893750_634359178012422500
TRANSCRIPT
-
7/29/2019 893750_634359178012422500
1/55
Dr.T.V.Rao MD
Dr.T.V.Rao MD 1
-
7/29/2019 893750_634359178012422500
2/55
Sir George G. Stokes
The phenomenon offluorescence was knownby the middle of the
nineteenth century.British scientist SirGeorge G. Stokes firstmade the observation thatthe mineral fluorsparexhibits fluorescence
when illuminated withultraviolet light, and hecoined the word"fluorescence"
Dr.T.V.Rao
MD 2
-
7/29/2019 893750_634359178012422500
3/55
Discovery of Fluorescence
Microbiology The fluorescence microscope
was devised in the early partof the twentieth century by
August Khler, CarlReichert, and HeinrichLehmann, among others.However, the potential ofthis instrument was notrealized for several decades,and fluorescence
microscopy is now animportant (and perhapsindispensable) tool incellular biology.
Dr.T.V.Rao MD 3
-
7/29/2019 893750_634359178012422500
4/55
Differences between Conventional
and Fluorescent Microscope
The Conventionalmicroscope usesvisible light (400-700nanometers) toilluminate and
produce amagnified image ofa sample.
A fluorescence microscope,uses a much higherintensity light sourcewhich excites afluorescent species in asample of interest. Thisfluorescent species inturn emits a lower
energy light of a longerwavelength thatproduces the magnifiedimage instead of theoriginal light source.
Dr.T.V.Rao MD 4
-
7/29/2019 893750_634359178012422500
5/55
What is Fluorescence?
Fluorescence is lightproduced by a
substance when it isstimulated by anotherlight. Fluorescence iscalled "cold light"because it does not
come from a hot sourcelike an incandescentlight bulb.
Dr.T.V
.Rao MD 5
-
7/29/2019 893750_634359178012422500
6/55
Fluorescence microscopy is a unique way of using a
microscope to discover facts about specimens that often
are not shown by standard bright field microscopy. Inbright field microscopy, specimens are illuminated fromoutside, below or above, and dark objects are seen againsta light background. In fluorescence microscopy,specimens are self-illuminated by internal light, so bright
objects are seen in vivid color against a dark background.Bright objects against dark backgrounds are more easilyseen. This characteristic of fluorescence microscopymakes it very sensitive and specific.
What is FluorescenceMicroscopy?
Dr.T.V.Rao MD 6
-
7/29/2019 893750_634359178012422500
7/55
Most cellular components are colorless and cannot
be clearly distinguished under a microscope. Thebasic premise of fluorescence microscopy is to stain
the components with dyes. Fluorescent dyes, alsoknown as fluorophores of fluorochromes, aremolecules that absorb excitation light at a givenwavelength (generally UV), and after a short delayemit light at a longer wavelength. The delay between
absorption and emission is negligible, generally onthe order of nanoseconds. The emission light canthen be filtered from the excitation light to reveal thelocation of the fluorophores.
Principle of Fluorescent
Microscopy
Dr.T.V.Rao MD 7
-
7/29/2019 893750_634359178012422500
8/55
Principle of Fluorescent
Microscopy Fluorescence microscopy
uses a much higher intensitylight to illuminate the
sample. This light excitesfluorescence species in thesample, which then emitlight of a longer wavelength.The image produced isbased on the second lightsource or the emissionwavelength of thefluorescent species -- ratherthan from the lightoriginally used to
illuminate, and excite, thesam le.
Dr.T.V.Rao MD 8
-
7/29/2019 893750_634359178012422500
9/55
Works on Faster
Transmission of Light Fluorescence, describes
light emission thatcontinues only duringthe absorption of theexcitation light. The timeinterval between absorptionof excitation light and
emission of re-radiated lightin fluorescence is ofextraordinarily shortduration, usually less than amillionth of a second.
Dr.T.V.Rao MD 9
-
7/29/2019 893750_634359178012422500
10/55
Works on Principles of Light
Pathways Specifically, a dichroic
mirror is used toseparate the excitation
and emission lightpaths. Within theobjective, theexcitation/emissionshare the same optics.
In a fluorescencemicroscope, thedichroic mirrorseparates the lightpaths.
Dr.T.V.Rao MD 10
-
7/29/2019 893750_634359178012422500
11/55
Fluorescence microscopy is the most popular method for
studying the dynamic behavior exhibited in live cellimaging. This stems from its ability to isolate individual
proteins with a high degree of specificity amidst non-fluorescing material.
The sensitivity is high enough to detect as few as 50molecules per cubic micrometer.
