amber - a novel, non-invasive approach to direct neutrino mass measurement

1
AMBER - A novel, non-invasive approach to direct neutrino mass measurement J.A.Thornby , M.J.Hadley, A.Lovejoy, Y.A.Ramachers Department of Physics, University of Warwick, Coventry, CV4 7AL AMBER (Anti-neutrino Mass Balance using Electrostatics and Radioactivity) is a novel technique for direct neutrino mass measurement. It is a beta decay endpoint experiment but differs in two important regards from conventional experiments, such as KATRIN [1]. AMBER will use a Ni-63 source, as opposed to Tritium, and substitutes traditional energy measurements for high precision Voltage measurements. SOURCE BIAS & HIGH VOLTAGE SUPPLY • Collecting enough electrons to yield the Q-value potential of 66.9 KeV would be very slow. • The AMBER source is biased (and electronics are floated) at a high voltage near the Q-value. → Only have to complete the last part of the decay spectrum. • Therefore AMBER requires a stable High Voltage controller – a limiting factor of the experiment. • KATRIN at the limit of technology here [1], but Warwick are developing a new such device in-house. CURRENT RESULTS & THE FUTURE OF AMBER • AMBER still in its infancy, only a few months old! But numerous design iterations to date. • Still much to be done: → DAQ/F-E design; Calibration using Cd-109; HT stability; Measure “zero” for absolute reference. • However a very versatile technique, with other secondary applications: → Search for tensor currents; Creation of High Voltage calibration standard based on nuclear physics; Possible sensitivity to moon phases; Gas conductivity measurements. • Some time has been spent testing AMBER in air, where there is leakage to ground. • Unusual voltage decay characteristics have prompted further study. • Added ability to set ball voltage before levitating – in order to observe decay curves. → Measure conductivity of air – not very well understood in atmospheric physics! • Shown below are a series of decay curves – note the unusual dips for decays from positive voltages. → More work needed to understand the technique – systematic effect or new physics? MEASURING VOLTAGES – The Kelvin Technique • AMBER’s capacitor “plate” is a levitating ball – suspended by two permanent magnets and an electromagnet. • Levitation serves multiple purposes and is controlled by unique electronics. • Levitation coils (see circuits) use only μW of power to maintain the ball’s equilibrium. • Combined with vacuum conditions ~ 10 -5 mbar means: → Ball has no mechanical couplings to anything and has no leakage path to ground. • Ball’s potential measured non-invasively relative to pickup plate (see circuits) serving as an inverse Kelvin probe. • Alternating current in the levitation coil causes the ball to oscillate vertically at ~ 11 Hz. • Oscillation causes change of capacitance between ball and pickup plate. • Electronics measures ball’s potential (wrt pickup plate) in response to capacitance change due to oscillation. BASIC PRINCIPLE • Ni-63 beta decay source used as an electron current source to charge a capacitor. Voltage measurements are converted to collected charge as a function of time. → Yields an integrated beta decay spectrum, used to reconstruct beta spectrum. • Process is self quenching - only the most energetic electrons are able to overcome the electrostatic repulsion of the amassed charges. • Eventually charge on the plate stabilises when no more electrons can be gathered. Potential created by stored charge corresponds to the beta decay endpoint energy. ELIMINATING EXTERNALS • Ball has a potential with respect to all of its surroundings: Coat the ball and its container in Nickel to eliminate contact potentials, using surface evaporation techniques. → AMBER’s electronics drive the container at a voltage that follows the ball’s potential. → Net result: Ball sees no external potentials and has no capacitance to its surroundings – except to Ni-63 source. t C source C dt t U I C U 0 )) ( ( 1 PRECISION High precision essential for accurate endpoint and spectrum determination. • AMBER prototype currently has 10mV precision, 1mV precision is tenable. • KATRIN overall sensitivity to neutrino mass of 0.2 eV. • Beta decay of Ni-63 has a Q-value of 66.945 keV. • Kurie plot below shows endpoint energy resolution effects. Using new “magic wand” technology (pictured right, patent pending) and a modified “Cockcroft Walton ladder”. Above: The AMBER prototype vacuum chamber. Left: The AMBER prototype module, demonstrating: 1) Levitating ball 2) Pickup plate 3) Permanent magnets 4) EM levitation coil 5) AMBER electronics. M iniVoltage D ecays -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 1 101 201 301 401 501 601 701 801 901 Tim e (10 s) V oltage (100 V ) REFERENCES [1] KATRIN Design Report, KATRIN Collaboration, 2004. ΔE = 0.01 eV ΔE = 0.93 eV 1 2 3 3 4 5

Upload: trista

Post on 06-Jan-2016

16 views

Category:

Documents


0 download

DESCRIPTION

Δ E = 0.01 eV. Δ E = 0.93 eV. AMBER - A novel, non-invasive approach to direct neutrino mass measurement J.A.Thornby , M.J.Hadley, A.Lovejoy, Y.A.Ramachers Department of Physics, University of Warwick, Coventry, CV4 7AL. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: AMBER - A novel, non-invasive approach to direct neutrino mass measurement

AMBER - A novel, non-invasive approach to direct

neutrino mass measurementJ.A.Thornby, M.J.Hadley, A.Lovejoy, Y.A.Ramachers

Department of Physics, University of Warwick, Coventry, CV4 7AL

AMBER (Anti-neutrino Mass Balance using Electrostatics and Radioactivity) is a novel technique for direct neutrino mass measurement. It is a beta decay endpoint experiment but differs in two important regards from conventional experiments, such as KATRIN [1].

