SERIAL NUMBERS This manual applies directly to HP 8510 network analyzers. and HP 8SllA. HP 8512A HP 8533A HP 8534A and HP 851SA test SCIS m,ith serial number prettxes as follows:

HP 85102A 2427 2402

HP SS13A HP 8514A

2345 2343

HP 851 IA HP 8512A

2345 2336

HP 8515A 2345

For additional important information about serial numbers. see INSTRUMENTS COVERED BY M4NUAL in Section 1 ofthis Operating and Service Manual. or Section VII Backdating.

FIRMWARE AND SOi=Ti4fARE REVISIONS This manual applies directly to the following softwarc and firm- ware supplied N,ith the HP 8510 system:

Part Number Revision

8SlOl-loo01 (syslem firmware) 85101-loo02 (system firmware w/option 010) 85103-loo03 (adjustments and diagnostic) 08510-10001 (adjustments) OS510-loo06 (performance tesl HP 8510A/8511A) 08510-loo02 (perroramncr ~CSI HP 8SlOA/8512A) 08510-loo03 (performance test HP 8510AIP513A) 08510-10004 (performance test HP 85lOA!8514A) 08510-loo05 (performance test lW8510iV8515A)

A.O1.OO A.O1.OO

1.0 1.0 1.0 1.0 1.0 1.0 1.0

For later revisions oithese programs, 5 Manual Changes pack- age will be included with this manual.

m Copyright HEWLF,-WACl‘ARD COMPANY 1984 1400 FOUNTAINGROVE PARKWAY. SANTA ROSA, CA 95401 U.S.A.

MANUAL PART NO. 05510-90001 Micmfiehe Pan No. 08510-90002 Printed: MARCH 1984

@iI HEWLETT PACKARD

COPYRIGHT AND DISCLAIMER NOTICE

Copyright - Agilent Technologies, Inc. Reproduced with the permission of Agilent Technologies Inc. Agilent Technologies, Inc. makes no warranty of any kind with regard to this material including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies, Inc. is not liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material or data.

GIDEP Metrology
Cropping of the black borders around each page was not accomplished, as it increases file sizes in the database. However, this action will be accomplished on files in future versions of Metrology DVD.

‘X-RAY RADIATIDN NOTICE

pi-z-1 Mode, FrS,n, A pKq

Wrihrend de6 Betriebs erzeugt dieses C&r&t Rb’ntgenstrahlung. Das G&it ist 80 abgeschirmt. dap die Dosisleistung weniger als 36 pA/kg (0.5 mR/h) in 5cm Abstand van der Obetfkche der Katodenstrahlr6hre b&r&$ Somit sind die Sichcr heitohatimmungen verschiedener I%nder. u.A der deutschen Rktgenverordnung eingehalten.

Die Stirke der IQintgenstrahlung hk-@ im Wesent. lichen van der Bauart der Katodenstrahlr’dhre sb. sotie van den’.Spannungen, w&he an dieser anhegen. Urn Men tichan Be&b zu gewahrleien, diirfen die Einsiellungen der Niederspannuegs. und des Hochspannungsnett~ils nur nach der Anleitung in Kapitel V des Handbuches vorgenommen werden.

When operating, this instrument emits r.rays; however, it is well shielded and meets safety and health requiremente of various countries, such 8s the X-ray Radiation Act of Germany.

Radiation emitted by this instrument ia less fian 0.5 mR!hr at B diaiance of five (5) centimetere from the surface of the eathcderay tube. The xmy radiation primarily depends on the charac- teristics of the cathode-ray tube and its asawiated low-voltage and high-voltage circuitry. To ensure safe operation of the instrument. adjust both the low-voltage and high+oltage power supplies a6 outlined in Section V of this manual (if appli. able).

Die Katodenstrahlrirhre darf nur durch die gleiche Type emetzt wverden. &he Kapirel Vi fiir KP - ErS*tZteile).

Replace the cathode-ray tube with an identical CRT only. Refer to Section VI for proper HP pan number.

