Sr. No Evaluation Parameters M/s DWP M/s MediquipsM/s Shirazi
TradingM/s Siemens
01 Complete Package/Tender (Mandatory) Yes Yes Yes No
Technical Evaluation
Dated: 05-10-2017Package No/Tender Number: PC/115677/01/CH&ICH
Name of the Equipment and Qty:- Echo Cardiography as Package
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
(To be evaluated by Purchase Department)
01 Complete Package/Tender (Mandatory) Yes Yes Yes No
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes No
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Non Responsive
Sr. # Evaluation Parameters M/s DWP M/s MediquipsM/s Shirazi
TradingM/s Siemens
01Exclusive Authorization / Sole Agent Certificate
by the ManufacturerYes Yes Yes Yes
Technical & Engineering capability (As defined for
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
02Technical & Engineering capability (As defined for
the specific tender in specifications)Yes Yes Yes Yes
03
Certificate from the Manufacturer about the
after sales services through agent or itself (In
casespecifically demanded in the specifications)
Yes Yes Yes Yes
04 Bidder Past performance Yes Yes Yes Yes
05Availability of relevant Tools and Testing/
Calibration EquipmentYes Yes Yes Yes
06 Warranty Yes Yes Yes No
Remarks: Responsive Responsive Responsive Non Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/01/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.Brand Philips Toshiba GE
ModelAffinity 70
EPIQ-7Aplio - 400
Aplio - I 900
Vivid S 60 N
Vivid E 95
Netherland Japan USA
USA (Affinity 70
EPIQ-7)
Japan/Malaysia
Aplio - 400
Aplio - I 900
Norway
(Vivid S 60 N
Vivid E 95)
Compliance Compliance Compliance
Specification Compliance features wise: Accepted Accepted Accepted
Specifications: Accepted Accepted Accepted
PART - III
Specification Compliance / Evaluation Parameters
1Echo Cardiography as Package
Country of Manufacturer
Compliance with defined quality standards
Country of Origin of Product
/Model Number
Specifications: Accepted Accepted Accepted
Accepted Accepted Accepted
Accepted Accepted Accepted
Responsive Responsive Responsive
Siemens Acuson S 1000 Acuson SC 2000
USA USA, Japan, RSK & Netherlands
(Acuson S 1000 Acuson SC 2000) Compliance Not Accepted Warranty offered is not
according to the required tender specifications Not Accepted Not Accepted
Non Responsive
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Engr. Nazir Zaidi Electromedical Engineer Mayo Hospital, Lahore
Dr. Asif ur Rehman Associate Professor Cardiology
PIC, Lahore
Dr. Zubair Akram Prof. of Cardiology/ HOD
Allama Iqbal Medical College Jinnah Hospital, Lahore
Sr. No Evaluation Parameters M/s DWP M/s Shirazi Trading
01 Complete Package/Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/14/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Portable Echo Cardiography
KNOCK DOWN CRITERIA - (BID EVALUATION)
Part - I
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Responsive
Sr. # Evaluation Parameters M/s DWP M/s Shirazi Trading
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
03Certificate from the Manufacturer about the after sales services through
agent or itself (In casespecifically demanded in the specifications)Yes Yes
04 Bidder Past performance Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes
06 Warranty Yes Yes
Remarks: Responsive Responsive
PART - III
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
(All evaluation parmeters defined below are mandatory for compliance.)
Item Sr. No.
M/s DWP M/s Shirazi Trading
Brand Philips GE
Model CX 50 Vivid IQ
Netherland USA
Singapore (CX 50) China (Vivid IQ)
Compliance Compliance
Specification Compliance features wise: Compliance Compliance
Specifications: Accepted Accepted
Accepted Compliance
Accepted Accepted
Responsive Responsive
Package No/Tender Number: PC/115677/14/CH&ICH Dated: 05-10-2017
Technical Eligibility of Firm:
14Portable Echo
Machine
Technical Status:
Country of Origin of Product/Model Number
Specification Compliance / Evaluation Parameters
Country of Manufacturer
Compliance with defined quality standards
Technical Eligibility of Product:
Responsive ResponsiveTechnical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead
Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Engr. Nazir Zaidi Electromedical Engineer Mayo Hospital, Lahore
Dr. Asif ur Rehman Associate Professor Cardiology
PIC, Lahore
Dr. Zubair Akram Prof. of Cardiology/ HOD
Allama Iqbal Medical College Jinnah Hospital, Lahore
Sr. No Evaluation Parameters M/s Shirazi Trading M/s Mediquips
01 Complete Package/Tender (Mandatory) Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/24/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- SonoSite Portable Ultrasound
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
(To be evaluated by Purchase Department)
01 Complete Package/Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Responsive
Sr. # Evaluation Parameters M/s Shirazi Trading M/s Mediquips
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02Technical & Engineering capability (As defined for the specific tender
in specifications)Yes Yes
03
Certificate from the Manufacturer about the after sales services
through agent or itself (In casespecifically demanded in the
specifications)
Yes Yes
04 Bidder Past performance Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes
06 Warranty Yes Yes
Remarks: Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATIONPackage No/Tender Number: PC/115677/24/CH&ICH Dated: 05-10-2017(All evaluation parmeters defined below are mandatory for compliance.)
Item Sr. No.
M/s Shirazi Trading M/s Mediquips
Brand GE Konica Minolta
ModelVenue 50
Sonosite Portable
Ultrasound Mahchine
SONIMAGE HS-1
USA Japan
China
(Venue 50
Sonosite Portable Ultrasound
Mahchine)
Japan (SONIMAGE HS-1)
Compliance Compliance
Compliance Compliance
Country of Origin of Product/Model Number
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
24 Sonosite Portable Ultrasound Machine
Compliance with defined quality standards
Specification Compliance features wise:
Compliance Compliance
Accepted Accepted
Accepted Accepted
Responsive Responsive
Specification
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Part - I
Sr. No Evaluation Parameters M/s Hospicare M/s Amtronech M/s Ozawa Traders
01 Complete Package/Tender (Mandatory) No Yes No
Technical Evaluation
Package No/Tender Number: PC/115677/02/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Holter Monitor with Event Recorder
KNOCK DOWN CRITERIA - (BID EVALUATION)
(To be evaluated by Purchase Department)
01 Complete Package/Tender (Mandatory) No Yes No
02 Original Receipt of Tender (Mandatory) Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes
Remarks: Non Responsive Responsive Non Responsive
Sr. # Evaluation Parameters M/s Hospicare M/s Amtronech M/s Shirazi Trading
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes
02Technical & Engineering capability (As defined for the specific tender
Yes Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
02Technical & Engineering capability (As defined for the specific tender
in specifications)Yes Yes Yes
03
Certificate from the Manufacturer about the after sales services
through agent or itself (In casespecifically demanded in the
specifications)
Yes Yes Yes
04 Bidder Past performance Yes Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes Yes
06 Warranty Yes Yes Yes
Remarks: Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATIONPackage No/Tender Number: PC/115677/02/CH&ICH Dated: 05-10-2017(All evaluation parmeters defined below are mandatory for compliance.)
