7526A Precision Temperature/Pressure Process Calibrator
7526A Precision Temperature/Pressure Process Calibrator
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 2
Introducing the7526A Precision Process CalibratorIntroducing the7526A Precision Process Calibrator
• Value proposition
• Typical applications
• Key features & benefits
• Product overview
• Product features
• Summary specs
• Options & accessories
A “best fit” for process instrumentation calibration
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 3
7526A Value Proposition7526A Value Proposition
• Balances precision and versatility with cost
• The 7526A is less expensive than high-end multi-product calibrators, yet more precise and versatile than handheld field calibrators
• Calibrates wide ranging workload
Versatility, precision & valueThe 7526A offers an isolated measurement channel allowing it to precisely, and simultaneously, source and measure temperature & electrical parameters. Pressure measurement can be viewed in both displays in different units of measure.
7526A
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 4
Typical ApplicationsTypical Applications
IndustryIndustry Applications Applications Calibration workloadCalibration workloadPharmaceutical/Bio-tech
Monitor temp/pressure process, monitor freezers/ovens/incubators, freeze-drying, SIP
Temp/pressure transmitter, pressure gauges, handheld calibrator, temp readout, chart recorder
Food Monitor process temp, food dryers, monitor cold storage, monitor refrigeration equipment, SIP
Temp/pressure transmitter, temp readout, limit alarm, PID controller
Oil & Gas Monitor pipeline provers, custody transfer
Temp/pressure transmitter, pressure gauges, flow transmitter
Chemicals Monitor temp/pressure reaction conditions, refrigeration, boilers
Temp/pressure transmitter, limit alarms, pressure gauges
Electronic Equipment Manufacturer
Manufacturing electronic devices
Automatic test station
3rd Party Calibration Lab Service industries listed above
All of the above
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 5
7526A Key Features & Benefits7526A Key Features & Benefits
Feature Feature BenefitBenefit
Improved thermocouple accuracy
Can calibrate Documenting Process Calibrators (2:1 TUR) and nearly all handheld temperature & pressure calibrators
Switch test (unique to 7526A)
Increased workload. Not every temperature source, such as a drywell, includes a switch-hold function. Pressure sources do not.
Simultaneous source and measure
Calibrate more DUTs in less time. See results of an adjustment immediately.
MET/CAL® Calibration Software compatible
Write automated procedures, create reports.
Value Costs less than a DPC, highly versatile, precise but competitively priced.
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 6
7526A Product Overview7526A Product Overview
• Precision Temperature/PressureProcess Calibrator
• Sources and measures:
– Thermocouples (13 types)
– RTDs (9 types)
– Thermistors (YSI-400)
– Voltage (0 to 100 Vdc)
– Resistance (0 Ω to 4 kΩ)
– Current (100 mA output / 50 mA input)
• Measure only:
– Pressure (0 to 700 bar / 10,000 psi)
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 7
7526A Product Overview (Left Half)7526A Product Overview (Left Half)
DC voltage output terminals0 Vdc to 100 VdcAccuracy: 30 ppm (+3 μV)
DC current output terminals0 mA to 100 mAAccuracy: 50 ppm
RTD/Ω output terminals (two-wire)5 Ω to 4 kΩ, Accuracy: ± 0.05 ºC, –200 to 630 ºC
Four-wire RTD/Ω input terminalsAccuracy: ±0.02 ºC Pt 385, 100 Ω, –80 to 100 ºC
Isolated pressure module inputAccepts Fluke 700 & 525A series pressure modules
Thermocouple input/output terminalAccuracy: ± 0.1 ºC–100 to 800 ºC (type K)
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 8
7526A Product Overview (Right Half)7526A Product Overview (Right Half)
Isolated input terminals Allows simultaneous source and measure dc voltage (0 to 100 V)dc current measurement (0 to 50 mA)Switch-test input 24 V dc loop power supply
Numeric and secondary function keys
Isolated input controlsdc voltage/current Switch test mode24 Vdc power supply250 Ω HART resistorPressure input mode
Cursor controls
Input/Output Function Keysdc voltage/currentTC/RTD Pressure modeUnit typeStby/Operate
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 9
7526A Product Overview (Rear Panel)7526A Product Overview (Rear Panel)
RS-232 Port GPIB IEEE-488 Service port Chassis ground
Power line voltage selector and fuse compartment
AC power inlet 120/240 AC
Power switch
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 10
7526A Product Features7526A Product Features
STBY/OPR •In Standby mode, changes to the output value are not applied until you select the Operate mode•In Operate mode, each change to the output value is applied immediately•Voltages more than 30 V are not applied. The mode automatically reverts to Standby for safety
• In Thermocouple mode, move through the thermocouple types (including millivolts)
• In RTD/Ohms mode, move through the RTD types (including ohms)
• In Pressure mode, move through the pressure units
Select a secondary function from the numeric keypad
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 11
7526A Product Features7526A Product Features
HART •Enables a 250 Ω internal resistor
•Without the loop resistance in place, the DC power supply would “short out” the communicator’s AC voltage signal
•The presence of a loop resistor in the circuit prevents the DC power supply from “loading” the AC voltage signal by the communicator.
