Gulf Coast Conference 2012
Page 1
A Dual Channel Pre-column Backflush System
Roger L. Firor Agilent Technologies 2850 Centerville Rd Wilmington, DE USA
AUX
MMI FID 2
FID 1
Pre-column
Column 2
restrictor
In oven
![Page 2: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/2.jpg)
Outline
• Introduction to backflush • System configuration with CFT (Capillary Flow
Technology) • Software tools • Hardware configurations • Application examples • Advanced setups
Page 2
Gulf Coast Conference 2012
![Page 3: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/3.jpg)
Why use Backflush ?
Those that need: • Retention time precision • Improved spectral quality • Longer lasting calibrations • Improved quantitation • Shorter analysis time • Increased column lifetime • Reduced source maintenance
Page 3
Gulf Coast Conference 2012
![Page 4: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/4.jpg)
Pre-column Backflush
• Eliminate bake out times at end of runs • Dirty samples • Late eluting compounds of no analytical interest • Need to protect column from high boilers • Interest in maximizing lab throughput and minimizing
cost of operations • Compatible with GC and GC/MSD systems • Alternative for established methods
Page 4
Gulf Coast Conference 2012
![Page 5: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/5.jpg)
Advantages of Pre-column Backflush
Backflush is carried out while analysis is occurring on the analytical column
Midpoint pressure control allows analytical column to run at desired flow while pre-column is backflushed
Use of an uncoated pre-column allows transfer of components at a relatively low temperature (i.e. 35 C) and faster backflushing of the heavier components
Sample matrix has little influence
Wide choice of pre-column and analytical column dimensions and phase ratios possible to meet specific application needs
Gulf Coast Conference 2012
Page 5
![Page 6: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/6.jpg)
Backflush with Coated Pre-columns
Pre-columns with thin coating (i.e. 0.10 um to 0.25 um films) can be used
Provides finer control of compounds that are backflushed
Backflush times will be much longer typically
Backflush of heavy compounds remaining on the pre-column will take longer
Better inertness
Gulf Coast Conference 2012
Page 6
![Page 7: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/7.jpg)
Use Flow Start with working method
Extract pertinent method information from method files
Select BF configuration
Set up and verify hardware
Install columns and test configuration
Screen standard on new configuration
Lock with recasted RTL calibration if RTL method
Pick last peak of interest using graphical tool, from which BF timing and conditions are recommended
Verify BF works
Fine tune BF conditions if necessary
Page 7
Gulf Coast Conference 2012
![Page 8: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/8.jpg)
Principle Of Backflushing
S/SL Inlet
Capillary Flow Device
1 psi
75 psi
Detector(s)
Split Vent
During GC Run
After GC Run
S/SL Inlet
Capillary Flow Device
25 psi
3.8 psi
Detector(s)
Aux EPC Split Vent
Aux EPC
Page 8
Gulf Coast Conference 2012
![Page 9: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/9.jpg)
Principle Of Backflushing
Concurrent During GC
Run S/SL Inlet
Backflush Device
1 psi
Appropriate column
Pressure maintained
Detector(s)
Aux EPC Split Vent
Page 9
Gulf Coast Conference 2012
![Page 10: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/10.jpg)
Backflushing After Each Injection
Backflush 20 sec Flow
Backflush 30 sec Flow
Backflush 40 sec Flow
Backflush 60 sec Flow
Column Inlet
End of/or during Run Flow
Backflushing removes heavy materials after each injection.
Page 10
Gulf Coast Conference 2012
![Page 11: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/11.jpg)
AUX EPC
Pre-column
Purged Ultimate Union
Multimode Inlet Temperature programmed split
1.6 mL/min 2.0 mL/min
Analytical Column
140 mL/min
FID
Basic System Overview
![Page 12: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/12.jpg)
Three Configurations and Two Modes of Backflush
Post-run Easiest to set up
Coated Pre-Column (split column) More set-up - Concurrent Faster Turn-Around-Time - Post Run
Uncoated Pre-Column More set up - Concurrent Even Faster Turn-Around-Time
Page 12
Gulf Coast Conference 2012
Pressure zeros should be set on the GC for the inlet and Aux channels before use
Post and Concurrent
![Page 13: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/13.jpg)
BF Wizard Overview
• Software based tools to: – Guide end-user to a proposed BF configuration – Illustrate step-by step procedure to configure BF hardware /
plumbing – Develop a BF enabled method
• Automated BF development for > 8 peaks in a chromatogram
• Comparison mode for 3 ≤ # Peaks ≤ 8 • < 3 peaks in chromatogram is a manual process
Page 13
Gulf Coast Conference 2012
![Page 14: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/14.jpg)
Backflush Method Setup
14 Gulf Coast Conference 2012
![Page 15: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/15.jpg)
Backflush Method Setup
15 Gulf Coast Conference 2012
![Page 16: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/16.jpg)
Backflush Setup
Gulf Coast Conference 2012
Page 16
![Page 17: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/17.jpg)
Backflush Setup
Gulf Coast Conference 2012
Page 17
![Page 18: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/18.jpg)
Backflush Setup
Gulf Coast Conference 2012
Page 18
![Page 19: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/19.jpg)
Backflush Setup
Gulf Coast Conference 2012
Page 19
![Page 20: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/20.jpg)
Backflush Setup
Gulf Coast Conference 2012
Page 20
![Page 21: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/21.jpg)
Backflush Setup
Gulf Coast Conference 2012
Page 21
![Page 22: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/22.jpg)
Backflush Setup
Gulf Coast Conference 2012
Page 22
![Page 23: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/23.jpg)
Backflush Setup
Gulf Coast Conference 2012
Page 23
![Page 24: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/24.jpg)
What Kind of Devices can be used for Backflushing?
