Adventures in
Sample Introduction
for ICP-OES and ICP-MS
Adventures in
Sample Introduction
for ICP-OES and ICP-MS
Geoffrey N. ColemanMeinhard Glass Products
A Division of Analytical Reference Materials International
![Page 2: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/2.jpg)
2
Sample Introduction Components
•ICP Torches
•Spray Chambers
•Nebulizers•Conventional
•High Efficiency
•Direct injection
•Accessories
![Page 3: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/3.jpg)
3
Overview
•Brief review
•Components•Torches
•Spray chambers
•Nebulizers
•What’s new....
![Page 4: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/4.jpg)
4
References
Richard F. Browner, Georgia Institute of Technology
Anders G.T. Gustavsson, Swedish Institute of Technology
Jean-Michel Mermet, Universite Claude Bernard-Lyon, France
Akbar Montaser, George Washington University
John W. Olesik, Ohio State University
Barry L. Sharp, Macauley Land Use Institute, Scotland “Pneumatic Nebulizers and Spray Chambers for Inductively Coupled Plasma Spectroscopy”, Journal of
Analytical Atomic Spectrometry, 1988, 3, 613 – 652 (Part 1); 939 – 963 (Part 2).
![Page 5: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/5.jpg)
5
Processes
Starting with a “homogeneous” solution sample....
•Nebulization
•Desolvation
•Dissociation
•ExcitationAll require energy and time.There is a “domino” effect.
![Page 6: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/6.jpg)
6
Interferences
•Nebulization
•Desolvation
•Dissociation
•Excitation
Probably 85% of significant interferences occur at nebulization, due to changes in surface tension, density, and viscosity.These are multiplicative interferences.
![Page 7: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/7.jpg)
7
Mean Droplet Size
d3,2
0.5
0.5
0.45 3l
g
1.5585
V597
10 Q
Q +
( )
NUKIYAMA AND TANASAWA EQUATION
d3,2 = Sauter mean diameter - (m)
V = Velocity difference of gas-liquid - (m/s)
= Surface tension - (dyn/cm)
= Liquid density - (g/cm3)
= Liquid viscosity - (Poise or dyn·s/cm2)
Ql = Volume flowrate, liquid - (cm3/s)
Qg = Volume flowrate, gas - (cm3/s)
S. Nukiyama and Y. Tanasawa, Trans. Soc. Mech. Eng., Tokyo, 1938-40, Vol. 4 – 6, Reports 1 – 6.
![Page 8: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/8.jpg)
8
Rule-of-Thumb
When the Total Dissolved Solids exceeds about 1000 ppm, changes in surface tension, density, and viscosity begin to affect the droplet size distribution and, thus, the slope of the analytical calibration curve.
![Page 9: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/9.jpg)
9
Interferences
Control by:
•Matrix Removal – usually not practical
•Swamping – risk of contamination
•Matrix Matching – probably most useful
• Internal Standard – line selection
•Method of Standard Additions – most tedious and time-consuming
![Page 10: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/10.jpg)
10
Single Droplet Studies
• Desolvation begins
• Evaporation from surface
• Droplet diameter diminishes
• Crust forms as solvent evaporates
•Internal pressure builds
•Droplet explodes
•Escaping water vapor cools immediate surroundings
•Particles dehydrate
•Particles evaporate
![Page 11: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/11.jpg)
11
Implications
•Large Surface Area/Volume
•Small Droplets•Faster desolvation and vaporization
•Narrow Size Distribution•Consistent desolvation and vaporization
•Well-defined excitation/observation zones
•Virtually no signal comes from droplets larger than 8 - 10 m
•Most signal comes from < 3 m.
![Page 12: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/12.jpg)
12
ICP Plasma Torches
•Tg 6000 – 9000 K
•Skin Effect•Electric
•Magnetic
•Pressure/Temperature
•Injection Velocity3 – 5 m/sec to overcome skin effectsInjector diameter 1.0 – 2.4 mm i.d.Carrier at 0.7 – 1.0 L/min
•Residence Time
•Highly Volatile Solvents
•Chemical Interferences
•Viewing Zone
![Page 13: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/13.jpg)
13
ICP Plasma Torches
End-on Viewing
•Must remove “tail flame”
•Ground state atoms
•Molecular species
•Larger injector diameters – longer residence time
•Significant chemical interferences
•Significant sensitivity improvement – up to 10x
![Page 14: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/14.jpg)
14
ICP Plasma Torches
•Outside: 16 – 18 mm
•Inner – Outer Gap: 0.5 – 1.0 mm
•Injector: 1.0 – 4.0 mm• 1.0 mm for volatile solvents
• 2.0 mm general purpose radial torch
• 2.4 mm general purpose axial torch
•Demountable Injectors• Ceramic (alumina) or sapphire for HF
• Flexibility
• Complexity
• Cost
![Page 15: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/15.jpg)
15
ICP Spray Chambers
Aerosol Conditioning• Remove droplets larger than 20 um
•Gravitational settling
•Inertial impaction
•Evaporation
•Recombination
• Reduce aerosol concentration
• Modify aerosol phase equilibria
• Modify aerosol charge equilibria
• Reduce turbulence of nebulization
![Page 16: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/16.jpg)
16
ICP Spray Chambers
Particle Motion in a Spray Chamber
![Page 17: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/17.jpg)
17
ICP Spray Chambers
Scott Double-Pass• Large volume (> 100 mL)
• Large surface area•Phase equilibria
• Stagnant areas
• Long stabilization time
• Long washout
• Drainage
![Page 18: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/18.jpg)
18
ICP Spray Chambers
Cyclonic with Baffle• Moderate volume: 50 mL
• Moderate surface area
• Entire volume swept by carrier flow
• Fast equilibration
• Fast washout
• Sensitivity enhanced by 1.2 – 1.5x
• Now most common type
![Page 19: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/19.jpg)
19
ICP Spray Chambers
•Desolvation begins in the spray chamber•Extent affects droplet size
•Affects amount transported to the plasma
•Maintain constant temperature
•Liquid on the walls must equilibrate with vapor•Minimize surface area
•Drain away excess quickly
![Page 20: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/20.jpg)
20
ICP Spray Chambers
•Speciation begins in the spray chamber•Volatile species in gas phase are more
efficiently transported than droplets
•Nebulization does not control the rate of sample introduction
•Cool spray chamber (especially for organic solvents)
•Minimize surface area
![Page 21: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/21.jpg)
21
Nebulizers
•Pneumatic•Self-aspirating
• Concentric
• Cross-flow
•Non-aspirating• Babington
• V-groove
• GEM Cone
• MiraMist
• Grid
• Fritted
•Other•Ultrasonic nebulizer
•Thermospray
•Spark ablation
•Laser ablation
•Specialty•HEN, MCN, MicroMist
•DIHEN, DIN
![Page 22: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/22.jpg)
22
Mean Droplet Size
NUKIYAMA AND TANASAWA EQUATION
d3,2 = Sauter mean diameter - (m)
V = Velocity difference of gas-liquid - (m/s)
= Surface tension - (dyn/cm)
= Liquid density - (g/cm3)
= Liquid viscosity - (Poise or dyn·s/cm2)
Ql = Volume flowrate, liquid - (cm3/s)
Qg = Volume flowrate, gas - (cm3/s)