Item Sr. No.
M/s Amtronech M/s Shirazi Trading
Brand Mortara BTL Industries
Model DR 200/Ea BTL Cardio-Point- Holter-08-H300
USA UK
USA
(DR 200/Ea)Bulghria
(BTL Cardio-Point- Holter-08-H300)
Compliance Compliance
Specification Compliance features wise:
Compliance 1- Event loop recorder (Qty 03) as
mentioned in bid
not quoted
Compliance Not accepted
Compliance with defined quality standards
Specifications:
Country of Origin of Product/Model Number
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
2Holter Monitor
with event recorder
Compliance Not acceptedCompliance Not acceptedCompliance Not accepted
M/s Hospicare Cardioline
Walk 400H
Itlay Itlay
(Walk 400H)
Compliance 1- Event loop recorder (Qty 03)
as mentioned in bid not quoted NNoott aacccceepptteedd Not accepted Not accepted
Non Responsive Responsive Non Responsive
Specifications:Technical Eligibility of Product:Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular
Surgery
Engr. Nazir Zaidi Electromedical Engineer Mayo Hospital, Lahore
Dr. Asif ur Rehman Associate Professor Cardiology
PIC, Lahore
Dr. Zubair Akram Prof. of Cardiology/ HOD
Allama Iqbal Medical College Jinnah Hospital, Lahore
Part - I
Sr. No Evaluation ParametersM/s Total
TechnologyM/s Dymedix M/s Medi Urge
M/s Noor
International
01 Complete Package/Tender (Mandatory) Yes No Yes No
Technical Evaluation
Package No/Tender Number: PC/115677/03/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Fluid & Blood Warmer
KNOCK DOWN CRITERIA - (BID EVALUATION)
(To be evaluated by Purchase Department)
01 Complete Package/Tender (Mandatory) Yes No Yes No
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Remarks: Responsive Non Responsive Responsive Responsive
Sr. # Evaluation ParametersM/s Total
TechnologyM/s Dymedix M/s Medi Urge
M/s Noor
International
01Exclusive Authorization / Sole Agent Certificate
Yes Yes Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
01Exclusive Authorization / Sole Agent Certificate
by the ManufacturerYes Yes Yes Yes
02Technical & Engineering capability (As defined for
the specific tender in specifications)Yes Yes Yes Yes
03
Certificate from the Manufacturer about the
after sales services through agent or itself (In
casespecifically demanded in the specifications)
Yes Yes Yes Yes
04 Bidder Past performance Yes Yes Yes Yes
05Availability of relevant Tools and Testing/
Calibration EquipmentYes Yes Yes Yes
06 Warranty Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/03/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Total
TechnologyM/s Dymedix M/s Medi Urge
M/s Noor
International
BrandBelmont Instrumen
Co.Enthermics Stihler Electronic The Surgical Company
Model Buddy 2 IV NOW1 Astoflo Plus ECO Fluido Compact
USA Not mentioned Germany Netherland
USA
(Buddy 2)
USA
(IV NOW1)
Germany (Astoflo
Plus ECO)
Netherland (Fluido
Compact)Compliance Compliance Compliance Compliance
Specification Compliance features wise: Non compliance Non compliance Compliance Non compliance
Fluid & Blood
Warmer
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
Online system is not
Additional cost of
disposable
Additional cost of
disposable
Country of Origin of Product/Model Number
Compliance with defined quality standards
3
Not accepted Not accepted Compliance Not acceptedNot accepted Not accepted Compliance Not accepted
Non Responsive Non Responsive Responsive Non Responsive
Technical Eligibility of Product:Technical Eligibility of Firm:
Technical Status:
Online system is not
Quoted
Under Specs
disposable
set of cassette
and patient IV line
and as per End User
Requirement
disposable
set of cassette
and patient IV line per
End User
Requirement
Compliance Specifications:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Part - I
Sr. No Evaluation Parameters M/s Vertex Medical M/s Shirazi Trading
Technical Evaluation
Package No/Tender Number: PC/115677/06/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Anesthesia Workstation
KNOCK DOWN CRITERIA - (BID EVALUATION)
(To be evaluated by Purchase Department)
01 Complete Package/Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Responsive
Sr. # Evaluation Parameters M/s Vertex Medical M/s Shirazi Trading
01Exclusive Authorization / Sole Agent Certificate by the
ManufacturerYes Yes
02Technical & Engineering capability (As defined for the specific
Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
02Technical & Engineering capability (As defined for the specific
tender in specifications)Yes Yes
03
Certificate from the Manufacturer about the after sales services
through agent or itself (In casespecifically demanded in the
specifications)
Yes Yes
04 Bidder Past performance Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes
06 Warranty Yes Yes
Remarks: Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/06/CH&ICH Dated: 05-10-2017
Evaluated by Technical Evaluation Committee
Item
Sr. No.M/s Vertex Medical M/s Shirazi Trading
Brand Drager Werk GE
Model Primus+IACS C700 Avance CS2+B850Germany USA
Germany/USA(Primus) USA/Europe/Mexico(Avance CS2+B850)Compliance Compliance
Specification Compliance features wise: Compliance Compliance Accepted AcceptedAccepted AcceptedAccepted Accepted
Responsive Responsive
Specifications:Technical Eligibility of Product:Technical Eligibility of Firm:
Technical Status:
Compliance with defined quality standards Country of Origin of Product/Model Number
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
6Anesthesia
Workstation
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Part - I
Sr. No Evaluation ParametersM/s Hospitcare
Systems
M/s Human
Health Care
M/s Noor
International
M/s Biotech
Services
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/27/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Portable ECG
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Department
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
Sr. # Evaluation ParametersM/s Hospitcare
Systems
M/s Human
Health Care
M/s Noor
International
M/s Biotech
Services
01Exclusive Authorization / Sole Agent Certificate
by the ManufacturerYes Yes Yes Yes
02Technical & Engineering capability (As defined for
Yes Yes Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
02Technical & Engineering capability (As defined for
the specific tender in specifications)Yes Yes Yes Yes
03
Certificate from the Manufacturer about the
after sales services through agent or itself (In
casespecifically demanded in the specifications)
Yes Yes Yes Yes
04 Bidder Past performance Yes Yes Yes Yes
05Availability of relevant Tools and Testing/
Calibration EquipmentYes Yes Yes Yes
06 Warranty Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/27/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Hospitcare
Systems
M/s Human Health
Care
M/s Noor
InternationalBrand Cardioline Kalamed Nihon Kohden
Model ECG100+ KES-301 CardiofaxCItlay Germany Japan
Itlay(ECG100+) Germany (KES-301) Japan (CardiofaxC)Compliance Compliance Compliance
Specification Compliance features wise: Compliance Compliance Compliance
Specifications: Accepted Accepted Accepted
Compliance with defined quality standards
Country of Origin of Product/Model Number
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
27 Portable ECG
Specifications: Accepted Accepted Accepted
