X-Ray Photoelectron Spectroscopy: Theory and Practice
PHYS-481 (Fall 2014)
University of Illinois at Chicago RRC - Electron Microscopy Service
Contact Information for EMS in RRC-East
Alan Nicholls, PhD Director of Research Service Facility - Electron Microscopy Research Resources Center-East 845 West Taylor Street SES Building, Room 110 Email: nicholls@uic.edu Office: (312) 996-1227
Tad Daniel, PhD Senior Research Specialist - Electron Microscopy Research Resources Center-East 845 West Taylor Street SES Building, Room 112 Email: tad1289@uic.edu Office: (312) 355-2087
University of Illinois at Chicago RRC - Electron Microscopy Service
Outline of Lecture
(1) Background
(2) Vacuum 101
(3) Analytical Capabilities
(4) Instrumentation
(5) Spectrum Simulation
(6) Summary
University of Illinois at Chicago RRC - Electron Microscopy Service
Background
Photoelectric effect discovered by Albert Einstein
Nobel Prize
1921
Photoemission as an analytical tool demonstrated by Kai Siegbahn (Electron Spectroscopy for Chemical Analysis – ESCA)
Nobel Prize
1981
University of Illinois at Chicago RRC - Electron Microscopy Service
Quantum Numbers Spectroscopist Notation n l s j nlj 1 0 ± 1/2 1/2 1s1/2 2 0 ± 1/2 1/2 2s1/2 2 1 + 1/2 3/2 2p3/2 2 1 - 1/2 1/2 2p1/2 3 0 ± 1/2 1/2 3s1/2 3 1 + 1/2 3/2 3p3/2 3 1 - 1/2 1/2 3p1/2 3 2 + 1/2 5/2 3d5/2 3 2 - 1/2 3/2 3d3/2
j = |l ± s| j: Total angular momentum l: Orbital angular momentum s: Spin angular momentum
Background
University of Illinois at Chicago RRC - Electron Microscopy Service
XPS probes core-levels → Binding energies in the range of 10 – 103 eV → Kinetic energies of similar magnitudes when Al-Kα or Mg- Kα radiation is used → Electrons with such low KE easily scattered (REMEMBER THIS)
Background
University of Illinois at Chicago RRC - Electron Microscopy Service
1000
10
1
IMFP
, λ (n
m)
100
Kinetic Energy (eV)
‘Universal’ IMFP vs. KE Curve
Background
‘Universal Curve’ shows that photoelectrons with KE in the 10 – 103 eV range have inelastic mean-free-paths (IMFPs) from 1 – 3.5 nm IMFP depends on: (1) Material (atomic #, density) (2) Kinetic energy
University of Illinois at Chicago RRC - Electron Microscopy Service
95% of all photoelectrons detected are generated within 3λ of the surface = Sampling Depth (65% within 1λ). ‘3λ’ is used as the ‘benchmark’ definition for Sampling Depth in XPS. So the sampling depth for XPS is typically 3 - 10 nm → Surface-Sensitive! Instrument must be run under ultra-high vacuum!
Background
University of Illinois at Chicago RRC - Electron Microscopy Service
VISCOUS FLOW - When diameter of tube > 100λ, gas molecules more likely to bump into each other. Molecules in general move towards lower pressure end of tube. Unlikely to get backstreaming.
MOLECULAR FLOW - When diameter of tube is < λ, gas molecules are more likely to collide with the tube wall than each other. There is free movement of molecules in either direction, the numbers directly related to ratio of pressures at each end of tube. At high vacuum this ratio is likely to be close to 1. Backstreaming a concern.
TRANSITIONAL FLOW - Intermediate between Viscous and Molecular.