Different molecules can now be stained with different
colors, allowing multiple types of molecule to be trackedsimultaneously. These factors combine to givefluorescence microscopy a clear advantage over otheroptical imaging techniques, for both in vitro and in vivoimaging.
Advantages of Fluorescent
Microscopy
Dr.T.V.Rao MD 11
-
7/29/2019 893750_634359178012422500
12/55
Fluorescence Microscope
Fluorescencemicroscopy by
epi-illuminationis the mostcommonly usedmethod today
because it is simpleto do, needsrelatively simpleequipment, and isefficient.
Dr.T.V.Rao MD 12
-
7/29/2019 893750_634359178012422500
13/55
Epifluorescence Microscopy
Epifluorescence microscopy is amethod of fluorescencemicroscopy that is widely usedin life sciences The excitatory
light is passed from above (or,for inverted microscopes, frombelow), through the objectivelens and then onto thespecimen instead of passing itfirst through the specimen. The
fluorescence in the specimengives rise to emitted light whichis focused to the detector by thesame objective that is used forthe excitation
Dr.T.V.Rao MD 13
-
7/29/2019 893750_634359178012422500
14/55
The Specimens to be
StainedMost specimens for
fluorescence
microscopy must bestained. Fluorescentstains are called"fluorochromes."
Acridine orange,auramine O, andfluorescent antibody(FA) are thefluorochromes usedmost.
Dr.T.V.Rao MD 14
-
7/29/2019 893750_634359178012422500
15/55
Fluorescence Microscopy applied inMany Braches of Science and Medicine
uses of fluorescencemicroscopy are many
and varied. They are inmedicine, public health,biological research, andenvironment
monitoring.The most common
application is medicallaboratory
diagnosis.Dr.T.V.Rao MD 15
-
7/29/2019 893750_634359178012422500
16/55
How to Use a
FluorescenceMicroscope
Dr.T.V.Rao MD 16
-
7/29/2019 893750_634359178012422500
17/55
How to Use a FluorescenceMicroscope
The object to be studiedis marked with amolecule called a
fluorophore (a dye).When the florescentlight is activated, thelight used forillumination isseparated from theflorescent molecule (thefluorophore), which ismuch weaker. This is done
through an emission filter.Dr.T.V.Rao MD 17
-
7/29/2019 893750_634359178012422500
18/55
Step 1 Locate the light switch
on the side of themicroscope that turns
on the light. Turn themicroscope on.
Write down the exacttime you turn on the
light. The florescentlight is mercury-based,and a time log must bekept for exposure anduse of the light.
Dr.T.V.Rao MD 18
-
7/29/2019 893750_634359178012422500
19/55
Step 2
Locate the toggleswitch on the right
side of themicroscope betweenthe oculars andobjectives. Thisswitch controls the
shutter for themercury light to theobjective lens.
Dr.T.V.Rao MD 19
-
7/29/2019 893750_634359178012422500
20/55
Step 3
Select the appropriatedye for your object (thiswill depend entirely onwhat you are going tobe studying). The mostcommon dyes includeI3 (for use with CTC,
DTAF and fluorescein),A (for use with DAPIand f420), N21 (for usewith Rhoda mine) and
L3 (for use withfluorescein).Dr.T.V.Rao MD 20
-
7/29/2019 893750_634359178012422500
21/55
Step 4
Put the filter
(dye) into the trayoperated by thesilver slidingknob. To remove
the tray, simplypull the silverknob out.
Dr.T.V.Rao MD 21
-
7/29/2019 893750_634359178012422500
22/55
Step 5
Select the lens youwould like to use. The
63x objective lens willhave the highestnumerical aperture.The 100x objective lenswill have the highest
magnitude that can beused with the mercury-based florescent lightsource.
Dr.T.V.Rao MD 22
-
7/29/2019 893750_634359178012422500
23/55
Step 6
Turn the light off whenfinished, and mark thetime. Wait 30 minutesbefore turning the lightback on, or the lampcould explode. It is agood idea to keep track ofhow many hours the
lamp is in use and replaceit according to themanufacture's guidelines.
Dr.T.V.Rao MD 23
-
7/29/2019 893750_634359178012422500
24/55
Step 7
Clean off the
microscope lenswith lenspaper, or if
really dirty, usea cotton swapand glass
cleaner.Dr.T.V.Rao MD 24
-
7/29/2019 893750_634359178012422500
25/55
A specific antibody is labeled by chemically attaching a
fluorophore to form what is known as a conjugate, whichis then spread on a microscope slide containing the
suspected presence of a particular antigen known tostimulate production of the antibody. If the antigen ispresent, the labeled antibody conjugate binds to theantigen and remains bound to the specimen after it iswashed. The presence of the chemically attachedfluorescent conjugate and antigen is demonstrated when
the fluorophore is excited at its excitation peak, and thesubsequent emission intensities at various wavelengthscan then be observed visually or captured by a detectorsystem (digital or traditional camera).