Anything with an independent pressure or flow control
Purged Union
Purged Splitter
Deans Switch
24 Gulf Coast Conference 2012
![Page 25: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/25.jpg)
PUU Purged Ultimate Union (Pressure Controlled Tee)
– Primary use = Backflush – Pre-column and post column configurations
• Inlet to one detector – Note: pre-column includes mid-column
– Pre-column configurations can be implemented with diff pump
– Recommended for GC and GC/MS systems
25 Gulf Coast Conference 2012
![Page 26: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/26.jpg)
Effluent Splitters
• 2-way effluent splitter – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre, mid, post-column
• Not best practice: must plug one port – Don't use just for BF, use if > 1 detector
• 3-Way effluent splitter
– Primary use = post-column splitter to three detectors – Backflush – Can be used in other configurations e.g., pre, mid, post-column
• Not best practice: must plug one port – Don't use just for BF, use if > 2 detectors
26 Gulf Coast Conference 2012
![Page 27: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/27.jpg)
Gulf Coast Conference 2012
Page 27
Backflush with Purged Union: Polywax 500 Sample: C5-C18 plus PW500
C26, 10 min
C36, 14 min
C42, 16 min
2 m x 0.53 mm Uncoated retention gap
![Page 28: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/28.jpg)
Pre-column Backflush of C5-C18 Mix After C14
Gulf Coast Conference 2012
Page 28
Retention time STD’s < 0.0019 min
1M x 0.32mm retention gap, 100M x 0.25mm x 0.50um DB-PETRO
C14
C12
![Page 29: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/29.jpg)
Crude Oil Backflush with Coated 2 m x 0.32 mm x 0.10 um DB-1 Pre-column Conditions: Pre-column flow 1.6 ml/min, Analytical column flow 2.0 ml/min MMI: 250 °C(0.5 min) to 425 °C @ 200 °C/min Oven Program: 35 °C (4 min) to 200 °C (1 min) @ 2.0 °C/min Analytical column: 100m x 0.25mm x 0.5um DB-Petro
Backflush time: 4 min
Backflush time: 8 min
C10
C11
Page 29
Gulf Coast Conference 2012
![Page 30: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/30.jpg)
2M x 0.32mm x 0.10um Pre-column: 12 Run Overlay of Crude Oil
Gulf Coast Conference 2012
Page 30
![Page 31: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/31.jpg)
Single Channel Configuration
FID
He or H2
EPC1 P2
IN
Oven
Purged Union
30 m x 0.25 mm x 0.25 um HP-1
30 m x 0.25 mm x 0.50 um HP- INNOWax
Bleed
MMI
P1
restrictor
31 Gulf Coast Conference 2012
Improved stability at low flows
![Page 32: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/32.jpg)
CFT Splitters Configured for Analytical and Pre-column Monitor Channels
Gulf Coast Conference 2012
Page 32
Two un-purged splitters are used
AUX
MMI FID 2
FID 1
Pre-column
Column 2
restrictor
In oven
![Page 33: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/33.jpg)
CFT Splitter Configuration
Gulf Coast Conference 2012
Page 33
Pre-column in
Pressure source in
Pre-column monitor
Column 2
![Page 34: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/34.jpg)
Light Crude: Backflush after C14 Monitor channel in red, 3 m x 0.32 mm x 0.25 um pre-column
34 Gulf Coast Conference 2012
Analytical column: 100m x 0.25mm x 0.5um DB-Petro
![Page 35: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/35.jpg)
Iso-Paraffinic Kerosene
35 Gulf Coast Conference 2012
Analytical column: 100m x 0.25mm x 0.5um DB-Petro
Monitor channel in blue, 3 m x 0.32 mm x 0.10 um pre-column
![Page 36: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/36.jpg)
Synthetic Iso-Paraffinic Kerosene
BF = 6 min BF=8 min
36 Gulf Coast Conference 2012
Analytical column: 100m x 0.25mm x 0.5um DB-Petro Monitor channel in red, 3 m x 0.32 mm x 0.10 um pre-column
![Page 37: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/37.jpg)
SPK Sample Overlay of runs at 8 and 10 min backflush times
37 Gulf Coast Conference 2012
![Page 38: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/38.jpg)
D3606 Parameters with Hydrogen Carrier
Oven: 80 °C isothermal
Column 1: 30 m x 0.25 mm x 0.25 um HP-1
Column 2: 30 m x 0.25 mm x 0.50 um HP-INNOWax
Column 1: 22.9 psig
Column 2: 15.0 psig
Backflush at 4.10 min
38 Gulf Coast Conference 2012
HP-1/INNOWax System
![Page 39: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/39.jpg)
D3606 Parameters with Hydrogen Carrier
Oven: 65 °C isothermal
