a power adaptive, 1.22 pw/hz, 10 mhz read-out front-end for … · 2019-05-18 · a power adaptive,...

10
A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti and Kofi Odame Abstract—This paper presents a read-out front-end ASIC for the measurement of tissue impedances. The 10 mm 2 2-channel front-end ASIC is fabricated in a 0.18 μm CMOS technology. The measurement results show that the proposed ASIC operates over a frequency range of 100 Hz up to 10 MHz. The ASIC has 1.22 pW/Hz power performance, with an SNR over 72 dB for frequencies 1 MHz, and over 65 dB SNR for frequencies 1 MHz. The total measured power consumption of the read- out front-end is shown to range from 2.1 mW to 21.7 mW depending on the input frequency. To achieve a wide dynamic range, the IA has an adaptive gain control feature, and it employs programmable bias currents and reconfigurable structure to save power while processing low frequency signals. A dual- DAC hybrid SAR ADC is presented that reduces the power consumption of the ADC driver by 7 fold for frequencies between 4 kHz and 1 MHz. Index Terms—Power adaptive, Dual-DAC, hybrid, Bio- Impedance, Instrumentation amplifier (IA), SAR ADC, Wear- able, Low noise, Front-end. I. I NTRODUCTION A LOW-POWER analog/mixed-signal read-out front- end is a critical component for a multi-channel bio- impedance system-on-chip (SoC) [1], [2], [3], [4]. The ad- vantage of bio-impedance as a medical imaging technology is that it is radiation free, low cost and miniaturizable [5], [6]. In particular, wide bandwidth bio-impedance is useful in a variety of applications, from body fluid volume mea- surements [7], [8], and characterizing the internal anatomy [9], to distinguishing benign from cancerous tissue using electrical impedance tomography (EIT) [10], [11]. For optimal imaging contrast between benign and cancerous tissue, such as breast cancer detection, broadband EIT operations of up to 10 MHz is necessary [11]. To differentiate between two different impedance distributions, a signal-to-noise ratio (SNR) of 60-90 dB is required [11], [12]; higher values of SNR yield quality image reconstructions and allow for better distinguishability. These applications have high bandwidth, high signal-to- noise ratio (SNR) requirements, and meeting them in a multi- channel SoC means that the analog/mixed-signal read-out front-end must be designed carefully to avoid unnecessarily high power consumption and heat management problems. Figure 1 shows the architecture of a conventional bio- impedance analog read-out front-end (AFE). The instrumen- tation amplifier (IA) and analog-to-digital converter (ADC) Manuscript received XX, 2019. This work was supported by the National Science Foundation (NSF), under grant number IIS-1418497. M. Takhti was with the Thayer School of Engineering, Dartmouth College, Hanover, NH 03755 USA. He is now with Apple, Inc., Cupertino, CA 95014 USA, (e-mail: [email protected]). K. Odame is with the Thayer School of Engineering, Dartmouth College, Hanover, NH 03755 USA. both consume a moderate amount of power, but it is the ADC driver that demands the majority of the power budget in a conventional AFE. This is because of the significant capacitive loading from the flywheel filter, which provides a reservoir of charge to mitigate kickback from the ADC’s sampling circuit [13]. The ADC driver requires a high bias current to charge and discharge the large flywheel capacitor. In this paper, we present a novel bio-impedance analog read-out front-end that achieves a 10 dB improvement in SNR and a 3x improvement in power efficiency over the previous state-of-the-art [14]. To achieve a low-power read-out front- end, an ADC with pipelined sampling and conversion stages is presented, which eliminates the flywheel circuit and the associated high bias current of the ADC driver. Also, the new AFE employs frequency-dependent power adaptation in the IA, ADC driver and ADC. We implemented our AFE design in a 0.18 μm CMOS process, and we report on its impedance measurement performance. II. SYSTEM OVERVIEW A block diagram of the proposed read-out chain is shown in Fig. 1. The proposed chain comprises 1) a power adaptive instrumentation amplifier (IA) to amplify the signals induced on the recording electrodes, while rejecting the large-signal, in-band common-mode interference that is characteristic of bio-impedance measurements, 2) AC coupling capacitors and biasing resistors to properly bias the ADC driver, 3) a power adaptive ADC driver to drive the large capacitive input of the ADC, and 4) a dual-DAC hybrid successive-approximation- register (SAR) ADC that digitizes the induced voltages on the electrodes and passes them to the matched filter for amplitude and phase extractions. A. Instrumentation Amplifier Figure 2 shows the circuit schematic of the power adaptive, variable gain instrumentation amplifier. The IA has an open- loop current conveyor architecture, which is best known for wide bandwidth, low power, high common-mode-rejection- ratio (CMRR) applications [15], [16]. The gain of the IA is defined as R O /R I , which is made variable to increase the dynamic range of the system. Table I shows the resistor values chosen for different gain settings. It is shown in [11] that the excitation frequency of up to 10 MHz is required for some applications such as breast cancer detection. The IA accepts input signals ranging from 100 Hz up to 10 MHz. To avoid unnecessary power consumption while processing low frequency signals, the IA is implemented with programmable bias currents. The passive common-mode

Upload: others

Post on 29-Mar-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-OutFront-End for Bio-Impedance Measurement

Mohammad Takhti and Kofi Odame

Abstract—This paper presents a read-out front-end ASIC forthe measurement of tissue impedances. The 10 mm2 2-channelfront-end ASIC is fabricated in a 0.18 µm CMOS technology.The measurement results show that the proposed ASIC operatesover a frequency range of 100 Hz up to 10 MHz. The ASIC has1.22 pW/Hz power performance, with an SNR over 72 dB forfrequencies ≤ 1 MHz, and over 65 dB SNR for frequencies ≥1 MHz. The total measured power consumption of the read-out front-end is shown to range from 2.1 mW to 21.7 mWdepending on the input frequency. To achieve a wide dynamicrange, the IA has an adaptive gain control feature, and it employsprogrammable bias currents and reconfigurable structure tosave power while processing low frequency signals. A dual-DAC hybrid SAR ADC is presented that reduces the powerconsumption of the ADC driver by 7 fold for frequencies between4 kHz and 1 MHz.

