ahd1601-01
TRANSCRIPT
-
8/8/2019 AHD1601-01
1/24
Echocardiography relies on the use o ultrasoundhigh-requency sound wavestointerrogate bodily tissues. Ultrasound waves penetrate through and are reected by tissuesbeore returning to the transducer, where they are then converted into images. Modern ultra-
sound equipment uses three principal modalities. Tey include M-mode imaging, B-mode (or
two-dimensional imaging), and Doppler imaging. Tis chapter will introduce the basic physical
principles o ultrasound and tomographic anatomy in cardiac imaging with ultrasound.
Basic Principles of
Echocardiography andTomographic AnatomyBernard E. Bulwer, Jose Rivero, and Scott D. Solomon
Figure 1-1. Ultrasound waves used in medical ultrasound imaging are generated by the piezoelectric
elements housed within the transducer (left panel). In order to generate an ultrasound image, the ultra-
sound pulse must make a round trip between the piezoelectric elements (housed inside the transducer),the scan line, and the region o interest (imaged structure). Te ultrasound waves that are emitted
travel along scan lines and interact with tissue structures. Ultrasound waves or echoes are dierentially
reected rom cardiac structures and returned to transducer where they are then converted to electrical
impulses and subsequently converted to images by system components (eg, signal processor and scan
converters) housed within the ultrasound equipment. wo-dimensional images representing cardiac
cross-sectional anatomy are composed o se
scan lines (see Fig. 1-3).
Te principal method used in order to prod
ultrasound images is reerred to as B-mode or b
ness mode imaging (right panel). Echocardiogr
images are processed and displayed as pixels, orpicture elements, o varied shades o gray (black
white). Each pixel has its own anatomical addre
and echoreectivity and is mapped accordingly
Te major components o a modern ultra
scanner include the ollowing: 1) ransducer. T
transducer is the interace between the patien
and the beamormer. Electrical energy is conv
here into ultrasound and returning ultrasound
(echoes) into electrical energy. 2) Beamorme
beamormer has several components with var
unctions, including the pulser and the amplif
Receiver/signal processor. Returning echoes r
the imaged tissue must be processed and flte
Processor/ scan converter. Te scan converterverts echocardiograph data into echocardiog
images that can be displayed, stored, or urthe
processed. 5) Display monitors. Tese are cath
ray tubes or liquid cr ystal display (LCD) screen
display the image. PRFpulse repetition requ
-
8/8/2019 AHD1601-01
2/24
\ Atlas of Echocardiography
Figure 1-3. Modern ultrasound equipment
phased array transducers, which electronical
generate and steer the ultrasound beams gen
ated in sequence. Each beam results in the g
tion o a scan line. A defned sequence o sca
lines is swept through the angle o view in or
to generate a composite called the scan sect
scan sector still rame rom a phased array tr
ducer typically consists o 96 to 512 scan line
Ultrasound beams or standard B-mode im
are generated in pulses. Te requency o thepulses is called the pulse repetition requency
tory nature o sound waves is caused by the compression and rareactio
particles. Te properties o ultrasound (as is the case in all waves) includ
requency (): the number o cycles occurring per second; 2) wavelength
the length o the waveone cycle; and 3) velocity (c): propagation spee
velocity. Tese properties are related by the equation c = (top panel)Amplitude represents the height or power o the wave. Te intensi
represents the power in watts divided by the area in square centimeters
Ultrasound waves are subject to attenuationthe decrease in amplitudrelative to the distance traveled as sound waves weaken as they travel
through a medium.