Pressure (Torr)
Average distance gas molecules between
collisions λ
Time for formation of a monolayer of gas
Number of gas molecules per litre
Atmospheric Pressure 760 0.000066 3.3ns 2.5x1022
1 0.066 3.3 µsec 2.5x1019
10-1 660 µm 33 µsec 2.5x1018 Medium Vacuum 10-3 66 mm 3.3 msec 2.5x1016
High Vacuum 10-6 66 m 3.3 sec 2.5x1013
10-9 66 km 55 mins 2.5x1010 Ultra-high Vacuum 10-10 660 km 550 mins 2.5x109
Vacuum 101
University of Illinois at Chicago RRC - Electron Microscopy Service
So for an 60mm diameter tube
VISCOUS > 0.1 Torr > TRANSITIONAL > 1 mTorr > MOLECULAR
PUMPS • Used from atmosphere down to 0.1Pa
• Problems with corrosive of condensable gases (H2O)
• Potential source of oil contamination of vacuum system if pressure in line to system is not kept in viscous flow regime.
• At or near atmospheric pressure an oil mist is ejected through outlet valve. Must vent outside or through a filter.
• Used from 1Pa to 10-7 Pa
• Historically most widely used high vacuum pump, really a vapor jet pump.
• Pumping speed virtually constant below 10-1 Pa
• Major problem - Backstreaming; minimised by using a low vapour pressure oil.
• BAD NEWS - do not let air into a diffusion pump!
• Used from 10-1 Pa to 10-9 Pa
• Gas molecules that are pumped are trapped inside pump by the gettering action of the sputtered Ti - limited lifetime.
• Absolute freedom from oil contamination with no moving parts.
• Ideal for high vacuum systems but are not well suited on systems that are cycled frequently to atmosphere.
• Used from 10 Pa to 10-8 Pa
• Extremely high speed (10,000rpm) mechanical pumps typically with magnetic levitation bearings for EM use.
• Works efficiently in Molecular Flow region - needs to be backed.
• No backstreaming of oil when operating at full speed.
• Major concern - preventing physical damage to pump!
ROTARY DIFFUSION ION PUMP Turbo Molecular Pump OTHER PUMPS
• Cryosorption - oil free, capture, Atmosphere to 10-1 Pa
• Diaphragm - oil free, transfer, Atmosphere to 1Pa
• Claw pump - dry, transfer, Atmosphere to 10Pa
• Molecular Drag - dry, transfer, 10Pa to 10-6 Pa
• Sublimation - oil free, capture, 10-1 Pa to 10-9 Pa
REMEMBER -- No Pump exerts a force that drags or pulls gas molecules to it. Pumping is purely diffusion of gas molecules from high pressure to low pressure regions
Vacuum 101
University of Illinois at Chicago RRC - Electron Microscopy Service
Analytical Capabilities of XPS
(1) Identify elements/compounds (except H and He)
(2) Determine oxidation states (e.g. Ti3+ or Ti4+)
(3) Identify types of chemical bonds (e.g. Si-O or Si-C)
(4) Semi-quantitative analysis (10-15% error)
(5) Determine adsorbate/film thickness
(6) Highly surface-sensitive (3 – 10 nm from the surface) → Detection limit 0.1 to 1 at% → Ultra-high vacuum required!!! → Minimize/delay surface reactions and contaminations
University of Illinois at Chicago RRC - Electron Microscopy Service
KINETIC ENERGY, eV
Analytical Capabilities of XPS
Survey spectrum for element identification
University of Illinois at Chicago RRC - Electron Microscopy Service
Nickel
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
XPS spectra show characteristic "stepped" background. Due to inelastic processes (extrinsic losses) from deep in bulk. Electrons deeper in surface loose energy and emerge with reduced KE, apparent background increase at higher BE
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
Typical Features and ‘Artifacts’ of Core-Level Peaks Can get multiple peaks from core levels – must be aware of where they come from in order to carry out chemical analysis (not all may be present) 1. Spin orbit splitting leads to additional peaks (no splitting for s, splitting for
p,d,f etc.) 2. Additional peaks due to, for example, chemical shifts and oxidation
states 3. Ghost peaks at lower binding energies (achromatic X-ray only) – no
useful info! 4. Shake up/ off peaks at higher binding energies (result of energy being
transferred from the ejected photoelectron electron to a valence electron).
