ahd1601-01

Upload: scott-solomon

Post on 10-Apr-2018

231 views

Category:

Documents


4 download

TRANSCRIPT

  • 8/8/2019 AHD1601-01

    1/24

    Echocardiography relies on the use o ultrasoundhigh-requency sound wavestointerrogate bodily tissues. Ultrasound waves penetrate through and are reected by tissuesbeore returning to the transducer, where they are then converted into images. Modern ultra-

    sound equipment uses three principal modalities. Tey include M-mode imaging, B-mode (or

    two-dimensional imaging), and Doppler imaging. Tis chapter will introduce the basic physical

    principles o ultrasound and tomographic anatomy in cardiac imaging with ultrasound.

    Basic Principles of

    Echocardiography andTomographic AnatomyBernard E. Bulwer, Jose Rivero, and Scott D. Solomon

    Figure 1-1. Ultrasound waves used in medical ultrasound imaging are generated by the piezoelectric

    elements housed within the transducer (left panel). In order to generate an ultrasound image, the ultra-

    sound pulse must make a round trip between the piezoelectric elements (housed inside the transducer),the scan line, and the region o interest (imaged structure). Te ultrasound waves that are emitted

    travel along scan lines and interact with tissue structures. Ultrasound waves or echoes are dierentially

    reected rom cardiac structures and returned to transducer where they are then converted to electrical

    impulses and subsequently converted to images by system components (eg, signal processor and scan

    converters) housed within the ultrasound equipment. wo-dimensional images representing cardiac

    cross-sectional anatomy are composed o se

    scan lines (see Fig. 1-3).

    Te principal method used in order to prod

    ultrasound images is reerred to as B-mode or b

    ness mode imaging (right panel). Echocardiogr

    images are processed and displayed as pixels, orpicture elements, o varied shades o gray (black

    white). Each pixel has its own anatomical addre

    and echoreectivity and is mapped accordingly

    Te major components o a modern ultra

    scanner include the ollowing: 1) ransducer. T

    transducer is the interace between the patien

    and the beamormer. Electrical energy is conv

    here into ultrasound and returning ultrasound

    (echoes) into electrical energy. 2) Beamorme

    beamormer has several components with var

    unctions, including the pulser and the amplif

    Receiver/signal processor. Returning echoes r

    the imaged tissue must be processed and flte

    Processor/ scan converter. Te scan converterverts echocardiograph data into echocardiog

    images that can be displayed, stored, or urthe

    processed. 5) Display monitors. Tese are cath

    ray tubes or liquid cr ystal display (LCD) screen

    display the image. PRFpulse repetition requ

  • 8/8/2019 AHD1601-01

    2/24

    \ Atlas of Echocardiography

    Figure 1-3. Modern ultrasound equipment

    phased array transducers, which electronical

    generate and steer the ultrasound beams gen

    ated in sequence. Each beam results in the g

    tion o a scan line. A defned sequence o sca

    lines is swept through the angle o view in or

    to generate a composite called the scan sect

    scan sector still rame rom a phased array tr

    ducer typically consists o 96 to 512 scan line

    Ultrasound beams or standard B-mode im

    are generated in pulses. Te requency o thepulses is called the pulse repetition requency

    tory nature o sound waves is caused by the compression and rareactio

    particles. Te properties o ultrasound (as is the case in all waves) includ

    requency (): the number o cycles occurring per second; 2) wavelength

    the length o the waveone cycle; and 3) velocity (c): propagation spee

    velocity. Tese properties are related by the equation c = (top panel)Amplitude represents the height or power o the wave. Te intensi

    represents the power in watts divided by the area in square centimeters

    Ultrasound waves are subject to attenuationthe decrease in amplitudrelative to the distance traveled as sound waves weaken as they travel

    through a medium.