Ultrasound waves are primarily reected rom the boundaries betwee
tissues o dierent acoustic impedance (bottom panel). Tis type o reec
tion is reerred to as a specular reection. Although part o the ultrasoun
is reected as the echo, a percentage is transmitted (transmission) throug
the reector. Ultrasound waves can interact with tissue in the ollowing w
1) Reraction is a change in the direction o a non perpendicular incident
ultrasound beam without a change in requency. Tis results in the bendi
a portion o the incident ultrasound waves away rom the main beam axi
Scattering is a combination o reection and reraction o incident ultraso
waves within tissues, which diverts them in multiple directions. Interactio
tissues and tissue boundaries produces a pattern o echoes that are chara
teristic o the tissues imaged, which gives them their signature appearance
Hyperechoic structures produce greater amplitude o reections and appwhite. Hypoechoic tissues produce echoes with smaller amplitudes and a
as various shades o gray. Anechoic or echolucent structures appear echo
because they completely absorb all the incident ultrasound waves. 3) Att
ation reers to the loss o amplitude or beam intensity with distance trave
within the imaged tissue due to absorption and scattering. 4) Absorption
transmission o ultrasound waves with complete loss o acoustic energy a
with no reected echoes. A raction o the acoustic energy is lost as heat.
Figure 1-. Ultrasound waves are high-requency sound waves, which
have a general range o over a million cycles per second. In contrast, sound
that is audible to humans all within the 20 to 20,000 Hz range. Te oscilla-
-
8/8/2019 AHD1601-01
3/24
Basic Principles of Echocardiography
Figure 1-4. Te resolution o ultrasound images (ie, the ability o the ultra-
sound beam to detect and display anatomical details within the structures
imaged) is dependent on many actors, including the characteristics o the
transducer combined with the physical principles o ultrasound. Each ultra-
sound beam emitted has a defned slice thickness (left panel). Te components
o overall resolution include axial, lateral, elevational (slice thickness), temporal,
and contrast resolution. Axial resolution, which is also called depth, longitudi-
nal, linear, or range resolution, reers to the ability to resolve or detail echoes
that are reected rom two contiguous structures oriented along the axis o
the scan line. It is inuenced by transducer requency and wavelength (pulse
length) and is equivalent to one-hal spatial pulse length. Increased transducer
requency results in increased axial resolution, although this is a trade-o o
penetration, which decreases with increasing transducer requency. Lateralresolution reers to the ability to resolve or detail echoes rom two side-by-side
structures oriented perpendicularly to the scan line axis and is inuenced
the beam width and imaging depth. Spatial resolution reers to the ability
resolve detail in an ultrasound image (averaging 1 mm) and is dependent
both axial and lateral resolution. It is heavily inuenced by transducer req
(right panel). Slice thickness (elevational) resolution reers to the ability to
resolve details situated out o plane o the ultrasound beam or scan plan
emporal resolution reers to the ability to precisely capture still rames o
moving structures as they move rom instant to instant accuracy in time
averaging 50 m/sec. Te greater the pulse repetition requency (ie, the mo
pulses emitted rom the transducer per second), the greater the ability o
transducer to accurately capture a still rame o a moving structure. Con
resolution reers to the ability to discern dierential reectivity between d
ent tissues (eg, endocardial border) and the myocardium, and is inuencepre and post processing. (Far-right panel from Otto et al. [1]; with permiss
Figure 1-5. Harmonic imaging is an important development that has
greatly improved ultrasound image quality. Harmonic images rely on the act
that tissues may reverberate at requencies that are multiples o the original
ultrasound beam. Harmonic images are created rom returning echoes that
have requencies twice that o the undamental requency. Second harm
undergo less distortion than frst harmonics and result in increased sign
noise ratio. Image quality and endocardial defnition are considerably be
with harmonic imaging (right panel).
-
8/8/2019 AHD1601-01
4/24
4\ Atlas of Echocardiography
Figure 1-6. Contrast echocardiography has been an important advance-
ment in echocardiography. It has clinical use in the assessment o both
cardiac structure and unction. Encapsulated microspheres or microbubbles
measuring ~3 to 6 m in diameter can pass through capillaries. Microbubbles
contain a shell or gas comprised o one or a combination o air, lipid, poly-
mer, peruorocarbon, insoluble gas (eg, perutren lipid microsphere (Defn-
ity). Tese improve their echoreectivity. Microbubbles may have resonant
requencies ranging rom 2 to 10 MHz.
