accucheck meter manual

104
BLOOD GLUCOSE METER Reference Manual

Upload: igdirect

Post on 21-Apr-2015

143 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: AccuCheck Meter Manual

BLOOD GLUCOSE METER

Reference Manual

Page 2: AccuCheck Meter Manual
Page 3: AccuCheck Meter Manual

Whether the Accu-Chek Aviva Meter is your first blood glucose meter or you have used a meter for some time; please take the time to read this booklet carefully before you use your new meter. To use it correctly and dependably, you need to understand its operation, screen displays, and all individual features.

Should you have any questions, please call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

1

Page 4: AccuCheck Meter Manual

2 The Accu-Chek® Aviva SystemYour new Accu-Chek Aviva Meter is for quantitative blood glucose testing using Accu-Chek Aviva Test Strips.

The Accu-Chek Aviva Meter may be used by patients who test their own blood glucose as well as by healthcare professionals for monitoring patients’ blood glucose values.

Suitable for self-testing.

The system includes:• Accu-Chek Aviva Meter with battery• Accu-Chek Aviva Test Strips and code chip• Accu-Chek Aviva Control Solution

Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline – Third Edition; CLSI document M29-A3, 2005).

Why Regular Blood Glucose Testing Is ImportantTesting your blood glucose regularly can make a big difference in how you manage your diabetes every day. Therefore, we have made it as simple as possible.

▲!

Page 5: AccuCheck Meter Manual

Need Help?If you have any questions, please call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

Please remember to complete the warranty card and send it by post to Roche Diagnostics, so you receive the best customer service possible and news on product updates.

3

Page 6: AccuCheck Meter Manual

4 Important Information About Your New Meter• Your new meter is designed for testing fresh capillary whole blood samples (for example,

blood from your fingertip or forearm). The meter is for outside the body (in vitro) use. It should not be used to diagnose diabetes.

• Only use Accu-Chek Aviva Test Strips. Other test strips will give inaccurate results.

• The meter comes with a preset time and date. However, you may need to change the time to your time zone.

• If you have followed the steps in the booklet, but still have symptoms that do not seem to match your test results, or if you have questions, talk to your healthcare professional.

The Accu-Chek Multiclix Finger Pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from more than one person as this poses an infection risk.

Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline – Third Edition; CLSI document M29-A3, 2005).

▲!

▲!

Page 7: AccuCheck Meter Manual

ContentsChapter 1: Understanding Your New System .......................................................8

The Accu-Chek Aviva Meter ..............................................................................8 Coding the Meter ............................................................................................10 Adjusting the Time and Date – First Time Use.................................................12 Using the Accu-Chek Aviva System .................................................................14

Chapter 2: Testing Your Blood Glucose ..............................................................15

Performing a Blood Glucose Test ....................................................................15 Flagging Test Results ......................................................................................20 Alternative Site Testing (AST) ..........................................................................21

Chapter 3: Meter Memory, Setup, and Downloading ..........................................29

Memory ..........................................................................................................29Meter Setup ....................................................................................................31Setting the Time and Date...............................................................................32Setting the Beeper On/Off ...............................................................................35Setting the Alarm Clock Function ....................................................................38Setting the Hypoglycemic (Hypo) Alarm Function.............................................42Downloading Your Results to a Computer or PDA ............................................45

5

Page 8: AccuCheck Meter Manual

Chapter 4: Control Testing ..................................................................................47

Why Perform Control Tests ..............................................................................47 About the Control Solutions .............................................................................48 Performing a Control Test ...............................................................................49 Understanding Control Test Results .................................................................53

Chapter 5: Testing Blood Glucose in Multiple Patients ......................................56

Testing Blood Glucose .....................................................................................58Disinfecting the Meter .....................................................................................62What to Disinfect .............................................................................................62How to Disinfect ..............................................................................................62When to Disinfect ............................................................................................63Suitable Disinfecting/Cleaning Solutions ..........................................................63

Chapter 6: Maintenance and Troubleshooting ...................................................64

Changing the Battery.......................................................................................64Cleaning the Meter ..........................................................................................66Maintenance and Troubleshooting ...................................................................67Screen Messages and Troubleshooting ...........................................................68

6

Page 9: AccuCheck Meter Manual

Chapter 7: Technical Information .......................................................................75

Product Limitations .........................................................................................75Specifications ..................................................................................................75Product Safety Information ..............................................................................77Disposing of the Meter ....................................................................................77Warranty and Service Instructions ...................................................................80Additional Supplies ..........................................................................................81Information for Healthcare Professionals – Alternative Site Testing (AST) ........82

Chapter 8: The Accu-Chek Multiclix Finger Pricker ...........................................83

Loading a Lancet Drum ...................................................................................85Taking a Fingertip Blood Sample .....................................................................88Taking an Alternative Site Blood Sample .........................................................89Getting a Good Drop of Blood ..........................................................................92Changing the Lancet Drum ..............................................................................95Caring for Your Finger Pricker .........................................................................96Index ...............................................................................................................97

7

Page 10: AccuCheck Meter Manual

Chapter 1: Understanding Your New SystemThe Accu-Chek Aviva MeterDisplay – Shows results, messages, and results stored in memory.

Right and Left Arrow Buttons –Press to enter memory, adjust settings, and scroll through results.

Test Strip Slot –Insert test strip here.

Infrared (IR) Window –Transfers data from the meter to a computer or PDA.

On/Off/Set Button –Turns the meter on or off and sets options.

Battery Door – Flip open the battery door by pushing the tab in the direction of the arrow.

Code Chip Slot – Insert code chip into this opening.

8

Page 11: AccuCheck Meter Manual

Code Chip(for example)

Battery – Insert with (+) symbol facing up.

Test Strip Container Control Solution Bottle

Golden End – Insert this end of the test strip into the meter.

Yellow Window – Touch blood drop or control solution here.

Test Strip

Top View

On/Off/Set Button

Infrared (IR) Window

9

Page 12: AccuCheck Meter Manual

1. Make sure the meter is off.

2. Turn the meter over. 3. Remove the old code chip (if there is one in the meter) and discard it.

Coding the Meter

▲!Inserting the code chip sets the meter to match the test strips. This “codes” the meter. Change the code chip every time you open a new box of test strips.

10

Page 13: AccuCheck Meter Manual

4. Turn the code chip over so the code number faces away from you. Push it into the code chip slot until it stops.

5. Leave the code chip in the meter until you open a new box of test strips.

Notes:• Do not force the code chip into the meter – it is designed to go into the meter only one way.

• If you see a code “- - -” on the display, insert a code chip into the meter.

11

Page 14: AccuCheck Meter Manual

Adjusting the Time and Date – First Time UseSetting the correct time and date in the meter is important if you use the meter memory or if you want to download your results to a computer.

1. Press and release to turn the meter on. The time and date appear on the display. “Set-up” and the hour flash.

2. Press and release or to decrease or increase the hour. Press and hold or down to scroll faster.

3. Press and release to set the hour. The minutes flash.

decrease increase

| | | | ||

|||||||

|

| | | | ||

||||

12

Page 15: AccuCheck Meter Manual

4. Press and release or to adjust the

minutes. Press to set the minutes.

5. Repeat to set the day, month, and year.

6. Once you set the year, press and hold until the flashing test strip

symbol appears. The time and date are now set.