    Ultrasound waves are primarily reected rom the boundaries betwee

    tissues o dierent acoustic impedance (bottom panel). Tis type o reec

    tion is reerred to as a specular reection. Although part o the ultrasoun

    is reected as the echo, a percentage is transmitted (transmission) throug

    the reector. Ultrasound waves can interact with tissue in the ollowing w

    1) Reraction is a change in the direction o a non perpendicular incident

    ultrasound beam without a change in requency. Tis results in the bendi

    a portion o the incident ultrasound waves away rom the main beam axi

    Scattering is a combination o reection and reraction o incident ultraso

    waves within tissues, which diverts them in multiple directions. Interactio

    tissues and tissue boundaries produces a pattern o echoes that are chara

    teristic o the tissues imaged, which gives them their signature appearance

    Hyperechoic structures produce greater amplitude o reections and appwhite. Hypoechoic tissues produce echoes with smaller amplitudes and a

    as various shades o gray. Anechoic or echolucent structures appear echo

    because they completely absorb all the incident ultrasound waves. 3) Att

    ation reers to the loss o amplitude or beam intensity with distance trave

    within the imaged tissue due to absorption and scattering. 4) Absorption

    transmission o ultrasound waves with complete loss o acoustic energy a

    with no reected echoes. A raction o the acoustic energy is lost as heat.

    Figure 1-. Ultrasound waves are high-requency sound waves, which

    have a general range o over a million cycles per second. In contrast, sound

    that is audible to humans all within the 20 to 20,000 Hz range. Te oscilla-

  • 8/8/2019 AHD1601-01

    3/24

    Basic Principles of Echocardiography

    Figure 1-4. Te resolution o ultrasound images (ie, the ability o the ultra-

    sound beam to detect and display anatomical details within the structures

    imaged) is dependent on many actors, including the characteristics o the

    transducer combined with the physical principles o ultrasound. Each ultra-

    sound beam emitted has a defned slice thickness (left panel). Te components

    o overall resolution include axial, lateral, elevational (slice thickness), temporal,

    and contrast resolution. Axial resolution, which is also called depth, longitudi-

    nal, linear, or range resolution, reers to the ability to resolve or detail echoes

    that are reected rom two contiguous structures oriented along the axis o

    the scan line. It is inuenced by transducer requency and wavelength (pulse

    length) and is equivalent to one-hal spatial pulse length. Increased transducer

    requency results in increased axial resolution, although this is a trade-o o

    penetration, which decreases with increasing transducer requency. Lateralresolution reers to the ability to resolve or detail echoes rom two side-by-side

    structures oriented perpendicularly to the scan line axis and is inuenced

    the beam width and imaging depth. Spatial resolution reers to the ability

    resolve detail in an ultrasound image (averaging 1 mm) and is dependent

    both axial and lateral resolution. It is heavily inuenced by transducer req

    (right panel). Slice thickness (elevational) resolution reers to the ability to

    resolve details situated out o plane o the ultrasound beam or scan plan

    emporal resolution reers to the ability to precisely capture still rames o

    moving structures as they move rom instant to instant accuracy in time

    averaging 50 m/sec. Te greater the pulse repetition requency (ie, the mo

    pulses emitted rom the transducer per second), the greater the ability o

    transducer to accurately capture a still rame o a moving structure. Con

    resolution reers to the ability to discern dierential reectivity between d

    ent tissues (eg, endocardial border) and the myocardium, and is inuencepre and post processing. (Far-right panel from Otto et al. [1]; with permiss

    Figure 1-5. Harmonic imaging is an important development that has

    greatly improved ultrasound image quality. Harmonic images rely on the act

    that tissues may reverberate at requencies that are multiples o the original

    ultrasound beam. Harmonic images are created rom returning echoes that

    have requencies twice that o the undamental requency. Second harm

    undergo less distortion than frst harmonics and result in increased sign

    noise ratio. Image quality and endocardial defnition are considerably be

    with harmonic imaging (right panel).

  • 8/8/2019 AHD1601-01

    4/24

    4\ Atlas of Echocardiography

    Figure 1-6. Contrast echocardiography has been an important advance-

    ment in echocardiography. It has clinical use in the assessment o both

    cardiac structure and unction. Encapsulated microspheres or microbubbles

    measuring ~3 to 6 m in diameter can pass through capillaries. Microbubbles

    contain a shell or gas comprised o one or a combination o air, lipid, poly-

    mer, peruorocarbon, insoluble gas (eg, perutren lipid microsphere (Defn-

    ity). Tese improve their echoreectivity. Microbubbles may have resonant

    requencies ranging rom 2 to 10 MHz.