Das Getit ist in Deutwhland zugelassen u&r

derNummer: 39+79,h/Rb Number of German License:

Table

l-1

1-2

l-3

1-4

1-5

1-6

l-7

l-8

1-9

1-10

l-11.

1-12

Figure

l-1

1-2

TABLES

Page

Modification Kits for HP 83500.Series RF Plug-ins .............. 1-5

Specifications for HP 8510 with HP 8512A Test Set .............. 1-7

Specifications for HP 8510 with HP 8514A Test Set.............. I-20

Supplemental Characteristics for HP 8510 System ................ 1-34

Time Domain S'upplemental Characteristics ....................... I-42

Accessories Available for the HP 8510 System................... I-45

Required Test Equipment for Board-Level On-Site Repair......... I-54

Test Equipment for Display/Processor Component Level Bench Repair................................................... 1-55

Test Equipment for IF/Detector and Test Sets Component Level Repair.........................................,.......,. I-56

Specifications for HP 8510 with HP 8511A Test Set...:.......... l-57

Specifications for HP 8510 with HP 8513A Test Set .............. 1-65

Specifications for HP 8510 with HP 8515A Test Set .............. l-80

ILLUSTRATIONS

Page

HP 8510 Network Analyzer System................................ 1-O

Typical Serial Number Label........................+........... I-I

CONTENTS

Introduction....................... Systems Controlled By This Manual.. Safety Considerations..............

........

........

System Specifications.......................

System Descriptions... HP 8510A/8511A System. HP 8510A/8512A System. HP 8510A/8513A System. HP 85lOA/8514A System. HP 8510A/8515A System.

. . * . . . .

. * . . . . .

. . .

. . .

.a . . . . .

. . .+ . . .

. . . . . . .

..*.*..

. . . . . .

. . . . . .

. ..*...

. ...*..

Sweep Oscillators Compatible with an HP 8510 HP 8340A Synthesized Sweeper Compatibility.. HP 83508 Sweep Oscillator Compatibility..... HP 83500-Series RF Plug-in Compatibility..+.

Options................ Option 101 Time Domain. Product Support Options HP 8510T t24D.......... HP 8510T t23N .......... HP 8350A t23L .......... HP 8510T t23G .......... HP 8510T t23B .......... HP 8510T t23A .......... HP 8510T t25D .......... HP 8510T t23U .......... System Option 910 ...... System Option 914 ......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

......

Equipment Required...................;......

Accessories Available.......................

Recommended Test Equipment..................

Warranty....................................

.,..,..... . . . . . .

. . . . . . . . . .

. . . . . . . . . .

. . . . . . . . . .

. . . . . . . . . .

. . . . . . . . . .

. . . . . . . . . .

System.... . . . . . . .

+.......a. . ...*.....

..........

..........

..........

..........

..........

..........

..........

..........

..........

..........

..........

. . . . ...<..

. . . . . . . . . .

Page

l-1 1-l 1-2

1-2

l-2 I-2 1-2 1-3 l-3 1-3

l-4 1-4 1-4 1-4

l-5 1-5 1-5 l-5 l-5 l-6 I-6 I-6 1-G l-6 I-6 1-6 1-6

1-6

1-6

1-6

l-6

1 SAFETY CONSIDERATIONS

;ENERAL SAFETY EARTH GROUND

T ‘his product and related documentation must be n <viewed for familiaritation with safety markings a nd instructions before operation. This product has b een dcsigne,d and tested in accordance with inter- n ational standards.

This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wh’fng ter- minals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative

;AFETV SYMBOLS and be secured against any unintended operation

A Instruction manual symbol: the product will be marked with this symbol when it is necessary for BEFORE APPLYING POWER

the user to refer to the instruc- tion manual (refer to Table of Verify that the product is configured to match the Contents). available main power source per the input power

configuration instructions provided in this manual

// Indicates hazardous voltages.