7. See Chapter 3 “Meter Setup” to set other options.

Note: When you install a new battery, the meter automatically prompts you to check the time and date when you turn it on.

| | | | ||

|||||||

|

| | | | ||

||||

|||||||

||

| | | | ||

||||||||||

||

| | | | ||

||||decrease increase

| |

|||

||

|

| | | |

||

|||

13

Page 16: AccuCheck Meter Manual

Using the Accu-Chek Aviva System• Only use Accu-Chek Aviva Test Strips.

• Change the code chip every time you open a new box of test strips.

• Store the unused test strips in their original container.

• Close the container tightly immediately after you take a test strip out. This helps keep the test strips dry.

• Use the test strip immediately after you take it out of the container.

• Be sure to check the expiry date on the test strip container. Do not use the test strips after that date.

• Store the test strip container and meter in a cool dry place, such as a bedroom.

• Store the test strips at 2°C to 32°C. Do not freeze.

• Do not apply blood or control solution to the test strip before you insert it into the meter.

▲!Do not store test strips in high heat and moisture areas (e.g., bathroom or kitchen)! Heat and moisture can damage the test strips.

14

Page 17: AccuCheck Meter Manual

Chapter 2: Testing Your Blood GlucosePerforming a Blood Glucose TestBefore you perform your first blood test, set up the meter correctly. You will need the meter, a test strip, a finger pricker, and a lancet.

15

The Accu-Chek Multiclix Finger Pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from more than one person as this poses an infection risk.

Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline – Third Edition; CLSI document M29-A3, 2005).

▲!

▲!

Page 18: AccuCheck Meter Manual

Performing a Blood Glucose Test (continued)

1. Wash and dry your hands thoroughly.

2. Prepare the finger pricker according to the manufacturer’s instructions.

3. Insert the test strip into the meter in the direction of the arrows. The meter turns on automatically.

4. Make sure the code number on the display matches the code number on the test strip container. If you miss the code number on the display, remove the test strip and reinsert it into the meter.

16

Page 19: AccuCheck Meter Manual

17

5. A test strip and flashing blood drop symbol appear on the display.

6. Let your arm hang down at your side briefly to allow blood to flow to your fingertips.

7. Obtain a blood sample with the aid of the finger pricker. Blood samples taken from the palm are equivalent to blood taken from the fingertip. For detailed information on how to obtain blood from the palm, refer to the Alternative Site Testing (AST) section, steps 5 and 6.

| | | | ||

|||||||

|

Page 20: AccuCheck Meter Manual

Performing a Blood Glucose Test (continued)

8. With your hand facing downwards, gently squeeze your finger to assist the flow of blood. This helps you get a blood drop.

9. Touch the blood drop to the front edge of the yellow window of the test strip. Do not put blood on top of the test strip. When you see flash, you have enough blood in the test strip. If you applied blood but do not see the flashing , you may reapply more blood within five seconds.

| | | | ||

|||||||

|

18

Page 21: AccuCheck Meter Manual

Note: After a successful test, the meter turns itself off five seconds after the test strip is removed.

Refer to Chapter 8 “Getting a Good Drop of Blood” for helpful tips on getting a drop of blood. To reduce the risk of infection, change the lancet after each test. Blood glucose results are displayed in mmol/L. If your meter displays mg/dL, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 7096000 (Ireland). Using the wrong meter can cause misinterpretation of the test result.

19

10. The result appears on the display. If you want to flag the test result for a special event, leave the test strip in the meter. (See the next section.) Otherwise, remove and discard the used test strip.

Page 22: AccuCheck Meter Manual

Flagging Test Results If you wish, you can “flag” your test result with an asterisk ( ) to mark a special result. You might want to flag an event such as an AST result or after performing exercise. When you review the results in the meter memory, this “flag” can help you remember what was different about the result.

Here is how to flag a result:

1. Perform a test.

2. With the test result on the display, and the test strip STILL IN THE METER, press or once. The appears in the lower right corner.

3. Remove and discard the used test strip.

flag

20

Page 23: AccuCheck Meter Manual

21

Alternative Site Testing (AST)You have the option of obtaining a blood sample from other sites on your body besides the fingertip. Blood obtained from a fingertip or palm, fleshy part of the hand under the thumb (thenar) and the little finger (hypothenar), can be used at any time to measure blood glucose. If blood from an alternative site, forearm, upper arm, thigh, or calf, is used, there are certain times when testing is not appropriate (see next page). This is because your glucose level changes quicker in your fingertip or palm than in the alternative sites. This difference may cause you to make the wrong therapeutic decisions producing adverse health effects.

Please read the following section before you try testing from other sites.

thenarhypothenar

Page 24: AccuCheck Meter Manual

22

IMPORTANT• Talk to your healthcare professional about Alternative Site Testing.

Alternative site testing may be done:• Immediately before a meal• FastingDO NOT test from an alternative site:• Two hours or less after eating• After exercising• If you are sick• If you think your blood glucose is low• If you often don’t notice when your blood glucose is low• During peak action time of short-acting insulin or rapid-acting insulin analogues• Up to two hours after injecting a short-acting insulin or a rapid-acting insulin analogue

Do not change your treatment because of just one result. NEVER ignore symptoms of high or low blood glucose. If your blood glucose result does not match how you feel, perform a fingertip/palm test to confirm your result. If the fingertip/palm result still does not match how you feel, call your healthcare professional.

▲!

Page 25: AccuCheck Meter Manual

23

To perform AST, you will need the meter, a test strip, a finger pricker designed for AST, and a lancet.

1. Prepare the finger pricker by attaching the converter cap and following the standard procedure.

2. Insert the test strip into the meter in the direction of the arrows. The meter turns on automatically.

3. Make sure the code number on the display matches the code number on the test strip container. If you miss the code number on the display, remove the test strip and reinsert it into the meter.

Page 26: AccuCheck Meter Manual

24

4. A test strip and flashing blood drop symbol appear on the display.

5. Press the finger pricker firmly against a fleshy area on the alternative site. Press the finger pricker up and down in a slow pumping motion to assist the flow of blood.

| | | | ||

|||||||

|

Alternative Site Testing (AST) (continued)

Page 27: AccuCheck Meter Manual

6. Trigger the finger pricker while keeping steady pressure on the site. Apply pressure to the site with the finger pricker to assist the flow of blood.

7. Touch the blood drop to the front edge of the yellow window of the test strip. When you see flash, you have enough blood in the test strip. If you applied a blood sample but do not see the flashing ,you may reapply more blood within five seconds. Flag the result or remove and discard the used test strip.

25

| | | | ||

|||||||

|

Note: If the blood drop is too small, reapply pressure to get a sufficient drop.

Page 28: AccuCheck Meter Manual

Troubleshooting Checks1. Check if the test strips were expired.

2. Check if the cap on the test strip container was always closed tightly.

3. Check if the test strip was out of the container for a long time.

4. Check if the test strips were stored in a cool, dry place.

Action Discard the test strips if they are past the expiry date.

Replace the test strips if you think they were uncapped for some time.

Repeat the test with a new test strip.

Repeat the test with a properly stored test strip.

