Introduction Abnormal nocturnal blood pressure (BP) such as non-dipping or nocturnal hypertension (reverse-dipping) represents a potent marker for current and future cardiovascular risks. Accurate BP measurement reflecting “sleep” BP is important. Standard cuff-based ambulatory nocturnal BP measurement yields limited data points potentially resulting in imprecise results. Hypothesis Younghoon Kwon 1 , Martin Baruch 2 , Sebastian Estrada 1 , Elizabeth Harrison 1 , Ronald Williams 1 , Yeilim Cho 1 , David Gerdt 2 , Heather Bonner 1 , Sula Mazimba 1 , Richard Berry 3 1 University of Virginia; 2 Caretaker Medical; 3 University of Florida Conclusion Adoption of Continuous Beat-to-Beat Nocturnal Blood Pressure Measurement in Ambulatory Blood Pressure Monitoring Disclosure: Younghoon Kwon has received research funding for the study from Caretaker Medical (Charlottesville, VA) Acknolwedgement: This work was partly supported by NIH (R21- HL140432) during this work Characteristics of study participants ( n = 13) Methods Spot check nocturnal BP does not capture true average nocturnal BP as measured by continuous beat to beat BP recording. Design: Prospective observational Subjects: Patients undergoing clinically indicated in-lab polysomnography (PSG) BP measurement: Beat-to-beat fashion by noninvasive Caretaker® device that uses a pulse decomposition analysis algorithm Sleep wakefulness status was ascertained by PSG Analysis: Comparison between “every 30 min periodic systolic BP data starting at the onset of BP recording ” and “ average BP data from entire 30 min continuous beat-to-beat BP measurement ” both by Caretaker® for entire recording time and by sleep state (sleep vs. awake) BP variability defined as SD of the rolling difference between the 100 point mean of systole on continuous BP measurement by the patient’s sleep state Results Age:52 (12) yo, Male (8/13), BMI 34, HTN (11/13) Diagnostic PSG: 9/13, Split night PSG: 4/13 Sleep characteristics of study participants ( n = 13) Any OSA (AHI>5) present: 10/13 Significant OSA (AHI>15) present: 6/13 Average total sleep time: 385 min Main results ( n = 13) Mean SBP of Sleep and Awake were similar [135.3 (19.8) vs. 134.5 (18.7) mmHg, P=0.45] ( Figure 1- Box Plot) Mean nocturnal SBP by periodic BP measurement was higher compared with beat-to-beat-derived average BP (139.1 mmHg [20.3] vs. 134.7 [19.5], p< 0.0001) (Figure 2 ) The difference between the two methods remained similar when continuous BP was derived from sleep vs. awake period (4.5 mmHg [5.9] vs. 5.0 [5.6], p= 0.75) BP variability was more pronounced during awake compared with sleep period (4.6 mmHg [1.3] versus 3.3 [1.3], p<0.0001) Figure 1 Sample BP trend by sleep/awake state (Example from 3 subjects) Figure 2 Awake Sleep 80 100 120 140 160 180 (mmHg) Mean SBP comparision P=0.45 0 50 100 150 200 SBP (mmHg) Intermittent measurement Continuous beat-to-beat measurement p < 0.0001 SBP comparison between hypothetical spot check BP vs. continuous beat-to-beat (Example) BP change in sleep in the context of respiratory events and wakefulness (Example) Figure 3 Figure 4 Figure 5 Figure 6 Acute BP change during sleep (example) Periodic BP measurement currently adopted by standard ambulatory nocturnal BP may not yield true nocturnal sleep BP pattern due to its spot-check nature and lack of sleep- awake information Incorporation of beat-to-beat continuous BP measurement along with sleep recording can provide more accurate and meaningful nocturnal BP information

Upload: others

Post on 19-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Adoption of Continuous Beat-to-Beat Nocturnal Blood ...€¦ · Introduction •Abnormal nocturnal blood pressure (BP) such as non-dipping or nocturnal hypertension (reverse-dipping)

Introduction• Abnormal nocturnal blood pressure (BP) such as non-dipping

or nocturnal hypertension (reverse-dipping) represents a potent marker for current and future cardiovascular risks.

