitter. This manual includes complete panel mounting instructions.Field mounting requires a separate mounting kit. The 3-8050 Universal kit enables the transmitter to be installed virtually anywhere. Detailed instructions for fi eld installation options are included with the 3-8050 Universal kit.

1.1 Panel Installation 1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive

template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch. 2. Place gasket on instrument, and install in panel. 3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument. 4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE. Press quick-clips outward and remove.

106 mm (4.2 in.)

42 mm (1.7 in.)

64 mm (2.5 in.)

SIDE VIEW

92 mm(3.6 in.)

97 mm(3.8 in.)

56 mm (2.2 in.)41 mm

(1.6 in.)

OptionalRearCover

Field Mount Panel Mount

FRONT VIEW

96 mm(3.8 in.)

96 mm(3.8 in.)

quick-clips

gasketpanel

terminalsmountingbracket

latchOutput -

Output +

System Pwr

Loop -

System Pwr

Loop +2

1

4

3

Sensr Gnd

(SHIELD)

Sensr IN

(RED)

Sensr V+

(BLACK)

7

6

5

SIDE VIEWField Mount &Panel Mount

Panel MountInstallation Detail

82 mm(3.23 in.) 8050

• Max loop impedance: 50 Ω max. @ 12 V 325 Ω max. @ 18 V 600 Ω max. @ 24 V• Update rate: 0.5 seconds• Accuracy: ±0.03 mA @ 25°C, 24 VOpen-collector output:• Isolated, 50 mA sink or source, 30 VDC max. pull-up voltage• Programmable for: • Hi or Lo w/adjustable hysteresis • Proportional pulse (400 pulses per minute maximum)

Environmental• Operating temperature: -10 to 70°C (14 to 158°F)• Storage temperature: -15 to 80°C (5 to 176°F)• Relative humidity: 0 to 95%, non-condensing• Maximum altitude: 2000 m (6562 ft)• Insulation category: II• Pollution degree: 2

Standards and Approvals• CE, UL listed• Immunity: EN50082-2• Emissions: EN55011 Class B• Manufactured under ISO 9001 and ISO 14001

*3-8750.090-1*

3-8750.090-1 Rev. F 3/06 English

English

Page 2: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

2 8750-1 pH/ORP Transmitter

Output -

Output +

System PwrLoop -System PwrLoop +

2

1 4

3

EarthGND

Sensr Gnd(SHIELD)

V-(BLACK)

V+(RED)

8

7

6

5

Temp -(WHITE)

Temp +(GREEN)

Iso. GND(BLUE)

mV Input(BROWN)

12

11

10

9

3. Electrical Connections

Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.

Wiring Procedure 1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end. 2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws. 3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out. 4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.

Wiring Removal Procedure 1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws. 2. When fully open, remove wire from terminal.

Wiring Tips:• Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.• Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.• Seal cable entry points to prevent moisture damage.• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.

2

1

Terminals 1 and 2: Loop Power12-24 VDC ±10% system power and current

loop output.Max. loop impedance:

50 Ω max. @ 12 V325 Ω max. @ 18 V600 Ω max. @ 24 V

Internal open-collectoroutput circuit

Output --

Output +Isolation

S

D

15Ω

Terminals 3 and 4: Open-collector OutputA transistor output, programmable in CALIBRATE menu as:

• High or Low setpoint with adjustable hysteresis• Proportional pulse up to 400 pulses per minute• May be disabled (Off) if not used.

Terminals 9-12: Sensor connections 9-10 are mV (pH or ORP) input from the electrode. 11-12 are Temperature input from the electrode

Terminals 5-8: Preamplifi er power and grounds5-6 are DC voltages from the 8750 to power the

preamplifi er.7-8 are ground terminals for the sensor and for earth

ground.