Index Terms—Power adaptive, Dual-DAC, hybrid, Bio-Impedance, Instrumentation amplifier (IA), SAR ADC, Wear-able, Low noise, Front-end.

I. INTRODUCTION

A LOW-POWER analog/mixed-signal read-out front-end is a critical component for a multi-channel bio-

impedance system-on-chip (SoC) [1], [2], [3], [4]. The ad-vantage of bio-impedance as a medical imaging technologyis that it is radiation free, low cost and miniaturizable [5],[6]. In particular, wide bandwidth bio-impedance is usefulin a variety of applications, from body fluid volume mea-surements [7], [8], and characterizing the internal anatomy[9], to distinguishing benign from cancerous tissue usingelectrical impedance tomography (EIT) [10], [11]. For optimalimaging contrast between benign and cancerous tissue, such asbreast cancer detection, broadband EIT operations of up to 10MHz is necessary [11]. To differentiate between two differentimpedance distributions, a signal-to-noise ratio (SNR) of 60-90dB is required [11], [12]; higher values of SNR yield qualityimage reconstructions and allow for better distinguishability.

These applications have high bandwidth, high signal-to-noise ratio (SNR) requirements, and meeting them in a multi-channel SoC means that the analog/mixed-signal read-outfront-end must be designed carefully to avoid unnecessarilyhigh power consumption and heat management problems.

Figure 1 shows the architecture of a conventional bio-impedance analog read-out front-end (AFE). The instrumen-tation amplifier (IA) and analog-to-digital converter (ADC)

Manuscript received XX, 2019. This work was supported by the NationalScience Foundation (NSF), under grant number IIS-1418497.

M. Takhti was with the Thayer School of Engineering, Dartmouth College,Hanover, NH 03755 USA. He is now with Apple, Inc., Cupertino, CA 95014USA, (e-mail: [email protected]).

K. Odame is with the Thayer School of Engineering, Dartmouth College,Hanover, NH 03755 USA.

both consume a moderate amount of power, but it is theADC driver that demands the majority of the power budgetin a conventional AFE. This is because of the significantcapacitive loading from the flywheel filter, which providesa reservoir of charge to mitigate kickback from the ADC’ssampling circuit [13]. The ADC driver requires a high biascurrent to charge and discharge the large flywheel capacitor.

In this paper, we present a novel bio-impedance analogread-out front-end that achieves a 10 dB improvement in SNRand a 3x improvement in power efficiency over the previousstate-of-the-art [14]. To achieve a low-power read-out front-end, an ADC with pipelined sampling and conversion stagesis presented, which eliminates the flywheel circuit and theassociated high bias current of the ADC driver. Also, the newAFE employs frequency-dependent power adaptation in theIA, ADC driver and ADC. We implemented our AFE designin a 0.18 µm CMOS process, and we report on its impedancemeasurement performance.

II. SYSTEM OVERVIEW

A block diagram of the proposed read-out chain is shownin Fig. 1. The proposed chain comprises 1) a power adaptiveinstrumentation amplifier (IA) to amplify the signals inducedon the recording electrodes, while rejecting the large-signal,in-band common-mode interference that is characteristic ofbio-impedance measurements, 2) AC coupling capacitors andbiasing resistors to properly bias the ADC driver, 3) a poweradaptive ADC driver to drive the large capacitive input of theADC, and 4) a dual-DAC hybrid successive-approximation-register (SAR) ADC that digitizes the induced voltages on theelectrodes and passes them to the matched filter for amplitudeand phase extractions.

A. Instrumentation Amplifier

Figure 2 shows the circuit schematic of the power adaptive,variable gain instrumentation amplifier. The IA has an open-loop current conveyor architecture, which is best known forwide bandwidth, low power, high common-mode-rejection-ratio (CMRR) applications [15], [16]. The gain of the IAis defined as RO/RI, which is made variable to increase thedynamic range of the system. Table I shows the resistor valueschosen for different gain settings.

It is shown in [11] that the excitation frequency of up to 10MHz is required for some applications such as breast cancerdetection. The IA accepts input signals ranging from 100 Hzup to 10 MHz. To avoid unnecessary power consumptionwhile processing low frequency signals, the IA is implementedwith programmable bias currents. The passive common-mode

Page 2: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

(a) (b)

Fig. 1. Block diagram of (a) traditional read-out front-end, and (b) the proposed read-out front-end, including: a programmable gain, power adaptive IA, anautomatic gain control unit, an AC coupling and resistor biasing to bias the ADC driver, a power adaptive ADC driver, and a dual-DAC hybrid SAR ADC.The ADC driver accounts for the majority of the power difference. In Fig. 1(b), the dual-DAC ADC permits a much longer sampling period (~12 times longerthan the conventional 12-bit ADC in Fig. 1(a)), which in turn allows us to use an ADC driver with a proportionately lower gain bandwidth product and hencea lower power consumption.

Fig. 2. Circuit schematic of the power adaptive, variable gain IA.

feedback (CMFB) is able to accommodate a large range ofprogrammable bias currents, because the M7L2\7R2 transistorsare switched in or out as needed, to maintain a constant currentdensity in the M7 transistors.

The gain of the IA is set through an automatic gain control(AGC) unit. The decision time of the AGC must be minimizedto reduce the measurement time, which is important forreducing the impact of electrode displacement and motionartifacts [17]. Our comparator-based AGC is similar to thatused in [18]. To limit the AGC decision time to a maximum ofa quarter period, we used a pair of comparators to detect bothpositive and negative voltage crossing thresholds of Vcm± Vth(Vth is the threshold voltage, and Vcm is the common-modevoltage).

