NEW KERARING SI-6FOR KERATOCONUS TREATMENT
ASCRS, BOSTON 2010
FRANCISCO SANCHEZ LEON MDINSTITUTO NOVAVISION CD. MÉXICOwww.novavision.compacornea@yahoo.comTEL +52555 3438882
NO FINANCIAL INTEREST TO MEDIPHACOS CO.
150°X 200m150°X 200m
150° X 300m150° X 300m
Pourpose
Evaluate the corneal remodeling using the new Keraring SI 6 for the Keratocous Treatment.
©1997 – 2006 IntraLase Corp. All rights reserved. Confidential Mkt Doc 470 Rev. A
Corneoplastic Techniques
Mechanical methods Intracorneal segmentsIncisional procedures
Biochemical methods Cross-linking UV lightConductive Keratoplasty (CK)
Mix methods
Excimer laser ablationInlays (Acufocus)
Optimized prismatic design
Scalene triangular shape SI-6 compared to Isosceles triangle in SI-5.
Improves effect and optimizes total internal reflection, reducing glare and halos
800µ base width instead of 600µ
Wider base provides 25% greater implant volume and compensates for reduction of effect in the larger optical zone
When should Keraring SI-6 implantation be considered?
Mesopic pupil diameter ≥ 5mm BCVA is ≥ 0.6 Moderete Sphere ≤ 5 D Cylinder value is ≤ 5 D Segment thickness should not exceed 60% of the
minimum pachymetry at the 5.5 optical ring zone When PRK is planned as a second procedure. When phakic IOL implantation is planned as a
second procedure
Different Arc Lengths for each Refraction and Topography appearance
New Keraring SI-6IS-6 IntraLase 60 Hz parameters
Depth in Cornea 80% at 5.5 mm Incision Axis at steepest meridian Entry Cut Length 1 mm
Inner Diameter 6.0 mm Outer Diameter 6.8 mm
Ring Energy 1.5 Entry Cut Energy 1.3
New Keraring SI-6150°Asymmetric Thickness Implantation
-3.50/-4.00X135°=20/60-3.50/-4.00X135°=20/60
-0.50/-1.25 X 30°=20/30-0.50/-1.25 X 30°=20/30
150°X 200m150°X 200m
150° X 300m150° X 300m
+1.50 / -7.50 X 20°= 20/60+1.50 / -7.50 X 20°= 20/60
Plano/-1.25 X 45°= 20/30Plano/-1.25 X 45°= 20/30
New Keraring SI-6
120°symmetric implantation
120°X 250m120°X 250m
Single Inferior Segment for PMD Corneoplasty Achivement
PreopPreop PostopPostop
New SI-6 KeraringEarly Conclusions Keraring SI-6 achieves Corneoplasty purposes. Keraring SI-6 is not intended to replace the IS-5,
but rather intends to offer the surgeon additional tools to customize the corneal remodeling according to each individual patient’s needs