SCIENTEK™
WARMING CABINET SWC Series
PRODUCT SPECIFICATION
Scientek Technology Corporation Tel: 604 273 9094 SWC Series Warming Cabinets Fax: 604 273 1262 Specifications, Rev 0, March 2005
1. GENERAL DESCRIPTION 3
1.1 CABINET CONSTRUCTION 3 1.2 DOOR ASSEMBLY 3 1.3 ELECTRIC HEATER AND FAN ASSEMBLY 3 1.4 CONTROL MODULE 3 1.5 WIRING 3
2. MODEL SPECIFICATIONS 4
2.1 SWC-015 & SWC-015G (GLASS DOOR) 5 2.2 SWC-028 & SWC-028G (GLASS DOOR) 6 2.3 SWC-041 & SWC-041G (GLASS DOOR) 7 2.4 SWC-061 & SWC-061G (GLASS DOOR) 8 2.5 SWC-015-015, SWC-015G-015, SWC015-015G, SWC-015G-015G 9 2.6 SWC-015-028, SWC-015G-028, SWC-015-028G, SWC-015G-028G 10 2.7 SWC-015-041, SWC-015G-041, SWC-015-041G, SWC-015G-041G 11 2.8 SWC-028-028, SWC-028G-028, SWC-028-028G, SWC-028G-028G 12
3. AGENCY APPROVALS 13
3.1 ELECTRICAL/ SAFETY 13 3.2 SEISMIC ANCHORAGE 13
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
2
1. GENERAL DESCRIPTION
Scientek warming cabinets are designed to be used in hospital and research laboratories for warming blankets and solutions. The temperature in the cabinet is controlled in the range 90°F -160°F
The unit is made of following major components: cabinet, door, electric heater and fan assembly, control module and wiring.
1.1 Cabinet Construction Formed stainless steel, double wall, insulated. The cabinet can be mounted on a seismic base for floor mounting, or fitted with adjustable rubber feet for countertop mounting, or mounted on another cabinet using mounting hardware.
1.2 Door Assembly The cabinet door assembly consists of a door panel, hinges, latches and the seal. It can be either solid stainless steel door or glass door. The stainless steel door is insulated double wall and the glass door is made of a stainless steel frame with tempered glass insert. The door can be mounted on either left or right hand side
1.3 Electric Heater and Fan Assembly The heater and fan assembly provide the hot air and circulate the air inside the chamber. The heater, fan, high temperature switch and K-type thermo probe are mounted inside the formed stainless steel channel, perforated to allow appropriate air circulation. The high temperature switch is provided to disconnect the heater in a case of the temperature raise over the limit. The temperature probe is provided to feed the signal to temperature control unit.
1.4 Control Module The control module consists of the illuminated OFF/ON switch, temperature control unit, control relay and alarm buzzer.
Appropriate temperature control unit and a control relay are (electromechanical or solid state) are used to maintain the temperature inside the cabinet at the set point.
The alarm buzzer is turned on by the temperature control unit to notify of a temperature deviation inside the cabinet beyond specified alarm limits.
1.5 Wiring The wiring includes power supply cord with the plug and internal wiring connecting electrical components. The wiring is suitable for current load and the temperature range.
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
3
2. MODEL SPECIFICATIONS
The warming cabinets are designed in four base models and multiple modular models.
The base models are a single cabinet style and use the same control module, heater and fan assembly. Only difference is the height of the cabinet, while the width and depth are identical.
The modular style models consist of two base models combined into one unit. The modules share the same power supply but have separate control modules and heater and fan assemblies.
The warming cabinet models with glass door have a suffix ‘G’ following the three-digit part of the model number.