Encapsulation enables gas to stay within blood vessels, and to not be lost to
diusion. Microbubbles create increased echogenicity during interaction with
ultrasound waves. Left panel, At the low-pressure (rareaction) regions o
incident ultrasound wave, microbubbles expand (may expand 2- to 10-o
the crest o the ultrasound wavesthe regions o peak pressuresthe m
bubbles are compressed. Tis creates bubble oscillations, which produce n
echoes. Tese echoes contain harmonic requencies o the incident ultras
that improve their echoreectivity, and hence image analysis. Te dieren
acoustic impedance o microspheres is lower than that o blood, which le
increased linear backscattering and reectivity at the microsphere-blood
ace. Te microspheres produce echoes in response. Right panels above
below, Contrast harmonic imaging with generation o microbubble harm
this figure could be
removed if you
need space(covered in
contrast chapter)
-
8/8/2019 AHD1601-01
5/24
Basic Principles of Echocardiography
Causes o Ultrasound ArtiactsFaculty equipment
Instrument malunction
Intererence rom other equipment (eg, electrocautery during surgery);malunctioning transducer
Improper instrument settings
Too much time gain compensation (gain)
No harmonic imaging (undamental requency only)
Suboptimal imaging technique: inadequate transducer requency,
sector width, depth, dynamic range, and power outputImproper technique or patient characteristics
Sonographer inexperience
Suboptimal imaging due to patient movement, breathing, obesity,COPD, post-chest surgery
Suboptimal imaging technique: inadequate transducer requency,sector width, depth, dynamic range, and power output
Acoustic or sonographic artiacts
Attenuation artiacts: shadowing and enhancement
Propagation artiacts: Poor spatial resolution, section/slice thickness,speckling, reverberation, comet-tail or ring down, mirror image,range ambiguity, and slide lobe artiacts
Figure 1-7. All imaging modalities are subject to artiacts, which can h
adverse eects on the interpretation o studies. Artiacts are alse image
can be caused by aulty equipment, improper technique, or by acoustic
phenomenon or intererence. Understanding artiacts and their mecha-
nisms is important is important in echocardiography because they comm
occur. Teir recognition is crucial to proper interpretation because they
cause unnecessary alarm and unwarranted clinical studies. COPDchro
obstructive pulmonary disease.
Figure 1-8. Reverberation artiacts are caused by internal reverberations,
or back-and-orth echoes within the imaged tissue itsel, which result in
equipment miscalculation o the return trip time o the emitted pulses.
Reverberation artiacts occur at greater depths than the true image, but at
distances that are multiples o the true image. Let panel, Te ultrasound
is both reected rom and transmitted through a structure (ie, an area ocalcifcation) beore reecting o a second structure. Te reected echo is
re-reected in the direction o the original beam returning to the deeper
structure and is then reected back to the transducer. An artiact is gen
erated at a distance that is a multiple o the original reecting structur
Right panel, Te image resulting rom the inow conduit o an LV ass
device positioned in the LV apex. With reverberation artiacts, the arti
are equally spaced, located parallel to the sound beam axis, and are see
depths deeper than the true reector. Acoustic shadowing or dropoutacts appear as hypoechoic or anechoic regions below structures with
attenuation (eg, calcifed structures).
-
8/8/2019 AHD1601-01
6/24
6\ Atlas of Echocardiography
Figure 1-10. Upper left , Side lobes degrade lateral resolution. Artiactsappear side by side at the same depth. Grating lobe artiacts ollow a
similar principle in their maniestation, but they arise rom the structural
arrangement o multiple transducer elements within the transducer.
Upper right, his side-lobe artiact, arising rom the pericardium in this
parasternal long-axis view, gives the impression o a linear echodensity in
the let atrium (LA). Lower left, he hyperechoic linear echodensity seen
in this let atrial appendage (LAA) on transesophageal echocardiography
does not share the same echorelectivity pattern as thrombus or the tinate muscles, which would appear isoechoic with the LAA wall. Ne
does it exhibit the characteristics o the wararin ridge that appear a
linear shadowing within the LAA lumen, which moves synchronously
the old o tissue separating the LAA rom the let upper pulmonary
It is actually a side lobe artiact that is mapped into the LAA lumen.
right, Color low Doppler angiogram shows no evidence o a physi
presence within the LAA lumen.
