    Intro to Ultrasonicspan> Phased Array

    by Tom Nelligan and Dan Kass

    Many people are familiar with the medical applicationsof ultrasonic imaging, in which high frequency sound waves are used to create

    highly detailed cross-sectional pictures of internal organs. Medical sonograms are commonly made with specialized multi-element

    transducers known as phased arrays and their accompanying hardware and software. But the applications of ultrasonic phased array

    technology are not limited to medical diagnosis. In recent years, phased array systems have seen increasing use in industrial settings to

    provide new levels of information and visualization in common ultrasonic tests that include weld inspection, bond testing, thickness

    profiling, and in-service crack detection. This paper provides a brief introduction to how phased array systems work and how they can

    be employed in industrial ultrasonic nondestructive testing.

    What is a phased array system?

    Conventional ultrasonic transducers for NDT commonly consist of either a singleactive element that both generates and receives high

    frequency sound waves, or two paired elements, one for transmitting and one for receiving. Phased array probes, on the other hand,

    typically consist of a transducer assembly with from 16 to as many as 256 small individual elements that can each be pulsed separately.

    These may be arranged in a strip (linear array), a ring (annular array), a circular matrix (circular array), or a more complex shape. As is

    the case with conventional transducers, phased array probes may be designed for direct contact use, as part of an angle beam

    assembly with a wedge, or for immersion use with sound coupling through a water path. Transducer frequencies are most commonly in

    the range from 2 MHz to 10 MHz. A phased array system will also include a sophisticated computer-based instrument that is capable of

    driving the multi-element probe, receiving and digitizing the returning echoes, and plotting that echo information in various standard

    formats. Unlike conventional flaw detectors, phased array systems can sweep a sound beam through a range of refracted angles or

    along a linear path, or dynamically focus at a number of different depths, thus increasing both flexibility and capability in inspection

    setups.

    Typical phased array probe assemblies

    Typical multi-element construction

    How do they work?

    In the most basic sense, a phased array system utilizes the wave physics principle of phasing, varying the time between a series of

    outgoing ultrasonic pulses in such a way that the individual wave fronts generated by each element in the array combinewith each other

    to add or cancel energy in predictable ways that effectively steer and shape the sound beam.

    This is accomplished by pulsing the individual probe elements at slightly different times. Frequently the elements will be pulsed in

    groups of 4 to 32 in order to improve effective sensitivity by increasing aperture, which reduces unwanted beam spreading and enables

    sharper focusing. Software known as a focal law calculator establishes specific delay times for firing each group of elements in order to

    generate the desired beam shape, taking into account probe and wedge characteristics as well as the geometry and acoustical

    properties of the test material. The programmed pulsing sequence selected by the instrument's operating software then launches a

    number of individual wave fronts in the test material. These wave fronts in turn combine constructively and destructively into a single

    primary wave front that travels through the test material and reflects off cracks, discontinuities, back walls, and other material

    boundaries like any conventional ultrasonic wave. The beam can be dynamically steered through various angles, focal distances, and

    focal spot sizes in such a way that a single probe assembly is capable of examining the test material across a range of different

    perspectives. This beam steering happens very quickly, so that a scan from multiple angles or with multiple focal depths can be

    performed in a small fraction of a second.

    The returning echoes are received by the various elements or groups of elements and time-shifted as necessary to compensate for

    varying wedge delays and then summed. Unlike a conventional single element transducer, which will effectively merge the effects of all

    beam components that strike its area, a phased array transducer can spatially sort the returning wavefront according to the arrival time

    and amplitude at each element. When processed by instrument software, each returned focal law represents the reflection from a

    particular angular component of the beam, a particular point along a linear path, and/or a reflection from a particular focal depth. The

    echo information can then be displayed in any of several formats.

    Home About Us Careers Site Map Global Si te

    Home> Knowledge> Theory > Introduction to Olympus Technologies

    ENGLISH

    o to Ultrasonic Phased Array / Olympus http://www.olympus-ims.com/en/ultrasonics/intro-to-pa/

    6 28/2/2012 3:57 PM

  • 8/12/2019 A B C S Scan Olympus

    2/6

    Example of angled beam generated by flat probe by means of variable delay

    Example of focused linear scan beam

    What do the images look like?

