Loop systems – T-coil. Conny Andersson IEC 60118-4. Conny.Andersson@edin.se. Loop System – Wireless system. Receiver = T-coil in Hearing Instrument (HI) No current consumption Transmitter = Loop Amplifier and Loop World Wide Standard. Acoustic and T-coil input. Acoustic Output. Input. - PowerPoint PPT Presentation

TRANSCRIPT

www.univox.eu

Loop systems – T-coilConny Andersson

IEC 60118-4

Conny.Andersson@edin.se

www.univox.eu

Loop System – Wireless system

Receiver = T-coil in Hearing Instrument (HI)No current consumption

Transmitter = Loop Amplifier and Loop

World Wide Standard

www.univox.eu

Input

Acoustic and T-coil inputAcousticOutput

switch

MT

HearingInstrument

Loop Level:Fixed @ 400mA/mfor all signals

Speech level=70dB@1m distance

130dB

Gain +60dB

Coil Sensitivity130dB

Same output level for M and T-position

www.univox.eu

Relative T-coil sensitivitySwitching between M and T will give

approximately the same output level

It could be stated differently:0dB difference between M and T-position

Related to acoustic level: The acoustic output is the same for T and M-position with a 70dB acoustical input signal

www.univox.eu

International Standards, IEC

Hearing Instrument (HI) Many parameters are fitted for the

individual. Can not be standardized

Hearing Loop system All parameters are standardized

InternationalElectrotechnicalCommission

www.univox.eu

ANSI S3.22-2009American National Standard Institute has

many abbreviations for T-coil functions3.4 Terms relating to measurements with inductive inputs3.4.1 high frequency average SPL for an inductive telephone simulator (HFA-SPLITS). The high frequencyaverage of the SPLITS levels (see Clause 3.4.5). The abbreviation for this term is HFASPLITS.3.4.2 high frequency average SPL in a vertical magnetic field (HFA-SPLIV). The high-frequencyaverage of the SPLIV levels (see Clause 3.4.6). The abbreviation for this term is HFA-SPLIV.3.4.3 relative simulated equivalent telephone sensitivity (RSETS). Difference in decibels obtainedby subtracting the RTG + 60 dB SPL from the HFA-SPLITS. The abbreviation for this term is RSETS.3.4.4 relative test loop sensitivity (RTLS). Difference in decibels obtained by subtracting the RTG +60 dB SPL from the HFA-SPLIV. The abbreviation for this term is RTLS.3.4.5 sound pressure level for an inductive telephone simulator (SPLITS). Sound pressure leveldeveloped in the coupler by a hearing aid with the gain control at the RTS when the input is themagnetic field generated by a Telephone Magnetic Field Simulator (refer to Clause 3.2.5). Theabbreviation for this term is SPLITS.3.4.6 sound pressure level in a vertical magnetic field (SPLIV). Sound pressure level developed inthe coupler by a hearing aid with the gain control at the RTS when the input is a 31.6 mA/m sinusoidalalternating magnetic field parallel to the vertical reference and the input selector of the hearing aid setto T-position. The abbreviation for this term is SPLIV.

Good News, ANSI S3.22 committee has decided to make T-coil specification mandatory for quality test inspection

www.univox.eu

Standardized Parameters

Magnetic Field Strength, 400mA/m

Coverage area/space (sitting and standing),Should have minimum variations (+-3dB)

www.univox.eu

Field distribution – level/coverage

Variation less than +-3dB

www.univox.eu

Standardized Parameters (hum)

Magnetic field strength from electrical appliances

Received directly by the T-coilNOT by the hearing loop system

Less problematic due to stronger regulations of electrical environment

www.univox.eu

Hearing Instr. frequency response

Mic

T-coil

Hearing instrument, microphone/telecoil response

www.univox.eu

Often reduced low freq. T-coil

Mic

Tele-coil

Reduced low frequency sensitivity

25dB

Hearing instrument, microphone/telecoil response

www.univox.eu

HI has large variation in freq resp

Mic

Tele-coil

Hearing intrument, microphone/telecoil response Hearing instrument, microphone/telecoil response

www.univox.eu

Loop frequency response is flat

Mic

Tele-coil

Hearing instrument, microphone/telecoil response

Within +-3dB +-3dB

www.univox.eu

Loop specifications is regulated

It is therefore impossible to judge the full quality of a loop system by using a Hearing Instrument

Measurements are mandatory

www.univox.eu

Vertical Field Strength

The specifications for most of the parameters for loops are for vertical directions

This corresponds to HI manufacturers placing the T-coil in vertical position

www.univox.eu

-3dB (@45°)-3dB (@45°)

Conventional loops - tilt angel

0dB=full level

Silence!

Vertical field

Silence!

www.univox.eu

Loops/T-coil with TWO directions

full levelfull level

Vertical AND Horizontal Field0dB=full level

-3dB (@45°)-3dB (@45°)

www.univox.eu

Conclusion and a `T-coil wish`Increased low frequency responseHum harmonics removerDefault sensitivity should follow the standard

of 400mA/m (125ms, RMS) If the sensitivity is set higher in the Hearing

Instrument there is a strong risk for high distortion, clipping/saturation when entering a correct installed Loop System.A weak loop must not be compensated by increasing the T-coil sensitivity with a decreased Signal to Noise ratio.

www.univox.eu

Conclusion and a `T-coil wish`Several channels and modulated/coded

loop system.

One “standard loop system for world wide use in theatres etc

Other channels for school and domestic use. Decreased Signal to Noise ratio.

www.univox.eu

Thank you for your attention!

www.univox.eu

Thank you for your attention!

www.univox.eu

Input

Acoustic and T-coil inputAcousticOutput

switch

MT

HearingInstrument

Loop Level:Fixed @ 400mA/mfor all signals

Speech level=70dB@1m distance

130dB

Gain +60dB

Coil Sensitivity130dB

Same output level for M and T-position

www.univox.eu

Input

Acoustic and T-coil inputAcousticOutput

switch

MT

HearingInstrument

Loop Level:Fixed @ 100mA/mfor all signals

Speech level=70dB@1m distance

130dB

Gain +60dB

Coil Sensitivity130dB

Same output level for M and T-position

Long-term signals:

www.univox.eu

Input

Acoustic and T-coil inputAcousticOutput

switch

MT

HearingInstrument

Loop Level:Fixed @ 31.6 mA/mfor all signals

Speech level=60dB@1m distance

120dB

Gain +60dB

Coil Sensitivity120dB

Same output level for M and T-position

Long-term signals: