Mona D. Jensen, PhD
R&D Manager
IDEXX Laboratories, Inc.
July 2009
perFOrmance OF IDEXX Catalyst Dx® Chemistry Analyzer
iDexx technical supportU.S./Canada/Latin America 1-800-248-2483
Europe 00800 1234 3399Australia 1800 655 978
New Zealand 0800 102 084
Spreading layerFilters sample to minimize interference from lipemic, icteric and hemolyzed samples.
reagent/regiStration layer(S)Chemical reaction takes place here
Support layer
Objective: to correlate accuracy and precision of the ideXX Catalyst dx® Chemistry analyzer with a nationally recognized reference standard, the VitroS® 250 Chemistry System from ortho-Clinical diagnostics, a Johnson & Johnson company.
summary:Canine, feline and equine serum, plasma and whole blood samples were tested, as well as urine samples for uCre and upro. Samples were chosen to cover the entire dynamic range of individual chemistry tests.
patient samples were analyzed on three Catalyst dx Chemistry analyzers and in duplicate on one VitroS 250 Chemistry System for 24 different chemistries.
a high degree of correlation was observed for all chemistries between the VitroS 250 Chemistry System and the Catalyst dx Chemistry analyzer. Correlation coefficients ranged from 0.959 to 0.998
clinical implicatiOns: the Catalyst dx Chemistry analyzer is used to perform clinical chemistry analyses in a timely manner. it is important that such technologies meet reference methodology standards. By comparing the Catalyst dx Chemistry analyzer results to those of the VitroS 250 Chemistry System, which is referenced against the international Federation of Clinical Chemistry (iFCC), the Clinical and laboratory Standards institute (ClSi) and the Center for disease Control (CdC) recommended and/or approved methods, Catalyst dx Chemistry analyzer accuracy has been established.
Dry-sliDe technOlOgy:Both the VitroS 250 Chemistry System and the Catalyst dx Chemistry analyzer utilize dry-slide technology, which offers many advantages. the dry slide is made of several layers that help minimize interferences from lipemic, icteric and hemolyzed samples. dry slides are easier to use, have less user variability, provide flexible testing options and are more economical than liquid reagent analyzers. Chemical reactions that take place on dry slides are read optically and processed by either endpoint or rate protocols.
insiDe the catalyst Dx® chemistry analyzer sliDe:
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
R2 = 0.97Slope = 1.03
Intercept = -0.08n = 2736
7
6
5
4
3
2
1
00 1 2 3 4 5 6 7
albbun
Cat
alys
t Dx
Con
cent
ratio
n (U
/L)
VITROS 250 Concentration (U/L)
2000
1500
1000
500
00 500 1000 1500 2000
R2 = 0.98Slope = 1.01
Intercept = -3.37n = 2741
alKp
Cat
alys
t Dx
Con
cent
ratio
n (U
/L)
VITROS 250 Concentration (U/L)
1000
900800700
600500400
300200
1000
0 100 200 300 400 500 600 700 800 900
R2 = 0.99Slope = 0.93
Intercept = 9.31n = 2436
alt
Cat
alys
t Dx
Con
cent
ratio
n (U
/L)
VITROS 250 Concentration (U/L)
2400
2000
1600
1200
800
400
00 200 600 1000 1400 1800 2200
R2 = 0.97Slope = 1.02
Intercept = 1.14n = 1349
amyl
Cat
alys
t Dx
Con
cent
ratio
n (U
/L)
VITROS 250 Concentration (U/L)
1000
800
600
400
200
00 200 400 600 800 1000
R2 = 0.99Slope = 0.94
Intercept = 1.99n = 699
ast
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
140
120
100
80
60
40
20
00 10 20 30 40 50 60 70 80 90 110 130
R2 = 0.99Slope = 0.99
Intercept = 0.84n = 2610
bun
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
16
14
12
10
8
6
4
2
0
R2 = 0.98Slope = 0.95
Intercept = 0.15n = 2352
0 2 4 6 8 10 12 14 16
ca
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
500450400350300250200150100
500
0 50 100 150 200 250 300 350 400 450 500
R2 = 0.99Slope = 0.99
Intercept = 3.50n = 1737
chOl
Cat
alys
t Dx
Con
cent
ratio
n (U
/L)
VITROS 250 Concentration (U/L)
200018001600140012001000
800600400200
00 200 400 600 800 1200 1600 2000
R2 = 0.99Slope = 1.02
Intercept = -7.85n = 591
cK
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
14
12
10
8
6
4
2
00 1 2 3 4 5 6 7 8 9 10 11 12 13 14
R2 = 0.99Slope = 0.99
Intercept = 0.03n = 2687
crea
Cat
alys
t Dx
Con
cent
ratio
n (U
/L)
VITROS 250 Concentration (U/L)
1000
900
800700
600500400300
200
1000
0 100 200 300 400 500 600 700 800 900
R2 = 0.99Slope = 0.93
Intercept = 2.04n = 1862
ggt
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
700
600
500
400
300
200
100
00 100 200 300 400 500 600 700
R2 = 0.99Slope = 1.01
Intercept = -1.01n = 2500
glu
materials and methods: Canine, feline and equine serum, plasma and whole blood samples were evaluated on both the VitroS® 250 Chemistry System and the Catalyst dx® Chemistry analyzer according to the applicable users’ guides. urine samples were similarly assayed for upro and uCre. Samples were run on three or four Catalyst dx Chemistry analyzers. reference concentration is the mean of two runs on the VitroS 250 Chemistry System. data was collected and plotted on graphs to determine regression statistics.