S. Nukiyama and Y. Tanasawa, Trans. Soc. Mech. Eng., Tokyo, 1938-40, Vol. 4 – 6, Reports 1 – 6.
d3,2
0.5
0.5
0.45 3l
g
1.5585
V597
10 Q
Q +
( )
![Page 23: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/23.jpg)
23
Self-Aspirating Nebulizers
•Concentric•Gouy design (1897)
•Efficiency approaching 3%
•Glass
•Quartz
•Teflon
•Cross-flow•Efficiency approaching 2.5%
•Glass
•Sapphire
![Page 24: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/24.jpg)
24
Self-Aspirating Nebulizers
Glass Concentric
![Page 25: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/25.jpg)
25
Self-Aspirating Nebulizers
Glass Concentric
![Page 26: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/26.jpg)
26
Self-Aspirating Nebulizers
![Page 27: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/27.jpg)
27
Self-Aspirating Nebulizers
![Page 28: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/28.jpg)
28
Self-Aspirating Nebulizers
Cross-flow
![Page 29: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/29.jpg)
29
Non-aspirating Nebulizers
•Original Babington Design (1973)
•Very inefficient
•Could nebulize “anything”
•V-groove (Suddendorf, 1978)
•Much improved efficiency, > 1%
•Best choice for analysis of slurries
•Best choice for analysis of used oils
•Grid (Hildebrand, 1986)
•Efficiency approaching 4.5%
•Very difficult to maintain
![Page 30: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/30.jpg)
30
Non-aspirating Nebulizers
V-groove (Babington)
![Page 31: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/31.jpg)
31
Non-aspirating Nebulizers
• GEM Cone (PerkinElmer)
•Efficiency ~ 1.2%
• MiraMist/Parallel-Path (Burgener)
•Efficiency approaching 3 %
![Page 32: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/32.jpg)
32
Non-aspirating Nebulizers
• MiraMistParallel-Path
![Page 33: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/33.jpg)
33
Non-aspirating Nebulizers
Ultrasonic Nebulizer• Efficiency approaches 30%
• Sensitivity improves ~10x
• Droplet size < 5 m
• Potentially heavy solvent load
• Desolvation essentialMembrane separator available
• Desolvation interferences occur (eg., As III vs. As IV)
• Does not handle high solids well
![Page 34: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/34.jpg)
34
Sample Introduction Accessories
Desolvation: Apex Q from Elemental Scientific
• Sensitivity improves ~10x
• Uses concentric nebulizer and cyclonic spray chamber
• Desolvation interferences
• High solids problematic
• Available in HF-resistant version
![Page 35: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/35.jpg)
35
Sample Introduction Accessories
Spray Chamber Cooling: PC3 from Elemental Scientific• Sensitivity improves
• Reduces solvent loading
• Reduces oxide interferences in ICPMS
• Uses concentric nebulizer and cyclonic spray chamber
• Available in HF-resistant version
![Page 36: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/36.jpg)
36
Sample Introduction Accessories
•Fit Kits couple liquid and gas supplies to the nebulizer
•Especially useful for high pressure nebulizers
![Page 37: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/37.jpg)
37
The MEINHARD®
Nebulizer
Type A •Lapped ends – capillary
and nozzle flush
•Simple, monolithic design
Type C •Recessed capillary for
higher TDS tolerance
•Vitreous, fire-polished ends
•Stronger suction
Type K •Recessed capillary
•Lapped ends
•Lower Ar flow: 0.7 L/min
![Page 38: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/38.jpg)
38
The MEINHARD®
Nebulizer
0
0.2
0.4
0.6
0.8
1
1.2
%R
SD
Cd Cu Fe Mn
Element
TR-30-A3(8)
TR-30-A3(1)
TR-30-C1(12)
TR-30-K2/3(22)
0
20
40
60
80
100
120
140
160
180
200
PP
B
Cd Cu Fe Mn
Element
TR-30-A3(8)
TR-30-A3(1)
TR-30-C1(12)
TR-30-K2/3(22)
0
0.5
1
1.5
2
2.5
3
3.5
4
PP
B
Cd Cu Fe Mn
Element
TR-30-A3(8)
TR-30-A3(1)
TR-30-C1(12)
TR-30-K2/3(22)
0
2000
4000
6000
8000
10000
12000
Co
un
ts
Cd Cu Fe Mn
Element
TR-30-A3(8)
TR-30-A3(1)
TR-30-C1(12)
TR-30-K2/3(22)
Inte
nsit
y,
40
pp
bP
recis
ion
, 40
pp
bB
EC D
L
![Page 39: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/39.jpg)
39
The MEINHARD®
Nebulizer
Type A •Lapped ends – capillary
and nozzle flush
•Simple, monolithic design
Type C •Recessed capillary for
higher TDS tolerance
•Vitreous, fire-polished ends
