IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
L 6: X Ray production
IAEAInternational Atomic Energy Agency
IntroductionA review is made of: The main elements of the of X Rays tube: cathode and anode structure The technology constraints of the anode and cathode material The rating charts and X Ray tube heat loading capacitiesIAEA
6: X Ray production
2
Topics Basic elements of an X Ray source assembly Cathode structure Anode structure Rating chart X Ray generator Automatic exposure control
IAEA
6: X Ray production
3
Overview To become familiar with the technologicalprinciples of the X Ray production
IAEA
6: X Ray production
4
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Part 6: X Ray productionTopic 1: Basic elements of an X Ray source assembly
IAEAInternational Atomic Energy Agency
Basic elements of the X Ray source assembly Generator : powercircuit supplying the required potential to the X Ray tube X Ray tube and collimator: device producing the X Ray beamIAEA
6: X Ray production
6
X Ray tubes
IAEA
6: X Ray production
7
X Ray tube components Cathode: heated filament which is the sourceof the electron beam directed towards the anode tungsten filament
Anode (stationary or rotating): impacted by
electrons, emits X Rays Metal tube housing surrounding glass (or metal) X Ray tube (electrons are traveling in vacuum) Shielding material (protection against scattered radiation)IAEA
6: X Ray production
8
X Ray tube componentshousing cathode
1: mark of focal spotIAEA
1: long tungsten filament 2 : short tungsten filament 3 : real size cathode 6: X Ray production 9
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Part 6: X Ray production
Topic 2: Cathode structure
IAEAInternational Atomic Energy Agency
Cathode structure (I)Cathode includes filament(s) and associated circuitry tungsten material : preferred because of its highmelting point (3370C) slow filament evaporation no arcing minimum deposit of W on glass envelope
To reduce evaporation the emission temperature of the cathode is reached just before the exposure in stand-by, temperature is kept at 1500C so that2700C emission temperature can be reached within a second IAEA
6: X Ray production
11
Example of a cathode
IAEA
6: X Ray production
12
Cathode structure (I)
Modern tubes have two filaments
a long one : higher current/lower resolution a short one : lower current/higher resolution
Coulomb interaction makes the electron beam divergent on the travel to the anodelack of electrons producing X Rays larger area of target used focal spot increased lower image resolution
Focalisation of electrons is crucial !IAEA
6: X Ray production
13
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Part 6: X Ray production
Topic 3: Anode structure
IAEAInternational Atomic Energy Agency
X Ray tube characteristics Anode mechanical constraints Material : tungsten, rhenium, molybdenum, graphite Focal spot : surface of anode impacted by electrons Anode angle Disk and annular track diameter (rotation frequency from 3,000 to 10,000 revolutions/minute) Thickness mass and material (volume) heat capacity
Anode thermal constraints Instantaneous power load (heat unit) Heat loading time curve Cooling time curveIAEA
6: X Ray production
15
Anode angle (I)
The Line-Focus principle
Anode target plate has a shape that is more rectangular or ellipsoidal than circular
the shape depends on : filament size and shape focusing cups and potential distance between cathode and anode
This conflict is solved by slantingthe target faceIAEA
Image resolution requires a small focal spot Heat dissipation requires a large spot
6: X Ray production
16
Anode characteristic
1 : anode track 2 : anode track
IAEA
6: X Ray production
17
Anode angle (II) Angle Angle
Incident electron beam width
Actual focal spot size Incident electron beam width Apparent focal spot size Film
Actual focal spot size Increased apparent focal spot size Film
THE SMALLER THE ANGLE THE BETTER THE RESOLUTIONIAEA
6: X Ray production
18
Anode heel effect (I) Anode angle (from 7 to 20) induces avariation of the X Ray output in the plane comprising the anode-cathode axis Absorption by anode of X photons with low emission angle The magnitude of influence of the heel effect on the image depends on factors such as : anode angle size of film focus to film distance
Anode aging increases heel effectIAEA
6: X Ray production
19
Anode heel effect (II) The heel effect is not always a negativefactor It can be used to compensate for different attenuation through parts of the body For example: thoracic spine (thicker part of the patienttowards the cathode side of the tube) mammography6: X Ray production
IAEA
20
Focal spot size and imaging geometry Focal spot finite size image unsharpened Improving sharpness small focal spot size For mammography focal spot size 0.4 mm nominal Small focal spot size reduced tube output (longerexposure time)