Figure 1-9. Left panel, Comet tail or ring down artiacts (arrows)
on the parasternal long-axis view (PLAX). hese app ear as hyperechoic,
comet tailed, and linear in appearance and are located parallel to the
ultrasound beam axis. hey arise rom merger o closely spaced reverbera-
tions (middle panel). Mirror image artiacts are commonly seen whe
imaging the thoracic aorta on transesophageal echocardiography (
his is a long-axis view o the descending thoracic aorta long-axis vi
(right panel). Etransthoracic echocardiography.
-
8/8/2019 AHD1601-01
7/24
Basic Principles of Echocardiography
Figure 1-11. Left panel, he Doppler principle as applied to normal
blood low dynamics within the thoracic aorta. he Doppler requency
shit, sometimes reerred to as Doppler requency, is a shit in the re-quency o the returning echoes compared with the requency o emitted
ultrasound waves. he Doppler requency is itsel a wave with its own
requency. Echoes relected rom blood lowing away rom the transducer
are at lower requencies compared with those emitted rom the transduc-
er ( left panel, blue column). Echoes relected rom blood lowing towards
the transducer are at higher requencies compared with those emitted
rom the transducer (left panel, red column). Echoes relected rom blood
lowing perpendicular to the transducer show no changes in requency
compared with those emitted rom the transducer ( left panel, white
column). Right panel, Doppler assessment o blood low hemodynam
Normal laminar low exhibits a range o requencies that are detectabspectral Doppler, which is so named or its ability to exhibit the spect
o requencies. Note the narrow range or band o requencies in the (
panel, blue and red columns) and their relationship to the baseline, w
creates a window between the Doppler display and the baseline. Ri
panel, deep red column, During turbulent low, there is a wider ran
blood low velocities seen on the spectral Doppler display (spectral b
ening) with a illed-in window. Right panel, blue column, Flow aw
rom transducer. Right panel, red column, Flow towards transducer
This figure could
also be removed
-
8/8/2019 AHD1601-01
8/24
-
8/8/2019 AHD1601-01
9/24
Basic Principles of Echocardiography
Figure 1-14. Te Doppler equation is used to convert Doppler requency
shits into blood ow velocities. However, the accuracy o the Doppler
requency shit is directly inuenced by the beam-vessel alignment or the
Doppler angle. Te smaller the Doppler angle (), the more accurate themeasurement. Although the actual angle can be used in the equation to
correct Doppler beams that are not exactly parallel to the interrogated
(most ultrasound equipment has the capability to do this), the urther
angle is rom 0 degrees, the less accurate the measurement will be. Cs
soundblood
; Vvelocity; FRrequency o the received or returning echo
Frequency o the transmitted ultrasound pulses
Figure 1-15. In order to construct an ultrasound image (eg, a scan sec-
tor), the ultrasound pulse must make a round trip between the transducer
and the target. In pulsed Doppler studies, the transducer must listen or
the Doppler signal ollowing transmission. A new ultrasound pulse cannot
be transmitted beore the previous pulse has returned. he pulse rep
tion requency (PRF) is the number o pulses transmitted each secon
he urther the imaging depth, the slower the PRF must be to allow
the round trip. Ddepth.
-
8/8/2019 AHD1601-01
10/24
10\ Atlas of Echocardiography
Figure 1-16. In pulsed-wave Doppler (upper left panel) the ultrasound signal
is emitted repeatedly rom the transducer and the return signal is listened
or by the transducer. Sending out Doppler in pulses allows the equipment
to interrogate velocities at particular depths by gating, or only listening at a
certain time ater the pulse is emitted. By ocusing on a particular scan line, a
region o interest (sample volume) can be defned by both the depth using
time gating and the lateral location. Because pulsed Doppler is sent out and
received in pulses, the Doppler velocities are essentially being sampled at
the pulse repetition requency and are thereore subject to sampling issues,
which can limit the velocity o the blood ow that can be interrogated.All
spectral Doppler signals (upper right panel) show time on the X-a xis a
velocity on the Y-axis. When the pulsed-wave Doppler signal appears
low (Doppler window), it is indicative o laminar low where the ma
o red blood cells are traveling within a narrow range o velocities.