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
4
2.1 SWC-015 & SWC-015G (Glass Door)
Compartment Size
4 ft3
25.5”Wx17.5”Dx15.5”H
Overall Size 30”W x 24”D x 28”H
Electrical 120VAC, 6Amp, 60Hz
8’ cord, 15A provided
Heat Loss Max 810 Btu/hr
Standard Features Digital Temperature Control
ºF or ºC Display
Accuracy +/- 1ºF (ºC)
Uniform Heating
One adjustable shelf
Reverseable Door Hinges
Door Lock
Stackable with SWC-015, -028 & -041
Low-E Tempered Glass (SWC-015G)
Options Seismic Anchoring
Casters (2 with brakes, total 4)
Leveling Feet (total 4)
Shelf SS (additional)
Left Side Door Hinge
Audible Alarm (high & low temperature)
Trim SS on One Side
Trim SS on Two Sides
Power Plug Hospital Grade
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
5
2.2 SWC-028 & SWC-028G (Glass Door)
Compartment Size
7.4 ft3
25.5”Wx17.5”Dx28.5”H
Overall Size 30”W x 24”D x 41”H
Electrical 120VAC, 6Amp, 60Hz
8’ cord, 15A provided
Heat Loss Max 1180 Btu/hr
Standard Features Digital Temperature Control
ºF or ºC Display
Accuracy +/- 1ºF (ºC)
Uniform Heating
One adjustable shelf
Reverseable Door Hinges
Door Lock
Stackable with SWC-015, -028
Low-E Tempered Glass (SWC-028G)
Options Seismic Anchoring
Casters (2 with brakes, total 4)
Leveling Feet (total 4)
Shelf SS (additional)
Left Side Door Hinge
Audible Alarm (high & low temperature)
Trim SS on One Side
Trim SS on Two Sides
Power Plug Hospital Grade
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
6
2.3 SWC-041 & SWC-041G (Glass Door)
Compartment Size
10.75 f
25.5”Wx17.5”Dx41.5”H
Overall Size 30”W x 24”D x 54”H
Electrical 120VAC, 6Amp, 60Hz
8’ cord, 15A provided
Heat Loss Max 1550 Btu/hr
Standard Features Digital Temperature Control
ºF or ºC Display
Accuracy +/- 1ºF (ºC)
Uniform Heating
Two adjustable shelves
Reverseable Door Hinges
Door Lock
Stackable with SWC-015
Low-E Tempered Glass (SWC-041G)
Options Seismic Anchoring
Casters (2 with brakes, total 4)
Leveling Feet (total 4)
Shelf SS (additional)
Left Side Door Hinge
Audible Alarm (high & low temperature)
Trim SS on One Side
Trim SS on Two Sides
Power Plug Hospital Grade
t3
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
7
2.4 SWC-061 & SWC-061G (Glass Door)
Compartment Size
15.9 ft3
25.5”Wx17.5”Dx61.5”H
Overall Size 30”W x 24”D x 74”H
Electrical 120VAC, 12Amp, 60Hz
8’ cord, 15A provided
Heat Loss Max 2120 Btu/hr
Standard Features Digital Temperature Control
ºF or ºC Display
Accuracy +/- 1ºF (ºC)
Uniform Heating
Three adjustable shelves
WC-061G)
Reverseable Door Hinges
Door Lock
Low-E Tempered Glass (S
Options
s, total 4)
low temperature)
ade
Seismic Anchoring
Casters (2 with brake
Leveling Feet (total 4)
Shelf SS (additional)
Left Side Door Hinge
Audible Alarm (high &
Trim SS on One Side
Trim SS on Two Sides
Power Plug Hospital Gr
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
8
2.5 SWC-015-015, SWC-015G-015, SWC015-015G, SWC-015G-015G
Top Cabinet
Size 4 ft3
25.5”Wx17.5”Dx15.5”H
Top Cabinet Size
4 ft3
25.5”Wx17.5”Dx15.5”H
30”W x 24”D x 48”H
Electrical 120VAC, 12Amp, 60Hz
8’ cord, 15A provided
Heat Loss Max 1620 Btu/hr
Standard Features ntrol
ble shelf – Top Cabinet
Reverseable Door Hinges
ock
lass (glass door)