Column 1: 30 m x 0.25 mm x 0.25 um HP-1
Column 2: 50 m x 0.25 mm x 0.40 um CP-TCEP
Column 1: 26.8 psig
Column 2: 20.5 psig
Backflush at 4.00 min
39 Gulf Coast Conference 2012
HP-1/TCEP System
![Page 40: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/40.jpg)
Column Setup for Backflush
Gulf Coast Conference 2012 40
HP-1/TCEP System for D3606
![Page 41: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/41.jpg)
D3606
Gulf Coast Conference 2012 41
Benzene
Toluene
HP-1/INNOWax System, 80 °C
![Page 42: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/42.jpg)
Overlay of Calibration Mix and Shell 89
Gulf Coast Conference 2012 42
HP-1/TCEP System for D3606 BF at 4.00 min
![Page 43: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/43.jpg)
Overlay Calibration Mix with Shell 89
Gulf Coast Conference 2012 43
Benzene
Toluene
Column 1: 30 m x 0.25 mm x 0.25 um HP-1 Column 2: 50 m x 0.25 mm x 0.4 um TCEP
![Page 44: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/44.jpg)
Calibration: HP-1/INNOWax System
y = 0.404x R² = 1
0
0.5
1
1.5
2
2.5
0 1 2 3 4 5 6
Benzene
y = 0.4214x R² = 1
0 1 2 3 4 5 6 7 8 9
0 5 10 15 20 25
Toluene
%LV %LV
Hydrogen carrier
44 Gulf Coast Conference 2012
![Page 45: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/45.jpg)
Hydrogen Carrier, Isothermal at 80 °C
Run Sample Amt Benzene Vol %
Amt Toluene Vol %
1 Reference 0.82 5.31 2 Reference 0.85 5.31
45 Gulf Coast Conference 2012
Column Set
Column 1: 30 m x 0.25 mm x 0.25 um HP-1 Column 2: 30 m x 0.25 mm x 0.50 um HP-INNOWax
All runs are independent sample preps
![Page 46: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/46.jpg)
Hydrogen Carrier, Isothermal at 65 °C
Run Sample Amt Benzene Vol %
Amt Toluene Vol %
1 Reference 0.86 5.58
2 Reference 0.87 5.40
1 Shell 89 1.03 5.12
2 Shell 89 1.02 5.25
3 Shell 89 1.06 5.27
46 Gulf Coast Conference 2012
Column Set
Column 1: 30 m x 0.25 mm x 0.25 um HP-1 Column 2: 50 m x 0.25 mm x 0.4 um TCEP
All runs are independent sample preps
![Page 47: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/47.jpg)
Advanced Configurations for Backflush
Gulf Coast Conference 2012 47
![Page 48: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/48.jpg)
Variable Split Bleed Saves The Day
Provides necessary bleed to allow pressure control solenoids to operate well (> 1 mL/min) Maintains low flow even during high-pressure backflush Uses normally unused pressure channel (EPC, Backpressure channel on PCM) Allows split flow to be turned OFF for leak checking Provides the opportunity to program bleed flow for custom applications
![Page 49: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/49.jpg)
Using 2nd EPC for Bleed During Run
Set P3 < P2
2 mL/min independent of
P2
Purged Union Column 1 Column 2
Det
SV Trap
He supply
EPC1 P2
IN
Oven
IN
EPC2 P3
atmosphere 1/16” Swagelok tee
Page 49
Gulf Coast Conference 2012
![Page 50: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/50.jpg)
Crude Oil GCxGC: Pre-column Backflush
Gulf Coast Conference 2012
Page 50
BF at 30 min, Oven programmed to 390 C
FID
PCM
FID
PCM
MMI
FID
Column 1
Column 2
PCM Micro Valve
PCM: Pressure control mode,
Modulator
AUX
Restrictor/ Pre-column
G3186B
Aux: Controls Column 1 in constant flow
21-25 ml/min typical
![Page 51: A Dual Channel Pre-column Backflush System · 2016. 9. 2. · – Primary use = post-column splitter to two detectors – Backflush – Can be used in other configurations e.g., pre,](https://reader035.vdocument.in/reader035/viewer/2022071105/5fdf1d7f03a41571fa017ebf/html5/thumbnails/51.jpg)
Summary
• CFT devices with an Aux pressure/flow source can be used for backflush applications • Relies on precise EPC midpoint pressure control
• Backflush can be concurrent while analytical column runs at constant flow
• Excellent thermal control since all parts are in the oven
• Heavy petroleum including crude oils can be routinely analyzed with complete protection of the high resolution capillary column
• The specific carbon number where backflush begins can be easily controlled and fine tuned
• Coated pre-columns offer additional control for precise compound cuts
•First or pre-column can be anything from short (2 m) retention gap to high res cap column
• Dual channel configurations allow direct monitoring of the first column
• Straight forward configurations for ASTM D3606
51 Gulf Coast Conference 2012