The IA has two power settings: 3.3 mW and 0.47 mW. Thelatter is set for input frequencies less than roughly 500 kHz,and the former is to process higher frequency input signals.

TABLE IRESISTORS RI AND RO FOR DIFFERENT GAIN SETTINGS

IA Gain RI Ro

1 10 kΩ 10 kΩ

10 1 kΩ 10 kΩ

100 1 kΩ 100 kΩ

Fig. 3. Measured IA transfer function for its two power gain settings.

The transfer function of the IA for its two power gain settingsis shown in Fig. 3. The IA has a measured bandwidth of 1MHz and 10 MHz for its two power settings, 0.47 mW and 3.3mW, respectively. Measurement results show a gain of 100.6,10.4, and 1 V/V for the 3.3 mW power consumption mode,and 100.3, 10, and 1 V/V for the 0.47 mW mode.

In terms of dynamic range, the IA’s input-referred noise ismeasured as 200, 20 and 10 nV/

√Hz, respectively for gains

of 1, 10.4, 100.6 (see Fig. 4), while the THD varies between-56 dB and -35 dB, depending on the gain setting and inputfrequency (Fig. 5). For the gain of 10 setting, the RI resistoris reduced to 1 kΩ. This increases the current demands ontransistors M1L\1R, leading to a higher THD. (Note: RI is 1kΩ for the gain of 100 setting, but THD is lower in this casebecause of the smaller amplitude input.)

Maintaining a high CMRR throughout the bandwidth ofoperation is specifically important for bio-impedance applica-tions, as there is a large common-mode signal in the frequencyof operation that cannot be filtered out with DC blockingcapacitors [19]. Therefore, a current conveyor-based IA isimplemented in this design, as it is well-known for providingactive isolation of common-mode signals [20]. The measuredresults in Fig. 6 show a maximum CMRR of 66 dB at 10 MHzwith a 1-VPP common-mode input signal.

B. Successive-approximation-register ADC

A large voltage difference between the ADC samplingcapacitor and input signal may result in large spikes at the

Page 3: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

Fig. 4. Measured IA input-referred noise for the gain settings of 1, 10, and100.

Fig. 5. Measured IA THD with a full swing output of 2 − VPP and threedifferent gain settings.

Fig. 6. Measured common-mode-rejection-ratio of the IA for a 1-VPPcommon-mode input signal.

beginning of the sampling phase [13]. If these spikes do not dieout within the sampling phase, some error will be introducedto the sampled signal. To reduce the depth of the spike, alarge reservoir capacitance can be placed at the output of theADC driver [13]. Due to the large input capacitance of theSAR ADC and the flywheel filter, the ADC driver’s powerconsumption can easily be several times that of the ADC itself[21]. The ADC driver’s power consumption could be reducedwith duty cycling, but this would be useful only for low samplerate applications [22]. To reduce the gain bandwidth and powerconsumption requirements of the ADC driver, we developeda hybrid R-C SAR ADC with two pairs of capacitive DACs.Fig. 7 shows the structure of the proposed dual-DAC hybridresistive-capacitive (R-C) SAR ADC. From one sample to thenext, the role of sampling DAC alternates between DAC1± and

DAC2± . Also, while one pair of DACs samples the currentinput, the other pair of DACs is used in the SAR algorithm toconvert the previous sample into a digital word. In this manner,the relaxed sampling time allows the current spikes generatedat the output of the ADC driver to die out well within thesampling period.

In the hybrid R-C SAR ADC, the capacitive DAC is the onlyportion of the DAC that is replicated, while a single resistiveDAC is shared. Although the total area of the capacitive DACdoubles in a dual-DAC hybrid R-C SAR ADC structure, thepower requirement of the ADC driver reduces significantly.

Fig. 8 shows the circuit schematic for the proposed ADCstructure, along with a fully differential difference amplifier(FDDA) as the ADC driver. The 1st DAC (DAC1±) is shownin solid black line capacitors, and the solid gray capacitorsrepresent the 2nd CAP DAC (DAC2±). The switches shownin Fig. 8 that are controlled with DAC Selector (DAC_Sel)and Comparator Connector (Comp_Con) are used to switchin and out DAC1± and DAC2±, during the conversion andsampling phases. The DAC_Sel* signal is the early fall-offversion of the DAC_Sel, which is used to implement bottomplate sampling to prevent input-dependent charge injection[23]. VCM, VRP, and VRN are respectively the common mode,positive- and negative-reference voltages. The switches in Fig.8 that are controlled by DAC_Sel, DAC_Sel* and Comp_Conare positioned to place the 1st and 2nd CAP DACs, respectively,

(a)

(b)

Fig. 7. Block diagram of the proposed dual-DAC hybrid R-C SAR ADCwhile (a) CAP DAC1± & DAC2± are in the sampling and conversion phases,respectively, and (b) CAP DAC1± & DAC2± are respectively in conversionand sampling phases.

Page 4: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

Fig. 8. Circuit schematic of the proposed hybrid R-C SAR ADC following an FDDA configured in unity gain feedback as the ADC driver.

Fig. 9. Timing diagram for an N -bit dual-DAC hybrid SAR ADC.

in the conversion and sampling phases. A simplified timingdiagram of the main clock and control signals for an N -bitdual-DAC hybrid SAR ADC is shown in Fig. 9. The mainclock (Main Clk) frequency is equal to the conversion rate ofthe ADC, and each period of the main clock is divided intoN+1 periods of the internal clock (Int. Clk) for an N -bit ADC.

Referring to the timing diagram in Fig. 9, the Comp_Concontrol line toggles one clock cycle before the DAC_Sel line.This is because the parasitic input capacitor of the comparator,Cp, holds a voltage Vx from the previous conversion cycle (seeFig. 10). When the Comp_Con line switches DAC2+ onto thecomparator input, some time is needed for the DAC to recoverfrom the addition of Cp and its extra charge. This recoveryphase is the extra clock period between Comp_Con togglingand DAC_Sel toggling.