With continuous wave (CW) Doppler (lower left panel) the Doppler si
emitted continuously (similar to a continuous tone) and continuously list
or. For this reason, CW Doppler can be used to identiy peak velocities, b
cannot localize velocities to particular depths. Note the flled in appearan
o the CW spectral Doppler display in the ( lower right panel). Tis reects
wide spectrum o velocities detected within the much larger sample volu
-
8/8/2019 AHD1601-01
11/24
-
8/8/2019 AHD1601-01
12/24
1\ Atlas of Echocardiography
Figure 1-18. Left panel, Apical ive-chamber view in a patient with
dynamic LV outlow tract (LVO) obstruction showing accelerated low
on color low Doppler ( inset). Pulsed wave (PW) Doppler showed aliased
velocities with baseline set just above 2 m/s. Right panel, Nyquist limit
and aliasing; baseline adjustment. he same patient as in the (left pa
with baseline set at 6.0 m /s and switch to continuous wave Doppler
mode, the entire spectral display and peak velocities are now unamb
ously displayed.
Figure 1-19. Te law o conservation o mass states that the ow in one
part o the heart has to be equivalent to the ow in another part; ie, ow
rate in = ow rate out ( left panel). Doppler ultrasonography harnesses this
principle in order to calculate valve area (eg, in aortic stenosis). Blood ow
velocities increase at sites where blood vessel diameter narrows. Te dier-
ential ow velocities are detected by Doppler and used to calculate the valve
areas. Te continuity equation suggests that the area at a particular location
multiplied by either the velocity o ow or the time integral o owveloc-
ity time integral (VI) represents the area under the curve o the Doppler
ow proflewill be similar in two locations assuming no shunt is prese
For example, in order to calculate the valve area in a patient with suspec
aortic stenosis (right panel), the cross sectional area (CSA) at the level o
LV outow tract is calculated (in actuality, the diameter is measured and
area is calculated), and multiplied with the velocity or VI obtained at t
same location using pulsed wave (PW) Doppler. Tis product should eq
the product o the velocity or VI across the valve itsel, which is estima
using continuous wave (CW) Doppler to fnd the highest velocity and t
aortic valve area (AVA), which is then calculated. LVOLV outow tra
-
8/8/2019 AHD1601-01
13/24
Basic Principles of Echocardiography
Figure 1-0. Te Bernoulli principle serves as the basis o calculating pressure gradients (driving pres-
sure) across a narrowed orifce. Te Bernoulli principle broadly states that as the velocity o a moving uidincreases, the pressure within the uid decreases. It assumes the conservation o energy, though in reality
Figure 1-1. One o the advances in cardiac ultrasound has been the development o color low
Doppler (CFD) imaging. It has dramatically improved the assessment o cardiovascular hemodynam-
ics, and complements the cross-sectional and spectral Doppler examination. CFD is a pulse wave
(PW)based modality that, unlike PW, uses multiple gates ( ie , multiple sample volumes) simultane-
ously instead o a single gate and provides spatial display o real-time color-coded velocities super-
imposed upon real-time two-dimensional (2D) imaging. CFD thus provides inormation on low
velocity and direction. CFD imaging is subject to the same physical principles as PW Doppler and is
some energy is lost through heat, turbulent r
and vortex ormation. Doppler-derived veloci
are inserted into the modifed or simplifed ve
o the Bernoulli equation in order to calculate
sure gradients across cardiac chambers and va
As a blood vessel or orifce narrows or becom
stenosed, the velocity o blood downstream
the stenosis must increase. Blood proximal to
stenosis (location 1) will thereore accelerate a
ows downstream rom the stenosis (location(law o conservation o energy).