Page 3: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

8750-1 pH/ORP Transmitter 3

3.2 System Power/Loop Connections

12111098765

White (Temp)Green (Temp)Blue (ISO GND)Brown (mV input)

Silver (Shield)Black (V-)Red (V+)

Preamp:Signet 2720

* (connection of terminal 8 to Earth GND may reduce electrical interference)

Terminals

*

Twist-LockElectrodes

DryLocElectrodes

Preamp:Signet 2760-1Signet 2760-2

Stand-alone application, no current loop used

Transmitter

Terminals

Power

Supply

12-24 VDC

Connection to a PLC with built-in power supply

Internal

PLC Connection

Connection to a PLC/Recorder, separate supply

Example: Two transmitters connected to PLC/Recorder with separate power supply

-

+

PLC or Recorder

Loop Input

4-20 mA

System Pwr Loop -

System Pwr Loop +

4

3

2

1

Transmitter

Terminals

Power

Supply

12-24 VDC

-

+

+

-

PLC or Recorder

Channel 1

4-20 mA in

Channel 2

4-20 mA in

+

-

-

+Transmitter 1

Terminals

System Pwr Loop -

System Pwr Loop +

Transmitter 2

Terminals

PLC Terminals

Loop Input

4-20 mA

TransmitterTerminals

+-

Power

Supply

12-24 VDC

+-

Power

Supply

12-24 VDC

-

+

System Pwr Loop -

System Pwr Loop +

4

3

2

1

System Pwr Loop -

System Pwr Loop +

4

3

2

1

System Pwr Loop -

System Pwr Loop +

4

3

2

1

4

3

2

1

3.1 Sensor Input Connections

Transmitter

≤ 400 ft. (122 m)

Signet pH/ORPTransmitter

10.20 pH25.0C

ENTER

Page 4: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

4 8750-1 pH/ORP Transmitter

Signet pH/ORPTransmitter

10.20 pH25.0C

ENTER

VIEW Menu• During normal operation, ProcessPro displays the VIEW menu.• When using the CALIBRATE or OPTIONS menus, ProcessPro will return to the VIEW menu if no activity

occurs for 10 minutes.• To select the item you want displayed, press the UP or DOWN arrow keys. The items will scroll in a

continuous loop. Changing the display selection does not interrupt system operations.• No key code is necessary to change display selection.• Output settings cannot be edited from the VIEW menu.

• Proportional PulsingThe Open Collector output will generate a 100 mS pulse at the rate defi ned by settings in the CALIBRATE menu (see page 6) In the example below:• The output will be 0 pulses/min. at pH values less than 5.0.• The output will be 50 pulses/min. at 7.5 pH.• The output will be 100 pulses/min. at pH values above 10 pH.

3.3 Open Collector OutputThe Open Collector output can be used as a switch that responds when the process value moves above or below a setpoint, or it can be used to generate a pulsing signal with a rate proportional to the process value.

• Low:Output triggers when process variable is less than the setpoint. The output will relax when the process moves above the setpoint plus the hysteresis value.

• High:Output triggers when process variable is greater than the setpoint. The output will relax when the process variable moves below the setpoint plus the hysteresis value.

Hysteresis

Time

Low Setpoint

Process

Relay energizedRelay relaxed

Hysteresis

Time

High Setpoint

Process

105

Open C

ollec

tor O

utput

Rate:

0 to 1

00 Puls

es/m

in.

endpoint

Puls

e ra

te

0 pulses

100 pulses

Starting point

7.00 pH 12.6 ºC

Monitor the Temperature input from the sensor.This is the permanent view display.

All of the displays below are temporary. The permanent display will return after ten minutes.

Input: 307 mV

Loop Output: 14.16 mA

Last CAL: 06-30-01 >

EASY CAL: >

Monitor the millivolt input from the electrode. Use this display to determine the relative condition of your electrode during periodic calibration. (7 pH buffer = 0 mV, ±50 mV) Monitor the 4-20 mA Loop output.

Monitor date for scheduled maintenance or date of last calibration.