Unusual Test ResultsIf your blood glucose result does not match how you feel, follow these steps:

26

Normal Blood Glucose ValuesThese test strips are calibrated to deliver whole blood results. The normal fasting blood glucose range for an adult without diabetes is 3.5-5.3 mmol/L.1 For people with diabetes: Please consult your diabetes team for the blood glucose range appropriate for you.

You should treat your low or high blood glucose as recommended by your healthcare professional.

Page 29: AccuCheck Meter Manual

Troubleshooting Checks5. Check if you followed the testing

steps.

6. Check if the code number on the meter display matches the code number on the test strip container.

7. If you are still unsure of the problem…

ActionRead Chapter 2 “Testing Your Blood Glucose” and test again. If you still have problems, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

If they do not match, insert the correct code chip into the meter and test again.

Repeat the test with a new test strip and run a control test. If you still have problems, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

27

Page 30: AccuCheck Meter Manual

Symptoms of High or Low Blood GlucoseBeing aware of the symptoms of high or low blood glucose can help you understand your test results and decide what to do if they seem unusual. Here are the most common symptoms:

High blood glucose (hyperglycemia): fatigue, increased appetite or thirst, frequent urination, blurred vision, headache, or general aching.

Low blood glucose (hypoglycemia): sweating, trembling, blurred vision, rapid heartbeat, tingling, or numbness around mouth or fingertips.

▲!If you are experiencing any of these symptoms, test your blood glucose. If your blood glucose result is displayed as LO or HI, call your healthcare professional immediately.

28

Page 31: AccuCheck Meter Manual

Chapter 3: Meter Memory, Setup, and DownloadingMemoryStoring Test ResultsThe meter automatically stores up to 500 blood glucose test results with the time and date of the test. You can review them at any time. Test results are stored from the newest to the oldest. It is very important to set the correct time and date in the meter. Having the correct time and date setting helps ensure appropriate interpretation of stored blood glucose results by you and your healthcare team.

Notes: • Do not change your therapy based on one individual result in memory.• The memory is not lost when you replace the battery. You do need to check that the time and

date are correct after you replace the battery. See Chapter 1, “Adjusting the Time and Date.”• Once 500 results are in memory, adding a new result causes the oldest one to be deleted.• Hold or down to scroll through the results faster.• Up to 20 control results are stored in memory, but cannot be reviewed on the meter. The

stored control results must first be downloaded to a compatible software application. For product availability, please call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

• Control results are not included in the 7, 14, and 30 day averages.

29

Page 32: AccuCheck Meter Manual

Viewing Test ResultsPress either or to enter memory. The most recent result appears. Press to view previous results in order.

OR, Press to look at the 7, 14, or 30 day averages, in that order.

30

Page 33: AccuCheck Meter Manual

Meter SetupUsing the Set-up ModeBy using the set-up mode, you can personalize the meter to suit your lifestyle. Here are the features you can customize –

Time and Date – set the time and date

Beeper – select on or off

Alarm clock – select on or off. If you choose on, select 1 to 4 times a day for a reminder to test.

Hypo alarm – select on or off. If you choose on, choose the blood glucose level for the alarm.

Using the set-up mode is easy. The has three functions for the set-up mode:

• With the meter on, press and hold to enter the set-up mode – about four seconds – until “set-up” flashes on the display.

• Press and release to set the feature you have chosen.

• You can exit the set-up mode at any time by pressing and holding for about four seconds until you see the flashing test strip symbol.

31

Page 34: AccuCheck Meter Manual

TIME / DATE(hrs, min, day, month, year)

BEEPER (on/off)

ALARM CLOCK (A-1, A-2, A-3, A-4)

HYPO ALARM(off, on, level)

You are here

decrease increase

| | | | ||

|||||||

|

| | ||

||||

1. Press to turn the meter on. The flashing test strip symbol appears.

2. Enter the set-up mode (press and hold for about four seconds). “Set-up” and the hour flash on the display.

3. Press and release or to decrease or increase the hour.

32

Setting the Time and Date| |

|||

||

|

| | | |

||

|||

Page 35: AccuCheck Meter Manual

4. Press and release to set the hour. The minutes flash.

5. Press and release or to adjust the minutes. Press to set the minutes.

Note: Press and hold or down to scroll faster.

decrease increase 33

Page 36: AccuCheck Meter Manual

6. Repeat to set the day, month, and year. 7. If you want to set up more options, press and release . If you want to exit, press and hold until you see the flashing test strip symbol.

Setting the Time and Date (continued)

34

| | | | ||

|||||||

|

| | | | ||

||||

|||||||

||

| | | | ||

|||

||||||||

|

| | | | ||

||||

| |

|||

||

|

| | | |

||

|||

Page 37: AccuCheck Meter Manual

TIME / DATE(hrs, min, day, month, year)

BEEPER (on/off)

ALARM CLOCK (A-1, A-2, A-3, A-4)

HYPO ALARM(off, on, level)

You are here

Setting the Beeper On/OffYour new meter has the beeper preset to “On.” You can set the beeper to “OFF,” if you prefer – this will not affect your test results.

The beeper is helpful because it prompts you:

• To apply blood or control solution to the test strip

• When enough blood or control solution is drawn into the test strip

• When the test is complete

• When a button is pressed

• When it is time to test (if you set the alarm clock)

• If an error occurred while testing (even if the beeper is set to off, it still beeps for an error)

35

Page 38: AccuCheck Meter Manual

Setting the Beeper On/Off (continued)

1. Press to turn the meter on. The flashing test strip symbol appears.

2. Enter the set-up mode (press and hold for about four seconds). “Set-up” and the hour flash on the display.

3. Press and release repeatedly until you reach the display with the flashing beeper symbol and “On” or “OFF.”

| | | | ||

|||||||

|

| | | | ||

||||

| | | | ||

|||||||

|

| | | | ||

||||

| | | | ||

|||||||

|

| | | | ||

||||

36

| |

|||

||

|

| | | |

||

|||

Page 39: AccuCheck Meter Manual

4. Press or to switch between “On” or “OFF.”

5. If you want to set up more options, press and release . If you want to exit, press and hold until you see the flashing test strip symbol.

37

| |

|||

||

|

| | | |

||

|||

Page 40: AccuCheck Meter Manual

TIME / DATE(hrs, min, day, month, year)

BEEPER (on/off)

ALARM CLOCK (A-1, A-2, A-3, A-4)

HYPO ALARM(off, on, level)

You are here

Setting the Alarm Clock FunctionThe alarm clock function is a handy way to remind you to test. You can set up to four alarms per day. The meter beeps every two minutes – up to three times. You can turn the alarm off by inserting a test strip or pressing any button. You must have the beeper set to “On” for the alarm to sound.Your new meter is preset with the alarm clock function set to “OFF.” You must turn it “On” to use this feature.If you turn A-1, A-2, A-3, and A-4 on, the meter is preset with the following times for your convenience. You can adjust the times to suit your needs.A-1 8:00 A-2 12:00 A-3 18:00 A-4 22:00

Notes: • If you tested within 30 minutes of an alarm, the alarm does not occur.• If the meter is on at the alarm time, the alarm does not occur.• Exposure to cold conditions may disable alarms until the meter is turned on.

38

Page 41: AccuCheck Meter Manual

When you are setting the time for the alarm clock function, the bell symbol remains on the display and “set-up” flashes continuously.