Accepted Accepted Accepted
Accepted Accepted Accepted
Responsive Responsive Responsive
M/s Biotech Services
Suzuken KenzCardico 306 Japan Japan (Kenz306 ) Compliance Compliance AAcccceepptteedd Accepted Accepted
Responsive
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Part - I
Sr. No Evaluation Parameters M/s Endo Kare M/s Almed Solutions
Technical Evaluation
Package No/Tender Number: PC/115677/12/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Fiber Optic Bronchoscope
KNOCK DOWN CRITERIA - (BID EVALUATION)
(To be evaluated by Purchase Department)
01 Complete Package/Tender (Mandatory) No Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Non Responsive Responsive
Sr. # Evaluation Parameters M/s Endo Kare M/s Almed Solutions
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
02 Technical & Engineering capability (As defined for the specific tender in specifications) Yes Yes
03Certificate from the Manufacturer about the after sales services through agent or
itself (In casespecifically demanded in the specifications)Yes Yes
04 Bidder Past performance Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes
06 Warranty Yes Yes
Remarks: Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/12/CH&ICH Dated: 05-10-2017
Evaluated by Technical Committee
Item
Sr. No.M/s Endo Kare M/s Almed Solutions
Brand Olympus Karl StorzModel (LF-GP/TP/DP) C-Mac
Japan GermanyJapan & Germany (LF-GP/TP/DP) Japan (C-Mac)
Compliance Compliance Specification Compliance features wise: Non compliance Compliance
Incomplete offer ComplianceNot accepted AcceptedNot accepted Accepted
Non Responsive Responsive
Technical Eligibility of Product:Technical Eligibility of Firm:
Technical Status:
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer Country of Origin of Product/Model Number
Compliance with defined quality standards
Specifications:
12F. Optic
Bronchoscope
Non Responsive ResponsiveTechnical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Part - I
Sr. No Evaluation Parameters M/s Dymedix System M/s Human Health Care
01 Complete Package/Tender (Mandatory) Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/15/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- BP Appratus
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
01 Complete Package/Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Responsive
Sr. # Evaluation Parameters M/s Dymedix System M/s Human Health Care
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
Technical & Engineering capability (As defined for the specific tender in
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Yes
03Certificate from the Manufacturer about the after sales services through
agent or itself (In casespecifically demanded in the specifications)Yes Yes
04 Bidder Past performance Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes
06 Warranty Yes Yes
Remarks: Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/15/CH&ICH Dated: 05-10-2017
Evaluated by Technical Committee
Item
Sr. No.
M/s Dymedix System M/s Human Health Care
Brand W.A Baun Co.
Model 250USA
USA (0250)Compliance
Specification Compliance features wise: Accepted
Accepted
Accepted
Accepted
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
Country of Origin of Product/Model Number
Compliance with defined quality standards
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
15 BP Apparatus
Accepted
Responsive
Yamasu Yamasu
Japan Japan Compliance Accepted Accepted Accepted
Accepted
ResponsiveTechnical Status:
Technical Eligibility of Firm:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Part - I
Sr. No Evaluation Parameters M/s Amtronech M/s Hospicare
Systems
M/s Reliance
Medical
01 Complete Package/Tender (Mandatory) No Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/04/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Near IR Spectroscopy
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
01 Complete Package/Tender (Mandatory) No Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes
Remarks: Non Responsive Responsive Responsive
Sr. # Evaluation Parameters M/s Amtronech M/s Hospicare
Systems
M/s Reliance
Medical
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes
Technical & Engineering capability (As defined for the specific tender in
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
Evaluated by Technical Committee
PART - II
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Yes Yes
03Certificate from the Manufacturer about the after sales services
through agent or itself (In casespecifically demanded in the
specifications)
Yes Yes Yes
04 Bidder Past performance Yes Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes Yes
06 Warranty Yes Yes Yes
Remarks: Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/04/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.
M/s Hospicare Systems M/s Reliance Medical
Brand Medtronic Cobidien Casemed
ModelINOVOS Cerebral
OximeterFore-Sight Elite
USA USA
USA (INOVOS)
Compliance
Specification Compliance features wise: Compliance
Accepted
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
Country of Origin of Product/Model Number
Compliance with defined quality standards
Specifications:
4Near Infrared
Spectroscopy
Accepted
Accepted
Accepted
M/s Amtronech
Masimo
Masimo Root Systems
Maxico USA & Maxico (Masimo Root Systems)
Compliance Non Compliance
SSeennssoorrss ffoorr neonates aarree nnoott available Not Accepted Accepted
Non Responsive Responsive
USA (Fore-Sight Elite)
Compliance Compliance
Accepted
Accepted Accepted
Responsive
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Specifications:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Sr. # Evaluation ParametersM/s
Forecare
M/s Eastern
Medical
M/s Vertex
Medical
M/s Sigma
InternationalM/s Meditech
M/s Human
Healthcare
01 Complete Package/Tender (Mandatory) Yes Yes No Yes Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/17/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Suction Machine (Portable)
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Technical Committee
01 Complete Package/Tender (Mandatory) Yes Yes No Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes Yes Yes
Remarks: Responsive Responsive Non Responsive Responsive Responsive Responsive
Sr. # Evaluation ParametersM/s
Forecare
M/s Eastern
Medical
M/s Vertex
Medical
M/s Sigma
InternationalM/s Meditech
M/s Human
Healthcare
01Exclusive Authorization / Sole Agent Certificate
by the ManufacturerYes Yes Yes Yes Yes Yes
02Technical & Engineering capability (As defined for
Yes Yes Yes Yes Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
02Technical & Engineering capability (As defined for
the specific tender in specifications)Yes Yes Yes Yes Yes Yes
03
Certificate from the Manufacturer about the
after sales services through agent or itself (In
casespecifically demanded in the specifications)
Yes Yes Yes Yes Yes Yes
04 Bidder Past performance Yes Yes Yes Yes Unsatisfactory Yes
05Availability of relevant Tools and Testing/
Calibration EquipmentYes Yes Yes Yes Yes Yes
06 Warranty Yes Yes Yes Yes Yes Yes
Remarks: Resposive Resposive Responsive Resposive Non Resposive Resposive
KNOCK DOWN CRITERIA - PRODUCT EVALUATIONPackage No/Tender Number: PC/115677/17/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.
