afe4490 with phototransistor – test and measurement results · ti healthtech engineering...
TRANSCRIPT
![Page 1: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/1.jpg)
TI HealthTech Engineering components for life.
Health, Fitness, Imaging
AFE4490 with DCM03 – Measurement Results
Aug 27, 2013
![Page 2: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/2.jpg)
The slide set presents the PPG data using DCM03 (Dual LEDs + Photodiode module) with AFE4490.
The tests were carried out on the thumb and the wrist of 1 subject.
The PPG results were captured using a DCM03 sensor board with watch board.
Results:
Summary
Left Thumb Left Wrist Set current (mA)
Actual current (mA) TIA gain DC shift
2nd stage gain IR DC (V)
IR pk-pk (mV) TIA gain DC shift
2nd stage gain IR DC (V)
IR pk-pk (mV)
DCM03
0.25 0.294 500k 0uA 4 0.601 20 1M 0uA 4 0.598 2 1 0.88 100k 0uA 4 0.525 10 250k 0uA 4 0.658 3
2.5 2.35 100k 0uA 4 0.89 6 5 5 50k 0uA 4 1.14 5
10 10 25k 0uA 4 1.14 5
![Page 3: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/3.jpg)
Measurement Setup Hardware
![Page 4: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/4.jpg)
DCM03 Sensor board
![Page 5: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/5.jpg)
Sensor board & Watch board setup
Rear side of watch setup Front side of watch setup
![Page 6: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/6.jpg)
Hardware setup on Thumb
![Page 7: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/7.jpg)
PPG Data – Thumb (0.294mA)
Test Conditions: PRF = 500 Hz Duty cycle = 25 % Rf = 500 KΩ Cf = 5 pF ILED = 0.294 mA Stage 2 Gain = 4 Test Subject = Subject1 Meas. Site = Thumb
![Page 8: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/8.jpg)
PPG Data – Thumb (0.88mA)
Test Conditions: PRF = 500 Hz Duty cycle = 25 % Rf = 100 KΩ Cf = 5 pF ILED = 0.88 mA Stage 2 Gain = 4 Test Subject = Subject1 Meas. Site = Thumb
![Page 9: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/9.jpg)
Hardware setup on wrist
![Page 10: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/10.jpg)
PPG Data – Wrist (0.294mA)
Test Conditions: PRF = 500 Hz Duty cycle = 25 % Rf = 1MΩ Cf = 5 pF ILED = 0.294 mA Stage 2 Gain = 4 Test Subject = Subject1 Meas. Site = Wrist
![Page 11: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/11.jpg)
PPG Data – Wrist (0.88mA)
Test Conditions: PRF = 500 Hz Duty cycle = 25 % Rf = 250KΩ Cf = 5 pF ILED = 0.88 mA Stage 2 Gain = 4 Test Subject = Subject1 Meas. Site = Wrist
![Page 12: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/12.jpg)
PPG Data – Wrist (2.35mA)
Test Conditions: PRF = 500 Hz Duty cycle = 25 % Rf = 100KΩ Cf = 5 pF ILED = 2.35 mA Stage 2 Gain = 4 Test Subject = Subject1 Meas. Site = Wrist
![Page 13: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/13.jpg)
PPG Data – Wrist (5mA)
Test Conditions: PRF = 500 Hz Duty cycle = 25 % Rf = 50KΩ Cf = 5 pF ILED = 5 mA Stage 2 Gain = 4 Test Subject = Subject1 Meas. Site = Wrist
![Page 14: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/14.jpg)
PPG Data – Wrist (10mA)
Test Conditions: PRF = 500 Hz Duty cycle = 25 % Rf = 25KΩ Cf = 5 pF ILED = 10 mA Stage 2 Gain = 4 Test Subject = Subject1 Meas. Site = Wrist
![Page 15: AFE4490 with Phototransistor – Test and measurement results · TI HealthTech Engineering components for life. Health, Fitness, Imaging AFE4490 with DCM03 – Measurement Results](https://reader036.vdocument.in/reader036/viewer/2022071017/5fd139bd52663827b87c1bac/html5/thumbnails/15.jpg)
Left Thumb Left Wrist Set current (mA)
Actual current (mA) TIA gain DC shift 2nd stage IR DC (V) IR pk-pk (mV) TIA gain DC shift 2nd stage IR DC (V) IR pk-pk (mV)
DCM03
0.25 0.294 500k 0 4 0.601 20 1M 0 4 0.598 2 1 0.88 100k 0 4 0.525 10 250k 0 4 0.658 3
2.5 2.35 100k 0 4 0.89 6 5 5 50k 0 4 1.14 5
10 10 25k 0 4 1.14 5
Summary data