Referring to Fig. 1, there is a matched filter following theADC block. To extract the amplitude and phase of a signal, thematched filter processes 1600 samples of the digitized signal.The matched filter requires an integer number of periods.Therefore, the maximum sampling frequency of the ADC (6.5MHz) would result in an excessive number of data pointsfor frequencies lower than ~4 kHz (~6.5 MHz/1600). As aresult, for input signal frequencies lower than 4 kHz, the ADC

Fig. 10. Simplified circuit operation of the ADC considering parasiticcapacitance at the input node of the comparator in a) DAC1+ performsconversion and DAC2+ is connected to the input signal for sampling, b)conversion on DAC1+ sample is done and while DAC2+ is still connectedto the input signal, the comparator input is being switched from DAC1+

to DAC2+, and c) The comparator input is connected to the DAC2+ whileDAC2+ is still connected to the input signal.

sampling rate (Fs) is reduced to Fs = 1600 × fIn, where fInis the input signal frequency. As shown in Fig. 11, the ADCpower consumption thus ranges from 1.46 mW to 4.41 mW.To process signals with frequencies higher than 3.25 MHz,which fall outside of the first Nyquist zone of the ADC, anunder-sampling technique [24] is adopted. This method allowsany input frequency of fIn that is above the Nyquist rate ofthe ADC to be aliased back into the first Nyquist zone ofthe ADC. The aliased signal is at a much lower frequencyequal to |fIn−nFs|, where n is an integer number, and Fsis the sampling frequency. As a result, power hungry highspeed ADCs can be avoided, which helps to keep the powerconsumption of the read-out front-end low. This comes at theprice of a higher noise level, which is also aliased into the

Page 5: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

Fig. 11. Measured power consumption of the ADC versus input frequency.

first Nyquist zone of the ADC, and can reduce the SNR.The proposed scheme interleaves only the capacitive part

of the entire ADC, sharing the other blocks, which mitigatessome of the complications that come with the fully time-interleaved ADCs but nonetheless inherits other sources oferror. There are three main mismatches that impact the perfor-mance of the time-interleaved ADCs: offset, gain, and timingmismatches.

Employing the same comparator makes it such that the dual-DAC hybrid R-C SAR ADC, in contrast to traditional timeinterleaved ADCs, does not suffer from the offset mismatch.

The gain and timing mismatches in two-channel tradi-tional time-interleaved ADCs cause amplitude modulation andgenerate an image at Fs/2 − fIn [25], [26], where Fs andfIn are the sampling and input frequencies. The capacitiveportion of the DAC in our structure may also suffer from gainmismatch, which can be avoided by proper sizing of the unitcapacitor. Compared to the traditional time-interleaved ADC,our structure is physically less dispersed, since the dual-DAChybrid R-C SAR ADC only interleaves its capacitive part.This reduces clock routing, and hence timing mismatches, inour structure. Furthermore, in the EIT application, there isa narrow filter at the end of the chain (see Fig. 1) called amatched filter that is centered at the input frequency, whichfilters out the distortion. Therefore, as long as the tone atFs/2 − fIn, generated by the gain/timing mismatch, does notoverlap with fIn, the gain/timing mismatch does not affect ourresults in the EIT system. Note that in the EIT application, theinput frequency is selected arbitrarily by the user such that itcovers several points in the desired frequency range which donot overlap with any generated tones.

In the following we investigate how the dual-DAC structureaffects the matching requirement of the capacitive DAC partswith respect to differential nonlinearity (DNL). As the twotraditional and dual-DAC hybrid structures share the same re-sistive DAC, here we consider only the matching requirementfor the capacitive part. We model each unit capacitor as thesum of a nominal capacitance, Co , and some random capaci-tance, δc, which is a normally-distributed random variable witha variance of σ2

c . In a traditional N -bit binary weighted DAC,the worst case DNL occurs at the mid range, transitioningfrom V100...000 to V011...111. At this transition, the DNL erroris due to 2N − 1 capacitors (2N−1 capacitors from the MSB,

and 2N−1 − 1 capacitors from the first N − 1 bits). In thisscenario, it was shown that the standard deviation of the worstcase DNL in a single capacitive DAC structure is defined as[27]:

σS,DNL ≈

√2N

σcCo

LSB (1)

In order to have the worst case DNL be less than half of theLSB, using (1) one can size the minimum unit capacitance.

In our proposed dual-DAC structure, the worst case DNLoccurs at the transition from the last digital word to thesecond to last one, transitioning from VDAC1, 111...111 toVDAC2, 111...110. At this transition, the DNL error is due to2×2N capacitors (2N capacitors from each of the two DACs).In the same way that (1) was derived, one can find the worstcase DNL for the dual-DAC structure:

σD,DNL ≈

√2N+1

σcCo

LSB (2)

By comparing (1) and (2), and only considering the mismatcheffect, one can see that, in order to have the worst case DNL beless than half of the LSB, the unit capacitance in a dual-DACshould be

√2 larger than the one in the single DAC structure

to yield the same result. Fig. 12 shows the comparison ofthe standard deviation of the worst case DNL for both thesingle and dual capacitive DACs, sized to have the same yield(3σDNL < ±0.5LSB) using (1) and (2). It should be notedthat the two equations (1) and (2) only consider the matchingrequirement; however, the smallest unit capacitance can alsobe determined by parasitic capacitance (mostly input of thecomparator), design rules, and noise requirement [27]. In our

(a)

(b)

Fig. 12. Monte Carlo simulation comparison of the worst case DNL in a(a) single DAC with the unit capacitor sized using (1), and (b) dual-DACstructure with the unit capacitor sized using (2), which is

√2 larger than that

of the single DAC.

Page 6: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

(a)

Fig. 13. Measured DNL and INL of the ADC.