M/s Forecare M/s Eastern Medical M/s Vertex Medical M/s Sigma International M/s MeditechM/s Human
Healthcare
Brand Medist Cheiron Medela Italian Medical ARDO SIEM - Nova
Model ME VACS M - 90 Victoria Basic Suction Pump New Hospivack 400 Ardo Suction Pump 611-A1 Max
Slovakia Czech Republic Switzerland Itlay UK Itlay
Slovakia
(ME VACS M - 90)
Czech
Republic(Victoria)
Switzerland
(Basic Suction Pump)Itlay (New Hospivack 400)
UK/Switzerland
(Ardo Suction Pump)Itlay (611-A1 Max)
Compliance Compliance Compliance Compliance Compliance Compliance
Specification Compliance features wise: Accepted Accepted Not accepted Accepted Accepted Accepted
Accepted Accepted
Country of Origin of Product
/Model Number
Compliance with defined quality standards
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
1Suction
Machine(Portable)
specs
Accepted Accepted
Accepted Accepted Not accepted Not accepted Accepted Accepted
Accepted Accepted Not accepted Accepted
Not Accepted
(On previous
performance based)
Accepted
Responsive Responsive Non Responsive Non Responsive Non Responsive Responsive
Technical Eligibility of Firm:
Technical Status:
Technical Eligibility of Product:
Plastic body, not reliable
Sppecifications: Accepted Acceppted
to operate on hhiighh
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Qty & Specs not as per requirement, I.e. 12 Nos. & 50 L/M
load
Part - I
Sr. No Evaluation Parameters M/s Hospicare M/s AmtronechM/s Bio Tech
Services
M/s Al Kareem
Medical Tech
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/16/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Pulse Oximeter
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
Sr. # Evaluation Parameters M/s Hospicare M/s AmtronechM/s Bio Tech
Services
M/s Al Kareem
Medical Tech
01Exclusive Authorization / Sole Agent Certificate
Yes Yes Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Technical Committee
Evaluted by Technical Committee
PART - II
01Exclusive Authorization / Sole Agent Certificate
by the ManufacturerYes Yes Yes Yes
02Technical & Engineering capability (As defined for
the specific tender in specifications)Yes Yes Yes Yes
03
Certificate from the Manufacturer about the
after sales services through agent or itself (In
casespecifically demanded in the specifications)
Yes Yes Yes Yes
04 Bidder Past performance Yes Yes Yes Yes
05Availability of relevant Tools and Testing/
Calibration EquipmentYes Yes Yes Yes
06 Warranty Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/16/CH&ICH Dated: 05-10-2017
Evaluted by Technical Committee
Item
Sr. No.
M/s HospicareM/s
AmtronechM/s Bio Tech Services
Brand Medtronic Masimo Medlab
Model Nellcor PM100N Rad-97 Vitromap
USA USA Germany
USA (Nellcor
PM100N)USA (Rad-97) Germany (Vitromap)
Compliance Compliance Compliance
Compliance Compliance Compliance
Accepted Accepted Accepted
M/s Al Kareem Medical Tech Andromeda VT-900 Series USA USA, China & Germany
(VT-900 Series) Compliance Compliance Accepted AAcccceepptteedd Accepted
Responsive
Accepted Accepted Accepted
Country of Origin of Product/Model Number
Specification Compliance features wise:
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
16 Pulse Oximeter
Technical Eligibility of Product:
Compliance with defined quality standards
Specifications:
Accepted Accepted Accepted
Accepted Accepted Accepted
Responsive Responsive Responsive
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Responsive
Part - I
Sr. # Evaluation ParametersM/s Hoora
Pharma
M/s Meditec
Instruments
M/s Sigma
International
M/s Global
Marketing
Services
M/s AM Sales &
ServicesM/s Roche
M/s S Ejazuddin &
Co
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes Yes Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/21/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- ABG Machine
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Department
03 Affidavit from Bidder Yes Yes Yes Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive Responsive Responsive Responsive
Sr. # Evaluation ParametersM/s Hoora
Pharma
M/s Meditec
Instruments
M/s Sigma
International
M/s Global
Marketing
Services
M/s AM Sales &
ServicesM/s Roche
M/s S Ejazuddin &
Co
01Exclusive Authorization / Sole Agent Certificate
by the ManufacturerYes Yes Yes Yes Yes Yes Yes
02Technical & Engineering capability (As defined for
the specific tender in specifications)Yes Yes Yes Yes Yes Yes Yes
03
Certificate from the Manufacturer about the
after sales services through agent or itself (In
casespecifically demanded in the specifications)
Yes Yes Yes Yes Yes Yes Yes
04 Bidder Past performance Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory)
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
Evaluated by Technical Committee
Part-II
05Availability of relevant Tools and Testing/
Calibration EquipmentYes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory) Yes (Satisfactory)
06 Warranty Yes Yes Yes Yes Yes No Yes
Remarks: Responsive Responsive Responsive Responsive Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/21/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.
M/s Hoora PharmaM/s Meditec
Instruments
M/s Sigma
International
M/s Global
Marketing
Services
M/s AM Sales &
ServicesM/s Roche
M/s S Ejazuddin
& Co
Brand SIEMENS Medica Co. Convergent OPTI Medical Nova Biomedical Cobas Radio Medical
Model Rapid Lab 348Easy State Blood
Gas AnalyserLiquical OPTI CCA - TS2
State Profile pHOx
PlusCobas B 221
ABL 80 Flex
Germany USA Germany UK USA Switzerland Denmark
USA (Rapid Lab
348)
USA (Easy State Blood
Gas Analyser)
Germany
(Liquical)
UK (OPTI CCA -
TS2)
USA (State Profile
pHOx Plus)EEC/Japan/USA
(Cobas B 221)
Denmark (ABL 80
Flex)
Compliance Compliance Complaince Complaince Complaince Complaince Complaince
Specification Compliance features wise: Accpted Accpted Accpted Accpted Accpted Accpted Accpted
Accpted Accpted Accpted Accpted Accpted Accpted Accpted
Accpted Accpted Accpted Accpted Accpted Accpted Accpted
Accpted Accpted Accpted Accpted Accpted Accpted Accpted
Responsive Responsive Responsive Responsive Responsive Responsive Responsive
Country of Origin of Product
/Model Number
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
21 ABG Machine
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Compliance with defined quality standards
Specifications:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Part - I