•Some minimum loop impedance is needed so that the HART signal can be ‘seen’ or ‘read’ by a HART master, like a HART communicator or HART modem.
LOOP PWR•Turns on an internal 24 Vdc power supply in series with the current measurement circuit
•Used to power 4 – 20 mA transmitters
∿
↓∿ ↓
HARTtransmitter
HART communicator
250 < R < 1100
Powersupply
Computer
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 12
7526A Product Features7526A Product Features
RTD and Ohms Measure/Source•Measures & Sources all common RTD types
•Stores CVD coefficients for up to five RTDs
•Stores coefficients for one SPRT (ITS-90)
•Resistance from 0 Ω to 4 kΩ.
RTD types:
•Pt 385 100 Ω, 200 Ω, 500 Ω, 1000 Ω•Pt 3926 100 Ω•Pt 3916 (JIS) 100 Ω•Ni120 120 Ω•Cu 427 (Minco) 10 Ω•YSI 400 thermistor
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 13
7526A Product Features7526A Product Features
Thermocouple (TC) Measure/SourceMeasure & sources all common thermocouple types and basic millivolts from -10.0 to 75.0 mV
Thermocouple types: B, C, E, J, K, L, N, R, S, T, U, XK, BP
Pressure Measure•The isolated pressure display uses the same physical pressure connector as the primary display•Both displays can show pressure at the same time •The same source can be shown in different pressure units
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 14
7526A Pressure Modules7526A Pressure Modules
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 15
7526A Product Features7526A Product Features
Preset setpoints
Nine preset output setpoints can be stored and recalled for each of these output modes:
•Voltage•Current•Each thermocouple type (including mV)•Each RTD type (includes each of the five custom curves)
Switch Test Mode•When in Switch Test mode the Calibrator can sense a contact transition (closed or opened)
•At the time of the transition, the Calibrator records the value and units shown on the primary display
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 16
7526A Product Features7526A Product Features
Remote Interface
•RS-232 or GPIB (IEEE-488) can control 7526A from a PC
•A serial to USB adapter cable is included as standard with the 7526A
•Individual commands can be typed into terminal emulator software
•MET/CAL compatible – write automated procedures and generate calibration reports
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 17
7526A Workload7526A Workload
Product TypesTwo wire transmittersI/P convertersP/I convertersFourwire conditionersPressure switchesPressure GaugesLimit alarmsPanel metersPID controllersThermometersIsolatorsDMMs Chart Recorders
CompaniesRosemountMoore industriesWilkerson InstrumentsM-SystemsAPI OmegaNewportYokogawaEuthrothermWestinAnalog DevicesAction InstrumentsCR Magnetics
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 18
7526A Summary Specs7526A Summary Specs
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 19
7526A Summary Specs (cont.)7526A Summary Specs (cont.)
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 20
Convert Spec in PPM to % ReadingConvert Spec in PPM to % Reading
Convert a spec in ppm to % error of reading
1. Divide accuracy spec (30 ppm) by 1,000,000=30 ppm/1,000,000 = 0.00003
2. Multiply reading (10 V) by result above=10 V x 0.00003 = 0.0003 V
3. Add floor offset to voltage error (100 µV in this case)=0.0003 V + 0.0001 V = 0.0004
4. Add combined voltage error to reading= 0.0004 ± 10 V = 10.0004 or 9.9996
5. Divide result by reading = 10.004 V / 10 V = ± 0.004%
Example:
A part per million simply means a fraction of 1/1,000,000 Accuracy is expressed as ppm of reading + floor offsetAssume we are using the 0 to 10 V range and reading 10 V
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 21
7526A Standard & Optional Equipment7526A Standard & Optional Equipment
Included in the box:
•7526A Precision Process Calibrator•Users Manual CD-ROM•Getting Started•AC Mains Cord•Thermocouple Shorting Jumper•NIST Traceable Calibration Report•USB to Serial Adapter Cable
Options and accessories:
•5520A-525A Leads kit•Y7526A Rack Mount kit•Fluke 700 and 525A-P series pressure modules•MET/CAL with 7526A Function Select Code (FSC)•MET/CAL 7526A calibration procedure•Optional accredited calibration
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 22
QuestionsQuestions
?
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 23
Appendix – P/I TransmitterAppendix – P/I Transmitter
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 24
Appendix – I/P TransmitterAppendix – I/P Transmitter
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 25
Appendix – V/I TransmitterAppendix – V/I Transmitter
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 26
Appendix – RTD TransmitterAppendix – RTD Transmitter
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 27
Appendix – TC TransmitterAppendix – TC Transmitter
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 28
Appendix – RTD IndicatorAppendix – RTD Indicator
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 29
Appendix – Temperature Switch TestAppendix – Temperature Switch Test
©2012 Fluke Corporation Fluke 7526A Precision Process Calibrator 30
Appendix – Pressure Switch TestAppendix – Pressure Switch Test