Te modifed Bernoulli equation requires
knowing velocities both proximal and distal
stenosis. In the majority o cases in physiolog
conditions, we can ignore the proximal veloc
Tis is not true, however, when velocities pro
to the stenosis are elevated. For example, i w
were to use the Bernoulli equation in order t
assess the gradient across a stenotic aortic va
physiologic velocities proximal to the valve w
be relatively low compared with the velocitie
tal to the valve. However, i that same patien
outow tract obstruction due to hypertroph
cardiomyopathy or a sub aortic membrane, t
proximal velocities would have to be consideLVOLV outow tract.
thus subject to aliasing when velocities are
enough that the Doppler shit is greater tha
hal the pulse repetition requency.
Left panel, Schema o apical our-chamb
(A4C) view showing CFD sector scan superim
posed upon real-time two-dimensional gray
image. Te active color scan sector employs
ple gates (trains or ensemble) that are color-
velocities. Right panel, A4C view showing c
Doppler superimposed on B-mode two-dim
sional imaging. Flow direction during early syreveals blue ow along the LV outow tract.
ow indicating ow momentum rom let at
into LV is also visualized.
CFD provides an intuitive assessment o c
diovascular hemodynamics because it permi
angiographic view o the ow, and is there
more intuitive than spectral Doppler. It allow
rapid qualitative assessment o abnormal o
patterns. It is an important guide to PW and
continuous wave Doppler position and align
LAlet atrium; RVright atrium.
-
8/8/2019 AHD1601-01
14/24
14\ Atlas of Echocardiography
Figure 1-3. Te mosaic color pattern shown in this image reects hi
velocity turbulent ow (arrow) in a patient with severe mitral regurgita
Note the mosaic pattern o bright hues indicative o aliasing due to tur
lence and high-velocity ow.
Comparison o Doppler ModesContinuous wave Doppler Pulsed wave Doppler Color fow Doppler
Sample volume Large (with spectral broadening) Small (with spectral window due tonarrow range o velocities); samplevolume adjustable (best 25 mm)
Large with multigating; adjustable actcolor window in color scan sector
Detection o peak/maximal velocities
Detects highest velocities along scanline; peak velocity measurements
Peak velocity measurements, but aliasingoccurs at high velocities
Mean velocity measurements; aliasingoccurs as it is a PW-based modality
Aliasing No aliasing Aliasing (Nyquist limit) A PW Doppler-based technique; alias
Sensitivity Best sensitivity Good sensitivity Good sensitivity(jet size very sensitive to gain setting
Depth resolution/range ambiquity
Range ambiguity Range resolution; detects location o fow Range resolution;real time 2D anatomy o fow
Spectrum analyzed Wide (spectral broadening) Narrow; spectral broadening seen withturbulent fow and large sample volumes
Wide color-encoded Doppler shits oconventional *BART scale and color
hues, saturation, and luminance*Blue away, red towards.
Figure 1-4. Comparison: continuous wave, pulsed wave, and color ow Doppler.
Figure 1-. Images obtained by color ow Doppler (CFD) rames must
be interpreted with reerence to the velocity scales displayed on the image.
Tese reerence maps or scales provide inormation on average ow velocities
based on three characteristics o color: 1) hue, 2) luminance, and 3) satura-
tion. wo major modes o display are velocity mode and variance mode.Left panel, Conventional velocity mode: ow towards transducer is
coded (by convention) in shades o red. Flow away rom transducer is color-
coded in shades o blue (BAR: blue away, red towards). Black color on the
sale indicates no ow. Lighter shades o BAR indicate aster ow.
Right panel, Variance mode maps provide inormation in addition to
velocity and direction. Tey can help distinguish laminar rom turbulent ow.
In laminar ow, the average velocities are not much dierent rom the peak
velocities. Tey are close together (ie, they exhibit minimal variance). When
blood ow is turbulent, there are much greater dierences between the
average and peak velocities, namely greater variance. Deeper shades indicate
slower laminar ow, while lighter shades may represent aster, turbulent ow.