Sr. No Evaluation Parameters M/s Claris MedicalM/s Alliance
Medical
01 Complete Package/Tender (Mandatory) Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/08/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Dual Chamber Pacemaker
KNOCK DOWN CRITERIA - (BID EVALUATION)
(To be evaluated by Purchase Department)
01 Complete Package/Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Responsive
Sr. No Evaluation Parameters M/s Claris MedicalM/s Alliance
Medical01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Yes
Certificate from the Manufacturer about the after sales services through agent or
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
03Certificate from the Manufacturer about the after sales services through agent or
itself (In casespecifically demanded in the specifications)Yes Yes
04 Bidder Past performance Yes (Satisfactory) Yes (Satisfactory)
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes (Satisfactory) Yes (Satisfactory)
06 Warranty Yes Yes
Remarks: Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/08/CH&ICH Dated: 05-10-2017
Evaluated by Technical Committee
Item
Sr. No.M/s Claris Medical M/s Alliance Medical
Brand Osypka Medical Osypka AG
Model PACE 203 PACE 203 H
USA, Germany Germany
USA, Germany (PACE 203) Germany (PACE 203 H)
Compliance Compliance
Specification Compliance features wise: Compliance Compliance
Accepted Accepted
Accepted Accepted
Accepted Accepted
Country of Origin of Product/Model Number
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
8Dual Chamber
Pacemaker
Technical Eligibility of Product:
Technical Eligibility of Firm:
Compliance with defined quality standards
Specifications:
Responsive ResponsiveTechnical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Part - I
Sr. NoEvaluation ParametersM/s Crocus
Scientific Trading M/s MediTech M/s Medi Serve
M/s Iqbal &
Company
M/s Medi
UrgeM/s Bbraun
01 Complete Package/Tender (Mandatory) Yes No No Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes Yes Yes
Technical Evaluation
Package No/Tender Number: PC/115677/09/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Infusion Pumps
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes Yes Yes
Remarks: Responsive Non Responsive Non Responsive Responsive Responsive Responsive
PART - II
Sr. # Evaluation ParametersM/s Crocus
Scientific Trading M/s MediTech M/s Medi Serve
M/s Iqbal &
Company
M/s Medi
UrgeM/s Bbraun
01Exclusive Authorization / Sole Agent Certificate by
the ManufacturerYes Yes Yes Yes Yes Yes
02Technical & Engineering capability (As defined for
the specific tender in specifications)Yes
Yes but not
enoughYes Yes Yes Yes
03
Certificate from the Manufacturer about the after
sales services through agent or itself (In Yes
Yes but not
satisfactory Yes Yes Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
(To be evaluated by Technical Evaluation Committee)
Evaluated by Technical Committee
03 sales services through agent or itself (In
casespecifically demanded in the specifications)
Yes satisfactory
performance
Yes Yes Yes Yes
04 Bidder Past performance Doubtful Poor Yes (Satisfactory)Yes
(Satisfactory)
Yes
(Satisfactory)
Yes
(Satisfactory)
05Availability of relevant Tools and Testing/
Calibration EquipmentNot mentioned No Yes Yes Yes Yes
06 Warranty Yes Yes No Yes Yes Yes
Remarks: Non Responsive Non Responsive Non Responsive Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/011/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.
M/s Crocus Scientific
Trading M/s MediTech M/s Medi Serve
M/s Iqbal &
CompanyM/s Medi Urge M/s Bbraun
Brand SCO MED Codan Argus Smiths Medical Carefusion Fresenius Bbraun
Model IFP-1800
Argus 717 V
Wolumetric Infusion
Pump
Grasevy 1200
Infusion PumpAlaris - GW Agilia VP Infusomat P
Germany Switzerland Czech Republic Switzerland France Germany
Germany (IFP-1800)
Switzerland
(Argus 717 V
Wolumetric Infusion
Pump)
UK
(Grasevy 1200
Infusion Pump)
UK (Alaris - GW)France (Agilia
VP)
Germany
(Infusomat P)
Compliance Compliance Compliance Compliance Compliance Compliance
Specification Compliance features wise: Not Accepted Not Accepted Not Accepted Accepted Accepted Accepted
Country of Origin of Product/Model Number
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
9 Infusion Pump
Compliance with defined quality standards
Specification Compliance features wise: Not Accepted Not Accepted Not Accepted Accepted Accepted Accepted
Not Accepted Not Accepted Not Accepted Accepted Accepted Accepted
Not Accepted Not Accepted Not Accepted Accepted Accepted Accepted
Non Resposive Non Resposive Non Resposive Resposive Responsive Responsive
Specifications:
Price is mentioned in
technical bid
No past performance
record found
Accepted
Technical Eligibility of Product:
Technical Status:
1-Display parameters
not according to
requirement
2-Warranty one year
only
1-Very poor past
performance
2-Infusion pump
operates on only 5
brands IVs
Infusion set (not
Accepted Accepted
Technical Eligibility of Firm:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Package No/Tender Number: PC/115677/11/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Infant Radiant Warmer
Part - I
Sr. No Evaluation Parameters M/s Medi UrgeM/s Dymedix
Systems
M/s Med
Equips
M/s Alam
Medix
M/s Shirazi
Tading
Technical Evaluation
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive Responsive
Sr. # Evaluation Parameters M/s Medi Urge M/s Med
Equips
M/s Alam
Medix
M/s Shirazi
Trading
01Exclusive Authorization / Sole Agent Certificate
by the ManufacturerYes Yes Yes Yes
02Technical & Engineering capability (As defined for
the specific tender in specifications)Yes Yes Yes Yes
03
Certificate from the Manufacturer about the
after sales services through agent or itself (In
casespecifically demanded in the specifications)
Yes Yes Yes Yes
Yes Yes Yes
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Technical Committee
04 Bidder Past performanceYes
(Satisfactory)
Yes
(Satisfactory)
Yes
(Satisfactory)Yes (Satisfactory)
05Availability of relevant Tools and Testing/
Calibration Equipment
Yes
(Satisfactory)
Yes
(Satisfactory)
Yes
(Satisfactory)Yes (Satisfactory)
06 Warranty Yes
M/s Dymedix Systems
Yes
Satisfactory
Yes
Yes
Yes
(Satisfactory)
Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive Responsive
Package No/Tender Number: PC/115677/011/CH&ICH Dated: 05-10-2017
Item
Sr. No.
