vts098 pull down - empty ambient = 20°c, set point = -82°c ... · test environment: controlled...
Post on 27-Jul-2020
0 Views
Preview:
TRANSCRIPT
-100
-80
-60
-40
-20
0
20
40
0,00 0,50 1,00 1,50 2,00 2,50 3,00
Tem
pe
ratu
re °
C
Time hour
VTS098 Pull down - EmptyAmbient = 20°C, Set point = -82°C, Humidity = 60%
Back in bottom Center in bottom Front in bottom Back in top Center in top Front in top
Tr 17123
Pull down test from ambient temperature – VTS098
Test Conditions:
Result of test:
Starting at ambient temperature, the set point was reached after 173 min.
Model: VTS098
SN: 64938807
Test environment: Controlled conditions
Ambient temperatur: +20°C ± 1°C
Ambient humidity: 60% ± 5%
Temperature set-point in cabinet: -82°C
Cabinet load: None
Door openings: None
Sensors used: 25g tinned brass formed as a cylinder with diameter of 15.2mm
Sensors position: 6 points measure distributed diagonal in the cabinet layers. Sensors a placed with one in front where the door opens, one in the middle and one in the back opposite of the front sensor
Installation: Appliance installed according to the instruction manual condition
Refrigerant: Nature R
-85
-84
-83
-82
-81
-80
-79
0,00 2,00 4,00 6,00 8,00 10,00 12,00
Tem
pe
ratu
re °
C
Time hour
VTS098 Cyclic operation - EmptyAmbient = 20°C, Set point = -82°C, Humidity = 60%
Back in bottom Center in bottom Front in bottom
Back in top Center in top Back in top
Tr 18108
Cyclic operation – VTS098
Test Conditions:
Result of test:
At an ambient temperature at 20°C, the energy for 24 hour is 6,780 kWh
The temperature difference between top and bottom is 5,4°C
Model: VTS098
SN: 64938807
Test environment: Controlled conditions
Ambient temperatur: +20°C ± 1°C
Ambient humidity: 60% ± 5%
Temperature set-point in cabinet: -82°C
Cabinet load: None
Door openings: None
Sensors used: 25g tinned brass formed as a cylinder with diameter of 15.2mm
Sensors position: 6 points measure distributed diagonal in the cabinet layers. Sensors a placed with one in front where the door opens, one in the middle and one in the back opposite of the front sensor
Installation: Appliance installed according to the instruction manual condition
Refrigerant: Nature R
-90
-85
-80
-75
-70
-65
-60
0 0,2 0,4 0,6 0,8 1 1,2 1,4 1,6 1,8 2
VTS098 Door open - EmptyAmbient = 20°C, Set point = -82°C, Humidity = 60%
Back in top Center in top Front in top Back in bottom Center in bottom Front in bottom
Tr 18108
Door open 1 minute – VTS098
Test Conditions:
Result of test:
Starting at set point, after a door opening of 1 minute, the temperature recovered to set point after 85 min
Model: VTS098
SN: 64938807
Test environment: Controlled conditions
Ambient temperatur: +20°C ± 1°C
Ambient humidity: 60% ± 5%
Temperature set-point in cabinet: -82°C
Cabinet load: None
Door openings: 1 in 1 minute
Sensors used: 25g tinned brass formed as a cylinder with diameter of 15.2mm
Sensors position: 6 points measure distributed diagonal in the cabinet layers. Sensors a placed with one in front where the door opens, one in the middle and one in the back opposite of the front sensor
Installation: Appliance installed according to the instruction manual condition
Refrigerant: Nature R
-90
-80
-70
-60
-50
-40
-30
0,00 0,20 0,40 0,60 0,80 1,00 1,20 1,40 1,60
Tem
pe
ratu
re °
C
Time hour
VTS098 Hold over - EmptyAmbient = 20°C, Set point = -82°C, Humidity = 60%
Back in bottom Center in bottom Front in bottom Back in top Center in top Front in top
Tr 17123
Hold over time from -82°C to -60°C – VTS098
Test Conditions:
Result of test:
Starting at set point temperature, the temperature rise above -60°C between 58min and 80min.
Model: VTS098
SN: 64938807
Test environment: Controlled conditions
Ambient temperatur: +20°C ± 1°C
Ambient humidity: 60% ± 5%
Temperature set-point in cabinet: -82°C
Cabinet load: None
Door openings: None
Sensors used: 25g tinned brass formed as a cylinder with diameter of 15.2mm
Sensors position: 6 points measure distributed diagonal in the cabinet layers. Sensors a placed with one in front where the door opens, one in the middle and one in the back opposite of the front sensor
Installation: Appliance installed according to the instruction manual condition
Refrigerant: Nature R
-90
-80
-70
-60
-50
-40
-30
0,00 2,00 4,00 6,00 8,00 10,00 12,00 14,00 16,00
Tem
per
atu
re °
C
Time hour
VTS098 Hold over- FullAmbient = 20°C, Set point = -82°C, Humidity = 60%
Back in bottom Center in bottom Front in bottom Back in top Center in top Front in top
Tr 17040
Hold over time from -82°C to -60°C – VTS098
Test Conditions:
Result of test:
Starting at set point temperature, the temperature rise above -60°C between 220min and 870min.
Model: VTS098
SN: 64938807
Test environment: Controlled conditions
Ambient temperatur: +20°C ± 1°C
Ambient humidity: 60% ± 5%
Temperature set-point in cabinet: -82°C
Cabinet load: Full
Door openings: None
Sensors used: 25g tinned brass formed as a cylinder with diameter of 15.2mm
Sensors position: 6 points measure distributed diagonal in the cabinet layers. Sensors a placed with one in front where the door opens, one in the middle and one in the back opposite of the front sensor
Installation: Appliance installed according to the instruction manual condition
Refrigerant: Nature R