-
8/8/2019 AHD1601-01
15/24
Basic Principles of Echocardiography
Te Normal Echocardiographic Examination and omographic Anatom
Figure 1-5. ransthoracic two-dimensional echocar
raphy: imaging planes, transducer positions, and standviews. A2Capical two-chamber view; A3Capical th
chamber view; A4Capical our-chamber view; A5C
cal fve-chamber view; PLAXparasternal long-axis vie
PSAXparasternal short-axis view; PSAX-AVparaste
short-axis view-aortic valve level; PSAX-MVparastern
short-axis view-mitral valve level; PSAX-PMparastern
short-axis view papillary muscle level; SC-4Csubcosta
view (our-chamber).
Would be better for the figure to occupy the majority of the page and the text below it - i.
e., stretch figure down and out - with caption below
-
8/8/2019 AHD1601-01
16/24
-
8/8/2019 AHD1601-01
17/24
Basic Principles of Echocardiography
CFigure 1-7. (Continued)
Figure 1-8. Echocardiographic parasternal
inow scan plane and corresponding anatom
IVCinerior vena cava; LLPVlet lower pul
monary vein; LUPVlet upper pulmonary ve
RAright atrium; RAAright atrial appenda
RVORV outow tract; SVCsuperior ven
-
8/8/2019 AHD1601-01
18/24
18\ Atlas of Echocardiography
Figure 1-30. Echocardiographic short-axis s
planes and corresponding anatomy.
Figure 1-9. wo-dimensional long-axis view showing structures visua
on the RV inow view. RAright atrium.
-
8/8/2019 AHD1601-01
19/24
Basic Principles of Echocardiography
Figure 1-31. wo-dimensional parasternal short-view showing the
structures visualized at the level o the aortic valve. RAright atrium;
RVORV outow tract.
Figure 1-3. Views o the high parasternal
short-axis scan plane at the level o the pul-
monary arteries and corresponding anatom
Aoaorta; DAdescending thoracic aort
PApulmonary artery.
-
8/8/2019 AHD1601-01
20/24
0\ Atlas of Echocardiography
Figure 1-34. Anatomical structures correspo
to the parasternal short-axis (PSAX) scan plan
the level mitral valve (MV) and papillary musc
Figure 1-33. wo-dimensional echocardiographic rame showing the s
tures visualized on high parasternal short-axis view at the level o the ma
pulmonary artery (PA) and biurcation into the right and let pulmonar
arteries. Aoaorta; RAright atrium; RVORV outow tract.
are really
sting space heree figure above
34) could be
de bigger - theis minimal
-
8/8/2019 AHD1601-01
21/24
Basic Principles of Echocardiography
Figure 1-36. Echocardiographic our-chamb
scan planes and corresponding anatomy. RA
atrium; SVCsuperior vena cavity.
Figure 1-35. wo-dimensional short-axis rames showing structures visual-
ized on the parasternal short-axis (PSAX) views at the mitral valve level (left
panel), mid ventricular or papillary muscle level (right panel), and apic
level (right panel).
-
8/8/2019 AHD1601-01
22/24
\ Atlas of Echocardiography
Figure 1-39. wo-dimensional rame showing structures
alized on the suprasternal notch view.
Figure 1-38. Echocardiographic suprastern
notch scan plane and corresponding anatom
LAlet atrium; PApulmonary artery; R A
right atrium; SVCsuperior vena cavity.
Figure 1-37. wo-dimensional our-chamber rames showing the structures
visualized on the apical our-chamber (A4C) ( left panel), apical fve-chamber
(A5C), (middle panel), and subcostal our-chamber views (right panel).
LAlet atrium; RAright atrium.
