acetic acid ch co h no · acetic acid ch3co2h no.81 2 5 10 20 30 40 50 ppm n=1 81 detecting layer...
TRANSCRIPT
-
2 5 10 20 30 40 50 ppm
n=1
81
Acetic Acid CH3CO2H No.81
0.2 ppm(2 pump strokes)Pink → YellowTemperature 0 to 40 ℃ (32 to 104 ℉) correction not usedRelative humidity 0 to 80 % correction used10 %(for 2 to 10 ppm), 5 %(for 10 to 50 ppm)10 tubes for 10 tests36 months
Detecting limit :Colour change :Operating conditions :
Relative standard deviation :Tube quantity and number of tests per box :Shelf life :
Performance
1 to 2 ppm 2 to 50 ppm 50 to 100 ppm 2(200 mL) 1(100 mL) 1/2(50 mL) 1/2 1 2 1.5 min 45 sec 30 sec
Measuring rangeNumber of pump strokesCorrection factorSampling time
CH3CO2H + Base → Reaction productReaction principle
Possible coexisting substances and their interferences
Chlorine ≧ 1/2 +Hydrogen chloride ≧ 3 times +Hydrogen cyanide ≧ 3 times + YellowNitric acid ≧ 3 times +Nitrogen dioxide ≧ 1/2 +
Sulphur dioxide ≧ 1/2 +
Substance Concentration Interference Changes colour by itself to
Other substances measurable with this detector tube
Acetic anhydride Factor : 0.3 1 0.6 to 15 ppmAcrylic acid Factor : 1.0 1 .02 to 50 ppmFormic acid Factor : 2.6 1 5.2 to 130 ppmIsovaleric acid Factor : 1.0 1 .02 to 50 ppmMaleic anhydride Factor : 0.4 1 0.8 to 20 ppmMethacrylic acid Factor : 0.9 1 1.8 to 45 ppmPropionic acid Factor : 1.5 1 .03 to 75 ppm
Substance Correction No. of pump strokes Measuring range
Diffusion tube methodCalibration gas generation
Scale