    Encapsulation enables gas to stay within blood vessels, and to not be lost to

    diusion. Microbubbles create increased echogenicity during interaction with

    ultrasound waves. Left panel, At the low-pressure (rareaction) regions o

    incident ultrasound wave, microbubbles expand (may expand 2- to 10-o

    the crest o the ultrasound wavesthe regions o peak pressuresthe m

    bubbles are compressed. Tis creates bubble oscillations, which produce n

    echoes. Tese echoes contain harmonic requencies o the incident ultras

    that improve their echoreectivity, and hence image analysis. Te dieren

    acoustic impedance o microspheres is lower than that o blood, which le

    increased linear backscattering and reectivity at the microsphere-blood

    ace. Te microspheres produce echoes in response. Right panels above

    below, Contrast harmonic imaging with generation o microbubble harm

    this figure could be

    removed if you

    need space(covered in

    contrast chapter)

  • 8/8/2019 AHD1601-01

    5/24

    Basic Principles of Echocardiography

    Causes o Ultrasound ArtiactsFaculty equipment

    Instrument malunction

    Intererence rom other equipment (eg, electrocautery during surgery);malunctioning transducer

    Improper instrument settings

    Too much time gain compensation (gain)

    No harmonic imaging (undamental requency only)

    Suboptimal imaging technique: inadequate transducer requency,

    sector width, depth, dynamic range, and power outputImproper technique or patient characteristics

    Sonographer inexperience

    Suboptimal imaging due to patient movement, breathing, obesity,COPD, post-chest surgery

    Suboptimal imaging technique: inadequate transducer requency,sector width, depth, dynamic range, and power output

    Acoustic or sonographic artiacts

    Attenuation artiacts: shadowing and enhancement

    Propagation artiacts: Poor spatial resolution, section/slice thickness,speckling, reverberation, comet-tail or ring down, mirror image,range ambiguity, and slide lobe artiacts

    Figure 1-7. All imaging modalities are subject to artiacts, which can h

    adverse eects on the interpretation o studies. Artiacts are alse image

    can be caused by aulty equipment, improper technique, or by acoustic

    phenomenon or intererence. Understanding artiacts and their mecha-

    nisms is important is important in echocardiography because they comm

    occur. Teir recognition is crucial to proper interpretation because they

    cause unnecessary alarm and unwarranted clinical studies. COPDchro

    obstructive pulmonary disease.

    Figure 1-8. Reverberation artiacts are caused by internal reverberations,

    or back-and-orth echoes within the imaged tissue itsel, which result in

    equipment miscalculation o the return trip time o the emitted pulses.

    Reverberation artiacts occur at greater depths than the true image, but at

    distances that are multiples o the true image. Let panel, Te ultrasound

    is both reected rom and transmitted through a structure (ie, an area ocalcifcation) beore reecting o a second structure. Te reected echo is

    re-reected in the direction o the original beam returning to the deeper

    structure and is then reected back to the transducer. An artiact is gen

    erated at a distance that is a multiple o the original reecting structur

    Right panel, Te image resulting rom the inow conduit o an LV ass

    device positioned in the LV apex. With reverberation artiacts, the arti

    are equally spaced, located parallel to the sound beam axis, and are see

    depths deeper than the true reector. Acoustic shadowing or dropoutacts appear as hypoechoic or anechoic regions below structures with

    attenuation (eg, calcifed structures).

  • 8/8/2019 AHD1601-01

    6/24

    6\ Atlas of Echocardiography

    Figure 1-10. Upper left , Side lobes degrade lateral resolution. Artiactsappear side by side at the same depth. Grating lobe artiacts ollow a

    similar principle in their maniestation, but they arise rom the structural

    arrangement o multiple transducer elements within the transducer.

    Upper right, his side-lobe artiact, arising rom the pericardium in this

    parasternal long-axis view, gives the impression o a linear echodensity in

    the let atrium (LA). Lower left, he hyperechoic linear echodensity seen

    in this let atrial appendage (LAA) on transesophageal echocardiography

    does not share the same echorelectivity pattern as thrombus or the tinate muscles, which would appear isoechoic with the LAA wall. Ne

    does it exhibit the characteristics o the wararin ridge that appear a

    linear shadowing within the LAA lumen, which moves synchronously

    the old o tissue separating the LAA rom the let upper pulmonary

    It is actually a side lobe artiact that is mapped into the LAA lumen.

    right, Color low Doppler angiogram shows no evidence o a physi

    presence within the LAA lumen.