Easy Cal is the fastest and simplest periodic calibration method. Requires 4 pH, 7 pH and 10 pH. (Any two)

View Menu for pHDisplay Description

Page 5: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

8750-1 pH/ORP Transmitter 5

Output Setpnt:10.00 pH

Output Setpnt:00.00 pH

ENTER

Output Setpnt:00.00 pH

Output Setpnt:09.00 pH

Output Setpnt:Saving

Output Setpnt: 9.00 pH >

10.0 pH >Output Setpnt:

Temperature >Set:

CALIBRATE: ----Enter Key Code

CALIBRATE: *---Enter Key Code

CALIBRATE: **--Enter Key Code

CALIBRATE: ***-Enter Key Code

Temperature >Set:

Notes on Steps 5 and 6:• All output functions remain active during editing.• Only the fl ashing element can be edited.• RIGHT ARROW key advances the fl ashing element in a continuous loop.• Edited value is effective immediately after pressing ENTER key.• If no key is pressed for 10 minutes unit will restore the last saved value and return to step 3.• Step 6 (pressing ENTER key) always returns you to Step 3.• Repeat steps 3-6 until all editing is completed.

Notes on Step 2:If no key is pressed for 5 minutes while display is showing "Enter Key Code", the display will return to the VIEW menu.

Notes on Steps 3 and 4:• Refer to pages 6 and 7 for complete listing of menu items and their use.• From the Step 3 display, pressing the UP and DOWN keys simultaneously will return the display to the VIEW menu.• If no key is pressed for 10 minutes, display will also return to the VIEW menu.

ProcessPro Editing Procedure:Step 1. Press and hold ENTER key: • 2 seconds to select the CALIBRATE menu • 5 seconds to select the OPTIONS menu.Step 2. The Key Code is UP-UP-UP-DOWN keys in sequence. • After entering the Key Code, the display will show the fi rst item in the selected menu.Step 3. Scroll menu with UP or DOWN arrow keys.Step 4. Press RIGHT ARROW key to select menu item to be edited. • The fi rst display element will begin fl ashing.Step 5. Press UP or DOWN keys to edit the fl ashing element. • RIGHT ARROW key advances the fl ashing element.Step 6. Press ENTER key to save the new setting and return to Step 3.

OPTIONS

CALIBRATE

VIEW

2s 5s

Press &hold for access:

ENTER

Step 5

Step 6

Notes on Step 1:• The View Menu is normally displayed.• The CALIBRATE and OPTIONS menus require a KEY CODE.

Step 4

First item in CALI-BRATE menu

Step 3

Step 3: Finished Editing?Press the UP and DOWN keys simultaneously after saving the last setting to return to normal operation.

Press the UP and DOWN keys simultaneously while any element is fl ashing. This will recall the last saved value of the item being edited and return you to Step 3.

Step 5: Made an Error?

Page 6: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

6 8750-1 pH/ORP Transmitter

Provides a maximum 25ºC offset to match temperature measurement to external reference.

Applies a linear offset to the pH measurement. The ideal value is the average pH of your application. (A sample of your application at process temperature is recommended.)

Applies a slope to the pH measurement. The slope value and the standard value must be at least 2 pH units apart.The ideal values are the minimum and maximum values of your process.

Select the minimum and maximum values for the 4-20 mA Current loop output.

Select pH or Temperature as the source for the Open Collector Output.

Select the mode of operation for the Open Collector output: High, Low or proportional Pulse.The signal may be disabled if not in use.

In Low or High Mode, the Open Collector output will be activated when the pH reaches this value.

In Low or High mode, the Open Collector output will be deactivated at Setpoint ± Hysteresis, depending on High or Low Setpoint selection. (See details on page 4.)

If the output is in PULSE mode, set the start and end point of the range and also set the maximum pulse rate. (The maximum PULSE rate setting is 400 pulses per minute.)

The combined Output Range and Pulse rate settings shown here indicate:"Start pulsing when the pH value is 4 and increase the pulse rate up to the maximum of 120 pulses per minute when the pH value reaches 8"

Use this “note pad” to record important dates, such as annual recertifi cation or scheduled maintenance.

Set: Temperature >

Set: Standard >

Set: Slope > Loop Range: pH 0.00 → 14.00 >

Output Source: pH >

Output Mode: Low > Output Setpnt: 4.00 pH >

Output Hys: 0.50 pH >

Output Range: pH 4.00 → 8.00 > Output PlsRate: 120 Pulses/min > Last CAL: 12-9-06 >

Calibrate Menu for pHDisplay

(Factory settings shown) Description

Page 7: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

8750-1 pH/ORP Transmitter 7

Contrast:

3 >

Averaging:

Off >

Output Active:

Low >

Loop 1 Adjust:

4.00 mA >

Loop 1 Adjust:

20.00 mA >

Test Loop1:

>

Test Output1:

>

Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. Select lower contrast if the display is in warmer ambient surroundings.