Direct Immunofluorescence
Dr.T.V.Rao MD 25
-
7/29/2019 893750_634359178012422500
26/55
Indirect immunofluorescence here serum possibly containing
unlabeled antibody and its related, but known, antigen areincubated together. A fluorochrome conjugated to an anti-
human antibody (if the specimen being tested is human) is thenplaced on the slide containing the unlabeled antibody-antigen.If indeed, there has been an antigen-antibody reaction, thefluorochrome-labeled anti-human antibody attaches itself to thecomplex formed by the antigen and antibody. Subsequently,the labeled grouping of antigen, antibody, and fluorochromelabeled anti-human antibody is excited at the peak wavelength
intensity for that fluorochrome and any resulting emission isobserved. The indirect immunofluorescence technique reducesthe necessity of keeping in stock large numbers of labeledantibodies, and also usually results in greater fluorescenceintensity. fluorescence,
Indirect Immunofluorescence
Dr.T.V.Rao MD 26
-
7/29/2019 893750_634359178012422500
27/55
Advantages of fluorescence microscopy are
due to its sensitivity, specificity, rapid
testing, and easy use. It is easy to set up anddo, provides rapid diagnostic tests, and canbe very specific. Modern technology allowsconversion of most compound microscopeseasily and economically into effectivefluorescence microscopes.
Fluorescence Microbiology Modernises theDiagnostic Laboratories
Dr.T.V.Rao MD 27
-
7/29/2019 893750_634359178012422500
28/55
Dr.T.V.Rao MD 28
-
7/29/2019 893750_634359178012422500
29/55
Fluorescent Staining in
Tuberculosis The Auramine-rhodamine
process uses a yellowfluorescent dye to visualizeMycobacterium tuberculosisunder a fluorescencemicroscope. Potassiumpermanganate or acridineorange can be used as acounterstain. Under the lens,the bacterial cells will appeargreen.
The Auramine-rhodaminestain is more sensitivethan the Zhiel-Neelsonand more cost effective.
Dr.T.V.Rao MD 29
-
7/29/2019 893750_634359178012422500
30/55
Auramine-phenol solution
Auramine O 25 gEthanol 3000 mlPhenol 250 g
Distilled water 5300 ml
Suspend 25 g auramine-O in 3000 ml of ethanol in a 5-litre conical flask. Add a magnet and place on amagnetic stirrer until solution is complete.
Dissolve 250 g phenol in 5300 ml of distilled water. Mix the phenol solution with alcoholic auraminesolution.
Store in an amber colored bottle at room temp. for upto 3 mths.
Filter before use.
Reagents for staining
Dr.T.V.Rao MD 30
-
7/29/2019 893750_634359178012422500
31/55
Acid-alcohol for decolorization
Sodium chloride 20 gHydrochloric acid, A.R. 20 mlDistilled water 500 mlEthanol 1500 ml
Dissolve sodium chloride in distilled water
Add conc. hydrochloric acid, mix thoroughly.
Add alcohol
Can be stored at room temperature for 3 months.
Reagents for staining
Dr.T.V.Rao MD 31
-
7/29/2019 893750_634359178012422500
32/55
Counter stain
Potassium permanganate 1 gmDistilled water 1000 ml
Dissolve and store in an amber colored bottle
Stays at room temperature for up to 3 months.
Reagents for staining
Dr.T.V.Rao MD 32
-
7/29/2019 893750_634359178012422500
33/55
Use of 1000x magnification
Use of oil immersion
Examination of 300 microscopicfields
About 15 minutes to examine onenegative smear
Examination by an experienced
microscopist
Fuchsin-stained smears require
-
Dr.T.V.Rao MD 33
-
7/29/2019 893750_634359178012422500
34/55
Place the slides on a staining rack, with thesmeared side facing up, the slides not touchingeach other
Flood the slides with freshly filtered auramine-phenol. Let stand for 7-10 min.