AMBER will use a Ni-63 source, as opposed to Tritium, and substitutes traditional energy measurements for high precision Voltage measurements.

SOURCE BIAS & HIGH VOLTAGE SUPPLY

• Collecting enough electrons to yield the Q-value potential of 66.9 KeV would be very slow.

• The AMBER source is biased (and electronics are floated) at a high voltage near the Q-value.

→ Only have to complete the last part of the decay spectrum.

• Therefore AMBER requires a stable High Voltage controller – a limiting factor of the experiment.

• KATRIN at the limit of technology here [1], but Warwick are developing a new such device in-house.

CURRENT RESULTS & THE FUTURE OF AMBER

• AMBER still in its infancy, only a few months old! But numerous design iterations to date.

• Still much to be done:

→ DAQ/F-E design; Calibration using Cd-109; HT stability; Measure “zero” for absolute reference.

• However a very versatile technique, with other secondary applications:

→ Search for tensor currents; Creation of High Voltage calibration standard based on nuclear physics; Possible sensitivity to moon phases; Gas conductivity measurements.

• Some time has been spent testing AMBER in air, where there is leakage to ground.

• Unusual voltage decay characteristics have prompted further study.

• Added ability to set ball voltage before levitating – in order to observe decay curves.

→ Measure conductivity of air – not very well understood in atmospheric physics!

• Shown below are a series of decay curves – note the unusual dips for decays from positive voltages.

→ More work needed to understand the technique – systematic effect or new physics?

MEASURING VOLTAGES – The Kelvin Technique

• AMBER’s capacitor “plate” is a levitating ball – suspended by two permanent magnets and an electromagnet.

• Levitation serves multiple purposes and is controlled by unique electronics.

• Levitation coils (see circuits) use only μW of power to maintain the ball’s equilibrium.

• Combined with vacuum conditions ~ 10-5 mbar means:

→ Ball has no mechanical couplings to anything and has no leakage path to ground.

• Ball’s potential measured non-invasively relative to pickup plate (see circuits) serving as an inverse Kelvin probe.

• Alternating current in the levitation coil causes the ball to oscillate vertically at ~ 11 Hz.

• Oscillation causes change of capacitance between ball and pickup plate.

• Electronics measures ball’s potential (wrt pickup plate) in response to capacitance change due to oscillation.

BASIC PRINCIPLE

• Ni-63 beta decay source used as an electron current source to charge a capacitor.

• Voltage measurements are converted to collected charge as a function of time.

→ Yields an integrated beta decay spectrum, used to reconstruct beta spectrum.

• Process is self quenching - only the most energetic electrons are able to overcome the electrostatic repulsion of the amassed charges.

• Eventually charge on the plate stabilises when no more electrons can be gathered.

• Potential created by stored charge corresponds to the beta decay endpoint energy.

ELIMINATING EXTERNALS

• Ball has a potential with respect to all of its surroundings:

→ Coat the ball and its container in Nickel to eliminate contact potentials, using surface evaporation techniques.

→ AMBER’s electronics drive the container at a voltage that follows the ball’s potential.

→ Net result: Ball sees no external potentials and has no capacitance to its surroundings – except to Ni-63 source.

t

CsourceC dttUIC

U0

))((1

PRECISION

• High precision essential for accurate endpoint and spectrum determination.

• AMBER prototype currently has 10mV precision, 1mV precision is tenable.

• KATRIN overall sensitivity to neutrino mass of 0.2 eV.

• Beta decay of Ni-63 has a Q-value of 66.945 keV.

• Kurie plot below shows endpoint energy resolution effects.

Using new “magic wand” technology (pictured right, patent pending) and a modified “Cockcroft Walton ladder”.

Above: The AMBER prototype vacuum chamber.

Left: The AMBER prototype module, demonstrating: 1) Levitating ball 2) Pickup plate 3) Permanent magnets 4) EM levitation coil 5) AMBER electronics.

Mini Voltage Decays

-2

-1.5

-1

-0.5

0

0.5

1

1.5

2

1 101 201 301 401 501 601 701 801 901

Time (10 s)

Vo

ltag

e (

100 V

)

REFERENCES

[1] KATRIN Design Report, KATRIN Collaboration, 2004.

ΔE = 0.01 eV ΔE = 0.93 eV

1

2

3

3

4

5