OFF provides the most instantaneous output response to changes in input value. LOW averaging = 4 seconds, HIGH averaging = 8 seconds of input signal.Longer averaging produces more stable display and output response.

Active HIGH: This setting is used to turn a device (pump, valve) ON at the setpoint. Active LOW: This setting is used to turn a device OFF at the setpoint.

Adjust the minimum and maximum current output. The display value represents the precise current output.Adjustment limits:• 3.80 mA < 4.00 mA > 5.00 mA • 19.00 mA < 20.00 mA > 21.00 maUse this setting to match the 8750 loop output to any external device.

Press UP or DOWN keys to manually order any output current value from 3.6 mA to 21.00 mA to test current loop output.

Press UP or DOWN keys to manually toggle the state of the open collector output.

Options Menu for pH

Display(Factory settings shown) Description

EASY CAL Procedure - pH• This procedure simplifi es system calibration using standard 4.0, 7.0, 10.0 pH buffers only. If these pH buffers are not available,

calibrate the system via the CALIBRATE menu, using the STANDARD and SLOPE settings.• Access the CALIBRATE menu and set sensor temperature before performing EASY CAL for new electrode installations.• Access EASY CAL menu from the view menu.

Theoretical mV values pH @ 25°C mV 2 +296 3 +237 4 +177 5 +118 6 +59 7 +0 8 -59 9 -118 10 -177 11 -237 12 -296

To Calibrate: Response: To Accept:

Place electrode tip in first pH bufferpH 7.0 = 0 mVpH 4.0 = 177pH 10 = -177Limit ± 50 mV

Allow for stabilization to accept

Place electrode tipin second pH buffer.

Allow for stabilization Press to acceptsecond buffer calibration.

ENTER

ENTER

* 7.00 pH-005 mV

* 6.90 pH-005 mV

* 3.93 pH+179 mV

1

2

30 seconds*

30 seconds*

Press UP, UP, UP, DOWN buttons in sequence to enter menu, XXXX will appear during code entry.

EASY CAL: ----Enter Key Code

Place Sensor inpH Buffer #1

Place Sensor inpH Buffer #2

* 6.90 pH-005 mV

* 3.93 pH+179 mV

* 4.00 pH+179 mV

_ _ _ _

To exit menus and return to VIEW press UP and DOWN button at the same time

Display returns to VIEWMenu in 10 minutes or when ENTER is pressed

Good Easy CalPress <ENTER>

Page 8: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

8 8750-1 pH/ORP Transmitter

Set:

Standard >

Set:

Slope >

Loop Range: mV

-1000 → +1000 >

Output Mode:

Off >

Output Setpnt:

-500 mV >

Output Hys:

10 mV >

Output Range: mV

-500 → +500 > Output PlsRate:

120 pulses/min >

Last Cal:

06-30-01 >

Calibrate Menu for ORPDisplay

(Factory settings shown) Description

Applies a linear offset to the ORP measurement. The ideal value is the average pH of your application. (A sample of your application at process temperature is recommended.)

Applies a slope to the ORP measurement. The slope value and the standard value must be at least 120 mV apart. The ideal values are the minimum and maximum values of your application. Select the minimum and maximum ORP values for the 4-20 mA Current loop output. Minimum range is -1000 mV; Maximum range is +2000 mV.

Select the desired mode of operation for the Open Collector output: High, Low or proportional Pulse. The signal may also be disabled if not in use.

In Low or High Mode, the Open Collector output will be deactivated when the ORP reaches this value.

In Low or High mode, the Open Collector output will be deactivated at Setpoint ± Hysteresis, depending on High or Low Setpoint selection. (See details on page 4.)

If the output is in PULSE mode, set the start and end point of the range and also set the maximum pulse rate. (The maximum PULSE rate setting is 400 pulses per minute.)