    Figure 1-9. Left panel, Comet tail or ring down artiacts (arrows)

    on the parasternal long-axis view (PLAX). hese app ear as hyperechoic,

    comet tailed, and linear in appearance and are located parallel to the

    ultrasound beam axis. hey arise rom merger o closely spaced reverbera-

    tions (middle panel). Mirror image artiacts are commonly seen whe

    imaging the thoracic aorta on transesophageal echocardiography (

    his is a long-axis view o the descending thoracic aorta long-axis vi

    (right panel). Etransthoracic echocardiography.

  • 8/8/2019 AHD1601-01

    7/24

    Basic Principles of Echocardiography

    Figure 1-11. Left panel, he Doppler principle as applied to normal

    blood low dynamics within the thoracic aorta. he Doppler requency

    shit, sometimes reerred to as Doppler requency, is a shit in the re-quency o the returning echoes compared with the requency o emitted

    ultrasound waves. he Doppler requency is itsel a wave with its own

    requency. Echoes relected rom blood lowing away rom the transducer

    are at lower requencies compared with those emitted rom the transduc-

    er ( left panel, blue column). Echoes relected rom blood lowing towards

    the transducer are at higher requencies compared with those emitted

    rom the transducer (left panel, red column). Echoes relected rom blood

    lowing perpendicular to the transducer show no changes in requency

    compared with those emitted rom the transducer ( left panel, white

    column). Right panel, Doppler assessment o blood low hemodynam

    Normal laminar low exhibits a range o requencies that are detectabspectral Doppler, which is so named or its ability to exhibit the spect

    o requencies. Note the narrow range or band o requencies in the (

    panel, blue and red columns) and their relationship to the baseline, w

    creates a window between the Doppler display and the baseline. Ri

    panel, deep red column, During turbulent low, there is a wider ran

    blood low velocities seen on the spectral Doppler display (spectral b

    ening) with a illed-in window. Right panel, blue column, Flow aw

    rom transducer. Right panel, red column, Flow towards transducer

    This figure could

    also be removed

  • 8/8/2019 AHD1601-01

    8/24

  • 8/8/2019 AHD1601-01

    9/24

    Basic Principles of Echocardiography

    Figure 1-14. Te Doppler equation is used to convert Doppler requency

    shits into blood ow velocities. However, the accuracy o the Doppler

    requency shit is directly inuenced by the beam-vessel alignment or the

    Doppler angle. Te smaller the Doppler angle (), the more accurate themeasurement. Although the actual angle can be used in the equation to

    correct Doppler beams that are not exactly parallel to the interrogated

    (most ultrasound equipment has the capability to do this), the urther

    angle is rom 0 degrees, the less accurate the measurement will be. Cs

    soundblood

    ; Vvelocity; FRrequency o the received or returning echo

    Frequency o the transmitted ultrasound pulses

    Figure 1-15. In order to construct an ultrasound image (eg, a scan sec-

    tor), the ultrasound pulse must make a round trip between the transducer

    and the target. In pulsed Doppler studies, the transducer must listen or

    the Doppler signal ollowing transmission. A new ultrasound pulse cannot

    be transmitted beore the previous pulse has returned. he pulse rep

    tion requency (PRF) is the number o pulses transmitted each secon

    he urther the imaging depth, the slower the PRF must be to allow

    the round trip. Ddepth.

  • 8/8/2019 AHD1601-01

    10/24

    10\ Atlas of Echocardiography

    Figure 1-16. In pulsed-wave Doppler (upper left panel) the ultrasound signal

    is emitted repeatedly rom the transducer and the return signal is listened

    or by the transducer. Sending out Doppler in pulses allows the equipment

    to interrogate velocities at particular depths by gating, or only listening at a

    certain time ater the pulse is emitted. By ocusing on a particular scan line, a

    region o interest (sample volume) can be defned by both the depth using

    time gating and the lateral location. Because pulsed Doppler is sent out and

    received in pulses, the Doppler velocities are essentially being sampled at

    the pulse repetition requency and are thereore subject to sampling issues,

    which can limit the velocity o the blood ow that can be interrogated.All

    spectral Doppler signals (upper right panel) show time on the X-a xis a

    velocity on the Y-axis. When the pulsed-wave Doppler signal appears

    low (Doppler window), it is indicative o laminar low where the ma

    o red blood cells are traveling within a narrow range o velocities.