    In most typical flaw detection and thickness gaging applications, the ultrasonic test data will be based on time and amplitude information

    derived from processed RF waveforms. These waveforms and the information extracted from them will commonly be presented in one or

    more of four formats: A-scans, B-scans, C-scans, or S-scans. This section shows some examples of image presentations from both

    conventional flaw detectors and phased array systems.

    A-Scan displaysAn A-scan is a simple RF waveform presentation showing the time and amplitude of an ultrasonic signal, as commonly provided by

    conventional ultrasonic flaw detectors and waveform display thickness gages. An A-scan waveform represents the reflections from one

    sound beam position in the test piece. The flaw detector A-scan below shows echoes from two side-drilled holes in a steel reference

    block. The columnar sound beam from a common single-element contact transducer intercepts two out of the three of the holes and

    generates two distinct reflections at different times that are proportional to the depth of the holes.

    Generalized beam profile Straight beam A-scan image

    A single-element angle beam transducer used with a conventional flaw detector will generate a beam along one angular path. While

    beam spreading effects will cause the beam diameter to increase with distance, the area of coverage or field of vision of a conventional

    angle beam will still basically be limited to one angular path. In the example below, a 45 degree wedge at one fixed position is able to

    detect two of the side-drilled holes in the test block because they fall within its beam, but it is not possible to detect the third without

    moving the transducer forward.

    Generalized beam profile Angle beam A-scan image

    A phased array system will display similar A-scan waveforms for reference, however in most cases they will be supplemented by

    B-scans, C-scans, or S-scans as seen below. These standard imaging formats aid the operator in visualizing the type and position of

    flaws in a test piece.

    B-Scan displaysA B-scan is an image showing a cross-sectional profile through one vertical slice of the test piece, showing the depth of reflectors with

    respect to their linear position. B-scan imaging requires that the sound beam be scanned along the selected axis of the test piece, either

    mechanically or electronically, while storing relevant data. In the case below, the B-scan shows two deep reflectors and one shallower

    reflector, corresponding to the positions of the side drilled holes in the test block. With a conventional flaw detector, the transducer must

    be moved laterally across the test piece.

    o to Ultrasonic Phased Array / Olympus http://www.olympus-ims.com/en/ultrasonics/intro-to-pa/

    6 28/2/2012 3:57 PM

  • 8/12/2019 A B C S Scan Olympus

    3/6

    Generalized beam profile Typical b-scan image showing relative hole depth

    A phased array system, on the other hand, can use electronic scanning along the length of a linear array probe to similarly create a

    cross-sectional profile without moving the transducer:

    Electronic linear scan (B-scan) image showing relative hole position and depth across the length of a linear array

    C-Scan displays

    A C-scan is a two dimensional presentation of data displayed as a top or planar view of a test piece, similar in its graphic perspective to

    an x-ray image, where color represents the gated signal amplitude at each point in the test piece mapped to its x-y position. With

    conventional instruments, the single-element transducer must be moved in an x-y raster scan pattern over the test piece. With phased

    array systems, the probe is typically moved physically along one axis while the beam electronically scans along the other. Encoders will

    normally be used whenever precise geometrical correspondence of the scan image to the part must be maintained, although unencoded

    manual scans can also provide useful information in many cases.

    The images that follow show C-scans of a reference block made with a conventional immersion scanning system with a focused

    immersion transducer, and with a portable phased array system using an encoded hand scanner and a linear array. While the graphic

    resolution is not fully equivalent, there are other considerations. The phased array system is field portable, which the conventional

    system is not, and costs about one-third the price. Additionally, the phased array image was made in a few seconds, while the

    conventional immersion scan took several minutes.

    o to Ultrasonic Phased Array / Olympus http://www.olympus-ims.com/en/ultrasonics/intro-to-pa/

    6 28/2/2012 3:57 PM

  • 8/12/2019 A B C S Scan Olympus

    4/6

    General ized beam profi le and direction of motion Conventional C-scan image showing hole position

    General ized beam profi le and direction of motion Phased array C-scan image showing hole position

    S-Scan displays

    An S-scan or sectorial scan image represents a two-dimensional cross-sectional view derived from a series of A-scans that have been

    plotted with respect to time delay and refracted angle. The horizontal axis corresponds to test piece width, and the vertical axis to depth.