1. Press to turn the meter on. The flashing test strip symbol appears.

2. Enter the set-up mode (press and hold for about four seconds). “Set-up” and the hour flash on the display.

3. Press and release repeatedly until you reach the display with the bell symbol, “OFF,” and the flashing “set-up” and “A-1.”

| | | | ||

|||||||

|

| | | | ||

||||

| | | | ||

||||||||||||

| ||

| | | | ||

||||||||||||

| || 39

| |

|||

||

|

| | | |

||

|||

Page 42: AccuCheck Meter Manual

Setting the Alarm Clock Function (continued)

4. Press and release or to switch

between “On” or “OFF.” Press and release to set your choice.

5. If you select “On,” the hour flashes. “A-1” and the bell symbol remain on the display.

6. Press and release or to select the

hour. Press and release to set the hour.

| | | | ||

|||||||

|

| | | | ||

||||

| | | | ||

|||||||

|

| | | | ||

||||

40

Page 43: AccuCheck Meter Manual

7. The minutes flash. Press and release or to select 00, 15, 30, or 45. These are the only choices.

8. Press and release to set the minutes.

9. The next alarm “A-2” and “set-up” flash on the display with “OFF” and the bell symbol. You can either set a second alarm or press and hold until you see the flashing test strip symbol to exit the set-up mode.

| | | | ||

|||||||

|

| | | | ||

||||

| | | | ||

||||

||

|||||| |

|

41

Page 44: AccuCheck Meter Manual

TIME / DATE(hrs, min, day, month, year)

BEEPER (on/off)

ALARM CLOCK (A-1, A-2, A-3, A-4)

HYPO ALARM(off, on, level)

You are here

Setting the Hypoglycemic (Hypo) Alarm Function

1. Press to turn the meter on. The flashing test strip symbol appears.

| | | | ||

||||||||||

| || |

You can set the meter to let you know when your blood glucose is possibly too low. You can also select what blood glucose level you want this alarm to have (3.3 to 4.4 mmol/L). Before you set the hypo alarm, talk to your healthcare professional to help you decide what blood glucose level is your hypo level. Your new meter is preset to “OFF” for the hypo alarm. If you want to turn it “On,” follow these steps.

42

Page 45: AccuCheck Meter Manual

2. Enter the set-up mode (press and hold for about four seconds). “Set-up” and the hour flash on the display.

3. Press and release repeatedly until you reach the display with “OFF” and the flashing “set-up” and .

4. Press and release or to switch

between “On” and “OFF.” Press and release to set your choice.

| | | | ||

|||||||

|

| | | | ||

||||

| | | | ||

|||||||

|

| | | | ||

||||

43

Page 46: AccuCheck Meter Manual

Setting the Hypoglycemic (Hypo) Alarm Function (continued)

5. If you select “On,” “set-up” and flash. The display shows 3.9 mmol/L.

6. Press and release or to select the

level you want. Press and release to set it.

7. Press and hold until you see the flashing test strip symbol to exit the set-up mode.

| | | | ||

||||

| | | | ||

|||||||

|

| | | | ||

||||

| | | | ||

|||||||

| | | | | ||

||||||||||

| || |

▲!This function is no substitute for hypoglycemia training by your healthcare professional.

44

Page 47: AccuCheck Meter Manual

Downloading Your Results to a Computer or PDAYou can transfer your stored results to a computer to track, identify patterns, and print.

Transferring Data Directly to a Computer or PDA Using Specialized Software and Infrared Cable1. Install the software according to the instructions. To transfer

the results to a computer, connect the infrared meter cable according to the instructions.

2. Run the software program and follow the instructions about how to download information. Make sure the software is ready to accept data from the meter.

3. With the meter off, press and hold both and until two arrows on the display alternately flash.

4. Locate the infrared (IR) window on the top of the meter.

5. Locate the IR window on either the infrared cable (computer) or PDA.

Infrared (IR) Window

45

Page 48: AccuCheck Meter Manual

6. Put the meter on a flat surface. Point the two IR windows toward each other. They should be 3 to 10 cm apart.

7. Do not move the infrared cable (computer), PDA, or meter during the transfer.

8. Follow the prompts on the software.

9. The software program may shut off the meter automatically when the data transfer is complete.

Notes: • If the data did not transfer successfully, try again. If you still have problems, call the

Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland). • To make the most of the download feature, you must set the time and date in the meter

correctly.• To find out more about downloading your results and what tools are available, call the

Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

Downloading Your Results to a Computer or PDA (continued)

l l l l ||

|||||||

|

l l l l ||

|||||||

|

46

Page 49: AccuCheck Meter Manual

Chapter 4: Control TestingWhy Perform Control TestsPerforming a control test lets you know that the meter and test strips are working correctly to give reliable results. You should perform a control test when:

• You open a new box of test strips

• You left the test strip container open

• You want to check the performance of the meter and test strips

• Your test strips were stored in extreme temperatures and/or humidity

• You dropped the meter

• Your test result does not agree with how you feel

• You want to check if you are testing correctly

47

Page 50: AccuCheck Meter Manual

About the Control Solutions• Only use Accu-Chek Aviva Control Solutions.

• The meter automatically recognizes the Accu-Chek Aviva Control Solution.

• The control solution results are not displayed in memory.

• Write the date you opened the bottle on the bottle label. The solution can be used for three months from that date or until the expiry date on the bottle label, whichever comes first.

• Do not use control solution that is past the expiry date.

• The solution can stain clothing. If you spill it, wash your clothes with soap and water.

• Close the bottle tightly after use.

• Store the bottle at 2°C to 32°C. Do not freeze.

48

Page 51: AccuCheck Meter Manual

Performing a Control TestYou will need the meter, a test strip, and control solution Level 1 and Level 2. The control level is printed on the bottle label.

1. Insert the test strip into the meter in the direction of the arrows. The meter turns on automatically.

2. Make sure the code number on the display matches the code number on the test strip container. If you miss seeing the code number, remove the test strip and reinsert it into the meter.

49

Page 52: AccuCheck Meter Manual

3. Select the control solution you want to test. You will enter the level later in the test.

4. Put the meter on a flat surface, like a table.

5. Remove the control bottle cap. Wipe the tip of the bottle with a tissue.

OR

Performing a Control Test (continued)

50

Page 53: AccuCheck Meter Manual

6. Squeeze the bottle until a tiny drop forms at the tip. Touch the drop to the front edge of the yellow window of the test strip. When you see flash, you have enough control solution in the test strip. Wipe the tip of the bottle with a tissue then cap the bottle tightly.

7. The result appears on the display, along with the control bottle symbol and flashing “L.” Do not remove the test strip yet. Press once to mark the result as a Level 1. If you tested the Level 2 control, press a second time.

| | | | ||

|||||||

|

| | | | ||

|||||||

|

51

Page 54: AccuCheck Meter Manual

8. Press to set the control level in the meter.

9. “OK” and the control result alternate on the display if the result is in range. The range is printed on the test strip container label. “Err” and the control result alternate on the display if the result is not in range. Remove and discard the used test strip.

Performing a Control Test (continued)

52

Page 55: AccuCheck Meter Manual

Range (mmol/L)

Level 1 1.4–3.1

Level 2 14.2–19.1

Understanding Control Test Results

The label on the test strip container shows the acceptable ranges for both the Level 1 and Level 2 control solutions. The result you get should be inside this range. Make sure you compare the result to the correct level of control.