(a)

Fig. 14. Circuit schematic of the power adaptive fully differential differenceamplifier used as the core of the ADC driver.

design, it is parasitic capacitance, and not matching concerns,that are a constraint on unit capacitor sizing. So, the unitcapacitor is sized the same in the dual-DAC ADC as it wouldhave been in a conventional, single-DAC ADC. Of course,the total area of the DAC doubles in a dual-DAC SAR ADCstructure – but the benefit is that the power requirement of theADC driver reduces significantly.

In this work, a 12-bit, 6.5 MSPS SAR ADC is implementedwith respectively 7- and 5-bit capacitive and resistive DACs.From the measured DNL and INL shown in Fig. 13, the ADCexhibited a linearity of 10 bits. Trimming would be requiredfor higher resolution applications.

The digital control logic circuitry block used in this designis a modified version of the scheme presented in [28]. Thisblock implements the binary search algorithm, controls theDAC switches, and outputs the ADC’s digital word. A masterhigh frequency clock is generated and transmitted into thefront-end using low-voltage differential signaling (LVDS). Allnecessary clocks inside the chip are generated from the masterclock using clock dividers and combinational logic circuitry.

C. ADC Driver

Due to the limited driving capability of the IA, an ADCdriver is placed between the IA and the ADC. Fig. 14 showsthe circuit schematics of the core of the ADC driver. The coreof the ADC driver is a fully differential difference amplifier(FDDA), configured in a unity gain feedback configuration.The bias currents of the FDDA core and its common-modefeedback (CMFB) are programmable, varied in tandem to

accommodate different ADC sampling rates and slew raterequirements. The biasing current of the ADC driver variesover a wide range. Therefore, the biasing approach presentedin [29] is established that is well suited for all-level currentbiasing.

The measured transfer function of the ADC driver is plottedin Fig. 15 for two representative different bias current levels.

Typically, the ADC sampling time is a fraction of theconversion time. In our new dual-DAC structure, however, theADC sampling time is equal to the entire ADC conversiontime, which significantly relaxes the bandwidth requirementof the ADC driver. The power consumption of the ADC nowis set with the more stringent requirement by the slew rate andbandwidth requirements of the ADC driver.

D. Gain Bandwidth Product of ADC Driver

The requirements for the slew rate and gain bandwidth arerelated to the non-linear settling time segment and the linearsettling time segment, respectively [30]. For a first-order ADCdriver to operate with N-bit resolution, it must have a 3-dBfrequency that satisfies ω > ln(2N+1)/ts, where ts is theADC sampling time. Assuming the driver is implemented asan opamp in unity gain feedback, the gain-bandwidth product(GBW) of this opamp is GBW = ω. Furthermore, to avoidslewing, the opamp needs to provide sufficient current to theload when large input is applied.

For the traditional hybrid structure, the sampling time (ts)is equal to Tc/(N+1), where Tc is the conversion time of theADC. However, the sampling time for the proposed structureis equal to the entire conversion time. Therefore, the gainbandwidth of the ADC driver used for the hybrid structureis relaxed by a factor of (N + 1), compared to that of thetraditional structure. The slew rate requirement is similar inboth the traditional and dual-DAC hybrid structures.

The tail current (It = I1 in Fig. 14) of the ADC driver mustbe selected such that it meets the more stringent requirementfor the current set by the gain bandwidth (It,GBW), and slewrate (It,SR): It = max (It,GBW, It,SR). The requirement onthe current value set by the gain bandwidth and slew rate,respectively, can be calculated as:

It,GBW =VeffCtω

2(3)

Fig. 15. Measured transfer function of the ADC for two representativedifferent biasing currents.

Page 7: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

Fig. 16. Measured power consumption of the ADC driver versus inputfrequency compared between our design and ref. [14].

It,SR =SR

Ct(4)

where Veff is the effective voltage of the input transistors, Ctis the total capacitance driven by the ADC driver, and SR isthe required slew rate. In the traditional hybrid structure in[14], (3) is always the dominant factor that sets the currentrequirement. In the dual-DAC hybrid structure, however, (3)is relaxed by a factor of (N +1). Therefore, in our design, forinput frequencies lower than 1 MHz, (3) sets the requirement,whereas for frequencies above 1 MHz, the current is set basedon the slew-rate requirement, which is determined by (4). AsFig. 16 shows, the ADC driver power consumption is reduced7 fold for frequencies between 4 kHz and 1 MHz when adual-DAC (versus conventional [14]) SAR ADC is used.

III. READ-OUT FRONT-END MEASUREMENT RESULTS

The proposed ASIC read-out front-end including two identi-cal front-end EIT channels is implemented in 0.18 µm CMOStechnology and its micrograph is shown in Fig. 17. Each read-out front-end channel comprises an instrumentation amplifier(IA), an automatic gain control (AGC) unit, an AC couplingwith a biasing resistor, an ADC driver and a SAR ADC.

One of the main performance metrics of EIT imagingsystems is signal-to-noise ratio (SNR). This metric shows howrepeatable the measurement results are. To calculate the SNR,a known signal is applied to the read-out front-end and theextracted amplitude (i.e., the output of the matched filter) atthe end of the chain is recorded [11]. This measurement isrepeated with all measurement conditions kept constant, withthe SNR is defined as the ratio of the mean squared over thevariance of the repeated measurements:

SNR = 10 log10

[ ∑ln=1 (Vn/l)

2∑ln=1

(Vn − V

)2/l

](5)

Here V is the mean of all measurements, l is the numberof repeated measurements, and Vn is the value of n-th mea-surement. The numerator and denominator in (5) representthe mean and variance of all measurements, respectively. Tocalculate the SNR, 23 different frequencies covering a rangeof 100 Hz up to 10 MHz are selected and 80 repeatedmeasurements (l = 80) are captured for each frequency. Toextract the amplitude and phase of a signal, the matched

Fig. 17. Die micrograph of the two-channel EIT read-out front-end. The chiparea including the bonding pads is 10 mm2.

filter processes 1600 samples of the digitized signal. Themeasurements are performed with a 20 mVPP input signal withan IA gain of 100 V/V, with the results plotted in Fig. 18 (a).The read-out front-end chain can handle up to 2 VPP, with anIA gain of 1 V/V.