The combined Output Range and Pulse rate settings shown here indicate:"Start pulsing when the ORP value is -500 and increase the pulse rate up to the maximum of 120 pulses per minute when the ORP value reaches +500"

Use this “note pad” to record important dates, such as annual recertifi cation or scheduled maintenance.

Page 9: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

8750-1 pH/ORP Transmitter 9

Contrast:

3 >

Averaging:

Off >

Output Active:

Low >

Loop 1 Adjust:

4.00 mA >

Loop 1 Adjust:

20.00 mA >

Test Loop1:

>

Test Output1:

>

Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. Select lower contrast if the display is in warmer ambient surroundings.

OFF provides the most instantaneous output response to changes in input value. LOW averaging = 4 seconds, HIGH averaging = 8 seconds of input signal.Longer averaging produces more stable display and output response.

Active HIGH: This setting is used to turn a device (pump, valve) ON at the setpoint. Active LOW: This setting is used to turn a device OFF at the setpoint.

Adjust the minimum and maximum current output. The display value represents the precise current output.Adjustment limits:• 3.80 mA < 4.00 mA > 5.00 mA • 19.00 mA < 20.00 mA > 21.00 maUse this setting to match the 8750 loop output to any external device.

Press UP or DOWN keys to manually order any output current value from 3.6 mA to 21.00 mA to test current loop output.

Press UP or DOWN keys to manually toggle the state of the open collector output.

Options Menu for ORP

Display(Factory settings shown) Description

EASY CAL Procedure - ORP• This procedure simplifi es system calibration using standard 4.0 pH and 7.0 pH buffers saturated with Quinhydrone. To calibrate

using any other ORP buffer solutions, use the Standard and Slope functions in the CALIBRATE menu.• Access EASY CAL menu from the view menu.

* For best results, gently stir the submerged electrode for approximately 5 seconds during the stabilization period.

Large temperature differences from pro-cess fl uids to buffers may require longer stabilization time.

Technical notes:The difference between the actual mV and value shown is a good indication of the condition of the electrode. Differences in excess of 50 mV may indi-cate a need to service the electrode.

Press UP, UP, UP, DOWN buttons in sequence to enter menu, will appear during code entry.

To Calibrate: Response: To Accept:

Place electrode tip in first pH buffer;

pH 7.0 87 mVpH 4.0 264 mV

Allow for stabilization to accept

Place electrode tipin second (different) pH buffer.

pH 4.0 264 mVpH 7.0 87 mV

Allow for stabilization Press to acceptsecond buffer calibration.

ENTER

ENTER

* ORP: + 84 mVInput: + 82 mV

1

2

30 seconds*

30 seconds*

EASY CAL: ----Enter Key Code

Place Sensor inORP Buffer #1

Place Sensor inORP Buffer #2

* ORP: + 84 mVInput: + 82 mV

* ORP: + 87 mVInput: + 82 mV

* ORP: +262 mVInput: +260 mV

* ORP: +262 mVInput: +260 mV

* ORP: +264 mVInput: +260 mV

To exit menus and return to VIEW press UP and DOWN button at the same time

Display returns to VIEWMenu in 10 minutes or when ENTER is pressed

Good Easy CalPress <ENTER>

Page 10: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

10 8750-1 pH/ORP Transmitter

The mV value from the sensor when placed in a 7 pH buffer represents the sensor offset. Signet recommends servicing/replacing the sensor when the offset exceeds 50 mV.The 3-2759 pH/ORP system tester allows simple system troubleshooting.

Suggested Solutions

1. Use pH 4, 7, 10 buffers 2. Clean probe and retry EASY CAL. 3. Use Manual calibration for Standard and

Slope if mV offset exceeds 50 mV.

1. Place sensor in correct buffer solution. 2. Use fresh buffer.

1. Use pH values at least 2 pH units apart. 2. Clean pH sensor; replace if necessary 3. Use fresh buffer.

1. Use pH values at least 2 pH units apart. 2. Clean pH sensor; replace if necessary 3. Use fresh buffer.

1. Check all wiring, contacts in preamplifi er. 2. Verify sensor is securely installed. 3. Replace pH sensor.

1. Recalibrate system. 2. Replace pH sensor 3. Replace preamplifi er.

1. Check all wiring, contacts in preamplifi er. 2. Verify sensor is securely installed. 3. Replace pH sensor.

1. Required 4, 7 or 10 buffers not being used.

2. Sensor is depleted too severely to use EasyCal

Sensor was not moved from buffer #1 to buffer #2.