5. Plasmon loss peaks (due to electron excitations) 6. Photon-induced Auger peaks 7. Effects of charging of non conductive specimens
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
537.0 534.0 531.0 528.0 525.0 Binding Energy (eV)
O 1s
No spin-orbital splitting for s Spin-orbital splitting for p, d, f
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
Ti 2p1/2 and 2p3/2 chemical shift for Ti and Ti4+. Charge withdrawn Ti → Ti4+ so 2p orbital relaxes to higher BE
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
Schematic of Ghost and Shake-up peaks
0
100
200
300
400
500
600
293.8
293.4
293
292.6
292.2
291.8
291.4
291
290.6
290.2
289.8
289.4
289
288.6
288.2
287.8
287.4
287
286.6
286.2
285.8
285.4
285
284.6
284.2
283.8
283.4
283
282.6
282.2
281.8
281.4
281
280.6
280.2
279.8
279.4
279
278.6
278.2
277.8
277.4
277
276.6
276.2
MonochromaticAchromatic
Binding Energy Kinetic Energy
Main peak
Shake-up Peak
Ghost Peak
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
Electrical insulators cannot dissipate charge generated by photoemission Process. Surface picks up excess positive charge - all peaks shift to higher BE Can be reduced by exposing surface to neutralizing flux of low energy electrons - "flood gun" or "neutralizer“. BUT must have good reference peak.
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
Is = Io exp (-d / λcosθ) Is: Intensity at surface Io: Intensity from infinitely-thick sample d: depth λ: Inelastic mean-free-path (IMFP) θ: Spectrometer take-off angle
Beer-Lambert relationship (Numerical expression Describing the photoelectron Intensity generated from a material)
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
⎥⎦
⎤⎢⎣
⎡⎟⎟⎠
⎞⎜⎜⎝
⎛−−= ∞
FilmFilmFilm
dIIλ
exp1⎟⎟⎠
⎞⎜⎜⎝
⎛−= ∞
FilmSubstrateSubstrateSubstrate
dII,
expλ
SiO2 Surface Layer
Si Substrate
Analytical Capabilities of XPS
Using the Beer-Lambert expression to estimate film thickness…
University of Illinois at Chicago RRC - Electron Microscopy Service
ISi: 1411.6
nm 0.58
1lncos 2
22
=
⎪⎭
⎪⎬
⎫
⎜⎜⎝
⎛+⎟⎟⎠
⎞
⎪⎩
⎪⎨
⎧
⎜⎜⎜
⎝
⎛
⎟⎟
⎠
⎞=
∞
∞
oxide
Si
SiO
SiO
SiSiOoxide
d
II
IId θλ
ISiO2: 241.4
When λ, θ and all the respective intensities are known, film thickness can be determined by taking the ratio of Ifilm to Isubstrate and solve for d
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
Semi-Quantitative Analysis Photoelectron intensity from a homogeneous material is also dependent on instrumental factors, and can be alternatively-described by
I = JCσζTλ J: X-ray flux C: Concentration of the element-of-interest σ: Ionization cross-section ζ: Spectrometer angular acceptance T: Spectrometer transmission function λ: IMFP of the element
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
C = I/(JσζTλ) = I/JF F is termed the Relative Sensitivity Factor: - incorporates all the terms associated with the spectrometer and material - empirically-determined by XPS manufacturer
- values are normalized against Fluorine
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
Atomic fraction of an element (A) in a multi-component material (ABCD…) can be estimated using the following,
Atomic % A = (IA/FA) / Σ (In/Fn) Quantification using this expression is valid only if: (1) Material is homogeneous, (2) Material surface is smooth and flat.
Analytical Capabilities of XPS
University of Illinois at Chicago RRC - Electron Microscopy Service
Ion Gun
Charge Neutralizer (built into lenses)
Instrumentation
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Kratos Axis-165 XPS system in RRC-East
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Total resolution (i.e. peak’s full-width-half-maximum) of instrument is convolution of: (1) X-ray energy spread, (2) Spectrometer broadening, and (3) Intrinsic line-width of the element-of-interest.