•Stronger suction
Type K •Recessed capillary
•Lapped ends
•Lower Ar flow: 0.7 L/min
![Page 40: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/40.jpg)
40
Glass Concentric Nebulizer
•Advantages•Simple, single piece desgin
•All glass design, inert
•Permanently aligned - self aligning
• Easy to use
•Disadvantages•Low efficiency ( ~3%)
•Glass attacked by HF
•High or undissolved solids may clog capillary
![Page 41: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/41.jpg)
41
HF-Resistant Nebulizers
•Concentric nebulizers in Teflon PFA and Polypropylene from Elemental Scientific
•Typical flows: 50 – 700 L/min; 1 L/min
•Integral or demountable solution tubing
•Efficiency: 2 – 3%
MicroFLOW PFA
PolyPro
![Page 42: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/42.jpg)
42
HF-Resistant Kits
Complete Kits include:
•Demountable Torch
•Pt or Sapphire Injector
•Adapter
•Teflon PFA Spray Chamber
•Teflon PFA or Polypropylene Nebulizer
![Page 43: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/43.jpg)
43
Nebulizers
![Page 44: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/44.jpg)
44
Nebulizers
![Page 45: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/45.jpg)
45
Mean Droplet Size
NUKIYAMA AND TANASAWA EQUATION
d3,2 = Sauter mean diameter - (m)
V = Velocity difference of gas-liquid - (m/s)
Ql = Volume flowrate, liquid - (cm3/s)
Qg = Volume flowrate, gas - (cm3/s)
•Adjust annulus to increase V, but maintain Qg
•Adjust capillary to decrease Ql
d3,2
0.5
0.5
0.45 3l
g
1.5585
V597
10 Q
Q +
( )
![Page 46: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/46.jpg)
46
High Efficiency Nebulizer
Type A HEN
![Page 47: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/47.jpg)
47
High Efficiency Nebulizer
![Page 48: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/48.jpg)
48
High Efficiency Nebulizer
PN: TR-30-A3 MicroConcentric Nebulizer (Cetac) MicroMist (Glass Expansion)
![Page 49: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/49.jpg)
49
High Efficiency Nebulizer
•The HEN normally aspirates 30 – 300 L/min
•Design gas flow is 1 L/min of argon
•Normal operating pressure is 170 psi, 150 and 90 psi versions are available.
![Page 50: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/50.jpg)
50
High Efficiency Nebulizer
•Under normal operating conditions, a HEN exhibits a D3,2 of 1.2 – 1.5 m
•“Starved” TR-30-A3 exhibits D3,2 of 3.2 – 4.2 m
•Normal operating conditions for a TR-30-A3 yield a mean droplet size of about 15 m
![Page 51: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/51.jpg)
51
High Efficiency Nebulizer
![Page 52: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/52.jpg)
52
High Efficiency Nebulizer
•Type A Nozzle Geometry
•Smaller Sample Uptake Capillary
Liquid flow rate from 10-1200 l/min
•Small Bore Sample InputLow Dead Volume Connection (LC, CZE)
•Smaller Gas Annular AreaHigher Ar pressure - 150 psig
![Page 53: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/53.jpg)
53
High Efficiency Nebulizer
Applications:•Chromatography detection
•Capillary electrophoresis•Liquid chromatography
•Limited sample volume
•Minimize speciation interferences•Very high analyte transport•Much less discrimination between volatile
species and dissolved species
![Page 54: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/54.jpg)
54
Direct Injection HEN
•DIHEN is designed to be inserted directly into a demountable torch
•DIHEN is dimensionally similar to HEN (see table, slide 47)
•DIHEN is operationally similar to HEN, except•Normal carrier flow is 0.2 – 0.4 L/min
•Minimize speciation interferences• Easily introduce highly volatile solvents• Essentially 100% transport• Large-Bore version less prone to clogging, but noisy
![Page 55: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/55.jpg)
55
DIHEN
•Typical demountable torch with DIHEN in place
•Detection limits better than conventional pneumatic nebulizer
•Detection limits not as good as HEN
![Page 56: Adventures in Sample Introduction for ICP-OES and ICP-MS Geoffrey N. Coleman Meinhard Glass Products A Division of Analytical Reference Materials International](https://reader033.vdocument.in/reader033/viewer/2022061614/56649d835503460f94a68e46/html5/thumbnails/56.jpg)