    With continuous wave (CW) Doppler (lower left panel) the Doppler si

    emitted continuously (similar to a continuous tone) and continuously list

    or. For this reason, CW Doppler can be used to identiy peak velocities, b

    cannot localize velocities to particular depths. Note the flled in appearan

    o the CW spectral Doppler display in the ( lower right panel). Tis reects

    wide spectrum o velocities detected within the much larger sample volu

  • 8/8/2019 AHD1601-01

    11/24

  • 8/8/2019 AHD1601-01

    12/24

    1\ Atlas of Echocardiography

    Figure 1-18. Left panel, Apical ive-chamber view in a patient with

    dynamic LV outlow tract (LVO) obstruction showing accelerated low

    on color low Doppler ( inset). Pulsed wave (PW) Doppler showed aliased

    velocities with baseline set just above 2 m/s. Right panel, Nyquist limit

    and aliasing; baseline adjustment. he same patient as in the (left pa

    with baseline set at 6.0 m /s and switch to continuous wave Doppler

    mode, the entire spectral display and peak velocities are now unamb

    ously displayed.

    Figure 1-19. Te law o conservation o mass states that the ow in one

    part o the heart has to be equivalent to the ow in another part; ie, ow

    rate in = ow rate out ( left panel). Doppler ultrasonography harnesses this

    principle in order to calculate valve area (eg, in aortic stenosis). Blood ow

    velocities increase at sites where blood vessel diameter narrows. Te dier-

    ential ow velocities are detected by Doppler and used to calculate the valve

    areas. Te continuity equation suggests that the area at a particular location

    multiplied by either the velocity o ow or the time integral o owveloc-

    ity time integral (VI) represents the area under the curve o the Doppler

    ow proflewill be similar in two locations assuming no shunt is prese

    For example, in order to calculate the valve area in a patient with suspec

    aortic stenosis (right panel), the cross sectional area (CSA) at the level o

    LV outow tract is calculated (in actuality, the diameter is measured and

    area is calculated), and multiplied with the velocity or VI obtained at t

    same location using pulsed wave (PW) Doppler. Tis product should eq

    the product o the velocity or VI across the valve itsel, which is estima

    using continuous wave (CW) Doppler to fnd the highest velocity and t

    aortic valve area (AVA), which is then calculated. LVOLV outow tra

  • 8/8/2019 AHD1601-01

    13/24

    Basic Principles of Echocardiography

    Figure 1-0. Te Bernoulli principle serves as the basis o calculating pressure gradients (driving pres-

    sure) across a narrowed orifce. Te Bernoulli principle broadly states that as the velocity o a moving uidincreases, the pressure within the uid decreases. It assumes the conservation o energy, though in reality

    Figure 1-1. One o the advances in cardiac ultrasound has been the development o color low

    Doppler (CFD) imaging. It has dramatically improved the assessment o cardiovascular hemodynam-

    ics, and complements the cross-sectional and spectral Doppler examination. CFD is a pulse wave

    (PW)based modality that, unlike PW, uses multiple gates ( ie , multiple sample volumes) simultane-

    ously instead o a single gate and provides spatial display o real-time color-coded velocities super-

    imposed upon real-time two-dimensional (2D) imaging. CFD thus provides inormation on low

    velocity and direction. CFD imaging is subject to the same physical principles as PW Doppler and is

    some energy is lost through heat, turbulent r

    and vortex ormation. Doppler-derived veloci

    are inserted into the modifed or simplifed ve

    o the Bernoulli equation in order to calculate

    sure gradients across cardiac chambers and va

    As a blood vessel or orifce narrows or becom

    stenosed, the velocity o blood downstream

    the stenosis must increase. Blood proximal to

    stenosis (location 1) will thereore accelerate a

    ows downstream rom the stenosis (location(law o conservation o energy).