When the control result is inside the range on the test strip container, the test strips and meter are working properly.

Example

Example

53

Page 56: AccuCheck Meter Manual

If the control result is not inside the acceptable range, here are some things you can do to resolve the problem:

Troubleshooting Checks1. Check if the test strip or control

solutions were expired.

2. Check if you wiped the tip of the control solution bottle before and after use.

3. Check if the caps on the test strip container and the control solution bottle were always closed tightly.

4. Check if the test strip was out of the container for a long time.

5. Check if the test strips and control solutions were stored in a cool, dry place.

ActionDiscard the test strips or control solutions if either is past the expiry date. If the control solution was opened more than three months ago, discard it.

Wipe the tip of the bottle with a tissue. Repeat the control test with a new test strip and a fresh drop of control solution.

Replace the test strips or control solutions if you think either was uncapped for some time.

Repeat the control test with a new test strip.

Repeat the control test with a properly stored test strip or control solution.

54

Page 57: AccuCheck Meter Manual

Troubleshooting Checks6. Check if you followed the testing

steps.

7. Check if you chose the correct control solution level, either 1 or 2, when you performed the test.

8. Check if the code number on the meter display matches the code number on the test strip container.

9. If you are still unsure of the problem…

ActionRead Chapter 4 “Control Testing” and test again. If you still have problems, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

If you chose the wrong control solution level, you can still compare the control result to the range printed on the test strip container.

If they do not match, insert the correct code chip into the meter and test again.

Repeat the control test with a new test strip. If you still have problems, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

55

Page 58: AccuCheck Meter Manual

Chapter 5: Testing Blood Glucose in Multiple PatientsOnly healthcare professionals are allowed to perform blood glucose tests in a multiple-patient testing facility.

When using the meter always follow the recognised procedures for handling objects that are potentially contaminated with human material.

Practice the health and safety policy of your laboratory or institution. Also, to prevent contamination of your meter, please follow the directions printed here.

For healthcare professionals we recommend the Accu-Chek Safe-T-Pro Plus or Accu-Chek Safe-T-Pro Uno sterile single-use lancing devices.56

Page 59: AccuCheck Meter Manual

Patients and healthcare workers are potentially at risk of becoming infected when the same blood glucose meter is used to test blood glucose in more than one patient.

Any patient with an infection or suffering from an infectious disease and any patient who is a carrier of a multi-resistant organism must be assigned his/her own meter. This also applies if a patient is suspected of being one of the above and, for as long as the suspicion exists, the meter must not be used to test any other patient.

The Accu-Chek Multiclix Finger Pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from more than one person as this poses an infection risk.

Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline – Third Edition; CLSI document M29-A3, 2005).

Note: • The hypo level may vary from person to person. Therefore we recommend leaving the hypo

alarm function turned off when using the meter in a professional setting.

57

Page 60: AccuCheck Meter Manual

Testing Blood Glucose

1. Wear protective gloves.

2. Wash or disinfect the patient’s finger and allow it to dry thoroughly. The drop of blood does not spread as easily over dry skin.

3. Use only a finger pricker that is approved for use by healthcare professionals.

4. Use a finger pricker with adjustable penetration depth settings. If using the Accu-Chek Safe-T-Pro Plus sterile single-use lancing device, select a penetration depth suitable for the patient’s skin. Start with a shallow depth setting.

5. Select a puncture site along the side of a fingertip.

6. Insert a test strip into the meter.

58

Page 61: AccuCheck Meter Manual

7. Make sure the code number on the meter display matches the code number on the test strip container.

8. Lance the side of the fingertip.

9. Using gentle pressure, massage the finger towards the fingertip to encourage a drop of blood to form. A proper but nevertheless small drop of blood (approx. 0.6 µL) must form.

10. Immediately after the drop of blood has formed, touch the drop to the front edge of the yellow window of the test strip (do not put blood on top of the test strip).

11. The test strip draws up blood.

59

Page 62: AccuCheck Meter Manual

12. When you see flash, you have enough blood in the test strip. If you applied blood but do not see the flashing , you may reapply more blood, if applied IMMEDIATELY.

13. The result appears on the display.

14. Remove and discard the used test strip.

Testing blood glucose (continued)

60

| | | | ||

|||||||

|

Page 63: AccuCheck Meter Manual

Notes: • To test blood glucose approximately 0.6 µL of blood is required. Avoid producing large drops

of blood on the patient’s finger. Where the drops of blood are too large, the meter may come into contact with the blood. Wipe off any large drops of blood. Test using the drop of blood that forms again at the puncture site.

• Residues of water or disinfectant on the skin can dilute the drop of blood producing false results.

• After a successful test the meter turns itself off 5 seconds after the test strip is removed.

• Dispose of used lancets, test strips or single-use lancing devices according to local health and safety regulations.

• The test strip cannot properly draw up blood that has spread over the skin.61

Page 64: AccuCheck Meter Manual

Disinfecting the MeterThe Accu-Chek Aviva meter may be disinfected. Disinfection is performed according to the degree of soiling (see “When to Disinfect”).

What to DisinfectIn the unlikely event that the meter becomes contaminated, the following parts of the meter may be disinfected:

• The area around the test strip slot• The display screen• The outer casing

How to Disinfect• Make sure the meter is off.• Gently wipe the meter’s surface.• Cotton buds, wipes and cloths lightly moistened with disinfectant are all suitable for

disinfecting the meter.• Allow areas you have wiped time to dry thoroughly.• Remove any fluff or lint that may remain.• Refer to the section “Cleaning the Meter” for further information.

62

Page 65: AccuCheck Meter Manual

Notes: • Make sure that no liquid enters the meter. Do not spray anything onto the meter and do

not immerse it in liquid. Doing so may damage the meter’s internal components and stop it from working properly.

When to DisinfectEach time after you have tested a patient’s blood glucose: wipe the area around the test strip slot. If you find blood on the cloth, also wipe the test strip slot itself and the display.If there is blood on the meter: wipe the meter.If blood has entered the meter: wipe the area around the test strip slot and the display screen.Allow areas you have wiped time to dry thoroughly.The meter, including the test strip slot and display screen, must be thoroughly cleaned and disinfected at least once a day after use. Carefully check all recesses, grooves and gaps.

Suitable Disinfecting/Cleaning SolutionsProducts suitable for use:

• Super Sani-Cloth®

• 70% isopropyl alcohol• Mild washing up liquid mixed with water• 10% household bleach solution (1 part bleach plus 9 parts water) made the same day

63

Page 66: AccuCheck Meter Manual

Chapter 6: Maintenance and TroubleshootingChanging the Battery

1. Open the battery door on the back of the meter by pushing the tab in the direction of the arrow and pulling the door up. Remove the old battery.

2. Insert the new battery with the + side up.

3. Put the battery door back in place and snap it closed.

64

Page 67: AccuCheck Meter Manual

Notes:• The meter uses one 3-volt lithium battery, type 2032. This type of battery can be found

in many stores. It is a good idea to have a spare battery available.

• Be sure the battery goes in + side up or facing you.

• After you change the battery, the meter prompts you to confirm the meter’s time and date settings. See Chapter 1 “Adjusting the Time and Date.” All test results are saved in memory.