M/s Medi UrgeM/s Dymedix
Systems
M/s Med
Equips
M/s Alam
Medix
M/s Shirazi
TadingBrand Cobams SRL Shvabe Zurich Atom Medicor GE
ModelLR - 90 (COD
07A)SNO UOMZ Sun Flower VLR - 2101 Lullabay Warmer
Itlay Switzerland Japan Hungary USA
Itlay
(LR - 90 (COD
07A))
Switzerland,
Russia (SNO
UOMZ)
Japan
(Sun Flower)
Hungary
(VLR - 2101)
India
(Lullabay
Warmer)
Compliance Compliance Compliance Compliance Compliance
Evaluated by Technical Committee
PART - III
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Country of Origin of Product/Model Number
Specification Compliance / Evaluation Parameters
Country of Manufacturer
11Infant Radiant
Warmer
Compliance with defined quality standards
Specification Compliance features wise: Compliance Compliance Compliance Compliance Compliance
Accepted Accepted Accepted Accepted Accepted
Accepted Accepted Accepted Accepted Accepted
Responsive Responsive Responsive Responsive Responsive
AcceptedAccepted Accepted Accepted AcceptedTechnical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Specifications:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Package No/Tender Number: PC/115677/23/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Head Light
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation Parameters M/s Medi Urge M/s Dymedix Systems
01 Complete Package/Tender (Mandatory)
Technical Evaluation
01 Complete Package/Tender (Mandatory) Yes NO
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Non Responsive
Sr. # Evaluation Parameters M/s Medi Urge M/s Dymedix Systems
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Yes
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
02specifications)
Yes Yes
03Certificate from the Manufacturer about the after sales services through
agent or itself (In casespecifically demanded in the specifications)Yes Yes
04 Bidder Past performance Yes (Satisfactory) Average
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes (Satisfactory) Yes (Satisfactory)
06 Warranty Yes Yes
Remarks: Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/23/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Medi Urge M/s Dymedix Systems
Brand Integra Ecleris
Model Luxtec MLX 300 LUXUSA
Country of Origin of Product/Model Number USA & Canada (Luxtec MLX 300) USA (LUX)Compliance Compliance
Specification Compliance features wise: Acceptable Not Accepted
Acceptable Under specsAcceptable Not Accepted
Acceptable Not Accepted
Responsive Non Responsive
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Compliance with defined quality standards
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
23 Head Light
Responsive Non ResponsiveTechnical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Package No/Tender Number: PC/115677/22/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Non-Invasive Monitor
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation ParametersM/s Vertex
MedicalM/s Mediquips M/s Al-Kareem M/s Vital Care
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes
Technical Evaluation
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
Sr. # Evaluation ParametersM/s Vertex
Medical
M/s
Mediquips
M/s
Al-Kareem
M/s
Vital Care
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes Yes
02Technical & Engineering capability (As defined for the specific tender
in specifications)Yes Yes
Yes
(Satisfactory)Yes
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
03Certificate from the Manufacturer about the after sales services
through agent or itself (In casespecifically demanded in the
specifications)
Yes Yes Yes Yes
04 Bidder Past performanceYes
(Satisfactory)
Yes
(Satisfactory)
Yes
(Satisfactory)
Yes
(Satisfactory)
05 Availability of relevant Tools and Testing/ Calibration EquipmentYes
(Satisfactory)
Yes
(Satisfactory)
Yes
(Satisfactory)
Yes
(Satisfactory)
06 Warranty Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/22/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Vertex
MedicalM/s Mediquips M/s Al-Kareem M/s Vital Care
Brand Drager Nihon Kohden Veiele Heyer Medical AG
Model Vista 120 Life Scope SE 1500 VizOR-17Germany Japan Sweden Germany
Country of Origin of Product/Model Number China(Vista 120) Japan(Life Scope) Sweden(SE 1500) Germany(VizOR-17)
Compliance Compliance Compliance Compliance
Specification Compliance features wise: Acceptable Acceptable Acceptable Acceptable
Acceptable Acceptable Acceptable AcceptableAccepted Accepted Accepted Accepted
Accepted Accepted Accepted Accepted
Responsive Responsive Responsive Responsive
Technical Eligibility of Firm:
Technical Status:
Specification Compliance / Evaluation Parameters
22Non-Invasive
Monitor
PART - III
Country of Manufacturer
Compliance with defined quality standards
Specifications:Technical Eligibility of Product:
Responsive Responsive Responsive ResponsiveTechnical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Package No/Tender Number: PC/115677/20/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Defibrillator
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation ParametersM/s Human
HealthcareM/s Mediquips M/s Biotech M/s Sigma
01 Complete Package/Tender (Mandatory) No Yes Yes Yes
Technical Evaluation
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Remarks: Non Responsive Responsive Responsive Responsive
Sr. # Evaluation ParametersM/s Human
HealthcareM/s Mediquips M/s Biotech M/s Sigma
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes Yes
02Technical & Engineering capability (As defined for the specific tender
in specifications)Yes Yes Yes Yes
03Certificate from the Manufacturer about the after sales services
through agent or itself (In casespecifically demanded in the Yes Yes Yes Yes
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
(All Evaluation parameters defined below are mandatory for compliance.)
03 through agent or itself (In casespecifically demanded in the
specifications)
Yes Yes Yes Yes
04 Bidder Past performanceYes
(Satisfactory)
Yes
(Satisfactory)Yes (Satisfactory)
Yes
(Satisfactory)
05 Availability of relevant Tools and Testing/ Calibration EquipmentYes
(Satisfactory)
Yes
(Satisfactory)Yes (Satisfactory)
Yes
(Satisfactory)
06 Warranty Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/20/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Human
HealthcareM/s Mediquips M/s Biotech M/s Sigma
BrandUS Defib Medical
Technologies
Nihon Kohden
Cardio LifeInnomed Bexen Cardio
Model Cardiostart TEC-5631K Cardio aid 360-B Reanibex-800Japan Hungary Spain
Country of Origin of Product/Model NumberJapan /
TEC-5631K
Hungary /
Cardio aid 360-B
Spain /
Reanibex-800Compliance Compliance Compliance
Specification Compliance features wise: Acceptable Acceptable Acceptable
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
Compliance with defined quality standards
22 Defibrillator
Acceptable Acceptable Acceptable
Acceptable Acceptable Acceptable
Acceptable Acceptable Acceptable
USA
USA / Cardiostart
Compliance Non Compliance Not as per
requirement , 1- Local Trolley quoted 2-Technical data not attached)
Not Acceptable Acceptable Non Responsive Responsive Responsive Responsive
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Package No/Tender Number: PC/115677/07/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Ventilator
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation ParametersM/s Noor
Internation
M/s Eastern
Medical
M/s
Hospicare
M/s Shirazi
Trading
M/s Vertex
Medical
M/s Total
Technologies
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes Yes Yes
Technical Evaluation
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive Responsive Responsive
Sr. # Evaluation ParametersM/s Noor
Internation
M/s Eastern
MedicalM/s Hospicare
M/s Shirazi
Trading
M/s Vertex
Medical
M/s Total
Technologies
01Exclusive Authorization / Sole Agent Certificate by the
ManufacturerYes Yes Yes Yes Yes Yes
02Technical & Engineering capability (As defined for the
specific tender in specifications)Yes Yes Yes Yes Yes Yes
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
(All Evaluation parameters defined below are mandatory for compliance.)
02specific tender in specifications)
Yes Yes Yes Yes Yes Yes
03
Certificate from the Manufacturer about the after sales
services through agent or itself (In casespecifically
demanded in the specifications)
Yes Yes Yes Yes Yes Yes
04 Bidder Past performanceYes
(Staisfactory)
Yes
(Staisfactory)
Yes
(Staisfactory)
Yes
(Staisfactory)
Yes
(Staisfactory)
Yes
(Staisfactory)
05Availability of relevant Tools and Testing/ Calibration
Equipment
Yes
(Staisfactory)
Yes
(Staisfactory)
Yes
(Staisfactory)
Yes
(Staisfactory)
Yes
(Staisfactory)
Yes
(Staisfactory)
06 Warranty Yes Yes Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/07/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.