    Te modifed Bernoulli equation requires

    knowing velocities both proximal and distal

    stenosis. In the majority o cases in physiolog

    conditions, we can ignore the proximal veloc

    Tis is not true, however, when velocities pro

    to the stenosis are elevated. For example, i w

    were to use the Bernoulli equation in order t

    assess the gradient across a stenotic aortic va

    physiologic velocities proximal to the valve w

    be relatively low compared with the velocitie

    tal to the valve. However, i that same patien

    outow tract obstruction due to hypertroph

    cardiomyopathy or a sub aortic membrane, t

    proximal velocities would have to be consideLVOLV outow tract.

    thus subject to aliasing when velocities are

    enough that the Doppler shit is greater tha

    hal the pulse repetition requency.

    Left panel, Schema o apical our-chamb

    (A4C) view showing CFD sector scan superim

    posed upon real-time two-dimensional gray

    image. Te active color scan sector employs

    ple gates (trains or ensemble) that are color-

    velocities. Right panel, A4C view showing c

    Doppler superimposed on B-mode two-dim

    sional imaging. Flow direction during early syreveals blue ow along the LV outow tract.

    ow indicating ow momentum rom let at

    into LV is also visualized.

    CFD provides an intuitive assessment o c

    diovascular hemodynamics because it permi

    angiographic view o the ow, and is there

    more intuitive than spectral Doppler. It allow

    rapid qualitative assessment o abnormal o

    patterns. It is an important guide to PW and

    continuous wave Doppler position and align

    LAlet atrium; RVright atrium.

  • 8/8/2019 AHD1601-01

    14/24

    14\ Atlas of Echocardiography

    Figure 1-3. Te mosaic color pattern shown in this image reects hi

    velocity turbulent ow (arrow) in a patient with severe mitral regurgita

    Note the mosaic pattern o bright hues indicative o aliasing due to tur

    lence and high-velocity ow.

    Comparison o Doppler ModesContinuous wave Doppler Pulsed wave Doppler Color fow Doppler

    Sample volume Large (with spectral broadening) Small (with spectral window due tonarrow range o velocities); samplevolume adjustable (best 25 mm)

    Large with multigating; adjustable actcolor window in color scan sector

    Detection o peak/maximal velocities

    Detects highest velocities along scanline; peak velocity measurements

    Peak velocity measurements, but aliasingoccurs at high velocities

    Mean velocity measurements; aliasingoccurs as it is a PW-based modality

    Aliasing No aliasing Aliasing (Nyquist limit) A PW Doppler-based technique; alias

    Sensitivity Best sensitivity Good sensitivity Good sensitivity(jet size very sensitive to gain setting

    Depth resolution/range ambiquity

    Range ambiguity Range resolution; detects location o fow Range resolution;real time 2D anatomy o fow

    Spectrum analyzed Wide (spectral broadening) Narrow; spectral broadening seen withturbulent fow and large sample volumes

    Wide color-encoded Doppler shits oconventional *BART scale and color

    hues, saturation, and luminance*Blue away, red towards.

    Figure 1-4. Comparison: continuous wave, pulsed wave, and color ow Doppler.

    Figure 1-. Images obtained by color ow Doppler (CFD) rames must

    be interpreted with reerence to the velocity scales displayed on the image.

    Tese reerence maps or scales provide inormation on average ow velocities

    based on three characteristics o color: 1) hue, 2) luminance, and 3) satura-

    tion. wo major modes o display are velocity mode and variance mode.Left panel, Conventional velocity mode: ow towards transducer is

    coded (by convention) in shades o red. Flow away rom transducer is color-

    coded in shades o blue (BAR: blue away, red towards). Black color on the

    sale indicates no ow. Lighter shades o BAR indicate aster ow.

    Right panel, Variance mode maps provide inormation in addition to

    velocity and direction. Tey can help distinguish laminar rom turbulent ow.

    In laminar ow, the average velocities are not much dierent rom the peak

    velocities. Tey are close together (ie, they exhibit minimal variance). When

    blood ow is turbulent, there are much greater dierences between the

    average and peak velocities, namely greater variance. Deeper shades indicate

    slower laminar ow, while lighter shades may represent aster, turbulent ow.