1. pH Standard value within 2 pH units of Slope value

2. pH Sensor effi ciency is inadequate

1. pH Slope value within 2 pH units of Standard value

2. pH Sensor effi ciency is inadequate

1. No temperature or mV signal from sensor detected.

2. No connection between pH sensor and preamplifi er.

mV input is less than 0 pH or greater than 15 pH.

1. No temperature or mV signal from sensor detected.

2. No connection between pH sensor and preamplifi er.

During EasyCal: "Out of Range Use Manual CAL"

During EasyCal: "Same Buffer"

During CALIBRATE Std:"Standard too close to Slope!"

During CALIBRATE Slope:"Slope too close to Standard!"

During CALIBRATE: "Out of Range Check Sensor"

During normal operation: Constant "15.00 pH" or constant "0.00 pH"with good temp value

During normal operation: "Check Sensor?"

Troubleshooting - pH

Display Condition Possible causes

Page 11: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

8750-1 pH/ORP Transmitter 11

1. Use pH 4, 7 buffers saturated with quinhydrone.

2. Clean probe and retry EASY CAL. 3. Use Manual calibration for Standard and

Slope if mV offset exceeds 50 mV. 1. Place sensor in correct buffer solution. 2. Use fresh buffer.

1. Use ORP values at least 120 mV apart. 2. Clean ORP sensor; replace if necessary. 3. Use fresh buffer.

1. Use ORP values at least 120 mV apart. 2. Clean ORP sensor; replace if necessary 3. Use fresh buffer.

1. Check all wiring, contacts in preamplifi er. 2. Verify sensor is securely installed. 3. Replace ORP sensor. 1. Recalibrate system. 2. Replace pH sensor. 3. Replace preamplifi er.

1. Check all wiring, contacts in preamplifi er. 2. Verify sensor is securely installed. 3. Replace ORP sensor.

1. Required 4, 7 buffers with quinhydrone not being used.

2. Sensor is depleted too severely to use EasyCal.

Sensor was not moved from buffer #1 to buffer #2.

1. ORP Standard value within 120 mV of Slope value.

2. ORP Sensor effi ciency is inadequate.

1. ORP Slope value within 120 mV of Standard value.

2. ORP Sensor effi ciency is inadequate.

1. No mV signal or sensor id from

sensor detected. 2. No connection between ORP sensor

and preamplifi er.

mV input is less than -999 or greater than +1999.

1. No temperature or mV signal from sensor detected.

2. No connection between ORP sensor and preamplifi er.

During EasyCal: "Out of Range Use Manual CAL"

During EasyCal: "Same Buffer"

During CALIBRATE Std:"Standard too close to Slope!"

During CALIBRATE Slope:"Slope too close to Standard!"

During CALIBRATE: "Out of Range Check Sensor"

During normal operation: Constant "-1000" or constant ""+2000"

During normal operation: "Check Sensor?"

Troubleshooting - ORP

Display Condition Possible causes Suggested Solutions

Page 12: 8750-1 English manual - Instrumart · 2007-01-22 · Brown (mV input) Silver (Shield) Black (V-) Red (V+) Preamp: Signet 2720 * (connection of terminal 8 to Earth GND may reduce electrical

George Fischer Signet, Inc., 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090

3-8750.090-1 Rev. F 3/06 English © George Fischer Signet, Inc. 2001 Printed in U.S.A. on recycled paper

Ordering InformationMfr. Part No. Code Description3-8750-1 159 000 053 pH/ORP transmitter Field mount3-8750-1P 159 000 054 pH/ORP transmitter Panel mount3-8750-2 159 000 055 pH/ORP transmitter Field mount with relays3-8750-2P 159 000 056 pH/ORP transmitter Panel mount with relays3-8750-3 159 000 057 pH/ORP transmitter Field mount with single input/dual output3-8750-3P 159 000 058 pH/ORP transmitter Panel mount with single input/dual output