results: the Catalyst dx Chemistry analyzer has a high degree of correlation for each of the 24 chemistries with the VitroS 250 Chemistry System. Below are correlation graphs for individual chemistries showing correlation coefficient (r2), slope, intercept, and number of samples (n).
legend: Canine + Equine x Feline
the cOmparisOnthe evaluatiOn
Cat
alys
t Dx
Con
cent
ratio
n (m
mol
/L)
VITROS 250 Concentration (mmol/L)
11
9
7
5
32
00 1 2 3 4 5 6 7 8 9 10 11 12
R2 = 0.99Slope = 0.99
Intercept = 0.01n = 137
lac
Cat
alys
t Dx
Con
cent
ratio
n (U
/L)
VITROS 250 Concentration (U/L)
2800
2400
2000
1600
1200
800
400
00 200 600 1000 1400 1800 2200 2600
R2 = 0.99Slope = 0.97
Intercept = 9.29n = 594
lDh
Cat
alys
t Dx
Con
cent
ratio
n (U
/L)
VITROS 250 Concentration (U/L)
6000
5000
4000
3000
2000
1000
00 1000 2000 3000 4000 5000 6000
R2 = 0.96Slope = 1.05
Intercept = 81.62n = 1316
lipa
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
5
4
3
2
1
00 1 2 3 4 5
R2 = 0.99Slope = 0.95
Intercept = 0.10n = 223
mgC
atal
yst D
x C
once
ntra
tion
(µm
ol/L
)
VITROS 250 Concentration (µmol/L)
1000
900
800700
600500
400
300200
1000
0 100 200 300 400 500 600 700 800 900
R2 = 0.96Slope = 1.12
Intercept = 7.97n = 590
nh3
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
16
14
12
10
8
6
4
2
00 2 4 6 8 10 12 14 16
R2 = 0.99Slope = 1.03
Intercept = -0.20n = 1908
phOs
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
28
24
20
16
12
8
4
00 2 4 6 8 10 12 14 16 18 20 22 24 26 28
R2 = 0.99Slope = 1.02
Intercept = -0.01n = 1848
tbil
Cat
alys
t Dx
Con
cent
ratio
n (g
/dL)
VITROS 250 Concentration (g/dL)
12
10
8
6
4
2
00 2 4 6 8 10 12
R2 = 0.97Slope = 0.99
Intercept = 0.18n = 2424
tp
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
350
300
250
200
150
100
50
00 50 100 150 200 250 300 350
R2 = 0.99Slope = 0.98
Intercept = 0.93n = 92
trig
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
350
300
250
200
150
100
50
00 50 100 150 200 250 300 350
R2 = 0.99Slope = 0.96
Intercept = 2.28n = 188
ucre
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
400
350
300
250
200
150
100
50
00 50 100 150 200 250 300 350 400
R2 = 0.98Slope = 1.06
Intercept = -10.00n = 365
uprO
Cat
alys
t Dx
Con
cent
ratio
n (m
g/dL
)
VITROS 250 Concentration (mg/dL)
20
15
10
5
00 5 10 15 20
R2 = 0.99Slope = 1.01
Intercept = -0.52n = 21
uric
cOmparisOn cont.
the ideXX Catalyst dx® Chemistry analyzer provides accurate and consistent analysis of blood and urine components as compared to nationally recognized standards, the international Federation of Clinical Chemistry (iFCC), the Clinical and laboratory Standards institute (ClSi) and the Center for disease Control (CdC). the ideXX Catalyst dx Chemistry analyzer provides unmatched diagnostic decision support for in-house analysis.
© 2009 IDEXX Laboratories, Inc. All rights reserved. • 09-68136-00VitroS is a registered trademark of ortho-Clinical diagnostics. all other ®/tM marks are owned by ideXX laboratories, inc. or its affiliates in the united States and/or other countries.
cOnclusiOn