The measured total power consumption of the read-outfront-end is plotted in Fig. 18 (b). The total power consump-tion of the read-out chain ranges from 2.1 mW to 21.7 mW. Inour previous 10 MHz read-out front-end [14], the total powerconsumption varied between 6.9 mW and 21.8 mW for thesame frequency range, while the power reaches it maximum21.8 mW for any frequency over ~4 kHz.

To quantify the measurement accuracy of the read-out chain,it is required to perform in-vitro phantom measurement withknown loads as suggested in [11], [14], [31], [32]. Fig. 19shows the in-vitro phantom measurement setup, where thetissue is modeled as a parallel R-C circuit (ZT), connectedto two ‘electrodes’ modeled as series resistors (RE). Thefrequency of the voltage source Vin connected to the phantomis swept from 100 Hz to 10 MHz. The voltage that developsacross the tissue model (VT) is fed to the read-out chain andthe output of the ADC is fed to a matched filter block. Theextracted amplitude and phase of the matched filter are usedto generate a Cole-Cole plot. The x-axis and y-axis showthe resistance, RT (ω), and reactance, XT (ω), of the load,respectively, where we have:

ZT (ω) =R

1 + (ωRC)2− j

(ωR2C

)1 + (ωRC)2

= RT (ω)− jXT (ω)

(6)The values of R, C and RE are 1 kΩ, 1 nF and 1 kΩ respec-

tively [11]. Fig. 20 (a) shows the Cole-Cole measurement plot,with the arrow indicating the direction of increasing frequency.The accuracy of the extracted amplitude and that of the phaseused in the Cole-Cole plot are calculated and plotted in Fig. 20(b). The accuracy of the extracted amplitude varies between99.9% and 92.5% and the phase error varies between ∼ 0 and0.15 degrees, depending on the operating frequency.

Page 8: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

IV. DISCUSSION

A comparison of the performance of the proposed read-outfront-end with other ASIC bio-impedance measurement worksis presented in Table II. The figure-of-merit (FOM) used in

(a)

(b)

Fig. 18. (a) Measured SNR of the read-out front-end versus frequency for23 different frequencies, with a 20 mVPP input signal with an IA gain of100 V/V, and (b) measured total power of the chain versus input frequencycompared between our design and ref. [14].

Fig. 19. The phantom measurement setup, with the tissue and electrode beingmodeled as ZT and RE respectively. The induced voltage on the tissue model,VT, is connected to the input of the read-out front-end, and the extractedamplitude and phase are used in the Cole-Cole plot.

Table II is defined as:

FOM =P

BW × 10(SNR/20)

(WHz

)(7)

where P is the power consumption of the AFE, BW representsthe bandwidth of the system, and SNR is the minimum signal-to-noise ratio the system achieves in the BW bandwidth. Allof the systems presented in Table II are single-frequency sys-tems, meaning they can process only one frequency at a time.The ASIC in [33] covers a frequency range of up to 90 kHzfor cardiac monitoring. The analog demodulation approach hasbeen used in this system, where the signal is down convertedusing analog mixers. Employing analog demodulation reducesthe speed requirement of the ADC (since the ADC sees onlyDC or low frequency signal), however this approach is proneto accuracy degradation [35]. Furthermore, it is shown in [36]that, in order to meet a high SNR target, analog systems mustburn more power than their digital counterparts. This ASIChas a decent power efficiency of 4.44 pW/Hz; however, thissystem has a relatively low SNR and low bandwidth. TheEIT system for cardiac monitoring presented in [2] extracts

TABLE IIPERFORMANCE COMPARISON WITH PREVIOUS WORKS

JSSC’11 [33] JSSC’15 [2] ISSCC’17 [4] TBioCAS’18 [34] TBioCAS’18 [14] This Work

Process 180 nm CMOS 180 nm CMOS 65 nm CMOS 180 nm CMOS 180 nm CMOS 180 nm CMOS

Frequency Range Up to 90 kHz 10 - 200 kHz 10 - 300 kHz 500 Hz - 700 kHz 100 Hz - 10 MHz 100 Hz - 10 MHz

Demodulation Type Analog Analog Analog Digital Digital Digital

Front-EndInput Noise 58 nV/

√Hz 42 nV/

√Hz 21 nV/

√Hz 14 nV/

√Hz N/A 10 nV/

√Hz

SNR 40 dB 56.3 dB † N/A 71 dB64 dB @ ≤1 MHz

56 dB @ ≤10 MHz

72 dB @ ≤1 MHz

65 dB @ ≤10 MHz

FOM 4.44 pW/Hz ? 13.97 pW/Hz ‡ N/A 4.09 pW/Hz ?13.7 pW/Hz @ 1 MHz ?

3.45 pW/Hz @ 10 MHz ?

2.26 pW/Hz @ 1 MHz ?

1.22 pW/Hz @ 10 MHz ?

† SNR averaged over measurement index.? Front-end.‡ Includes current injection block.

Page 9: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

(a)

(b)

Fig. 20. (a) Measured Cole-Cole plot over a frequency range of 100 Hz- 10 MHz, and accuracy of the extracted amplitude and phase used in theCole-Cole plot are plotted in (b).

both amplitude and phase, using analog demodulation. Thissystem also has a relatively low bandwidth (200 kHz). TheASIC EIT for lung ventilation monitoring in [4] uses 65 nmtechnology that is fairly new and uses analog demodulationto extract both amplitude and phase. The proposed read-outfront-end compared to our previous work [14] covers thesame frequency range, and employs the digital demodulation.However, the proposed work shows over 8 dB improvementcompared to its predecessor [14]. The FOM of the proposedchain is almost three times better than that observed in [14].Besides, the proposed chain has a power adaptive structurethat significantly reduces its power consumption for lowerfrequencies as compared to the work presented in [14], whichis shown in Fig. 18 (b). The combined area of the ADC andADC driver in the proposed work is ~1.14 mm2, which is 54%larger than its predecessor [14] with an area of ~0.74 mm2.The increased area is mainly due to the dual DAC structureand adaptive biasing. However, despite the increase in area, thedual DAC structure and adaptive biasing significantly improvethe power performance of the proposed work.

