a12report
TRANSCRIPT
-
8/12/2019 A12Report
1/3
Student Number: 25154893 16/03/2014
Measuring Feigenbaums First Constant
using an Electric Circuit
Brittany [email protected]
Personal Tutor: Graham Reed
Abstract: Any system exhibiting period doubling should produce Feigenbaums constant, theratio 1, at 4.23 from 2
to 8 periods. This ratio should be present in electrical systems, but has been found somewhat inaccurate in
previous experiments. Six results were gathered, only one of which found amplitudes relating
Feigenbaums Constant to an appropriate approximation of the ratio, at 4.24. Although this system was
exhibiting chaotic behaviour, through an exponential relationship between the amount of periods present
in the system and the input amplitude, it was not conclusively exhibiting the ratio 1.It was concluded
that an electrical system is markedly difficult to produce consistent behaviour in, due to its variance withinput, and the inaccuracy when making visual measurements of period doubling.
1. Introduction
Feigenbaum described period doubling towards chaos as
a naturally occurring phenomenon, in which one-
dimensional maps for a system that shows a single
quadratic maximum will bifurcate at the same rate:
+1
+2 +1
(1)
The value is not absolute, and approaches the ratio ,
defined as 4.699 quickly as per each period doublingiteration of the system. This value was described as
being present in all period doubling systems[1].
This theory should therefore apply to electrical systems,
and should be observable through the use of electronic
measurement tools. By analysing a series system
comprising an EMF, inductor and diode, the amplitude
of voltage measured across the diode should exhibit
period doubling behaviour, as described in previous
experiments[2].
The experiment undertaken sought to clarify the value
by relating input voltage to observed bifurcation in
output voltage across the diode. Thereby, period
doubling would be inferred. As the amplitude of a
sinusoidal signal is variant, and all elements of the
system tend towards a higher level of uncertainty, the
amount of periods present in a system becomes difficult
to determine, i.e. chaotic. For this reason, 1 is most
accurately gauged electronically, where 1 is given as
approximately 4.23. Therefore, the three results gathered
must be at 2, 4 and 8 periods present in the system, inorder to gain the ratio 1.
2. Background
When the EMF of the circuit is applied within the
forward-bias region of the diode, the diode will act as a
voltage source. The diode will respond in a capacitive
manner when the EMF is applied within the reverse-bias
region of the diode. The alternation between conduction
and capacitance is not an immediate effect when the
current reaches zero, it continues conduction for a shortperiod of time .
is exponentially dependent on the current applied to the
system. As a result, with a certain applied current, the
system no longer has a single period in which it is
conducting, but the previous period will have an effect
on a succinctly occurring current. Therefore, period
doubling occurs, and determines an exponential
relationship between each succinct period doubling. A
more succinct, mathematical analysis of this behaviour is
given in the paper Chaos In A Diode[3].
Previous experiments have well described this electrical
circuit, but have often failed to draw results within an
accuracy range of 5%[4].
Figure 1: Period doubling exponentially tending towards chaos,
applicable to an electronic system.[5]
mailto:[email protected]:[email protected]:[email protected] -
8/12/2019 A12Report
2/3
Student Number: 25154893 16/03/2014
3. Experimental Design
An electric circuit was built comprising an unspecified
diode, 10 20%copper wire coil inductor, wiredusing a breadboard. An EMF was supplied using a
TG550 function generator, with a marked 1% error
margin given on all functional parameters used in thistest, as provided in its data booklet.The function
generator was set to provide a 182 signal, wherevoltage was used as a parameter to investigate period
doubling.Wires used in connecting these components
were either electrically insulated or kept short to avoid
overriding inductance, which could have given
unspecified results.
Figure 2: Schematic of analysed circuit
The measurement equipment comprised of two x10
attenuation oscilloscope probes, and a TDS1000C-EDU
oscilloscope. The probes were calibrated using the
oscilloscopes calibration function to provide a< 1%error in measurement. The oscilloscopes used measured
across the diode for channel 1, and across the functiongenerator for channel 2, with ground as reference, as
given in figure 1.The oscilloscope scale was adjusted as
per each value, to give a precise reading for period
doubling.
The system was analysed by slowly adjusting the
function generators voltage amplitude, and then taking
two measurements, 1 ,2 , per each observed perioddoubling. The first measurement was taken slightly
before period doubling, and the second measurement
slightly after.
2 1
2 (2)
The result using equation (2) would give an average
amplitude, , as to where period doubling had occurred.This average was used due to inconsistency generated by
treating period doubling as a point, as opposed to a
region.
The amplitude of period doubling was measured three
times per system, in order to gain an approximation of
Feigenbaums first constant, using equation 1.The
experiment was repeated multiple times in order to better
approximate .
4. Results
The same system was analysed in six experiments, and
produced a variety of different amplitudes at which
period doubling occurred.
To determine whether or not period doubling hadoccurred, a result would be gathered at a notable change
in the oscilloscopes reading of output amplitude.
Figures 3 and 4 describe the visual changes observed to
gather these amplitudes.