65

Page 68: AccuCheck Meter Manual

Cleaning the MeterCaring for the Accu-Chek Aviva Meter is easy – just keep it free of dust. If you need to clean it, follow these guidelines carefully to help you get the best performance possible:

Do• Make sure the meter is off

• Gently wipe the meter’s surface with a soft cloth slightly dampened (wring out any excess liquid) with one of these cleaning solutions:

• Super Sani-Cloth®

• 70% isopropyl alcohol

• Mild washing up liquid mixed with water

• 10% household bleach solution (1 part bleach plus 9 parts water) made the same day

Do Not• Get any moisture in the code chip slot or

test strip slot

• Spray any cleaning solution directly onto the meter

• Put the meter under water or liquid

• Pour liquid into the meter

66

Page 69: AccuCheck Meter Manual

Maintenance and TroubleshootingThe meter needs little or no maintenance with normal use. It automatically tests its own systems every time you turn it on and lets you know if something is wrong. (See Chapter 6 on “Screen Messages and Troubleshooting.”) If you drop the meter or think it is not giving accurate results, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland). To make sure the display is working properly, turn the meter off, then press and hold to see the complete display. All the segments should be clear and look exactly like the picture below. If not, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1800 709600 (Ireland).

67

Page 70: AccuCheck Meter Manual

Screen Messages and Troubleshooting

▲!Never make treatment decisions based on an error message. If you have any concerns, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

The meter will not turn on or the display is blank.• Battery is dead –

Insert new battery• Display is damaged –

Call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

• Meter is defective – Call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

• Extreme temperatures – Move the meter to a more temperate area

68

Page 71: AccuCheck Meter Manual

| | | | ||

|||||||

|

| | | | ||

||||

| | | | ||

||||

Battery power is low. Change the battery soon. See Chapter 6 “Changing the Battery.”

The meter is in set-up mode, waiting for you to change or confirm settings. See Chapter 3 “Meter Setup.”

| | | | ||

||||||||||

| || |

| | | | ||

|||||||

|

The meter is ready for you to insert a test strip.

The meter is ready for a drop of blood or control solution.

69

Page 72: AccuCheck Meter Manual

Blood glucose may be higher than the measuring range of the system.

Blood glucose may be lower than the measuring range of the system.

The test result was flagged. See Chapter 2 “Flagging Test Results”.

Blood glucose is below the defined hypo (low blood glucose) level.

70

Page 73: AccuCheck Meter Manual

| | | | ||

||||

| | | | |||||||||||

|| |

| |

The meter is not coded or the code chip is not inserted. Turn off the meter and recode it. See Chapter 1 “Coding the Meter.”

The test strips expire at the end of the current month. Before the end of the month, insert a new code chip from a new box of test strips and ensure the code chip number matches the code number on the test strip container. Make sure the time and date in the meter are correct.

71

Note:• If you see any other error screen, please call the Accu-Chek Customer Careline on

0800 701000 (UK) or 1 800 709600 (Ireland).

Page 74: AccuCheck Meter Manual

Your blood glucose may be extremely low, or the test strip may be damaged or not properly inserted. If you see this error message after you applied blood to the test strip, see Chapter 2 “Unusual Test Results.” If you see this error message before you applied blood to the test strip, remove the test strip and reinsert it, or replace it if damaged. If the message reappears, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

| | | | ||

|||||||

|

The code chip is incorrect. Turn the meter off and insert a new code chip. If this does not fix the problem, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

An error occurred during the test. Discard the test strip and repeat the test. See Chapter 1 “Understanding Your New System.”

| | | | ||

||||

||||||

| | | |

72

Page 75: AccuCheck Meter Manual

Not enough blood or control solution was drawn into the test strip for measurement or was applied after the test had started. Discard the test strip and repeat the test.

The code chip is from an expired lot of test strips. Ensure the code chip number matches the code number on the test strip container. Make sure the time and date in the meter are correct.

| | | | ||

|||||||||

|| | | |

Blood or control solution was applied to the test strip before the flashing blood drop symbol appeared on the display. Discard the test strip and repeat the test.

An electronic error occurred, or, in rare cases, a used test strip was removed and reinserted. Turn the meter off and on, or take the battery out for 10 seconds and reinsert it. Perform a blood glucose or control test. If the problem persists, call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland).

| | | | ||

||||

||||||

| | | |

| | | | ||

|||||||

|

73

Page 76: AccuCheck Meter Manual

| | | | ||

|||||||

|

| | | | ||

|||||||

|

The temperature is above or below the proper range for the meter (6°C to 44°C). Move to an area between 6°C and 44°C, wait five minutes and repeat the test. Do not artificially heat or cool the meter.

The battery is almost out of power. Change the battery now. See Chapter 6 “Changing the Battery.”

The time and date settings may be incorrect. Make sure the time and date are correct and adjust, if necessary. See Chapter 3 “Setting the Time and Date.”

74

Page 77: AccuCheck Meter Manual

Chapter 7: Technical InformationProduct LimitationsPlease read the literature packaged with your test strips to find the latest information on product specifications and limitations.

Specifications Blood volume 0.6 µL

Sample type Fresh capillary whole blood

Measuring time 5 seconds

Measuring range 0.6 to 33.3 mmol/L

Test strip storage conditions 2°C to 32°C

Meter storage conditions Temperature: -25°C to 70ºC

System operating conditions 6°C to 44°C 10% to 90% relative humidity

Memory capacity 500 blood glucose and 20 control results with time and date

Automatic power off 2 minutes

Power supply One 3-volt lithium battery (type 2032)

Display LCD

Dimensions 94 x 53 x 22 mm (LWH)

75

Page 78: AccuCheck Meter Manual

Weight Approx. 60 g (with battery)

Construction Hand-held

Protection class III

Meter type The Accu-Chek Aviva Meter is suitable for continuous operation

Control solution storage conditions 2°C to 32°C

Electromagnetic CompatibilityThis meter meets the electromagnetic immunity requirements as per EN ISO 15197 Annex A. The chosen basis for electrostatic discharge immunity testing was basic standard IEC 61000-4-2. In addition, it meets the electromagnetic emissions requirements as per EN 61326. Its electromagnetic emission is thus low. Interference from other electrically driven equipment is not to be anticipated.

Performance AnalysisThe performance data for the Accu-Chek Aviva System (Accu-Chek Aviva Meter with Accu-Chek Aviva Test Strips) were obtained using capillary blood from diabetic patients (method comparison, accuracy), venous blood (repeatability) and control solution (reproducibility). The system is calibrated with venous blood containing various levels of glucose. The reference values are obtained using the hexokinase method. For method comparison, the results were compared with results obtained using the hexokinase method with deproteinization (automatic analyzer). The hexokinase method is traceable to an NIST standard.The Accu-Chek Aviva System meets the EN ISO 15197 requirements.

Measuring PrincipleBlood glucose concentrations may be measured in whole blood or plasma. Although you always apply whole blood to the test strip, the meter displays blood glucose results that relate either to whole blood or plasma. To see whether your blood glucose meter displays results relating to whole blood or plasma, see the package insert that came with your test strips. You will also find information on how the system works, on the test principle and on reference methods.

76

Page 79: AccuCheck Meter Manual

Product Safety InformationStrong electromagnetic fields may interfere with the proper operation of the meter. Do not use this meter close to sources of strong electromagnetic radiation.

