hid validation: the 3500 series genetic...

Post on 04-Jun-2020

2 Views

Category:

Documents

1 Downloads

Preview:

Click to see full reader

TRANSCRIPT

HID Validation: The 3500 Series Genetic Analyzers

2

HID Validation: The 3500 Series Genetic Analyzers

Outline

• Scope• Materials and Reagents

• Validation Elements• Selection of Validation Data

• Additional Sources of Validation Information• User Bulletin

3

HID Validation: The 3500 Series Genetic Analyzers

Validation ScopeHardware and Reagents

• Instruments• (3) 8 Capillary Instruments (3500)• (3) 24 Capillary Instruments (3500xL)

• Size Standard• GeneScan™ 600 LIZ® Size Standard v2.0• GeneScan™ 500 ROX™ Size Standard

• Profiler Plus® / Cofiler® do not use normalization

• Polymer and Array• POP-4™ and 36 cm Capillary Array

4

HID Validation: The 3500 Series Genetic Analyzers

Validation ScopeSample Summary

Test Name Samples Input Replicates

Genotype Concordance and Reproducibility

•40 male and 42 female gDNA samples•4 racial groups•Controls

1-2 ng total DNA •1 replicate per sample•2 injections

Sizing Precision and Accuracy

Allelic ladder 1 µL per well •See below

Sensitivity and Normalization

•007 control DNA•3 gDNA samples

•STR: 0.125ng, 0.5ng, 1ng, 6ng

•5 replicates for DNA samples•3 replicates for NTC•4 injections•Minifiler: 0.125ng,

0.25ng, 0.5ng, .75ng

Mixture Analysis •4 pairs of male and female gDNA•Mixture ratios•1:0, 1:1, 1:3, 1:5, 1:9, and 0:1

•Recommended concentrations per kit

1 replicate per sample

Resolution •Allelic Ladder•Research material

•1 µL per well 24 replicates

5

HID Validation: The 3500 Series Genetic Analyzers

Validation ScopeAmpFℓSTR® Kits

• Identifiler®

• Minifiler™

• SGM Plus®

• Yfiler®

• Sinofiler™

• SEfiler Plus™

• Profiler Plus®

• Cofiler®

• Identifiler® Direct• Identifiler® Plus• NGM™

Precision and Accuracy

7

HID Validation: The 3500 Series Genetic Analyzers

Precision and AccuracyExample Configurations for Standard Deviation

Within Injection - 3500

• 8 data points per mean• 12 injections per plate• 12 columns

• 3 Instruments

• 12 data points per mean• 4 injections per plate• 8 rows

• 3 Instruments

Across Injections – 3500xL

8

HID Validation: The 3500 Series Genetic Analyzers

3500 Sizing Precision Study:Within an Injection of Identifiler® Allelic Ladder

9

HID Validation: The 3500 Series Genetic Analyzers

3500 Sizing Precision Study:Across Multiple Injections of Identifiler® Allelic Ladder Samples

96 Replicates of Identifiler® Allelic Ladder

3500 -12 Injections3500xL - 4 Injections

Introduction to Normalization

11

HID Validation: The 3500 Series Genetic Analyzers

Sources of CE Variation (3100)

• Numerous optical components, capillary array

• Electrophoresis, injection

• Sample and sample preparation

Raman Intensity, 3100file: INSTRUMENT DATA Mod2B.xls

y = -2.7202x + 6042.8R2 = 0.0059

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

0 20 40 60 80 100 120 140

Instrument

A/D

Units

Series1Linear (Series1)

median= 6026.50Stdev = 808.54Average = 6054.11Stdev/ average = 0.13

Almost 2-fold difference in signal from over 100 instruments from the factory floor

12

HID Validation: The 3500 Series Genetic Analyzers

Normalization: What is it?

• This feature normailzes signal variations associated with instrument, capillary array, sample salt load, and injection variability among capillaries and instruments

• Normalization can be applied during primary analysis of the data

Before normalization After normalization

13

HID Validation: The 3500 Series Genetic Analyzers

• Normalization Target (NT)• The expected average RFU for eleven peaks in the

GeneScan 600 LIZ® v2 size standard used for normalization

• Normalization Factor• Normalization Target divided by the Average peak height of

the subset of peaks in the GeneScan 600 LIZ® v2 size standard used for normalization

• Normalization Factor Threshold• The passing range for Normalization Factor (default range is

0.3 to 3.0)

Normalization: Definitions

14

HID Validation: The 3500 Series Genetic Analyzers

Sources of Variation by Normalization Method

No NormFactory only

Int Std onlyFactor+IntStd

Injection to Injection

Cap to Cap

Within

Instrument to Instrument

Total

39%

33%

25%

14%

31%

24% 24%

13%

20%19%

5%5%

12%11%

4%4%3%

3%

0%0%

0%

5%

10%

15%

20%

25%

30%

35%

40%

%CV

Levels of Normalization

Sources of Variation

• Instrument to Instrument largest source of variation

• Internal standard Normalization greatest reduction in variation

Sensitivity and GeneScan™ 600 LIZ® Size Standard v2.0

16

HID Validation: The 3500 Series Genetic Analyzers

Sensitivity Study:Average Peak Height across a Range of DNA Input Amounts AmpFℓSTR®