Total FWHM = {FWHMx-ray2 + FWHMspectrometer
2 + FWHMintrinsic2}1/2
Inte
nsity
Energy
50 %
100 %
Total FWHM
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Binding energy shifts due to different chemical states or bonding configurations can be subtle (1eV or less) for certain elements. Say, if achievable total FWHM of a peak is 2 eV, the XPS instrument will not be able to resolve 2 peaks that are separated only by 0.5eV!!! Total FWHM can only be decreased by minimizing the FWHM of: (1) X-ray; (2) Spectrometer * Intrinsic line-widths is a non-controllable term!
University of Illinois at Chicago RRC - Electron Microscopy Service
Ideal Candidates for X-ray source: Al- and Mg-Kα (1) High energy
(2) Narrow energy spread
Instrumentation
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Twin-Anode Achromatic X-ray source: Bombard metallic anode with 10-25kV electrons with ~10mA of current to generate X-rays. Can generate high X-ray flux producing high signal BUT specimen may be damaged by heat generated by the X-rays and continuum radiation and source emits X-ray satellites (additional weak lines at lower binding energies) Simple, relatively inexpensive High flux (1010-1012 photons·s-1) Beam size ~ 1cm
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Rowland Circle
Quartz Crystal
X-ray Source
Sample
Electron Spectrometer
Monochromatic X-Ray source: Diffraction from bent SiO2 crystal focusing primary λ at specimen. Other λ 's focused at different points in space (filtered). Always use Al Kα which is diffracted from quartz (no equivalent crystal for Mg Kα)
Beam size ~ 1 cm to 50 mm
Eliminates satellites peaks – simpler spectra
Decreases FWHM of X-ray energy
Flux decreases at least an order of magnitude leading to less damage, improves S/N (no X-ray continuum) but lower signal More complicated and expensive
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Most common type of electrostatic deflection-type analyzer: Concentric Hemispherical Analyzer (CHA) or spherical sector analyzer Energy resolution dependant on radius. Capable of collecting photoelectrons of larger angular distribution. Photoelectrons of a specific energy are focused by the lens at the slit of the spectrometer. Lens also controls sampling area. Photoelectrons travel through a circular path and exit into a series of channeltrons (electron multipliers).
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
- Hollow glass tube with semiconducting layer on inner surface
- Electrons/ions entering Channeltron produce secondary electrons (SEs)
- ‘Avalanche’ effect as SEs accelerate down the tube under HV
- Signal amplified by ~ 106 to 108
Single-Channel Electron Multipliers (Channeltron)
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Negative potential on two hemispheres V2 > V1
Potential of mean path, Ro through analyzer is
Vo = (V1R1+V2R2)/2Ro
An electron of kinetic energy eV = Vo will travel a circular orbit through hemispheres at radius Ro Since Ro, R1 and R2 are fixed, in principle changing V1 and V2 will allow electrons of different KE to be detected.
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
But ΔE/Eo = S/2Ro
ΔE = (S/2Ro)Eo
Peak FWHM Spectrometer
Term (constant)
Photoelectron KE
Spectrometer broadening (i.e. FWHMspectrometer) is a function of photoelectron KE entering spectrometer…
i.e. Resolution is non-uniform across XPS energy spectrum!
University of Illinois at Chicago RRC - Electron Microscopy Service
To circumvent the problem… (1) Hemisphere potentials are fixed to allow electrons with a fixed KE (the pass energy, Ep) to reach the electron detectors (Channeltrons); (2) Electrostatic lens before the slit decelerate photoelectrons of a particular KE to Ep; (3) Magnetic lens focus these electrons with Ep at the slit, so that only electrons with that pass energy are allowed to enter the slit. Scanning of the energy-scale is achieved by varying the decelerating potential on the electrostatic lens instead of the hemisphere potentials → Fixed energy-resolution across the energy-scale!!!