To avoid electrostatic discharge, do not use the meter in a very dry environment, especially one in which synthetic materials are present.

Disposing of the MeterDuring blood glucose measurement, the meter itself may come into contact with blood. Used meters therefore carry a risk of infection. Please dispose of your used meter – after removing the battery – according to the regulations applicable in your country. For information about correct disposal, please contact your local council and authority.

The meter falls outside the scope of the European Directive 2002/96/EC – Directive on waste electrical and electronic equipment (WEEE).

77

▲!

▲!

Page 80: AccuCheck Meter Manual

Explanation of SymbolsOn packaging, on the type plate, and in the instructions for your Accu-Chek Aviva Meter, you may encounter the following symbols, shown here with their meaning.

Consult the instructions for use Caution (refer to accompanying documents). Please refer to safety-related notes

in the manual accompanying this instrument.

Store at

Manufacturer

Catalogue number

Listed by Underwriter’s Laboratories, Inc.© in accordance with UL 61010A-1 and CAN/CSA C22.2 No.1010-1.

For in vitro diagnostic use

This product fulfills the requirements of Directive 98/79/EC on in vitro diagnostic medical devices

Dispose in domestic waste

3V type 2032

78

Page 81: AccuCheck Meter Manual

The Accu-Chek Multiclix Finger PrickerOn the finger pricker itself and on the packaging of the Accu-Chek Multiclix Finger Pricker, you will encounter the following symbols, shown here with their meaning.

For single use only

Use by / Expiry Date

Lot number

Sterilized using irradiation (with at least 25 kGy)

This product fulfills the requirements of Directive 93/42/EEC on medical devices

2

STERILE R

79

Page 82: AccuCheck Meter Manual

Warranty and Service InstructionsRoche Diagnostics warrants the meter against any material and manufacturing defects that may arise during proper use of the meter. For precise terms and conditions, please see the warranty card in the box. This warranty does not affect any statutory or other rights. The warranty ceases to apply if the meter has been misused or tampered with. Roche undertakes to repair faulty components free of charge or, at their discretion, to provide a fault-free replacement meter. Be sure to fill out the warranty form that came with your system and return it to Roche Diagnostics.

Statutory and other rights: This manufacturer’s warranty is in addition to any statutory or other rights which may be acquired by the customer through his dealings with the meter supplier. The statutory provisions on rights in consumer goods sales in the country of purchase shall apply.

Accu-Chek Multiclix Finger Pricker GuaranteeThe Accu-Chek Multiclix Finger Pricker is guaranteed for life from the date of purchase against defects in materials or workmanship. Please fill in the enclosed warranty card and return it to Roche Diagnostics Limited. Our liability is explicitly limited to the demonstrable manufacturing defects. This guarantee is rendered invalid if the instrument has been improperly operated or otherwise tampered with. The guarantee period is not affected by any claim made under this guarantee.

80

Page 83: AccuCheck Meter Manual

If you have a claim under this guarantee, please return the Accu-Chek Multiclix Finger Pricker and the original sales receipt to:

Customer Services Department, Roche Diagnostics LimitedCharles AvenueBurgess HillWest SussexRH15 9RY Guarantee: The statutory provisions on rights in consumer goods sales in the country of purchase shall apply.

Additional SuppliesTest strips are available from your local Pharmacist only and the control solutions can be obtained via the Accu-Chek Customer Careline only.

Test Strips Accu-Chek Aviva Test Strips

Control Solutions Accu-Chek Aviva Control Solutions

Lancet Drums Accu-Chek Multiclix Lancet Drums

81

Page 84: AccuCheck Meter Manual

Information for Healthcare Professionals – Alternative Site Testing (AST)

Healthcare Professionals: Follow the infection control procedures appropriate for your facility.

Decisions about whether to recommend alternative site testing (AST) should take into account the motivation and knowledge level of the patient and his or her ability to understand the considerations relative to diabetes and AST. If you are considering recommending AST for your patients, you need to understand that there is a potential for a significant difference between fingertip/palm and alternative site blood glucose test results. The difference in capillary bed concentration and blood perfusion throughout the body can lead to sample site-to-site differences in glucose results. These physiological effects vary between individuals and can vary within a single individual based upon his or her behaviour and relative physical condition. Our studies involving AST of adults with diabetes show that most persons will find their glucose level changes more quickly in the fingers’/palms’ blood than the alternative sites’ blood.* This is especially important when glucose levels are falling or rising rapidly. If your patient is used to making treatment decisions based upon fingertip/palm readings, he or she should consider the delay, or lag-time, affecting the reading obtained from an alternative site. *Data on file

References1. Stedman, TL, Stedman’s Medical Dictionary, Medical Dictionary, 27th Edition, 1999, p. 2082.

82

!

Page 85: AccuCheck Meter Manual

Chapter 8: The Accu-Chek Multiclix Finger Pricker

Comfort Dial with Depth Selection

Cap

Finger Pricker

Release Button

Priming Button

New LancetDrum

AST Converter Cap(for alternative site testing)

Lancet Counter

Depth Indicator

83

Page 86: AccuCheck Meter Manual

Notes:• Testing with blood from approved alternative sites requires the AST converter cap. This is

explained later in the chapter.• The best depth setting is the lowest number that lets you get enough blood for a test.

Try different settings to find the best one.• For virtually pain-free testing, always use a new lancet for every test.

The AST converter cap is for testing alternative sites on the body, not the fingertip.

The blue cap is for fingertip testing only.

The Accu-Chek Multiclix Finger Pricker (continued)

84

▲!Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline – Third Edition; CLSI document M29-A3, 2005).

The Accu-Chek Multiclix Finger Pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from more than one person as this poses an infection risk.▲!

Page 87: AccuCheck Meter Manual

Loading a Lancet DrumYou must first load the drum into the finger pricker to get it ready for use.

A used drum has two visible red stripes and the blue ring can rotate on the white drum. You cannot insert a used drum into the finger pricker.

A new drum has a blue ring that does not move.

1. Place your thumb to the side of the notch and remove the cap by pulling it straight off. It is not a twist cap.

85

Page 88: AccuCheck Meter Manual

2. Make sure the release button is not yellow. Insert a new drum, blue ring first, until it clicks into place.

3. Align the notch on the cap with the notch on the device and slide the cap on until it stops.

4. You are now ready to use the first lancet. The lancet counter shows five white bars, meaning you have five new lancets remaining.

Loading a Lancet Drum (continued)

86

Page 89: AccuCheck Meter Manual

5. The depth indicator shows the current depth selection. The higher the number, the deeper the penetration. Twist the comfort dial to adjust the setting. Start at 2 or 3 for soft skin. Try a higher setting if your skin is calloused or thick.

6. Press the priming button like a pen until it clicks. The release button turns yellow. Do not press the release button while pressing the priming button.

87

Page 90: AccuCheck Meter Manual

1. Wash and dry your hands.

2. Press the finger pricker firmly against the side of your fingertip and press the yellow release button.

3. Load the next lancet by twisting the plunger one-quarter turn forward (until it stops), then back again. The lancet counter decreases by one white bar. For safety reasons, once you advance to a new lancet, you cannot go back to a used lancet.