    This is the most common format for medical sonograms as well as for industrial phased array images. The sound beam sweeps through

    a series of angles to generate an approximately cone-shaped cross-sectional image. It should be noted that in this example, bysweeping the beam the phased array probe is able to map all three holes from a single transducer position.

    o to Ultrasonic Phased Array / Olympus http://www.olympus-ims.com/en/ultrasonics/intro-to-pa/

    6 28/2/2012 3:57 PM

  • 8/12/2019 A B C S Scan Olympus

    5/6

    Advanced NDT Solutions

    Weld Inspection Solutions

    Corrosion Inspection Solutions

    Composite Inspection Solutions

    Tube Inspection Solutions

    Industrial Scanners

    Aerospace Inspection Solutions

    Flaw Detectors

    Ultrasonic Flaw Detectors

    Phased Array

    Eddy Current Products

    Eddy Current Array Products

    BondMaster

    Probes and Transducers

    Pulser-Receivers

    Integrated InspectionSystems

    Bar Inspection Systems

    Tube Inspection Systems

    NDT Systems Instrumentation

    Thickness Gages

    Precision Thickness Gages

    Corrosion Thickness Gages

    Microscope Solutions

    Laser Confocal Microscopes

    Opto-digital Microscopes

    Semiconductor & Flat Panel DisplayInspection Microscopes

    Upright Metallurgical Microscopes

    Inverted Metallurgical Microscopes

    Modular Microscopes

    Polarizing Microscopes

    Measuring Microscopes

    Stereo Microscopes

    Objective Lenses

    Digital Cameras

    Image Analysis Software

    Optical Metrology

    Laser Confocal Microscopes

    Measuring Microscopes

    Interferometers

    Lens Decenter Measurement Devices

    Micro Spectrophotometer

    Videoscopes, Borescopes

    Industrial Videoscopes

    Industrial Fiberscope

    Industrial Rigid Borescopes

    Light Sources

    Turning tools

    High Speed Video Cameras

    i-SPEED LT

    i-SPEED 2

    i-SPEED TR

    i-SPEED 3

    i-SPEED FS

    i-SPEED PL

    i-SPEED Comparison

    i-SPEED Video Examples

    Innov-X XRF and XRDAnalyzers

    Alloy QA/QC

    Petrochemical & Power

    Precious Metals

    Scrap Recycling, Metal, Glass, &Plastics

    Geochemistry & Mining

    Oil & Concentrate Analyses

    Environmental & IH

    Consumer Products SafetyArchaeometry, Academic & R&D

    Pharmaceutical & Nutraceutical

    Government & Security

    Applications Solutions Key

    Applications

    Application Notes

    Applications Support

    Support

    Service Centers

    PDF Library

    Video Gallery

    Software Downloads

    Training Academy

    Obsolete Products

    ISO Certifications

    MSDS Datasheets

    A-Scan of a single angular component at left, composite sectorial scan at right. The cursor marking 49 degrees identifies the angularlocation of the displayed A-scan.

    Where are phased array systems used?

    Ultrasonic phased array systems can potentially be employed in almost any test where conventional ultrasonic flaw detectors have

    traditionally been used. Weld inspection and crack detection are the most important applications, and these tests are done across a

    wide range of industries including aerospace, power generation, petrochemical, metal billet and tubular goods suppliers, pipeline

    construction and maintenance, structural metals, and general manufacturing. Phased arrays can also be effectively used to profile

    remaining wall thickness in corrosion survey applications.

    The benefits of phased array technology over conventional UT come from its ability to use multiple elements to steer, focus and scan

    beams with a single transducer assembly. Beam steering, commonly referred to sectorial scanning, can be used for mapping

    components at appropriate angles. This can greatly simplify the inspection of components with complex geometries. The small footprint

    of the transducer and the ability to sweep the beam without moving the probe also aids inspection of such components in situations

    where there is limited access for mechanical scanning. Sectorial scanning is also typically used for weld inspection. The ability to test

    welds with multiple angles from a single probe greatly increases the probability of detection of anomalies. Electronic focusing permits

    optimizing the beam shape and size at the expected defect location, thus further optimizing probability of detection. T he ability to focus

    at multiple depths also improves the ability for sizing critical defects for volumetric inspections. Focusing can significantly improve signal-

    to-noise ratio in challenging applications, and electronic scanning across many groups of elements allows for C-Scan images to be

    produced very rapidly.

    Further information on phased array technology and instrumentation is available from Olympus NDT. Contact us for details.