If this product is to be energized via an autotrans former, make sure the common terminal is con netted to the neutral (grounded side of main: SUPPlY).

c Indicates earth (ground) terminal.

The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.

The CAUTION sign denotes a hazard. It calls attention to an operating procedure, practice, or the like, which, if not correctly performed or adhered to, could result in damage to or destruc- tion of part or all of the product. Do not proceed beyond a CAU- TION sign until the indicated conditions are fully understood and met.

SERVICING

pGz-1

Any servicing, adjustment. mainrenance, or repair of this product must be per- formed only by qualified personnel.

Adjustments described in this manual may be performed with power supplied to the product while protective covers are removed. Energy available at many puints may, if contacted, result in per- sonal injury.

Copocirors inside this product may still be charged even when disconnected from their power source

To avoid a fire hazard. only fuses with the required current rating and of rhe specified type (normal blow, time delay. etc.) are to be used for replacement.

I,, NOTE

Refer to ParagFaphs 1-15 and 1-29 for system configurations and acc.~~&es available. I

HP 8510 - General Informaticxl

-. rrgure 1-I. HP 8510 h’etwork Analyzer System

HP 8510 - General Information

SECTION I GENERAL INFORMATION

1-I. INTRODUCTION

1-2. The Hewlett-Packard Model 8510 network analyzer system (Figure l-1) con- sists of several separate Instruments: the display/processor, the IF/detector, single or multiple test sets, and an HP 8340A or HP 8350B sweep oscillator. These in- struments are interconnected to make up an HP 8510 system. The HP 8510 system includes all of the connecting cables, as well as various other accessories used with the system in calibr+tion and measurements. Additiofial system capabilities are obtained by adding an HP desktop computer (controller) and other peripheral equipment.

l-3. This manual covers system and instrument specifications, installat.ion, opera- tion, programming, performance verification, adjustments, replaceable parts, service, and repair of the HP 8510 system.

1-4. SYSTEMS COVERED BY THIS MANUAL

l-5. Attached to each instrument is a serial number label that has the instrument serial number printed on it. These numbers are recorded at the factory upofi ini- tial shipment so that service and warranty information can be maintained by the factory for the specific combination shipped.

l-6. A typical serial number label is shown in Figure 1-2. The serial number is in two parts. The first four digits followed by a letter comprise the serial number prefix. The last five digits form the sequential suffix that, is unique to each in- strument. The contents of this manual apply directly to instruments having the same serial number prefixes as those listed on the Title Page of this manual under SERIAL NUMBERS.

1:. SERIAL NUMBER r PREFIX SUFFIX ++

FACTORY) SER 1234A 12345

INSTALLED

J OPTIONS

Figure 1-2. Typical Serial Number Label

l-7. An instrument manufactured after the printing of this manual may have a serial prefix that is not listed on the Title Page. An unlisted serial prefix indi- cates that the Instrument is different from those documented In this manual. The manual for the Instrument is then supplied with a Manual Changes supplement that contains information that documents the differences.

1-8. In addition to change information, the Manual Changes supplement cmtairH information for correcting errom In the manual. To keep this manual as current as

HP 8510 _ General Information

possible, Hewlett-Packard recommends that you periodically obtain the latest Man- ual Changes supplement. The supplement for this manual is keyed to both the print date and part number of the manual; both appear on the Title Page. Copies of the Manual Changes supplement are available from Hewlett-Packard or by sub- scription as part of support option HP 85lOT +23U.