V. CONCLUSION

In this work, a power adaptive 2-channel read-out front-end ASIC for bio-impedance measurement is presented. Theread-out front-end covers a wide frequency range of 100 Hzto 10 MHz, which is necessary for distinguishing malignantfrom benign tissues. The IA has a reconfigurable structure witha passive common-mode-feedback that makes it low powerand power adaptive. The IA also employs an automatic gain

control unit to increase the dynamic range of the chain, andreduce the required resolution of the ADC. In order to reducethe total power consumption of the chain, a dual-DAC hybridSAR ADC is implemented that significantly reduces the powerconsumption of the ADC driver, and hence the chain. Thepresented read-out chain has 1.22 pW/Hz power performance,while maintaining an SNR between 81 and 65 dB. As a result,the proposed read-out chain is well suited for use in high-frequency, low power EIT imaging systems.

ACKNOWLEDGMENT

The authors would like to acknowledge the support pro-vided by the National Science Foundation (NSF), under grantnumber IIS-1418497.

REFERENCES

[1] P. Kassanos, L. Constantinou, I. F. Triantis, and A. Demosthenous, “Anintegrated analog readout for multi-frequency bioimpedance measure-ments,” IEEE Sensors Journal, vol. 14, no. 8, pp. 2792–2800, 2014.

[2] S. Hong, J. Lee, J. Bae, and H.-J. Yoo, “A 10.4 mW electrical impedancetomography SoC for portable real-time lung ventilation monitoringsystem,” IEEE Journal of Solid-State Circuits (JSSC), vol. 50, no. 11,pp. 2501–2512, 2015.

[3] S. Hong, K. Lee, U. Ha, H. Kim, Y. Lee, Y. Kim, and H.-J. Yoo, “A 4.9mΩ-sensitivity mobile electrical impedance tomography IC for earlybreast-cancer detection system,” IEEE Journal of Solid-State Circuits(JSSC), vol. 50, no. 1, pp. 245—-257, 2015.

[4] M. Kim, J. Jang, H. Kim, J. Lee, J. Lee, J. Lee, K.-R. Lee, K. Kim,Y. Lee, K. J. Lee et al., “A 1.4-mΩ-sensitivity 94-dB dynamic-rangeelectrical impedance tomography SoC and 48-channel hub-SoC for 3-D lung ventilation monitoring system,” IEEE Journal of Solid-StateCircuits (JSSC), vol. 52, no. 11, pp. 2829–2842, 2017.

[5] D. S. Holder, Electrical impedance tomography: methods, history andapplications. CRC Press, 2004.

[6] R. Bayford, “Bioimpedance tomography (electrical impedance tomog-raphy),” Annual Review of Biomedical Engineering, vol. 8, pp. 63–91,2006.

[7] M. Y. Jaffrin and H. Morel, “Body fluid volumes measurementsby impedance: A review of bioimpedance spectroscopy (BIS) andbioimpedance analysis (BIA) methods,” Medical engineering & physics,vol. 30, no. 10, pp. 1257–1269, 2008.

[8] B. C. Kwan, C.-C. Szeto, K.-M. Chow, M.-C. Law, M.-S. Cheng,C. B. Leung, W.-F. Pang, V. W. Kwong, and P. K. Li, “Bioimpedancespectroscopy for the detection of fluid overload in Chinese peritonealdialysis patients,” Peritoneal Dialysis International, pp. pdi–2013, 2014.

[9] A. Adler and A. Boyle, “Electrical impedance tomography: Tissueproperties to image measures,” Breast cancer, vol. 14, p. 31, 2017.

[10] R. J. Halter, “Development of a multi-high frequency electricalimpedance tomography system for breast imaging,” Ph.D. dissertation,Dartmouth College, 2006.

[11] R. J. Halter, A. Hartov, and K. D. Paulsen, “A broadband high-frequencyelectrical impedance tomography system for breast imaging,” IEEETransactions on Biomedical Engineering (TBME), vol. 55, no. 2, pp.650–659, 2008.

[12] B. H. Brown and A. D. Seagar, “The Sheffield data collection system,”Clinical Physics and Physiological Measurement, vol. 8, no. 4A, p. 91,1987.

[13] R. Downs and M. Oljaca, “Designing SAR ADC drive circuitry, Part II:Input behavior of SAR ADCs,” Analog Zone, 2005.

[14] M. Takhti, Y.-C. Teng, and K. Odame, “A 10 MHz read-out chain forelectrical impedance tomography,” IEEE Transactions on BiomedicalCircuits and Systems (TBioCAS), vol. 12, no. 1, pp. 222–230, 2018.

[15] N. Van Helleputte, S. Kim, H. Kim, J. P. Kim, C. Van Hoof, and R. F.Yazicioglu, “A 160 µA biopotential acquisition IC with fully integratedIA and motion artifact suppression,” IEEE Transactions on BiomedicalCircuits and Systems (TBioCAS), vol. 6, no. 6, pp. 552–561, 2012.

[16] Y.-C. Teng and K. M. Odame, “A 10 MHz 85 dB dynamic rangeinstrumentation amplifier for electrical impedance tomography,” in IEEEBiomedical Circuits and Systems Conference (BioCAS), 2014, pp. 632–635.

Page 10: A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for … · 2019-05-18 · A Power Adaptive, 1.22 pW/Hz, 10 MHz Read-Out Front-End for Bio-Impedance Measurement Mohammad Takhti

[17] S. Khan, P. Manwaring, A. Borsic, and R. Halter, “FPGA-based voltageand current dual drive system for high frame rate electrical impedancetomography,” IEEE Transactions on Medical Imaging, vol. 34, no. 4,pp. 888–901, 2015.