Wash well with running water
Decolorize with acid-alcohol for 1-2 min.Wash as before with water and slope the slides
to air dry
Counter stain with 0.1% KMNO4 for 30 seconds
Staining procedure
Dr.T.V.Rao MD 34
Staining procedure
-
7/29/2019 893750_634359178012422500
35/55
Avoid under-decolorization. AFBs are difficult to over-
decolorize since the decolorization procedure with acid-alcohol is relatively milder than the 25% H2SO4 used in Z-N.
Thick smears: Interfere with decolorization, and counterstain. Mask the presence of AFB and tendency to flake,resulting in loss of smear material and possible transfer ofmaterial to other slides
Strong counter stain: May mask the presence of AFB
Re Staining: Smears examined by FM may be restained by Z-N to confirm findings. However, Z-N stained smears cannot
be used for FM
Fading: Stained smears may fade on exposure to light. To bestored wrapped in brown or black paper and kept away fromlight
Staining procedure -Precautions
Dr.T.V.Rao MD 35
-
7/29/2019 893750_634359178012422500
36/55
EXAMINATION PROCEDURE
The mercury vapor lamp : It takes about 10
min. to reach full intensity
Examination: Using the low power objective
(100-150x) first examine a known pos. slide toensure that the microscope is correctly set up
Appearance: Bacilli appear as slender bright
yellow fluorescent rods, standing out clearly
against a dark background
Positive: Minimum of 4 AFB in the entiresmear.
Negative: Less than 4 bacilli ( No.of bacilli to
be recorded)
Dr.T.V.Rao MD 36
-
7/29/2019 893750_634359178012422500
37/55
Screen the Smear in DefinedPattern
Dr.T.V.Rao MD 37
-
7/29/2019 893750_634359178012422500
38/55
Identify the Acid Fast Bacilli
with Caution
Dr.T.V.Rao MD 38
-
7/29/2019 893750_634359178012422500
39/55
EXAMINATION
PROCEDURE The mercury vapor lamp : It takes about
10 min. to reach full intensity
Examination: Using the low power
objective (100-150x) first examine aknown pos. slide to ensure that the
microscope is correctly set up
Appearance: Bacilli appear as slender
bright yellow fluorescent rods, standing
out clearly against a dark background
Positive: Minimum of 4 AFB in theentire smear.
Negative: Less than 4 bacilli ( No.of
bacilli to be recorded)
Dr.T.V.Rao MD 39
-
7/29/2019 893750_634359178012422500
40/55
Grading is Essential to
Determine Prognosis Morphological Confirmation:
With a high power objective (400-
600x) - To be done with all
doubtful smears as well as scantypositives.
Examination: At least three
horizontal sweeps on the entire
smear.
Gradation: Grade positive smearsinto three degrees of positivity
using the high power objective
Dr.T.V.Rao MD 40
-
7/29/2019 893750_634359178012422500
41/55
QUANTIFICATION OF FLUOROCHROMESMEAR RESULTS
ZN
1000x
REPORT
FM Magnification
250x 450x 630x
0NO.AFB 0 0 0
1-9/100FIELD EXACT NO
DIVIDE
by 10 by 4 by 2
10-99/FIELD 1+
1-10/FIELD 2+
>10/FIELD 3+
Dr.T.V.Rao MD 41
-
7/29/2019 893750_634359178012422500
42/55
Examining and Reporting Acid-fast
SmearsNumber of AFB Observed
Report 200x,250x 400x,450x
No AFB seen 0 0
Doubtful: repeat 1-2/30F* 1-2/70F
1+ 1-9/10F 2-18/50F
2+ 1-9/F 4-36/10F
3+ 10-90/F 4-36/F
4+ >90/F >36/F
* number of acid-fast bacilli observed per microscopic field
Dr.T.V.Rao MD 42
-
7/29/2019 893750_634359178012422500
43/55
FM Grading
No. of bacilli per HPF GradeLess than 6 per field
6-100 bacilli per field
More than 100 per field or
large clumps
Dr.T.V.Rao MD 43
-
7/29/2019 893750_634359178012422500
44/55
More than 50 smears examined/day
Continuous availability of power
More sensitive. where small No. of bacilli arepresent
Majority of FM +ve samples are also +ve byculture
Does not yield more false positive than ZN
Doubtful smears to be re examined by ZN
Fluorescence MicroscopyAdvantages in Acid Fast Bacilli
Identification
Dr.T.V.Rao MD 44
-
7/29/2019 893750_634359178012422500
45/55
The quality of a photomicrograph, either digital or
recorded on film, is dependent upon the quality ofthe microscopy. Film is a stern judge of how good
the microscopy has been prior to capturing theimage. It is essential that the microscope beconfigured using Khler illumination, and that thefield and condenser diaphragms are adjustedcorrectly and the condenser height is optimized.