  • 8/8/2019 AHD1601-01

    15/24

    Basic Principles of Echocardiography

    Te Normal Echocardiographic Examination and omographic Anatom

    Figure 1-5. ransthoracic two-dimensional echocar

    raphy: imaging planes, transducer positions, and standviews. A2Capical two-chamber view; A3Capical th

    chamber view; A4Capical our-chamber view; A5C

    cal fve-chamber view; PLAXparasternal long-axis vie

    PSAXparasternal short-axis view; PSAX-AVparaste

    short-axis view-aortic valve level; PSAX-MVparastern

    short-axis view-mitral valve level; PSAX-PMparastern

    short-axis view papillary muscle level; SC-4Csubcosta

    view (our-chamber).

    Would be better for the figure to occupy the majority of the page and the text below it - i.

    e., stretch figure down and out - with caption below

  • 8/8/2019 AHD1601-01

    16/24

  • 8/8/2019 AHD1601-01

    17/24

    Basic Principles of Echocardiography

    CFigure 1-7. (Continued)

    Figure 1-8. Echocardiographic parasternal

    inow scan plane and corresponding anatom

    IVCinerior vena cava; LLPVlet lower pul

    monary vein; LUPVlet upper pulmonary ve

    RAright atrium; RAAright atrial appenda

    RVORV outow tract; SVCsuperior ven

  • 8/8/2019 AHD1601-01

    18/24

    18\ Atlas of Echocardiography

    Figure 1-30. Echocardiographic short-axis s

    planes and corresponding anatomy.

    Figure 1-9. wo-dimensional long-axis view showing structures visua

    on the RV inow view. RAright atrium.

  • 8/8/2019 AHD1601-01

    19/24

    Basic Principles of Echocardiography

    Figure 1-31. wo-dimensional parasternal short-view showing the

    structures visualized at the level o the aortic valve. RAright atrium;

    RVORV outow tract.

    Figure 1-3. Views o the high parasternal

    short-axis scan plane at the level o the pul-

    monary arteries and corresponding anatom

    Aoaorta; DAdescending thoracic aort

    PApulmonary artery.

  • 8/8/2019 AHD1601-01

    20/24

    0\ Atlas of Echocardiography

    Figure 1-34. Anatomical structures correspo

    to the parasternal short-axis (PSAX) scan plan

    the level mitral valve (MV) and papillary musc

    Figure 1-33. wo-dimensional echocardiographic rame showing the s

    tures visualized on high parasternal short-axis view at the level o the ma

    pulmonary artery (PA) and biurcation into the right and let pulmonar

    arteries. Aoaorta; RAright atrium; RVORV outow tract.

    are really

    sting space heree figure above

    34) could be

    de bigger - theis minimal

  • 8/8/2019 AHD1601-01

    21/24

    Basic Principles of Echocardiography

    Figure 1-36. Echocardiographic our-chamb

    scan planes and corresponding anatomy. RA

    atrium; SVCsuperior vena cavity.

    Figure 1-35. wo-dimensional short-axis rames showing structures visual-

    ized on the parasternal short-axis (PSAX) views at the mitral valve level (left

    panel), mid ventricular or papillary muscle level (right panel), and apic

    level (right panel).

  • 8/8/2019 AHD1601-01

    22/24

    \ Atlas of Echocardiography

    Figure 1-39. wo-dimensional rame showing structures

    alized on the suprasternal notch view.

    Figure 1-38. Echocardiographic suprastern

    notch scan plane and corresponding anatom

    LAlet atrium; PApulmonary artery; R A

    right atrium; SVCsuperior vena cavity.

    Figure 1-37. wo-dimensional our-chamber rames showing the structures

    visualized on the apical our-chamber (A4C) ( left panel), apical fve-chamber

    (A5C), (middle panel), and subcostal our-chamber views (right panel).

    LAlet atrium; RAright atrium.