MiniFilerTM PCR Amplification Kit

Normalization MethodFactory OnlyFactory + Internal Std

0 750 50 250 125

H1126

H1152

H1161 OO7

10000

5000

15000

00.125ng 0.25ng 0.50ng 0.75ng

10000

5000

15000

0

0.125ng 0.25ng 0.50ng 0.75ngGenomic DNA #2

Control DNA Standard 007

Genomic DNA #1

Genomic DNA #3

Peak

Hei

ght

17

HID Validation: The 3500 Series Genetic Analyzers

Sensitivity Study:Average Peak Height by Dye for 1ng DNA Input AmpF ℓSTR® Identifiler®

PCR Amplification Kit

2000

4000

8000

10000

6000

12000

14000

Internal Standard Normalization

20002000

Disabled Enabled

Peak

Hei

ght

18

HID Validation: The 3500 Series Genetic Analyzers

Size Standard Peak Height Consistency Study:Five AmpFℓSTR® PCR Amplification KitsGeneScan™ 600 LIZ® Size Standard v2.0

6000

5000

4000

3000

2000

1000

0

Lot 1 Lot 2 Lot 3

GS600 Peak Height

Identifiler®Identifiler® DirectIdentifiler® PlusMinifilerTM

NGMTM

7000

GS600 Lot Number

Peak

Hei

ght

19

HID Validation: The 3500 Series Genetic Analyzers

Size Standard Peak Height Consistency Study:%CV for 5 Different AmpFℓSTR® KitsGeneScan™ 600 LIZ® Size Standard v2.0

% CV

Size Standard Peak Height Consistency · 6 instruments · 3 lots of size standard

20

HID Validation: The 3500 Series Genetic Analyzers

Internal Standard Normalization: Injection to Injection Peak Height ConsistencyGeneScan™ 600 LIZ® Size Standard v2.0

% CV

Normalization MethodFactory OnlyFactory + Internal Std

40

30

20

10

0Lot 1 Lot 2 Lot 3

Size Standard Peak Height Consistency · 6 instruments · 3 lots of size standard

Resolution and Baseline Evaluation

22

HID Validation: The 3500 Series Genetic Analyzers

Resolution

• SEfiler Plus™ (SE33 triplet) and Identifiler® (TH01 9.3/10) Allelic Ladders

• 160 injections per instrument on 6 instruments

• Research STR markers greater than 300 base pairs were also developed with single base pair differences (data not shown)

23

HID Validation: The 3500 Series Genetic Analyzers

Baseline Noise - Developmental Evaluation G5 Average Noise (10x SD)

0

20

40

60

80

100

120

140

160

180

200

FAM VIC NED PET LIZ TOTAL

10x

Ave

rage

Std

. Dev

.

Data generated during the Sensitivity experiment was then used to confirm the level of pull up and the peak amplitude threshold setting.

175 RFU

24

HID Validation: The 3500 Series Genetic Analyzers

Pull-up360 Samples at 6 ng template

5.0%

20000

4000

8000

12000

16000

24000

Additional Studies and Validation Documentation

26

HID Validation: The 3500 Series Genetic Analyzers

Identifiler® Direct

• 82 gDNA, 25 cycles• Three 3500xl and three 3500 instruments.• Mother plate split into 6 daughter plates.

8000

8000

8 Capillary 3500

24 Capillary 3500xL

14000

14000

0

0

27

HID Validation: The 3500 Series Genetic Analyzers

Additional Studies

• Genotype Concordance• 100% Concordant with 3130xL

• GeneMapper® ID-X v1.2 Evaluation• Normalization and RFID documentation functionality performed as

expected• Average Mx values obtained in the Mixture Study were similar to

the 3130xl

28

HID Validation: The 3500 Series Genetic Analyzers

Validation Documentation

• AAFS and ISHI Posters

• Bode Conference Presentations

• User Bulletin containing developmental validation in progress

• www.appliedbiosystems.com/3500hid

29

HID Validation: The 3500 Series Genetic Analyzers

FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

©2009 Life Technologies Corporation. All rights reserved. The trademarks mentioned herein are the property of Life Technologies Corporation or their respective owners.

For those who require IVD-marked devices, the 3500 Dx and 3500xL Dx Genetic Analyzers and system accessories meet the requirements of the In Vitro Diagnostic Medical Devices Directive (98/79/EC). The 3500 Dx and 3500xL Dx systems are for distribution and use in specific European countries only. For more information about the 3500 Dx Series Systems, contact yourApplied Biosystems representative.

Legal

Thank you!

www.appliedbiosystems.com/3500HID

top related