Digital Temperature Co
ºF or ºC Display
Accuracy +/- 1ºF (ºC)
Uniform Heating
One adjusta
One adjustable shelf – Bottom Cabinet
Door L
Low-E Tempered G
Options Seismic Anchoring
Casters (2 with brakes, total 4)
temperature)
Trim SS on One Side
Trim SS on Two Sides
Power Plug Hospital Grade
Leveling Feet (total 4)
Shelf SS (additional)
Left Side Door Hinge
Audible Alarm (high & low
Overall Size
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
9
2.6 SWC-015-028, SWC-015G-028, SWC-015-028G, SWC-015G-028G
Top Cabinet 4 ft
Size
3
x17.5”Dx15.5”H 25.5”W
Size 7.4 ft3
25.5”W
30”W x 24”D x 61”H
Electrical 120VAC, 12Amp, 60Hz
8’ cord, 15A provided
Max 1990 Btu/hr
Standard Features Control
(ºC)
lf – Top Cabinet
et
pered Glass (glass door)
Digital Temperature
ºF or ºC Display
Accuracy +/- 1ºF
Uniform Heating
One adjustable she
One adjustable shelf – Bottom Cabin
Reverseable Door Hinges
Door Lock
Low-E Tem
Options
s, total 4)
low temperature)
ade
Seismic Anchoring
Casters (2 with brake
Leveling Feet (total 4)
Shelf SS (additional)
Left Side Door Hinge
Audible Alarm (high &
Trim SS on One Side
Trim SS on Two Sides
Power Plug Hospital Gr
Heat Loss
Top Cabinet x17.5”Dx28.5”H
Overall Size
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
10
2.7 SWC-015-041, SWC-015G-041, SWC-015-041G, SWC-015G-041G
Top Cabinet Size
4 ft3
x17.5”Dx15.5”H
25.5”W
10.75 ft
25.5”Wx17.5”Dx41.5”
verall Size 30”W x 24”D x 74”H
120VAC, 12Amp, 60H
8’ cord, 15A provi
tandard Features Digital Temperatu
ºF or ºC Display
Accuracy +/- 1ºF (
Uniform Heating
One adjustable shelf – Top Cabinet
Two adjustable shelves – B
Reverseable
Door Lock
Seismic Anchoring
Casters (2 with brakes,
Leveling Feet (total 4)
Shelf SS (additional)
Left Side Door Hinge
Audible Alarm (high &
Trim SS on One Side
Trim SS on Two Sides
Options
total 4)
low temperature)
Power Plug Hospital Grade
Top Cabinet Size
3
H
O
Electrical z
ded
Heat Loss Max 2360 Btu/hr
Sre Control
ºC)
ottom Cabinet
Door Hinges
Low-E Tempered Glass (glass door)
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
11
2.8 SWC-028-028, SWC-028G-028, SWC-028-028G, SWC-028G-028G
Top Cabinet Size
7.4 ft3
25.5”Wx17.5”Dx28.5”H
Top Cabinet Size
10.75 ft3
25.5”Wx17.5”Dx28.5”H
Overall Size 30”W x 24”D x 74”H
Electrical 120VAC, 12Amp, 60Hz
8’ cord, 15A provided
Heat Loss Max 2360 Btu/hr
Standard Features Digital Temperature Control
ºF or ºC Display
Accuracy +/- 1ºF (ºC)
Uniform Heating
One adjustable shelf – Top Cabinet
One adjustable shelf – Bottom Cabinet
Reverseable Door Hinges
Door Lock
Low-E Tempered Glass (glass door)
Options Seismic Anchoring
Casters (2 with brakes, total 4)
Leveling Feet (total 4)
Shelf SS (additional)
Left Side Door Hinge
Audible Alarm (high & low temperature)
Trim SS on One Side
Trim SS on Two Sides
Power Plug Hospital Grade
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
12
Scientek Technology Corporation SWC Series Warming Cabinets Specifications, Rev 0, March 2005
13
3.1 Electrical/ S c UL standard.
3.2 Seismic n
The seismic anchorage for all models conforms199 ing Code (UBC
3. AGENCY APPROVALS
afety All models ertified to
A chorage
code and 7 Uniform Build with the 2001 California Building
)