  • 8/8/2019 AHD1601-01

    23/24

    Basic Principles of Echocardiography

    Figure 1-40. Panoramic schema o coronary

    artery territories and LV segments using nom

    ture recommended by the American Heart A

    ation Writing Group on Myocardial Segment

    and Registration or Cardiac Imaging [2]. IVC

    rior vena cavity; LAlet atrium; laalet atr

    appendage; LADlet anterior descending aPApulmonary artery; LAOlet anterior o

    LCxlet circumex; RAOright anterior ob

    RCA/PDAright coronary/posterior descend

    SVCsuperior vena cavity.

    1. Otto C, ed: Principles o echocardiography imaging acquisition and Doppler

    analysis. In extbook of Clinical Echocardiography, edn 3. Philadelphia:

    WB Saunders; 2004.

    2. Cerqueira MD, Weissman NJ, Dilsizian V, et al.: Standardized myocardial segmen

    and nomenclature or tomographic imaging o the heart: a statement or health

    proessionals rom the Cardiac Imaging Committee o the Council on Clinical C

    ogy o the American Heart Association. Circulation 2002, 105:539542.

    References

    Recommended Reading

    Bushberg J, Siebert JA, Leidholdt EM Jr, Boone JM: Te Essential Physics of Medical

    Imaging, edn 2. Philadelphia: Lippincott Williams & Wilkins; 2002:469553.

    Cape EG, Yoganathan AP: Principles and instrumentation or Doppler. In Marcus

    Cardiac Imaging, edn 2. Philadelphia: WB Saunders; 1996:273291.

    Feigenbaum H: Echocardiography, edn 4. Malvern, PA: Lea and Febiger; 1986.

    Gardin JM, Adams DB, Douglas PS, et al.: Recommendations or a standardized repo

    adult transthoracic echocardiography: a report rom the American Society o E

    diographys Nomenclature and Standards Committee and ask Force or a Stan

    ized Echocardiography Report.J Am Soc Echocardiogr 2002, 15:275290.

  • 8/8/2019 AHD1601-01

    24/24

    Geiser EA: Echocardiography: physics and instrumentation. In Marcus Cardiac Imaging,

    edn. 2. Edited by Skorton DJ, Schelbert HR, Wol GL, Brundage BH.

    Philadelphia: WB Saunders; 1996:273291.

    Hatle L, Angelsen B: Doppler Ultrasound in Cardiology: Physical Principles and

    Clinical Applications, edn. 2. Philadelphia: Lea & Febiger; 1985.

    Henry WL, DeMaria A, Gramiak R, et al.: Report o the American Society o Echo-

    cardiography Committee on nomenclature and standards in two-dimensional

    echocardiography. Circulation 1980, 62:212217.

    Hung J, Lang R, Flachskamp F, et al.: 3D echocardiography: a review o the current

    status and uture directions.J Am Soc Echocardiogr 2007, 20:213233.

    Kremkau FW: Diagnostic Ultrasound: Principles and Instruments, edn 6. Philadelphia:

    WB Saunders; 2002.

    McAlpine WA: Heart and Coronary Arteries: an Anatomical Atlas for Clinical

    Diagnosis, Radiological Investigation, and Surgical reatment . Berlin, Heidelberg,

    New York: Springer-Verlag; 1975.

    Medical diagnostic ultrasound instrumentation and clinical interpretation: repo

    the ultrasonography task orce.JAMA 1991, 265:11551159.

    ORourke MF, Nichols WW: McDonalds Blood Flow in Arteries, edn 5. New York

    Oxord University Press; 2005.

    Quinones MA, Douglas PS, Foster E, et al.: ACC/AHA clinical competence statem

    on echocardiography: a report o the American College o Cardiology/Ame

    Heart Association/American College o Physicians - American Society o In

    Medicine ask Force on Clinical Competence. J Am Coll Cardiol 2003, 41:687

    Quinones MA, Otto CM, Stoddard M, et al.: Recommendations or quantifcatio

    Doppler echocardiography: a report rom the Doppler Quantifcation ask

    o the Nomenclature and Standards Committee o the American Society ocardiography. J Am Soc Echocardiogr 2002, 15:167184.

    Seghal CM: Principles o ultrasonic imaging and Doppler ultrasound. In extboo

    Echocardiography and Doppler in Adults and Children . Edited by St. John Su

    MG, Oldershaw PJ, Kotler MN. Oxord, UK: Blackwell Science; 1996:330.