-
8/8/2019 AHD1601-01
23/24
Basic Principles of Echocardiography
Figure 1-40. Panoramic schema o coronary
artery territories and LV segments using nom
ture recommended by the American Heart A
ation Writing Group on Myocardial Segment
and Registration or Cardiac Imaging [2]. IVC
rior vena cavity; LAlet atrium; laalet atr
appendage; LADlet anterior descending aPApulmonary artery; LAOlet anterior o
LCxlet circumex; RAOright anterior ob
RCA/PDAright coronary/posterior descend
SVCsuperior vena cavity.
1. Otto C, ed: Principles o echocardiography imaging acquisition and Doppler
analysis. In extbook of Clinical Echocardiography, edn 3. Philadelphia:
WB Saunders; 2004.
2. Cerqueira MD, Weissman NJ, Dilsizian V, et al.: Standardized myocardial segmen
and nomenclature or tomographic imaging o the heart: a statement or health
proessionals rom the Cardiac Imaging Committee o the Council on Clinical C
ogy o the American Heart Association. Circulation 2002, 105:539542.
References
Recommended Reading
Bushberg J, Siebert JA, Leidholdt EM Jr, Boone JM: Te Essential Physics of Medical
Imaging, edn 2. Philadelphia: Lippincott Williams & Wilkins; 2002:469553.
Cape EG, Yoganathan AP: Principles and instrumentation or Doppler. In Marcus
Cardiac Imaging, edn 2. Philadelphia: WB Saunders; 1996:273291.
Feigenbaum H: Echocardiography, edn 4. Malvern, PA: Lea and Febiger; 1986.
Gardin JM, Adams DB, Douglas PS, et al.: Recommendations or a standardized repo
adult transthoracic echocardiography: a report rom the American Society o E
diographys Nomenclature and Standards Committee and ask Force or a Stan
ized Echocardiography Report.J Am Soc Echocardiogr 2002, 15:275290.
-
8/8/2019 AHD1601-01
24/24
Geiser EA: Echocardiography: physics and instrumentation. In Marcus Cardiac Imaging,
edn. 2. Edited by Skorton DJ, Schelbert HR, Wol GL, Brundage BH.
Philadelphia: WB Saunders; 1996:273291.
Hatle L, Angelsen B: Doppler Ultrasound in Cardiology: Physical Principles and
Clinical Applications, edn. 2. Philadelphia: Lea & Febiger; 1985.
Henry WL, DeMaria A, Gramiak R, et al.: Report o the American Society o Echo-
cardiography Committee on nomenclature and standards in two-dimensional
echocardiography. Circulation 1980, 62:212217.
Hung J, Lang R, Flachskamp F, et al.: 3D echocardiography: a review o the current
status and uture directions.J Am Soc Echocardiogr 2007, 20:213233.
Kremkau FW: Diagnostic Ultrasound: Principles and Instruments, edn 6. Philadelphia:
WB Saunders; 2002.
McAlpine WA: Heart and Coronary Arteries: an Anatomical Atlas for Clinical
Diagnosis, Radiological Investigation, and Surgical reatment . Berlin, Heidelberg,
New York: Springer-Verlag; 1975.
Medical diagnostic ultrasound instrumentation and clinical interpretation: repo
the ultrasonography task orce.JAMA 1991, 265:11551159.
ORourke MF, Nichols WW: McDonalds Blood Flow in Arteries, edn 5. New York
Oxord University Press; 2005.
Quinones MA, Douglas PS, Foster E, et al.: ACC/AHA clinical competence statem
on echocardiography: a report o the American College o Cardiology/Ame
Heart Association/American College o Physicians - American Society o In
Medicine ask Force on Clinical Competence. J Am Coll Cardiol 2003, 41:687
Quinones MA, Otto CM, Stoddard M, et al.: Recommendations or quantifcatio
Doppler echocardiography: a report rom the Doppler Quantifcation ask
o the Nomenclature and Standards Committee o the American Society ocardiography. J Am Soc Echocardiogr 2002, 15:167184.
Seghal CM: Principles o ultrasonic imaging and Doppler ultrasound. In extboo
Echocardiography and Doppler in Adults and Children . Edited by St. John Su
MG, Oldershaw PJ, Kotler MN. Oxord, UK: Blackwell Science; 1996:330.