AccessoriesMfr. Part No. Code Description3-2714 198 844 300 Twist-Lock Flat pH electrode3-2714-HF 198 844 305 Twist-Lock Flat pH hydrofl uoric acid resistant electrode (<2%)3-2716 198 844 302 Twist-Lock Bulb pH electrode3-2716-DI 198 844 306 Twist-Lock Bulb pH electrode for process liquids (< I00 µs)3-2715 198 844 301 Twist-Lock Flat ORP Electrode3-2717 198 844 303 Twist-Lock Bulb ORP Electrode3-2720 198 864 602 Twist-Lock Preamplifi er, 1/4 in. NPT3-2720-2 198 864 603 Twist-Lock Preamplifi er, ISO 7/1-R 3/4P31542 198 801 630 Red sensor cap for in-line installationsP31542-3 159 000 464 Blue sensor cap for in-line installations3-2759 159 000 762 pH/ORP system tester (includes bypass adapter)3-2759.393 159 000 765 2720 Adapter CableP31515-0P200 159 000 630 PVC Pipe Adapter P31515-0C200 159 000 631 CPVC Pipe Adapter P31515-0V200 159 000 459 PVDF Pipe Adapter 1220-0021 198 801 186 O-ring, FPM (standard)1224-0021 198 820 006 O-ring, EPR1228-0021 198 820 007 O-ring, Kalrez®

3-8050 159 000 184 Universal mounting kit3-8050.395 159 000 186 Splashproof rear cover3-8050.396 159 000 617 RC Filter kit (for relay use)3-8052 159 000 188 3/4” Integral mounting kit3-8052-1 159 000 755 3/4" NPT mount junction box3-0000.596 159 000 641 Heavy duty wall mount bracket3-0700.390 198 864 403 pH buffer kit3-8050.392 159 000 640 Model 200 retrofi t adapter3-5000.598 198 840 225 Surface mount bracket3-5000.399 198 840 224 5 x 5 inch adapter plate for +GF+ Signet retrofi t3-9000.392 159 000 368 Liquid-tight connector kit, 3 sets, 1/2 in. NPT3-9000.392-1 159 000 839 Liquid-tight connector kit, 1 set, 1/2 in. NPT3-9000.392-2 159 000 841 Liquid-tight connector kit, 1 set, PG 13.57300-7524 159 000 687 24 VDC Power Supply 7.5 Ω, 300mA7300-1524 159 000 688 24 VDC Power Supply 15 Ω, 600mA7300-3024 159 000 689 24 VDC Power Supply 30 Ω, 1.3 A 7300-5024 159 000 690 24 VDC Power Supply 50 Ω, 2.1 A 7300-1024 159 000 691 24 VDC Power Supply 100 Ω, 4.2 A3-3719-11 159 000 804 pH/ORP Wet-Tap, 11/2 in. NPT3-3719-21 159 000 805 pH/ORP Wet-Tap, 2 in. NPT3-3719-12 159 000 806 pH/ORP Wet-Tap, ISO 7/1-R 1.53-3719-22 159 000 807 Wet-Tap Assembly, ISO 7/1-R 23-2716-WT 159 000 809 Twist-Lock pH Electrode (use with 3719 Wet-Tap)3-2717-WT 159 000 811 Twist-Lock ORP Electrode (use with 3719 Wet-Tap)2007-0225 159 000 812 PP Clamp-on Saddle, 2.5 in. x 11/2 in. (ASTM, NPT)2007-0230 159 000 813 PP Clamp-on Saddle, 3 in. x 11/2 in. (ASTM, NPT2007-0240 159 000 814 PP Clamp-on Saddle, 4 in. x 11/2 in. (ASTM, NPT)2007-0260 159 000 815 PP Clamp-on Saddle, 6 in. x 2 in. (ASTM, NPT)2007-0280 159 000 816 PP Clamp-on Saddle, 8 in. x 2 in. (ASTM, NPT)2007-0210 159 000 817 PP Clamp-on Saddle, 10 in. x 2 in. (ASTM, NPT)2007-0212 159 000 818 PP Clamp-on Saddle, 12 in. x 2 in. (ASTM, NPT)