an evaluation of the isotope ratio performance of an axial tof - icp - mass spectrometer f....
TRANSCRIPT
An evaluation of the isotope ratio performance of an axial TOF - ICP - mass spectrometer
F. Vanhaecke, L. Moens, R. Dams, L. Allen* and S. Georgitis*University of Ghent, Laboratory of Analytical Chemistry, Proeftuinstraat 86, B-9000 Gent, Belgium.* LECO Corporation, 3000 Lakeview Ave., St. Joseph, MI 49085 USA.
1. Instrumentation
LECO Renaissance TOF-ICPMS instrument
123
Acceleration
Extraction
Skimmer
Sampler
Detector
Ion Mirror
ICP Torch
Vacuum StagesFlight Tube
simultaneous extraction of ions from ICP automatic correction for instabilities
in ion source & during ion sampling process
on-axis acceleration
up to 30000 full spectra / s
modified ETP discrete dynode electron multiplier use of analog detection mode to avoid pulse counting saturation no correction for detector dead time required
Meinhard type C concentric nebulizer& double pass cyclonic spray chamber (PTFE) free aspiration of sample solution
1. Instrumentation (2)
203 204 205 206 207 208 209 210
m/z
0
0.5
1
1.5
2
2.5
3
analog signal intensity (mV)
20 microgram/L Pb
2. Data processingAccurate background correction application of 1 mV offset value
use of signal intensity at m/z = 210 as internal
reference (no ionic species present)
correction for signal drift & instrument instability improvement of both precision & accuracy
Isotope ratio precision
expressed as RSD% for n = 10
2. Data processing (2)
averageblank,
nrepsample,averageblank,nrepsample, (210)signal
(210)signal(M)signalM)(signal
3. Influence of acquisition time per replicate & analyte concentration on precision
figure A
0.5 s 1 s 5 s 10 s 30 s 60 s 120 s 300 s
acquisition time per replicate
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
RS
D%
fo
r 2
06
Pb
/20
7P
b n
= 1
0
500 microgram/L
50 microgram/L
500 microgram/L 0.33 0.13 0.089 0.042 0.023 0.022 0.022 0.033
50 microgram/L 0.7 0.42 0.14 0.18 0.09 0.17 0.071 0.059
figure B
50 ug/L 100 ug/L 250 ug/L 500 ug/L
analyte concentration (microgram/L)
0
0.02
0.04
0.06
0.08
0.1
0.12
0.14
RS
D%
fo
r 2
06
Pb
/20
7P
b n
= 1
0
30s / replicate
10s / replicate
30s / replicate 0.11 0.051 0.038 0.037
10s / replicate 0.13 0.12 0.09 0.047
3. Influence of acquisition time per replicate & analyte concentration on precision(2)
Despite use of analog mode,
trends in agreement with Poisson counting statistics average improvement 50 500 g/L = 3.5 10 = 3.2 (A) average improvement 1 30 s = 5.2 30 = 5.5 (A) average improvement 10 30 s = 1.8 3 = 1.7 (B)
No further improvement on increasing acquisition time > 30 s !!
3. Influence of acquisition time per replicate & analyte concentration on precision (3)
4. Typical isotope ratio precision at 30 s acquisition time / replicate (RSD% for n = 10)
50 g/L 500 g/L
63Cu/65Cu 0.12 0.03164Zn/66Zn 0.15 0.05666Zn/68Zn 0.21 0.06386Sr/87Sr 0.30 0.04486Sr/88Sr 0.26 0.038
107Ag/109Ag 0.040 0.033135Ba/138Ba 0.28 0.069137Ba/138Ba 0.14 0.051204Pb/208Pb 2.5 0.069206Pb/207Pb 0.090 0.023206Pb/208Pb 0.075 0.041
5. Isotope ratio precision for transient signals (RSD% for n = 5) - FI signals, FWHM = 6 s
50 g/L 500 g/L
63Cu/65Cu 0.21 0.1264Zn/66Zn 0.66 0.07286Sr/87Sr 0.47 0.1586Sr/88Sr 2.47 0.23
107Ag/109Ag 0.23 0.043137Ba/138Ba 0.31 0.10206Pb/207Pb 0.36 0.12206Pb/208Pb 0.48 0.044207Pb/208Pb 0.48 0.10
Isotope ratio precision at 500 µg/L still comparable to the best values ever reported for quadrupole-based ICPMS
6. Mass discrimination
1% per mass unit at mid-mass (Cd) comparable to quadrupole-based or sector field
instrumentation
7. AccuracyThe accuracy attainable was evaluated by comparing Pb isotopic results for a Merck Titrisol ('natural') Pb standard solution obtained by (1) TOF-ICPMS and (2) single collector TIMS.
For TOF-ICPMS, 'NIST 981 - Common lead isotopic standard' was used as an external standard.
TOF-ICPMS TIMS
1.3
1.325
1.35
1.375206Pb/207Pb
TOF-ICPMS TIMS0.5
0.525
0.55
0.575208Pb/206Pb
1.3295 ± 0.0056
1.3230 ± 0.0055
1.877 ± 0.0201.8779 ± 0.0032
For TOF-ICPMS, the result is the average of 3 subsequent determinations,. Each determination consisted of 5 replicate measurements (30 s) of the 'sample' solution and 5 replicate measurements of the NIST standard solution. For TIMS, the result is the average of 5 determinations. For both techniques, the uncertainties indicated are 95% confidence intervals.
7. Accuracy (2)
technique isotope ratio precision* additional comments
TIMS 0.01 - 0.05% isolation of analyte element required
quadrupole-based ICPMS 0.1 - 0.5%
sector field ICPMS 0.05% (LR) HR capabilities
multi-collector ICPMS 0.01 - 0.05% LR only – expensive
TOF-ICPMS 0.05% 'simultaneous' measurement of several ratios
* : at sufficiently high signal intensities and for isotope ratios close to unity
At sufficiently high signal intensities and for a moderate acquisition time, axial TOF-ICPMS yields a typical isotope ratio precision 0.05% RSD. This precision can be obtained for many ratios simultaneously. The mass discrimination is comparable to that observed with quadrupole-based and sector field instruments and when appropriately corrected for, accurate results can be obtained.
8. Conclusions