Instrumentation
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Most XPS systems also come equipped with an ion gun. Purpose: (1) Removes surface contaminants (usually oxides and hydrocarbons);
(2) Allows depth-profiling study. How does it ‘clean’ a surface? (1) Ionizes an inert gas (usually Ar)
(2) Focuses and accelerates the ions towards the specimen surface
(3) Rasters the ion beam across the surface
(4) Ions impart energy to surface contaminants → Sputtering them away!
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Atomic Fraction
of Cd
Atomic Fraction
of Te
Cd: Te Ratio
As-received 0.65 0.35 1.86
10 mins Sputter
0.51 0.49 1.04
20 mins Sputter
0.55 0.45 1.22
Example: Successful removal of Cr-contamination from CdTe
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Potential Problems Associated with Sputter-Cleaning: • Surface roughening;
• Alter oxidation states of some elements;
• Selective ‘etching’ of surface due to different sputtering rates in a multi-component material.
University of Illinois at Chicago RRC - Electron Microscopy Service
Instrumentation
Example: Oxidation-state changed in Pd after sputtering…
100
150
200
250
300
350
400
450
500
326328330332334336338340342344346348350352
Binding Energy (eV)
Inte
nsity
(CP
S)
Pd 3d (As-received)
Pd 3d (10 min Sputter)
Elemental Pd 3d5/2
PdO 3d5/2
Pd 3d3/2 Orbitals
University of Illinois at Chicago RRC - Electron Microscopy Service
Spectrum Simulation (Peak-Fitting)
Simulation of XPS spectra (i.e. peak fitting) to match experimentally-observed spectra. Purpose: (1) Background noise subtraction to reveal true peak intensities;
(2) Precise determination of peak position;
(3) Deconvolute spectra into individual components when 2 or more peaks are in close proximity.
University of Illinois at Chicago RRC - Electron Microscopy Service
Select Background
Model
Adding Synthetic Peak
Providing Initial-Guess For Peak Parameters: (1) Line position (2) Area (3) FWHM (4) Gaussian-Lorentzian Ratio (5) Asymetry (6) S.O.S.
Software Fitting
End of
Fitting
Modify Peak Parameters
Background Quality?
Good
Poor
Fit Quality?
Good
Poor
Option 1
Option 2
Spectrum Simulation (Peak-Fitting)
University of Illinois at Chicago RRC - Electron Microscopy Service
Types of Background: (1) Linear (2) Shirley (typically-used) (3) Tougaard
Shirley Tougaard
Linear
Spectrum Simulation (Peak-Fitting)
University of Illinois at Chicago RRC - Electron Microscopy Service
Types of Line-Shapes: (1) Gaussian Function
→ Describes the measurement process
(e.g. instrumental response, X-ray line-shape, Doppler and thermal broadening)
(2) Lorentzian Function
→ Describes lifetime broadening (intrinsic line-width)
XPS peaks can usually be described by varying G-L ratio
Spectrum Simulation (Peak-Fitting)
University of Illinois at Chicago RRC - Electron Microscopy Service
Peaks associated with pure metals may tend to be asymmetic… → Need to introduce a ‘Tail Modifier’ term to the G-L functions
Spectrum Simulation (Peak-Fitting)
University of Illinois at Chicago RRC - Electron Microscopy Service
XPSPEAK version 4.1 is a free windows-based XPS peak fitting program available online. Installation file and user manual downloadable at: http://www.phy.cuhk.edu.hk/~surface/XPSPEAK
You will need this software to complete the XPS assignments!
Spectrum Simulation (Peak-Fitting)
University of Illinois at Chicago RRC - Electron Microscopy Service
Summary
X-ray Photon Spectroscopy (XPS) (Originally called ESCA) is a surface-sensitive technique that probes the chemical properties of the top ~10 nm of a solid surface that must be UHV-compatible – no wet specimens! It is the most versatile and quantifiable of all the surface chemical analysis methods: elemental ID, oxidation-states, chemical bonds, film thickness, depth-profiling, semi-quantitative analysis… Limitations: (1) does not detect H or He; (2) Radiation damage possible (worse for achromatic sources); (3) Charge neutralization needed for insulating material; (4) Chemical analysis can be limited to functional groups and in some cases chemical shifts are not resolvable.