Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung 1 , Emilie Soisson 1,2 , Hugo Bouchard 3,4 , and Jan Seuntjens 1 1 Medical Physics Unit,McGill University, Montréal, Québec, Canada 2 Dept. of Medical Physics,McGill University Health Centre, Montréal, Québec, Canada 3 Département de Physique, Université de Montréal, Montréal, Québec, Canada 4 Département de Radio-Oncologie, Centre Hospitalier de l’Université de Montréal, Montréal, Québec, Canada 2010 IAEA IDOS Symposium, Vienna, Austria November 10 th , 2010

Upload: others

Post on 04-Jun-2020

16 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Advanced Dosimetry Techniques

for Accurate Verification of

Nonstandard Beams

Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4, and Jan Seuntjens1

1Medical Physics Unit, McGill University, Montréal, Québec, Canada

2Dept. of Medical Physics, McGill University Health Centre, Montréal, Québec, Canada

3Département de Physique, Université de Montréal, Montréal, Québec, Canada

4Département de Radio-Oncologie, Centre Hospitalier de l’Université de Montréal,

Montréal, Québec, Canada

2010 IAEA IDOS Symposium, Vienna, Austria

November 10th, 2010

Page 2: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Introduction

� Common use of small and composite non-uniform fields

in radiotherapy

� IMRT, tomotherapy, cyberknife, gammaknife, etc.

� Standard dosimetry protocols

� AAPM TG-51(1) and IAEA TRS-398(2)

� Large dosimetric errors for small and non-uniform field

calibration

(1) Almond et al., Med. Phys. 26:1847 (1999)(2) Andreo et al., International Atomic Energy Agency, Vienna, Austria (2000)

Page 3: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Fraser et al., J. Appl. Clin. Med. Phys. 10: 243 (2009)

Page 4: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Introduction

� New methods required for accurate beam calibration in small and non-uniform (nonstandard) fields

� Common use of small and composite non-uniform fields

in radiotherapy

� IMRT, tomotherapy, cyberknife, gammaknife, etc.

� Standard dosimetry protocols

� AAPM TG-51(1) and IAEA TRS-398(2)

� Large dosimetric errors for small and non-uniform field

calibration

(1) Almond et al., Med. Phys. 26:1847 (1999)(2) Andreo et al., International Atomic Energy Agency, Vienna, Austria (2000)

Page 5: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Objectives

� Apply established dosimetry techniques(3) for accurate

dose measurements in various small and nonstandard

fields.

� Determine the correction factors for these fields based on

the new dosimetry formalism.

(3) Chung et al., Med. Phys., 37:2404 (2010)

Page 6: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,
Page 7: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Reference fieldf ref

Machine specificreference field fmsr

Plan-class specific reference field fpcsr

(4) Stewart et al., Phys. Med. Biol., 52: 3089 (2007)

(4)Gafchromic ®

EBT filmDiamonddetector

GLIC-03 microLion(PTW)

Page 8: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Chung et al. Med. Phys. 37:2404 (2010)

Page 9: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

(c)TomoTherapy®

TomoTherapy®

(c)TomoTherapy®

85 cm

85 cm

Thomas et al., Med. Phys. 32, 1346-1353 (2005)

Page 10: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Characterization of reference detectors

Method: Bouchard et al., Med. Phys., 36:1931 (2008) Chung et al., Med. Phys. 37:3265 (2010)

Page 11: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Chung et al. Proc.

of IAEA IDOS,

submitted (2010)

Page 12: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Correction factor

(5)

(5) Med. Phys. 32:570-577 (2005)

Farmer-type chambers smaller ionization chambers

( fmsr= 5×10 cm2 )

Page 13: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Candidate pcsr fields

� A TomoTherapy®-based IMRT delivery

VolumeSize

(cm3)

Prescribed

Dose (Gy)Importance

DVH

Vol. (%)

Ave. (1SD)

(Gy)

Homogeneity

Index (HI)

PTV-HD 78.42 70 300 97 71.09 (0.45) 0.025

PTV-ED 585.22 56 300 97 58.78 (3.40) 0.248

spinal

cord21.00 25 50 50 26.55 (12.19) 1.432

40 % (28 Gy)50 % (35 Gy)60 % (42 Gy)70 % (49 Gy)80 % (56 Gy)90 % (63 Gy)100 % (70 Gy)

A

P

R L R L

S

I I

S

A P

PTV-HDPTV-ED

spinal cordaxial coronal sagittal

Page 14: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Ionization

ChamberExradin A12 NE2571

Exradin

A1SLExradin A14

PinPoint®

31006

Measurement

1

0.995

±0.0040.997

±0.0041.001

±0.0040.994

±0.0040.997

±0.004

Measurement

2

0.995

±0.0030.993

±0.0030.993

±0.0031.002

±0.0030.994

±0.004

Correction factor

Meas. 1

Meas. 2

Farmer-type chambers smaller ionization chambers

Chung et al. Proc.

of IAEA IDOS,

submitted (2010)

Page 15: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Various nonstandard field deliveries

100 %: Dave at the small PTV

20 %30 %40 %50 %60 %70 %80 %90 %95 %100 %

small PTV

Page 16: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Correction factor

Page 17: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Correction factor

Page 18: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Correction factor

Page 19: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,

Conclusions

� Correction factor measurement: within 0.4% 1σ uncertainty

� Msr correction factor

� agreed well between the measurement and calculation for

Farmer-type chambers

� Pcsr correction factor

� close to unity for homogeneous dose delivery

� large positioning uncertainty for heterogeneous dose delivery

� This work suggests the use of a qualifier of plan

homogeneity such as HI or MF to uniquely specify pcsr

correction factors.

Page 20: Advanced Dosimetry Techniques for Accurate Verification of ...Advanced Dosimetry Techniques for Accurate Verification of Nonstandard Beams Eunah Chung1, Emilie Soisson1,2, Hugo Bouchard3,4,