Taking a Fingertip Blood Sample

88

Page 91: AccuCheck Meter Manual

Taking an Alternative Site Blood Sample

1. Place your thumb to the side of the notch and remove the cap by pulling it straight off. It is not a twist cap.

2. Align the notch on the converter cap with the notch on the finger pricker and slide the cap on until it stops.

3. Dial the depth setting to 5 1/2 for forearm, upper arm, thigh, or calf testing. Dial the depth setting to 3 for palm testing. 89

Page 92: AccuCheck Meter Manual

4. Press the priming button like a pen until it clicks. The release button turns yellow. Do not press the release button while pressing the priming button.

5. Press the finger pricker firmly against a fleshy site and press the yellow release button. Immediately lift the finger pricker off the site. Remove the converter cap by pulling it straight off.

Taking an Alternative Site Blood Sample (continued)

90

Page 93: AccuCheck Meter Manual

6. Return the converter cap to the same position over the puncture site. Apply pressure to the site to help the blood drop form properly. Do not allow the cap to touch the blood drop.

7. Load the next lancet by twisting the plunger one-quarter turn forward (until it stops), then back again. The lancet counter decreases by one white bar. For safety reasons, once you advance to a new lancet, you cannot go back to a used lancet.

Note:Once you run a successful test, you may want to find the lowest depth setting that provided enough blood with the least amount of pain.

91

Page 94: AccuCheck Meter Manual

Getting a Good Drop of BloodFingertipWash your hands and dry them well. It helps to massage the fingertip to stimulate blood flow. Gently squeeze your finger at the knuckle, then release, squeeze, then release. This helps to increase the blood flow. You can let your hand hang by your side to increase blood flow.

Alternative Site Testing (AST)You have the option of obtaining a blood sample from other sites on your body besides the fingertip. Blood obtained from a fingertip or palm, fleshy part of the hand under the thumb (thenar) and the little finger (hypothenar), can be used at any time to measure blood glucose. If blood from an alternative site, forearm, upper arm, thigh, or calf, is used, there are certain times when testing is not appropriate (see next section). This is because your glucose level changes quicker in your fingertip or palm than in the alternative sites. This difference may cause you to make the wrong therapeutic decisions producing adverse health effects. Please read the following section before you try testing from other sites.92

thenar

hypothenar

Page 95: AccuCheck Meter Manual

93

IMPORTANT• Talk to your healthcare professional about Alternative Site Testing.

Treatment of the Puncture SiteWipe the puncture site with a clean, dry tissue. If it continues to bleed, apply gentle pressure to the site with a plaster to protect it from dirt or infection.

Page 96: AccuCheck Meter Manual

Do not change your treatment because of just one result. NEVER ignore symptoms of high or low blood glucose.

If your blood glucose does not match how you feel, perform a fingertip test to confirm your result. If the fingertip result still does not match how you feel, call your healthcare professional.

Alternative site testing may be done:• Immediately before a meal• Fasting

DO NOT test from an alternative site:• Two hours or less after eating• After exercising• If you are sick• If you think your blood glucose is low• If you often don’t notice when your blood glucose is low• During peak action time of short-acting insulin or rapid-acting insulin analogues• Up to two hours after injecting a short-acting insulin or a rapid-acting insulin analogue

▲!

94

Page 97: AccuCheck Meter Manual

95

Changing the Lancet DrumWhen you have used the sixth and last lancet, you should change the drum.

Notes:• A used drum cannot be reinserted into the finger pricker.• In rare instances, if a used drum is dropped on a hard surface the lancets may protrude

slightly from the drum. If you drop a drum, always pick it up at the red-striped end.

1. Place your thumb to the side of the notch and remove the cap by pulling it straight off. It is not a twist cap.

2. Hold the drum between your thumb and index finger and pull it straight out of the finger pricker. Two red stripes are visible on the white part, indicating that the drum has been used.

Page 98: AccuCheck Meter Manual

96

Caring for Your Finger PrickerDO• Wipe off any blood immediately• Clean the finger pricker and cap regularly• At least once a week, clean the finger pricker and inside

of the cap with a cloth or cotton bud moistened with: • 70% ethanol or 70% isopropanol (available in pharmacies) • Air dry thoroughly

DO NOT• Dip the cap in water or cleaning fluids. This can damage the cap.• Use any cleaning products other than 70% ethanol or 70% isopropanol.

Page 99: AccuCheck Meter Manual

Indexalarm clock function, setting, 38alternative site testing, (AST), 21, 82, 92appetite, increased, 28battery, changing, 64battery, installing, 64battery, type, 65, 75, 78beeper, setting, 35blood glucose, testing, 15blurred vision, 28button, on/off/set, 8code chip, 9, 10computer, connecting the meter to, 45control solution, 48

control test results, understanding, 53control test, acceptable range, 53control test, performing, 49control test, results, 53control test, unacceptable range, 53display check, 67disposing of the meter, 77error messages, 68–74expiry date, 14, 48fatigue, 28flagging test results, 20frequent urination, 28high blood glucose, 28hyperglycemia, 28hypoglycemia, 28

hypoglycemic alarm function, setting, 42low blood glucose, 28maintenance, meter, 64memory, meter, 29numbness, 28product limitations, 75product specifications, 75professional caregivers, 82rapid heartbeat, 28settings, meter, 31supplies, 81sweating, 28symbols, 78–79technical information, 75test results, unusual, 26test strips, Accu-Chek Aviva, 9, 14 97

Page 100: AccuCheck Meter Manual

98

thirst, increased, 28time and date, setting, 32tingling, 28trembling, 28troubleshooting, 68–74warranty, 80

Page 101: AccuCheck Meter Manual

Any Questions?We want to give you the help you need for long and reliable service of your meter.Review your test procedure and make sure that you are performing the steps correctly. If you still have questions, please call the Accu-Chek Customer Careline on 0800 701000 (UK) or 1 800 709600 (Ireland). Please have your Accu-Chek Aviva Meter, Accu-Chek Aviva Test Strips, and Accu-Chek Aviva Control Solutions available when you call.IMPORTANT – READ CAREFULLY BEFORE USEGUARANTEE: The statutory provisions on rights in consumer goods sales in the country of purchase shall apply.

Roche Diagnostics LimitedCharles AvenueBurgess HillWest SussexRH15 9RY +44 (0) 1444 256000

Accu-Chek Customer Careline*0800 701 000 (UK)1 800 709 600 (Ireland)

*Calls may be recorded for training purposes 99

Page 102: AccuCheck Meter Manual

100

The Accu-Chek Aviva System, including the meter, code chip, and test strips, is covered by the following patents: British Patent Nos. 0 505 504 and 2187191, and European Patent No. 0 505 504. British Patent No. 2187191 is licensed from Quadrant Holdings Cambridge Limited.

The Accu-Chek Multiclix Finger Pricker is covered by the following patents: European Patent Nos. 1 034 740, 0 565 970, and 0 782 838.

Page 103: AccuCheck Meter Manual
Page 104: AccuCheck Meter Manual

ACCU-CHEK® Aviva

©2008 Roche Diagnostics. All rights reserved. 04886046001(02)-0907

Roche Diagnostics GmbH,68298 Mannheim,Germanywww.accu-chek.com

ACCU-CHEK, ACCU-CHEK MULTICLIX, and ACCU-CHEK AVIVA are trademarks of Roche.

All other product names and trademarks are the property of their respective owners.