When properly adjusted, the microscope will yieldimages that have even illumination over the entirefield of view and display the best compromise ofcontrast and resolution.
How to get better Photomicrograph forDocumentation
Dr.T.V.Rao MD 45
-
7/29/2019 893750_634359178012422500
46/55
QBC Malaria TestUses the Principles of fluorescence
The QBC MalariaTest is afluorescencemicroscopy-based malaria
diagnostic testthat speeds andsimplifies malaria
detectionDr.T.V.Rao MD 46
-
7/29/2019 893750_634359178012422500
47/55
The method is centrifugation and therebyconcentration of the red blood cells in a predictablearea of the QBC tube, making detection easy and fast.Red cells containing Plasmodia are less dense than
normal ones and concentrate just below theleukocytes, at the top of the erythrocyte column. Thefloat forces all the surrounding red cells into the 40micron space between its outside circumference andthe inside of the tube. Since the parasites contain DNAwhich takes up the acridine orange stain, they appear
as bright specks of light among the non-fluorescingred cells. Virtually all of the parasites found in the 60microliter of blood can be visualized by rotating thetube under the microscope. A negative test can bereported within one minute and positive result withinminutes.
Quantitative Buffy Coat
(QBC) Test
Dr.T.V.Rao MD 47
-
7/29/2019 893750_634359178012422500
48/55
MICROSCOPIC EXAMINATION
0.1% Calcofluor White (W/V) Solution
(a) Use commercially available solution cellufluor17352 (Polysciences, Washington, PA), fluorescentbrightener 28.F6259 (Sigma, St. Louis, MO), 1 gm
(b) distilled water, 100 ml
(c) Gently heat if precipitate develops. Filter ifprecipitate persists. Store at 25 C in the dark.
(3) Commercially prepared kits are
EXAMINATION OF SPECIMENS forFungal Diseases
Dr.T.V.Rao MD 48
-
7/29/2019 893750_634359178012422500
49/55
Need Specific Protocols
Microscope filter system:An epifluorescentmicroscope equipped
with a mercury vaporlamp and either anultraviolet (UV) or blue-violet (BV) excitationfilters to achieve radiation
on the slide below 412 nmshould be used since themaximum absorbance ofCFW is 347 nm.
Dr.T.V.Rao MD 49
-
7/29/2019 893750_634359178012422500
50/55
Fluorescent stains for demonstratingCryptosporidium spp.
oocysts include Auramine- rhodamine, Auramine CarbolFuschin and Acridine orange . Confirmatory staining ofsuspected oocyst by another method may be required
Casemore 1984
Dr.T.V.Rao MD 50
-
7/29/2019 893750_634359178012422500
51/55
Fluorescent stains
Cryptosporidium spp. Fluorescent stains for
demonstratingCryptosporidium spp.
oocysts includeAuramine- rhodamine,Auramine CarbolFuschin and Acridineorange . Confirmatorystaining of suspected
oocyst by anothermethod may be required
Casemore 1984
Dr.T.V.Rao MD 51
-
7/29/2019 893750_634359178012422500
52/55
Immunofluorescent antibody (IFA) procedureemploying cryptosporidium- specific polyclonal or monoclonal antibodies has
been developed
Polyclonal AB, raised against 18 and 20 kDa C.parvum coproantigen, were used to react with C.parvum sporozoites in an immunofluorescence
assay.Monoclonal antibody reagents offer increasedsensitivity and an excellent alternative toconventional staining methods.
These reagents are helpful when screening largenumbers of patients or those with minimalsymptoms.Elimination of the problems of false-positive and
false-negative results with routine staining methods.Dr.T.V.Rao MD 52
-
7/29/2019 893750_634359178012422500
53/55
Giardia and Cryptosporidium
The sensitivity andspecificity of theMerifluor DFA test,have beenreportedto be 96 to 100%and 99.8 to 100%,
respectively, for
both Giardia andCryptosporidium
Dr.T.V.Rao MD 53
-
7/29/2019 893750_634359178012422500
54/55
Several Immuno Diagnostic Methods
are available Using the FluorescentTechniques, which needs specific
testing material and to follow specificprotocols. The Technologists should
be familiar with literature availablewith the Kits
Dr.T.V.Rao MD 54
Enumerable Possibilities withFlorescent Methods
-
7/29/2019 893750_634359178012422500
55/55
Created by Dr.T.V.Rao MD for e
learning resources forMicrobiologists in the Developingworld
Email [email protected]
Dr.T.V.Rao MD 55