• Accurate BP measurement reflecting “sleep” BP is important.

• Standard cuff-based ambulatory nocturnal BP measurement yields limited data points potentially resulting in imprecise results.

Hypothesis

Younghoon Kwon1, Martin Baruch2, Sebastian Estrada1, Elizabeth Harrison1, Ronald Williams1, Yeilim Cho1, David Gerdt2, Heather Bonner1, Sula Mazimba1, Richard Berry3

1University of Virginia; 2Caretaker Medical; 3University of Florida

Conclusion

Adoption of Continuous Beat-to-Beat Nocturnal Blood Pressure Measurement in Ambulatory Blood Pressure Monitoring

Disclosure: Younghoon Kwon has received research funding for the study from Caretaker Medical (Charlottesville, VA)

Acknolwedgement: This work was partly supported by NIH (R21- HL140432) during this work

Characteristics of study participants ( n = 13)

Methods

• Spot check nocturnal BP does not capture true average nocturnal BP as measured by continuous beat to beat BP recording.

• Design: Prospective observational

• Subjects: Patients undergoing clinically indicated in-lab polysomnography (PSG)

• BP measurement: Beat-to-beat fashion by noninvasive Caretaker® device that uses a pulse decomposition analysis algorithm

• Sleep wakefulness status was ascertained by PSG

• Analysis:

• Comparison between “every 30 min periodic systolic BP data starting at the onset of BP recording” and “average BP data from entire 30 min continuous beat-to-beat BP measurement” both by Caretaker® for entire recording time and by sleep state (sleep vs. awake)

• BP variability defined as SD of the rolling difference between the 100 point mean of systole on continuous BP measurement by the patient’s sleep state

Results

• Age:52 (12) yo, Male (8/13), BMI 34, HTN (11/13)• Diagnostic PSG: 9/13, Split night PSG: 4/13

Sleep characteristics of study participants ( n = 13)

• Any OSA (AHI>5) present: 10/13• Significant OSA (AHI>15) present: 6/13• Average total sleep time: 385 min

Main results ( n = 13)

• Mean SBP of Sleep and Awake were similar [135.3 (19.8) vs. 134.5 (18.7) mmHg, P=0.45] (Figure 1- Box Plot)

• Mean nocturnal SBP by periodic BP measurement was higher compared with beat-to-beat-derived average BP (139.1 mmHg [20.3] vs. 134.7 [19.5], p< 0.0001) (Figure 2)

• The difference between the two methods remained similar when continuous BP was derived from sleep vs. awake period (4.5

mmHg [5.9] vs. 5.0 [5.6], p= 0.75)• BP variability was more pronounced during awake compared with

sleep period (4.6 mmHg [1.3] versus 3.3 [1.3], p<0.0001)

Figure 1

Sample BP trend by sleep/awake state (Example from 3 subjects)

Figure 2

A w a k e S le e p8 0

1 0 0

1 2 0

1 4 0

1 6 0

1 8 0

(mm

Hg

)

M e a n S B P c o m p a r is io n

P = 0 .4 5

0

5 0

1 0 0

1 5 0

2 0 0

SB

P (

mm

Hg

)

In te rm it te n t m e a s u re m e n tC o n t in u o u s b e a t- to -b e a t m e a s u re m e n t

p < 0 .0 0 0 1

SBP comparison between hypothetical spot check BP vs. continuous beat-to-beat (Example)

BP change in sleep in the context of respiratory events and wakefulness (Example)

Figure 3

Figure 4

Figure 5

Figure 6

Acute BP change during sleep (example)

• Periodic BP measurement currently adopted by standard ambulatory nocturnal BP may not yield true nocturnal sleep BP pattern due to its spot-check nature and lack of sleep-awake information

• Incorporation of beat-to-beat continuous BP measurement along with sleep recording can provide more accurate and meaningful nocturnal BP information