1-9. For information concerning a serial number prefix that is not listed on the title page or in the Manual Changes supplement, contact your nearest Hewlett- Packard Sales Support Office.

l-10. SAFETY COKSIDERATIOXS

l-11. Refer to the’safety sheet at the front of this section for general safety instructions. Other cautions and warnings that appear throughout this manual should also be carefully followed to prevent bodily injury or damage to the equipment.

l-12. SYSTEM SPECIFICATIONS

l-13. Tables l-2, l-3, l-10, l-11, and l-12 list measurement specifications for the HP 85lOA in combination with an HP 8512A, 8514A, 8511A, 8513A or 8515A t,est set and an RF source. The specifications describe the warranted performance Of the system over the temperature range of 0 o C to 55O C, except where noted in the specification table. The specifications shown are those obtained when an HP 8340A synthesized sweeper is used as the system RF source.

l-14. Table l-4 and some entries In the specification tables denoted as “typical”, “nominal”, or “approximate” are the typical, but not warranted, operating charac- teristics of the system. Supplemental characteristics provide information useful in applying the instrument by giving ncnwarranted performance parameters.

l-15. SYSTEM DESCRIPTION

l-16. HP 85lOA/851lA System

l-17. The combination of the HP 8511A test set with the HP 8510A network an- alyzer and RF source provides a four channel receiver/signal processor that Oper- ates over the frequency range of 45 MHz to 26.5 GHz. This system provides cm- figuration flexibility of a user supplied signal separation network allowing custom measurement capabilities. The test set contains four RE to IF converters all of which can operate over the entire dynamic range of the system.

I-18. HP 8510A/8512A System

l-19. The combination of the HP 8512A test set with the WP 8510A network an- alyzer and RF source provides a system for making reflection sion ($21) measurements over the frequency range of 500 MHz

(S ,I and transmis- f! o 18 GHz. This

system provides a broad dynamic range and a simple test set architecture. The design of the test set includes three separate RF-to-IF converters which eliminate the need to mechanically switch between reflection and transmission measurements. This test set uses a directional coupler for signal separation.

l-20. The recommended accessories for the HP 85lOA/8512A system are as follows:

HP 8510 - General Information

HP 85132A 7mm test port return cable HP 85050A 7mm calibration klt HP 8505lA 7mm verification kit

l-21. HP 8510A/8513A

l-22. The combination of the HP 8513A test set with the HP 8510A network an- alyzer and source provides a system for making reflection (S I) and transmission (S21) measurement over the frequency range of 45 MHz to 26.5 Gl-lz. This system provides broad frequency coverage with wide dynamic range, and a simple test, set architecture. The design of the test set Includes three separate- RF to IF con- verters which eliminate the need to mechanically switch between relection and transmission measurem’enta. This test set uses a bridge for signal separation.

l-23. The recommended accessories for the HP 8510A/8513A system are as follows:

HP 85133A 3.5 mm test port return cable HP 85052A 3.5 mm calibration kit HP 85053A 3.5 mm verification kit HP 8513OA 3.5 mm to 7 mm adapter set HP 85050A 7 mm calibration kit llP 85051A 7 mm verification kit . HP 85132A 7 mm test port return cable

l-24. HP 8510A/8514A System

l-25. The combination of the HP 8514A test set with the HP 8510A network an- alyzer and source provides a system for making S-parameter measurements over the frequency range of 500 MHz to 18 GHz. This system is suited for making mea- surements ~1 two port devices where it is not convenient to physically reverse the test device to measure all four S-parameters. The dual fourport architecture of the test set develops a separate reference channel for each Incident port, thus re- ducing the number of Internal switches to one. This test set uses directional cou- plers for signal splitting. For measurements of active devices, the test set includes two 90 dB step attenuators for changing the incident power level at either port, as well as provisions for applying dc bias from au external power supply to the test port center conductor.

I-26. The recommended accessories for the HP 8510A/8514A system are as follows:

HP 851323 7mm test port return cable set HP 85050A 7mm calibration kit HP 85051A 7mm verification kit

The HP 85132A tes: mrt return cable is also suitable for use with the HP 8510A/ 8514A; however, some Increased sensitivity to frequency drift and residual FM may result.

l-27. HP 85lOA/8515A System

1-28. The combination of the HP 8515A test set with the HP 8510A network an- alyzer and source provides a system for making S-parameter measurements over the frequency range of 45 MHz to 26.5 ‘GHz. This system is suited for making mea- surements on two port devices where it is not convenient to physically reverse the