Figure 3: A system exhibiting two periods
Figure 4: A system exhibiting four periods
Experiment 1() 2() 3() 1
First 1.80 2.50 3.22 0.97
Second 1.80 2.70 3.08 2.37
Third 1.80 2.69 2.91 4.24
Fourth 1.96 2.47 2.88 1.24
Fifth 1.66 2.10 2.50 1.10
Sixth 1.76 2.52 2.84 2.375
Table 1: Output amplitude and ratio of period doubling
Due to the inherently chaotic nature of the system, and
the oscillatory function being provided, the results
gathered were found to have been within a 5%
tolerance range. These values varied on the oscilloscope
display, and were estimated using equation (2).
The values gathered attempting to confirmFeigenbaumsConstant varied between 77.10% and +1.03% of
Feigenbaums original value of 4.23.
-
8/12/2019 A12Report
3/3
Student Number: 25154893 16/03/2014
Of the six experiments, one result was found to have
been within a tolerable range given the error. Experiment
three produced appropriate results, however was not
carried out in a different manner to the rest of the
experiments. Due to the range of tolerance of each
amplitude, it was found by minor adjustment within the
5%range of each value that each experiment wouldyield a far closer value to 4.23.
Another measurement observed in this system was an
effect of blocking occurring in regards to whether or not
a system was exhibiting local period doubling. To move
from one local period doubling to the next, a certain
amplitude would have to be applied to the system.
However, to observe this change occurring in reverse,
the system would have to cross the same region, meaning
there was a voltage range for each period doubling in
which a system would be in either the higher period
doubling, or the lower, depending on which it hadpreviously been in. This value changed significantly per
each experiment.
Despite inaccurate results being drawn through most
experiments, for the specified component and
measurement tolerance ranges, experiment three yielded
a highly accurate result. Within 1.03% of the
Feigenbaums constant in a system exhibiting up to 8
periods, experiment three shows a working electrical
mode of Feigenbaums hypothesis.
5. Discussion
The system was determined to be difficult to analyse
using the provided hardware. The gathered results in
Table 1 for 1 were, in all experiments other than the
third, more than 40%out of the expected range, which
greatly exceeds the determined error margins of
components and measurement equipment.
The data drawn could have been misrepresented, as
visual observation of peak changes in the output may not
necessarily give an accurate range of when period
doubling is occurring. The first gathered term 1 is far
more consistent than
2and
3. This could show that thesystem is already becoming heavily affected by noise
due to the tendency towards chaos as the system tends to
more periods present. As such, the gathered data would
have no longer represented period doubling at certain
amplitude, but inconsistencies.
Other component conditions were not accounted for in
this experiment. The diodes forward and reverse regions
were not defined before the experiment begun, and may
contribute to an analytical approach which could better
estimate the expected amplitude at period doubling. This
would have in turn lead to more care being taken in
reading experimental results around the points ofexpected period doubling.
Component conditions such as variance with temperature
and inductance are unaccounted for in the experiment,
yet account for changes in resistance, and therefore
current. As a higher peak voltage is applied to the
system, its expected therefore that the diode will raise in
temperature, changing its forward-bias region. The
inductance of the system is also subject to change basedon temperature. As a result, the system may be changing
over time, meaning that the gathering of results is
inherently gathering from different conditional systems.
All of these changes account for minor variations in
gathering appropriate values for amplitude. The blocking
described in the results section also accounted for
variations in gathering results. Due to the variance
involved, and the visual approach to deducing period
doubling, results are heavily influenced by external
factors.
6. Conclusions
Five of the six experiments failed to conclusively prove
Feigenbaums hypothesis. Although period doubling is
observed, and the system tends towards chaos, with an
exponential rise in periods present in the system per
amplitude, the experiments have failed to draw
conclusive linear results.
The circuit conditions involved in this system are
observably difficult to resolve and quantify. These
unreliable factors lead to difficulty in observing specific
points of bifurcation.
As small values of variability produce large changes in
accuracy to approximation of Feigenbaums constant,
these small inaccuracies are most likely to be responsible
for the discrepancy in each experiment, including what
could be wishful thinking in the case of the experiment
that drew conclusively successful results.
References
[1] M. Feigenbaum, Universal Behaviour in NonlinearSystems, 1980, pp. 1-5.
[2] University of Wisconsin, Demonstration of Chaos, 2003,pp. 4.
[3] A. Missert, P. Thompson, Chaos In A Diode, 2008, pp. 1-
2.
[4] B. Prusha, Measuring Feigenbaumsin a Bifurcating
Electric Circuit, 1997, pp. 3.[5] University of Yale, Deterministic Chaos,
http://classes.yale.edu/fractals/chaos/Feigenbaum/Feigenbaum.html
http://classes.yale.edu/fractals/chaos/Feigenbaum/Feigenbaum.htmlhttp://classes.yale.edu/fractals/chaos/Feigenbaum/Feigenbaum.htmlhttp://classes.yale.edu/fractals/chaos/Feigenbaum/Feigenbaum.htmlhttp://classes.yale.edu/fractals/chaos/Feigenbaum/Feigenbaum.htmlhttp://classes.yale.edu/fractals/chaos/Feigenbaum/Feigenbaum.html