Large focal spot allows high output (shorter exposuretime)
Balance depends on organ movement (fast movingorgans may require larger focus)IAEA
6: X Ray production
21
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Part 6: X Ray production
Topic 4: Rating Chart
IAEAInternational Atomic Energy Agency
Heat loading capacities A procedure generates an amount of heat depending on: kV used, tube current (mA), length of exposure type of voltage waveform number of exposures taken in rapid sequence
Heat Unit (HU) [joule] :unit of potential x unit of tube current x unit of time
The heat generated by various types of X Ray circuits are: 1 phase units : 3 phase units, 6 pulse : 3 phase units, 12 pulse:IAEA
HU = kV x mA x s HU = 1.35 kV x mA x s HU = 1.41 kV x mA x s 23
6: X Ray production
X Ray tube rating chart (I) Tube cooling characteristics and focal spot size {mA - time} relationship at constant kV intensity decreases with increasing exposure time intensity increases with decreasing kV
Note: higher power reduced exposure time reduced motion unsharpness
IAEA
6: X Ray production
24
X Ray tube rating chart (II) Manufacturerscombine heat loading characteristics and information about the limits of their X Ray tubes in graphical representations called Tube Rating Charts Example: Tube A: a 300 mA, 0.5 s, 90 kV procedure woulddamage the system operated from a 1-phase half wave rectified generator (unacceptable)
Tube B: a 200 mA, 0.1 s, 120 kV procedure comply withthe technical characteristics of the system operated from a 3-phase fully rectified generator (acceptable)
IAEA
6: X Ray production
25
X Ray tube rating chart (III)X Ray tube A 1 half-wave rectified 3000 rpm 90 kV 1.0 mm effective focal spot
700 600
Tube current (mA)
500 400 300 200 100 0.01 0.05 0.1 0.5 1.0 5.0 10.0970
kV p
120 k 0 kVp Vp
Unacceptable
50 kVp
Exposure time (s)
IAEA
6: X Ray production
26
X Ray tube rating chart (IV)
700
Tube current (mA)
600 500 400 300 200 100 0.01 0.05 0.1
70 kV p 90 kV p 125 kV p
X Ray tube B 3 full-wave rectified 10.000 rpm 125 kV 1.0 mm effective focal spotUnacceptable
50 kV p
Acceptable
0.5
1.0
5.0
10.0
Exposure time (s)IAEA
6: X Ray production
27
Anode cooling chart (I) Heat generated is stored in the anode, anddissipated through the cooling circuit A typical cooling chart has: input curves (heat units stored as a function of time) anode cooling curve
The following graph shows that: a procedure delivering 500 HU/s can go onindefinitely if it is delivering 1000 HU/s it has to stop after 10 min if the anode has stored 120.000 HU, it will take 5 min to cool down completelyIAEA
6: X Ray production
28
Anode cooling chart (II)240
Maximum Heat Storage Capacity of AnodeHU 00 10500
Heat units stored (x 1000)
220 200 180 160 140 120 100 80 60 40 20
c /seec
Imput curve
/s HU
U/ 350 H
sec
ec 250 HU/s
Co oli
ng c
urv e
1
2
3
4
5
6
7
8
9
10
11
12
13
14
IAEA
Elapsed time (min)
6: X Ray production
29
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Part 6: X Ray production
Topic 5: X Ray generator
IAEAInternational Atomic Energy Agency
X-ray generator (I) It supplies the X-ray tube with : Current to heat the cathode filament Potential to accelerate electrons Automatic control of exposure (power application time) Energy supply 1000 X-ray beam energy (of which 99.9% is dissipated as thermal energy)IAEA
6: X Ray production
31
X-ray generator (II) Generator characteristics have a strong influence onthe contrast and sharpness of the radiographic image The motion unsharpness can be greatly reduced by a generator allowing an exposure time as short as achievable Since the dose at the image plane can be expressed as: D = k0 . Un . I . T
U: peak voltage (kV) I: mean current (mA) T: exposure time (ms) n: ranging from about 1.5 to 3 6: X Ray production 32
IAEA
X-ray generator (III) Peak voltage value has an influence on thebeam hardness It has to be related to medical question What is the anatomical structure to investigate ? What is the contrast level needed ? For a thorax examination : 140 - 150 kV is suitable tovisualize the lung structure While only 65 kV is necessary to see bone structure
The ripple r of a generator has to be as low aspossibleIAEA
r = [(U - Um )/U] x 100% in6: X Ray production 33
Tube potential wave form (I) Conventional generators single single three three 1-pulse (dental and some mobile systems) 2-pulse (double rectification) 6-pulse 12-pulse
Constant potential generators (CP) HF generators (use of DC choppers to convert50Hz mains into voltages with frequencies in the kHz range) Inverter technologyIAEA
6: X Ray production
34
Tube potential wave form (II)kV ripple (%) 100% 13% Three phase 6-pulse 4% Three phase 12-pulse Single phase single pulse Single phase 2-pulse
Line voltage
0.01 s 0.02 s
IAEA
6: X Ray production
35
The choice of the number of pulses (I) Single pulse : low power (