[18] S. Hong, K. Lee, U. Ha, H. Kim, Y. Lee, Y. Kim, and H.-J. Yoo,“A 4.9 mΩ-sensitivity mobile electrical impedance tomography IC forearly breast-cancer detection system,” in IEEE International Solid-StateCircuits Conference (ISSCC), 2014, pp. 316–317.

[19] J. Rosell and P. Riu, “Common-mode feedback in electrical impedancetomography,” Clinical Physics and Physiological Measurement, vol. 13,no. A, p. 11, 1992.

[20] N. Van Helleputte, M. Konijnenburg, J. Pettine, D.-W. Jee, H. Kim,A. Morgado, R. Van Wegberg, T. Torfs, R. Mohan, A. Breeschoten et al.,“A 345 µW multi-sensor biomedical SoC with bio-impedance, 3-channelECG, motion artifact reduction, and integrated DSP,” IEEE Journal ofSolid-State Circuits (JSSC), vol. 50, no. 1, pp. 230–244, 2015.

[21] B. Malki, T. Yamamoto, B. Verbruggen, P. Wambacq, and J. Craninckx,“A 70 dB DR 10 b 0-to-80 MS/s current-integrating SAR ADC withadaptive dynamic range,” IEEE Journal of Solid-State Circuits (JSSC),vol. 49, no. 5, pp. 1173–1183, 2014.

[22] R. F. Yazicioglu, P. Merken, R. Puers, and C. Van Hoof, “A 200µWeight-channel acquisition ASIC for ambulatory EEG systems,” in IEEEInternational Solid-State Circuits Conference (ISSCC). IEEE, 2008,pp. 164–165.

[23] R. J. Baker, CMOS: Circuit design, layout, and simulation. John Wiley& Sons, 2008, vol. 1.

[24] W. Kester et al., Mixed-signal and DSP design techniques. Newnes,2003.

[25] S. M. Jamal, D. Fu, N.-J. Chang, P. J. Hurst, and S. H. Lewis, “A 10-b120-Msample/s time-interleaved analog-to-digital converter with digitalbackground calibration,” IEEE Journal of Solid-State Circuits (JSSC),vol. 37, no. 12, pp. 1618–1627, 2002.

[26] B. Razavi, “Design considerations for interleaved ADCs,” IEEE Journalof Solid-State Circuits (JSSC), vol. 48, no. 8, pp. 1806–1817, 2013.

[27] M. Saberi, R. Lotfi, K. Mafinezhad, and W. A. Serdijn, “Analysis ofpower consumption and linearity in capacitive digital-to-analog convert-ers used in successive approximation ADCs,” IEEE Transactions onCircuits and Systems I: Regular Papers, vol. 58, no. 8, pp. 1736–1748,2011.

[28] M. D. Scott, B. E. Boser, and K. S. Pister, “An ultralow-energy ADCfor smart dust,” IEEE Journal of Solid-State Circuits (JSSC), vol. 38,no. 7, pp. 1123–1129, 2003.

[29] B. A. Minch, “A low-voltage MOS cascode bias circuit for all currentlevels,” in IEEE International Symposium on Circuits and Systems(ISCAS), vol. 3, 2002, pp. III–619–III–622.

[30] D. A. Johns and K. Martin, Analog integrated circuit design. JohnWiley & Sons, 2008.

[31] J. Xu, P. Harpe, J. Pettine, C. Van Hoof, and R. F. Yazicioglu, “Alow power configurable bio-impedance spectroscopy (BIS) ASIC withsimultaneous ECG and respiration recording functionality,” in IEEEEuropean Solid-State Circuits Conference (ESSCIRC), 2015, pp. 396–399.

[32] F. Seoane, J. Ferreira, J. J. Sanchéz, and R. Bragós, “An analogfront-end enables electrical impedance spectroscopy system on-chip forbiomedical applications,” Physiological Measurement, vol. 29, no. 6, p.S267, 2008.

[33] L. Yan, J. Bae, S. Lee, T. Roh, K. Song, and H.-J. Yoo, “A 3.9 mW 25-electrode reconfigured sensor for wearable cardiac monitoring system,”IEEE Journal of Solid-State Circuits (JSSC), vol. 46, no. 1, pp. 353–364,2011.

[34] A. Rao, Y.-C. Teng, C. Schaef, E. K. Murphy, S. H. Arshad, R. J. Halter,and K. M. Odame, “An analog front end ASIC for cardiac electricalimpedance tomography,” IEEE Transactions on Biomedical Circuits andSystems (TBioCAS), vol. 12, pp. 729–738, 2018.

[35] B. Razavi, “Design considerations for direct-conversion receivers,” IEEETransactions on Circuits and Systems II: Analog and Digital SignalProcessing, vol. 44, no. 6, pp. 428–435, 1997.

[36] R. Sarpeshkar, “Analog versus digital: Extrapolating from electronicsto neurobiology,” Neural Computation, vol. 10, no. 7, pp. 1601–1638,1998.

Mohammad Takhti received the B.S. degree inelectrical engineering from the University of Dezfulin Dezful, Iran, the M.S. degree in electrical engi-neering from the K. N. Toosi University of Technol-ogy (KNTU), in Tehran, Iran, and the Ph.D. degreein electrical engineering from the Thayer School ofEngineering at Dartmouth College, Hanover, NH,USA.

He worked in the Research Laboratory for Inte-grated Circuits and Systems (ICAS) at KNTU asan Associate Researcher from 2010 to 2013. He is

currently with Apple Inc., Cupertino, CA, USA.

Kofi Odame (S’06-M’08-SM’15) is an AssociateProfessor of Electrical Engineering at the ThayerSchool of Engineering at Dartmouth College. Kofi’sprimary interest is in analog integrated circuits fornonlinear signal processing. His work has appli-cations in low-power electronics for implantableand wearable biomedical devices, as well as inautonomous sensor systems.