M/s Noor
Internation
M/s Eastern
Medical
M/s Shirazi
Trading
M/s Vertex
Medical
M/s Total
Technologies
Brand Hamilton E Vent Medical GE Drager WorkLowenstein
MedicalModel G-5 Inspiration 7i Care Scape R860 Avita V 300 Elise 600
Switzerland USA USA Germany Germany
Country of Origin of Product/Model NumberSwitzerland (G-
5)
USA
(Inspiration 7i)
USA(Care Scape
R860)
Germany
(Avita V 300)
Germany (Elise
600)
Compliance Compliance Compliance Compliance Compliance
Specification Compliance features wise: Accepted Accepted Accepted Accepted Accepted
Accepted
Internal battery
M/s Hospicare
Covedian
New Port E 360 USA
USA (New Port E 360)
Compliance Accepted No Buit in Nebulizer
Model of Nebulizer,
Accepted Accepted AcceptedSpecifications:
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
Compliance with defined quality standards
7 Ventilator
Accepted
requirement
backup not Air Compressor, according to Humidifier not
Accepted Accepted Accepted
Accepted Not Accepted Accepted Accepted Accepted
Accepted Accepted Accepted Accepted Accepted
Responsive Not Responsive
Quoted
Not Accepted Accepted
Not Responsive Responsive Responsive Responsive
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Package No/Tender Number: PC/115677/26/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Forced Air Warmer
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation Parameters M/s Dymedix Systems M/s Reliance Medical
Technical Evaluation
01 Complete Package/Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Responsive
Sr. # Evaluation Parameters M/s Dymedix Systems M/s Reliance Medical
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
Technical & Engineering capability (As defined for the specific tender in
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Yes
03Certificate from the Manufacturer about the after sales services through
agent or itself (In casespecifically demanded in the specifications)Yes Yes
04 Bidder Past performance Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes
06 Warranty Yes Yes
Remarks: Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/26/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Dymedix Systems M/s Reliance Medical
Brand Cocoon IOB Medical
Model CWS 4000 IOB Warming UnitAustrailia USA
Country of Origin of Product/Model Number USA(IOB Warming Unit)Compliance
Specification Compliance features wise: Accepted
Accepted Compliance
Compliance
Responsive
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
PART - III
Specification Compliance / Evaluation Parameters
Country of Manufacturer
Compliance with defined quality standards
26Forced Air
Warmer
Australia (CWS 4000)
Compliance Accepted Accepted Compliance Compliance
RResponsiive ResponsiveTechnical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Dr. Khalid Hameed Zaighum DMS (Purchase)
Dr. Usaid Qureshi Assistant Professor Pead Cardiology
Dr. Muhammad Asim Assistant Professor of Cardiovasular
Dr. Saeedah Asaf Associate Professor Anesthesia
Cardiovascular Surgery
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular Surgery
Package No/Tender Number: PC/115677/29/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Laser Printer (Colour)
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation Parameters M/s Vertex M/s Google Tech
01 Complete Package/Tender (Mandatory) Yes Yes
Technical Evaluation
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Responsive
Sr. # Evaluation Parameters M/s Vertex M/s Google Tech
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer No Yes
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Yes
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
03Certificate from the Manufacturer about the after sales services through
agent or itself (In casespecifically demanded in the specifications)No Yes
04 Bidder Past performance Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes
06 Warranty Yes Yes
Remarks: Non Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/29/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Vertex M/s Google Tech
Brand HP Hp
Model No Model Mentioned HP/M452NWUSA USA
Country of Origin of Product/Model Number (Not Mentioned) ChinaCompliance Compliance
Specification Compliance features wise: Not Acceptable Acceptable
Not Acceptable Acceptable
Acceptable Acceptable
Acceptable Acceptable
Non Responsive Responsive
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
PART - III
Specification Compliance / Evaluation Parameters
29 Laser Printer
Country of Manufacturer
Compliance with defined quality standards
Non Responsive ResponsiveTechnical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Mazhar Iqbal Noor Computer Programmer
Mr. Riaz Shahid
I.T Incharge
Dr. Khalid Hameed Zaighum
(DMS (Purchase)
Dr. Muhammad Asim
Assistant Professor of Cardiovasular
Dr. Usaid Qureshi Assistant Professor
Pead Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular
Surgery
Package No/Tender Number: PC/115677/30/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Computer System
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation Parameters M/s VertexM/s Generation
System
M/s Fatemi
ImpexM/s Google Tech.
01 Complete Package/Tender (Mandatory) Yes Yes Yes Yes
Technical Evaluation
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Remarks: Responsive Responsive Responsive Responsive
Sr. # Evaluation Parameters M/s VertexM/s Generation
System
M/s Fatemi
ImpexM/s Google Tech.
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes Yes
02Technical & Engineering capability (As defined for the specific tender
in specifications)Yes Yes Yes Yes
PART - IIKNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
(All Evaluation parameters defined below are mandatory for compliance.)
02in specifications)
Yes Yes Yes Yes
03Certificate from the Manufacturer about the after sales services
through agent or itself (In casespecifically demanded in the
specifications)
Yes Yes Yes Yes
04 Bidder Past performance Yes Yes Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes Yes Yes
06 Warranty Yes Not Mentioned Yes Yes
Remarks: Responsive Non Responsive Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/30/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.
M/s VertexM/s Generation
SystemM/s Fatemi Impex M/s Google Tech.
Brand HP Lenovo Lenovo LenovoModel HP/800G3 Lenovo/M910 Lenovo /M910 Lenovo /M910
USA USA USA USA
Country of Origin of Product/Model Number China(HP/800G3)China(Lenovo/M
910)
China(Lenovo
/M910)
China(Lenovo
/M910)Compliance No Compliance Compliance
Specification Compliance features wise: Compliance Not Acceptable Acceptable Acceptable
CompliancePrinter without
Networking QoutedAcceptable Acceptable
Acceptable Not Acceptable Acceptable Acceptable
Specifications:
Technical Eligibility of Product:
PART - III
Specification Compliance / Evaluation Parameters
30 Computer System
Country of Manufacturer
Compliance with defined quality standards
Acceptable Not Acceptable Acceptable Acceptable
Responsive Non Responsive Responsive Responsive
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Mazhar Iqbal Noor Computer Programmer
Mr. Riaz Shahid
I.T Incharge
Dr. Khalid Hameed Zaighum
(DMS (Purchase)
Dr. Muhammad Asim
Assistant Professor of Cardiovasular
Dr. Usaid Qureshi Assistant Professor
Pead Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular
Surgery
Package No/Tender Number: PC/115677/31/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- DSL LINE wireless
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation Parameters M/s Optcl
Technical Evaluation
01 Complete Package/Tender (Mandatory) Yes
02 Original Receipt of Tender (Mandatory) Yes
03 Affidavit from Bidder Yes
04 Bid Security (Mandatory) Yes
05 Bid Validity (Mandatory) Yes
06 Delivery Period (Mandatory) Yes
Remarks: Responsive
Sr. # Evaluation Parameters M/s Optcl
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
02 Technical & Engineering capability (As defined for the specific tender in specifications) Yes
03Certificate from the Manufacturer about the after sales services through agent or itself (In
casespecifically demanded in the specifications)Yes
04 Bidder Past performance Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes
06 Warranty Yes
Remarks: Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/31/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item Sr. No.
