• pH observed with Honeywell pH electrodes

• Current Plan for Calibration and Cleaning

• Gel replacement• Checking with

Buffers

pH: SATURN-03; SATURN-04

2pH vs Temperature; SATURN-03

0.0 5.0 10.0 15.0 20.0 25.04.50

5.25

6.00

6.75

7.50

8.25

BC pH 7.00 buffer AC pH 7.00 bufferpH 7.00 buffer - temp line

pH in 7.00 Buffer BC and AC vs Temper-ature

0.0 5.0 10.0 15.0 20.0 25.08.50

8.70

8.90

9.10

9.30

9.50

9.70

9.90

10.10

10.30

10.50

BC pH buffer 10.00 AC pH buffer 10.00pH buffer 10.00 - temp line

• BC = before cleaning• AC = after cleaning• Calibration Analysis eleven times from 20120917 - 20130104

3Percent Drift SATURN-03

7 18 25 36 55 63 69 76 83 108

-20.00

-10.00

0.00

10.00

20.00

30.00

40.00

50.00

60.00

70.00

80.00

pH 7.00 % drift pH 10.00 % drift

days

SATURN-03 pH Percent Drift: 20120917 - 20130104

4pH vs Temperature; SATURN-04

0 2 4 6 8 10 12 14 16 18 206.94

6.96

6.98

7.00

7.02

7.04

7.06

7.08

7.10

7.12

BC pH 7.00 buffer AC pH 7.00 buffer

pH 7.00 buffer - temp line

pH 7.00 Buffer BC and AC vs Temperature

0 2 4 6 8 10 12 14 16 18 209.9

9.95

10

10.05

10.1

10.15

10.2

10.25

10.3

10.35

BC pH buffer 10.00 AC pH buffer 10.00

pH 10.00 buffer - temp line

pH 10.00 Buffer BC and AC vs Temperature

• BC = before cleaning• AC = after cleaning• Calibration Analysis six times from 21021002 – 20130107

5Percent Drift; SATURN-04

0 10 20 30 40 50 60 70 80 90 100

-8

-6

-4

-2

0

2

4

6

8

pH 7.00 % drift pH 10.00 % drift

SATURN-04 pH Percent Drift

days

6Current Calibration Plan (including gel replacement)• Calibrate once monthly, record pH before and after cleaning and track adjustments made• Check in buffers at least once weekly before and after cleaning• Track data and look for trends and when to replace gel