M/s Optcl
Brand CISCO
Model Linksys WAP300N
USA
Country of Origin of Product/Model Number USA
Compliance
Specification Compliance features wise: Acceptable
AcceptableAcceptable
Acceptable
Responsive
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
PART - III
Specification Compliance / Evaluation Parameters
31 DSL Line
Country of Manufacturer
Compliance with defined quality standards
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Mazhar Iqbal Noor Computer Programmer
Mr. Riaz Shahid
I.T Incharge
Dr. Khalid Hameed Zaighum
(DMS (Purchase)
Dr. Muhammad Asim
Assistant Professor of Cardiovasular
Dr. Usaid Qureshi Assistant Professor
Pead Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular
Surgery
Package No/Tender Number: PC/115677/32/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Video Conference
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation Parameters M/s Amtech M/s Commtel System
Technical Evaluation
01 Complete Package/Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Responsive
Sr. # Evaluation Parameters M/s Amtech M/s Commtel System
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
Technical & Engineering capability (As defined for the specific tender in
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Yes
03Certificate from the Manufacturer about the after sales services through
agent or itself (In casespecifically demanded in the specifications)Yes Yes
04 Bidder Past performance Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes
06 Warranty Yes Yes
Remarks: Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/32/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Amtech M/s Commtel System
Brand Polycom/Samsung Huawei/Sony
ModelPolycom RealPresence Group 500,
Samsung Smart LED 55" MU 7000Huawei TE 40, Sony KDL 55
Thialand China
Country of Origin of Product/Model Number USA ChinaCompliance Compliance
Specification Compliance features wise: Accepted Accepted
Not Accepted Specifications are not according to the AcceptedSpecifications:
PART - III
Specification Compliance / Evaluation Parameters
32 Video Conference
Country of Manufacturer
Compliance with defined quality standards
Specifications are not according to the
requirements
Accepted
Accepted Accepted
Accepted Accepted
Non Responsive Responsive
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Mazhar Iqbal Noor Computer Programmer
Mr. Riaz Shahid
I.T Incharge
Dr. Khalid Hameed Zaighum
(DMS (Purchase)
Dr. Muhammad Asim
Assistant Professor of Cardiovasular
Dr. Usaid Qureshi Assistant Professor
Pead Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular
Surgery
Package No/Tender Number: PC/115677/33/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Laptop Computer
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation ParametersM/s Google
TechM/s Fatemi Impex M/s Vertex
01 Complete Package/Tender (Mandatory) No No Yes
Technical Evaluation
02 Original Receipt of Tender (Mandatory) Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes
04 Bid Security (Mandatory) Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes
Remarks:Non
ResponsiveNon Responsive Responsive
Sr. # Evaluation ParametersM/s Google
TechM/s Fatemi Impex M/s Vertex
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes
02Technical & Engineering capability (As defined for the specific tender
in specifications)Yes Yes Yes
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
(All Evaluation parameters defined below are mandatory for compliance.)
03Certificate from the Manufacturer about the after sales services
through agent or itself (In casespecifically demanded in the
specifications)
Yes Yes Yes
04 Bidder Past performance Yes Yes Yes
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Yes Yes
06 Warranty Yes No Yes
Remarks: Responsive Non Responsive Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/33/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s Google Tech M/s Fatemi Impex M/s Vertex
Brand Lenovo HP HP
Model Lenovo IP320 Pro Book 450 G4Pro Book 450 G4 +
402 DNEUSA USA USA
Country of Origin of Product/Model Number China China China
Compliance Compliance
Specification Compliance features wise: Non Compliance Partial Compliance Compliance
Incomplete pakage Printer
not Quoted
Printer not Quoted giving 1
year warranty not according
to tender
Compliance
Not Acceptable Not Acceptable Acceptable
Not Acceptable Not Acceptable Acceptable
Specifications:
Technical Eligibility of Product:
Technical Eligibility of Firm:
PART - III
Specification Compliance / Evaluation Parameters
33 Laptop Computer
Country of Manufacturer
Compliance with defined quality standards
Not Acceptable Not Acceptable Acceptable
Non Responsive Non Responsive Responsive
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Mazhar Iqbal Noor Computer Programmer
Mr. Riaz Shahid
I.T Incharge
Dr. Khalid Hameed Zaighum
(DMS (Purchase)
Dr. Muhammad Asim
Assistant Professor of Cardiovasular
Dr. Usaid Qureshi Assistant Professor
Pead Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular
Surgery
Package No/Tender Number: PC/115677/35/CH&ICH Dated: 05-10-2017
Name of the Equipment and Qty:- Multimedia Projector
Part - I
KNOCK DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Sr. No Evaluation Parameters M/s National Industrial M/s Mecatech
Technical Evaluation
01 Complete Package/Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes No
04 Bid Security (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Remarks: Responsive Non Responsive
Sr. # Evaluation Parameters M/s National Industrial M/s Mecatech
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes No
Technical & Engineering capability (As defined for the specific tender in
(All Evaluation parameters defined below are mandatory for compliance.)
PART - II
KNOCK DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Techincal Committee
02Technical & Engineering capability (As defined for the specific tender in
specifications)Yes Not Satisfactory
03Certificate from the Manufacturer about the after sales services through
agent or itself (In casespecifically demanded in the specifications)Yes Not Attached
04 Bidder Past performance Yes Not Available
05 Availability of relevant Tools and Testing/ Calibration Equipment Yes Not Mentioned
06 Warranty Yes Not Mentioned
Remarks: Responsive Non Responsive
KNOCK DOWN CRITERIA - PRODUCT EVALUATION
Package No/Tender Number: PC/115677/35/CH&ICH Dated: 05-10-2017
(All evaluation parmeters defined below are mandatory for compliance.)
Item
Sr. No.M/s National Industrial M/s Mecatech
Brand NEC Not MentionedModel NEC P501 X SCX 4001 X 1
Japan Japan
Country of Origin of Product/Model Number Japan JapanYes No
Specification Compliance features wise: Compliance Non Compliance
ComplianceIncomplete bidding documents, No
certificates, Under Specification item
quoted
Specifications:
PART - III
Specification Compliance / Evaluation Parameters
35Multimedia
Projector
Country of Manufacturer
Compliance with defined quality standards
quotedAcceptable Not Accepted
Acceptable Not Accepted
Responsive Non Responsive
Technical Eligibility of Product:
Technical Eligibility of Firm:
Technical Status:
Engr. Ahmed Maqsood Electromedical Engineer
Engr. Muhammad Umair Kabeer Biomedical Engineer
Mazhar Iqbal Noor Computer Programmer
Mr. Riaz Shahid
I.T Incharge
Dr. Khalid Hameed Zaighum
(DMS (Purchase)
Dr. Muhammad Asim
Assistant Professor of Cardiovasular
Dr. Usaid Qureshi Assistant Professor
Pead Cardiology
Dr. Syed Najam Haider Associate Professor Pead Cardiovascular
Surgery